| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | PATHVYSION HER-2 DNA PROBE KIT |
| Classification Name | system, test, her-2/neu, nucleic acid or serum |
| Generic Name | neu gene assay kit |
| Applicant | ABBOTT MOLECULAR, INC. |
| PMA Number | P980024 |
| Supplement Number | S006 |
| Date Received | 10/17/2011 |
| Decision Date | 02/27/2012 |
| Product Code | |
| Advisory Committee |
Immunology |
| Supplement Type | 135 review track for 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for manufacturing process changes to the pathvysion her-2 dna probe kit, including changes associated with fish probe restriction enzyme digested gel electrophoresis (dna restriction analysis), implementation of an in-process validated test method for a260 absorbance readings, addition of an incoming screening test for dextran sulfate and revision to manufacturing procedure document for lsi her-2/neu aminated dna. |
|
|