| Class 2 Device Recall OraPlug Absorbable Collagen Wound Dressing for Dental Surgery | |
Date Initiated by Firm | March 28, 2013 |
Date Posted | May 10, 2013 |
Recall Status1 |
Terminated 3 on June 03, 2015 |
Recall Number | Z-1297-2013 |
Recall Event ID |
64778 |
PMA Number | P840062 |
Product Classification |
Material, dressing, surgical, polylactic acid - Product Code LPG
|
Product | Ora Plug Absorbable Collagen Wound Dressing for Dental Surgery
The Absorbable Collagen Wound Dressings for Dental Surgery are intended for use on moist or bleeding oral wounds where a stabilized blood clot can help protect the surface of the wound from further injury. Topical administration of the Absorbable Collagen Wound Dressings directly over the wound effectively controls bleeding usually within two to five minutes. |
Code Information |
1110144, 1110243 1110782, 1111645 |
Recalling Firm/ Manufacturer |
Integra LifeSciences Corporation 105 Morgan Ln Plainsboro NJ 08536-3339
|
For Additional Information Contact | 609-275-2700 |
Manufacturer Reason for Recall | Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products. |
FDA Determined Cause 2 | Process control |
Action | The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: FCA@integralife.com or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product.
"However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol."
For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723. |
Quantity in Commerce | 31,940 boxes |
Distribution | Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|