| Trade Name | STAR S4 ACTIVETRAK EXCIMER LASER SYSTEM AND WAVESCAN WAVEFRONT SYSTEM |
| Classification Name | excimer laser system |
| Generic Name | excimer laser for ophthalmic use |
| Applicant | VISX, INCORPORATED |
| PMA Number | P930016 |
| Supplement Number | S016 |
| Date Received | 10/01/2002 |
| Decision Date | 05/23/2003 |
| Product Code | |
| Docket Number | 03M-0333 |
| Notice Date | 07/30/2003 |
| Advisory Committee |
Ophthalmic |
| Supplement Type | panel track |
| Supplement Reason | change design/components/specifications - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
| Information About: |
Labeling, Approval Order, Summary of Safety and Effectiveness |
Approval Order Statement Approval for the vixx star s4 activetrak excimer laser system and wavescan wavefront system. The device uses a 6. 00 mm optical zone, an 8. 00 mm treatment zone, and is indicated for wavefront-guided laser assisted in situ keratomileusis (lasik) for: 1) the reduction or elimination of myopic astigmatism up to -6. 00 d mrse, with cylinder between 0. 00 and -3. 00 d; 2) patients 21 years of age or older; and 3) patients with documented evidence of a change in manifest refraction of no more than 0. 50 d (in both cylinder and sphere components) for at least one year prior to the date of pre-operative examination. |