| | Class 2 Recall DeVilbiss Model 515A Oxygen Concentrator |  |
| Date Posted |
December 30, 2008 |
| Recall Number |
Z-0617-2009 |
| Product |
DeVilbiss Model 515A Oxygen Concentrator
Intended to provide supplemental low flow oxygen therapy. These devices can be used in homes, nursing homes, and patient care facilities. |
| Code Information |
Model 515ADS - serial number range A561137DS - A562049DS; Model 515ADZ - serial number range A538908DZ - A538924DZ; Model 515AKS - serial number range A542879KS - A543722KS; and Model 515AKZ - serial number A501627KZ.
|
Recalling Firm/ Manufacturer |
Sunrise Medical 100 Devilbiss Dr Somerset, Pennsylvania 15501-2125 |
| For Additional Information Contact |
Customer Service 800-338-1988
|
Reason for Recall |
potential for sieve beds to have been incorrectly assembled
|
| Action |
On 10/30/08, the recalling firm began to telephone customers of the problem. The recalling firm followed up with Urgent: Medical Device Recall letters, dated 11/7/2008. The letters requested that users cease the use of the affected device and immediately contact their provider to make arrangements to have the Concentrator replaced. The firm has set up a toll free line to assist customers with identification of affected units and the return process. |
| Quantity in Commerce |
428 units |
| Distribution |
Worldwide Distribution --- USA including states of AL, CA, FL ,GA, IN, IL, MA, MD, MI, MS, NC, NE, NH, NM, NY, PA, SD, TN, TX, VA, and WI, and countries of Canada, Columbia, Germany, Mexico, Netherlands, Singapore, and United Kingdom. |
| |
|
|
|
|