Class 1 Device Recall Ventlab Manual Resuscitators

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Ventlab Manual Resuscitators see related information
Date Posted July 03, 2014
Recall Status1 Terminated on November 17, 2015
Recall Number Z-1948-2014
Recall Event ID 68242
510(K)Number K953546  K982215  K012842 
Product Classification Ventilator, emergency, manual (resuscitator) - Product Code BTM
Product Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.
Code Information Manual Resuscitator Bag Series:
AF1000, AF2000, AF5000, BT4000, VN2000,
VN5000, VT1000; Lot #s105147 -107609.
AF1000, AF2000, AF3000, AF4000, AF5000,
BT2000, BT3000, BT4000, PRO-1900, SC7000,
SC8120, SS3200, VN2000, VN3000, VN4000,
VN5000; Lot #s 106245 -107291.
VN2102; Lot #200349.
AF1000, AF2000, AF3000, AF4000, AF5000,
BT2000, BT3000, BT4000, BT5000, BVM700,
CPRM2000, CPRM3000, PRO-1000, PRO-2000,
SC7000, SC8020, VN2000, VN3000, VN4000; Lot #s 107029 -107634.
VN2002; Lot # 200492.
SC8000, SC9000; Lot #s 101441 -107461.
CPRM1000; Lot #s 99523 -107315.
Recalling Firm/
Manufacturer
Ventlab LLC
2710 Northridge Dr Nw
Suite A
Grand Rapids MI 49544-9112
For Additional Information Contact
800-237-5481
Manufacturer Reason
for Recall
The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.
FDA Determined
Cause 2
Process change control
Action The firm, Ventlab, issued a press dated May 14, 2014, stating Ventlab, LLC has initiated a voluntary medical device removal of a limited number of Ventlab" Resuscitator Bags. The bag series and lot numbers were included in a table as well as manufacture dates. In addition, Ventlab, LLC sent an "URGENT: MEDICAL DEVICE PRODUCT REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 19, 2014 to its distributors and customers. The letter described the product, problem and actions to be taken. The customers were instructed to review and examine their inventory for affected product; stop using them and immediately contact Ventlab, LLC for further instructions on the return of these products; and if product was further distributed, identify their customers/consignees and notify them of this product removal. The customers were also instructed to complete and return the attached Return Response Form via fax at 1-800-400-8820, Attn: Quality Department, or email to: PFA@ventlab.com as soon as possible and inform all affected personnel of this removal. Ventlab will send new corrected replacement resuscitation bags to customers once you return the affected product. If you have any questions regarding this action, please call Ventlab at 1-800-237-5481 between the hours of 8:30 AM to 5:00 PM (EST) Monday through Friday, or 1-844-635-5326 or via e-mail at PFA@ventlab.com.
Quantity in Commerce 251,384
Distribution US Nationwide Distribution in the states of OH, NY, TX, TN, FL, CT, and VA,
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = BTM and Original Applicant = MEDICAL MARKETING CONCEPTS
510(K)s with Product Code = BTM and Original Applicant = VENTLAB CORP.
-
-