 Class 2 Device Recall Kopf see related information
Date Posted November 06, 2012
Recall Status1 Terminated on April 16, 2014
Recall Number Z-0248-2013
Recall Event ID 63103
510(K)Number K053536 
Product Classification Prosthesis, hip, semi-constrained (metal uncemented acetabular component) - Product Code KWA
Product Kopf Adapter S
Sterile, Rx
Code Information REF Numbers: 01.00185.145, 146, 147, 148
2329720, 2345868, 2351558, 2351558, 2357964, 2357985, 2363710, 2365318, 2390727, 2401986, 2404475, 2420196, 2424824, 2424824, 2428509, 2449961, 2469199, 2476706, 2476706, 2476708, 2476708, 2486173, 2324764, 2335268, 2339607, 2342207, 2342207, 2342209, 2342211, 2354632, 2357988, 2365331, 2368841, 2385720, 2388044, 2396086, 2399440, 2404481, 2408183, 2414476, 2414480, 2420203, 2420203, 2424841, 2428528, 2430914, 2433313, 2455294, 2462031, 2462048, 2469203, 2469203, 2476713, 2476718, 2486178, 2486178, 2486180, 2499005, 2499005, 2509186, 2516387, 2516387, 2516389, 2314373, 2322686, 2329747, 2339609, 2339611, 2342218, 2350072, 2351571, 2351573, 2354637, 2356264, 2360204, 2430920, 2443367, 2455300, 2476723, 2482727, 2322688, 2322689, 2322689, 2335264, 2335264, 2336972, 2336972, 2399446, 2428549, 2428549, 2428549
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
574-372-4807
For Additional Information Contact
574-372-4807
Manufacturer Reason
for Recall
Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
FDA Determined
Cause 2
Error in labeling
Action Zimmer, Inc. sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated September 4, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Certificate of Acknowledgement Form was attached for customers to complete and return via fax to Zimmer, Inc. at 574 372-4265.
Quantity in Commerce 2011
Distribution Nationwide Distribution.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = KWA and Original Applicant = ZIMMER GMBH
-
-