| Trade Name | TAXUS EXPRESS2 PACLITAXEL CORONARY STENT SYSTEM |
| Classification Name | coronary drug-eluting stent |
| Generic Name | intravascular stent with delivery system |
| Applicant | BOSTON SCIENTIFIC CORP. |
| PMA Number | P030025 |
| Supplement Number | S021 |
| Date Received | 08/02/2005 |
| Decision Date | 09/24/2008 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | normal 180 day track |
| Supplement Reason | labeling change - indications |
| Expedited Review Granted? | No |
| Combination Product |
Yes
|
Approval Order Statement Approval for stent diameters which expand the product matrix to include 2. 25 mm and 4. 0mm taxus express2 paclitaxel-eluting coronary stent systems. The device, as modified, will be marketed under the trade name taxus express2 paclitaxel-eluting coronary stent system (monorail, mr and over-the-wire, otw) and is indicated for the following: the taxus express paclitaxel-eluting coronary stent system is indicated for improving luminal diameter for the treatment of de novo lesions in native coronary arteries >2. 25 to <4. 0mm in diameter in lesions < 28mm in length. |
| Post-Approval Study | Show Report Schedule and Study Progress |