Discrepant results (accuracy) comparison of inratio test with lab results provided by end-user at time complaint was filed: date: 01/05/07, inratio: 6.
1, lab: 3.
8, mean: 4.
95, confidence limits: 2.
8-7.
2.
Date: 01/15/07, inratio: 4.
3, lab: 2.
8, mean: 3.
55, confidence limits: 2.
2-5.
3.
Per internal procedure, the mean of the inratio meter and comparative system inr were calculated.
Both inratio and lab values are within the confidence limits for inr testing.
The results are not considered discrepant within the context of the documented variability for inr testing.
Therefore, further testing is not required at this time.
Per text "pt has antiphospholipid antibody syndrome, (cardiolipin-positive, lupus-negative).
" pt's condition may be causing interference.
|
Caller alleged discrepant results compared with the lab.
Results as follows: date: 01/05/07, inratio: 6.
1, lab: 3.
8.
Date: 01/15/07; inratio: 4.
3; lab: 2.
8.
Pt has antiphospholipid antibody syndrome (cardiolipin-positive, lupus-negative).
|