|
Class 2 Device Recall Kodak DirectView DR 7500 Diagnostic XRay System |
|
Date Initiated by Firm |
July 09, 2014 |
Date Posted |
August 10, 2014 |
Recall Status1 |
Terminated 3 on May 12, 2017 |
Recall Number |
Z-2177-2014 |
Recall Event ID |
68727 |
510(K)Number |
K051258
|
Product Classification |
System, x-ray, stationary - Product Code KPR
|
Product |
Kodak DirectView DR 7500 Diagnostic X-Ray System
Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. |
Code Information |
The device can be tracked according to service code. The service code for the DR 7500 system is 8087. |
Recalling Firm/ Manufacturer |
Carestream Health, Inc. 1049 W Ridge Rd Rochester NY 14615-2731
|
For Additional Information Contact |
585-781-1997
|
Manufacturer Reason for Recall |
An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.
|
FDA Determined Cause 2 |
Device Design |
Action |
Carestream will notify all DR 7500 customers of the issue and schedule a Field Engineer to inspect and repair the system if required. Specific instructions (Service Modification) for the inspection and repair will followed by the Carestream Field Engineer in order to complete the correction. Documented evidence of the Field Engineer visit will be noted in the Carestream service engineering record for each affected system. Carestream estimates that the plan of correction will be completed within 2-4 weeks |
Quantity in Commerce |
589 systems |
Distribution |
Worldwide Distribution - US Nationwide and the countries of (United Kingdom, United Arab Emirates, Turkey, Taiwan, Switzerland, Sweden, Spain, South Africa, Portugal, Poland, Oman, New Zealand, Netherlands, Kuwait, Italy, Israel, Hong Kong, Germany, France, Finland, Denmark, Dubai, Czech Republic, China, Korea, Belgium, Australia, and Amsterdam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPR and Original Applicant = EASTMAN KODAK COMPANY
|
|
|
|