Class 2 Device Recall OEC Uroview 2800

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall OEC Uroview 2800 see related information
Date Initiated by Firm November 20, 2006
Date Posting Updated June 22, 2012
Recall Status1 Terminated 3 on June 25, 2012
Recall Number Z-1865-2012
Recall Event ID 62016
510(K)Number K993687  
Product Classification System, x-ray, fluoroscopic, image-intensified - Product Code JAA
Product OEC Uroview 2800, Model number A349855.

Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures.
Code Information 0579-R , P6-0583-R , P6-0584-R , P6-0585-L , P6-0586-L , P6-0587-L , P6-0599-L , P6-0600-R , P6-0612-R , P6-0614-L , P6-0617-L , P6-0618-R , P6-0619-L , P6-0620-L , P6-0621-R , P6-0622-L , P6-0623-R , P6-0624-L , P6-0625-R , P6-0627-L , P6-0629-L , P6-0632-R , P6-0634-R , P6-0635-L , P6-0643-R , P6-0646-L , P6-0647-R , P6-0658-R , P6-0660-R , P6-0663-R , P6-0672-L , P6-0674-L , P6-0676-L , P6-0677-L , P6-0682-L , P6-0683-L , P6-0687-L , P6-0688-R , P6-0689-R , P6-0690-L , P6-0691-L , P6-0692-L , P6-0693-L , P6-0695-R , P6-0698-L , P6-0700-R , P6-0703-L , P6-0705-L , P6-0706-L , P6-0708-R , P6-0710-L , P6-0712-R , P6-0713-R , P6-0714-L , P6-0716-R , P6-0720-R , P6-0718-R , P6-0719-L , P6-0721-L , P6-0724-R , P6-0726-L , P4-0007-R-RC , P6-0009-L, and P4-0481-L.
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City UT 84116-2862
For Additional Information Contact
801-536-4516
Manufacturer Reason
for Recall
Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and "digital spot mode", and the system does not reboot.
FDA Determined
Cause 2
Software design
Action On 11/20/06 two written notifications were sent to consignees addressed to the Facility Administrator and the Radiology Administrator. The notifications advised them that GE Healthcare had identified several intermittent potential safety issues that may occur with the product. Interim solutions were provided for each issue. The letter stated that GEHC OEC would contact customers and remedy the issues free of charge when a permanent solution became available. An Urgent Recall Notice, dated April 13, 2007, was also sent to customers. Customers were to contact the service team at 800-874-7378 for further information, if there were any questions or concerns.
Quantity in Commerce 717 units
Distribution Worldwide Distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JAA and Original Applicant = GE DEC MEDICAL SYSTEMS
-
-