|
Class 2 Device Recall Terumo Advanced Perfusion System 1 |
|
Date Initiated by Firm |
June 21, 2012 |
Date Posted |
July 17, 2012 |
Recall Status1 |
Terminated 3 on March 26, 2013 |
Recall Number |
Z-2021-2012 |
Recall Event ID |
61753 |
510(K)Number |
K022947
|
Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
Product |
System 1 Base, 220/240V.
The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. |
Code Information |
Catalog Number: 801764 and Serial numbers: 0006-0066, 01 00-0311, 0313-0322,1001-1027,1100-1434,1436, 1440, and 1441. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor MI 48103-9586
|
For Additional Information Contact |
734-741-6173
|
Manufacturer Reason for Recall |
Terumo¿ Advanced Perfusion System 1 has displayed a Red X error message on multiple module icons indicating a loss or fluctuation of internal power on one side of the system's base. Terumo Cardiovascular Systems' initial investigation has determined that there may be more than one cause for the malfunction and is investigating further to determine the root cause.
Potential risks with the loss or
|
FDA Determined Cause 2 |
Process control |
Action |
Terumo sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 21, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to place the addendum in the Operator's Manual for Terumo System 1, at the beginning of the Central Control Module Section. Additionally, a Customer Response Form was attached for customers to complete and return to the firm. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this notice. |
Quantity in Commerce |
648 |
Distribution |
Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV. and the countries of ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DTQ and Original Applicant = TERUMO CARDIOVASCULAR SYSTEMS CORP.
|
|
|
|