• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall TECNIS CL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TECNIS CLsee related information
Date Initiated by FirmMay 13, 2014
Date PostedAugust 26, 2014
Recall Status1 Terminated 3 on May 18, 2017
Recall NumberZ-2281-2014
Recall Event ID 68346
PMA NumberP880081 
Product Classification intraocular lens - Product Code HQL
ProductAMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the visual correction of aphakia in adult patients. Tecnis CL Mono 13 mm with the following diopter sizes: 19.5-20 Product Usage: AMO TECNIS CL monofocal foldable IOLs with OptiEdge design are ultraviolet-absorbing posterior chamber intraocular lenses which compensate for corneal spherical aberrations. TECNIS CL Silicone Intraocular Lenses are indicated for primary implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction or phacoemulsification. These devices are intended to be placed in the capsular bag.
Code Information Serial Num 1014621311 1014631311 1014641311 1014651311 1014661311 1014671311 1014681311 1014691311 1014701311 1014711311 1014721311 1014731311 1014741311 1014751311 1014761311 1014771311 1014781311 1014791311 1014801311 1014811311 1014821311 1014831311 1014841311 1014851311 1014861311 1014871311 1014881311 1014891311 1014901311 1014911311 1014921311 1014931311 1014941311 1014951311 1014961311 1014971311 1014981311 1014991311 1015001311 1015011311 1015021311 1015031311 1015041311 1015051311 1015061311 1015071311 1015081311 1015091311 1015101311 1015111311 1015121311 1015131311 1015141311 1015151311 1015161311 1015171311 1015181311 1015191311 1015201311 1015211311 1015221311 1015231311 1015241311 1015251311 1015261311 1015271311 1015281311 1015291311 1015301311 1015311311 1015321311 1015331311 1015341311 1015351311 1015361311 1015371311 1015381311 1015391311 1015401311 1015411311 1015421311 1015431311 1015441311 1015451311 1015461311 1015471311 1015481311 1015491311 1015501311 1015511311 1015521311 1015531311 1015541311 1015551311 1015561311 1015571311 1015581311 1015591311 1015601311 1015611311 1016501310 1016521310 1016541310 1016561310 1016581310 1016601310 1016621310 1016641310 1016661310 1016671310 1016681310 1016691310 1016701310 1016711310 1016721310 1016731310 1016741310 1016751310 1016761310 1016771310 1016781310 1016791310 1016801310 1016811310 1016821310 1016831310 1016841310 1016851310 1016861310 1016871310 1016881310 1016891310 1016901310 1016911310 1016921310 1016931310 1016941310 1016951310 1016961310 1016971310 1016981310 1016991310 1017001310 1017011310 1017021310 1017031310 1017041310 1017051310 1017061310 1017071310 1017081310 1017091310 1017101310 1017111310 1017121310 1017131310 1017141310 1017151310 1017161310 1017171310 1017191310 1017211310 1017231310 1017241310 1017251310 1017261310 1017271310 1017281310 1017291310 1017301310 1017311310 1017321310 1017331310 1017341310 1017351310 1017371310 1017391310 1017411310 1017431310 1017451310 1017471310 1017491310 1017511310 1017531310 1017551310 1017571310 1017591310 1019801310 1019811310 1019821310 1019831310 1019841310 1019851310 1019861310 1019871310 1019881310 1019891310 1019901310 1019911310 1019921310 1019931310 1019941310 1019951310 1019961310 1019971310 1019981310 1019991310 1020001310 1020011310 1020021310 1020031310 1020041310 1020051310 1020061310 1020071310 1020081310 1020091310 1020101310 1020111310 1020121310 1020131310 1020141310 1020151310 1020161310 1020171310 1020181310 1020191310 1020201310 1020211310 1020221310 1020231310 1020241310 1020251310 1020261310 1020271310 1020281310 1020291310 1020301310 1020311310 1020321310 1020331310 1020341310 1020351310 1020361310 1020371310 1020381310 1020391310 1020401310 1020411310 1020421310 1020431310 1020441310 1020451310 1020461310 1020471310 1020481310 1020491310 1020501310 1020511310 1020521310 1020531310 1020541310 1020551310 1020561310 1020571310 1020581310 1020591310 1020601310 1020611310 1020621310 1020631310 1020641310 1020651310 1020661310 1020671310 1020681310 1020691310 1020701310 1020711310 1020721310 1020731310 1020741310 1020751310 1020761310 1020771310 1020781310 1020791310 1020801310 1020811310 1020821310 1020831310 1025301310 1025311310 1025321310 1025331310 1025341310 1025351310 1025361310 1025371310 1025381310 1025391310 1025401310 1025411310 1025421310 1025431310 1025441310 1025451310 1025461310 1025471310 1025481310 1025491310 1025501310 1025511310 1025521310 1025531310 1025541310 1025551310 1025561310 1025571310 1025581310 1025591310 1025601310 1025611310 1025621310 1025631310 1025641310 1025651310 1025661310 1025671310 1025681310 1025691310 1025701310 1025711310 1025721310 1025731310 1025741310 1025751310 1025761310 1025771310 1025781310 1025791310 1025801310 1025811310 1025821310 1025831310 1025841310 1025851310 1025861310 1025871310 1025881310 1025891310 1025901310 1025911310 1025921310 1025931310 1025941310 1025951310 1025961310 1025971310 1025981310 1025991310 1026001310 1026011310 1026021310 1026031310 1026041310 1026051310 1026061310 1026071310 1026081310 1026091310 1026101310 1026111310 1026121310 1026131310 1026141310 1026151310 1026161310 1026171310 1026181310 1026191310 1026201310 1026211310 1026221310 1026231310 1026241310 1026251310 1026261310 1026271310 1026281310 1026291310 1026301310 1026311310 1026321310 1026331310 1026341310 1026351310 1026361310 1026371310 1026381310 1029981312 1029991312 1030001312 1030011312 1030021312 1030031312 1030041312 1030051312 1030061312 1030071312 1030081312 1030091312 1030101312 1030121312 1030141312 1030161312 1030171312 1030181312 1030191312 1030211312 1030231312 1030251312 1030271312 1030291312 1030311312 1030331312 1030351312 1030371312 1030381312 1030391312 1030411312 1030421312 1030431312 1030441312 1030451312 1030461312 1030471312 1030481312 1030491312 1030501312 1030511312 1030521312 1030531312 1030551312 1030571312 1030591312 1030611312 1030631312 1030651312 1030671312 1030801312 1030821312 1030841312 1030861312 1030881312 1030901312 1030921312 1030941312 1030961312 1030981312 1030991312 1031001312 1031021312 1031031312 1031041312 1031051312 1031061312 1031071312 1031081312 1031091312 1031101312 1031111312 1031121312 1031131312 1031141312 1031151312 1031161312 1031171312 1031181312 1031191312 1031201312 1031211312 1031231312 1031251312 1031261312 1031281312 1031291312 1031301312 1031311312 1031321312 1031331312 1031341312 1031351312 1031361312 1031371312 1031381312 1031401312 1031411312 1031421312 1031441312 1031451312 1031461312 1031481312 1031501312 1031521312 1031541312 1031561312 1031581312 1031601312 1031831312 1031861312 1031871312 1031911312 1031931312 1031951312 1031971312 1031991312 1032011312 1032031312 1032051312 1032071312 1032971310 1032991310 1033011310 1033031310 1033051310 1033071310 1033491310 1033511310 1033531310 1033551310 1033571310 1033591310 1033611310 1033631310 1033651310 1033671310 1033691310 1033711310 1033731310 1033751310 1033771310 1033791310 1033801310 1033811310 1033821310 1033831310 1033841310 1033861310 1033881310 1033901310 1033921310 1033941310 1033961310 1033981310 1034001310 1034021310 1034041310 1036341310 1036381310 1036401310 1036471310 1036491310 1036511310 1036641310 1036681310 1036691310 1036701310 1036891310 1036911310 1036931310 1036951310 1036971310 1036991310 1037011310 1037031310 1037051310 1037071310 1037091310 1037121310 1037141310 1037161310 1037181310 1037251310 1041101309 1041181309 1041201309 1041221309 1041241309 1041261309 1041281309 1041301309 1041321309 1041331309 1041341309 1041351309 1041361309 1041371309 1041381309 1041391309 1041401309 1041411309 1041421309 1041431309 1041441309 1041441311 1041451309 1041461309 1041461311 1041481309 1041481311 1041501309 1041501311 1041521309 1041521311 1041531309 1041531311 1041541309 1041541311 1041551309 1041551311 1041561309 1041561311 1041571309 1041571311 1041581309 1041581311 1041591309 1041591311 1041601309 1041601311 1041611309 1041611311 1041621309 1041621311 1041631309 1041631311 1041641309 1041641311 1041651309 1041661309 1041661311 1041671309 1041681309 1041681311 1041691309 1041691311 1041701309 1041701311 1041711309 1041711311 1041721309 1041731309 1041741309 1041761309 1041771309 1041781309 1041781311 1041791309 1041801309 1041801311 1041821309 1041821311 1041841309 1041841311 1041851311 1041861309 1041861311 1041871311 1041881309 1041881311 1041891311 1041901309 1041901311 1041911309 1041911311 1041921309 1041921311 1041931309 1041931311 1041941309 1041941311 1041951309 1041951311 1041961309 1041961311 1041971309 1041971311 1041981309 1041981311 1041991309 1041991311 1042001309 1042011309 1042011311 1042021309 1042021311 1042031311 1042041309 1042051309 1042051311 1042061309 1042071309 1042071311 1042091311 1042111311 1042131311 1042151311 1042171311 1042191311 1042211311 1042271311 1042291311 1042311311 1042331311 1046331309 1046341309 1046351309 1046361309 1046371309 1046381309 1046391309 1046401309 1046411309 1046421309 1046431309 1046441309 1046451309 1046461309 1046471309 1046481309 1046491309 1046501309 1046511309 1046521309 1046531309 1046541309 1046551309 1046561309 1046571309 1046581309 1046591309 1046591311 1046601309 1046611309 1046611311 1046621309 1046631309 1046631311 1046641309 1046651309 1046651311 1046661309 1046671309 1046671311 1046681309 1046691309 1046691311 1046701309 1046711309 1046711311 1046721309 1046731309 1046731311 1046741309 1046751309 1046751311 1046761309 1046771309 1046771311 1046781309 1046791309 1046791311 1046801309 1046811309 1046811311 1046821309 1046831309 1046831311 1046841309 1046851309 1046851311 1046861309 1046871309 1046871311 1046881309 1046891309 1046891311 1046901309 1046911309 1046911311 1046921309 1046931309 1046931311 1046941309 1046951309 1046951311 1046961309 1046971309 1046971311 1046981309 1046981311 1046991309 1046991311 1047001309 1047001311 1047011311 1047021309 1047021311 1047031311 1047041309 1047041311 1047051311 1047061309 1047061311 1047071309 1047071311 1047081309 1047081311 1047091311 1047101309 1047101311 1047111311 1047121309 1047121311 1047131311 1047141309 1047141311 1047151311 1047161309 1047161311 1047171309 1047171311 1047181309 1047181311 1047191309 1047191311 1047201309 1047201311 1047211309 1047211311 1047221309 1047221311 1047231309 1047231311 1047241309 1047241311 1047251309 1047251311 1047261309 1047261311 1047271309 1047271311 1047281309 1047281311 1047291309 1047291311 1047301309 1047301311 1047311309 1047311311 1047321309 1047321311 1047331309 1047331311 1047341309 1047341311 1047351309 1047351311 1047361309 1047361311 1047371309 1047371311 1047381309 1047381311 1047391309 1047391311 1047401309 1047401311 1047411309 1047411311 1047421309 1047421311 1047431309 1047431311 1047441311 1047451311 1047461311 1047471311 1047481311 1047491311 1047501311 1047511311 1047521311 1047531311 1047541311 1047551311 1047561311 1047571311 1047581311 1047591311 1047601311 1047611311 1047621311 1047631311 1047641311 1047651311 1047661311 1047671311 1047681311 1047691311 1047701311 1047711311 1047721311 1047731311 1047741311 1047751311 1047761311 1050631309 1050641309 1050651309 1050661309 1050671309 1050681309 1050691309 1050701309 1050711309 1050721309 1050731309 1050741309 1050751309 1050761309 1050771309 1050781309 1050791309 1050801309 1050811309 1050821309 1050841309 1050861309 1050871309 1050881309 1050891309 1050901309 1050911309 1050921309 1050931309 1050941309 1050951309 1050961309 1050971309 1050981309 1050991309 1051001309 1051011309 1051021309 1051041309 1051051309 1051061309 1051071309 1051081309 1051091309 1051101309 1051111309 1051121309 1051131309 1051141309 1051151309 1051161309 1051171309 1051181309 1051191309 1051201309 1051211309 1051221309 1051231309 1051241309 1051251309 1051261309 1051271309 1051281309 1051291309 1051301309 1051311309 1051321309 1051331309 1051341309 1051351309 1051361309 1051371309 1051381309 1051391309 1051401309 1051421309 1051431309 1051441309 1051451309 1051461309 1051471309 1051481309 1052331311 1052351311 1052371311 1052391311 1052411311 1052431311 1052451311 1052471311 1052491311 1052511311 1052531311 1052551311 1052571311 1052581311 1052591311 1052601311 1052611311 1052621311 1052631311 1052641311 1052651311 1052661311 1052671311 1052681311 1052691311 1052711311 1052721311 1052731311 1052741311 1052751311 1052761311 1052771311 1052781311 1052791311 1052801311 1052811311 1052821311 1052831311 1052841311 1052851311 1052861311 1052871311 1052881311 1052891311 1052901311 1052911311 1052921311 1052931311 1052941311 1052951311 1052961311 1052971311 1052981311 1052991311 1053001311 1053011311 1053021311 1053031311 1053041311 1053051311 1053061311 1053071311 1053081311 1053091311 1053111311 1053121311 1053131311 1053141311 1053151311 1053161311 1053171311 1053181311 1053191311 1053201311 1053211311 1053221311 1053231311 1053241311 1053251311 1053261311 1053271311 1053281311 1053291311 1053301311 1053311311 1053321311 1053331311 1053341311 1053351311 1053361311 1053371311 1053381311 1057611310 1057621310 1057631310 1057641310 1057651310 1057661310 1057671310 1057681310 1057691310 1057701310 1057711310 1057721310 1057731310 1057741310 1057751310 1057761310 1057771310 1057781310 1057791310 1057801310 1057811310 1057821310 1057831310 1057841310 1057851310 1057861310 1057871310 1057881310 1057891310 1057901310 1057911310 1057921310 1057931310 1057941310 1057951310 1057961310 1057971310 1057981310 1057991310 1058001310 1058011310 1058031310 1058051310 1058071310 1058091310 1058111310 1058131310 1058141310 1058151310 1058161310 1058171310 1058181310 1058191310 1058201310 1058211310 1058221310 1058231310 1058241310 1058251310 1058261310 1058271310 1058281310 1058291310 1058301310 1058311310 1058321310 1058331310 1058341310 1058351310 1058351312 1058361310 1058371310 1058371312 1058381310 1058391310 1058391312 1058401310 1058411310 1058411312 1058431310 1058431312 1058451310 1058451312 1058471310 1058471312 1058491310 1058491312 1058511310 1058511312 1058531310 1058531312 1058541310 1058551310 1058551312 1058561310 1058571310 1058571312 1058581310 1058591310 1058591312 1058601310 1058611310 1058611312 1058621310 1058631310 1058631312 1058641310 1058651310 1058651312 1058661310 1058671310 1058671312 1058681310 1058681312 1058691310 1058691312 1058701310 1058701312 1058711310 1058711312 1058721312 1058731312 1058741312 1058751312 1058761312 1058771312 1058781312 1058791312 1058801312 1058811312 1058821312 1058831312 1058841312 1058851312 1058861312 1058871312 1058881312 1058891312 1058901312 1058911312 1058921312 1058931312 1058941312 1058951312 1058961312 1058971312 1058981312 1059001312 1059021312 1059041312 1059061312 1059081312 1059101312 1059111312 1059121312 1059131312 1059141312 1059151312 1059161312 1059171312 1059181312 1059191312 1059201312 1059211310 1059221310 1059221312 1059231310 1059231312 1059241310 1059241312 1059251310 1059251312 1059261310 1059261312 1059271310 1059271312 1059281310 1059281312 1059291312 1059301310 1059311312 1059321310 1059331310 1059341310 1059351310 1059361310 1059371310 1059381310 1059391310 1059401310 1059411310 1059421310 1059431310 1059441310 1059451310 1059461310 1059471310 1059481310 1059501310 1059511310 1059521310 1059531310 1059541310 1059551310 1059561310 1059571310 1059581310 1059591310 1059601310 1059611310 1059621310 1059631310 1059641310 1059651310 1059661310 1059671310 1059681310 1059691310 1059701310 1059711310 1059721310 1059731310 1059741310 1059751310 1059761310 1059771310 1059791310 1059801310 1059811310 1059821310 1059831310 1059841310 1059851310 1059861310 1059871310 1059881310 1059891310 1059901310 1059911310 1059921310 1059931310 1059941310 1059951310 1059961310 1059971310 1059981310 1060001310 1060011310 1060021310 1060031310 1060041310 1060051310 1060061310 1060071310 1060081310 1060101310 1060121310 1060131310 1060141310 1060151310 1060161310 1060171310 1060181310 1060191310 1060201310 1060211310 1060231310 1060251310 1060291310 1060311310 1060331310 1060351310 1060391310 1060411310 1060431310 1060451310 1060471310 1060491310 1060511310 1060531310 1060541310 1060551310 1060561310 1060571310 1060581310 1060591310 1060601310 1060611310 1060621310 1060641310 1060651310 1060661310 1060681310 1060701310 1060721310 1060741310 1060761310 1060771310 1060781310 1060791310 1060801310 1060811310 1060821310 1060831310 1060841310 1060851310 1060861310 1060871310 1060881310 1060891310 1060901310 1060911310 1060921310 1060931310 1062011310 1062031310 1062051310 1062061310 1062071310 1062081310 1062091310 1062101310 1062111310 1062131310 1062141310 1062151310 1062161310 1062171310 1062181310 1062191310 1062201310 1062211310 1062221310 1062231310 1062251310 1062271310 1062291310 1062311310 1062331310 1062351310 1062371310 1062391310 1062421310 1062441310 1062461310 1062481310 1062501310 1062511310 1062521310 1062541310 1062551310 1062561310 1062571310 1062581310 1062601310 1062621310 1062641310 1062661310 1062681310 1062691310 1062701310 1062711310 1062721310 1062731310 1062741310 1062751310 1062761310 1062771310 1062781310 1062791310 1062801310 1062811310 1062821310 1062831310 1062841310 1062851310 1062861310 1062871310 1062881310 1062891310 1062901310 1062911310 1062921310 1062931310 1062941310 1062951310 1062961310 1062971310 1062981310 1062991310 1063001310 1063011310 1063021310 1063031310 1063041310 1063051310 1063061310 1063071310 1063081310 1063091310 1063101310 1063111310 1070391309 1070401309 1070411309 1070421309 1070431309 1070441309 1070451309 1070461309 1070471309 1070481309 1070491309 1070501309 1070511309 1070521309 1070531309 1070541309 1070551309 1070561309 1070571309 1070581309 1070591309 1070601309 1070611309 1070621309 1070631309 1070641309 1070651309 1070661309 1070671309 1070681309 1070691309 1070701309 1070711309 1070721309 1070731309 1070741309 1070751309 1070761309 1070771309 1070781309 1070791309 1070801309 1070811309 1070821309 1070831309 1070841309 1070851309 1070861309 1070871309 1070881309 1070891309 1070901309 1070911309 1070921309 1070931309 1070941309 1070951309 1070961309 1070971309 1070981309 1070991309 1071001309 1071011309 1071021309 1071031309 1071041309 1071051309 1071061309 1071071309 1071081309 1071091309 1071101309 1071111309 1071121309 1071131309 1071141309 1071151309 1071161309 1071171309 1071181309 1071191309 1071201309 1071211309 1071221309 1071231309 1071241309 1071251309 1071261309 1071271309 1071281309 1071291309 1071301309 1071311309 1071321309 1071331309 1071341309 1071351309 1071361309 1071371309 1071381309 1071391309 1071401309 1071411309 1071421309 1071431309 1071441309 1071451309 1071461309 1071471309 1071481309 1071491309 1074641311 1074651311 1074661311 1074671311 1074681311 1074691311 1074701311 1074711311 1074731311 1074751311 1074771311 1074781311 1074791311 1074811311 1074821311 1074841311 1074851311 1074871311 1074891311 1074911311 1074931311 1074951311 1074971311 1074991311 1075011311 1075031311 1075051311 1075071311 1075091311 1075111311 1075121311 1075131311 1075141311 1075151311 1075161311 1075171311 1075181311 1075191311 1075201311 1075211311 1075221311 1075231311 1075241311 1075251311 1075261311 1075271311 1075281311 1075291311 1075301311 1075311311 1075331311 1075351311 1075371311 1075381311 1075391311 1075401311 1075411311 1075431311 1075441311 1075451311 1075461311 1075471311 1075481311 1075491311 1075501311 1075521311 1075541311 1075561311 1075581311 1075601311 1075621311 1075661311 1075721311 1084511311 1084571311 1084591311 1084611311 1084631311 1084651311 1084671311 1084691311 1084711311 1084721311 1084751311 1084771311 1084791311 1084811311 1084821311 1084841311 1084861311 1084881311 1084901311 1084911311 1084931311 1085041311 1085221311 1085241311 1085261311 1085281311 1085301311 1085321311 1085341311 1089931311 1089951311 1089971311 1089991311 1090031311 1090051311 1090071311 1090091311 1090131311 1090151311 1090171311 1090191311 1090211311 1090231311 1090251311 1090271311 1090291311 1090311311 1090331311 1090351311 1090361311 1090371311 1090381311 1090391311 1090401311 1090411311 1090421311 1090431311 1090441311 1090451311 1090461311 1090471311 1090481311 1090491311 1090501311 1090511311 1090521311 1090531311 1090541311 1090551311 1090561311 1090581311 1090591311 1090601311 1090611311 1090621311 1090631311 1090641311 1090651311 1090661311 1090681311 1090701311 1090721311 1090741311 1090761311 1090801311 1090811311 1090821311 1090841311 1090861311 1090881311 1090891311 1090901311 1090911311 1090921311 1090931311 1090941311 1090951311 1090961311 1090971311 1090981311 1090991311 1100811311 1100831311 1100851311 1100871311 1100891311 1100911311 1100931311 1100951311 1100971311 1100991311 1101011311 1101021311 1101031311 1101041311 1101051311 1101061311 1101071311 1101081311 1101091311 1101101311 1101121311 1101141311 1101161311 1101181311 1101201311 1101221311 1101331311 1101351311 1101371311 1101391311 1101411311 1101431311 1101451311 1101471311 1101491311 1101501311 1101511311 1101521311 1101531311 1101541311 1101551311 1101561311 1101571311 1101591311 1101611311 1101631311 1101651311 1101671311 1101691311 1101711311 1101731311 1101751311 1101771311 1101791311 1101811311 1108011312 1108031312 1108051312 1108071312 1108231312 1108251312 1108271312 1108291312 1108311312 1108331312 1108351312 1108371312 1108391312 1108411312 1108431312 1108451312 1108471312 1108491312 1108511312 1108531312 1108551312 1108571312 1108591312 1108611312 1108631312 1108711312 1108731312 1108751312 1108811312 1108831312 1108851312 1108871312 1108891312 1108911312 1108931312 1108951312 1108971312 1108991312 1109011312 1109031312 1110381310 1110401310 1110421310 1110441310 1110461310 1110481310 1110501310 1110521310 1110541310 1110561310 1110581310 1110601310 1110621310 1110641310 1110661310 1110681310 1110701310 1110721310 1110761310 1110831310 1110851310 1110871310 1110891310 1110981310 1111001310 1111021310 1111031310 1111071310 1111151310 1111171310 1111191310 1111211310 1111231310 1111251310 1111271310 1111291310 1111311310 1111331310 1111381310 1111391310 1111401310 1111411310 1111421310 1111431310 1111441310 1111451310 1111461310 1111471310 1111481310 1111491310 1111501310 1111511310 1111521310 1111531310 1111541310 1111551310 1111561310 1111571310 1111571311 1111581310 1111581311 1111591310 1111591311 1111601310 1111601311 1111611310 1111611311 1111621310 1111621311 1111631310 1111631311 1111641311 1111651310 1111651311 1111661310 1111661311 1111671310 1111681310 1111681311 1111691310 1111701310 1111701311 1111711310 1111721311 1111731310 1111731311 1111741311 1111751311 1111761310 1111761311 1111771310 1111771311 1111781310 1111781311 1111791311 1111801310 1111801311 1111811311 1111821310 1111821311 1111831311 1111841310 1111841311 1111851311 1111861310 1111861311 1111871311 1111881310 1111881311 1111891311 1111901310 1111901311 1111911311 1111921310 1111921311 1111941310 1111941311 1111961310 1111961311 1111981310 1111981311 1112001310 1112001311 1112021310 1112021311 1112041310 1112041311 1112061310 1112061311 1112081310 1112081311 1112101310 1112101311 1112121310 1112121311 1112141310 1112141311 1112161310 1112161311 1112181310 1112181311 1112191311 1112201310 1112201311 1112211311 1112221310 1112221311 1112231311 1112241310 1112241311 1112251311 1112261310 1112261311 1112271310 1112271311 1112281310 1112281311 1112291311 1112301310 1112301311 1112311311 1112321310 1112321311 1112331311 1112341310 1112341311 1112351311 1112361310 1112361311 1112371311 1112381310 1112381311 1112391311 1112401310 1112401311 1112411310 1112411311 1112421310 1112421311 1112431310 1112431311 1112441310 1112441311 1112451310 1112451311 1112461310 1112461311 1112471310 1112471311 1112481310 1112481311 1112491310 1112491311 1112501310 1112511311 1112531311 1112551311 1112571311 1112591311 1112611311 1112631311 1112651311 1112671311 1116751312 1116771312 1116951312 1116991312 1117011312 1117031312 1117051312 1117071312 1117081312 1117101312 1117121312 1117131312 1117141312 1117161312 1117181312 1117201312 1117261312 1117281312 1117301312 1117351312 1117371312 1117391312 1117411312 1117431312 1117451312 1117511312 1118091312 1118111312 1118131312 1118171312 1118191312 1118341312 1118351312 1118361312 1118371312 1118391312 1118431312 1118451312 1118491312 1118551312 1118591312 1118611312 1118631312 1118651312 1118671312 1119281312 1119301312 1119321312 1119341312 1119421312 1119441312 1119461312 1119481312 1119501312 1119581312 1119601312 1119621312 1119641312 1119661312 1119681312 1119701312 1119721312 1119741312 1119761312 1119781312 1119801312 1119821312 1127861310 1127871310 1127881310 1127891310 1127901310 1127911310 1127921310 1127931310 1127941310 1127951310 1127961310 1127971310 1127981310 1127991310 1128001310 1128011310 1128021310 1128031310 1128041310 1128051310 1128061310 1128071310 1128081310 1128091310 1128101310 1128111310 1128121310 1128131310 1128141310 1128151310 1128161310 1128171310 1128181310 1128191310 1128211310 1128231310 1128241310 1128251310 1128271310 1128281310 1128291310 1128301310 1128311310 1128321310 1128331310 1128341310 1128351310 1128361310 1128371310 1128381310 1128391310 1128401310 1128411310 1128421310 1128431310 1128441310 1128451310 1128461310 1128471310 1128481310 1128491310 1128501310 1128511310 1128521310 1128531310 1128541310 1128551310 1128561310 1128571310 1128581310 1128601310 1128621310 1128641310 1128651310 1128661310 1128671310 1128681310 1128691310 1128711310 1128721310 1128731310 1128741310 1128751310 1128761310 1128771310 1128781310 1128791310 1128801310 1128811310 1128821310 1128831310 1128841310 1128851310 1128861310 1128871310 1128881310 1128891310 1128901310 1128911310 1128921310 1128931310 1128941310 1128951310 1128961310 1128981310 1135791309 1135801309 1135811309 1135821309 1135831309 1135841309 1135851309 1135861309 1135871309 1135881309 1135891309 1135901309 1135911309 1135921309 1135931309 1135941309 1135961309 1135981309 1136001309 1136021309 1136041309 1136061309 1136081309 1136101309 1136121309 1136131309 1136141309 1136201309 1136241309 1136281309 1136341309 1136361309 1136381309 1136401309 1136421309 1136441309 1136461309 1136481309 1136501309 1136521309 1136541309 1136561309 1136581309 1136601309 1136621309 1136641309 1136661309 1136681309 1136721309 1136741309 1136761309 1136781309 1136801309 1136821309 1149131309 1149151309 1149171309 1149191309 1149211309 1149231309 1149251309 1149271309 1149291309 1149311309 1149331309 1149351309 1149361309 1149371309 1149391309 1149411309 1149431309 1149441309 1149451309 1149461309 1149471309 1149481309 1149491309 1149501309 1149511309 1149521309 1149531309 1149541309 1149551309 1149561309 1149571309 1149581309 1149591309 1149601309 1149611309 1149621309 1149631309 1149641309 1149651309 1149661309 1149671309 1149681309 1149691309 1149701309 1149711309 1149721309 1149731309 1149741309 1149751309 1149761309 1149771309 1149781309 1149791309 1149801309 1149811309 1149821309 1149891309 1150051309 1150071309 1150091309 1167481309 1167491309 1167501309 1167511309 1167521309 1167531309 1167541309 1167551309 1167561309 1167571309 1167581309 1167591309 1167601309 1167611309 1167621309 1167631309 1167641309 1167651309 1167661309 1167671309 1167681309 1167691309 1167701309 1167711309 1167721309 1167731309 1167741309 1167751309 1167761309 1167771309 1167781309 1167791309 1167801309 1167811309 1167821309 1167831309 1167841309 1167851309 1167861309 1167871309 1167881309 1167891309 1167901309 1167911309 1167921309 1167931309 1167941309 1167951309 1167961309 1167971309 1167981309 1167991309 1168001309 1168011309 1168021309 1168031309 1168041309 1168051309 1168061309 1168071309 1168081309 1168091309 1168101309 1168121309 1168141309 1168161309 1168181309 1168201309 1168221309 1168241309 1168261309 1168281309 1168301309 1168321309 1168341309 1168361309 1168381309 1168401309 1168421309 1172751309 1172761309 1172771309 1172781309 1172791309 1172801309 1172811309 1172821309 1172831309 1172841309 1172851309 1172861309 1172871309 1172881309 1172891309 1172901309 1172911309 1172921309 1172931309 1172941309 1172951309 1172961309 1172971309 1172981309 1172991309 1173001309 1173011309 1173021309 1173031309 1173041309 1173051309 1173061309 1173071309 1173081309 1173091309 1173101309 1173111309 1173121309 1173131309 1173141309 1173151309 1173161309 1173171309 1173181309 1173191309 1173201309 1173211309 1173221309 1173231309 1173241309 1173251309 1173261309 1173271309 1173281309 1173291309 1173301309 1173311309 1173321309 1173331309 1173341309 1173351309 1173361309 1173371309 1173381309 1173391309 1173401309 1173411309 1173421309 1173431309 1173441309 1173451309 1173461309 1173471309 1173481309 1173491309 1173501309 1173511309 1173521309 1173531309 1173541309 1173551309 1173561309 1173571309 1173581309 1173591309 1173601309 1173611309 1173621309 1173631309 1173631310 1173641309 1173651309 1173651310 1173661309 1173671309 1173671310 1173691309 1173691310 1173701309 1173711309 1173711310 1173721309 1173731309 1173731310 1173751309 1173751310 1173761309 1173771309 1173771310 1173781309 1173791310 1173811310 1173831310 1173851310 1173871310 1173891310 1173911310 1173931310 1173951310 1173971310 1173991310 1174011310 1174031310 1174041310 1174051310 1174061310 1174071310 1174081310 1174091310 1174101310 1174111310 1174121310 1174131310 1174141310 1174151310 1174161310 1174171310 1174181310 1174191310 1174201310 1174211310 1174221310 1174241310 1174261310 1174281310 1174301310 1174321310 1174341310 1174361310 1174381310 1174401310 1174411310 1174421310 1174441310 1174461310 1174471310 1174481310 1174491310 1174501310 1174511310 1174521310 1174531310 1174541310 1174551310 1174561310 1174571310 1174581310 1174591310 1174601310 1174611310 1174621310 1174631310 1174641310 1174651310 1174661310 1174671310 1174681310 1178901310 1178911310 1178921310 1178931310 1178941310 1178951310 1178961310 1178971310 1178981310 1178991310 1179001310 1179011310 1179021310 1179031310 1179041310 1179051310 1179061310 1179071310 1179081310 1179091310 1179101310 1179111310 1179121310 1179131310 1179141310 1179151310 1179161310 1179171310 1179181310 1179191310 1179201310 1179211310 1179221310 1179231310 1179241310 1179251310 1179261310 1179271310 1179281310 1179291310 1179301310 1179311310 1179321310 1179331310 1179341310 1179351310 1179361310 1179371310 1179381310 1179391310 1179401310 1179411310 1179421310 1179431310 1179441310 1179451310 1179461310 1179471310 1179481310 1179491310 1179501310 1179511310 1179521310 1179531310 1179541310 1179551310 1179561310 1179571310 1179581310 1179591310 1179601310 1179611310 1179621310 1179631310 1179641310 1179651310 1179661310 1179671310 1179681310 1179691310 1179701310 1179711310 1179721310 1179731310 1179741310 1179751310 1179761310 1179771310 1179781310 1179791310 1179801310 1179811310 1179821310 1179831310 1179841310 1179851310 1179861310 1179871310 1179881310 1179891310 1179901310 1179911310 1179921310 1179931310 1179941310 1179951310 1179961310 1179971310 1179981310 1179991310 1181771310 1181781310 1181791310 1181801310 1181811310 1181821310 1181831310 1181841310 1181851310 1181861310 1181871310 1181881310 1181891310 1181901310 1181911310 1181921310 1181931310 1181941310 1181951310 1181961310 1181971310 1181981310 1181991310 1182001310 1182011310 1182021310 1182031310 1182041310 1182051310 1182061310 1182071310 1182081310 1182091310 1182101310 1182111310 1182121310 1182131310 1182141310 1182151310 1182161310 1182171310 1182181310 1182191310 1182201310 1182211310 1182221310 1182231310 1182241310 1182251310 1182261310 1182271310 1182281310 1182291310 1182301310 1182311310 1182321310 1182331310 1182341310 1182351310 1182361310 1182371310 1182381310 1182391310 1182401310 1182411310 1182421310 1182431310 1182441310 1182451310 1182461310 1182471310 1182481310 1182491310 1182501310 1182511310 1182521310 1182531310 1182541310 1182551310 1182561310 1182571310 1182581310 1182591310 1182601310 1182611310 1182621310 1182631310 1182641310 1182651310 1182661310 1182671310 1182681310 1187061309 1187081309 1187091309 1187101309 1187121309 1187131309 1187141309 1187161309 1187171309 1187181309 1187201309 1187211309 1187221309 1187241309 1187251309 1187261309 1187281309 1187291309 1187321309 1187331309 1187341309 1187371309 1187381309 1187411309 1187421309 1187441309 1187451309 1187491309 1187521309 1187531309 1187541309 1187561309 1187571309 1187581309 1187601309 1187611309 1187621309 1187641309 1187651309 1187681309 1187691309 1187711309 1187731309 1187741309 1187751309 1187771309 1187791309 1187811309 1187821309 1187831309 1187851309 1187861309 1187871309 1187891309 1187901309 1187911309 1187931309 1187941309 1187951309 1187971309 1187981309 1187991309 1188011309 1188021309 1188031309 1188051309 1188061309 1188071309 1188081309 1188091309 1188101309 1188111309 1188121309 1188131309 1188141309 1188151309 1188161309 1189261309 1189281309 1189411309 1189431309 1189451309 1189471309 1189481309 1189501309 1189511309 1189521309 1189531309 1189541309 1189551309 1189561309 1189571309 1189581309 1189591309 1189601309 1189611309 1189621309 1189631309 1189641309 1189651309 1189661309 1189671309 1189681309 1189691309 1189701309 1189711309 1189721309 1189731309 1189741309 1189751309 1189761309 1189781309 1189801309 1189821309 1189881309 1189941309 1189991309 1190011309 1190031309 1190051309 1190071309 1190131309 1190151309 1190171309 1190191309 1190271309 1190291309 1190311309 1190331309 1190351309 1201731311 1201741311 1201751311 1201761311 1201771311 1201781311 1201791311 1201801311 1201811311 1201821311 1201831311 1201841311 1201851311 1201861311 1201871311 1201881311 1201891311 1201901311 1201911311 1201921311 1201931311 1201941311 1201951311 1201961311 1201971311 1201981311 1201991311 1202001311 1202011311 1202021311 1202031311 1202041311 1202051311 1202061311 1202071311 1202081311 1202091311 1202101311 1202111311 1202121311 1202131311 1202141311 1202151311 1202161311 1202171311 1202181311 1202191311 1202201311 1202211311 1202221311 1202231311 1202241311 1202251311 1202261311 1202271311 1202281311 1202291311 1202301311 1202311311 1202321311 1202331311 1202341311 1202351311 1202361311 1202371311 1202381311 1202391311 1202401311 1202411311 1202421311 1202431311 1202441311 1202451311 1202461311 1202471311 1202481311 1202491311 1202501311 1202511311 1202521311 1202531311 1202541311 1202551311 1202561311 1202571311 1202581311 1202591311 1202601311 1202611311 1202621311 1202631311 1202641311 1202651311 1202661311 1202671311 1202681311 1202691311 1202701311 1202711311 1202721311 1202731311 1202741311 1202751311 1202761311 1202771311 1202781311 1202791311 1208771311 1208781311 1208791311 1208801311 1208811311 1208821311 1208831311 1208841311 1208851311 1208861311 1208881311 1208901311 1208921311 1208941311 1208961311 1208981311 1209001311 1209041311 1209061311 1209081311 1209101311 1209121311 1209141311 1209161311 1209181311 1209201311 1209221311 1209241311 1209261311 1209281311 1209301311 1209321311 1209341311 1209361311 1209381311 1209401311 1209421311 1209441311 1209461311 1209481311 1209501311 1209521311 1209541311 1209561311 1209581311 1209601311 1209621311 1209631311 1209641311 1209651311 1209661311 1209671311 1209681311 1209691311 1209701311 1209711311 1209721311 1209731311 1209741311 1209751311 1209761311 1209771311 1209781311 1209791311 1209801311 1209811311 1209821311 1209831311 1209841311 1209851311 1209861311 1220741311 1220751311 1220761311 1220771311 1220781311 1220791311 1220801311 1220811311 1220821311 1220831311 1220841311 1220851311 1220861311 1220871311 1220881311 1220891311 1220901311 1220911311 1220921311 1220931311 1220941311 1220951311 1220961311 1220971311 1220981311 1220991311 1221011311 1221031311 1221041311 1221051311 1221061311 1221071311 1221081311 1221091311 1221101311 1221111311 1221121311 1221131311 1221141311 1221161311 1221181311 1221201311 1221221311 1221241311 1221261311 1221281311 1221301311 1221311311 1221321311 1221331311 1221341311 1221351311 1221361311 1221371311 1221381311 1221391311 1221401311 1221411311 1221421311 1221431311 1221441311 1221451311 1221461311 1221471311 1221481311 1221491311 1221501311 1221511311 1221521311 1221531311 1221541311 1221551311 1221561311 1221571311 1221581311 1221591311 1221601311 1221621311 1221641311 1221661311 1221681311 1221701311 1221721311 1221741311 1221761311 1221781311 1221791311 1221801311 1221811311 1221821311 1221831311 1221851311 1221861311 1221871311 1221881311 1221891311 1221901311 1221911311 1221921311 1221931311 1221941311 1221951311 1221961311 1221971311 1221981311 1221991311 1222001311 1222011311 1222021311 1222031311 1222041311 1222051311 1222061311 1222071311 1222081311 1222091311 1222101311 1222111311 1222121311 1222131311 1222141311 1222151311 1222161311 1222171311 1222181311 1222191311 1222201311 1222211311 1222221311 1222231311 1222241311 1222251311 1222261311 1222271311 1222281311 1222291311 1222301311 1222311311 1222321311 1222331311 1222341311 1222351311 1222361311 1222371311 1222381311 1222391311 1222401311 1222411311 1222421311 1222431311 1222441311 1222451311 1222461311 1222471311 1222481311 1222491311 1222501311 1222511311 1222521311 1222531311 1222541311 1222551311 1222561311 1222571311 1222581311 1222591311 1222601311 1222611311 1222621311 1222631311 1222641311 1222651311 1222661311 1222671311 1222681311 1222691311 1222701311 1222711311 1222721311 1222731311 1222741311 1222751311 1222761311 1222771311 1222781311 1229751310 1229761310 1229771310 1229781310 1229791310 1229801310 1229811310 1229821310 1229831310 1229841310 1229851310 1229861310 1229871310 1229881310 1229891310 1229901310 1229911310 1229921310 1229931310 1229941310 1229951310 1229961310 1229971310 1229981310 1229991310 1230001310 1230011310 1230021310 1230031310 1230041310 1230051310 1230061310 1230071310 1230081310 1230091310 1230101310 1230111310 1230121310 1230131310 1230141310 1230151310 1230161310 1230171310 1230181310 1230191310 1230201310 1230211310 1230221310 1230231310 1230241310 1230251310 1230261310 1230271310 1230281310 1230291310 1230301310 1230311310 1230321310 1230331310 1230341310 1230351310 1230361310 1230371310 1230381310 1230391310 1230401310 1230411310 1230421310 1230431310 1230441310 1230451310 1230461310 1230471310 1230481310 1230491310 1230501310 1230511310 1230521310 1230541310 1230551310 1230561310 1230571310 1230581310 1230591310 1230601310 1230611310 1230621310 1230631310 1230641310 1230651310 1230661310 1230671310 1230681310 1230691310 1230701310 1230711310 1230721310 1230731310 1230741310 1230751310 1230761310 1230771310 1230781310 1230791310 1230801310 1230811310 1230821310 1230831310 1230841310 1230851310 1231471311 1231491311 1231501311 1231511311 1231631311 1231671311 1231791311 1231811311 1231831311 1231871311 1231891311 1231901311 1231911311 1231921311 1231931311 1231941311 1231951311 1231961311 1231971311 1231981311 1231991311 1232001311 1232011311 1232021311 1232031311 1232041311 1232051311 1232061311 1232071311 1232081311 1232091311 1232111311 1232131311 1232151311 1232171311 1232191311 1232211311 1232231311 1232251311 1232501311 1232521311 1232541311 1232561311 1239711310 1239721310 1239731310 1239741310 1239751310 1239761310 1239771310 1239781310 1239791310 1239801310 1239811310 1239821310 1239831310 1239841310 1239851310 1239861310 1239871310 1239881310 1239891310 1239901310 1239911310 1239921310 1239931310 1239941310 1239951310 1239961310 1239971310 1239981310 1239991310 1240001310 1240011310 1240021310 1240031310 1240041310 1240051310 1240071310 1240081310 1240091310 1240111310 1240121310 1240131310 1240141310 1240151310 1240161310 1240171310 1240181310 1240191310 1240201310 1240211310 1240221310 1240231310 1240241310 1240251310 1240261310 1240271310 1240281310 1240291310 1240301310 1240311310 1240321310 1240331310 1240341310 1240351310 1240361310 1240371310 1240381310 1240391310 1240401310 1240421310 1240441310 1240461310 1240481310 1240501310 1240521310 1240531310 1240541310 1240551310 1240561310 1240571310 1240581310 1240591310 1240601310 1240611310 1240621310 1249351310 1249361310 1249371310 1249381310 1249391310 1249401310 1249411310 1249421310 1249431310 1249441310 1249451310 1249461310 1249471310 1249481310 1249491310 1249501310 1249511310 1249521310 1249531310 1249541310 1249551310 1249561310 1249571310 1249581310 1249591310 1249601310 1249611310 1249621310 1249631310 1249641310 1249651310 1249661310 1249671310 1249681310 1249691310 1249701310 1249711310 1249721310 1249731310 1249741310 1249751310 1249761310 1249771310 1249781310 1249791310 1249811310 1249831310 1249851310 1249861310 1249871310 1249891310 1249911310 1249931310 1249951310 1249961310 1249971310 1249981310 1249991310 1250001310 1250011310 1250021310 1250031310 1250041310 1250051310 1250061310 1250071310 1250081310 1250091310 1250101310 1250111310 1250121310 1250131310 1250141310 1250151310 1250161310 1250171310 1250181310 1250191310 1250201310 1250211310 1250221310 1250231310 1250241310 1250251310 1250261310 1250271310 1250281310 1250291310 1250301310 1250311310 1250321310 1250331310 1250341310 1250351310 1250361310 1250371310 1250381310 1250391310 1250401310 1250411310 1261181310 1261191310 1261201310 1261211310 1261221310 1261231310 1261241310 1261251310 1261261310 1261271310 1261281310 1261291310 1261301310 1261311310 1261321310 1261331310 1261341310 1261351310 1261361310 1261371310 1261381310 1261391310 1261401310 1261411310 1261421310 1261431310 1261441310 1261451310 1261461310 1261471310 1261481310 1261491310 1261501310 1261511310 1261521310 1261531310 1261541310 1261551310 1261561310 1261571310 1261581310 1261591310 1261601310 1261611310 1261621310 1261631310 1261651310 1261671310 1261691310 1261711310 1261731310 1261751310 1261771310 1261791310 1261801310 1261821310 1261831310 1261841310 1261851310 1261861310 1261871310 1261881310 1261891310 1261901310 1261911310 1261921310 1261931310 1261941310 1261951310 1261961310 1261971310 1261981310 1261991310 1262001310 1262011310 1262021310 1262031310 1262041310 1262051310 1262061310 1262071310 1262081310 1262101310 1262111310 1262121310 1262131310 1262141310 1262151310 1262161310 1262171310 1262181310 1262191310 1262201310 1262211310 1262221310 1262231310 1262241310 1281981308 1281991308 1282001308 1282011308 1282021308 1282031308 1282041308 1282051308 1282061308 1282071308 1282081308 1282091308 1282101308 1282111308 1282121308 1282131308 1282141308 1282151308 1282161308 1282171308 1282181308 1282191308 1282201308 1282211308 1282221308 1282231308 1282241308 1282251308 1282261308 1282271308 1282281308 1282291308 1282301308 1282311308 1282321308 1282331308 1282341308 1282351308 1282361308 1282371308 1282381308 1282391308 1282401308 1282411308 1282421308 1282431308 1282441308 1282451308 1282461308 1282471308 1282481308 1282491308 1282501308 1282511308 1282521308 1282531308 1282541308 1282551308 1282561308 1282571308 1282581308 1282591308 1282601308 1282611308 1282621308 1282631308 1282641308 1282651308 1282661308 1282671308 1282681308 1282691308 1282701308 1282711308 1282721308 1282731308 1282741308 1282751308 1282761308 1282771308 1282781308 1282791308 1282801308 1282811308 1282821308 1282841308 1282851308 1282861308 1282871308 1282881308 1282891308 1282901308 1282911308 1282921308 1282931308 1282941308 1282951308 1282961308 1282971308 1312941308 1312961308 1312981308 1313001308 1313021308 1313041308 1313061308 1313131308 1313151308 1313171308 1313191308 1313201308 1313211308 1313231308 1313251308 1313271308 1313281308 1313291308 1313301308 1313311308 1313321308 1313331308 1313341308 1313351308 1313361308 1313371308 1313381308 1313391308 1313401308 1313411308 1313421308 1313431308 1313441308 1313451308 1313461308 1313471308 1313481308 1313491308 1313501308 1313511308 1313521308 1313531308 1313541308 1313551308 1313561308 1313571308 1313581308 1313591308 1313601308 1313611308 1313621308 1313631308 1313641308 1313651308 1313661308 1313671308 1313681308 1313691308 1313701308 1313711308 1313721308 1313731308 1313741308 1313751308 1313761308 1313771308 1313781308 1313791308 1313801308 1313811308 1313821308 1313831308 1313841308 1313851308 1313861308 1313871308 1313881308 1313891308 1313901308 1313911308 1313921308 1313931308 1313941308 1313951308 1313961308 1313971308 1313981308 1313991308 1314001308 1314011308 1314021308 1314031308 1314041308 1314051308 1314061308 1320371308 1320381308 1320391308 1320401308 1320411308 1320421308 1320431308 1320441308 1320451308 1320461308 1320471308 1320481308 1320491308 1320501308 1320511308 1320521308 1320531308 1320541308 1320551308 1320561308 1320571308 1320581308 1320591308 1320601308 1320611308 1320621308 1320631308 1320641308 1320651308 1320661308 1320671308 1320681308 1320691308 1320701308 1320711308 1320721308 1320731308 1320741308 1320751308 1320761308 1320771308 1320781308 1320791308 1320801308 1320811308 1320821308 1320831308 1320841308 1320851308 1320861308 1320871308 1320881308 1320891308 1320901308 1320911308 1320921308 1320931308 1320941308 1320951308 1320961308 1320971308 1320981308 1320991308 1321001308 1321011308 1321021308 1321031308 1321041308 1321051308 1321061308 1321071308 1321081308 1321091308 1321101308 1321111308 1321121308 1321131308 1321141308 1321151308 1321161308 1321171308 1321181308 1321191308 1321201308 1321211308 1321221308 1321231308 1321241308 1321251308 1321261308 1321271308 1321281308 1321291308 1321301308 1321311308 1321321308 1321331308 1321341308 1321351308 1321361308 1321371308 1321381308 1321391308 1321401308 1321411308 1321421308 1321431308 1321441308 1321451308 1321461308 1321471308 1321481308 1321491308 1321501308 1321511308 1321521308 1005101310 1005111310 1005131310 1005151310 1005161310 1005171310 1005191310 1005211310 1005231310 1005251310 1005261310 1005271310 1005281310 1005291310 1005321310 1005341310 1005361310 1005381310 1005401310 1005421310 1005441310 1005451310 1005461310 1005471310 1005481310 1005491310 1005501310 1005521310 1005531310 1005541310 1005551310 1005561310 1005571310 1005581310 1005591310 1005601310 1005611310 1005621310 1005631310 1005641310 1005651310 1005661310 1005671310 1005681310 1005691310 1005701310 1005711310 1005721310 1005731310 1005741310 1005751310 1005761310 1005771310 1005781310 1005791310 1005801310 1005811310 1005821310 1005831310 1005841310 1005861310 1005881310 1005901310 1005911310 1005921310 1005931310 1005941310 1005951310 1005961310 1005971310 1005981310 1006001310 1006021310 1006041310 1006081310 1006101310 1006121310 1006141310 1010061310 1010071310 1010081310 1010091310 1010101310 1010111310 1010121310 1010131310 1010141310 1010151310 1010161310 1010171310 1010181310 1010191310 1010201310 1010211310 1010221310 1010231310 1010241310 1010251310 1010261310 1010271310 1010281310 1010291310 1010301310 1010311310 1010321310 1010331310 1010341310 1010351310 1010361310 1010371310 1010381310 1010391310 1010401310 1010411310 1010421310 1010431310 1010441310 1010451310 1010461310 1010471310 1010481310 1010491310 1010501310 1010511310 1010521310 1010531310 1010541310 1010551310 1010561310 1010571310 1010591310 1010601310 1010611310 1010621310 1010631310 1010641310 1010651310 1010661310 1010671310 1010681310 1010691310 1010701310 1010711310 1010721310 1010731310 1010741310 1010751310 1010761310 1010771310 1010781310 1010791310 1010801310 1010811310 1010821310 1010831310 1010841310 1010851310 1010871310 1010881310 1010891310 1010901310 1010911310 1010921310 1010931310 1010941310 1010951310 1010961310 1010971310 1010981310 1010991310 1011001310 1011011310 1011021310 1011031310 1017341309 1017351309 1017361309 1017371309 1017381309 1017391309 1017401309 1017411309 1017421309 1017431309 1017441309 1017451309 1017461309 1017471309 1017481309 1017501309 1017521309 1017531309 1017551309 1017571309 1017591309 1017611309 1017621309 1017631309 1017641309 1017651309 1017671309 1017691309 1017711309 1017731309 1017751309 1017781309 1017801309 1017821309 1017841309 1017851309 1017861309 1017871309 1017881309 1017891309 1017901309 1017911309 1017921309 1017941309 1017961309 1017981309 1017991309 1018001309 1018021309 1018031309 1018041309 1018051309 1018061309 1018081309 1018091309 1018101309 1018111309 1018121309 1018141309 1018161309 1018171309 1018191309 1018221309 1018231309 1018241309 1018251309 1018261309 1018271309 1018281309 1018291309 1018301309 1018311309 1018321309 1018331309 1018341309 1018351309 1018361309 1018381309 1018401309 1018421309 1018431309 1018441309 1018461309 1023011311 1023111311 1023131311 1023141311 1023151311 1023161311 1023171311 1023191311 1023211311 1023231311 1023251311 1023271311 1023291311 1023311311 1023331311 1023351311 1023371311 1023411311 1028601310 1028621310 1028641310 1028661310 1028681310 1028701310 1028721310 1028741310 1028761310 1028781310 1028801310 1029311310 1029331310 1029351310 1029371310 1029391310 1040361311 1040371311 1040381311 1040391311 1040401311 1040411311 1040421311 1040431311 1040441311 1040451311 1040461311 1040471311 1040481311 1040491311 1040501311 1040511311 1040521311 1040531311 1040541311 1040551311 1040561311 1040571311 1040581311 1040591311 1040601311 1040611311 1040621311 1040631311 1040641311 1040661311 1040671311 1040681311 1040701311 1040721311 1040731311 1040741311 1040751311 1040771311 1040791311 1040811311 1040831311 1040851311 1040861311 1040871311 1040881311 1040891311 1040901311 1040911311 1040921311 1040931311 1040941311 1040951311 1040961311 1040971311 1040981311 1040991311 1041001311 1041011311 1041021311 1041031311 1041041311 1041051311 1041061311 1041071311 1041081311 1041091311 1041101311 1041111311 1041121311 1041131311 1041151311 1041171311 1041191311 1041211311 1041231311 1041251311 1041271311 1041291311 1041311311 1041331311 1041351311 1041371311 1041391311 1041411311 1043821310 1043841310 1043861310 1043881310 1043901310 1043921310 1043941310 1043961310 1044041310 1044061310 1044081310 1044101310 1044121310 1044141310 1044161310 1044181310 1044201310 1044221310 1044241310 1044251310 1044261310 1044271310 1044281310 1044291310 1044311310 1044331310 1044351310 1044371310 1044391310 1044401310 1044411310 1044421310 1044431310 1044441310 1044451310 1044461310 1044471310 1044481310 1044491310 1044501310 1044511310 1044521310 1044531310 1044541310 1044551310 1044561310 1044571310 1044581310 1044591310 1044601310 1044621310 1044641310 1044661310 1044671310 1044681310 1044691310 1044701310 1044711310 1044721310 1044731310 1044741310 1044751310 1044761310 1044771310 1044781310 1044791310 1044801310 1044811310 1044821310 1044831310 1044841310 1044851310 1044861310 1052351310 1052371310 1052391310 1052411310 1052431310 1053141310 1053181310 1053201310 1053221310 1056511310 1056521310 1056531310 1056541310 1056551310 1056561310 1056571310 1056581310 1056591310 1056601310 1056611310 1056621310 1056631310 1056641310 1056651310 1056651311 1056661310 1056671310 1056671311 1056681310 1056691310 1056691311 1056701310 1056711310 1056711311 1056721310 1056731310 1056731311 1056741310 1056751310 1056751311 1056761310 1056771310 1056771311 1056781310 1056791310 1056791311 1056801310 1056811310 1056811311 1056821310 1056831310 1056831311 1056841310 1056851310 1056851311 1056861310 1056871310 1056871311 1056881310 1056881311 1056891310 1056891311 1056901310 1056911310 1056911311 1056921310 1056931310 1056931311 1056941310 1056951310 1056951311 1056961310 1056971310 1056971311 1056981310 1056991310 1056991311 1057001310 1057011310 1057021310 1057031310 1057041310 1057051310 1057061310 1057071310 1057081310 1057091310 1057101310 1057111310 1057121310 1057131310 1057141310 1057151310 1057161310 1057171310 1057181310 1057191310 1057201310 1057211310 1057221310 1057231310 1057241310 1057251310 1057261310 1057271310 1057281310 1057291310 1057301310 1057311310 1057321310 1057331310 1057341310 1057351310 1057361310 1057371310 1057381310 1057391310 1057401310 1057411310 1057421310 1057431310 1057441310 1057451310 1057461310 1057471310 1057481310 1057491310 1057501310 1057511310 1057521310 1057531310 1057541310 1057551310 1057561310 1057571310 1057581310 1057591310 1057601310 1061921311 1061941311 1062081311 1062111311 1062131311 1062141311 1062151311 1062161311 1062171311 1062181311 1062231311 1062251311 1062271311 1062281311 1062291311 1062301311 1062311311 1062321311 1062331311 1062341311 1062351311 1062361311 1062371311 1062381311 1062391311 1062401311 1062411311 1062421311 1062431311 1062441311 1062451311 1062461311 1062471311 1062481311 1062491311 1062501311 1062521311 1062541311 1062561311 1062581311 1062601311 1062621311 1062641311 1062661311 1062681311 1064971309 1064981309 1064991309 1065001309 1065011309 1065021309 1065031309 1065041309 1065051309 1065061309 1065061311 1065071309 1065081309 1065081311 1065091309 1065101309 1065101311 1065111309 1065121309 1065131309 1065141309 1065151309 1065161309 1065171309 1065181309 1065191309 1065201309 1065211309 1065221309 1065231309 1065241309 1065251309 1065261309 1065261311 1065271309 1065281309 1065281311 1065291309 1065301309 1065301311 1065311309 1065311311 1065321309 1065331309 1065331311 1065341309 1065351309 1065351311 1065361309 1065361311 1065371309 1065371311 1065381309 1065381311 1065391309 1065391310 1065391311 1065401309 1065401311 1065411309 1065411310 1065421309 1065421311 1065431309 1065431310 1065441309 1065451309 1065451310 1065461309 1065471309 1065471310 1065481309 1065491309 1065491310 1065501309 1065511309 1065511310 1065521309 1065531309 1065531310 1065541309 1065551309 1065551310 1065561309 1065571309 1065571310 1065581309 1065591309 1065591310 1065601309 1065611309 1065611310 1065621309 1065631309 1065631310 1065641309 1065651309 1065651310 1065661309 1065671309 1065671310 1065681309 1065691309 1065691310 1065701309 1065711309 1065711310 1065711311 1065721309 1065731309 1065731310 1065731311 1065741309 1065751309 1065751310 1065761309 1065771309 1065771310 1065781309 1065791309 1065801309 1065801311 1065811309 1065821309 1065831309 1065841309 1065851309 1065861309 1065871309 1065881309 1065891309 1065901309 1065911309 1065921309 1065921311 1065931309 1065941309 1065951309 1065961309 1065971309 1065981309 1065991309 1066001309 1066011309 1066021309 1066031309 1066041309 1066041311 1066051309 1066061309 1066061311 1066071309 1066081309 1066091309 1066101309 1066181310 1066191310 1066201310 1066211310 1066221310 1066231310 1066241310 1066251310 1066261310 1066271310 1066281310 1066291310 1066301310 1066311310 1066321310 1066331310 1066341310 1066351310 1066361310 1066371310 1066381310 1066391310 1066401310 1066411310 1066421310 1066431310 1067511310 1067521310 1067531310 1067541310 1067551310 1067561310 1067571310 1067591310 1067611310 1067631310 1067651310 1067671310 1067691310 1067701310 1067711310 1067721310 1067731310 1067741310 1067751310 1067761310 1067771310 1067781310 1067791310 1067801310 1067811310 1067821310 1067831310 1067841310 1067851310 1067861310 1067871310 1067881310 1067891310 1067901310 1067911310 1067921310 1067931310 1067941310 1067951310 1067971310 1067991310 1068011310 1068031310 1068051310 1068071310 1068091310 1068101310 1068111310 1068121310 1068131310 1068141310 1068151310 1068161310 1068171310 1068181310 1068191310 1068201310 1068211310 1068221310 1068231310 1068241310 1068251310 1068261310 1068271310 1068281310 1068291310 1068301310 1068311310 1068321310 1068331310 1068341310 1068351310 1068361310 1068371310 1068381310 1068391310 1068401310 1068411310 1068421310 1068431310 1068441310 1068451310 1068461310 1068471310 1068481310 1068491310 1068501310 1068511310 1068521310 1068531310 1068541310 1068551310 1068561310 1068571310 1068581310 1068591310 1068601310 1068611310 1068621310 1068631310 1068641310 1070381311 1070401311 1070631311 1070651311 1070671311 1070731311 1070751311 1070771311 1070791311 1070991311 1071011311 1071031311 1071051311 1071071311 1071091311 1071111311 1079201311 1079221311 1079241311 1079261311 1079281311 1079301311 1079321311 1079341311 1079361311 1079381311 1079401311 1079421311 1079441311 1079461311 1079481311 1079501311 1079521311 1079541311 1079571311 1079591311 1079611311 1079631311 1079651311 1079671311 1079691311 1079711311 1079731311 1079751311 1079771311 1079791311 1079811311 1079831311 1079851311 1079871311 1079891311 1079911311 1079931311 1079951311 1079971311 1079991311 1080001311 1080021311 1080031311 1081481309 1081501309 1081521309 1081541309 1081561309 1081581309 1081601309 1081621309 1081641309 1081651309 1081671309 1081691309 1081711309 1081731309 1081751309 1081771309 1081791309 1081811309 1081831309 1081851309 1081871309 1081891309 1081901309 1081911309 1081921309 1081941309 1081961309 1081981309 1082001309 1082611309 1082651309 1082691309 1082731309 1082771309 1082811309 1082851309 1082891309 1082931309 1082971309 1083011309 1083041309 1083051309 1083091309 1083271309 1083311309 1083321309 1083341309 1083351309 1083361309 1083381309 1083401309 1083421309 1083481309 1083501309 1083521309 1083531309 1083541309 1083551309 1083561309 1083571309 1083581309 1083601309 1083621309 1083631309 1083641309 1083661309 1083681309 1087701309 1087711309 1087721309 1087731309 1087741309 1087751309 1087761309 1087771309 1087781309 1087791309 1087801309 1087811309 1087821309 1087831309 1087841309 1087851309 1087861309 1087871309 1087881309 1087891309 1087901309 1087911309 1087921309 1087931309 1087941309 1087951309 1087961309 1087971309 1087981309 1087991309 1088001309 1088011309 1088021309 1088031309 1088041309 1088051309 1088061309 1088071309 1088081309 1088091309 1088101309 1088111309 1088121309 1088131309 1088141309 1088151309 1088161309 1088171309 1088191309 1088211309 1088231309 1088251309 1088261309 1088271309 1088281309 1088291309 1088311309 1088331309 1088351309 1088371309 1088381309 1088391309 1088401309 1088411309 1088421309 1088431309 1088441309 1088451309 1088461309 1088471309 1088481309 1088491309 1088501309 1088511309 1088521309 1088531309 1088541309 1088551309 1088561309 1088571309 1088581309 1088591309 1088601309 1088611309 1088621309 1088631309 1088651309 1088661309 1088671309 1088681309 1088691309 1088701309 1088711309 1088721309 1088731309 1088751309 1088761309 1088771309 1088781309 1101831311 1101851311 1101871311 1101891311 1101911311 1101931311 1101951311 1101971311 1101991311 1102011311 1102031311 1102051311 1102071311 1102091311 1102111311 1102131311 1102151311 1102171311 1102191311 1102211311 1102231311 1102251311 1102271311 1102291311 1102311311 1102331311 1102351311 1102371311 1102391311 1102411311 1102431311 1102451311 1102461311 1102471311 1102481311 1102491311 1102501311 1102511311 1102521311 1102541311 1102561311 1102581311 1102601311 1102681311 1102701311 1102721311 1102741311 1102761311 1102781311 1102801311 1105981312 1105991312 1106001312 1106011312 1106021312 1106031312 1106041312 1106051312 1106061312 1106081312 1106101312 1106121312 1106141312 1106161312 1106181312 1106201312 1106221312 1106241312 1106261312 1106281312 1106301312 1106321312 1106341312 1106361312 1106371312 1106381312 1106401312 1106421312 1106441312 1106461312 1106481312 1106501312 1106511312 1106521312 1106531312 1106541312 1106551312 1106561312 1106571312 1106581312 1106591312 1106601312 1106621312 1106641312 1106651312 1106661312 1106671312 1106681312 1106691312 1106701312 1106711312 1106721312 1106731312 1106751312 1106761312 1106771312 1106781312 1106791312 1106801312 1106811312 1106821312 1106841312 1106851312 1106861312 1106931312 1107311312 1107331312 1107391312 1107811312 1109151312 1109171312 1109191312 1109211312 1109231312 1109251312 1109261310 1109271310 1109271312 1109281310 1109291310 1109291312 1109301310 1109311310 1109311312 1109321310 1109331310 1109331312 1109341310 1109341312 1109351310 1109351312 1109361310 1109371310 1109371312 1109381310 1109391310 1109391312 1109401310 1109411310 1109411312 1109421310 1109421312 1109431310 1109431312 1109441312 1109451310 1109451312 1109461310 1109461312 1109471310 1109471312 1109481310 1109491310 1109491312 1109501310 1109511310 1109511312 1109521310 1109531310 1109531312 1109541310 1109551310 1109551312 1109561310 1109571310 1109571312 1109581310 1109591310 1109591312 1109601310 1109611310 1109611312 1109621310 1109631310 1109641310 1109651310 1109661310 1109671310 1109681310 1109691310 1109701310 1109711310 1109721310 1109731310 1109741310 1109751310 1109761310 1109771310 1109791310 1109801310 1109811310 1109821310 1109831310 1109841310 1109841312 1109851310 1109861310 1109871310 1109881310 1109881312 1109891310 1109901310 1109901312 1109911310 1109921310 1109921312 1109931310 1109941310 1109941312 1109951310 1109951312 1109961310 1109961312 1109971310 1109971312 1109981310 1109981312 1109991310 1109991312 1110001310 1110001312 1110011310 1110011312 1110021310 1110021312 1110031310 1110031312 1110041310 1110041312 1110051310 1110051312 1110061310 1110061312 1110071310 1110071312 1110081310 1110091310 1110091312 1110101310 1110111310 1110111312 1110121310 1110121312 1110131310 1110131312 1110141310 1110141312 1110151310 1110151312 1110161310 1110171310 1110171312 1110181310 1110191310 1110191312 1110201310 1110211310 1110211312 1110221310 1110231310 1110241310 1110251310 1110261310 1110271310 1110281310 1110291310 1110301310 1110311310 1110321310 1110331310 1110341310 1110351310 1110361310 1110371310 1113621310 1113631310 1113641310 1113651310 1113661310 1113671310 1113681310 1113691310 1113701310 1113711310 1113721310 1113731310 1113741310 1113751310 1113761310 1113771310 1113781310 1113791310 1113801310 1113811310 1113821310 1113831310 1113841310 1113851310 1113861310 1113871310 1113881310 1113891310 1113901310 1113911310 1113921310 1113931310 1113941310 1113951310 1113961310 1113971310 1113981310 1113991310 1114001310 1114011310 1114021310 1114031310 1114041310 1114051310 1114061310 1114071310 1114081310 1114091310 1114101310 1114111310 1114121310 1114131310 1114141310 1114151310 1114161310 1114171310 1114181310 1114191310 1114201310 1114211310 1114221310 1114231310 1114241310 1114251310 1114261310 1114271310 1114281310 1114291310 1114301310 1114311310 1114321310 1114331310 1114341310 1114351310 1114361310 1114371310 1114381310 1114391310 1114401310 1114411310 1114421310 1114431310 1114441310 1114451310 1114461310 1114471310 1114481310 1114491310 1114501310 1114511310 1114521310 1114531310 1114541310 1114551310 1114561310 1114571310 1114581310 1114591310 1114601310 1114611310 1114621310 1114631310 1114641310 1114651310 1114661310 1114671310 1114681310 1114681312 1114691310 1114701310 1114701312 1114711310 1114711312 1114721310 1114721312 1114731310 1114731312 1114741310 1114741312 1114751310 1114751312 1114761310 1114761312 1114771310 1114771312 1114781310 1114781312 1114791312 1114811312 1114831312 1114841312 1114851312 1114861312 1114871312 1114881312 1114891312 1114901312 1114911312 1114921312 1114931312 1114941312 1114951312 1114971312 1114991312 1115001312 1115011312 1115021312 1115031312 1115151312 1115171312 1115191312 1115211312 1115231312 1115271312 1115291312 1115391312 1115411312 1118091310 1118101310 1118111310 1118121310 1118131310 1118141310 1118151310 1118161310 1118171310 1118181310 1118191310 1118201310 1118211310 1118221310 1118231310 1118241310 1118251310 1118261310 1118271310 1118281310 1118291310 1118301310 1118311310 1118321310 1118331310 1118341310 1118351310 1118361310 1118371310 1118381310 1118391309 1118391310 1118401310 1118411309 1118411310 1118421309 1118421310 1118431309 1118431310 1118441309 1118441310 1118451309 1118461309 1118461310 1118471309 1118481309 1118481310 1118491309 1118501309 1118501310 1118511309 1118511310 1118521309 1118521310 1118531309 1118531310 1118541309 1118541310 1118551309 1118551310 1118561309 1118561310 1118571309 1118571310 1118581309 1118581310 1118591309 1118591310 1118601309 1118601310 1118611309 1118611310 1118621309 1118621310 1118631310 1118641309 1118641310 1118651310 1118661309 1118661310 1118671310 1118681309 1118681310 1118691310 1118701309 1118701310 1118711310 1118721309 1118721310 1118731310 1118741309 1118741310 1118751310 1118761309 1118761310 1118771310 1118781309 1118781310 1118791309 1118791310 1118801309 1118801310 1118811309 1118811310 1118821309 1118821310 1118831309 1118831310 1118841310 1118851309 1118851310 1118861310 1118871309 1118871310 1118881310 1118891309 1118891310 1118901310 1118911309 1118911310 1118921310 1118931310 1118941310 1118951309 1118951310 1118961310 1118971309 1118971310 1118981309 1118991309 1118991310 1119001309 1119011310 1119021309 1119031310 1119041309 1119041310 1119051310 1119061309 1119061310 1119081309 1119101309 1119121309 1119141309 1119161309 1119181309 1119201309 1119221309 1119241309 1119261309 1119281309 1128171309 1128191309 1128211309 1128231309 1128251309 1128261309 1128271309 1128281309 1128291309 1128301309 1128311309 1128321309 1128331309 1128341309 1128361309 1128381309 1128391309 1128401309 1128411309 1128431309 1128451309 1128471309 1128491309 1128511309 1128531309 1128551309 1128571309 1128591309 1128601309 1128611309 1128621309 1128631309 1128641309 1128651309 1128661309 1128671309 1128681309 1128691309 1128701309 1128711309 1128721309 1128731309 1128741309 1128751309 1128761309 1128771309 1128781309 1128791309 1128801309 1128811309 1128831309 1128841309 1128851309 1128861309 1128871309 1128891309 1128911309 1128991309 1129031309 1129041309 1129051309 1129061309 1129071309 1129081309 1129091309 1129101309 1129111309 1129121309 1129131309 1129141309 1129151309 1129161309 1129171309 1129181309 1129191309 1129201309 1129221309 1129241309 1129251309 1130231309 1130241309 1130251309 1130261309 1130271309 1130281309 1130291309 1130301309 1130311309 1130321309 1130331309 1130341309 1130351309 1130361309 1130371309 1130381309 1130391309 1130401309 1130411309 1130421309 1130431309 1130441309 1130451309 1130461309 1130471309 1130481309 1130491309 1130501309 1130511309 1130521309 1130531309 1130541309 1130551309 1130561309 1130571309 1130581309 1130591309 1130601309 1130611309 1130621309 1130631309 1130641309 1130651309 1130671309 1130681309 1130691309 1130701309 1130711309 1130721309 1130731309 1130741309 1130751309 1130761309 1130771309 1130791309 1130801309 1130811309 1130831309 1130841309 1130851309 1130861309 1130871309 1130881309 1130891309 1130901309 1130911309 1130921309 1130931309 1130941309 1130951309 1130961309 1130971309 1130981309 1130991309 1131011309 1131021309 1131031309 1131051309 1131061309 1131071309 1131081309 1131091309 1131101309 1131111309 1131121309 1131131309 1131141309 1131151309 1131161309 1131171309 1131181309 1131191309 1131211309 1131221309 1131231309 1131251309 1131261309 1131271309 1131291309 1131301309 1131311309 1131331309 1131341309 1147151311 1147171311 1147191311 1147211311 1147231311 1147251311 1147271311 1147291311 1147311311 1147331311 1147351311 1147371311 1147391311 1147411311 1147431311 1147441311 1147451311 1147461311 1147471311 1147491311 1147501311 1147511311 1147531311 1147541311 1147551311 1147571311 1147581311 1147591311 1147611311 1147621311 1147631311 1147651311 1147661311 1147671311 1147691311 1147701311 1147711311 1147731311 1147741311 1147751311 1147761311 1147771311 1147781311 1147791311 1147801311 1147811311 1147821311 1147831311 1147841311 1147851311 1147861311 1147871311 1147881311 1147891311 1147901311 1147911311 1147921311 1147931311 1147941311 1147951309 1147951311 1147961309 1147961311 1147971309 1147971311 1147981309 1147981311 1147991309 1148001309 1148001311 1148011309 1148021309 1148021311 1148031309 1148041309 1148041311 1148051309 1148061309 1148061311 1148071309 1148081309 1148081311 1148091309 1148101309 1148101311 1148111309 1148111311 1148121309 1148121311 1148131309 1148141309 1148141311 1148151309 1148161309 1148161311 1148171309 1148181309 1148181311 1148191309 1148191311 1148201309 1148201311 1148221309 1148221311 1148241309 1148261309 1148281309 1148301309 1148321309 1148341309 1148361309 1148401309 1148511309 1148531309 1148541309 1148551309 1148561309 1148571309 1148581309 1148591309 1148601309 1148611309 1148621309 1148631309 1148641309 1148651309 1148671309 1148691309 1148711309 1148731309 1148741309 1148751309 1148761309 1148771309 1148781309 1148791309 1148801309 1148821309 1148831309 1148841309 1148851309 1148861309 1148871309 1148881309 1148891309 1148901309 1148911309 1148921309 1148931309 1148941309 1148961309 1148981309 1149001309 1149011309 1149021309 1149031309 1149041309 1149051309 1149061309 1149071309 1149381311 1149401311 1149411311 1149421311 1149431311 1149441311 1149451311 1149461311 1149471311 1149481311 1149491311 1149501311 1149511311 1149521311 1149531311 1149541311 1149551311 1149561311 1149571311 1149581311 1149591311 1149601311 1149611311 1149621311 1149631311 1149641311 1149651311 1149661311 1149671311 1149681311 1149691311 1149701311 1149711311 1149721311 1149731311 1149741311 1149751311 1149761311 1149771311 1149781311 1149791311 1149801311 1149811311 1149821311 1149831311 1149841311 1149861311 1149881311 1149901311 1149911311 1149921311 1149941311 1149961311 1149981311 1150001311 1150021311 1150041311 1150051311 1150071311 1150081311 1150091311 1150101311 1150111311 1150121311 1150131311 1150141311 1150161311 1150171311 1150181311 1150191311 1150201311 1150211311 1150221311 1150231311 1150251311 1150261311 1150271311 1150291311 1150311311 1150331311 1150341311 1150351311 1150361311 1150371311 1150391311 1150401311 1150411311 1151581311 1151601311 1151621311 1151641311 1151661311 1151681311 1151691311 1151701311 1151711311 1151721311 1151731311 1151741311 1151751311 1151761311 1151771311 1151781311 1151791311 1151801311 1151811311 1151821311 1151831311 1151841311 1151851311 1151861311 1151881311 1151891311 1151901311 1151911311 1151921311 1151931311 1151941311 1151951311 1151961311 1151971311 1151991311 1152011311 1152031311 1152051311 1152071311 1152091311 1152111311 1152131311 1152151311 1152171311 1152191311 1152211311 1152231311 1152251311 1152271311 1152291311 1152311311 1152331311 1152341311 1152351311 1152361311 1152371311 1152391311 1152411311 1152431311 1152451311 1152471311 1152491311 1152511311 1152531311 1152551311 1152571311 1152591311 1152601311 1152611311 1152621311 1152631311 1152641311 1152651311 1152661311 1152671311 1152681311 1152691311 1152701311 1152711311 1152721311 1152731311 1152741311 1152751311 1152761311 1152771311 1152781311 1152791311 1152811311 1152831311 1152851311 1152871311 1152891311 1152911311 1152921311 1152931311 1152941311 1152951311 1152961311 1152971311 1152981311 1152991311 1153001311 1153011311 1153021311 1153031311 1153041311 1153051311 1153061311 1153071311 1153081311 1153091311 1153101311 1153111311 1153121311 1153131311 1153141311 1153151311 1153161311 1153171311 1153181311 1153201311 1153211311 1153221311 1153231311 1153241311 1153251311 1153261311 1153271311 1153281311 1153291311 1153301311 1153311311 1153321311 1153331311 1153341311 1153351311 1153361311 1153371311 1153381311 1153391311 1153401311 1153411311 1153421311 1153431311 1153441311 1153451311 1153461311 1153471311 1153481311 1153491311 1153501311 1153511311 1153521311 1153531311 1153541311 1153551311 1153561311 1153581311 1153601311 1153621311 1153631311 1153641311 1153651311 1153661311 1153671311 1153681311 1153691311 1153701311 1153711311 1153721311 1153731311 1153741311 1153761311 1153771311 1153781311 1153791311 1153801311 1153811311 1153821311 1153831311 1153841311 1153851311 1153861311 1153871311 1153881311 1153891311 1153901311 1153911311 1153921311 1153931311 1153941311 1153951311 1153961311 1153971311 1153981311 1153991311 1154001311 1154011311 1154021311 1154031311 1154041311 1154051311 1154061311 1154071311 1154081311 1154091311 1154101311 1154111311 1154121311 1154131311 1154141311 1154151311 1154161311 1154171311 1154181311 1154191311 1154201311 1154211311 1154221311 1154231311 1154241311 1154251311 1154261311 1154271311 1154281311 1154291311 1154301311 1154311311 1154321311 1154331311 1154341311 1154351311 1154361311 1154371311 1154381311 1154391311 1154401311 1154411311 1154421311 1154431311 1154441311 1154451311 1154461311 1154471311 1154481311 1154491311 1154501311 1154511311 1154521311 1154531311 1154541311 1154551311 1154561311 1154571311 1154581311 1154591311 1154601311 1154611311 1154621311 1154631311 1154641311 1154651311 1154661311 1154671311 1154681311 1154691311 1154701311 1154711311 1154721311 1154731311 1154741311 1154751311 1154761311 1154771311 1154781311 1154791311 1154801311 1154811311 1154821311 1154831311 1154841311 1154851311 1154861311 1156111310 1156121310 1156131310 1156141310 1156151310 1156161310 1156171310 1156181310 1156191310 1156201310 1156211310 1156221310 1156231310 1156241310 1156251310 1156261310 1156271310 1156281310 1156291310 1156301310 1156311310 1156321310 1156331310 1156341310 1156351310 1156361310 1156371310 1156381310 1156391310 1156401310 1156411310 1156421310 1156431310 1156441310 1156451310 1156461310 1156471310 1156481310 1156491310 1156501310 1156511310 1156521310 1156531310 1156541310 1156551310 1156561310 1156571310 1156581310 1156591310 1156601310 1156611310 1156621310 1156631310 1156641310 1156651310 1156661310 1156671310 1156681310 1156691310 1156701310 1156711310 1156721310 1156731310 1156741310 1156751310 1156761310 1156771310 1156781310 1156791310 1156801310 1156811310 1156821310 1156831310 1156841310 1156851310 1156861310 1156871310 1156881310 1156891310 1156901310 1156911310 1156921310 1156931310 1156941310 1156951310 1156961310 1156971310 1156981310 1156991310 1157001310 1157011310 1157021310 1157031310 1157041310 1157051310 1157061310 1157071310 1157081310 1157091310 1157101310 1157111310 1157121310 1157131310 1157141310 1157151310 1157161310 1157171310 1157181310 1157191310 1157201310 1157211310 1158031311 1158041311 1158051311 1158061311 1158071311 1158081311 1158091311 1158101311 1158111311 1158121311 1158131311 1158141311 1158151311 1158171311 1158181311 1158191311 1158201311 1158211311 1158221311 1158231311 1158241311 1158251311 1158261311 1158271311 1158281311 1158291311 1158301311 1158311311 1158321311 1158331311 1158341311 1158351311 1158361311 1158371311 1158381311 1158391311 1158401311 1158411311 1158421311 1158431311 1158441311 1158451311 1158461311 1158471311 1158481311 1158491311 1158501311 1158511311 1158521311 1158531311 1158541311 1158551311 1158561311 1158571311 1158581311 1158591311 1158601311 1158611311 1158621311 1158631311 1158641311 1158651311 1158661311 1158671311 1158681311 1158691311 1158701311 1158711311 1158721311 1158731311 1158741311 1158751311 1158761311 1158771311 1158781311 1158791311 1158801311 1158811311 1158821311 1158831311 1158841311 1158851311 1158861311 1158871311 1158881311 1158891311 1158901311 1158911311 1158921311 1158931311 1158941311 1158951311 1158961311 1158971311 1158981311 1158991311 1159001311 1159011311 1159021311 1159031311 1159041311 1159051311 1159061311 1159071311 1159081311 1159091311 1159101311 1159121311 1159131311 1159141311 1159151311 1170061311 1170071311 1170081311 1170091311 1170101311 1170111311 1170121311 1170131311 1170141311 1170151311 1170161311 1170171311 1170181311 1170191311 1170201311 1170211311 1170221311 1170231311 1170241311 1170251311 1170261311 1170271311 1170281311 1170291311 1170301311 1170311311 1170321311 1170331311 1170351311 1170361311 1170371311 1170381311 1170401311 1170411311 1170421311 1170431311 1170441311 1170451311 1170461311 1170471311 1170481311 1170491311 1170501311 1170511311 1170521311 1170531311 1170541311 1170551311 1170561311 1170571311 1170581311 1170591311 1170601311 1170611311 1170621311 1170631311 1170641311 1170651311 1170661311 1170671311 1170681311 1170691311 1170701311 1170711311 1170721311 1170731311 1170741311 1170751311 1170761311 1170771311 1170781311 1170791311 1170801311 1170811311 1170821311 1170831311 1170841311 1170851311 1170861311 1170871311 1170881311 1170891311 1170901311 1170911311 1170921311 1170931311 1170941311 1170951311 1170961311 1170971311 1170981311 1170991311 1171001311 1171011311 1171021311 1171031311 1171041311 1171061311 1171421310 1171431310 1171441310 1171451310 1171461310 1171471310 1171481310 1171491310 1171501310 1171511310 1171521310 1171531310 1171541310 1171551310 1171561310 1171571310 1171581310 1171591310 1171601310 1171621310 1171641310 1171661310 1171681310 1171701310 1171721310 1171741310 1171761310 1171781310 1171801310 1171821310 1171941310 1171971310 1171991310 1172011310 1172021310 1172041310 1172061310 1172081310 1172101310 1172121310 1172141310 1172161310 1172181310 1172201310 1172221310 1172241310 1172261310 1172281310 1172301310 1172321310 1172341310 1172361310 1172381310 1172391310 1172401310 1172411310 1172421310 1172431310 1172441310 1172451310 1172461310 1172471310 1172481310 1176571311 1176581311 1176591311 1176601311 1176611311 1176621311 1176631311 1176641311 1176651311 1176661311 1176671311 1176681311 1176691311 1176701311 1176711311 1176721311 1176731311 1176741311 1176751311 1176761311 1176771311 1176781311 1176791311 1176801311 1176811311 1176821311 1176831311 1176841311 1176851311 1176861311 1176871311 1176881311 1176891311 1176901311 1176911311 1176921311 1176931311 1176941311 1176951311 1176961311 1176971311 1176981311 1176991311 1177001311 1177011311 1177021311 1177031311 1177041311 1177051311 1177061311 1177071311 1177081311 1177091311 1177101311 1177111311 1177121311 1177131311 1177141311 1177151311 1177161311 1177181311 1177201311 1177221311 1177241311 1177261311 1177271311 1177281311 1177291311 1177301311 1177311311 1177321311 1177331311 1177341311 1177351311 1177361311 1177371311 1177381311 1177391311 1177401311 1177411311 1177421311 1177431311 1177441311 1177451311 1177461311 1177471311 1177481311 1177491311 1177501311 1177511311 1177521311 1177531311 1177541311 1177551311 1177561311 1177571311 1177581311 1177591311 1177601311 1177611311 1177621311 1177631311 1177641311 1177651311 1177661311 1177671311 1177681311 1177691311 1177701311 1177711311 1177771310 1177781310 1177791310 1177801310 1177811310 1177821310 1177841310 1177851310 1177861310 1177871310 1177881310 1177891310 1177901310 1177911310 1177921310 1177931310 1177941310 1177951310 1177961310 1177971310 1177981310 1177991310 1178001310 1178011310 1178021310 1178031310 1178041310 1178051310 1178061310 1178071310 1178081310 1178091310 1178101310 1178111310 1178121310 1178131310 1178141310 1178151310 1178161310 1178171310 1178181310 1178191310 1178201310 1178211310 1178221310 1178231310 1178241310 1178251310 1178261310 1178271310 1178281310 1178291310 1178301310 1178311310 1178321310 1178331310 1178341310 1178351310 1178361310 1178371310 1178381310 1178391310 1178401310 1178411310 1178421310 1178431310 1178441310 1178451310 1178461310 1178471310 1178481310 1178491310 1178501310 1178521310 1178541310 1178561310 1178571310 1178581310 1178591310 1178601310 1178611310 1178621310 1178631310 1178641310 1178651310 1178661310 1178671310 1178681310 1178691310 1178701310 1178711310 1178721310 1178731310 1178741310 1178751310 1178761310 1178771310 1178781310 1178791310 1178801310 1178811310 1178821310 1178831310 1178841310 1178851310 1178861310 1178871310 1178881310 1178891310 1186811310 1186821310 1186831310 1186841310 1186851310 1186861310 1186871310 1186881310 1186891310 1186901310 1186911310 1186921310 1186931310 1186941310 1186951310 1186961310 1186971310 1186981310 1186991310 1187001310 1187011310 1187021310 1187031310 1187041310 1187051310 1187061310 1187071310 1187081310 1187101310 1187121310 1187141310 1187161310 1187181310 1187191310 1187201310 1187211310 1187221310 1187231310 1187241310 1187251310 1187261310 1187271310 1187281310 1187291310 1187301310 1187311310 1187321310 1187331310 1187341310 1187351310 1187361310 1187371310 1187381310 1187391310 1187401310 1187411310 1187421310 1187431310 1187441310 1187451310 1187461310 1187471310 1187481310 1187491310 1187501310 1187511310 1187521310 1187531310 1187541310 1187551310 1187561310 1187571310 1187581310 1187591310 1187601310 1187611310 1187621310 1187631310 1187641310 1187651310 1187661310 1187671310 1187681310 1187691310 1187701310 1187711310 1187721310 1187731310 1187741310 1187751310 1187761310 1187771310 1187781310 1187791310 1187801310 1187811310 1187821310 1187831310 1187841310 1187851310 1187861310 1187871310 1187881310 1187891310 1187901310 1187911310 1187921310 1191251311 1191271311 1191291311 1191311311 1191331311 1191351311 1191371311 1191381311 1191391311 1191401311 1191411311 1191421311 1191431311 1191441311 1191451311 1191461311 1191471311 1191481311 1191491311 1191501311 1191511311 1191521311 1191541311 1191561311 1191581311 1191601311 1191621311 1191641311 1191661311 1191681311 1191701311 1191721311 1191741311 1191761311 1191781311 1191801311 1191951311 1191961311 1191971311 1191981311 1191991311 1192001311 1192011311 1192021311 1192031311 1192041311 1192051311 1192061311 1192071311 1192081311 1192091311 1192101311 1192111311 1192121311 1192131311 1192141311 1192151311 1192161311 1192171311 1192181311 1192191311 1192201311 1192211311 1192221311 1192231311 1192241311 1192251311 1192261311 1192271311 1192281311 1192291311 1192301311 1192311311 1192321311 1192331311 1192341311 1192351311 1192361311 1192371311 1192381311 1192401311 1192421311 1192431311 1192441311 1192451311 1192461311 1192471311 1192481311 1192491311 1192501311 1192511311 1192521311 1192531311 1192541311 1192551311 1192561311 1192571311 1192581311 1192591311 1192601311 1192611311 1192621311 1192631311 1192641311 1192651311 1192661311 1192671311 1192681311 1192691311 1192701311 1192711311 1192721311 1192731311 1192741311 1192751311 1192761311 1192771311 1192781311 1192791311 1192801311 1192811311 1192821311 1192831311 1192841311 1192851311 1192861311 1192871311 1192881311 1192891311 1192901311 1192911311 1192921311 1192931311 1192941311 1192951311 1192961311 1192971311 1192981311 1192991311 1193001311 1193011311 1193021311 1193031311 1193041311 1193051311 1193061311 1193071311 1194191311 1194201311 1194211311 1194221311 1194231311 1194241311 1194251311 1194261311 1194271311 1194281311 1194291311 1194301311 1194311311 1194321311 1194331311 1194341311 1194351311 1194361311 1194371311 1194381311 1194391311 1194401311 1194411311 1194421311 1194431311 1194441311 1194451311 1194461311 1194471311 1194481311 1194491311 1194501311 1194511311 1194521311 1194531311 1194541311 1194551311 1194561311 1194571311 1194581311 1194591311 1194601310 1194601311 1194611310 1194611311 1194621310 1194621311 1194631310 1194631311 1194641310 1194641311 1194651310 1194651311 1194661310 1194661311 1194671310 1194671311 1194681310 1194681311 1194691310 1194691311 1194701310 1194701311 1194711310 1194711311 1194721310 1194721311 1194731310 1194731311 1194741310 1194751310 1194751311 1194761310 1194771310 1194771311 1194781310 1194791310 1194791311 1194801310 1194811310 1194811311 1194821310 1194831310 1194831311 1194841310 1194851310 1194851311 1194861310 1194871310 1194871311 1194881310 1194891310 1194891311 1194901310 1194911310 1194911311 1194921310 1194931310 1194941310 1194951310 1194961310 1194971310 1194981310 1194991310 1194991311 1195001310 1195001311 1195011310 1195011311 1195021310 1195021311 1195031310 1195031311 1195041310 1195041311 1195051310 1195051311 1195061310 1195061311 1195071310 1195071311 1195081310 1195081311 1195091310 1195091311 1195101310 1195101311 1195111310 1195111311 1195121310 1195121311 1195131310 1195131311 1195141310 1195141311 1195151310 1195151311 1195161310 1195161311 1195171310 1195171311 1195181310 1195181311 1195191310 1195191311 1195201310 1195201311 1195211310 1195211311 1195221310 1195221311 1195231310 1195231311 1195241310 1195241311 1195251310 1195251311 1195261310 1195261311 1195271310 1195271311 1195281310 1195281311 1195291310 1195301310 1195311310 1195321310 1195331310 1195341310 1195351310 1195361310 1195371310 1195381310 1195391310 1195401310 1195411310 1195421310 1195431310 1195441310 1195451310 1195461310 1195471310 1195481310 1195491310 1195501310 1195511310 1195521310 1195531310 1195541310 1195551310 1195561310 1195571310 1195581310 1195591310 1195601310 1195611310 1195621310 1195631310 1195641310 1195651310 1195661310 1195671310 1195681310 1195691310 1195701310 1195711310 1195721310 1195731310 1195741310 1195751310 1195981309 1196001309 1196011309 1196031309 1196041309 1196051309 1196061309 1196071309 1196081309 1196091309 1196111309 1196131309 1196141309 1196151309 1196151311 1196161309 1196171309 1196171311 1196181309 1196191309 1196191311 1196201309 1196211309 1196211311 1196231309 1196231311 1196251309 1196251311 1196271309 1196271311 1196291309 1196291311 1196301309 1196311309 1196311311 1196331309 1196331311 1196351309 1196351311 1196371309 1196371311 1196391309 1196391311 1196411309 1196411311 1196421309 1196431309 1196431311 1196441309 1196451309 1196451311 1196471309 1196471311 1196481309 1196491309 1196491311 1196501309 1196511309 1196511311 1196531309 1196531311 1196551309 1196551311 1196561309 1196571309 1196571311 1196581309 1196581311 1196591309 1196591311 1196601309 1196601311 1196611309 1196611311 1196621309 1196621311 1196631309 1196631311 1196641311 1196651311 1196661311 1196671311 1196681311 1196691311 1196701311 1196711311 1196721311 1196731311 1196741309 1196751309 1196761309 1196771309 1196781309 1196801309 1196821309 1196831309 1196851309 1196871309 1196891309 1196911309 1196931309 1196951309 1196971309 1197001311 1197021311 1197041311 1197061311 1197081311 1197101311 1197121311 1197141311 1197161311 1197181311 1197201311 1197221311 1197241311 1197261311 1197281311 1203471309 1203491309 1203511309 1203531309 1203551309 1203571309 1203591309 1203611309 1203631309 1203651309 1203671309 1203691309 1203711309 1203731309 1203751309 1203771309 1203791309 1203811309 1203831309 1203851309 1203871309 1203891309 1203911309 1203931309 1203951309 1203971309 1203991309 1204011309 1204031309 1204051309 1204071309 1204091309 1204111309 1204131309 1204151309 1204171309 1204311309 1204321309 1204331309 1204341309 1204351309 1204361309 1204371309 1204381309 1204391309 1204401309 1204411309 1204421309 1204431309 1204441309 1204451309 1204461309 1204471309 1204481309 1210181309 1210191309 1210201309 1210211309 1210221309 1210231309 1210241309 1210251309 1210261309 1210271309 1210281309 1210291309 1210301309 1210311309 1210321309 1210331309 1210341309 1210361309 1210371309 1210381309 1210391309 1210401309 1210411309 1210421309 1210431309 1210441309 1210451309 1210471309 1210481309 1210491309 1210501309 1210511309 1210521309 1210531309 1210541309 1210551309 1210561309 1210571309 1210581309 1210591309 1210601309 1210611309 1210621309 1210631309 1210641309 1210651309 1210661309 1210671309 1210681309 1210691309 1210701309 1210711309 1210721309 1210731309 1210741309 1210751309 1210761309 1210771309 1210781309 1210791309 1210801309 1210811309 1210821309 1210831309 1210841309 1210851309 1210861309 1210871309 1210881309 1210891309 1210901309 1210911309 1210921309 1210931309 1210941309 1210951309 1210961309 1210971309 1210981309 1210991309 1211001309 1211011309 1211021309 1211031309 1211041309 1211051309 1211061309 1211071309 1211081309 1211091309 1211101309 1211111309 1211121309 1211131309 1211141309 1211151309 1211161309 1211171309 1211181309 1227581310 1227591310 1227601310 1227611310 1227621310 1227631310 1227641310 1227651310 1227661310 1227671310 1227681310 1227691310 1227701310 1227711310 1227721310 1227731310 1227741310 1227751310 1227761310 1227771310 1227781310 1227791310 1227801310 1227811310 1227821310 1227831310 1227841310 1227851310 1227861310 1227871310 1227881310 1227891310 1227901310 1227911310 1227921310 1227931310 1227941310 1227951310 1227971310 1227991310 1228011310 1228021310 1228031310 1228041310 1228051310 1228061310 1228071310 1228091310 1228111310 1228131310 1228151310 1228171310 1228181310 1228191310 1228201310 1228211310 1228231310 1228251310 1228261310 1228271310 1228281310 1228291310 1228301310 1228311310 1228321310 1228331310 1228341310 1228351310 1228361310 1228371310 1228381310 1228391310 1228401310 1228411310 1228421310 1228431310 1228441310 1228451310 1228461310 1228471310 1228481310 1228491310 1228501310 1228511310 1228521310 1228531310 1228541310 1241771310 1241851310 1241861310 1241881310 1241921310 1241941310 1241961310 1241981310 1242001310 1242021310 1242381310 1242461310 1242541310 1242621310 1242631310 1242701310 1242731310 1243311311 1243321311 1243331311 1243351311 1243371311 1243391311 1243421311 1243441311 1243461311 1243481311 1243501311 1243521311 1243531311 1243541311 1243551311 1243561311 1243571311 1243581311 1243601311 1243611311 1243621311 1243631311 1243641311 1243651311 1243661311 1243671311 1243681311 1243691311 1243701311 1243711311 1243721311 1243731311 1243741311 1243751311 1243761311 1243771311 1243791311 1243801311 1243811311 1243821311 1243831311 1243841311 1243851311 1243861311 1243871311 1243931311 1243951311 1243971311 1243981311 1244001311 1244011311 1244021311 1244031311 1244041311 1244051311 1244061311 1244071311 1244081311 1244091311 1244101311 1244121311 1244141311 1244151311 1244161311 1244181311 1244201311 1244211311 1244221311 1244241311 1244261311 1244281311 1244331311 1244341311 1244371311 1244381311 1244401311 1244421311 1244441311 1244461311 1244481311 1244501311 1244521311 1244541311 1244561311 1244581311 1244601311 1244621311 1244641311 1244661311 1244781311 1244931311 1244951311 1244971311 1244991311 1245011311 1245031311 1245051311 1245071311 1245091311 1245111311 1245131311 1245151311 1245171311 1245191311 1245211311 1245231311 1245251311 1245271311 1245291311 1245311311 1245331311 1245521311 1245531311 1245541311 1245551311 1245561311 1245571311 1245581311 1245591311 1245601311 1245611311 1245621311 1245631311 1245641311 1245651311 1245661311 1245671311 1245681311 1245701311 1245721311 1245741311 1245761311 1245781311 1245801311 1245821311 1245841311 1245861311 1245871311 1245881311 1245901311 1245921311 1245941311 1245961311 1246041311 1246061311 1246081311 1246101311 1246121311 1246141311 1246161311 1246171311 1246181311 1246191311 1246201311 1246211311 1246221311 1246231311 1246241311 1246251311 1246261311 1246271311 1246291311 1246311311 1246331311 1246391311 1246411311 1246431311 1246561311 1246581311 1246601311 1246621311 1246641311 1246661311 1246671311 1248601311 1248611311 1248621311 1248631311 1248641311 1248651311 1248661311 1248671311 1248681311 1248691311 1248701311 1248711311 1248721311 1248741311 1248761311 1248781311 1248801311 1248821311 1248841311 1248861311 1248881311 1248901311 1248911311 1248921311 1248931311 1248941311 1248951311 1248961311 1248971311 1248981311 1249001311 1249011311 1249031311 1249051311 1249071311 1249341311 1249401311 1249411311 1249421311 1249441311 1249461311 1249471311 1249481311 1249491311 1249501311 1249511311 1249521311 1249531311 1249541311 1249551311 1249561311 1249571311 1249581311 1249591311 1249601311 1249611311 1249621311 1249631311 1249641311 1249651311 1249661311 1249671311 1256961310 1257001310 1257021310 1257051310 1257071310 1257091310 1257111310 1257131310 1257151310 1257161310 1257171310 1257191310 1257201310 1257211310 1257231310 1257251310 1257271310 1257281310 1257291310 1257301310 1257311310 1257321310 1257331310 1257341310 1257351310 1257361310 1257371310 1257381310 1257401310 1257421310 1257441310 1257461310 1257481310 1257501310 1257521310 1257541310 1257561310 1257641310 1257681310 1257721310 1257761310 1257781310 1257801310 1257821310 1257831310 1257851310 1257871310 1257891310 1257901310 1257911310 1257931310 1257951310 1257961310 1257971310 1257981310 1257991310 1258001310 1258011310 1268511308 1268521308 1268531308 1268541308 1268551308 1268561308 1268571308 1268581308 1268591308 1268601308 1268611308 1268621308 1268631308 1268641308 1268651308 1268661308 1268671308 1268681308 1268691308 1268701308 1268711308 1268721308 1268731308 1268741308 1268751308 1268761308 1268771308 1268781308 1268791308 1268801308 1268811308 1268821308 1268831308 1268841308 1268851308 1268861308 1268871308 1268881308 1268891308 1268901308 1268911308 1268921308 1268931308 1268941308 1268951308 1268961308 1268971308 1268981308 1268991308 1269001308 1269011308 1269021308 1269031308 1269041308 1269051308 1269061308 1269071308 1269081308 1269091308 1269111308 1269131308 1269141308 1269151308 1269161308 1269171308 1269181308 1269191308 1269201308 1269211308 1269221308 1269231308 1269241308 1269251308 1269261308 1269271308 1269291308 1269311308 1269331308 1269351308 1269371308 1269391308 1269411308 1269431308 1269451308 1269471308 1269491308 1269511308 1269531308 1269541308 1269551308 1269561308 1269571308 1269581308 1276921308 1276931308 1276941308 1276951308 1276961308 1276971308 1276981308 1276991308 1277001308 1277011308 1277021308 1277031308 1277041308 1277051308 1277061308 1277071308 1277081308 1277091308 1277101308 1277111308 1277121308 1277131308 1277141308 1277151308 1277161308 1277171308 1277181308 1277191308 1277201308 1277211308 1277221308 1277231308 1277241308 1277251308 1277261308 1277271308 1277281308 1277291308 1277301308 1277311308 1277321308 1277331308 1277341308 1277351308 1277361308 1277371308 1277381308 1277391308 1277401308 1277411308 1277421308 1277431308 1277441308 1277451308 1277461308 1277471308 1277481308 1277491308 1277501308 1277511308 1277521308 1277531308 1277541308 1277551308 1277561308 1277571308 1277581308 1277591308 1277601308 1277611308 1277621308 1277631308 1277641308 1277651308 1277661308 1277671308 1277681308 1277691308 1277701308 1277711308 1277721308 1277731308 1277741308 1277751308 1277761308 1277771308 1277781308 1277791308 1277801308 1277811308 1277821308 1277831308 1277841308 1277851308 1277861308 1277871308 1277881308 1277891308 1277901308 1277911308 1277921308 1277931308 1277941308 1277951308 1277961308 1277971308 1277981308 1277991308 1278001308 1278011308 1278021308 1278031308 1278041308 1305431308 1305441308 1305451308 1305461308 1305471308 1305481308 1305491308 1305501308 1305511308 1305521308 1305531308 1305541308 1305551308 1305561308 1305571308 1305581308 1305591308 1305601308 1305611308 1305621308 1305631308 1305641308 1305651308 1305661308 1305671308 1305681308 1305691308 1305701308 1305711308 1305721308 1305731308 1305741308 1305751308 1305761308 1305771308 1305781308 1305791308 1305801308 1305811308 1305821308 1305831308 1305841308 1305851308 1305861308 1305871308 1305881308 1305891308 1305901308 1305911308 1305921308 1305931308 1305941308 1305951308 1305961308 1305971308 1305981308 1305991308 1306001308 1306011308 1306021308 1306031308 1306041308 1306051308 1306061308 1306071308 1306081308 1306091308 1306101308 1306111308 1306121308 1306131308 1306141308 1306151308 1306161308 1306171308 1306181308 1306191308 1306201308 1306211308 1306221308 1306231308 1306241308 1306251308 1306261308 1306271308 1306281308 1306291308 1306301308 1306311308 1306321308 1306331308 1306341308 1306351308 1306361308 1306371308 1306381308 1306391308 1306401308 1306411308 1306421308 1306431308 1308701308 1308711308 1308721308 1308731308 1308741308 1308751308 1308761308 1308771308 1308781308 1308791308 1308801308 1308811308 1308821308 1308831308 1308851308 1308871308 1308891308 1308911308 1308931308 1308941308 1308951308 1308971308 1308991308 1309011308 1309021308 1309031308 1309041308 1309051308 1309061308 1309071308 1309081308 1309091308 1309101308 1309111308 1309131308 1309141308 1309151308 1309161308 1309171308 1309191308 1309201308 1309211308 1309221308 1309231308 1309241308 1309251308 1309261308 1309271308 1309281308 1309291308 1309301308 1309311308 1309321308 1309331308 1309341308 1309351308 1309361308 1309381308 1309401308 1309421308 1309441308 1309461308 1309481308 1309501308 1309521308 1309541308 1309561308 1309581308 1309601308 1309621308 1309641308 1309661308 1309681308 1309691308 1309701308 
Recalling Firm/
Manufacturer
AMO Puerto Rico Manufacturing, Inc.
Rd # 402, Km 4.2
Anasco PR 00610
Manufacturer Reason
for Recall
Units may be labeled with the incorrect diopter power.
FDA Determined
Cause 2
Error in labeling
ActionAMO sent an Urgent Medical Device Recall dated May 13, 2014 to the affected customers. The letter identified the affected product, problem and actions to be taken. This recall only affects the specific Silicone Tecnis IOLs Model Z9002. Implanted IOLs with appropriate post-op refractive outcomes are not affected by this recall. Please underatake the following actions: 1. Stop using and remove from your inventory the affected product. 2.Complete and return the included Facsimile Form as AMO. If you have inventory to return, please complete the Facsimile Form, and returned to AMO. For questions contact AMO Customer Service Representative 877-266-4543.
Quantity in Commerce77,055 IOLs in total
DistributionWorldwide Distribution - Nationwide (USA and Puerto Rico), and foreign: Latin America, Asia Pacific, Canada, Europe, Africa and Middle East, and Japan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = HQL
-
-