| Trade Name | ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT |
| Classification Name | system, endovascular graft, aortic aneurysm treatment |
| Applicant | COOK INCORPORATED |
| PMA Number | P020018 |
| Supplement Number | S040 |
| Date Received | 10/07/2011 |
| Decision Date | 04/04/2012 |
| Product Code | |
| Docket Number | 12M-0371 |
| Notice Date | 04/17/2012 |
| Advisory Committee |
Cardiovascular |
| Clinical Trials |
NCT00875563
|
| Supplement Type | panel track |
| Supplement Reason | labeling change - indications |
| Expedited Review Granted? | Yes |
| Combination Product |
No
|
| Information About: |
Labeling, Approval Order, Summary of Safety and Effectiveness |
Approval Order Statement Approval for the zenith fenestrated aaa endovascular graft (with the adjunctive zenith alignment stent). The zenith fenestrated aaa endovascular graft is indicated for the endovascular treatment of patients with abdominal aortic or aortoiliac aneurysms having morphology suitable for endovascular repair, including: 1) adequate iliac/femoral access compatible with required introduction systems; 2) nonaneurysmal infrarenal aortic segment (neck) proximal to the aneurysms with: a) length >= 4 mm and unsuitable for a non-fenestrated graft. B) diameter <= 31 mm and >= 19 mm. C) angle < 45 degrees relative to long axis of aneurysm. D) angle < 45 degrees relative to axis of suprarenal aorta. 3) ipsilateral iliac artery fixation site > 30 mm in length and between 9 - 21 mm in diameter; and contralateral iliac artery distal fixation site > 30 mm in length and between 7 - 21 mm in diameter. The zenith alignment stent is indicated for use as an adjunct to the zenith fenestrated aaa endovascular graft to secure positive alignment of fenestrations or scallops with the orifice of aortic branch vessels having diameters ranging from 3 to 8 mm. |
| Approval Order |
Approval Order
|
| Post-Approval Study | Show Report Schedule and Study Progress |