Date Initiated by Firm |
February 04, 2013 |
Create Date |
June 27, 2016 |
Recall Status1 |
Terminated 3 on September 12, 2016 |
Recall Number |
Z-2055-2016 |
Recall Event ID |
63990 |
510(K)Number |
K041550 K110901
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral administration; and various solution through enteral, intravenous, intra-arterial, subcutaneous, and pathways. |
Code Information |
14339017 14339062 14339063 14339066 14339071 14339072 14339076 14339078 14339079 14339084 14339086 14339087 14339089 14339090 14339091 14339095 14339097 14339098 14339099 14339102 14339104 14339105 14339107 14339108 14339111 14339113 14339114 14339115 14339116 14339119 14339121 14339125 14339128 14339130 14339133 14339136 14339138 14339139 14339140 14339141 14339142 14339145 14339146 14339149 14339150 14339152 14339154 14339155 14339158 14339168 14339171 14339172 14339179 14339180 14339185 14339190 14339192 14339194 14339195 14339200 14339201 14339205 14339216 14339228 14339230 14339231 14339233 14339236 14339243 14339246 14339248 14339250 14339251 14339254 14339262 14339271 14339274 14339280 14339283 14339287 14339291 14339293 14339299 14339302 14339304 14339308 14339322 14339327 14339328 14339344 14339346 14339348 14339349 14339351 14339358 14339359 14339361 14339362 14339367 14339369 14339370 14339371 14339372 14339373 14339377 14339378 14339380 14339384 14339388 14339389 14339391 14339392 14339394 14339395 14339396 14339399 14339401 14339407 14339408 14339409 14339410 14339411 14339413 14339414 14339415 14339422 14339423 14339424 14339426 14339504 14339517 14339519 14339522 14339524 14339529 14339534 14339538 14339540 14339545 14339546 14339551 14339552 14339553 14339557 14339569 14339572 14339584 14339652 14339659 14339662 14339664 14339669 14339680 14339681 14339692 14339696 14339705 14339707 14339710 14339712 14339718 14339719 14339723 14339725 14339727 14339733 14339738 14339745 14339748 14339761 14339762 14339763 14339769 14339771 14339774 14339776 14339779 14339787 14339789 14339815 14339823 14339841 14339849 14339866 14339867 14339874 14339876 14339884 14339915 14339921 14339932 14339933 14339935 14339937 14339938 14339945 14339950 14339951 14339952 14339955 14339957 14339960 14339963 14339967 14339969 14339979 14339981 14339983 14339985 14339986 14339987 14339989 14339990 14339991 14339992 14339993 14339996 14339997 14339998 14339999 14340000 14340002 14340003 14340004 14340006 14340008 14340009 14340011 14340017 14340018 14340023 14340027 14340029 14340034 14340036 14340037 14340038 14340039 14340041 14340042 14340044 14340048 14340052 14340053 14340058 14340059 14340060 14340061 14340062 14340063 14340065 14340066 14340067 14340069 14340070 14340071 14340072 14340074 14340075 14340076 14340077 14340078 14340079 14340080 14340082 14340083 14340084 14340088 14340089 14340090 14340094 14340095 14340096 14340098 14340101 14340104 14340105 14340106 14340108 14340110 14340111 14340112 14340113 14340114 14340115 14340116 14340117 14340118 14340119 14340120 14340121 14340123 14340124 14340125 14340126 14340127 14340128 14340129 14340130 14340131 14340132 14340133 14340134 14340135 14340136 14340137 14340138 14340140 14340141 14340142 14340143 14340144 14340146 14340147 14340148 14340149 14340150 14340151 14340153 14340154 14340156 14340157 14340159 14340160 14340162 14340163 14340164 14340165 14340166 14340167 14340169 14340170 14340171 14340172 14340173 14340177 14340179 14340180 14340181 14340182 14340185 14340186 14340187 14340188 14340189 14340190 14340191 14340192 14340193 14340194 14340196 14340197 14340198 14340200 14340203 14340204 14340205 14340206 14340207 14340208 14340209 14340210 14340211 14340213 14340214 14340216 14340217 14340218 14340219 14340220 14340221 14340223 14340224 14340225 14340226 14340227 14340228 14340230 14340231 14340232 14340233 14340234 14340235 14340236 14340238 14340239 14340240 14340244 14340245 14340246 14340247 14340248 14340249 14340250 14340251 14340253 14340255 14340256 14340257 14340259 14340261 14340262 14340263 14340264 14340265 14340268 14340271 14340273 14340274 14340275 14340276 14340277 14340278 14340279 14340280 14340281 14340282 14340283 14340284 14340285 14340286 14340287 14340288 14340289 14340290 14340291 14340292 14340293 14340294 14340296 14340300 14340301 14340302 14340304 14340305 14340306 14340307 14340317 14340318 14340319 14340320 14340321 14340323 14340325 14340376 14340377 14340382 14340383 14340384 14340386 14340387 14340388 14340389 14340390 14340393 14340394 14340396 14340399 14340407 14340411 14340412 14340413 14340414 14340416 14340417 14340420 14340422 14340423 14340425 14340426 14340430 14340433 14340434 14340437 14340440 14340442 14340443 14340448 14340449 14340450 14340451 14340453 14340456 14340458 14340461 14340463 14340465 14340467 14340469 14340473 14340474 14340475 14340479 14340480 14340481 14340486 14340489 14340493 14340495 14340496 14340497 14340498 14340499 14340501 14340502 14340503 14340505 14340506 14340507 14340508 14340510 14340511 14340513 14340514 14340515 14340517 14340518 14340523 14340526 14340528 14340529 14340533 14340534 14340536 14340540 14340541 14340542 14340544 14340545 14340546 14340549 14340550 14340551 14340552 14340553 14340555 14340557 14340558 14340559 14340561 14340562 14340563 14340569 14340570 14340571 14340572 14340574 14340575 14340577 14340579 14340581 14340583 14340584 14340585 14340587 14340588 14340589 14340590 14340591 14340592 14340593 14340596 14340599 14340600 14340601 14340605 14340609 14340610 14340612 14340613 14340615 14340617 14340618 14340619 14340623 14340627 14340628 14340629 14340631 14340634 14340635 14340639 14340642 14340644 14340650 14340652 14340655 14340658 14340659 14340662 14340665 14340669 14340670 14340672 14340673 14340675 14340676 14340677 14340679 14340683 14340690 14340691 14340692 14340697 14340701 14340702 14340709 14340710 14340715 14340718 14340719 14340720 14340722 14340723 14340725 14340728 14340733 14340735 14340737 14340738 14340739 14340741 14340742 14340745 14340746 14340747 14340749 14340751 14340753 14340754 14340755 14340756 14340757 14340759 14340760 14340764 14340766 14340767 14340768 14340769 14340770 14340771 14340775 14340777 14340778 14340782 14340783 14340784 14340786 14340790 14340792 14340794 14340795 14340799 14340800 14340801 14340802 14340803 14340806 14340810 14340811 14340813 14340814 14340816 14340817 14340823 14340825 14340827 14340829 14340832 14340833 14340836 14340839 14340843 14340845 14340848 14340849 14340852 14340853 14340857 14340859 14340862 14340864 14340865 14340866 14340868 14340874 14340876 14340877 14340879 14340880 14340881 14340882 14340884 14340885 14340886 14340888 14340889 14340891 14340892 14340896 14340897 14340899 14340900 14340901 14340903 14340904 14340905 14340906 14340911 14340913 14340915 14340916 14340922 14340924 14340926 14340927 14340930 14340931 14340933 14340935 14340937 14340939 14340943 14340945 14340946 14340947 14340948 14340951 14340953 14340954 14340955 14340956 14340957 14340959 14340960 14340961 14340963 14340967 14340969 14340970 14340971 14340972 14340974 14340976 14340987 14340988 14340990 14340993 14340995 14340996 14340997 14340998 14340999 14341000 14341001 14341002 14341003 14341011 14341014 14341015 14341016 14341017 14341020 14341022 14341026 14341030 14341037 14341040 14341042 14341043 14341046 14341048 14341050 14341051 14341052 14341057 14341062 14341066 14341067 14341068 14341069 14341070 14341076 14341079 14341081 14341088 14341089 14341090 14341092 14341094 14341096 14341099 14341101 14341103 14341108 14341109 14341117 14341118 14341120 14341121 14341123 14341124 14341126 14341127 14341129 14341131 14341141 14341144 14341151 14341152 14341153 14341156 14341157 14341158 14341160 14341161 14341162 14341165 14341169 14341172 14341174 14341176 14341177 14341178 14341180 14341181 14341182 14341183 14341184 14341188 14341189 14341190 14341191 14341192 14341193 14341194 14341195 14341197 14341199 14341200 14341202 14341203 14341204 14341205 14341206 14341207 14341208 14341209 14341210 14341211 14341212 14341213 14341215 14341216 14341219 14341220 14341221 14341222 14341223 14341224 14341225 14341226 14341227 14341228 14341229 14341230 14341231 14341233 14341234 14341235 14341237 14341238 14341243 14341245 14341246 14341247 14341248 14341249 14341250 14341251 14341252 14341253 14341254 14341255 14341257 14341259 14341260 14341261 14341262 14341263 14341267 14341268 14341270 14341273 14341276 14341277 14341278 14341281 14341283 14341287 14341288 14341290 14341291 14341292 14341293 14341294 14341297 14341299 14341301 14341304 14341305 14341308 14341311 14341314 14341315 14341316 14341319 14341321 14341323 14341324 14341325 14341326 14341327 14341328 14341329 14341330 14341331 14341332 14341333 14341335 14341336 14341338 14341339 14341340 14341341 14341342 14341343 14341344 14341345 14341346 14341347 14341348 14341349 14341350 14341352 14341354 14341355 14341359 14341360 14341361 14341362 14341363 14341364 14341365 14341366 14341367 14341368 14341369 14341370 14341372 14341375 14341376 14341378 14341379 14341380 14341382 14341383 14341384 14341385 14341387 14341388 14341389 14341390 14341391 14341392 14341393 14341394 14341395 14341397 14341398 14341399 14341401 14341402 14341403 14341404 14341405 14341406 14341407 14341409 14341411 14341412 14341414 14341416 14341417 14341418 14341419 14341420 14341422 14341423 14341424 14341426 14341427 14341428 14341429 14341430 14341432 14341433 14341434 14341435 14341438 14341439 14341441 14341442 14341443 14341444 14341445 14341446 14341447 14341448 14341449 14341451 14341452 14341453 14341455 14341456 14341457 14341459 14341460 14341461 14341462 14341463 14341465 14341466 14341467 14341468 14341469 14341470 14341471 14341472 14341473 14341476 14341477 14341478 14341479 14341480 14341481 14341482 14341483 14341485 14341489 14341490 14341491 14341493 14341494 14341495 14341500 14341501 14341502 14341503 14341504 14341505 14341506 14341508 14341509 14341510 14341511 14341512 14341513 14341514 14341515 14341517 14341519 14341520 14341522 14341523 14341524 14341526 14341527 14341528 14341529 14341530 14341531 14341532 14341533 14341535 14341537 14341543 14341546 14341549 14341550 14341553 14341558 14341559 14341563 14341566 14341569 14341572 14341573 14341574 14341576 14341578 14341579 14341581 14341586 14341587 14341590 14341591 14341592 14341595 14341600 14341606 14341613 14341615 14341618 14341623 14341624 14341627 14341632 14341633 14341637 14341638 14341640 14341641 14341642 14341643 14341644 14341645 14341646 14341647 14341648 14341649 14341650 14341651 14341652 14341653 14341655 14341656 14341657 14341658 14341659 14341660 14341661 14341662 14341663 14341664 14341665 14341667 14341668 14341669 14341670 14341671 14341672 14341673 14341674 14341675 14341676 14341677 14341679 14341680 14341681 14341682 14341683 14341684 14341685 14341686 14341687 14341688 14341691 14341692 14341693 14341697 14341699 14341701 14341702 14341703 14341705 14341707 14341708 14341709 14341711 14341712 14341713 14341714 14341717 14341718 14341719 14341720 14341721 14341722 14341723 14341724 14341725 14341726 14341727 14341729 14341730 14341731 14341732 14341735 14341737 14341738 14341739 14341740 14341741 14341742 14341743 14341744 14341745 14341746 14341747 14341748 14341749 14341751 14341752 14341755 14341756 14341757 14341758 14341760 14341761 14341762 14341763 14341764 14341769 14341770 14341771 14341772 14341773 14341775 14341776 14341777 14341779 14341780 14341781 14341782 14341783 14341784 14341785 14341786 14341788 14341789 14341790 14341791 14341792 14341793 14341795 14341797 14341799 14341800 14341802 14341803 14341805 14341806 14341807 14341808 14341809 14341810 14341811 14341812 14341814 14341815 14341816 14341819 14341820 14341821 14341822 14341823 14341824 14341825 14341826 14341827 14341829 14341830 14341831 14341832 14341834 14341837 14341839 14341840 14341841 14341842 14341843 14341844 14341846 14341848 14341849 14341851 14341852 14341853 14341854 14341858 14341859 14341861 14341863 14341864 14341865 14341866 14341867 14341868 14341869 14341870 14341871 14341872 14341873 14341876 14341878 14341879 14341880 14341881 14341883 14341884 14341885 14341886 14341887 14341889 14341890 14341893 14341894 14341895 14341896 14341897 14341898 14341899 14341901 14341902 14341903 14341904 14341905 14341906 14341908 14341910 14341912 14341913 14341914 14341915 14341916 14341917 14341918 14341920 14341921 14341922 14341924 14341925 14341927 14341928 14341930 14341931 14341932 14341933 14341934 14341936 14341937 14341938 14341939 14341940 14341941 14341943 14341944 14341946 14341947 14341948 14341950 14341951 14341954 14341955 14341956 14341957 14341959 14341960 14341961 14341962 14341963 14341965 14341967 14341968 14341969 14341970 14341971 14341972 14341974 14341976 14341978 14341979 14341980 14341981 14341982 14341983 14341984 14341985 14341986 14341987 14341988 14341989 14341991 14341992 14341993 14341994 14341995 14341997 14341998 14341999 14342000 14342001 14342002 14342003 14342004 14342006 14342008 14342009 14342010 14342012 14342013 14342015 14342017 14342018 14342019 14342020 14342021 14342022 14342023 14342025 14342026 14342027 14342028 14342029 14342030 14342032 14342033 14342034 14342035 14342037 14342038 14342039 14342040 14342042 14342043 14342045 14342046 14342047 14342048 14342050 14342051 14342052 14342053 14342055 14342056 14342057 14342058 14342059 14342060 14342061 14342062 14342063 14342064 14342065 14342066 14342067 14342068 14342069 14342071 14342072 14342074 14342075 14342076 14342077 14342078 14342079 14342080 14342081 14342082 14342085 14342086 14342088 14342089 14342090 14342092 14342093 14342094 14342095 14342096 14342097 14342098 14342099 14342100 14342101 14342102 14342104 14342105 14342107 14342108 14342109 14342110 14342111 14342112 14342113 14342114 14342115 14342116 14342119 14342120 14342121 14342122 14342123 14342124 14342125 14342126 14342127 14342128 14342129 14342130 14342131 14342133 14342134 14342135 14342136 14342137 14342139 14342140 14342141 14342142 14342143 14342145 14342146 14342147 14342148 14342149 14342150 14342151 14342152 14342153 14342154 14342155 14342156 14342158 14342159 14342160 14342161 14342162 14342164 14342165 14342166 14342167 14342168 14342169 14342172 14342173 14342174 14342176 14342177 14342178 14342179 14342180 14342181 14342182 14342183 14342184 14342186 14342187 14342188 14342190 14342191 14342192 14342194 14342196 14342197 14342199 14342200 14342201 14342204 14342205 14342206 14342207 14342208 14342209 14342210 14342211 14342213 14342214 14342215 14342216 14342217 14342218 14342219 14342220 14342221 14342222 14342223 14342224 14342225 14342227 14342228 14342229 14342230 14342231 14342233 14342234 14342236 14342237 14342240 14342241 14342242 14342243 14342244 14342245 14342247 14342249 14342250 14342251 14342252 14342253 14342254 14342255 14342256 14342257 14342258 14342259 14342260 14342261 14342262 14342263 14342264 14342267 14342269 14342272 14342273 14342274 14342275 14342276 14342277 14342279 14342280 14342281 14342282 14342283 14342284 14342285 14342286 14342287 14342289 14342292 14342293 14342295 14342296 14342297 14342298 14342299 14342300 14342301 14342302 14342303 14342304 14342305 14342306 14342307 14342308 14342309 14342310 14342311 14342312 14342314 14342315 14342316 14342317 14342318 14342319 14342320 14342322 14342324 14342325 14342327 14342329 14342330 14342332 14342334 14342335 14342336 14342339 14342340 14342341 14342342 14342343 14342345 14342346 14342348 14342349 14342352 14342353 14342355 14342356 14342357 14342358 14342360 14342361 14342362 14342364 14342365 14342366 14342367 14342368 14342369 14342370 14342371 14342372 14342373 14342374 14342375 14342376 14342378 14342379 14342380 14342381 14342382 14342383 14342384 14342385 14342386 14342387 14342388 14342389 14342390 14342393 14342394 14342395 14342396 14342397 14342398 14342399 14342400 14342401 14342402 14342403 14342404 14342405 14342406 14342407 14342408 14342409 14342410 14342411 14342412 14342413 14342414 14342415 14342417 14342419 14342420 14342421 14342422 14342423 14342424 14342425 14342426 14342428 14342429 14342430 14342431 14342432 14342433 14342434 14342435 14342436 14342437 14342438 14342439 14342440 14342441 14342442 14342443 14342445 14342446 14342448 14342449 14342450 14342452 14342453 14342454 14342455 14342456 14342457 14342458 14342459 14342460 14342461 14342463 14342464 14342465 14342466 14342468 14342469 14342470 14342471 14342472 14342473 14342474 14342475 14342476 14342478 14342479 14342480 14342481 14342483 14342484 14342485 14342486 14342487 14342488 14342489 14342491 14342492 14342493 14342494 14342495 14342496 14342497 14342498 14342499 14342502 14342503 14342505 14342507 14342508 14342509 14342510 14342511 14342512 14342513 14342514 14342515 14342516 14342517 14342518 14342519 14342520 14342521 14342523 14342524 14342525 14342526 14342527 14342528 14342530 14342531 14342532 14342533 14342534 14342535 14342536 14342537 14342538 14342539 14342540 14342541 14342542 14342543 14342544 14342545 14342546 14342547 14342548 14342549 14342550 14342551 14342552 14342553 14342554 14342555 14342556 14342557 14342558 14342559 14342560 14342561 14342562 14342563 14342564 14342567 14342568 14342569 14342570 14342571 14342574 14342575 14342576 14342577 14342578 14342579 14342581 14342582 14342583 14342584 14342586 14342587 14342589 14342590 14342592 14342593 14342594 14342595 14342596 14342598 14342599 14342600 14342601 14342602 14342603 14342604 14342605 14342606 14342607 14342608 14342609 14342610 14342611 14342613 14342614 14342615 14342616 14342617 14342619 14342620 14342622 14342623 14342624 14342625 14342626 14342627 14342628 14342630 14342631 14342632 14342633 14342634 14342635 14342636 14342637 14342638 14342639 14342640 14342642 14342643 14342644 14342645 14342646 14342648 14342650 14342651 14342652 14342653 14342654 14342655 14342656 14342657 14342659 14342660 14342661 14342662 14342664 14342665 14342666 14342667 14342668 14342670 14342671 14342672 14342673 14342674 14342675 14342676 14342677 14342678 14342679 14342681 14342682 14342683 14342684 14342685 14342686 14342687 14342688 14342689 14342690 14342691 14342692 14342693 14342694 14342695 14342696 14342697 14342699 14342700 14342701 14342702 14342703 14342704 14342705 14342706 14342707 14342708 14342709 14342710 14342711 14342712 14342713 14342714 14342715 14342716 14342717 14342718 14342720 14342722 14342723 14342724 14342725 14342726 14342727 14342728 14342729 14342730 14342731 14342732 14342733 14342734 14342736 14342738 14342739 14342740 14342742 14342743 14342744 14342745 14342746 14342747 14342748 14342751 14342752 14342753 14342754 14342756 14342757 14342758 14342759 14342760 14342762 14342763 14342764 14342765 14342766 14342767 14342768 14342769 14342771 14342774 14342775 14342776 14342777 14342778 14342779 14342782 14342783 14342784 14342785 14342787 14342788 14342789 14342790 14342791 14342793 14342794 14342796 14342797 14342798 14342799 14342800 14342801 14342802 14342803 14342804 14342806 14342807 14342808 14342810 14342811 14342812 14342814 14342815 14342816 14342817 14342818 14342821 14342822 14342825 14342828 14342830 14342832 14342833 14342834 14342835 14342836 14342838 14342839 14342840 14342841 14342842 14342843 14342844 14342845 14342846 14342847 14342848 14342849 14342850 14342851 14342852 14342853 14342854 14342856 14342857 14342858 14342859 14342860 14342862 14342863 14342864 14342865 14342867 14342868 14342869 14342871 14342872 14342873 14342874 14342875 14342876 14342877 14342878 14342879 14342880 14342881 14342883 14342884 14342885 14342886 14342887 14342888 14342889 14342890 14342892 14342893 14342895 14342897 14342898 14342899 14342900 14342901 14342902 14342903 14342904 14342905 14342906 14342907 14342909 14342910 14342911 14342912 14342913 14342914 14342915 14342916 14342917 14342918 14342919 14342920 14342922 14342923 14342924 14342925 14342926 14342927 14342928 14342929 14342930 14342931 14342932 14342933 14342935 14342936 14342937 14342938 14342939 14342941 14342942 14342943 14342944 14342946 14342947 14342948 14342951 14342952 14342953 14342954 14342955 14342956 14342957 14342958 14342959 14342960 14342961 14342962 14342963 14342964 14342965 14342966 14342967 14342968 14342970 14342971 14342972 14342973 14342974 14342975 14342976 14342977 14342978 14342980 14342982 14342983 14342985 14342986 14342987 14342989 14342990 14342992 14342993 14342994 14342995 14342996 14342997 14342998 14343000 14343001 14343002 14343003 14343004 14343006 14343007 14343008 14343009 14343010 14343011 14343012 14343013 14343014 14343015 14343016 14343017 14343018 14343019 14343020 14343021 14343023 14343024 14343025 14343026 14343027 14343029 14343030 14343031 14343032 14343034 14343035 14343036 14343037 14343039 14343040 14343041 14343043 14343045 14343046 14343048 14343050 14343051 14343052 14343053 14343054 14343055 14343056 14343057 14343058 14343059 14343060 14343061 14343062 14343063 14343064 14343065 14343067 14343068 14343069 14343070 14343071 14343072 14343073 14343074 14343075 14343076 14343077 14343078 14343080 14343081 14343082 14343083 14343084 14343085 14343086 14343087 14343088 14343089 14343090 14343091 14343092 14343093 14343094 14343095 14343096 14343097 14343098 14343099 14343100 14343101 14343102 14343104 14343105 14343106 14343107 14343108 14343110 14343111 14343112 14343113 14343114 14343115 14343116 14343118 14343120 14343121 14343122 14343123 14343125 14343126 14343127 14343128 14343129 14343130 14343131 14343133 14343134 14343135 14343137 14343138 14343139 14343140 14343141 14343142 14343143 14343144 14343145 14343146 14343147 14343148 14343149 14343150 14343151 14343152 14343153 14343155 14343156 14343157 14343158 14343159 14343160 14343161 14343163 14343164 14343166 14343167 14343168 14343169 14343170 14343171 14343172 14343173 14343174 14343175 14343176 14343177 14343178 14343179 14343180 14343181 14343182 14343183 14343184 14343185 14343186 14343187 14343188 14343189 14343190 14343191 14343192 14343193 14343194 14343195 14343196 14343197 14343198 14343199 14343200 14343201 14343202 14343203 14343204 14343205 14343206 14343207 14343208 14343209 14343210 14343211 14343212 14343214 14343215 14343217 14343218 14343219 14343220 14343221 14343222 14343223 14343225 14343226 14343228 14343229 14343230 14343231 14343232 14343233 14343235 14343236 14343237 14343238 14343239 14343240 14343241 14343242 14343243 14343244 14343246 14343247 14343248 14343250 14343254 14343255 14343257 14343258 14343260 14343261 14343262 14343263 14343264 14343265 14343266 14343267 14343268 14343269 14343270 14343271 14343272 14343273 14343274 14343275 14343276 14343277 14343278 14343279 14343280 14343281 14343282 14343285 14343286 14343287 14343288 14343291 14343292 14343293 14343294 14343295 14343296 14343297 14343298 14343300 14343301 14343302 14343303 14343304 14343305 14343306 14343307 14343308 14343309 14343310 14343311 14343312 14343313 14343314 14343315 14343316 14343317 14343318 14343319 14343321 14343322 14343323 14343325 14343326 14343327 14343328 14343329 14343331 14343332 14343333 14343334 14343335 14343336 14343337 14343338 14343340 14343341 14343343 14343344 14343345 14343346 14343347 14343348 14343349 14343350 14343351 14343352 14343353 14343354 14343355 14343357 14343358 14343359 14343360 14343361 14343362 14343365 14343366 14343368 14343369 14343370 14343371 14343372 14343373 14343374 14343375 14343376 14343377 14343378 14343379 14343380 14343382 14343383 14343384 14343385 14343386 14343387 14343388 14343389 14343390 14343391 14343393 14343394 14343395 14343396 14343398 14343399 14343400 14343402 14343403 14343404 14343405 14343406 14343407 14343408 14343409 14343410 14343411 14343413 14343414 14343415 14343416 14343417 14343418 14343419 14343420 14343421 14343422 14343423 14343424 14343425 14343426 14343427 14343428 14343429 14343430 14343431 14343432 14343433 14343435 14343436 14343437 14343438 14343439 14343441 14343442 14343444 14343445 14343446 14343447 14343448 14343449 14343450 14343451 14343452 14343453 14343454 14343455 14343456 14343457 14343458 14343459 14343460 14343461 14343462 14343463 14343464 14343465 14343467 14343468 14343469 14343470 14343471 14343472 14343473 14343474 14343476 14343478 14343479 14343480 14343481 14343482 14343483 14343485 14343486 14343487 14343488 14343489 14343490 14343491 14343492 14343493 14343497 14343499 14343500 14343501 14343502 14343503 14343504 14343505 14343506 14343507 14343508 14343509 14343510 14343511 14343512 14343513 14343514 14343515 14343516 14343517 14343518 14343519 14343520 14343521 14343522 14343523 14343524 14343525 14343526 14343527 14343528 14343529 14343530 14343531 14343532 14343534 14343535 14343536 14343537 14343538 14343539 14343540 14343541 14343542 14343543 14343544 14343545 14343546 14343547 14343548 14343549 14343550 14343551 14343552 14343553 14343554 14343555 14343556 14343557 14343558 14343559 14343560 14343561 14343562 14343563 14343564 14343565 14343566 14343567 14343568 14343569 14343571 14343572 14343573 14343575 14343576 14343577 14343578 14343580 14343581 14343582 14343583 14343584 14343585 14343586 14343587 14343588 14343589 14343590 14343591 14343593 14343594 14343595 14343596 14343598 14343599 14343601 14343602 14343603 14343604 14343605 14343606 14343607 14343608 14343609 14343610 14343611 14343612 14343613 14343614 14343615 14343616 14343617 14343619 14343620 14343621 14343622 14343623 14343624 14343626 14343628 14343630 14343631 14343632 14343633 14343634 14343635 14343636 14343637 14343638 14343639 14343640 14343642 14343643 14343644 14343645 14343646 14343647 14343648 14343649 14343650 14343651 14343652 14343653 14343654 14343655 14343656 14343657 14343658 14343659 14343660 14343661 14343662 14343663 14343664 14343665 14343666 14343667 14343668 14343669 14343670 14343671 14343672 14343673 14343674 14343675 14343676 14343677 14343679 14343681 14343682 14343683 14343684 14343685 14343686 14343687 14343688 14343689 14343690 14343691 14343692 14343693 14343694 14343695 14343696 14343697 14343698 14343700 14343701 14343703 14343704 14343705 14343706 14343707 14343708 14343709 14343710 14343711 14343712 14343713 14343714 14343715 14343716 14343717 14343718 14343719 14343720 14343721 14343722 14343723 14343724 14343725 14343726 14343727 14343729 14343730 14343731 14343732 14343735 14343737 14343738 14343739 14343740 14343741 14343742 14343743 14343744 14343745 14343746 14343747 14343748 14343749 14343750 14343751 14343752 14343753 14343754 14343755 14343756 14343758 14343759 14343760 14343761 14343763 14343764 14343765 14343766 14343767 14343768 14343769 14343770 14343771 14343772 14343773 14343774 14343775 14343776 14343777 14343778 14343779 14343780 14343781 14343783 14343784 14343785 14343786 14343787 14343788 14343789 14343790 14343791 14343792 14343793 14343794 14343795 14343796 14343797 14343798 14343799 14343801 14343802 14343803 14343804 14343806 14343807 14343809 14343810 14343811 14343812 14343813 14343814 14343815 14343816 14343817 14343818 14343819 14343820 14343822 14343823 14343824 14343825 14343826 14343827 14343828 14343829 14343830 14343831 14343832 14343833 14343835 14343837 14343838 14343839 14343840 14343842 14343843 14343844 14343845 14343847 14343849 14343850 14343851 14343852 14343853 14343854 14343855 14343856 14343857 14343858 14343859 14343860 14343861 14343862 14343863 14343864 14343865 14343866 14343867 14343868 14343869 14343870 14343871 14343874 14343875 14343876 14343877 14343878 14343880 14343881 14343882 14343884 14343885 14343886 14343887 14343888 14343889 14343890 14343891 14343893 14343894 14343895 14343896 14343897 14343898 14343899 14343900 14343901 14343902 14343904 14343906 14343907 14343908 14343910 14343911 14343913 14343915 14343916 14343917 14343918 14343919 14343920 14343921 14343922 14343923 14343924 14343925 14343926 14343927 14343928 14343929 14343930 14343931 14343933 14343934 14343936 14343937 14343938 14343939 14343940 14343942 14343943 14343945 14343946 14343947 14343948 14343950 14343951 14343952 14343953 14343954 14343955 14343957 14343958 14343959 14343960 14343961 14343962 14343965 14343966 14343967 14343968 14343970 14343971 14343973 14343974 14343975 14343976 14343977 14343978 14343979 14343981 14343982 14343983 14343984 14343985 14343986 14343987 14343988 14343990 14343992 14343993 14343994 14343996 14343997 14343998 14343999 14344001 14344002 14344003 14344005 14344006 14344007 14344008 14344009 14344011 14344012 14344013 14344014 14344015 14344016 14344017 14344018 14344019 14344020 14344022 14344023 14344024 14344025 14344026 14344027 14344029 14344030 14344031 14344032 14344033 14344035 14344036 14344038 14344039 14344040 14344041 14344042 14344043 14344044 14344045 14344046 14344047 14344048 14344049 14344050 14344051 14344052 14344053 14344054 14344055 14344056 14344057 14344058 14344060 14344061 14344063 14344064 14344065 14344066 14344067 14344068 14344070 14344071 14344072 14344073 14344074 14344075 14344077 14344078 14344079 14344081 14344082 14344083 14344084 14344085 14344086 14344087 14344088 14344090 14344091 14344092 14344093 14344095 14344096 14344097 14344098 14344099 14344100 14344101 14344102 14344103 14344104 14344105 14344106 14344107 14344108 14344110 14344111 14344112 14344114 14344115 14344116 14344117 14344118 14344119 14344120 14344121 14344122 14344123 14344124 14344125 14344126 14344127 14344129 14344130 14344131 14344132 14344133 14344134 14344135 14344136 14344137 14344138 14344139 14344140 14344142 14344144 14344145 14344146 14344147 14344148 14344149 14344150 14344152 14344153 14344154 14344155 14344156 14344157 14344158 14344159 14344160 14344161 14344162 14344163 14344164 14344165 14344167 14344168 14344169 14344170 14344171 14344172 14344173 14344174 14344175 14344176 14344178 14344181 14344182 14344183 14344184 14344185 14344186 14344187 14344189 14344190 14344191 14344193 14344194 14344195 14344196 14344197 14344198 14344199 14344200 14344201 14344202 14344203 14344204 14344205 14344206 14344207 14344208 14344209 14344210 14344211 14344212 14344213 14344214 14344215 14344216 14344217 14344218 14344219 14344220 14344221 14344222 14344223 14344224 14344225 14344227 14344228 14344229 14344232 14344233 14344234 14344235 14344236 14344237 14344239 14344240 14344241 14344243 14344244 14344246 14344247 14344248 14344249 14344251 14344252 14344253 14344254 14344255 14344256 14344257 14344258 14344260 14344261 14344262 14344263 14344264 14344265 14344266 14344267 14344268 14344269 14344270 14344271 14344272 14344273 14344274 14344275 14344276 14344277 14344278 14344279 14344281 14344282 14344283 14344286 14344287 14344288 14344289 14344290 14344291 14344292 14344294 14344296 14344298 14344299 14344300 14344301 14344302 14344303 14344304 14344305 14344306 14344307 14344308 14344309 14344310 14344311 14344312 14344313 14344314 14344315 14344316 14344317 14344318 14344319 14344320 14344321 14344322 14344323 14344324 14344325 14344326 14344328 14344329 14344330 14344332 14344333 14344334 14344335 14344336 14344337 14344338 14344339 14344340 14344341 14344343 14344344 14344345 14344346 14344347 14344348 14344349 14344350 14344351 14344352 14344353 14344354 14344355 14344356 14344357 14344359 14344360 14344361 14344362 14344364 14344365 14344366 14344367 14344368 14344369 14344371 14344372 14344373 14344374 14344375 14344376 14344377 14344378 14344379 14344380 14344381 14344382 14344383 14344384 14344385 14344386 14344388 14344389 14344390 14344391 14344392 14344393 14344394 14344395 14344396 14344397 14344399 14344400 14344401 14344402 14344403 14344404 14344405 14344406 14344407 14344409 14344410 14344411 14344412 14344413 14344414 14344415 14344416 14344417 14344418 14344419 14344420 14344422 14344423 14344424 14344425 14344426 14344427 14344428 14344429 14344430 14344431 14344432 14344433 14344434 14344435 14344436 14344437 14344438 14344439 14344440 14344441 14344443 14344444 14344445 14344447 14344448 14344449 14344450 14344451 14344452 14344453 14344454 14344455 14344456 14344457 14344458 14344459 14344460 14344461 14344462 14344463 14344464 14344466 14344467 14344468 14344469 14344470 14344471 14344472 14344473 14344474 14344475 14344476 14344477 14344478 14344479 14344480 14344481 14344482 14344483 14344484 14344485 14344486 14344487 14344488 14344489 14344491 14344493 14344494 14344497 14344498 14344499 14344500 14344501 14344502 14344503 14344504 14344505 14344506 14344508 14344509 14344510 14344511 14344512 14344514 14344515 14344516 14344517 14344518 14344519 14344520 14344523 14344524 14344525 14344526 14344527 14344528 14344529 14344530 14344532 14344533 14344534 14344535 14344536 14344537 14344538 14344539 14344540 14344541 14344542 14344543 14344544 14344545 14344546 14344547 14344548 14344549 14344550 14344551 14344552 14344553 14344554 14344555 14344556 14344557 14344559 14344560 14344561 14344562 14344563 14344564 14344566 14344567 14344568 14344569 14344570 14344572 14344574 14344575 14344576 14344577 14344578 14344579 14344580 14344581 14344582 14344583 14344584 14344585 14344586 14344587 14344588 14344589 14344591 14344592 14344593 14344594 14344595 14344596 14344598 14344599 14344600 14344601 14344602 14344603 14344604 14344605 14344606 14344607 14344608 14344609 14344610 14344611 14344612 14344613 14344614 14344615 14344616 14344617 14344618 14344619 14344620 14344621 14344622 14344623 14344624 14344625 14344626 14344627 14344628 14344629 14344630 14344631 14344632 14344633 14344634 14344635 14344636 14344637 14344638 14344639 14344640 14344641 14344642 14344643 14344644 14344645 14344646 14344647 14344648 14344649 14344650 14344651 14344652 14344653 14344654 14344655 14344656 14344657 14344658 14344659 14344660 14344661 14344662 14344663 14344664 14344665 14344666 14344667 14344668 14344669 14344670 14344671 14344672 14344673 14344674 14344675 14344676 14344677 14344678 14344679 14344680 14344681 14344682 14344683 14344684 14344686 14344687 14344688 14344689 14344690 14344691 14344692 14344693 14344694 14344695 14344696 14344697 14344698 14344699 14344700 14344701 14344702 14344703 14344704 14344705 14344706 14344707 14344708 14344709 14344710 14344711 14344712 14344713 14344714 14344715 14344716 14344717 14344718 14344719 14344720 14344721 14344722 14344723 14344724 14344725 14344726 14344727 14344728 14344729 14344730 14344731 14344732 14344733 14344734 14344735 14344736 14344737 14344738 14344739 14344740 14344741 14344742 14344743 14344744 14344745 14344746 14344747 14344748 14344749 14344750 14344751 14344752 14344753 14344754 14344755 14344756 14344757 14344758 14344759 14344760 14344761 14344762 14344763 14344764 14344765 14344766 14344767 14344768 14344769 14344770 14344771 14344772 14344773 14344775 14344776 14344777 14344778 14344779 14344780 14344781 14344782 14344783 14344784 14344785 14344786 14344788 14344789 14344790 14344791 14344792 14344793 14344794 14344795 14344796 14344797 14344798 14344799 14344800 14344801 14344802 14344803 14344804 14344805 14344806 14344807 14344808 14344809 14344810 14344811 14344812 14344813 14344814 14344815 14344816 14344817 14344818 14344819 14344820 14344821 14344823 14344824 14344825 14344826 14344827 14344828 14344829 14344830 14344831 14344833 14344834 14344835 14344837 14344838 14344839 14344840 14344841 14344842 14344843 14344844 14344845 14344846 14344847 14344848 14344849 14344850 14344851 14344852 14344853 14344854 14344855 14344856 14344857 14344858 14344859 14344860 14344861 14344862 14344863 14344864 14344865 14344866 14344867 14344868 14344869 14344870 14344871 14344872 14344873 14344874 14344875 14344876 14344877 14344878 14344879 14344880 14344881 14344882 14344883 14344884 14344885 14344887 14344888 14344889 14344890 14344891 14344892 14344893 14344894 14344895 14344897 14344898 14344899 14344900 14344902 14344903 14344904 14344905 14344906 14344907 14344908 14344909 14344910 14344911 14344912 14344913 14344914 14344915 14344916 14344918 14344919 14344920 14344921 14344922 14344923 14344924 14344925 14344926 14344927 14344928 14344929 14344930 14344931 14344932 14344933 14344934 14344935 14344936 14344937 14344938 14344939 14344940 14344941 14344942 14344943 14344944 14344945 14344946 14344948 14344950 14344951 14344952 14344955 14344956 14344957 14344959 14344960 14344961 14344962 14344963 14344965 14344966 14344967 14344970 14344971 14344975 14344978 14344982 14344985 14344986 14344988 14344989 14344991 14344993 14344994 14344996 14344997 14344998 14344999 14345000 14345001 14345002 14345003 14345006 14345007 14345008 14345009 14345012 14345013 14345014 14345015 14345016 14345017 14345018 14345019 14345020 14345021 14345022 14345023 14345027 14345028 14345029 14345032 14345034 14345036 14345037 14345041 14345044 14345046 14345048 14345049 14345050 14345051 14345052 14345054 14345055 14345056 14345057 14345061 14345062 14345063 14345066 14345067 14345068 14345069 14345070 14345071 14345074 14345075 14345076 14345077 14345079 14345080 14345081 14345082 14345084 14345085 14345088 14345089 14345091 14345092 14345093 14345094 14345095 14345096 14345097 14345098 14345099 14345100 14345101 14345102 14345103 14345104 14345105 14345106 14345107 14345109 14345110 14345111 14345112 14345113 14345114 14345116 14345117 14345119 14345120 14345122 14345124 14345125 14345126 14345127 14345128 14345129 14345130 14345131 14345132 14345134 14345135 14345136 14345138 14345139 14345140 14345141 14345142 14345145 14345146 14345147 14345148 14345149 14345150 14345151 14345153 14345154 14345155 14345156 14345157 14345158 14345159 14345160 14345161 14345162 14345164 14345165 14345166 14345167 14345168 14345169 14345170 14345171 14345172 14345174 14345175 14345176 14345177 14345180 14345181 14345182 14345183 14345184 14345185 14345186 14345187 14345188 14345189 14345190 14345191 14345192 14345193 14345194 14345195 14345196 14345197 14345198 14345199 14345200 14345201 14345202 14345203 14345204 14345205 14345206 14345207 14345208 14345209 14345210 14345211 14345212 14345213 14345214 14345215 14345216 14345217 14345218 14345219 14345220 14345222 14345223 14345224 14345225 14345226 14345227 14345228 14345229 14345230 14345231 14345232 14345233 14345234 14345235 14345236 14345237 14345238 14345239 14345240 14345241 14345243 14345244 14345245 14345246 14345247 14345248 14345249 14345250 14345251 14345252 14345253 14345254 14345255 14345256 14345257 14345258 14345259 14345260 14345261 14345262 14345263 14345264 14345265 14345266 14345267 14345268 14345269 14345270 14345271 14345272 14345273 14345274 14345275 14345276 14345277 14345278 14345279 14345280 14345281 14345282 14345283 14345284 14345285 14345286 14345287 14345288 14345289 14345290 14345291 14345292 14345293 14345294 14345295 14345296 14345297 14345299 14345300 14345301 14345302 14345303 14345304 14345305 14345306 14345307 14345308 14345309 14345310 14345311 14345312 14345313 14345314 14345315 14345317 14345318 14345319 14345320 14345321 14345322 14345323 14345324 14345325 14345326 14345327 14345328 14345329 14345330 14345331 14345332 14345333 14345334 14345335 14345336 14345337 14345338 14345339 14345340 14345341 14345343 14345344 14345345 14345346 14345347 14345348 14345349 14345350 14345351 14345352 14345353 14345354 14345355 14345356 14345357 14345358 14345359 14345360 14345361 14345362 14345363 14345365 14345366 14345367 14345368 14345369 14345370 14345371 14345372 14345374 14345375 14345376 14345378 14345379 14345381 14345384 14345385 14345387 14345388 14345389 14345390 14345391 14345392 14345393 14345394 14345395 14345396 14345397 14345398 14345399 14345400 14345401 14345402 14345403 14345404 14345405 14345406 14345407 14345408 14345409 14345410 14345411 14345412 14345413 14345414 14345415 14345416 14345417 14345419 14345420 14345421 14345422 14345424 14345425 14345426 14345427 14345428 14345430 14345431 14345432 14345433 14345434 14345435 14345436 14345437 14345441 14345442 14345443 14345444 14345445 14345446 14345447 14345448 14345449 14345450 14345451 14345452 14345453 14345454 14345455 14345456 14345457 14345460 14345461 14345462 14345463 14345464 14345465 14345466 14345467 14345468 14345469 14345470 14345471 14345472 14345473 14345474 14345475 14345477 14345478 14345479 14345480 14345481 14345483 14345485 14345486 14345487 14345488 14345489 14345490 14345492 14345493 14345495 14345496 14345497 14345498 14345499 14345502 14345503 14345505 14345506 14345507 14345508 14345509 14345510 14345511 14345512 14345513 14345514 14345515 14345516 14345517 14345518 14345519 14345521 14345522 14345523 14345524 14345525 14345526 14345527 14345528 14345530 14345533 14345534 14345536 14345537 14345540 14345541 14345548 14345549 14345550 14345553 14345555 14345556 14345558 14345559 14345561 14345563 14345564 14345565 14345566 14345567 14345568 14345569 14345570 14345572 14345573 14345574 14345575 14345576 14345577 14345578 14345579 14345580 14345581 14345582 14345584 14345585 14345587 14345589 14345590 14345591 14345592 14345593 14345594 14345595 14345596 14345598 14345601 14345602 14345603 14345605 14345606 14345607 14345608 14345609 14345610 14345611 14345612 14345613 14345614 14345615 14345616 14345617 14345618 14345619 14345620 14345622 14345623 14345624 14345625 14345627 14345628 14345629 14345630 14345631 14345632 14345633 14345634 14345635 14345641 14345642 14345643 14345646 14345647 14345648 14345649 14345650 14345651 14345654 14345655 14345659 14345661 14345662 14345663 14345664 14345665 14345666 14345668 14345670 14345671 14345673 14345675 14345678 14345679 14345681 14345682 14345683 14345686 14345687 14345688 14345689 14345690 14345691 14345692 14345694 14345695 14345696 14345699 14345700 14345703 14345704 14345705 14345706 14345707 14345708 14345709 14345710 14345712 14345713 14345714 14345715 14345716 14345718 14345720 14345722 14345723 14345725 14345726 14345729 14345730 14345731 14345732 14345733 14345735 14345737 14345738 14345739 14345740 14345741 14345742 14345743 14345744 14345745 14345746 14345748 14345749 14345750 14345751 14345752 14345753 14345754 14345755 14345757 14345758 14345759 14345760 14345761 14345762 14345763 14345764 14345765 14345766 14345767 14345768 14345769 14345770 14345771 14345772 14345773 14345774 14345775 14345776 14345777 14345778 14345779 14345780 14345781 14345782 14345783 14345784 14345785 14345786 14345787 14345788 14345789 14345790 14345791 14345793 14345794 14345795 14345796 14345797 14345798 14345799 14345800 14345801 14345802 14345803 14345804 14345805 14345808 14345809 14345810 14345811 14345812 14345813 14345814 14345815 14345816 14345817 14345818 14345819 14345820 14345821 14345822 14345823 14345824 14345825 14345826 14345827 14345828 14345829 14345830 14345831 14345832 14345833 14345834 14345835 14345836 14345837 14345838 14345839 14345840 14345841 14345842 14345843 14345845 14345847 14345848 14345849 14345850 14345851 14345852 14345853 14345854 14345855 14345856 14345857 14345859 14345860 14345861 14345862 14345864 14345865 14345867 14345868 14345869 14345871 14345872 14345873 14345874 14345875 14345876 14345877 14345878 14345879 14345880 14345881 14345882 14345884 14345885 14345886 14345887 14345888 14345889 14345890 14345891 14345892 14345893 14345894 14345895 14345896 14345897 14345898 14345899 14345900 14345901 14345902 14345903 14345904 14345905 14345906 14345907 14345908 14345909 14345910 14345913 14345914 14345915 14345916 14345918 14345919 14345920 14345921 14345922 14345923 14345924 14345925 14345926 14345927 14345928 14345929 14345930 14345931 14345932 14345933 14345934 14345935 14345936 14345938 14345939 14345940 14345941 14345942 14345943 14345944 14345945 14345946 14345947 14345949 14345950 14345951 14345952 14345953 14345954 14345955 14345956 14345957 14345958 14345959 14345960 14345961 14345962 14345963 14345964 14345965 14345966 14345967 14345968 14345969 14345970 14345971 14345972 14345973 14345975 14345976 14345977 14345978 14345979 14345980 14345981 14345982 14345983 14345984 14345985 14345986 14345987 14345988 14345989 14345990 14345991 14345992 14345993 14345994 14345995 14345996 14345998 14345999 14346000 14346001 14346002 14346003 14346004 14346005 14346006 14346007 14346008 14346009 14346010 14346011 14346012 14346013 14346014 14346015 14346016 14346017 14346018 14346019 14346020 14346022 14346023 14346024 14346025 14346026 14346028 14346029 14346030 14346031 14346032 14346034 14346035 14346036 14346037 14346038 14346039 14346040 14346041 14346042 14346043 14346044 14346045 14346046 14346047 14346048 14346049 14346050 14346051 14346052 14346053 14346054 14346055 14346056 14346057 14346058 14346059 14346060 14346061 14346062 14346063 14346064 14346065 14346066 14346067 14346068 14346069 14346070 14346071 14346072 14346073 14346074 14346075 14346076 14346077 14346078 14346079 14346080 14346081 14346082 14346083 14346084 14346085 14346086 14346087 14346088 14346089 14346090 14346091 14346092 14346093 14346094 14346095 14346096 14346097 14346098 14346099 14346100 14346101 14346102 14346103 14346104 14346105 14346106 14346107 14346108 14346109 14346110 14346111 14346112 14346113 14346114 14346115 14346116 14346117 14346118 14346119 14346120 14346121 14346122 14346123 14346124 14346125 14346126 14346127 14346128 14346129 14346130 14346131 14346132 14346133 14346134 14346135 14346136 14346137 14346138 14346139 14346140 14346141 14346142 14346143 14346144 14346145 14346146 14346148 14346149 14346150 14346151 14346152 14346153 14346154 14346155 14346156 14346157 14346158 14346159 14346160 14346161 14346162 14346163 14346166 14346168 14346169 14346170 14346171 14346172 14346173 14346174 14346175 14346176 14346177 14346178 14346179 14346180 14346182 14346183 14346186 14346188 14346189 14346190 14346191 14346192 14346193 14346195 14346197 14346198 14346199 14346200 14346201 14346202 14346203 14346204 14346205 14346206 14346207 14346208 14346209 14346210 14346211 14346212 14346213 14346214 14346215 14346216 14346218 14346219 14346220 14346221 14346222 14346223 14346224 14346225 14346226 14346227 14346229 14346231 14346232 14346233 14346234 14346235 14346237 14346238 14346239 14346240 14346241 14346242 14346244 14346245 14346246 14346247 14346249 14346250 14346251 14346252 14346253 14346254 14346255 14346256 14346257 14346258 14346259 14346260 14346262 14346263 14346264 14346265 14346266 14346267 14346268 14346269 14346270 14346272 14346273 14346274 14346275 14346277 14346278 14346279 14346280 14346282 14346283 14346284 14346285 14346286 14346287 14346288 14346289 14346291 14346292 14346293 14346296 14346297 14346298 14346299 14346300 14346301 14346303 14346305 14346306 14346307 14346308 14346309 14346310 14346311 14346312 14346313 14346314 14346315 14346317 14346318 14346319 14346320 14346321 14346322 14346323 14346324 14346327 14346328 14346329 14346330 14346331 14346333 14346334 14346335 14346337 14346338 14346339 14346340 14346341 14346342 14346343 14346344 14346345 14346347 14346348 14346350 14346351 14346352 14346353 14346354 14346355 14346356 14346357 14346358 14346359 14346360 14346362 14346363 14346364 14346365 14346366 14346367 14346368 14346369 14346371 14346372 14346374 14346375 14346376 14346377 14346378 14346379 14346380 14346382 14346383 14346384 14346385 14346386 14346390 14346391 14346392 14346393 14346394 14346395 14346396 14346397 14346398 14346399 14346400 14346401 14346402 14346403 14346405 14346406 14346409 14346410 14346411 14346412 14346413 14346414 14346415 14346416 14346417 14346418 14346420 14346421 14346422 14346423 14346424 14346425 14346427 14346429 14346430 14346431 14346432 14346433 14346434 14346435 14346436 14346437 14346438 14346439 14346441 14346442 14346443 14346444 14346445 14346446 14346448 14346449 14346451 14346452 14346453 14346456 14346457 14346458 14346459 14346460 14346461 14346462 14346463 14346464 14346465 14346466 14346468 14346469 14346470 14346471 14346472 14346473 14346474 14346476 14346477 14346478 14346479 14346480 14346482 14346483 14346484 14346485 14346486 14346487 14346490 14346491 14346492 14346495 14346496 14346497 14346499 14346501 14346502 14346503 14346505 14346506 14346507 14346508 14346509 14346510 14346511 14346512 14346513 14346514 14346515 14346516 14346517 14346518 14346519 14346520 14346521 14346522 14346523 14346525 14346526 14346527 14346528 14346529 14346530 14346531 14346532 14346533 14346535 14346536 14346537 14346538 14346539 14346541 14346542 14346543 14346544 14346545 14346546 14346547 14346548 14346549 14346550 14346551 14346552 14346553 14346554 14346555 14346562 14346564 14346565 14346566 14346570 14346571 14346572 14346573 14346574 14346575 14346576 14346577 14346578 14346579 14346580 14346581 14346582 14346583 14346584 14346585 14346586 14346587 14346588 14346590 14346591 14346593 14346594 14346595 14346596 14346597 14346598 14346599 14346600 14346601 14346602 14346603 14346605 14346606 14346608 14346609 14346611 14346612 14346613 14346614 14346615 14346616 14346617 14346618 14346619 14346620 14346621 14346622 14346623 14346624 14346626 14346627 14346628 14346629 14346632 14346633 14346634 14346635 14346637 14346638 14346640 14346641 14346642 14346643 14346644 14346646 14346647 14346648 14346649 14346650 14346651 14346652 14346653 14346654 14346655 14346656 14346657 14346658 14346659 14346660 14346662 14346664 14346666 14346667 14346668 14346669 14346670 14346671 14346672 14346674 14346675 14346676 14346678 14346679 14346680 14346681 14346682 14346683 14346684 14346685 14346686 14346687 14346689 14346690 14346691 14346693 14346694 14346695 14346696 14346697 14346698 14346699 14346700 14346702 14346703 14346704 14346705 14346706 14346708 14346710 14346711 14346712 14346713 14346714 14346715 14346716 14346717 14346718 14346719 14346720 14346721 14346722 14346724 14346725 14346726 14346728 14346729 14346730 14346731 14346732 14346734 14346735 14346738 14346740 14346741 14346742 14346743 14346744 14346745 14346746 14346747 14346748 14346749 14346750 14346751 14346752 14346753 14346754 14346755 14346756 14346758 14346759 14346760 14346761 14346762 14346763 14346765 14346766 14346767 14346769 14346771 14346772 14346773 14346774 14346775 14346776 14346777 14346778 14346779 14346780 14346781 14346782 14346783 14346784 14346785 14346786 14346787 14346788 14346789 14346791 14346793 14346794 14346795 14346796 14346797 14346798 14346800 14346801 14346802 14346803 14346804 14346805 14346806 14346807 14346810 14346811 14346812 14346813 14346815 14346816 14346819 14346820 14346821 14346823 14346825 14346827 14346828 14346829 14346831 14346834 14346835 14346836 14346837 14346838 14346839 14346840 14346841 14346843 14346844 14346845 14346846 14346848 14346849 14346850 14346851 14346852 14346853 14346854 14346855 14346857 14346859 14346860 14346861 14346862 14346863 14346864 14346865 14346866 14346867 14346868 14346869 14346870 14346871 14346872 14346874 14346875 14346876 14346878 14346879 14346880 14346881 14346882 14346884 14346885 14346886 14346888 14346889 14346890 14346891 14346893 14346894 14346898 14346899 14346901 14346903 14346904 14346905 14346906 14346907 14346909 14346910 14346911 14346913 14346914 14346917 14346918 14346919 14346920 14346921 14346922 14346923 14346925 14346928 14346929 14346931 14346933 14346934 14346936 14346937 14346939 14346941 14346942 14346950 14346951 14346953 14346955 14346958 14346960 14346962 14346964 14346965 14346969 14346970 14346971 14346972 14346974 14346975 14346976 14346979 14346983 14346984 14346985 14346988 14346989 14346992 14346996 14346998 14346999 14347003 14347006 14347007 14347010 14347012 14347015 14347017 14347018 14347019 14347020 14347025 14347031 14347032 14347033 14347034 14347038 14347039 14347042 14347044 14347045 14347047 14347048 14347049 14347051 14347056 14347060 14347061 14347062 14347063 14347066 14347067 14347068 14347069 14347070 14347071 14347072 14347074 14347075 14347076 14347077 14347078 14347079 14347080 14347101 14347109 14347112 14347113 14347115 14347116 14347118 14347120 14347122 14347124 14347125 14347126 14347130 14347132 14347134 14347136 14347138 14347139 14347140 14347141 14347142 14347143 14347144 14347145 14347146 14347148 14347149 14347150 14347151 14347152 14347153 14347154 14347155 14347156 14347157 14347158 14347160 14347161 14347162 14347163 14347164 14347165 14347167 14347168 14347169 14347170 14347171 14347172 14347173 14347175 14347176 14347179 14347180 14347182 14347184 14347185 14347186 14347187 14347188 14347190 14347191 14347192 14347193 14347194 14347197 14347198 14347199 14347200 14347201 14347202 14347205 14347206 14347207 14347208 14347209 14347210 14347211 14347212 14347213 14347214 14347215 14347216 14347218 14347221 14347222 14347223 14347224 14347225 14347227 14347229 14347230 14347231 14347232 14347233 14347234 14347238 14347240 14347241 14347242 14347244 14347245 14347246 14347247 14347251 14347252 14347253 14347257 14347258 14347260 14347264 14347265 14347266 14347267 14347268 14347270 14347272 14347276 14347278 14347280 14347281 14347283 14347286 14347287 14347288 14347291 14347293 14347295 14347296 14347298 14347301 14347302 14347303 14347304 14347305 14347306 14347307 14347308 14347310 14347311 14347312 14347316 14347317 14347318 14347319 14347320 14347321 14347322 14347323 14347325 14347326 14347328 14347329 14347330 14347331 14347333 14347336 14347337 14347338 14347339 14347340 14347341 14347342 14347343 14347344 14347345 14347346 14347347 14347348 14347350 14347351 14347354 14347355 14347356 14347357 14347358 14347359 14347360 14347361 14347362 14347363 14347364 14347365 14347366 14347368 14347369 14347370 14347372 14347374 14347376 14347377 14347378 14347379 14347380 14347381 14347382 14347383 14347384 14347385 14347386 14347388 14347389 14347390 14347392 14347393 14347395 14347396 14347397 14347398 14347401 14347404 14347405 14347406 14347407 14347408 14347409 14347410 14347411 14347412 14347413 14347414 14347418 14347419 14347420 14347422 14347423 14347424 14347426 14347427 14347428 14347430 14347431 14347432 14347433 14347434 14347436 14347437 14347439 14347442 14347444 14347445 14347446 14347447 14347448 14347450 14347451 14347452 14347454 14347455 14347456 14347457 14347459 14347460 14347463 14347464 14347465 14347466 14347467 14347468 14347469 14347470 14347473 14347474 14347477 14347478 14347479 14347480 14347481 14347482 14347483 14347484 14347485 14347486 14347487 14347489 14347491 14347493 14347497 14347498 14347499 14347501 14347502 14347503 14347505 14347509 14347510 14347515 14347516 14347519 14347523 14347524 14347526 14347527 14347532 14347533 14347535 14347537 14347538 14347540 14347541 14347543 14347544 14347545 14347547 14347548 14347551 14347557 14347561 14347563 14347564 14347565 14347566 14347567 14347568 14347569 14347570 14347571 14347572 14347573 14347574 14347576 14347578 14347579 14347581 14347582 14347587 14347588 14347589 14347592 14347594 14347596 14347597 14347599 14347600 14347601 14347603 14347606 14347607 14347609 14347613 14347614 14347617 14347618 14347620 14347624 14347626 14347627 14347631 14347634 14347636 14347637 14347650 14347651 14347653 14347654 14347655 14347662 14347663 14347664 14347668 14347670 14347681 14347683 14347688 14347690 14347698 14347700 14347705 14347707 14347709 14347711 14347712 14347713 14347714 14347715 14347716 14347717 14347718 14347719 14347720 14347721 14347722 14347724 14347725 14347726 14347727 14347728 14347729 14347730 14347731 14347732 14347733 14347734 14347736 14347737 14347738 14347739 14347740 14347741 14347742 14347743 14347744 14347745 14347746 14347748 14347749 14347750 14347751 14347752 14347753 14347754 14347755 14347756 14347757 14347758 14347759 14347760 14347761 14347762 14347763 14347764 14347765 14347766 14347767 14347768 14347769 14347770 14347771 14347772 14347773 14347774 14347775 14347776 14347777 14347778 14347779 14347780 14347781 14347783 14347784 14347785 14347787 14347790 14347791 14347793 14347795 14347796 14347797 14347798 14347799 14347801 14347803 14347804 14347805 14347807 14347808 14347809 14347810 14347811 14347812 14347814 14347815 14347816 14347817 14347818 14347820 14347821 14347824 14347826 14347828 14347829 14347830 14347831 14347832 14347833 14347834 14347835 14347836 14347837 14347838 14347839 14347840 14347841 14347842 14347844 14347845 14347848 14347851 14347852 14347853 14347854 14347856 14347857 14347858 14347859 14347860 14347861 14347862 14347864 14347865 14347866 14347868 14347871 14347872 14347873 14347874 14347875 14347877 14347879 14347880 14347881 14347882 14347884 14347885 14347886 14347887 14347890 14347891 14347892 14347893 14347894 14347895 14347896 14347897 14347898 14347899 14347901 14347902 14347903 14347904 14347905 14347906 14347908 14347910 14347913 14347914 14347915 14347916 14347917 14347919 14347920 14347921 14347924 14347925 14347926 14347927 14347929 14347931 14347933 14347934 14347935 14347937 14347938 14347940 14347942 14347943 14347944 14347945 14347946 14347947 14347948 14347951 14347952 14347953 14347954 14347955 14347956 14347957 14347958 14347959 14347961 14347962 14347963 14347964 14347965 14347966 14347968 14347969 14347970 14347971 14347973 14347974 14347975 14347978 14347979 14347981 14347982 14347983 14347986 14347988 14347990 14347992 14347993 14347994 14347996 14347997 14347998 14347999 14348000 14348002 14348003 14348004 14348006 14348007 14348008 14348009 14348010 14348011 14348012 14348013 14348014 14348015 14348016 14348018 14348019 14348020 14348021 14348022 14348023 14348024 14348025 14348026 14348027 14348028 14348029 14348030 14348031 14348032 14348033 14348035 14348036 14348037 14348039 14348040 14348041 14348042 14348044 14348045 14348046 14348047 14348048 14348049 14348050 14348051 14348052 14348053 14348054 14348055 14348056 14348057 14348058 14348059 14348060 14348061 14348062 14348063 14348064 14348066 14348067 14348069 14348071 14348072 14348073 14348074 14348075 14348076 14348077 14348079 14348080 14348082 14348083 14348084 14348085 14348086 14348087 14348088 14348089 14348090 14348091 14348092 14348093 14348094 14348095 14348097 14348098 14348099 14348100 14348101 14348102 14348103 14348104 14348105 14348108 14348109 14348111 14348112 14348113 14348114 14348115 14348116 14348117 14348118 14348119 14348120 14348121 14348122 14348123 14348124 14348125 14348128 14348130 14348131 14348132 14348133 14348134 14348135 14348137 14348140 14348142 14348144 14348145 14348146 14348147 14348148 14348149 14348151 14348152 14348153 14348154 14348156 14348157 14348158 14348159 14348160 14348161 14348162 14348163 14348164 14348166 14348169 14348170 14348171 14348172 14348173 14348174 14348175 14348176 14348177 14348179 14348180 14348181 14348184 14348185 14348191 14348192 14348193 14348194 14348195 14348197 14348198 14348199 14348200 14348201 14348202 14348203 14348204 14348207 14348208 14348209 14348210 14348211 14348213 14348214 14348215 14348216 14348217 14348219 14348220 14348221 14348222 14348223 14348225 14348226 14348229 14348231 14348232 14348235 14348236 14348237 14348238 14348240 14348241 14348242 14348243 14348244 14348245 14348247 14348250 14348251 14348252 14348253 14348255 14348257 14348258 14348259 14348260 14348261 14348263 14348265 14348266 14348267 14348268 14348269 14348270 14348271 14348272 14348273 14348274 14348275 14348277 14348278 14348279 14348280 14348281 14348282 14348283 14348284 14348285 14348287 14348289 14348291 14348292 14348293 14348294 14348295 14348296 14348299 14348300 14348301 14348302 14348303 14348304 14348306 14348307 14348309 14348310 14348311 14348312 14348313 14348314 14348315 14348317 14348318 14348319 14348320 14348321 14348322 14348323 14348324 14348325 14348326 14348327 14348328 14348329 14348330 14348332 14348333 14348334 14348335 14348336 14348337 14348338 14348339 14348341 14348343 14348345 14348346 14348347 14348348 14348349 14348350 14348351 14348353 14348354 14348355 14348356 14348357 14348358 14348359 14348364 14348365 14348366 14348367 14348368 14348369 14348370 14348372 14348373 14348374 14348375 14348377 14348378 14348379 14348380 14348381 14348382 14348384 14348385 14348386 14348387 14348388 14348389 14348390 14348391 14348392 14348393 14348394 14348395 14348396 14348397 14348398 14348399 14348400 14348401 14348402 14348403 14348404 14348405 14348406 14348408 14348409 14348410 14348411 14348412 14348413 14348414 14348415 14348417 14348418 14348421 14348422 14348423 14348426 14348428 14348429 14348430 14348432 14348433 14348435 14348436 14348438 14348439 14348440 14348441 14348442 14348443 14348446 14348447 14348448 14348449 14348450 14348451 14348452 14348453 14348454 14348455 14348456 14348457 14348458 14348459 14348460 14348461 14348462 14348463 14348464 14348466 14348467 14348468 14348470 14348471 14348472 14348473 14348474 14348475 14348476 14348478 14348479 14348480 14348482 14348483 14348484 14348485 14348486 14348487 14348488 14348489 14348490 14348491 14348492 14348493 14348494 14348495 14348496 14348497 14348498 14348499 14348500 14348502 14348503 14348504 14348505 14348506 14348507 14348508 14348509 14348510 14348511 14348512 14348514 14348515 14348516 14348517 14348518 14348519 14348520 14348521 14348523 14348524 14348525 14348526 14348527 14348529 14348531 14348532 14348533 14348534 14348535 14348536 14348539 14348540 14348541 14348542 14348545 14348546 14348547 14348548 14348549 14348550 14348551 14348552 14348553 14348554 14348555 14348556 14348557 14348558 14348559 14348560 14348561 14348562 14348563 14348568 14348569 14348570 14348572 14348573 14348575 14348576 14348577 14348578 14348580 14348581 14348582 14348584 14348585 14348586 14348587 14348588 14348589 14348590 14348591 14348592 14348593 14348594 14348595 14348596 14348597 14348598 14348599 14348601 14348602 14348604 14348605 14348608 14348609 14348610 14348611 14348614 14348616 14348617 14348618 14348619 14348620 14348621 14348622 14348623 14348624 14348625 14348626 14348627 14348628 14348629 14348630 14348631 14348632 14348633 14348634 14348635 14348636 14348637 14348639 14348640 14348641 14348642 14348643 14348644 14348645 14348646 14348647 14348649 14348650 14348651 14348653 14348654 14348655 14348656 14348657 14348660 14348661 14348662 14348664 14348665 14348666 14348667 14348668 14348670 14348672 14348673 14348675 14348676 14348677 14348678 14348679 14348680 14348682 14348683 14348686 14348687 14348688 14348691 14348692 14348693 14348694 14348695 14348696 14348697 14348698 14348701 14348702 14348704 14348705 14348707 14348708 14348709 14348711 14348713 14348714 14348716 14348719 14348720 14348721 14348722 14348723 14348724 14348725 14348726 14348727 14348728 14348731 14348732 14348733 14348735 14348736 14348737 14348738 14348739 14348741 14348742 14348744 14348746 14348747 14348748 14348749 14348750 14348751 14348752 14348754 14348755 14348757 14348758 14348760 14348761 14348762 14348763 14348764 14348765 14348766 14348769 14348770 14348772 14348773 14348776 14348778 14348779 14348780 14348781 14348782 14348783 14348786 14348788 14348789 14348790 14348791 14348792 14348796 14348799 14348800 14348801 14348802 14348803 14348805 14348807 14348808 14348810 14348811 14348813 14348814 14348817 14348821 14348822 14348823 14348824 14348825 14348826 14348827 14348828 14348829 14348830 14348831 14348832 14348833 14348836 14348837 14348838 14348839 14348840 14348841 14348842 14348843 14348844 14348845 14348846 14348848 14348849 14348850 14348851 14348852 14348854 14348855 14348856 14348857 14348858 14348859 14348860 14348861 14348862 14348863 14348864 14348865 14348866 14348867 14348868 14348869 14348870 14348872 14348873 14348875 14348876 14348877 14348879 14348880 14348881 14348882 14348884 14348885 14348887 14348888 14348890 14348891 14348893 14348894 14348895 14348896 14348899 14348901 14348902 14348903 14348905 14348906 14348907 14348910 14348912 14348913 14348914 14348915 14348917 14348918 14348919 14348920 14348921 14348922 14348923 14348924 14348925 14348926 14348927 14348928 14348929 14348930 14348931 14348932 14348933 14348935 14348936 14348939 14348941 14348942 14348943 14348944 14348946 14348948 14348949 14348950 14348951 14348952 14348953 14348954 14348955 14348956 14348958 14348959 14348960 14348961 14348962 14348963 14348965 14348967 14348968 14348969 14348970 14348972 14348974 14348977 14348978 14348979 14348980 14348981 14348983 14348984 14348985 14348987 14348988 14348989 14348992 14348993 14348995 14348998 14348999 14349000 14349001 14349002 14349003 14349004 14349005 14349006 14349007 14349008 14349011 14349012 14349013 14349014 14349015 14349016 14349017 14349018 14349019 14349020 14349021 14349022 14349025 14349027 14349028 14349031 14349032 14349033 14349034 14349035 14349038 14349039 14349040 14349042 14349043 14349044 14349045 14349046 14349047 14349048 14349049 14349051 14349052 14349053 14349054 14349055 14349056 14349057 14349058 14349059 14349060 14349062 14349065 14349066 14349068 14349069 14349070 14349072 14349073 14349074 14349075 14349076 14349078 14349081 14349082 14349084 14349085 14349087 14349088 14349089 14349092 14349096 14349097 14349098 14349099 14349100 14349102 14349103 14349104 14349108 14349109 14349113 14349114 14349116 14349117 14349118 14349119 14349120 14349123 14349124 14349125 14349126 14349127 14349128 14349129 14349130 14349133 14349135 14349136 14349137 14349138 14349139 14349140 14349141 14349142 14349143 14349144 14349145 14349146 14349147 14349149 14349150 14349151 14349152 14349153 14349154 14349156 14349157 14349158 14349159 14349160 14349161 14349163 14349164 14349165 14349166 14349167 14349168 14349170 14349172 14349173 14349174 14349175 14349176 14349177 14349178 14349179 14349180 14349182 14349183 14349184 14349187 14349188 14349189 14349190 14349192 14349193 14349194 14349195 14349196 14349197 14349198 14349199 14349200 14349201 14349202 14349205 14349207 14349208 14349209 14349210 14349211 14349212 14349214 14349216 14349217 14349218 14349219 14349221 14349222 14349223 14349224 14349226 14349227 14349228 14349229 14349230 14349231 14349232 14349233 14349236 14349237 14349238 14349242 14349245 14349246 14349247 14349248 14349250 14349251 14349253 14349254 14349255 14349256 14349258 14349259 14349261 14349263 14349264 14349265 14349266 14349268 14349269 14349270 14349271 14349272 14349273 14349274 14349275 14349276 14349277 14349279 14349283 14349284 14349285 14349286 14349287 14349288 14349290 14349291 14349294 14349295 14349296 14349298 14349299 14349300 14349301 14349303 14349304 14349305 14349306 14349307 14349308 14349309 14349310 14349311 14349312 14349313 14349314 14349315 14349316 14349318 14349320 14349322 14349323 14349324 14349325 14349326 14349327 14349328 14349329 14349330 14349332 14349333 14349334 14349336 14349338 14349340 14349342 14349343 14349344 14349345 14349347 14349348 14349349 14349351 14349352 14349353 14349355 14349356 14349357 14349358 14349360 14349361 14349362 14349366 14349367 14349368 14349369 14349370 14349373 14349374 14349376 14349379 14349380 14349381 14349382 14349383 14349384 14349385 14349386 14349387 14349388 14349389 14349390 14349391 14349392 14349394 14349395 14349396 14349397 14349400 14349401 14349404 14349405 14349407 14349408 14349409 14349410 14349411 14349412 14349413 14349414 14349415 14349416 14349417 14349419 14349420 14349421 14349422 14349423 14349426 14349428 14349429 14349430 14349432 14349433 14349435 14349436 14349437 14349439 14349441 14349442 14349443 14349444 14349445 14349447 14349448 14349450 14349451 14349453 14349455 14349457 14349458 14349459 14349460 14349461 14349462 14349464 14349465 14349467 14349470 14349471 14349472 14349473 14349474 14349477 14349478 14349480 14349481 14349483 14349484 14349485 14349486 14349489 14349490 14349491 14349492 14349495 14349496 14349498 14349502 14349503 14349505 14349506 14349508 14349510 14349511 14349513 14349517 14349518 14349519 14349520 14349523 14349524 14349525 14349526 14349527 14349529 14349530 14349532 14349534 14349535 14349536 14349537 14349540 14349541 14349544 14349545 14349546 14349547 14349548 14349552 14349556 14349695 14349696 14349698 14349700 14349701 14349703 14349704 14349705 14349706 14349707 14349708 14349709 14349711 14349712 14349713 14349714 14349715 14349716 14349717 14349718 14349719 14349720 14349721 14349722 14349723 14349724 14349725 14349726 14349728 14349729 14349730 14349731 14349733 14349734 14349735 14349737 14349738 14349739 14349740 14349742 14349743 14349744 14349745 14349746 14349747 14349748 14349749 14349750 14349751 14349752 14349753 14349754 14349755 14349757 14349758 14349759 14349760 14349766 14349768 14349769 14349770 14349771 14349774 14349776 14349778 14349779 14349781 14349782 14349785 14349786 14349787 14349789 14349790 14349791 14349792 14349794 14349795 14349796 14349798 14349799 14349800 14349801 14349802 14349805 14349806 14349808 14349809 14349811 14349812 14349813 14349814 14349816 14349817 14349819 14349820 14349823 14349824 14349825 14349826 14349828 14349835 14349836 14349837 14349838 14349839 14349840 14349841 14349844 14349845 14349851 14349852 14349853 14349854 14349856 14349857 14349859 14349861 14349862 14349863 14349864 14349865 14349867 14349871 14349872 14349873 14349875 14349876 14349877 14349878 14349879 14349880 14349881 14349882 14349883 14349885 14349886 14349887 14349888 14349889 14349890 14349891 14349892 14349894 14349895 14349896 14349897 14349898 14349899 14349900 14349901 14349902 14349903 14349904 14349905 14349907 14349908 14349909 14349910 14349911 14349912 14349913 14349915 14349916 14349917 14349920 14349924 14349925 14349927 14349928 14349929 14349930 14349934 14349936 14349938 14349940 14349943 14349944 14349948 14349949 14349950 14349951 14349952 14349955 14349957 14349961 14349962 14349963 14349964 14349965 14349967 14349968 14349969 14349972 14349977 14349978 14349979 14349983 14349984 14349985 14349986 14349987 14349988 14349989 14349990 14349992 14349993 14349995 14349996 14349997 14349999 14350000 14350001 14350002 14350004 14350008 14350010 14350011 14350013 14350014 14350015 14350016 14350017 14350018 14350019 14350020 14350022 14350023 14350024 14350025 14350026 14350027 14350029 14350030 14350032 14350034 14350036 14350038 14350041 14350042 14350043 14350044 14350045 14350050 14350052 14350055 14350057 14350060 14350061 14350063 14350065 14350066 14350067 14350068 14350069 14350071 14350072 14350073 14350074 14350078 14350079 14350082 14350083 14350087 14350088 14350090 14350091 14350093 14350094 14350095 14350096 14350097 14350098 14350099 14350100 14350101 14350102 14350104 14350105 14350106 14350107 14350109 14350110 14350112 14350115 14350117 14350119 14350121 14350122 14350123 14350125 14350126 14350129 14350131 14350133 14350134 14350135 14350136 14350137 14350138 14350139 14350140 14350141 14350142 14350143 14350144 14350145 14350146 14350147 14350148 14350151 14350152 14350153 14350156 14350158 14350159 14350160 14350167 14350169 14350171 14350172 14350173 14350174 14350175 14350178 14350179 14350180 14350181 14350182 14350186 14350187 14350188 14350189 14350190 14350191 14350193 14350194 14350195 14350196 14350199 14350200 14350201 14350202 14350203 14350204 14350205 14350206 14350208 14350209 14350210 14350211 14350213 14350214 14350216 14350217 14350218 14350219 14350220 14350222 14350223 14350224 14350225 14350228 14350231 14350234 14350235 14350237 14350239 14350241 14350244 14350245 14350246 14350249 14350250 14350252 14350255 14350258 14350259 14350260 14350264 14350265 14350266 14350268 14350270 14350271 14350272 14350278 14350282 14350283 14350284 14350285 14350287 14350288 14350290 14350291 14350292 14350293 14350294 14350296 14350299 14350303 14350305 14350306 14350307 14350309 14350311 14350312 14350313 14350316 14350317 14350319 14350320 14350323 14350326 14350327 14350329 14350330 14350336 14350337 14350338 14350339 14350340 14350341 14350342 14350343 14350344 14350345 14350348 14350351 14350352 14350353 14350355 14350357 14350358 14350361 14350362 14350364 14350367 14350368 14350369 14350371 14350372 14350378 14350382 14350383 14350384 14350385 14350386 14350387 14350389 14350390 14350392 14350393 14350395 14350396 14350397 14350398 14350399 14350400 14350401 14350404 14350407 14350408 14350409 14350410 14350411 14350412 14350415 14350416 14350417 14350418 14350423 14350424 14350425 14350426 14350427 14350428 14350429 14350430 14350432 14350433 14350435 14350436 14350437 14350438 14350441 14350442 14350445 14350446 14350449 14350451 14350452 14350453 14350456 14350458 14350459 14350460 14350461 14350463 14350464 14350465 14350466 14350468 14350469 14350475 14350476 14350477 14350478 14350479 14350480 14350481 14350483 14350485 14350489 14350490 14350491 14350492 14350496 14350498 14350499 14350500 14350501 14350502 14350503 14350504 14350505 14350506 14350507 14350510 14350511 14350512 14350514 14350515 14350516 14350517 14350518 14350520 14350521 14350522 14350523 14350524 14350525 14350527 14350528 14350529 14350530 14350532 14350536 14350537 14350545 14350547 14350550 14350551 14350553 14350556 14350557 14350558 14350560 14350564 14350565 14350566 14350567 14350568 14350569 14350570 14350571 14350575 14350576 14350577 14350578 14350579 14350580 14350582 14350585 14350586 14350587 14350592 14350597 14350598 14350599 14350601 14350602 14350604 14350605 14350606 14350608 14350610 14350611 14350612 14350613 14350614 14350615 14350616 14350619 14350620 14350621 14350625 14350628 14350630 14350631 14350632 14350635 14350637 14350642 14350647 14350648 14350649 14350652 14350653 14350654 14350657 14350662 14350663 14350664 14350665 14350667 14350669 14350670 14350671 14350674 14350675 14350676 14350680 14350681 14350682 14350683 14350686 14350687 14350688 14350689 14350691 14350693 14350694 14350695 14350699 14350700 14350702 14350703 14350704 14350706 14350707 14350713 14350714 14350716 14350717 14350718 14350720 14350725 14350726 14350727 14350730 14350732 14350733 14350734 14350736 14350738 14350739 14350740 14350742 14350743 14350745 14350751 14350753 14350754 14350755 14350756 14350758 14350761 14350763 14350764 14350765 14350768 14350769 14350770 14350772 14350773 14350774 14350775 14350776 14350777 14350778 14350779 14350781 14350782 14350783 14350785 14350790 14350792 14350793 14350794 14350795 14350796 14350797 14350799 14350800 14350802 14350803 14350804 14350805 14350806 14350808 14350809 14350810 14350811 14350813 14350814 14350815 14350817 14350819 14350822 14350824 14350825 14350828 14350829 14350830 14350831 14350833 14350838 14350841 14350847 14350849 14350850 14350852 14350853 14350854 14350859 14350862 14350866 14350867 14350868 14350869 14350870 14350873 14350876 14350878 14350879 14350880 14350881 14350883 14350884 14350885 14350886 14350887 14350888 14350889 14350890 14350892 14350893 14350894 14350896 14350897 14350899 14350901 14350902 14350903 14350904 14350906 14350907 14350908 14350910 14350911 14350914 14350915 14350917 14350919 14350920 14350923 14350924 14350925 14350926 14350927 14350928 14350929 14350931 14350932 14350933 14350934 14350935 14350936 14350939 14350940 14350941 14350942 14350944 14350946 14350948 14350949 14350950 14350951 14350952 14350953 14350954 14350956 14350957 14350960 14350961 14350964 14350965 14350969 14350971 14350974 14350975 14350976 14350977 14350978 14350979 14350980 14350981 14350982 14350983 14350984 14350986 14350989 14350990 14350991 14350992 14350994 14350995 14350996 14350997 14350999 14351000 14351001 14351003 14351004 14351005 14351007 14351009 14351010 14351011 14351013 14351015 14351016 14351017 14351018 14351020 14351021 14351022 14351023 14351024 14351025 14351026 14351028 14351029 14351030 14351033 14351034 14351038 14351040 14351044 14351045 14351048 14351049 14351053 14351054 14351056 14351058 14351061 14351062 14351064 14351065 14351066 14351067 14351070 14351072 14351073 14351076 14351078 14351080 14351081 14351083 14351086 14351088 14351089 14351090 14351092 14351093 14351094 14351097 14351098 14351099 14351101 14351102 14351103 14351105 14351106 14351107 14351110 14351113 14351115 14351116 14351117 14351118 14351120 14351123 14351124 14351125 14351132 14351133 14351135 14351139 14351141 14351143 14351145 14351150 14351155 14351156 14351157 14351161 14351162 14351163 14351164 14351167 14351168 14351172 14351178 14351179 14351180 14351188 14351192 14351194 14351195 14351198 14351201 14351202 14351205 14351206 14351207 14351209 14351211 14351214 14351215 14351216 14351217 14351218 14351219 14351220 14351223 14351224 14351225 14351226 14351227 14351228 14351230 14351233 14351234 14351235 14351236 14351239 14351240 14351241 14351242 14351243 14351244 14351246 14351253 14351254 14351255 14351256 14351257 14351258 14351259 14351260 14351262 14351264 14351267 14351269 14351270 14351272 14351273 14351275 14351277 14351278 14351280 14351281 14351283 14351284 14351285 14351287 14351288 14351289 14351290 14351294 14351296 14351298 14351300 14351301 14351303 14351305 14351306 14351307 14351308 14351309 14351311 14351312 14351315 14351316 14351317 14351318 14351319 14351320 14351323 14351324 14351325 14351326 14351329 14351330 14351331 14351332 14351333 14351336 14351339 14351340 14351341 14351344 14351345 14351346 14351347 14351350 14351352 14351353 14351354 14351355 14351357 14351358 14351361 14351363 14351364 14351365 14351366 14351367 14351369 14351370 14351372 14351373 14351374 14351375 14351380 14351381 14351382 14351384 14351386 14351387 14351388 14351390 14351393 14351394 14351396 14351399 14351400 14351401 14351402 14351404 14351405 14351406 14351408 14351409 14351410 14351412 14351414 14351415 14351416 14351417 14351420 14351421 14351422 14351423 14351424 14351425 14351430 14351432 14351433 14351434 14351435 14351437 14351439 14351441 14351442 14351443 14351444 14351445 14351446 14351447 14351449 14351450 14351451 14351452 14351453 14351455 14351456 14351457 14351458 14351459 14351460 14351461 14351463 14351464 14351465 14351466 14351467 14351468 14351471 14351473 14351476 14351477 14351480 14351481 14351482 14351483 14351485 14351488 14351489 14351491 14351492 14351493 14351496 14351498 14351500 14351501 14351502 14351504 14351505 14351506 14351507 14351509 14351510 14351512 14351513 14351514 14351515 14351518 14351519 14351521 14351522 14351523 14351525 14351526 14351530 14351533 14351536 14351537 14351540 14351541 14351542 14351544 14351545 14351547 14351548 14351552 14351553 14351554 14351555 14351556 14351557 14351558 14351559 14351560 14351562 14351563 14351564 14351565 14351566 14351567 14351568 14351569 14351570 14351571 14351572 14351573 14351578 14351581 14351583 14351584 14351585 14351586 14351588 14351589 14351591 14351592 14351593 14351595 14351596 14351598 14351600 14351602 14351603 14351604 14351605 14351609 14351610 14351614 14351617 14351618 14351619 14351620 14351622 14351623 14351626 14351627 14351630 14351631 14351632 14351633 14351634 14351635 14351637 14351638 14351639 14351640 14351643 14351644 14351645 14351648 14351652 14351653 14351654 14351655 14351656 14351660 14351662 14351664 14351665 14351667 14351670 14351671 14351672 14351675 14351677 14351682 14351683 14351685 14351689 14351690 14351693 14351696 14351697 14351698 14351699 14351703 14351707 14351711 14351712 14351713 14351715 14351717 14351718 14351719 14351720 14351724 14351725 14351727 14351728 14351729 14351731 14351736 14351737 14351739 14351740 14351741 14351742 14351744 14351746 14351747 14351748 14351749 14351750 14351751 14351752 14351753 14351754 14351755 14351756 14351757 14351759 14351760 14351761 14351762 14351763 14351766 14351767 14351769 14351772 14351774 14351775 14351776 14351777 14351778 14351781 14351782 14351783 14351784 14351785 14351787 14351788 14351789 14351790 14351791 14351793 14351796 14351797 14351799 14351801 14351804 14351805 14351806 14351807 14351808 14351810 14351812 14351814 14351816 14351818 14351819 14351820 14351821 14351822 14351825 14351826 14351827 14351828 14351829 14351831 14351833 14351836 14351837 14351838 14351840 14351841 14351842 14351843 14351844 14351845 14351847 14351850 14351851 14351852 14351853 14351854 14351856 14351857 14351861 14351862 14351863 14351864 14351865 14351866 14351867 14351868 14351869 14351870 14351872 14351873 14351875 14351877 14351878 14351879 14351880 14351881 14351882 14351883 14351884 14351885 14351886 14351887 14351889 14351890 14351892 14351894 14351897 14351898 14351899 14351900 14351901 14351903 14351904 14351905 14351906 14351908 14351909 14351910 14351911 14351912 14351913 14351915 14351917 14351918 14351919 14351920 14351922 14351923 14351924 14351926 14351928 14351929 14351930 14351931 14351933 14351935 14351936 14351937 14351939 14351941 14351942 14351943 14351946 14351947 14351949 14351953 14351955 14351957 14351958 14351964 14351968 14351969 14351971 14351973 14351974 14351976 14351980 14351981 14351982 14351984 14351986 14351989 14351991 14351994 14351995 14351996 14351997 14351998 14351999 14352000 14352001 14352003 14352005 14352006 14352008 14352009 14352010 14352011 14352012 14352013 14352015 14352017 14352023 14352028 14352029 14352030 14352033 14352034 14352035 14352036 14352039 14352041 14352045 14352047 14352048 14352050 14352051 14352052 14352054 14352055 14352056 14352058 14352059 14352060 14352066 14352067 14352072 14352075 14352076 14352077 14352078 14352079 14352080 14352081 14352082 14352085 14352086 14352088 14352091 14352093 14352095 14352096 14352098 14352101 14352102 14352103 14352104 14352105 14352110 14352111 14352113 14352114 14352115 14352116 14352117 14352118 14352120 14352121 14352122 14352123 14352124 14352125 14352126 14352127 14352128 14352129 14352130 14352131 14352132 14352133 14352135 14352136 14352138 14352139 14352141 14352142 14352143 14352144 14352145 14352146 14352147 14352148 14352149 14352150 14352151 14352152 14352153 14352154 14352155 14352158 14352159 14352161 14352162 14352164 14352165 14352166 14352167 14352168 14352169 14352170 14352171 14352172 14352173 14352174 14352175 14352177 14352178 14352179 14352181 14352182 14352183 14352185 14352187 14352188 14352189 14352190 14352191 14352193 14352194 14352196 14352197 14352198 14352199 14352201 14352202 14352203 14352204 14352205 14352206 14352210 14352214 14352217 14352218 14352222 14352223 14352225 14352226 14352227 14352230 14352231 14352232 14352234 14352235 14352236 14352237 14352238 14352239 14352240 14352241 14352242 14352243 14352244 14352246 14352247 14352248 14352249 14352255 14352257 14352258 14352259 14352260 14352262 14352263 14352264 14352265 14352266 14352267 14352268 14352269 14352270 14352271 14352272 14352273 14352274 14352275 14352276 14352277 14352281 14352282 14352284 14352285 14352286 14352287 14352288 14352291 14352293 14352294 14352295 14352296 14352298 14352299 14352301 14352302 14352305 14352306 14352307 14352309 14352312 14352313 14352314 14352316 14352317 14352318 14352320 14352321 14352322 14352326 14352327 14352330 14352331 14352332 14352336 14352337 14352338 14352339 14352340 14352341 14352342 14352343 14352345 14352346 14352347 14352348 14352349 14352350 14352351 14352352 14352355 14352358 14352359 14352360 14352361 14352362 14352364 14352366 14352367 14352369 14352371 14352373 14352374 14352375 14352376 14352377 14352378 14352380 14352382 14352383 14352385 14352386 14352387 14352389 14352392 14352395 14352397 14352398 14352399 14352400 14352402 14352403 14352404 14352405 14352406 14352407 14352409 14352410 14352412 14352413 14352414 14352415 14352416 14352417 14352419 14352421 14352424 14352427 14352429 14352430 14352432 14352433 14352434 14352437 14352438 14352439 14352440 14352441 14352442 14352448 14352449 14352450 14352451 14352452 14352453 14352454 14352455 14352456 14352459 14352460 14352461 14352462 14352464 14352465 14352466 14352467 14352468 14352469 14352470 14352471 14352472 14352475 14352476 14352477 14352478 14352481 14352482 14352485 14352486 14352487 14352488 14352489 14352490 14352493 14352494 14352495 14352496 14352497 14352499 14352503 14352504 14352505 14352507 14352508 14352509 14352510 14352511 14352512 14352513 14352514 14352516 14352517 14352518 14352519 14352521 14352523 14352524 14352525 14352527 14352528 14352530 14352531 14352533 14352534 14352535 14352536 14352537 14352538 14352539 14352541 14352542 14352544 14352546 14352549 14352550 14352551 14352552 14352553 14352556 14352557 14352558 14352559 14352560 14352563 14352564 14352565 14352566 14352567 14352568 14352569 14352570 14352571 14352572 14352573 14352574 14352575 14352576 14352577 14352578 14352579 14352580 14352581 14352582 14352583 14352585 14352587 14352588 14352589 14352590 14352591 14352592 14352593 14352594 14352595 14352596 14352600 14352601 14352602 14352603 14352604 14352606 14352607 14352608 14352609 14352610 14352611 14352612 14352613 14352614 14352615 14352617 14352621 14352622 14352623 14352625 14352626 14352628 14352629 14352630 14352631 14352633 14352638 14352640 14352641 14352642 14352643 14352644 14352645 14352647 14352652 14352653 14352654 14352655 14352656 14352657 14352658 14352659 14352660 14352663 14352665 14352667 14352668 14352669 14352670 14352671 14352672 14352673 14352674 14352675 14352676 14352677 14352678 14352680 14352681 14352683 14352684 14352685 14352686 14352687 14352688 14352689 14352690 14352691 14352693 14352695 14352698 14352699 14352701 14352702 14352703 14352704 14352705 14352706 14352707 14352708 14352710 14352711 14352712 14352713 14352715 14352716 14352717 14352719 14352720 14352725 14352726 14352727 14352728 14352729 14352730 14352731 14352735 14352737 14352738 14352740 14352741 14352744 14352745 14352746 14352749 14352750 14352751 14352752 14352755 14352756 14352757 14352758 14352761 14352762 14352763 14352764 14352765 14352766 14352768 14352769 14352770 14352771 14352772 14352773 14352776 14352778 14352779 14352781 14352782 14352783 14352784 14352785 14352787 14352788 14352789 14352790 14352791 14352792 14352793 14352794 14352795 14352796 14352798 14352799 14352803 14352804 14352805 14352807 14352808 14352809 14352810 14352811 14352813 14352814 14352815 14352816 14352818 14352819 14352821 14352822 14352824 14352826 14352828 14352829 14352830 14352831 14352832 14352833 14352834 14352836 14352838 14352839 14352840 14352841 14352842 14352843 14352844 14352845 14352847 14352848 14352850 14352851 14352852 14352853 14352854 14352855 14352864 14352865 14352866 14352867 14352868 14352870 14352872 14352873 14352874 14352875 14352876 14352878 14352880 14352881 14352882 14352883 14352884 14352885 14352886 14352887 14352888 14352889 14352891 14352892 14352893 14352895 14352896 14352897 14352898 14352899 14352900 14352901 14352902 14352903 14352904 14352906 14352907 14352908 14352909 14352910 14352911 14352912 14352913 14352914 14352915 14352916 14352917 14352918 14352919 14352920 14352922 14352923 14352925 14352926 14352927 14352928 14352929 14352930 14352932 14352933 14352935 14352936 14352938 14352939 14352940 14352941 14352942 14352944 14352946 14352947 14352948 14352949 14352950 14352952 14352953 14352954 14352955 14352956 14352957 14352958 14352960 14352961 14352962 14352963 14352964 14352965 14352966 14352967 14352968 14352969 14352970 14352972 14352973 14352974 14352975 14352976 14352977 14352978 14352980 14352982 14352983 14352984 14352986 14352987 14352988 14352989 14352990 14352991 14352992 14352994 14352995 14352997 14352998 14353000 14353001 14353002 14353004 14353005 14353006 14353007 14353009 14353012 14353013 14353014 14353015 14353016 14353017 14353018 14353019 14353020 14353021 14353022 14353024 14353025 14353026 14353027 14353028 14353029 14353030 14353031 14353032 14353033 14353034 14353035 14353036 14353037 14353038 14353039 14353040 14353042 14353043 14353044 14353045 14353047 14353048 14353049 14353051 14353052 14353054 14353055 14353057 14353058 14353059 14353060 14353061 14353062 14353064 14353066 14353067 14353068 14353069 14353070 14353071 14353072 14353073 14353074 14353075 14353076 14353077 14353078 14353079 14353080 14353081 14353082 14353083 14353084 14353085 14353086 14353087 14353088 14353089 14353090 14353091 14353092 14353094 14353095 14353096 14353097 14353098 14353099 14353100 14353101 14353102 14353103 14353104 14353105 14353106 14353107 14353108 14353109 14353110 14353112 14353113 14353114 14353115 14353116 14353117 14353118 14353120 14353123 14353124 14353127 14353128 14353129 14353130 14353131 14353135 14353136 14353137 14353138 14353139 14353140 14353141 14353142 14353143 14353145 14353146 14353149 14353150 14353151 14353152 14353153 14353155 14353156 14353157 14353158 14353159 14353160 14353161 14353162 14353164 14353165 14353167 14353168 14353169 14353170 14353172 14353173 14353174 14353175 14353176 14353178 14353179 14353180 14353181 14353182 14353183 14353184 14353185 14353186 14353187 14353188 14353189 14353190 14353191 14353192 14353193 14353194 14353195 14353196 14353198 14353199 14353200 14353201 14353202 14353203 14353204 14353205 14353207 14353208 14353210 14353211 14353212 14353213 14353215 14353216 14353217 14353218 14353220 14353221 14353222 14353223 14353225 14353226 14353227 14353228 14353229 14353230 14353231 14353233 14353234 14353235 14353236 14353238 14353239 14353240 14353244 14353245 14353246 14353247 14353248 14353249 14353250 14353251 14353252 14353253 14353254 14353255 14353256 14353257 14353258 14353259 14353260 14353261 14353262 14353263 14353264 14353265 14353267 14353268 14353269 14353272 14353273 14353274 14353275 14353276 14353278 14353279 14353280 14353282 14353283 14353284 14353285 14353286 14353287 14353289 14353292 14353294 14353295 14353296 14353297 14353298 14353300 14353302 14353303 14353304 14353305 14353306 14353307 14353309 14353311 14353312 14353313 14353314 14353316 14353317 14353318 14353319 14353320 14353321 14353322 14353323 14353325 14353326 14353327 14353328 14353329 14353330 14353331 14353332 14353334 14353335 14353336 14353337 14353339 14353340 14353341 14353344 14353345 14353346 14353347 14353348 14353349 14353350 14353351 14353352 14353353 14353354 14353357 14353359 14353360 14353361 14353362 14353363 14353364 14353365 14353366 14353367 14353369 14353370 14353373 14353374 14353375 14353376 14353377 14353379 14353381 14353382 14353383 14353384 14353386 14353387 14353388 14353389 14353390 14353391 14353393 14353394 14353395 14353396 14353397 14353398 14353399 14353400 14353401 14353402 14353403 14353404 14353405 14353406 14353408 14353409 14353410 14353411 14353413 14353414 14353415 14353417 14353418 14353421 14353423 14353424 14353425 14353426 14353427 14353428 14353429 14353430 14353431 14353432 14353433 14353434 14353435 14353436 14353437 14353439 14353440 14353441 14353444 14353445 14353448 14353449 14353450 14353452 14353453 14353455 14353456 14353457 14353458 14353459 14353460 14353462 14353463 14353464 14353465 14353466 14353467 14353469 14353470 14353471 14353472 14353473 14353474 14353475 14353477 14353478 14353481 14353482 14353483 14353486 14353487 14353488 14353489 14353490 14353491 14353492 14353493 14353494 14353495 14353497 14353498 14353499 14353500 14353501 14353502 14353503 14353504 14353505 14353506 14353507 14353510 14353511 14353514 14353515 14353516 14353517 14353518 14353519 14353522 14353523 14353525 14353526 14353528 14353529 14353530 14353531 14353534 14353535 14353536 14353537 14353539 14353540 14353541 14353543 14353544 14353545 14353546 14353548 14353550 14353552 14353553 14353554 14353555 14353558 14353559 14353561 14353563 14353564 14353566 14353567 14353568 14353569 14353570 14353571 14353572 14353573 14353574 14353575 14353576 14353577 14353578 14353579 14353582 14353583 14353585 14353586 14353587 14353588 14353590 14353592 14353593 14353595 14353596 14353597 14353599 14353600 14353601 14353603 14353604 14353605 14353607 14353609 14353610 14353611 14353612 14353613 14353614 14353615 14353617 14353619 14353620 14353622 14353623 14353624 14353625 14353626 14353630 14353631 14353632 14353633 14353635 14353636 14353637 14353638 14353639 14353640 14353641 14353642 14353643 14353645 14353646 14353647 14353648 14353649 14353650 14353652 14353653 14353654 14353656 14353657 14353658 14353659 14353660 14353661 14353665 14353666 14353667 14353669 14353670 14353671 14353673 14353675 14353676 14353678 14353679 14353680 14353681 14353682 14353684 14353685 14353687 14353689 14353690 14353692 14353693 14353694 14353697 14353698 14353699 14353700 14353701 14353702 14353707 14353708 14353709 14353715 14353716 14353717 14353718 14353720 14353721 14353722 14353725 14353726 14353728 14353730 14353731 14353732 14353734 14353735 14353736 14353737 14353738 14353741 14353742 14353744 14353745 14353747 14353748 14353749 14353752 14353753 14353754 14353757 14353758 14353759 14353760 14353761 14353762 14353763 14353764 14353765 14353766 14353767 14353768 14353769 14353770 14353772 14353773 14353774 14353775 14353778 14353782 14353783 14353784 14353785 14353787 14353788 14353789 14353792 14353793 14353794 14353799 14353800 14353801 14353802 14353803 14353805 14353809 14353810 14353811 14353812 14353813 14353814 14353816 14353817 14353819 14353821 14353823 14353824 14353825 14353826 14353828 14353829 14353831 14353833 14353835 14353837 14353838 14353839 14353840 14353841 14353845 14353847 14353848 14353849 14353850 14353851 14353852 14353853 14353854 14353856 14353857 14353858 14353860 14353861 14353862 14353864 14353866 14353868 14353869 14353871 14353874 14353875 14353876 14353879 14353880 14353881 14353883 14353884 14353885 14353888 14353891 14353892 14353893 14353895 14353896 14353897 14353898 14353901 14353902 14353904 14353905 14353906 14353907 14353908 14353909 14353912 14353913 14353914 14353915 14353916 14353917 14353918 14353919 14353921 14353922 14353923 14353924 14353925 14353926 14353927 14353928 14353929 14353930 14353931 14353933 14353934 14353935 14353936 14353937 14353939 14353940 14353941 14353942 14353943 14353948 14353949 14353950 14353954 14353956 14353958 14353959 14353960 14353961 14353963 14353965 14353967 14353969 14353970 14353971 14353972 14353974 14353975 14353977 14353979 14353980 14353981 14353982 14353983 14353985 14353988 14353989 14353990 14353991 14353992 14353993 14353996 14353997 14353998 14354000 14354001 14354003 14354004 14354005 14354006 14354007 14354008 14354009 14354010 14354011 14354012 14354013 14354015 14354016 14354018 14354019 14354021 14354022 14354023 14354024 14354026 14354028 14354030 14354032 14354033 14354034 14354035 14354036 14354037 14354038 14354040 14354041 14354042 14354043 14354044 14354045 14354046 14354047 14354051 14354054 14354055 14354057 14354058 14354059 14354060 14354061 14354065 14354066 14354067 14354068 14354069 14354070 14354072 14354073 14354075 14354076 14354077 14354078 14354079 14354081 14354082 14354083 14354084 14354089 14354090 14354091 14354092 14354093 14354094 14354095 14354096 14354097 14354098 14354099 14354100 14354101 14354102 14354104 14354105 14354106 14354107 14354109 14354112 14354114 14354116 14354118 14354120 14354121 14354125 14354127 14354129 14354130 14354132 14354133 14354134 14354135 14354136 14354137 14354140 14354143 14354146 14354148 14354150 14354151 14354153 14354154 14354155 14354158 14354159 14354160 14354161 14354164 14354165 14354168 14354170 14354173 14354174 14354177 14354178 14354179 14354180 14354181 14354182 14354190 14354192 14354193 14354194 14354195 14354198 14354199 14354200 14354202 14354206 14354210 14354212 14354213 14354215 14354216 14354217 14354218 14354219 14354220 14354222 14354223 14354224 14354225 14354229 14354230 14354231 14354232 14354233 14354234 14354235 14354236 14354238 14354239 14354240 14354241 14354242 14354243 14354244 14354245 14354247 14354248 14354249 14354251 14354253 14354256 14354257 14354259 14354260 14354261 14354262 14354264 14354265 14354267 14354268 14354269 14354271 14354274 14354275 14354276 14354277 14354278 14354279 14354280 14354281 14354283 14354285 14354287 14354289 14354291 14354292 14354294 14354300 14354301 14354304 14354510 14354511 14354512 14354513 14354514 14354515 14354516 14354517 14354518 14354519 14354521 14354523 14354524 14354525 14354526 14354528 14354529 14354530 14354531 14354532 14354533 14354534 14354537 14354538 14354539 14354540 14354541 14354542 14354543 14354544 14354545 14354546 14354547 14354548 14354549 14354550 14354551 14354552 14354553 14354554 14354555 14354556 14354557 14354558 14354559 14354560 14354561 14354563 14354565 14354566 14354567 14354568 14354569 14354571 14354573 14354574 14354575 14354576 14354578 14354579 14354580 14354581 14354583 14354584 14354585 14354586 14354587 14354588 14354589 14354590 14354591 14354593 14354594 14354595 14354597 14354598 14354599 14354601 14354602 14354603 14354605 14354606 14354607 14354609 14354610 14354612 14354613 14354614 14354615 14354616 14354617 14354618 14354619 14354620 14354621 14354622 14354623 14354625 14354626 14354627 14354628 14354630 14354631 14354632 14354633 14354634 14354635 14354639 14354640 14354643 14354644 14354645 14354647 14354648 14354649 14354650 14354652 14354653 14354654 14354655 14354656 14354657 14354658 14354659 14354660 14354661 14354662 14354663 14354664 14354665 14354666 14354667 14354668 14354669 14354670 14354671 14354672 14354674 14354676 14354678 14354680 14354681 14354682 14354683 14354684 14354685 14354687 14354688 14354689 14354690 14354691 14354692 14354693 14354694 14354695 14354696 14354697 14354698 14354699 14354700 14354701 14354702 14354704 14354705 14354707 14354708 14354710 14354712 14354713 14354714 14354717 14354718 14354720 14354721 14354722 14354724 14354726 14354728 14354729 14354730 14354731 14354732 14354733 14354734 14354735 14354737 14354738 14354739 14354740 14354741 14354742 14354743 14354746 14354747 14354749 14354750 14354751 14354752 14354753 14354754 14354755 14354756 14354757 14354758 14354761 14354762 14354763 14354764 14354765 14354766 14354767 14354768 14354769 14354771 14354772 14354773 14354774 14354775 14354776 14354777 14354778 14354779 14354780 14354782 14354783 14354784 14354785 14354786 14354787 14354788 14354789 14354791 14354792 14354796 14354798 14354799 14354800 14354801 14354802 14354804 14354805 14354806 14354807 14354808 14354810 14354811 14354812 14354813 14354814 14354815 14354817 14354818 14354819 14354820 14354821 14354822 14354825 14354826 14354827 14354829 14354832 14354833 14354834 14354835 14354836 14354837 14354838 14354840 14354841 14354843 14354844 14354845 14354846 14354847 14354848 14354849 14354852 14354853 14354854 14354855 14354856 14354857 14354858 14354859 14354860 14354861 14354862 14354863 14354864 14354867 14354868 14354869 14354870 14354871 14354874 14354875 14354876 14354877 14354878 14354879 14354880 14354882 14354883 14354885 14354886 14354887 14354888 14354889 14354890 14354892 14354893 14354894 14354895 14354896 14354897 14354898 14354900 14354901 14354902 14354903 14354904 14354905 14354907 14354908 14354909 14354910 14354911 14354912 14354913 14354914 14354915 14354916 14354917 14354918 14354920 14354921 14354922 14354923 14354924 14354925 14354927 14354928 14354929 14354930 14354931 14354932 14354933 14354934 14354936 14354938 14354939 14354940 14354943 14354944 14354945 14354947 14354948 14354949 14354951 14354952 14354954 14354955 14354956 14354957 14354959 14354960 14354962 14354963 14354964 14354965 14354966 14354968 14354969 14354970 14354971 14354972 14354973 14354974 14354976 14354978 14354979 14354981 14354982 14354983 14354985 14354986 14354987 14354988 14354989 14354990 14354991 14354992 14354993 14354995 14354996 14354997 14354998 14354999 14355000 14356421 14356422 14356424 14356425 14356426 14356429 14356430 14356432 14356433 14356434 14356435 14356440 14356444 14356445 14356447 14356448 14356450 14356451 14356453 14356455 14356456 14356458 14356459 14356460 14356461 14356462 14356463 14356465 14356467 14356469 14356471 14356472 14356474 14356475 14356476 14356478 14356480 14356484 14356485 14356486 14356487 14356489 14356490 14356491 14356492 14356493 14356495 14356496 14356499 14356500 14356501 14356502 14356503 14356506 14356507 14356508 14356509 14356510 14356511 14356513 14356514 14356515 14356517 14356518 14356519 14356522 14356525 14356526 14356527 14356528 14356530 14356531 14356532 14356533 14356535 14356536 14356537 14356538 14356540 14356541 14356542 14356543 14356544 14356545 14356547 14356548 14356549 14356550 14356551 14356553 14356554 14356555 14356556 14356559 14356560 14356561 14356562 14356563 14356564 14356565 14356566 14356567 14356568 14356569 14356572 14356573 14356574 14356579 14356580 14356581 14356582 14356583 14356584 14356585 14356587 14356588 14356590 14356591 14356592 14356593 14356594 14356595 14356596 14356597 14356598 14356599 14356600 14356602 14356603 14356605 14356606 14356607 14356608 14356609 14356610 14356612 14356614 14356617 14356618 14356619 14356620 14356621 14356622 14356623 14356624 14356625 14356626 14356627 14356628 14356629 14356630 14356632 14356634 14356635 14356636 14356637 14356638 14356639 14356640 14356641 14356642 14356643 14356644 14356645 14356646 14356647 14356648 14356649 14356650 14356651 14356652 14356653 14356654 14356655 14356656 14356657 14356658 14356659 14356660 14356661 14356662 14356663 14356664 14356665 14356666 14356667 14356668 14356669 14356671 14356672 14356673 14356674 14356675 14356676 14356677 14356678 14356679 14356680 14356681 14356682 14356683 14356684 14356685 14356686 14356687 14356688 14356689 14356690 14356691 14356692 14356693 14356694 14356695 14356696 14356697 14356698 14356699 14356700 14356701 14356702 14356703 14356704 14356705 14356706 14356707 14356708 14356709 14356710 14356711 14356712 14356713 14356714 14356715 14356716 14356717 14356718 14356719 14356720 14356721 14356722 14356723 14356724 14356725 14356727 14356728 14356729 14356730 14356731 14356732 14356733 14356734 14356735 14356736 14356737 14356738 14356739 14356740 14356741 14356742 14356743 14356744 14356745 14356746 14356747 14356748 14356749 14356750 14356751 14356752 14356753 14356754 14356755 14356756 14356757 14356758 14356759 14356760 14356761 14356762 14356763 14356764 14356765 14356766 14356767 14356768 14356769 14356770 14356771 14356772 14356773 14356774 14356775 14356776 14356777 14356778 14356779 14356780 14356781 14356782 14356783 14356784 14356785 14356786 14356787 14356788 14356789 14356790 14356791 14356792 14356793 14356794 14356795 14356796 14356797 14356798 14356799 14356800 14356801 14356802 14356803 14356804 14356805 14356806 14356807 14356808 14356809 14356810 14356811 14356812 14356813 14356814 14356815 14356816 14356817 14356818 14356819 14356820 14356821 14356822 14356823 14356824 14356825 14356826 14356827 14356828 14356829 14356830 14356831 14356832 14356833 14356834 14356835 14356836 14356837 14356838 14356839 14356840 14356841 14356842 14356843 14356845 14356846 14356847 14356848 14356849 14356850 14356851 14356852 14356853 14356854 14356855 14356856 14356857 14356858 14356859 14356860 14356861 14356862 14356863 14356864 14356865 14356866 14356867 14356868 14356869 14356870 14356871 14356872 14356873 14356874 14356875 14356876 14356877 14356878 14356879 14356880 14356881 14356882 14356883 14356885 14356886 14356887 14356888 14356889 14356890 14356891 14356892 14356893 14356894 14356895 14356896 14356897 14356898 14356899 14356900 14356901 14356903 14356904 14356905 14356906 14356907 14356908 14356909 14356910 14356911 14356912 14356913 14356914 14356915 14356916 14356917 14356918 14356919 14356920 14356921 14356923 14356924 14356925 14356926 14356927 14356928 14356929 14356930 14356931 14356932 14356933 14356934 14356935 14356937 14356938 14356940 14356941 14356942 14356943 14356944 14356945 14356946 14356947 14356948 14356949 14356952 14356953 14356954 14356955 14356957 14356958 14356959 14356960 14356961 14356962 14356963 14356964 14356965 14356966 14356967 14356968 14356969 14356970 14356971 14356972 14356973 14356974 14356975 14356976 14356977 14356978 14356979 14356980 14356981 14356982 14356983 14356985 14356986 14356987 14356988 14356989 14356990 14356991 14356992 14356993 14356994 14356995 14356996 14356997 14356998 14356999 14357000 14357001 14357002 14357003 14357005 14357006 14357007 14357008 14357009 14357010 14357011 14357013 14357014 14357017 14357018 14357019 14357020 14357021 14357022 14357023 14357024 14357025 14357026 14357027 14357028 14357029 14357030 14357032 14357034 14357036 14357037 14357038 14357039 14357040 14357041 14357042 14357043 14357044 14357045 14357046 14357048 14357049 14357050 14357051 14357052 14357054 14357055 14357056 14357058 14357060 14357061 14357062 14357063 14357066 14357067 14357068 14357069 14357070 14357071 14357074 14357077 14357079 14357082 14357084 14357085 14357086 14357087 14357089 14357090 14357091 14357092 14357093 14357094 14357095 14357096 14357097 14357098 14357099 14357100 14357101 14357102 14357103 14357104 14357105 14357106 14357107 14357108 14357110 14357111 14357112 14357113 14357114 14357115 14357116 14357118 14357120 14357121 14357122 14357123 14357124 14357126 14357128 14357129 14357130 14357131 14357132 14357134 14357135 14357136 14357137 14357138 14357139 14357141 14357142 14357143 14357144 14357145 14357146 14357147 14357148 14357149 14357151 14357152 14357154 14357156 14357157 14357158 14357159 14357160 14357162 14357163 14357164 14357165 14357166 14357167 14357168 14357169 14357171 14357172 14357175 14357176 14357177 14357178 14357179 14357182 14357183 14357184 14357186 14357187 14357189 14357190 14357191 14357192 14357193 14357194 14357195 14357196 14357198 14357199 14357200 14357201 14357202 14357203 14357205 14357206 14357207 14357209 14357210 14357211 14357212 14357213 14357214 14357215 14357216 14357217 14357218 14357220 14357222 14357223 14357224 14357225 14357228 14357230 14357232 14357234 14357235 14357237 14357240 14357241 14357242 14357243 14357244 14357245 14357246 14357248 14357249 14357250 14357251 14357252 14357253 14357254 14357256 14357258 14357259 14357260 14357261 14357264 14357265 14357266 14357267 14357269 14357270 14357271 14357274 14357275 14357276 14357277 14357279 14357281 14357282 14357283 14357284 14357285 14357287 14357288 14357289 14357290 14357291 14357292 14357294 14357295 14357296 14357297 14357298 14357299 14357300 14357302 14357303 14357305 14357306 14357307 14357308 14357309 14357311 14357312 14357313 14357314 14357316 14357319 14357320 14357321 14357322 14357324 14357327 14357330 14357331 14357334 14357335 14357337 14357341 14357343 14357345 14357346 14357347 14357349 14357350 14357352 14357353 14357354 14357355 14357356 14357357 14357359 14357360 14357361 14357363 14357364 14357365 14357366 14357368 14357369 14357372 14357373 14357374 14357375 14357376 14357377 14357378 14357379 14357380 14357381 14357382 14357384 14357385 14357386 14357387 14357388 14357389 14357391 14357392 14357393 14357394 14357395 14357396 14357397 14357398 14357400 14357401 14357402 14357403 14357404 14357406 14357407 14357408 14357410 14357411 14357412 14357413 14357414 14357415 14357417 14357418 14357419 14357420 14357421 14357422 14357425 14357426 14357428 14357429 14357433 14357436 14357438 14357440 14357441 14357443 14357444 14357448 14357449 14357450 14357451 14357452 14357454 14357455 14357456 14357458 14357459 14357460 14357461 14357462 14357463 14357465 14357466 14357467 14357468 14357469 14357472 14357474 14357475 14357477 14357478 14357479 14357480 14357481 14357483 14357484 14357485 14357486 14357487 14357488 14357489 14357490 14357491 14357492 14357494 14357495 14357496 14357497 14357498 14357501 14357506 14357508 14357509 14357510 14357511 14357512 14357513 14357515 14357516 14357518 14357519 14357520 14357521 14357523 14357524 14357526 14357527 14357529 14357530 14357532 14357533 14357535 14357536 14357537 14357538 14357539 14357540 14357541 14357542 14357545 14357546 14357547 14357548 14357549 14357551 14357553 14357554 14357557 14357559 14357561 14357562 14357564 14357565 14357566 14357567 14357568 14357569 14357570 14357571 14357573 14357574 14357575 14357576 14357577 14357579 14357580 14357581 14357582 14357583 14357584 14357585 14357587 14357588 14357590 14357591 14357592 14357593 14357595 14357596 14357597 14357598 14357599 14357600 14357601 14357602 14357603 14357604 14357605 14357606 14357608 14357609 14357611 14357612 14357613 14357614 14357615 14357617 14357618 14357620 14357621 14357622 14357623 14357624 14357625 14357626 14357627 14357628 14357629 14357630 14357631 14357632 14357633 14357634 14357635 14357636 14357637 14357638 14357639 14357640 14357641 14357642 14357643 14357644 14357645 14357646 14357647 14357648 14357649 14357650 14357651 14357652 14357653 14357654 14357655 14357656 14357657 14357658 14357659 14357660 14357661 14357662 14357664 14357665 14357666 14357667 14357668 14357669 14357670 14357671 14357672 14357674 14357675 14357676 14357677 14357678 14357679 14357680 14357681 14357682 14357683 14357684 14357685 14357686 14357687 14357688 14357689 14357690 14357691 14357692 14357693 14357695 14357697 14357698 14357699 14357701 14357702 14357703 14357704 14357705 14357706 14357707 14357710 14357711 14357714 14357715 14357717 14357718 14357719 14357720 14357721 14357722 14357723 14357724 14357725 14357726 14357727 14357728 14357729 14357730 14357731 14357733 14357734 14357735 14357736 14357737 14357738 14357741 14357742 14357743 14357744 14357745 14357746 14357747 14357749 14357750 14357751 14357752 14357753 14357754 14357755 14357756 14357757 14357758 14357760 14357761 14357762 14357763 14357764 14357766 14357767 14357768 14357769 14357770 14357771 14357774 14357775 14357776 14357778 14357779 14357781 14357785 14357786 14357787 14357788 14357789 14357790 14357791 14357792 14357793 14357794 14357795 14357796 14357797 14357798 14357799 14357800 14357801 14357802 14357803 14357804 14357806 14357807 14357808 14357809 14357810 14357812 14357814 14357815 14357816 14357817 14357818 14357819 14357821 14357822 14357823 14357824 14357825 14357826 14357827 14357828 14357829 14357832 14357833 14357834 14357835 14357839 14357840 14357842 14357843 14357847 14357849 14357851 14357853 14357857 14357858 14357859 14357860 14357861 14357862 14357863 14357864 14357865 14357866 14357868 14357869 14357870 14357871 14357872 14357873 14357874 14357876 14357877 14357878 14357879 14357880 14357881 14357882 14357883 14357884 14357886 14357887 14357888 14357889 14357890 14357891 14357893 14357894 14357895 14357896 14357898 14357899 14357900 14357903 14357904 14357905 14357907 14357911 14357912 14357917 14357923 14357924 14357926 14357927 14357929 14357931 14357932 14357933 14357934 14357935 14357936 14357937 14357939 14357940 14357941 14357942 14357943 14357944 14357945 14357946 14357947 14357949 14357950 14357951 14357952 14357953 14357954 14357955 14357956 14357957 14357959 14357960 14357962 14357963 14357964 14357965 14357966 14357968 14357969 14357970 14357971 14357972 14357973 14357974 14357975 14357976 14357977 14357978 14357979 14357980 14357981 14357982 14357983 14357985 14357987 14357989 14357990 14357991 14357992 14357994 14357995 14357996 14357997 14357998 14358000 14358001 14358002 14358003 14358005 14358006 14358007 14358008 14358009 14358010 14358011 14358012 14358013 14358014 14358015 14358016 14358017 14358018 14358019 14358020 14358021 14358024 14358026 14358027 14358028 14358029 14358030 14358031 14358032 14358033 14358034 14358036 14358037 14358038 14358040 14358041 14358042 14358043 14358044 14358045 14358046 14358047 14358048 14358049 14358050 14358051 14358053 14358054 14358056 14358057 14358058 14358059 14358061 14358064 14358066 14358067 14358068 14358070 14358074 14358075 14358076 14358077 14358079 14358080 14358081 14358083 14358084 14358085 14358086 14358087 14358088 14358089 14358090 14358092 14358094 14358096 14358097 14358099 14358101 14358102 14358103 14358104 14358107 14358108 14358109 14358110 14358112 14358115 14358116 14358117 14358118 14358119 14358120 14358121 14358122 14358124 14358125 14358126 14358127 14358128 14358129 14358130 14358131 14358132 14358133 14358134 14358135 14358137 14358138 14358139 14358140 14358141 14358142 14358143 14358144 14358145 14358146 14358147 14358148 14358149 14358151 14358152 14358154 14358156 14358158 14358159 14358161 14358162 14358163 14358164 14358165 14358166 14358168 14358169 14358170 14358171 14358172 14358173 14358174 14358175 14358176 14358177 14358178 14358180 14358181 14358182 14358183 14358184 14358185 14358186 14358187 14358188 14358189 14358191 14358192 14358193 14358194 14358195 14358196 14358197 14358198 14358199 14358200 14358201 14358202 14358203 14358204 14358205 14358206 14358207 14358208 14358209 14358211 14358212 14358213 14358214 14358215 14358216 14358217 14358218 14358219 14358220 14358221 14358222 14358223 14358224 14358225 14358226 14358227 14358228 14358229 14358230 14358231 14358232 14358233 14358235 14358236 14358237 14358238 14358239 14358240 14358241 14358242 14358243 14358244 14358245 14358246 14358247 14358249 14358250 14358251 14358252 14358254 14358255 14358256 14358257 14358258 14358259 14358260 14358261 14358262 14358263 14358264 14358265 14358266 14358267 14358268 14358269 14358270 14358271 14358272 14358274 14358276 14358277 14358279 14358280 14358281 14358282 14358283 14358284 14358285 14358286 14358287 14358288 14358289 14358290 14358291 14358292 14358293 14358294 14358295 14358296 14358297 14358298 14358299 14358300 14358301 14358302 14358303 14358304 14358305 14358306 14358307 14358308 14358309 14358310 14358311 14358312 14358313 14358314 14358315 14358316 14358317 14358318 14358319 14358320 14358321 14358322 14358323 14358324 14358325 14358326 14358327 14358328 14358329 14358330 14358331 14358332 14358333 14358334 14358335 14358336 14358337 14358338 14358339 14358340 14358341 14358342 14358343 14358344 14358345 14358346 14358347 14358348 14358349 14358350 14358351 14358352 14358354 14358355 14358356 14358358 14358359 14358360 14358361 14358362 14358363 14358364 14358365 14358366 14358367 14358368 14358369 14358370 14358371 14358372 14358373 14358374 14358375 14358376 14358377 14358378 14358379 14358380 14358381 14358382 14358383 14358384 14358385 14358386 14358387 14358388 14358389 14358390 14358391 14358392 14358393 14358394 14358395 14358396 14358397 14358398 14358399 14358400 14358401 14358402 14358403 14358404 14358405 14358406 14358407 14358408 14358409 14358410 14358411 14358412 14358413 14358414 14358415 14358416 14358417 14358418 14358419 14358420 14358421 14358422 14358423 14358424 14358425 14358426 14358428 14358429 14358430 14358431 14358432 14358433 14358434 14358435 14358436 14358437 14358438 14358439 14358440 14358441 14358442 14358443 14358444 14358445 14358446 14358447 14358448 14358449 14358450 14358451 14358452 14358453 14358454 14358455 14358456 14358457 14358458 14358459 14358460 14358461 14358462 14358463 14358464 14358465 14358466 14358467 14358468 16452201 16452203 16452204 16452206 16452207 16452208 16452209 16452210 16452211 16452212 16452213 16452214 16452215 16452216 16452217 16452218 16452219 16452220 16452221 16452222 16452223 16452224 16452225 16452226 16452227 16452228 16452229 16452232 16452233 16452234 16452235 16452236 16452238 16452239 16452240 16452241 16452242 16452244 16452245 16452247 16452248 16452250 16452251 16452252 16452253 16452254 16452255 16452256 16452257 16452258 16452259 16452260 16452261 16452262 16452263 16452264 16452266 16452267 16452268 16452269 16452270 16452271 16452274 16452275 16452276 16452277 16452278 16452279 16452281 16452282 16452283 16452285 16452286 16452287 16452288 16452290 16452291 16452293 16452295 16452296 16452297 16452298 16452300 16452301 16452302 16452304 16452305 16452306 16452307 16452308 16452309 16452310 16452311 16452312 16452313 16452314 16452315 16452317 16452318 16452319 16452322 16452323 16452324 16452326 16452327 16452328 16452329 16452330 16452331 16452332 16452334 16452335 16452336 16452337 16452338 16452339 16452340 16452341 16452342 16452343 16452346 16452347 16452348 16452349 16452350 16452351 16452352 16452353 16452355 16452356 16452357 16452359 16452360 16452361 16452362 16452363 16452364 16452365 16452366 16452371 16452372 16452373 16452375 16452376 16452377 16452379 16452380 16452382 16452385 16452386 16452387 16452388 16452390 16452393 16452395 16452396 16452397 16452398 16452400 16452402 16452403 16452404 16452405 16452407 16452409 16452411 16452413 16452415 16452416 16452418 16452419 16452420 16452421 16452422 16452423 16452424 16452425 16452426 16452427 16452428 16452431 16452432 16452433 16452434 16452435 16452439 16452441 16452442 16452443 16452444 16452445 16452446 16452448 16452449 16452450 16452451 16452453 16452455 16452456 16452457 16452458 16452459 16452460 16452461 16452462 16452464 16452468 16452469 16452470 16452471 16452472 16452474 16452476 16452477 16452479 16452480 16452481 16452484 16452485 16452487 16452489 16452490 16452491 16452492 16452493 16452494 16452495 16452496 16452497 16452498 16452499 16452500 16452501 16452502 16452503 16452504 16452505 16452506 16452507 16452508 16452509 16452510 16452511 16452513 16452514 16452516 16452518 16452519 16452520 16452521 16452522 16452523 16452524 16452525 16452528 16452530 16452531 16452534 16452536 16452537 16452538 16452540 16452541 16452542 16452544 16452545 16452548 16452549 16452553 16452554 16452557 16452558 16452559 16452560 16452561 16452562 16452564 16452566 16452567 16452569 16452571 16452573 16452574 16452576 16452577 16452578 16452579 16452580 16452581 16452583 16452584 16452586 16452587 16452588 16452589 16452591 16452592 16452595 16452599 16452600 16452602 16452603 16452604 16452607 16452608 16452611 16452612 16452614 16452616 16452618 16452621 16452622 16452623 16452630 16452632 16452633 16452634 16452635 16452636 16452637 16452638 16452639 16452640 16452641 16452642 16452643 16452644 16452645 16452646 16452647 16452648 16452649 16452652 16452657 16452658 16452659 16452660 16452661 16452662 16452665 16452666 16452667 16452668 16452670 16452671 16452673 16452674 16452675 16452676 16452677 16452679 16452680 16452681 16452683 16452685 16452686 16452687 16452688 16452689 16452692 16452693 16452694 16452695 16452697 16452698 16452699 16452700 16452701 16452702 16452703 16452704 16452705 16452706 16452707 16452708 16452710 16452711 16452713 16452714 16452715 16452716 16452717 16452719 16452720 16452721 16452722 16452724 16452725 16452727 16452728 16452730 16452731 16452732 16452734 16452735 16452736 16452737 16452738 16452740 16452741 16452743 16452744 16452745 16452748 16452749 16452750 16452751 16452752 16452753 16452755 16452756 16452757 16452758 16452759 16452760 16452761 16452762 16452763 16452764 16452765 16452766 16452768 16452769 16452770 16452771 16452772 16452773 16452776 16452777 16452778 16452779 16452780 16452781 16452782 16452783 16452785 16452788 16452789 16452790 16452791 16452792 16452793 16452794 16452795 16452796 16452797 16452798 16452799 16452800 16452801 16452803 16452804 16452805 16452806 16452807 16452808 16452809 16452811 16452812 16452813 16452814 16452815 16452817 16452820 16452821 16452822 16452823 16452824 16452825 16452826 16452828 16452829 16452830 16452831 16452832 16452833 16452834 16452835 16452837 16452838 16452839 16452840 16452841 16452842 16452843 16452844 16452845 16452848 16452849 16452850 16452851 16452853 16452855 16452856 16452857 16452858 16452859 16452860 16452861 16452862 16452863 16452864 16452865 16452866 16452867 16452868 16452869 16452870 16452871 16452872 16452873 16452875 16452876 16452877 16452878 16452879 16452880 16452881 16452885 16452886 16452887 16452888 16452889 16452893 16452895 16452896 16452898 16452899 16452900 16452901 16452902 16452903 16452904 16452905 16452907 16452908 16452910 16452911 16452912 16452913 16452914 16452915 16452916 16452919 16452920 16452923 16452925 16452926 16452927 16452929 16452930 16452931 16452932 16452933 16452934 16452935 16452936 16452940 16452941 16452943 16452944 16452946 16452948 16452951 16452953 16452955 16452957 16452959 16452960 16452961 16452963 16452964 16452966 16452967 16452970 16452971 16452972 16452973 16452974 16452975 16452976 16452977 16452980 16452981 16452982 16452984 16452985 16452986 16452987 16452988 16452989 16452990 16452991 16452992 16452993 16452994 16452996 16452997 16452999 16453000 16453001 16453002 16453003 16453004 16453005 16453006 16453007 16453008 16453009 16453011 16453013 16453014 16453016 16453019 16453022 16453023 16453026 16453027 16453030 16453035 16453039 16453041 16453042 16453043 16453048 16453049 16453050 16453055 16453058 16453060 16453063 16453070 16453079 16453081 16453085 16453086 16453088 16453089 16453091 16453092 16453094 16453101 16453104 16453108 16453109 16453111 16453114 16453115 16453123 16453126 16453127 16453136 16453141 16453144 16453151 16453152 16453153 16453154 16453156 16453158 16453160 16453164 16453165 16453172 16453174 16453176 16453177 16453178 16453179 16453180 16453181 16453182 16453183 16453184 16453187 16453189 16453191 16453192 16453194 16453195 16453196 16453199 16453202 16453205 16453207 16453208 16453209 16453211 16453212 16453213 16453214 16453215 16453216 16453218 16453220 16453222 16453223 16453226 16453227 16453229 16453230 16453231 16453232 16453233 16453236 16453238 16453239 16453241 16453242 16453243 16453246 16453247 16453248 16453250 16453251 16453252 16453253 16453254 16453256 16453257 16453258 16453260 16453261 16453263 16453264 16453266 16453269 16453270 16453272 16453273 16453274 16453275 16453276 16453277 16453278 16453279 16453280 16453281 16453282 16453283 16453284 16453285 16453286 16453287 16453290 16453291 16453292 16453293 16453294 16453297 16453298 16453299 16453303 16453304 16453307 16453314 16453316 16453317 16453318 16453321 16453323 16453325 16453329 16453331 16453335 16453337 16453339 16453340 16453342 16453344 16453345 16453346 16453347 16453353 16453354 16453355 16453360 16453361 16453367 16453372 16453374 16453375 16453376 16453377 16453378 16453381 16453382 16453386 16453387 16453388 16453390 16453391 16453392 16453393 16453394 16453395 16453397 16453398 16453399 16453400 16453401 16453402 16453405 16453408 16453410 16453412 16453414 16453415 16453417 16453419 16453422 16453423 16453427 16453428 16453429 16453430 16453431 16453432 16453433 16453435 16453438 16453440 16453445 16453446 16453448 16453449 16453453 16453455 16453456 16453457 16453458 16453461 16453463 16453464 16453466 16453468 16453469 16453474 16453475 16453480 16453481 16453484 16453485 16453488 16453492 16453493 16453495 16453496 16453497 16453499 16453502 16453503 16453505 16453508 16453510 16453514 16453518 16453519 16453520 16453522 16453524 16453526 16453527 16453528 16453530 16453531 16453533 16453534 16453537 16453540 16453541 16453542 16453544 16453547 16453549 16453550 16453554 16453556 16453559 16453563 16453564 16453566 16453568 16453571 16453573 16453576 16453577 16453582 16453584 16453585 16453586 16453590 16453593 16453597 16453599 16453602 16453603 16453605 16453606 16453608 16453612 16453613 16453615 16453619 16453620 16453627 16453630 16453632 16453633 16453635 16453638 16453642 16453646 16453648 16453649 16453650 16453651 16453657 16453658 16453660 16453661 16453662 16453663 16453664 16453667 16453668 16453672 16453673 16453674 16453675 16453676 16453677 16453679 16453681 16453682 16453683 16453684 16453686 16453688 16453689 16453690 16453691 16453692 16453693 16453694 16453697 16453699 16453701 16453705 16453707 16453709 16453711 16453714 16453720 16453725 16453726 16453728 16453731 16453733 16453736 16453738 16453739 16453740 16453741 16453744 16453745 16453750 16453751 16453753 16453754 16453755 16453756 16453757 16453758 16453760 16453762 16453764 16453765 16453768 16453770 16453774 16453776 16453780 16453782 16453783 16453784 16453786 16453787 16453792 16453795 16453796 16453797 16453800 16453801 16453802 16453803 16453804 16453805 16453806 16453807 16453810 16453812 16453813 16453815 16453817 16453819 16453820 16453821 16453826 16453827 16453828 16453832 16453833 16453836 16453837 16453838 16453841 16453843 16453844 16453845 16453848 16453849 16453851 16453852 16453855 16453856 16453859 16453862 16453864 16453865 16453866 16453867 16453868 16453869 16453870 16453871 16453872 16453873 16453874 16453875 16453877 16453878 16453882 16453885 16453887 16453888 16453889 16453891 16453893 16453894 16453897 16453900 16453903 16453904 16453905 16453907 16453908 16453911 16453914 16453916 16453919 16453920 16453921 16453922 16453924 16453926 16453928 16453929 16453931 16453932 16453939 16453940 16453945 16453946 16453948 16453949 16453950 16453951 16453952 16453953 16453954 16453956 16453959 16453962 16453964 16453965 16453966 16453968 16453975 16453976 16453977 16453978 16453979 16453984 16453988 16453989 16453991 16453993 16453996 16453997 16453999 16454004 16454005 16454006 16454007 16454010 16454011 16454013 16454014 16454015 16454016 16454019 16454022 16454024 16454026 16454032 16454033 16454035 16454036 16454038 16454039 16454040 16454044 16454047 16454048 16454049 16454052 16454054 16454056 16454057 16454058 16454059 16454060 16454061 16454062 16454064 16454065 16454066 16454069 16454074 16454075 16454079 16454082 16454083 16454086 16454087 16454093 16454095 16454096 16454100 16454101 16454103 16454109 16454112 16454116 16454117 16454118 16454119 16454122 16454125 16454129 16454130 16454131 16454132 16454135 16454137 16454138 16454139 16454140 16454141 16454142 16454143 16454144 16454145 16454147 16454148 16454149 16454151 16454152 16454153 16454154 16454155 16454157 16454159 16454160 16454161 16454163 16454164 16454165 16454168 16454169 16454170 16454172 16454173 16454175 16454176 16454178 16454180 16454181 16454183 16454184 16454185 16454186 16454187 16454188 16454189 16454190 16454192 16454193 16454194 16454195 16454196 16454197 16454198 16454199 16454200 16454201 16454204 16454206 16454207 16454209 16454210 16454211 16454213 16454214 16454215 16454216 16454217 16454218 16454219 16454220 16454221 16454222 16454223 16454224 16454226 16454227 16454229 16454230 16454231 16454232 16454233 16454234 16454235 16454236 16454237 16454238 16454241 16454243 16454244 16454247 16454248 16454249 16454250 16454251 16454252 16454253 16454254 16454255 16454256 16454257 16454259 16454260 16454261 16454262 16454263 16454264 16454267 16454268 16454269 16454270 16454271 16454272 16454273 16454274 16454275 16454276 16454277 16454278 16454279 16454280 16454281 16454282 16454283 16454284 16454285 16454287 16454288 16454289 16454291 16454292 16454293 16454294 16454295 16454296 16454297 16454299 16454300 16454301 16454302 16454303 16454304 16454306 16454309 16454311 16454313 16454314 16454316 16454317 16454318 16454319 16454320 16454321 16454323 16454324 16454325 16454327 16454329 16454330 16454332 16454333 16454334 16454335 16454336 16454339 16454340 16454342 16454343 16454345 16454346 16454348 16454349 16454351 16454352 16454353 16454354 16454355 16454358 16454360 16454361 16454363 16454364 16454365 16454367 16454368 16454369 16454370 16454371 16454373 16454374 16454376 16454377 16454378 16454379 16454380 16454381 16454382 16454384 16454386 16454387 16454388 16454389 16454390 16454393 16454394 16454396 16454397 16454398 16454399 16454400 16454401 16454403 16454404 16454409 16454410 16454412 16454414 16454417 16454418 16454419 16454421 16454422 16454423 16454425 16454427 16454428 16454429 16454430 16454431 16454432 16454433 16454434 16454435 16454436 16454437 16454438 16454439 16454441 16454443 16454444 16454445 16454446 16454447 16454448 16454451 16454452 16454453 16454454 16454455 16454456 16454457 16454458 16454459 16454461 16454462 16454463 16454464 16454465 16454466 16454468 16454469 16454470 16454471 16454472 16454473 16454475 16454476 16454477 16454478 16454480 16454481 16454482 16454483 16454484 16454485 16454486 16454487 16454488 16454489 16454490 16454491 16454492 16454495 16454496 16454497 16454499 16454500 16454501 16454502 16454503 16454504 16454505 16454506 16454507 16454509 16454510 16454511 16454512 16454513 16454515 16454516 16454517 16454518 16454519 16454521 16454522 16454523 16454524 16454525 16454526 16454527 16454529 16454530 16454531 16454532 16454534 16454535 16454536 16454538 16454540 16454541 16454543 16454545 16454546 16454547 16454548 16454549 16454550 16454551 16454552 16454553 16454554 16454555 16454556 16454557 16454558 16454560 16454561 16454562 16454563 16454564 16454565 16454566 16454568 16454569 16454571 16454572 16454573 16454575 16454576 16454577 16454578 16454579 16454580 16454581 16454583 16454584 16454587 16454588 16454590 16454591 16454592 16454593 16454595 16454596 16454597 16454598 16454599 16454601 16454602 16454603 16454605 16454606 16454607 16454608 16454610 16454612 16454614 16454615 16454616 16454619 16454620 16454621 16454623 16454624 16454625 16454626 16454627 16454628 16454629 16454630 16454633 16454634 16454635 16454636 16454637 16454638 16454640 16454641 16454643 16454645 16454646 16454647 16454648 16454649 16454650 16454651 16454654 16454655 16454656 16454657 16454658 16454660 16454661 16454663 16454667 16454668 16454669 16454670 16454671 16454672 16454673 16454674 16454676 16454677 16454678 16454679 16454681 16454683 16454684 16454685 16454686 16454687 16454688 16454690 16454691 16454692 16454693 16454695 16454698 16454699 16454700 16454701 16454702 16454703 16454707 16454708 16454710 16454712 16454713 16454715 16454717 16454721 16454722 16454723 16454724 16454725 16454727 16454728 16454729 16454730 16454731 16454732 16454733 16454734 16454735 16454736 16454737 16454738 16454739 16454740 16454741 16454742 16454743 16454744 16454745 16454746 16454747 16454748 16454749 16454752 16454753 16454755 16454758 16454759 16454760 16454762 16454763 16454764 16454766 16454767 16454769 16454772 16454773 16454774 16454775 16454776 16454777 16454778 16454779 16454783 16454784 16454785 16454786 16454787 16454788 16454789 16454790 16454791 16454792 16454793 16454794 16454798 16454799 16454800 16454803 16454805 16454807 16454809 16454811 16454813 16454816 16454817 16454818 16454820 16454822 16454825 16454827 16454831 16454833 16454835 16454836 16454838 16454839 16454841 16454843 16454844 16454845 16454846 16454847 16454852 16454854 16454857 16454858 16454859 16454860 16454861 16454862 16454863 16454864 16454865 16454867 16454868 16454871 16454873 16454874 16454875 16454876 16454878 16454879 16454880 16454883 16454884 16454885 16454886 16454888 16454890 16454892 16454894 16454895 16454897 16454899 16454900 16454901 16454903 16454904 16454905 16454907 16454908 16454909 16454910 16454911 16454912 16454913 16454914 16454915 16454916 16454918 16454919 16454920 16454921 16454923 16454924 16454925 16454926 16454928 16454929 16454930 16454933 16454934 16454935 16454937 16454938 16454939 16454940 16454941 16454943 16454944 16454945 16454946 16454950 16454952 16454953 16454954 16454955 16454956 16454957 16454958 16454961 16454962 16454963 16454964 16454965 16454967 16454968 16454970 16454971 16454973 16454974 16454975 16454976 16454977 16454978 16454979 16454980 16454981 16454982 16454983 16454984 16454985 16454986 16454987 16454988 16454990 16454991 16454992 16454994 16454996 16454997 16454998 16455000 16455003 16455005 16455007 16455008 16455012 16455014 16455015 16455016 16455017 16455018 16455020 16455021 16455022 16455025 16455026 16455028 16455029 16455030 16455032 16455033 16455034 16455035 16455036 16455037 16455040 16455041 16455044 16455045 16455046 16455047 16455049 16455050 16455051 16455053 16455056 16455057 16455058 16455059 16455060 16455061 16455065 16455066 16455067 16455068 16455073 16455078 16455080 16455081 16455083 16455084 16455086 16455088 16455089 16455090 16455091 16455093 16455094 16455095 16455097 16455098 16455102 16455103 16455104 16455105 16455108 16455111 16455113 16455114 16455116 16455118 16455119 16455120 16455121 16455122 16455123 16455124 16455126 16455128 16455130 16455131 16455134 16455135 16455136 16455138 16455139 16455142 16455145 16455146 16455147 16455148 16455149 16455150 16455152 16455153 16455154 16455155 16455156 16455157 16455161 16455163 16455164 16455165 16455168 16455171 16455172 16455173 16455175 16455176 16455178 16455179 16455181 16455182 16455183 16455184 16455185 16455186 16455187 16455188 16455189 16455191 16455192 16455194 16455195 16455196 16455197 16455198 16455199 16455200 16455201 16455202 16455204 16455205 16455206 16455207 16455208 16455209 16455211 16455212 16455213 16455214 16455216 16455219 16455220 16455221 16455223 16455224 16455225 16455226 16455227 16455228 16455229 16455230 16455231 16455232 16455233 16455234 16455235 16455236 16455237 16455238 16455239 16455241 16455243 16455244 16455245 16455246 16455247 16455248 16455250 16455251 16455252 16455254 16455256 16455257 16455259 16455260 16455261 16455262 16455263 16455264 16455265 16455267 16455268 16455269 16455270 16455271 16455274 16455275 16455276 16455278 16455279 16455282 16455283 16455285 16455286 16455287 16455288 16455289 16455291 16455294 16455296 16455298 16455301 16455302 16455303 16455304 16455305 16455306 16455308 16455309 16455310 16455312 16455313 16455315 16455316 16455317 16455318 16455319 16455320 16455321 16455322 16455323 16455324 16455325 16455327 16455328 16455330 16455332 16455334 16455338 16455340 16455343 16455344 16455345 16455346 16455347 16455351 16455352 16455355 16455357 16455358 16455361 16455362 16455365 16455367 16455370 16455372 16455376 16455378 16455379 16455382 16455384 16455386 16455388 16455389 16455390 16455391 16455392 16455393 16455395 16455396 16455397 16455398 16455399 16455400 16455402 16455403 16455406 16455407 16455409 16455410 16455412 16455413 16455415 16455417 16455418 16455420 16455421 16455422 16455424 16455425 16455426 16455429 16455430 16455432 16455433 16455434 16455435 16455436 16455437 16455438 16455439 16455440 16455441 16455442 16455443 16455444 16455445 16455446 16455450 16455452 16455453 16455454 16455455 16455456 16455457 16455459 16455461 16455462 16455463 16455464 16455465 16455466 16455468 16455469 16455470 16455472 16455474 16455475 16455477 16455478 16455479 16455481 16455482 16455483 16455484 16455485 16455486 16455487 16455489 16455492 16455494 16455495 16455496 16455498 16455499 16455500 16455502 16455503 16455505 16455506 16455507 16455508 16455509 16455512 16455513 16455514 16455515 16455519 16455520 16455523 16455530 16455532 16455533 16455535 16455536 16455537 16455538 16455539 16455540 16455541 16455542 16455545 16455546 16455547 16455548 16455550 16455551 16455552 16455554 16455555 16455556 16455557 16455558 16455559 16455561 16455563 16455564 16455567 16455569 16455570 16455571 16455572 16455573 16455574 16455575 16455577 16455580 16455583 16455584 16455585 16455586 16455587 16455588 16455589 16455590 16455591 16455593 16455594 16455595 16455597 16455598 16455599 16455601 16455603 16455604 16455605 16455606 16455607 16455610 16455613 16455615 16455616 16455625 16455626 16455628 16455630 16455632 16455633 16455634 16455638 16455639 16455640 16455641 16455642 16455644 16455646 16455647 16455648 16455649 16455651 16455652 16455653 16455654 16455656 16455657 16455659 16455662 16455664 16455665 16455666 16455669 16455670 16455671 16455673 16455674 16455676 16455677 16455680 16455685 16455687 16455690 16455693 16455697 16455699 16455700 16455703 16455708 16455710 16455711 16455712 16455713 16455716 16455718 16455719 16455720 16455722 16455724 16455725 16455726 16455727 16455728 16455730 16455733 16455734 16455740 16455741 16455742 16455744 16455745 16455747 16455748 16455751 16455753 16455754 16455756 16455759 16455760 16455761 16455762 16455764 16455766 16455768 16455769 16455770 16455771 16455774 16455775 16455779 16455780 16455783 16455784 16455786 16455787 16455790 16455791 16455792 16455793 16455794 16455795 16455796 16455798 16455799 16455800 16455803 16455877 16455878 16456657 16456658 16456659 16456660 16456661 16456662 16456663 16456664 16456665 16456666 16456667 16456668 16456669 16456670 16456671 16456672 16456673 16456674 16456675 16456676 16456677 16456678 16456679 16456680 16456681 16456682 16456683 16456684 16456685 16456756 16456757 16456760 16456769 16456770 16456771 16456772 16456775 16456796 16456797 16456798 16456802 16456806 16456807 16456818 16456821 16456822 16456825 16456833 16456834 16456836 16456899 16456924 16456930 16456968 16457069 16457070 16457071 16457072 16457073 16457074 16457075 16457076 16457077 16457078 16457079 16457081 16457082 16457083 16457084 16457085 16457086 16457087 16457088 16457089 16457090 16457092 16457093 16457097 16457098 16457102 16457105 16457106 16457107 16457110 16457111 16457114 16457115 16457116 17147401 17147402 17147403 17147404 17147405 17147406 17147408 17147409 17147410 17147411 17147412 17147413 17147414 17147415 17147416 17147417 17147418 17147419 17147420 17147421 17147423 17147425 17147426 17147427 17147428 17147429 17147430 17147431 17147433 17147434 17147435 17147436 17147437 17147438 17147439 17147440 17147441 17147443 17147445 17147446 17147447 17147448 17147449 17147450 17147451 17147454 17147455 17147456 17147457 17147458 17147459 17147460 17147461 17147462 17147463 17147465 17147467 17147469 17147471 17147473 17147476 17147477 17147478 17147479 17147480 17147481 17147482 17147483 17147484 17147486 17147487 17147488 17147489 17147490 17147491 17147492 17147493 17147494 17147495 17147497 17147499 17147501 17147502 17147503 17147504 17147506 17147508 17147509 17147510 17147512 17147514 17147515 17147516 17147517 17147518 17147519 17147520 17147521 17147522 17147524 17147525 17147526 17147527 17147528 17147529 17147530 17147531 17147532 17147533 17147534 17147535 17147536 17147537 17147538 17147539 17147540 17147541 17147542 17147543 17147544 17147545 17147546 17147547 17147548 17147549 17147550 17147551 17147552 17147553 17147554 17147555 17147556 17147557 17147558 17147559 17147560 17147561 17147562 17147563 17147564 17147565 17147566 17147567 17147568 17147569 17147570 17147571 17147572 17147573 17147574 17147575 17147576 17147577 17147578 17147579 17147580 17147581 17147582 17147583 17147584 17147585 17147586 17147587 17147588 17147589 17147591 17147592 17147593 17147594 17147595 17147596 17147597 17147598 17147599 17147600 17147601 17147602 17147603 17147604 17147605 17147606 17147607 17147608 17147611 17147612 17147613 17147614 17147615 17147616 17147617 17147618 17147619 17147621 17147622 17147624 17147625 17147626 17147627 17147628 17147630 17147631 17147632 17147633 17147634 17147635 17147636 17147637 17147639 17147640 17147643 17147644 17147645 17147646 17147648 17147649 17147650 17147651 17147652 17147654 17147655 17147656 17147657 17147658 17147659 17147660 17147661 17147662 17147663 17147664 17147665 17147666 17147667 17147668 17147669 17147670 17147671 17147672 17147673 17147674 17147675 17147676 17147677 17147678 17147679 17147680 17147681 17147682 17147683 17147684 17147685 17147686 17147687 17147688 17147689 17147690 17147691 17147692 17147693 17147694 17147695 17147696 17147697 17147698 17147699 17147700 17147701 17147702 17147703 17147704 17147705 17147706 17147707 17147708 17147709 17147712 17147713 17147714 17147715 17147716 17147717 17147718 17147719 17147720 17147721 17147722 17147723 17147724 17147725 17147726 17147727 17147728 17147729 17147730 17147731 17147732 17147733 17147734 17147735 17147736 17147737 17147738 17147739 17147740 17147741 17147742 17147743 17147744 17147745 17147747 17147748 17147749 17147751 17147752 17147753 17147754 17147755 17147756 17147757 17147758 17147759 17147760 17147761 17147762 17147763 17147764 17147765 17147766 17147767 17147768 17147769 17147770 17147771 17147773 17147774 17147775 17147776 17147777 17147778 17147779 17147781 17147782 17147783 17147784 17147785 17147786 17147787 17147788 17147789 17147790 17147791 17147792 17147793 17147794 17147795 17147796 17147797 17147798 17147799 17147800 17147801 17147802 17147803 17147804 17147805 17147806 17147807 17147808 17147809 17147811 17147812 17147813 17147814 17147815 17147816 17147818 17147819 17147820 17147823 17147824 17147826 17147827 17147828 17147829 17147831 17147832 17147833 17147834 17147835 17147836 17147837 17147839 17147840 17147841 17147844 17147845 17147847 17147848 17147849 17147851 17147852 17147853 17147854 17147855 17147856 17147857 17147858 17147859 17147860 17147861 17147862 17147863 17147864 17147865 17147866 17147867 17147868 17147869 17147870 17147871 17147872 17147873 17147874 17147875 17147876 17147877 17147878 17147879 17147880 17147881 17147882 17147883 17147884 17147885 17147886 17147887 17147888 17147889 17147891 17147892 17147893 17147894 17147895 17147896 17147897 17147898 17147899 17147901 17147902 17147904 17147906 17147907 17147908 17147909 17147910 17147913 17147914 17147915 17147916 17147917 17147918 17147919 17147920 17147921 17147923 17147925 17147926 17147927 17147928 17147929 17147931 17147932 17147933 17147934 17147935 17147937 17147941 17147943 17147944 17147945 17147946 17147947 17147949 17147950 17147951 17147952 17227405 17227406 17227407 17227408 17227409 17227410 17227411 17227412 17227413 17227414 17227415 17227417 17227418 17227419 17227420 17227421 17227422 17227423 17227424 17227425 17227426 17227427 17227428 17227429 17227430 17227431 17227433 17227434 17227435 17227436 17227438 17227439 17227440 17227442 17227443 17227444 17227445 17227446 17227447 17227448 17227449 17227450 17227451 17227452 17227453 17227454 17227455 17227456 17227457 17227458 17227459 17227460 17227461 17227462 17227463 17227464 17227465 17227466 17227467 17227468 17227469 17227470 17227471 17227472 17227473 17227474 17227475 17227476 17227477 17227478 17227479 17227480 17227481 17227482 17227483 17227485 17227487 17227488 17227489 17227490 17227491 17227492 17227493 17227494 17227495 17227496 17227497 17227498 17227499 17227500 17227501 17227502 17227503 17227504 17227505 17227506 17227507 17227508 17227509 17227510 17227511 17227513 17227514 17227515 17227516 17227517 17227518 17227519 17227520 17227521 17227522 17227523 17227524 17227525 17227526 17227527 17227528 17227529 17227530 17227531 17227532 17227533 17227534 17227535 17227536 17227537 17227538 17227539 17227540 17227541 17227542 17227543 17227544 17227545 17227546 17227547 17227548 17227549 17227550 17227551 17227552 17227553 17227554 17227555 17227556 17227557 17227558 17227559 17227560 17227561 17227562 17227563 17227564 17227565 17227566 17227567 17227568 17227569 17227570 17227571 17227572 17227573 17227574 17227575 17227576 17227577 17227578 17227579 17227580 17227581 17227582 17227583 17227584 17227585 17227586 17227587 17227588 17227589 17227590 17227591 17227592 17227593 17227594 17227595 17227597 17227598 17227599 17227600 17227601 17227602 17227603 17227604 17227605 17227606 17227607 17227608 17227609 17227610 17227611 17227612 17227613 17227614 17227615 17227616 17227617 17227618 17227619 17227620 17227621 17227622 17227623 17227624 17227625 17227626 17227627 17227628 17227629 17227630 17227631 17227632 17227633 17227634 17227635 17227636 17227637 17227638 17227639 17227640 17227641 17227642 17227643 17227644 17227645 17227646 17227647 17227648 17227650 17227652 17227653 17227654 17227655 17227656 17227657 17227658 17227659 17227660 17227661 17227662 17227663 17227664 17227666 17227667 17227668 17227669 17227670 17227671 17227672 17227673 17227674 17227675 17227676 17227677 17227678 17227679 17227681 17227682 17227683 17227684 17227685 17227686 17227687 17227688 17227689 17227690 17227691 17227692 17227693 17227694 17227695 17227696 17227697 17227698 17227699 17227700 17227701 17227702 17227703 17227704 17227705 17227706 17227707 17227708 17227709 17227710 17227711 17227712 17227713 17227714 17227715 17227716 17227717 17227718 17227719 17227720 17227721 17227722 17227723 17227724 17227725 17227726 17227727 17227728 17227729 17227730 17227731 17227732 17227733 17227734 17227735 17227736 17227737 17227738 17227739 17227740 17227741 17227742 17227743 17227744 17227746 17227747 17227748 17227749 17227750 17227751 17227752 17227753 17227754 17227755 17227756 17227757 17227758 17227759 17227760 17227761 17227762 17227763 17227764 17227765 17227766 17227768 17227769 17227770 17227771 17227772 17227773 17227774 17227775 17227776 17227777 17227778 17227779 17227780 17227781 17227782 17227783 17227784 17227785 17227786 17227787 17227788 17227789 17227790 17227791 17227792 17227793 17227794 17227795 17227796 17227797 17227798 17227799 17227800 17227801 17227803 17227804 17227805 17227806 17227807 17227809 17227810 17227811 17227812 17227813 17227814 17227816 17227817 17227818 17227819 17227820 17227821 17227822 17227823 17227824 17227825 17227826 17227827 17227828 17227830 17227831 17227832 17227833 17227834 17227835 17227836 17227837 17227838 17227839 17227840 17227841 17227842 17227843 17227844 17227845 17227846 17227847 17227848 17227849 17227850 17227851 17227852 17227853 17227854 17227855 17227856 17227858 17227859 17227861 17227862 17227863 17227864 17227865 17227866 17227867 17227868 17227869 17227870 17227871 17227872 17227873 17227874 17227881 17227882 17227889 17227890 17227893 17227898 17227900 17227901 17227903 17227912 17227913 17227918 17227927 17227932 17227938 17227940 17227944 17227950 17227957 17227975 17227976 17227978 17227979 17227980 17227981 17227983 17228000 17228011 17228012 17228020 17228028 17228035 17228036 17228037 17228043 17228044 17228049 17228054 17228055 17228057 17228058 17228059 17228060 17228062 17228069 17228074 17228078 17228088 17228090 17228091 17228100 17228103 17228104 17228105 17228106 17228107 17228108 17228110 17228113 17228114 17228116 17228127 17228129 17228130 17228131 17228132 17228134 17228137 17228139 17228140 17228141 17228142 17228143 17228147 17228148 17228151 17228154 17228155 17228156 17228157 17228158 17228159 17228162 17228163 17228164 17228166 17228167 17228168 17228170 17228171 17228172 17228174 17228176 17228177 17228178 17228179 17228180 17228181 17228183 17228184 17228187 17228189 17228190 17228192 17228193 17228196 17228197 17228198 17228199 17228200 17228202 17228204 17228205 17228208 17228209 17228210 17228211 17228213 17228214 17228215 17228216 17228218 17228219 17228220 17228221 17228223 17228224 17228225 17228226 17228227 17228228 17228229 17228230 17228233 17228234 17228235 17228236 17228238 17228241 17228242 17228246 17228247 17228248 17228251 17228252 17228253 17228254 17228255 17228256 17228259 17228260 17228262 17228263 17228264 17228266 17228267 17228269 17228272 17228273 17228274 17228275 17228276 17228280 17228281 17228282 17228283 17228284 17228285 17228286 17228288 17228289 17228291 17228292 17228293 17228294 17228295 17228296 17228297 17228298 17228299 17228301 17228304 17228306 17228307 17228311 17228312 17228313 17228315 17228317 17228318 17228319 17228320 17228323 17228324 17228325 17228326 17228329 17228331 17228333 17228334 17228335 17228336 17228337 17228338 17228339 17228342 17228343 17228344 17228345 17228347 17228350 17228351 17228352 17228354 17228355 17228357 17228358 17228359 17228361 17228364 17228365 17228366 17228367 17228371 17228372 17228373 17228374 17228375 17228377 17228378 17228380 17228382 17228384 17228386 17228388 17228389 17228390 17228391 17228392 17228393 17228394 17228395 17228396 17228399 17228401 17228403 17228405 17228406 17228407 17228413 17228421 17228423 17228424 17228425 17228426 17228427 17228428 17228430 17228432 17228434 17228435 17228436 17228438 17228439 17228440 17228441 17228443 17228446 17228447 17228448 17228450 17228451 17228452 17228453 17228454 17228455 17228457 17228458 17228459 17228460 17228461 17228462 17228463 17228464 17228465 17228466 17228467 17228468 17228469 17228470 17228471 17228472 17228473 17228474 17228475 17228476 17228477 17228478 17228479 17228480 17228481 17228482 17228483 17228484 17228485 17228486 17228487 17228488 17228489 17228490 17228491 17228492 17228493 17228494 17228495 17228496 17228497 17228498 17228499 17228500 17228501 17228502 17228503 17228504 17228505 17228506 17228507 17228508 17228509 17228510 17228511 17228512 17228513 17228514 17228515 17228516 17228517 17228518 17228519 17228520 17228521 17228522 17228523 17228524 17228525 17228527 17228528 17228529 17228530 17228531 17228532 17228533 17228534 17228535 17228536 17228537 17228538 17228539 17228540 17228541 17228542 17228543 17228544 17228545 17228546 17228547 17228548 17228549 17228550 17228551 17228552 17228553 17228554 17228555 17228556 17228557 17228558 17228559 17228560 17228561 17228562 17228563 17228564 17228565 17228566 17228567 17228568 17228569 17228570 17228571 17228572 17228573 17228575 17228577 17228578 17228579 17228580 17228581 17228582 17228583 17228584 17228585 17228586 17228587 17228588 17228589 17228590 17228591 17228592 17228593 17228594 17228595 17228596 17228597 17228598 17228599 17228600 17228601 17228602 17228603 17228604 17228605 17228606 17228607 17228608 17228609 17228610 17228611 17228612 17228613 17228614 17228615 17228616 17228617 17228618 17228619 17228620 17228621 17228622 17228623 17228624 17228625 17228626 17228627 17228628 17228629 17228630 17228631 17228632 17228633 17228634 17228635 17228636 17228637 17228638 17228639 17228640 17228641 17228642 17228643 17228644 17228645 17228646 17228647 17228648 17228649 17228650 17228651 17228652 17228653 17228654 17228655 17228656 17228657 17228658 17228659 17228660 17228661 17228662 17228663 17228664 17228665 17228666 17228667 17228668 17228669 17228670 17228671 17228672 17228673 17228674 17228675 17228676 17228677 17228678 17228679 17228680 17228681 17228682 17228683 17228685 17228686 17228687 17228688 17228689 17228690 17228692 17228693 17228694 17228695 17228696 17228697 17228698 17228699 17228700 17228701 17228702 17228703 17228704 17228705 17228706 17228707 17228708 17228709 17228710 17228711 17228712 17228713 17228714 17228716 17228717 17228718 17228719 17228720 17228721 17228722 17228723 17228724 17228725 17228726 17228727 17228728 17228729 17228730 17228731 17228732 17228733 17228734 17228735 17228736 17228737 17228738 17228739 17228740 17228741 17228742 17228743 17228744 17228745 17228746 17228747 17228748 17228750 17228751 17228752 17228753 17228754 17228755 17228756 17228757 17228758 17228759 17228760 17228761 17228762 17228763 17228764 17228765 17228766 17228767 17228768 17228769 17228770 17228771 17228772 17228773 17228774 17228775 17228776 17228777 17228778 17228779 17228780 17228781 17228782 17228783 17228784 17228785 17228786 17228787 17228788 17228789 17228790 17228791 17228792 17228793 17228794 17228795 17228796 17228797 17228798 17228799 17228800 17228802 17228803 17228804 17228805 17228806 17228807 17228808 17228809 17228810 17228811 17228812 17228813 17228814 17228815 17228816 17228817 17228818 17228819 17228820 17228821 17228822 17228823 17228824 17228825 17228826 17228827 17228828 17228829 17228830 17228831 17228832 17228833 17228834 17228835 17228836 17228837 17228838 17228839 17228840 17228841 17228842 17228844 17228845 17228846 17228847 17228848 17228849 17228850 17228851 17228852 17228853 17228854 17228855 17228856 17228857 17228858 17228859 17228860 17228861 17228862 17228863 17228864 17228865 17228866 17228867 17228868 17228869 17228870 17228871 17228872 17228873 17228874 17228876 17228877 17228878 17228879 17228880 17228881 17228882 17228883 17228884 17228885 17228886 17228887 17228888 17228889 17228890 17228891 17228892 17228893 17228894 17228895 17228896 17228897 17228898 17228899 17228900 17228901 17228902 17228903 17228904 17228905 17228906 17228907 17228908 17228909 17228910 17228911 17228912 17228913 17228914 17228915 17228916 17228917 17228918 17228919 17228921 17228922 17228924 17228925 17228926 17228927 17228928 17228929 17228930 17228931 17228932 17228933 17228934 17228936 17228937 17228938 17228939 17228940 17228941 17228942 17228943 17228944 17228945 17228946 17228947 17228948 17228949 17228950 17228951 17228952 17228953 17228954 17228955 17228956 17228957 17228958 17228959 17228960 17228961 17228962 17228963 17228964 17228965 17228966 17228967 17228968 17228969 17228970 17228971 17228972 17228973 17228974 17228975 17228976 17228977 17228978 17228979 17228980 17228981 17228982 17228983 17228984 17228985 17228986 17228987 17228988 17228989 17228990 17228991 17228992 17228993 17228994 17228995 17228996 17228997 17228998 17228999 17229000 17229001 17229002 17229003 17229005 17229006 17229007 17229008 17229009 17229010 17229011 17229013 17229014 17229015 17229016 17229017 17229018 17229019 17229020 17229021 17229022 17229023 17229024 17229025 17229026 17229027 17229028 17229029 17229030 17229031 17229032 17229033 17229034 17229035 17229036 17229037 17229038 17229039 17229040 17229042 17229043 17229044 17229045 17229046 17229047 17229048 17229049 17229050 17229051 17229052 17229053 17229054 17229055 17229056 17229057 17229058 17229059 17229060 17229062 17229063 17229064 17229065 17229066 17229067 17229068 17229069 17229071 17229072 17229073 17229074 17229075 17229076 17229079 17229080 17229081 17229082 17229084 17229085 17229087 17229089 17229090 17229091 17229092 17229093 17229094 17229095 17229096 17229097 17229099 17229100 17229102 17229103 17229104 17229105 17229106 17229107 17229108 17229109 17229110 17229111 17229112 17229113 17229114 17229115 17229116 17229117 17229118 17229119 17229120 17229122 17229124 17229125 17229126 17229127 17229129 17229130 17229131 17229133 17229135 17229136 17229137 17229138 17229139 17229140 17229141 17229142 17229143 17229144 17229145 17229146 17229147 17229148 17229149 17229150 17229151 17229152 17229153 17229154 17229155 17229156 17229157 17229158 17229159 17229160 17229161 17229162 17229163 17229164 17229165 17229166 17229167 17229168 17229169 17229170 17229171 17229172 17229173 17229174 17229175 17229176 17229177 17229178 17229179 17229180 17229181 17229182 17229183 17229184 17229185 17229186 17229187 17229188 17229189 17229190 17229191 17229192 17229193 17229194 17229195 17229197 17229199 17229201 17229202 17229203 17229204 17229205 17229206 17229207 17229208 17229209 17229210 17229211 17229213 17229214 17229215 17229216 17229217 17229218 17229219 17229221 17229222 17229223 17229224 17229225 17229226 17229227 17229228 17229229 17229230 17229231 17229232 17229233 17229234 17229235 17229236 17229237 17229238 17229239 17229240 17229241 17229242 17229243 17229244 17229245 17229246 17229247 17229248 17229249 17229250 17229251 17229252 17229253 17229254 17229255 17229256 17229257 17229258 17229259 17229260 17229261 17229262 17229263 17229264 17229266 17229267 17229268 17229269 17229270 17229271 17229272 17229273 17229274 17229275 17229276 17229277 17229278 17229279 17229280 17229281 17229282 17229283 17229284 17229285 17229286 17229287 17229288 17229289 17229290 17229291 17229292 17229293 17229294 17229295 17229296 17229297 17229298 17229299 17229300 17229301 17229302 17229303 17229304 17229305 17229306 17229307 17229308 17229309 17229310 17229311 17229312 17229313 17229314 17229315 17229316 17229317 17229318 17229319 17229320 17229321 17229322 17229323 17229324 17229325 17229326 17229327 17229328 17229329 17229330 17229331 17229332 17229333 17229334 17229335 17229336 17229337 17229338 17229339 17229340 17229341 17229342 17229343 17229344 17229345 17229347 17229348 17229349 17229350 17229351 17229353 17229354 17229355 17229356 17229357 17229358 17229359 17229360 17229361 17229362 17229363 17229364 17229366 17229367 17229368 17229369 17229370 17229371 17229372 17229373 17229374 17229375 17229376 17229377 17229378 17229379 17229380 17229381 17229382 17229383 17229384 17229385 17229386 17229387 17229388 17229389 17229391 17229392 17229393 17229394 17229395 17229396 17229397 17229398 17229399 17229400 17229401 17229402 17229403 17229404 17229405 17229406 17229407 17229408 17229409 17229410 17229411 17229412 17229413 17229414 17229415 17229416 17229417 17229418 17229419 17229421 17229422 17229423 17229424 17229425 17229426 17229428 17229429 17229430 17229431 17229432 17229433 17229434 17229435 17229436 17229437 17229438 17229439 17229440 17229441 17229442 17229443 17229444 17229446 17229448 17229449 17229450 17229451 17229452 17229453 17229454 17229455 17229456 17229457 17229458 17229459 17229460 17229462 17229463 17229464 17229465 17229466 17229467 17229468 17229469 17229470 17229471 17229472 17229473 17229474 17229475 17229476 17229477 17229478 17229479 17229480 17229481 17229482 17229483 17229484 17229485 17229486 17229487 17229488 17229489 17229490 17229491 17229492 17229493 17229494 17229495 17229496 17229497 17229498 17229499 17229500 17229501 17229502 17229503 17229504 17229505 17229506 17229507 17229508 17229509 17229510 17229511 17229512 17229513 17229514 17229515 17229516 17229517 17229518 17229519 17229520 17229521 17229522 17229523 17229524 17229525 17229526 17229527 17229528 17229529 17229530 17229531 17229532 17229533 17229536 17229537 17229538 17229539 17229540 17229541 17229542 17229543 17229544 17229545 17229546 17229547 17229548 17229549 17229550 17229551 17229552 17229553 17229554 17229555 17229557 17229558 17229559 17229560 17229561 17229562 17229563 17229564 17229565 17229566 17229567 17229568 17229569 17229570 17229571 17229572 17229573 17229574 17229575 17229576 17229577 17229578 17229579 17229580 17229581 17229582 17229583 17229584 17229585 17229586 17229587 17229588 17229589 17229590 17229592 17229593 17229594 17229595 17229596 17229597 17229599 17229602 17229604 17229605 17229606 17229608 17229610 17229611 17229612 17229614 17229616 17229617 17229619 17229621 17229622 17229623 17229624 17229627 17229628 17229629 17229630 17229631 17229632 17229634 17229635 17229636 17229637 17229638 17229639 17229640 17229641 17229642 17229643 17229645 17229646 17229647 17229648 17230499 17230500 17230501 17230502 17230503 17230504 17230505 17230506 17230507 17230508 17230509 17230510 17230511 17230512 17230513 17230514 17230515 17230516 17230517 17230518 17230519 17230520 17230521 17230522 17230523 17230524 17230525 17230526 17230527 17230528 17230530 17230531 17230532 17230533 17230534 17230535 17230536 17230537 17230538 17230539 17230540 17230541 17230542 17230543 17230544 17230545 17230546 17230547 17230548 17230549 17230550 17230551 17230552 17230553 17230554 17230555 17230556 17230557 17230558 17230559 17230560 17230561 17230562 17230563 17230564 17230565 17230566 17230567 17230568 17230569 17230570 17230571 17230572 17230573 17230574 17230575 17230576 17230577 17230578 17230579 17230580 17230581 17230582 17230583 17230584 17230585 17230586 17230587 17230588 17230589 17230590 17230591 17230592 17230593 17230594 17230595 17230596 17230597 17230598 17230599 17230600 17230601 17230602 17230603 17230604 17230605 17230606 17230607 17230608 17230609 17230610 17230611 17230612 17230613 17230614 17230615 17230616 17230617 17230618 17230619 17230620 17230621 17230622 17230623 17230624 17230625 17230626 17230627 17230628 17230629 17230630 17230631 17230632 17230633 17230634 17230635 17230636 17230637 17230638 17230639 17230640 17230641 17230642 17230643 17230644 17230645 17230646 17230647 17230648 17230649 17230650 17230651 17230652 17230653 17230654 17230655 17230656 17230657 17230658 17230659 17230660 17230661 17230662 17230663 17230664 17230665 17230666 17230667 17230668 17230669 17230670 17230671 17230674 17230675 17230676 17230677 17230678 17230679 17230680 17230681 17230682 17230683 17230684 17230685 17230686 17230687 17230688 17230689 17230690 17230691 17230692 17230694 17230695 17230696 17230697 17230698 17230699 17230700 17230701 17230702 17230703 17230704 17230705 17230706 17230707 17230708 17230709 17230710 17230711 17230712 17230713 17230714 17230716 17230717 17230718 17230719 17230720 17230722 17230723 17230724 17230725 17230726 17230727 17230728 17230729 17230731 17230732 17230733 17230734 17230735 17230736 17230737 17230738 17230739 17230740 17230741 17230742 17230743 17230744 17230745 17230746 17230747 17230748 17230749 17230750 17232290 17232372 17232530 17232552 17232671 17232743 17232865 17232909 17232912 17232915 17232920 17232923 17232973 17232995 17233171 17233196 17233259 17233343 17233390 17233538 17233552 17233720 17233750 17233824 17233833 17233842 17233843 17233910 17233947 17233959 17234007 17234037 17234187 17234329 17234387 17234389 17234391 17234394 17234398 17234406 17234407 17234410 17234413 17234414 17234416 17234417 17234418 17234424 17234429 17234431 17234432 17234435 17234437 17234438 17234442 17234444 17234456 17234457 17234459 17234461 17234463 17234465 17234466 17234471 17234472 17234473 17234474 17234479 17234481 17234484 17234487 17234489 17234491 17234496 17234497 17234499 17234501 17234502 17234503 17234505 17234508 17234509 17234510 17234512 17234514 17234525 17234526 17234529 17234530 17234532 17234539 17234540 17234542 17234549 17234551 17234553 17234554 17234559 17234560 17234562 17234573 17234578 17234579 17234580 17234581 17234975 17234976 17234978 17234979 17234980 17234981 17234982 17234983 17234984 17234985 17234986 17234988 17234990 17234991 17234992 17234993 17234994 17234995 17234996 17234997 17234998 17234999 17235000 17235001 17235002 17235003 17235004 17235005 17235006 17235007 17235008 17235009 17235010 17235011 17235013 17235014 17235015 17235016 17235017 17235018 17235019 17235020 17235021 17235022 17235023 17235024 17235025 17235026 17235027 17235028 17235029 17235030 17235031 17235032 17235033 17235034 17235035 17235036 17235037 17235038 17235039 17235041 17235042 17235043 17235044 17235045 17235046 17235047 17235048 17235049 17235050 17235051 17235052 17235053 17235054 17235056 17235057 17235058 17235060 17235061 17235062 17235063 17235064 17235065 17235066 17235067 17235068 17235069 17235070 17235071 17235073 17235074 17235076 17235077 17235079 17235080 17235081 17235082 17235083 17235084 17235085 17235086 17235087 17235088 17235089 17235090 17235091 17235093 17235094 17235095 17235096 17235097 17235098 17235099 17235101 17235103 17235104 17235105 17235106 17235107 17235108 17235109 17235110 17235111 17235112 17235113 17235114 17235115 17235116 17235117 17235118 17235119 17235120 17235121 17235123 17235124 17235125 17235126 17235127 17235128 17235129 17235130 17235133 17235134 17235135 17235136 17235137 17235138 17235139 17235140 17235141 17235142 17235143 17235144 17235145 17235146 17235149 17235150 17235151 17235152 17235153 17235155 17235156 17235158 17235160 17235161 17235162 17235163 17235164 17235165 17235167 17235169 17235170 17235171 17235172 17235174 17235175 17235176 17235179 17235181 17235182 17235184 17235185 17235186 17235187 17235188 17235189 17235190 17235191 17235192 17235193 17235194 17235195 17235196 17235198 17235199 17235200 17235203 17235204 17235206 17235207 17235208 17235209 17235210 17235211 17235212 17235213 17235214 17235215 17235216 17235217 17235219 17235221 17235222 17235223 17235224 17235226 17235227 17235228 17235229 17235230 17235231 17235232 17235233 17235234 17235235 17235236 17235237 17235238 17235239 17235241 17235242 17235243 17235244 17235245 17235246 17235247 17235248 17235249 17235250 17235251 17235252 17235253 17235254 17235255 17235256 17235257 17235258 17235259 17235260 17235261 17235262 17235263 17235264 17235265 17235266 17235267 17235268 17235269 17235270 17235271 17235273 17235275 17235276 17235277 17235278 17235279 17235280 17235281 17235282 17235283 17235284 17235285 17235286 17235287 17235288 17235289 17235290 17235291 17235293 17235294 17235295 17235296 17235297 17235298 17235299 17235300 17235301 17235302 17235303 17235304 17235305 17235307 17235308 17235309 17235310 17235311 17235312 17235313 17235314 17235315 17235316 17235317 17235318 17235319 17235320 17235321 17235322 17235323 17235324 17235325 17235326 17235327 17235328 17235329 17235330 17235331 17235332 17235333 17235334 17235336 17235337 17235338 17235339 17235340 17235341 17235342 17235343 17235344 17235345 17235346 17235347 17235349 17235350 17235352 17235353 17235354 17235355 17235356 17235357 17235358 17235359 17235360 17235361 17235362 17235363 17235364 17235365 17235366 17235367 17235368 17235369 17235370 17235371 17235372 17235373 17235374 17235375 17235376 17235377 17235378 17235379 17235380 17235381 17235383 17235384 17235385 17235386 17235387 17235388 17235389 17235391 17235392 17235393 17235394 17235395 17235396 17235397 17235398 17235399 17235401 17235402 17235403 17235404 17235405 17235406 17235407 17235408 17235409 17235410 17235411 17235412 17235413 17235414 17235415 17235416 17235417 17235419 17235420 17235421 17235422 17235423 17235424 17235425 17235426 17235428 17235429 17235430 17235431 17235432 17235433 17235434 17235435 17235436 17235438 17235439 17235440 17235441 17235442 17235443 17235444 17235445 17235446 17235447 17235448 17235451 17235452 17235453 17235454 17235455 17235456 17235457 17235458 17235459 17235460 17235461 17235462 17235463 17235464 17235465 17235466 17235467 17235468 17235469 17235470 17235471 17235472 17235473 17235474 17235475 17235476 17235477 17235478 17235479 17235481 17235482 17235483 17235484 17235485 17235486 17235487 17235488 17235489 17235490 17235491 17235492 17235493 17235494 17235495 17235496 17235497 17235498 17235499 17235500 17235501 17235502 17235503 17235504 17235505 17235507 17235508 17235509 17235510 17235511 17235512 17235513 17235514 17235515 17235516 17235517 17235518 17235519 17235520 17235521 17235522 17235523 17235524 17235525 17235526 17235527 17235528 17235529 17235530 17235531 17235532 17235533 17235534 17235535 17235536 17235537 17235538 17235539 17235540 17235541 17235542 17235543 17235544 17235545 17235546 17235547 17235548 17235549 17235550 17235551 17235552 17235553 17235554 17235555 17235556 17235557 17235558 17235559 17235560 17235561 17235562 17235564 17235565 17235566 17235567 17235568 17235569 17235570 17235571 17235572 17235573 17235574 17235575 17235576 17235577 17235578 17235579 17235580 17235581 17235582 17235583 17235585 17235586 17235587 17235588 17235589 17235590 17235591 17235592 17235593 17235594 17235595 17235596 17235597 17235599 17235600 17235601 17235602 17235603 17235604 17235605 17235606 17235607 17235608 17235609 17235610 17235612 17235613 17235614 17235615 17235617 17235618 17235619 17235620 17235621 17235622 17235624 17235625 17235626 17235627 17235628 17235629 17235630 17235631 17235632 17235633 17235634 17235635 17235636 17235637 17235638 17235639 17235641 17235642 17235643 17235644 17235645 17235646 17235647 17235648 17235649 17235650 17235651 17235652 17235653 17235654 17235655 17235656 17235657 17235658 17235659 17235660 17235661 17235663 17235664 17235665 17235666 17235667 17235668 17235669 17235670 17235671 17235672 17235673 17235674 17235675 17235676 17235677 17235678 17235679 17235680 17235681 17235682 17235683 17235684 17235685 17235686 17235687 17235688 17235689 17235690 17235691 17235692 17235693 17235694 17235695 17235696 17235697 17235698 17235700 17235701 17235702 17235703 17235704 17235707 17235708 17235709 17235711 17235713 17235714 17235715 17235716 17235717 17235718 17235719 17235720 17235721 17235723 17235724 17235726 17235727 17235728 17235729 17235730 17235731 17235733 17235734 17235735 17235736 17235737 17235738 17235739 17235740 17235741 17235742 17235743 17235744 17235745 17235746 17235747 17235748 17235750 17235751 17235752 17235753 17235754 17235755 17235756 17235757 17235758 17235759 17235761 17235762 17235763 17235764 17235765 17235766 17235767 17235768 17235769 17235770 17235771 17235772 17235773 17235774 17235775 17235777 17235778 17235779 17235780 17235781 17235782 17235784 17235785 17235786 17235787 17235788 17235789 17235790 17235792 17235793 17235794 17235795 17235796 17235797 17235798 17235800 17235801 17235802 17235803 17235804 17235805 17235806 17235808 17235809 17235810 17235811 17235812 17235813 17235814 17235816 17235817 17235818 17235819 17235820 17235821 17235822 17235823 17235824 17235825 17235826 17235827 17235828 17235829 17235830 17235831 17235833 17235834 17235835 17235836 17235837 17235838 17235839 17235840 17235841 17235842 17235843 17235844 17235845 17235846 17235848 17235849 17235850 17235851 17235852 17235853 17235854 17235855 17235856 17235857 17235858 17235859 17235860 17235861 17235863 17235864 17235866 17235868 17235869 17235870 17235871 17235872 17235873 17235874 17235875 17235876 17235879 17235880 17235881 17235882 17235883 17235884 17235885 17235886 17235887 17235888 17235889 17235890 17235891 17235892 17235893 17235894 17235897 17235898 17235899 17235900 17235901 17235902 17235903 17235904 17235906 17235908 17235910 17235911 17235912 17235913 17235914 17235915 17235916 17235917 17235918 17235919 17235920 17235921 17235922 17235923 17235924 17235925 17235927 17235928 17235929 17235930 17235931 17235932 17235933 17235934 17235935 17235936 17235937 17235938 17235939 17235940 17235941 17235942 17235943 17235944 17235945 17235946 17235947 17235949 17235950 17235951 17235952 17235953 17235954 17235955 17235956 17235957 17235958 17235959 17235960 17235961 17235962 17235963 17235964 17235965 17235966 17235967 17235968 17235969 17235970 17235971 17235972 17235973 17235974 17235975 17235976 17235977 17235978 17235979 17235980 17235981 17235982 17235983 17235984 17235985 17235986 17235987 17235988 17235989 17235990 17235991 17235992 17235993 17235994 17235996 17235997 17235998 17235999 17236000 17236001 17236002 17236003 17236004 17236005 17236006 17236007 17236008 17236009 17236010 17236011 17236012 17236013 17236015 17236016 17236017 17236018 17236019 17236020 17236021 17236022 17236023 17236024 17236025 17236026 17236027 17236028 17236029 17236030 17236031 17236033 17236034 17236035 17236037 17236038 17236039 17236040 17236041 17236042 17236043 17236044 17236047 17236049 17236050 17236053 17236054 17236055 17236056 17236061 17236062 17236063 17236064 17236065 17236068 17236069 17236070 17236072 17236073 17236074 17236075 17236078 17236080 17236081 17236082 17236083 17236084 17236085 17236086 17236087 17236089 17236090 17236092 17236094 17236096 17236097 17236098 17236102 17236103 17236104 17236106 17236107 17236108 17236110 17236111 17236112 17236113 17236115 17236116 17236117 17236118 17236120 17236122 17236123 17236124 17236126 17236127 17236128 17236129 17236130 17236131 17236135 17236136 17236138 17236139 17236140 17236143 17236144 17236145 17236146 17236147 17236148 17236150 17236151 17236152 17236153 17236154 17236155 17236158 17236159 17236160 17236162 17236163 17236164 17236165 17236166 17236167 17236168 17236170 17236172 17236173 17236174 17236175 17236176 17236178 17236179 17236180 17236181 17236182 17236184 17236185 17236188 17236189 17236190 17236191 17236193 17236194 17236195 17236196 17236197 17236198 17236199 17236202 17236203 17236204 17236205 17236206 17236208 17236209 17236210 17236211 17236212 17236213 17236214 17236216 17236218 17236220 17236221 17236223 17236224 17236226 17236227 17236229 17236230 17236231 17236232 17236233 17236235 17236237 17236238 17236240 17236242 17236243 17236244 17236245 17236246 17236248 17236249 17236250 17236253 17236254 17236255 17236256 17236257 17236258 17236259 17236262 17236264 17236265 17236266 17236267 17236269 17236270 17236271 17236273 17236274 17236275 17236277 17236278 17236279 17236282 17236283 17236284 17236285 17236288 17236289 17236290 17236291 17236292 17236293 17236296 17236298 17236299 17236300 17236301 17236303 17236304 17236307 17236310 17236311 17236314 17236315 17236316 17236318 17236320 17236321 17236322 17236325 17236326 17236327 17236328 17236329 17236330 17236331 17236332 17236335 17236337 17236339 17236340 17236341 17236344 17236345 17236347 17236348 17236349 17236351 17236354 17236355 17236356 17236358 17236359 17236362 17236363 17236365 17236367 17236369 17236370 17236371 17236372 17236373 17236374 17236378 17236379 17236380 17236381 17236382 17236383 17236384 17236387 17236388 17236390 17236391 17236392 17236393 17236394 17236396 17236401 17236402 17236403 17236406 17236407 17236408 17236410 17236412 17236413 17236414 17236416 17236417 17236418 17236419 17236421 17236423 17236424 17236425 17236426 17236427 17236428 17236429 17236430 17236431 17236433 17236438 17236439 17236441 17236442 17236444 17236446 17236447 17236448 17236449 17236450 17236452 17236453 17236454 17236455 17236457 17236459 17236461 17236462 17236465 17236467 17236468 17236471 17236472 17236473 17236474 17236475 17236476 17236479 17236480 17236483 17236484 17236485 17236486 17236488 17236489 17236490 17236491 17236492 17236493 17236494 17236495 17236496 17236497 17236499 17236500 17236501 17236502 17236503 17236504 17236505 17236507 17236508 17236509 17236510 17236512 17236513 17236514 17236515 17236516 17236518 17236520 17236522 17236523 17236524 17236525 17236526 17236527 17236530 17236531 17236532 17236534 17236535 17236537 17236538 17236539 17236540 17236541 17236542 17236543 17236545 17236546 17236547 17236548 17236549 17236551 17236552 17236553 17236554 17236555 17236556 17236557 17236558 17236561 17236562 17236564 17236566 17236567 17236568 17236569 17236570 17236571 17236572 17236573 17236574 17236575 17236576 17236577 17236578 17236580 17236581 17236582 17236583 17236584 17236585 17236587 17236588 17236589 17236590 17236591 17236593 17236595 17236597 17236599 17236600 17236602 17236603 17236604 17236608 17236609 17236610 17236611 17236612 17236613 17236614 17236615 17236616 17236617 17236618 17236619 17236620 17236621 17236622 17236623 17236624 17236626 17236628 17236629 17236630 17236631 17236632 17236633 17236635 17236636 17236637 17236638 17236639 17236640 17236641 17236642 17236643 17236644 17236645 17236647 17236648 17236649 17236650 17236652 17236654 17236655 17236656 17236657 17236660 17236661 17236662 17236663 17236667 17236668 17236669 17236670 17236671 17236672 17236673 17236676 17236677 17236678 17236680 17236681 17236685 17236686 17236687 17236688 17236689 17236690 17236691 17236692 17236693 17236694 17236695 17236696 17236697 17236699 17236700 17236702 17236703 17236705 17236707 17236708 17236709 17236710 17236711 17236712 17236713 17236715 17236716 17236717 17236718 17236719 17236721 17236723 17236724 17236725 17236726 17236727 17236728 17236729 17236730 17236732 17236735 17236736 17236739 17236740 17236741 17236742 17236743 17236745 17236750 17236751 17236752 17236753 17236754 17236755 17236757 17236758 17236759 17236761 17236762 17236763 17236764 17236765 17236766 17236768 17236769 17236770 17236773 17236775 17236777 17236782 17236783 17236784 17236786 17236787 17236788 17236789 17236791 17236793 17236797 17236800 17236801 17236802 17236803 17236804 17236805 17236806 17236807 17236808 17236810 17236811 17236812 17236813 17236814 17236815 17236816 17236817 17236818 17236819 17236821 17236822 17236823 17236827 17236828 17236829 17236830 17236831 17236832 17236833 17236834 17236837 17236838 17236839 17236840 17236841 17236843 17236845 17236846 17236847 17236848 17236849 17236850 17236852 17236853 17236854 17236855 17236856 17236857 17236858 17236859 17236860 17236861 17236862 17236864 17236871 17236872 17236874 17236875 17236876 17236877 17236878 17236879 17236880 17236881 17236882 17236884 17236886 17236888 17236889 17236892 17236894 17236895 17236896 17236899 17236901 17236902 17236903 17236904 17236905 17236906 17236909 17236910 17236911 17236912 17236914 17236916 17236918 17236919 17236920 17236921 17236922 17236923 17236924 17236925 17236926 17236927 17236929 17236930 17236931 17236933 17236934 17236935 17236938 17236940 17236941 17236942 17236944 17236945 17236946 17236947 17236948 17236950 17236952 17236953 17236954 17236955 17236956 17236958 17236959 17236960 17236961 17236962 17236963 17236965 17236966 17236967 17236968 17236969 17236970 17236974 17236975 17236976 17236977 17236978 17236979 17236981 17236982 17236983 17236984 17236985 17236986 17236987 17236988 17236989 17236990 17236992 17236993 17236994 17236995 17236996 17236998 17236999 17237000 17237001 17237002 17237003 17237004 17237005 17237006 17237007 17237008 17237009 17237010 17237012 17237013 17237014 17237015 17237017 17237018 17237019 17237020 17237021 17237022 17237023 17237024 17237025 17237026 17237027 17237028 17237029 17237030 17237031 17237032 17237033 17237034 17237035 17237036 17237037 17237038 17237039 17237040 17237041 17237042 17237043 17237046 17237047 17237049 17237050 17237051 17237052 17237053 17237054 17237055 17237056 17237057 17237058 17237059 17237060 17237062 17237064 17237065 17237067 17237068 17237069 17237070 17237071 17237072 17237073 17237075 17237076 17237077 17237078 17237079 17237080 17237081 17237083 17237084 17237085 17237086 17237087 17237088 17237089 17237090 17237091 17237092 17237093 17237094 17237095 17237098 17237099 17237100 17237101 17237102 17237103 17237104 17237105 17237106 17237107 17237108 17237109 17237110 17237111 17237112 17237113 17237114 17237115 17237116 17237117 17237118 17237119 17237120 17237121 17237122 17237124 17237125 17237126 17237127 17237128 17237129 17237130 17237131 17237132 17237133 17237134 17237136 17237137 17237138 17237140 17237141 17237144 17237146 17237148 17237149 17237151 17237152 17237153 17237154 17237155 17237156 17237157 17237158 17237160 17237164 17237165 17237166 17237168 17237169 17237170 17237171 17237172 17237173 17237175 17237176 17237177 17237178 17237180 17237181 17237182 17237183 17237184 17237186 17237187 17237188 17237190 17237191 17237193 17237195 17237196 17237197 17237199 17237200 17237201 17237202 17237203 17237205 17237206 17237207 17237208 17237209 17237210 17237211 17237212 17237214 17237215 17237216 17237217 17237218 17237219 17237221 17237222 17237223 17237224 17237225 17237227 17237228 17237229 17237231 17237232 17237233 17237234 17237235 17237236 17237237 17237238 17237240 17237241 17237242 17237243 17237244 17237245 17237246 17237248 17237249 17237250 17237251 17237252 17237253 17237254 17237255 17237256 17237257 17237258 17237259 17237260 17237262 17237263 17237264 17237265 17237266 17237267 17237269 17237270 17237272 17237273 17237275 17237276 17237277 17237278 17237279 17237280 17237281 17237282 17237283 17237284 17237285 17237287 17237288 17237289 17237290 17237291 17237292 17237293 17237294 17237295 17237296 17237297 17237298 17237299 17237301 17237302 17237303 17237304 17237305 17237306 17237307 17237309 17237310 17237311 17237312 17237313 17237314 17237315 17237316 17237317 17237318 17237319 17237321 17237322 17237323 17237324 17237327 17237328 17237329 17237330 17237331 17237332 17237333 17237334 17237335 17237336 17237337 17237338 17237339 17237340 17237341 17237342 17237343 17237344 17237345 17237346 17237347 17237348 17237350 17237351 17237352 17237353 17237354 17237355 17237356 17237357 17237359 17237360 17237361 17237362 17237364 17237365 17237367 17237368 17237369 17237370 17237371 17237372 17237373 17237374 17237375 17237376 17237377 17237378 17237379 17237380 17237381 17237382 17237383 17237384 17237385 17237386 17237387 17237388 17237389 17237390 17237391 17237392 17237393 17237394 17237395 17237396 17237397 17237398 17237400 17237401 17237402 17237403 17237404 17237405 17237406 17237407 17237408 17237409 17237410 17237411 17237412 17237413 17237414 17237415 17237416 17237417 17237418 17237419 17237420 17237421 17237422 17237423 17237424 17237425 17237426 17237427 17237428 17237429 17237430 17237431 17237432 17237433 17237434 17237436 17237437 17237439 17237440 17237441 17237442 17237444 17237445 17237446 17237447 17237448 17237449 17237450 17237451 17237452 17237453 17237454 17237455 17237456 17237458 17237460 17237461 17237462 17237463 17237464 17237465 17237466 17237467 17237468 17237469 17237470 17237471 17237472 17237473 17237474 17237475 17237476 17237477 17237478 17237479 17237480 17237481 17237482 17237484 17237486 17237487 17237488 17237489 17237490 17237491 17237492 17237493 17237494 17237495 17237496 17237497 17237498 17237499 17237500 17237501 17237502 17237503 17237504 17237505 17237506 17237507 17237508 17237509 17237510 17237511 17237512 17237513 17237514 17237515 17237516 17237517 17237518 17237519 17237520 17237522 17237523 17237524 17237525 17237526 17237527 17237528 17237529 17237530 17237531 17237532 17237533 17237534 17237535 17237536 17237537 17237538 17237539 17237541 17237542 17237544 17237545 17237546 17237547 17237548 17237549 17237550 17237551 17237626 17237630 17237633 17237634 17237648 17237649 17237674 17237704 17237718 17237727 17237750 17237766 17237767 17237798 17237799 17237810 17237818 17237826 17237832 17237834 17237861 17237862 17237873 17237886 17237907 17237935 17237936 17237940 17237970 17237973 17237977 17237978 17237980 17237993 17237999 17238001 17238002 17238006 17238009 17238010 17238011 17238013 17238014 17238016 17238017 17238078 19019401 19019409 19019411 19019416 19019421 19019433 19019436 19019444 19019445 19019447 19019449 19019452 19019458 19019459 19019461 19019464 19019465 19019470 19019473 19019476 19019477 19019489 19019492 19019494 19019495 19019498 19019500 19019502 19019504 19019511 19019512 19019514 19019517 19019520 19514401 19514402 19514403 19514404 19514405 19514406 19514407 19514408 19514409 19514410 19514411 19514412 19514413 19514414 19514415 19514416 19514417 19514418 19514419 19514420 19514421 19514422 19514423 19514424 19514425 19514426 19514427 19514428 19514429 19514430 19514431 19514432 19514433 19514434 19514435 19514436 19514437 19514438 19514439 19514440 19514441 19514442 19514443 19514444 19514445 19514446 19514447 19514448 19514449 19514450 19514451 19514452 19514453 19514454 19514455 19514457 19514458 19514459 19514461 19514462 19514463 19514464 19514465 19514466 19514467 19514468 19514469 19514470 19514471 19514472 19514473 19514474 19514475 19514476 19514477 19514478 19514479 19514480 19514481 19514482 19514483 19514484 19514485 19514486 19514487 19514488 19514489 19514490 19514491 19514492 19514493 19514494 19514495 19514496 19514497 19514498 19514499 19514500 19514501 19514502 19514503 19514504 19514505 19514506 19514507 19514508 19514509 19514510 19514511 19514512 19514513 19514514 19514515 19514516 19514517 19514518 19514519 19514520 19514521 19514522 19514523 19514524 19514525 19514526 19514527 19514528 19514529 19514530 19514531 19514532 19514533 19514534 19514535 19514536 19514537 19514538 19514539 19514540 19514541 19514542 19514543 19514544 19514545 19514546 19514547 19514549 19514550 19514551 19514552 19514553 19514554 19514555 19514556 19514557 19514558 19514559 19514560 19514561 19514562 19514563 19514564 19514565 19514566 19514567 19514568 19514569 19514570 19514571 19514572 19514573 19514574 19514575 19514576 19514577 19514578 19514579 19514580 19514581 19514582 19514583 19514584 19514585 19514586 19514587 19514588 19514589 19514590 19514591 19514592 19514593 19514594 19514595 19514596 19514597 19514598 19514599 19514600 19514601 19514602 19514603 19514604 19514605 19514606 19514607 19514608 19514609 19514610 19514611 19514612 19514613 19514614 19514615 19514616 19514617 19514618 19514619 19514620 19514621 19514622 19514623 19514624 19514625 19514626 19514627 19514628 19514629 19514630 19514631 19514632 19514660 19514661 19514662 19514663 19514664 19514665 19514666 19514667 19514668 19514669 19514670 19514671 19514672 19514673 19514676 19514681 19514682 19514683 19514684 19514685 19514686 19514687 19514688 19514689 19514690 19514691 19514692 19514693 19514694 19514695 19514696 19514697 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
For Additional Information Contact |
Mr. Chris Eustace 224-212-4892
|
Manufacturer Reason for Recall |
Frequent Proximal Occlusion Alarms (POAs) occurring on Symbiq devices with list numbers 16026 and 16027. When there are false alarms, the pump will continue to alarm even though there is no occlusion present. False POAs could result in a delay/interruption in therapy, and require the clinician to reset the alarm multiple times or to replace the pump.
|
FDA Determined Cause 2 |
Software design |
Action |
Hospira sent letters on May 2, 2013, to customers informing them of the discontinuation of sales, service, and support for the Symbiq pump were sent via traceable mailing to all affected customers. For questions regarding this recall call 224-212-4892. |
Quantity in Commerce |
20,311 |
Distribution |
Worldwide Distribution - USA (nationwide) |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = ABBOTT LABORATORIES 510(K)s with Product Code = FRN and Original Applicant = HOSPIRA, INC.
|