| Trade Name | MEDTRONIC VASCULAR ENDURANT STENT GRAFT SYSTEM |
| Classification Name | system, endovascular graft, aortic aneurysm treatment |
| Applicant | MEDTRONIC VASCULAR |
| PMA Number | P100021 |
| Date Received | 06/04/2010 |
| Decision Date | 12/16/2010 |
| Product Code | |
| Docket Number | 11M-0040 |
| Notice Date | 01/25/2011 |
| Advisory Committee |
Cardiovascular |
| Clinical Trials |
NCT00870051
|
| Expedited Review Granted? | No |
| Combination Product |
No
|
| Information About: |
Labeling, Approval Order, Summary of Safety and Effectiveness |
Approval Order Statement Approval for the endurant stent graft system. The device is indicated for the endovascular treatment of infrarenal abdominal aortic or aorto-iliac aneurysms in patients with the following characteristics: 1) adequate iliac or femoral access that is compatible with vascular access techniques, devices, or accessories; 2) proximal neck length of >= 10mm; 3) infrarenal neck angulation <= 60 degrees; 4) distal fixation length of >= 15 mm; 5) aortic neck diameters with a range of 19 to 32 mm; 6) iliac diameters with a range of 8 to 25 mm; and 7) morphology suitable for aneurysm repair. |