Date Initiated by Firm |
May 13, 2014 |
Date Posted |
August 26, 2014 |
Recall Status1 |
Terminated 3 on May 18, 2017 |
Recall Number |
Z-2284-2014 |
Recall Event ID |
68346 |
PMA Number |
P880081 |
Product Classification |
intraocular lens - Product Code HQL
|
Product |
AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the visual correction of aphakia in adult patients. Tecnis CL Mono 13 mm with the following diopter sizes: 23, 23.5 & 24.
Product Usage: AMO TECNIS CL monofocal foldable IOLs with OptiEdge design are ultraviolet-absorbing posterior chamber intraocular lenses which compensate for corneal spherical aberrations. TECNIS CL Silicone Intraocular Lenses are indicated for primary implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction or phacoemulsification. These devices are intended to be placed in the capsular bag.
|
Code Information |
Serial Num 1006151310 1006191310 1006211310 1006231310 1006251310 1006271310 1006291310 1006311310 1006331310 1006351310 1006371310 1006371312 1006391310 1006391312 1006411310 1006411312 1006431310 1006431312 1006451310 1006451312 1006471310 1006471312 1006481310 1006491310 1006491312 1006501310 1006511310 1006511312 1006521310 1006531310 1006531312 1006541310 1006551310 1006561310 1006571310 1006591310 1006601310 1006611310 1006621310 1006631310 1006641310 1006651310 1006661310 1006671310 1006681310 1006691310 1006701310 1006711310 1006721310 1006731310 1006741310 1006751310 1006761310 1006771310 1006781310 1006791310 1006801310 1006811310 1006821310 1006831310 1006841310 1006851310 1006871310 1006881310 1006891310 1006901310 1006911310 1006921310 1006931310 1006931312 1006941310 1006951310 1006951312 1006961310 1006971310 1006981310 1007001310 1007221312 1007241312 1007261312 1007281312 1007301312 1007321312 1007341312 1007351312 1007361312 1007371312 1007381312 1007401312 1007421312 1007441312 1008211310 1008231310 1008351310 1008371310 1008391310 1008411310 1008431310 1008451310 1008471310 1008491310 1008511310 1008531310 1008541310 1008551310 1008561310 1008571310 1008581310 1008591310 1008601310 1008611310 1008621310 1008631310 1008641310 1008651310 1008661310 1008681310 1008701310 1008721310 1008741310 1008761310 1008781310 1008801310 1008821310 1008841310 1008861310 1008881310 1009001310 1009021310 1009041310 1009311309 1009391309 1009411309 1009431309 1009451309 1009521309 1009541309 1009551309 1009571309 1009591309 1009611309 1009631309 1009641309 1009651309 1009661309 1009671309 1009681309 1009701309 1009721309 1009771309 1009791309 1009811309 1009831309 1009851309 1009871309 1009891309 1009911309 1009941309 1009961309 1010011309 1010021309 1010031309 1010041309 1010051309 1010061309 1010071309 1010081309 1010091309 1010101309 1010111309 1010131309 1010151309 1010171309 1010181309 1010201309 1010211309 1010221309 1010231309 1010241309 1010251309 1010271309 1010281309 1010291309 1010301309 1010311309 1010321309 1010331309 1010341309 1010351309 1010361309 1010381309 1012701312 1012741312 1012781312 1012821312 1012861312 1012901312 1012941312 1012981312 1013021312 1013061312 1013081312 1013101312 1013121312 1013141312 1013161312 1013181312 1013201312 1013221312 1013261312 1013301312 1013341312 1013381312 1013421312 1013461312 1013471312 1013491312 1013501312 1013511312 1013531312 1013541312 1013551312 1013561312 1013571312 1013581312 1013591312 1013601312 1013611312 1013621312 1013631312 1013641312 1013661312 1013681312 1013691312 1013701312 1013721312 1013731312 1013741312 1013761312 1013771312 1015621311 1015641311 1015661311 1015681311 1015701311 1015721311 1015741311 1015761311 1015781311 1015801311 1015821311 1015841311 1015861311 1015881311 1015901311 1015921311 1015941311 1015961311 1015981311 1017651311 1017661311 1017671311 1017681311 1017691311 1017701311 1017711311 1017721311 1017731311 1017741311 1017751311 1017761311 1017771311 1017781311 1017791311 1017801311 1017811311 1017821311 1017831311 1017851311 1017871311 1018001311 1018021311 1018041311 1018061311 1018071311 1018081311 1018091311 1018101311 1018111311 1018121311 1018131311 1018141311 1018151311 1018161311 1018171311 1018181311 1018191311 1018201311 1018211311 1018221311 1018231311 1018241311 1018251311 1018261311 1018271311 1018281311 1018291311 1018301311 1018311311 1018321311 1018331311 1018341311 1018351311 1018361311 1018371311 1018381311 1018391311 1018401311 1018411311 1018421311 1018431311 1018441311 1018451311 1018461311 1018471311 1018481311 1018491311 1018501311 1018511311 1018521311 1018531311 1018541311 1018551311 1024161311 1024181311 1024191311 1024201311 1024211311 1024221311 1024241311 1024251311 1024271311 1024291311 1024311311 1024331311 1024351311 1024371311 1024391311 1024411311 1024461311 1024481311 1024501311 1024511311 1024521311 1024541311 1024761311 1024781311 1024851311 1024861311 1024871311 1024881311 1024891311 1024901311 1024911311 1024921311 1024931311 1024941311 1024951311 1024961311 1024971311 1024981311 1024991311 1025001311 1025011311 1025021311 1025031311 1025041311 1025051311 1025091311 1025111311 1025121311 1025131311 1025151311 1025171311 1025191311 1025211311 1029611312 1029781312 1029791312 1029891312 1029911312 1029931312 1029951312 1029971312 1030731311 1030741311 1030751311 1030761311 1030771311 1030781311 1030791311 1030801311 1030811311 1030821311 1030831311 1030841311 1030851311 1030861311 1030871311 1030881311 1030891311 1030901311 1030911311 1030921311 1030931311 1030941311 1030951311 1030961311 1030971311 1030981311 1030991311 1031001311 1031011311 1031021311 1031031311 1031041311 1031051311 1031061311 1031071311 1031081311 1031091311 1031101311 1031111311 1031121311 1031131311 1031141311 1031151311 1031161311 1031171311 1031181311 1031191311 1031201311 1031211311 1031221311 1031231311 1031241311 1031251311 1031261311 1031271311 1031281311 1031291311 1031301311 1031311311 1031321311 1031331311 1031341311 1031351311 1031361311 1031371311 1031381311 1031391311 1031401311 1031411311 1031421311 1031431311 1031441311 1031451311 1031461311 1031471311 1031481311 1031491311 1031501311 1031511311 1031521311 1031531311 1031541311 1031551311 1031561311 1031571311 1031581311 1031591311 1031601311 1031611311 1031621311 1031631311 1031641311 1031651311 1031661311 1031671311 1031681311 1031691311 1031701311 1031711311 1031721311 1031731311 1035371309 1035391309 1035481309 1035511309 1035531309 1035551309 1035581309 1035601309 1035621309 1035641309 1035651309 1035661309 1035671309 1035681309 1035691309 1035711309 1035721309 1035731309 1035741309 1035781309 1035801309 1035811309 1035831309 1035841309 1035851309 1035861309 1035871309 1035881309 1035891309 1035901309 1035911309 1035931309 1035941309 1035951309 1035971309 1035981309 1036131309 1036151309 1037401312 1037421312 1037441312 1037461312 1037481312 1037501312 1037521312 1037541312 1037561312 1037581312 1037601312 1037621312 1037631312 1037641312 1037661312 1037671312 1037681312 1037691312 1037701312 1037721312 1037731312 1037751312 1037771312 1037791312 1037811312 1037831312 1038001312 1038021312 1038041312 1038061312 1038081312 1038101312 1038121312 1038141312 1038161312 1038171312 1038181312 1038211312 1038231312 1038251312 1038271312 1038291312 1038451312 1038471312 1038491312 1058331309 1058341309 1058351309 1058361309 1058371309 1058381309 1058391309 1058401309 1058411309 1058421309 1058431309 1058441309 1058451309 1058461309 1058471309 1058481309 1058491309 1058501309 1058511309 1058521309 1058531309 1058541309 1058551309 1058561309 1058571309 1058581309 1058591309 1058601309 1058611309 1058621309 1058631309 1058641309 1058651309 1058661309 1058671309 1058681309 1058691309 1058701309 1058711309 1058721309 1058731309 1058741309 1058751309 1058761309 1058771309 1058781309 1058791309 1058801309 1058811309 1058821309 1058831309 1058841309 1058851309 1058861309 1058871309 1058881309 1058891309 1058901309 1058911309 1058931309 1058951309 1058961309 1058971309 1058981309 1058991309 1059001309 1059011309 1059021309 1059031309 1059051309 1059061309 1059071309 1059081309 1059091309 1059101309 1059111309 1059121309 1059131309 1059141309 1059151309 1059161309 1059171309 1059181309 1059191309 1059201309 1059211309 1059221309 1059231309 1059241309 1059251309 1059261309 1059271309 1059281309 1059291309 1059301309 1059311309 1059321309 1059321312 1059331309 1059341309 1059341312 1059351309 1059361309 1059361312 1059371309 1059381309 1059381312 1059391309 1059401309 1059401312 1059411309 1059421309 1059421312 1059431309 1059441309 1059441312 1059451309 1059461309 1059461312 1059481312 1059501312 1059521312 1059541312 1059551312 1059561312 1059571312 1059591312 1059611312 1059631312 1059641312 1059651312 1059661312 1059671312 1059681312 1059691312 1059701312 1059711312 1059721312 1059731312 1059741312 1059751312 1059851312 1059871312 1059891312 1059991312 1060081312 1060101312 1060351312 1060611309 1060621309 1060631309 1060641309 1060651309 1060661309 1060671309 1060681309 1060691309 1060701309 1060711309 1060721309 1060731309 1060741309 1060751309 1060761309 1060771309 1060781309 1060791309 1060801309 1060811309 1060821309 1060831309 1060841309 1060851309 1060861309 1060871309 1060881309 1060891309 1060901309 1060911309 1060921309 1060931309 1060941309 1060951309 1060971309 1060981309 1060991309 1061001309 1061011309 1061021309 1061031309 1061041309 1061051309 1061061309 1061071309 1061081309 1061091309 1061101309 1061111309 1061121309 1061131309 1061141309 1061151309 1061171309 1061181309 1061191309 1061201309 1061211309 1061221309 1061231309 1061241309 1061251309 1061261309 1061271309 1061281309 1061291309 1061301309 1061311309 1061321309 1061331309 1061341309 1061351309 1061361309 1061371309 1061381309 1061391309 1061391312 1061401309 1061401312 1061411309 1061411312 1061421309 1061421312 1061431309 1061431312 1061441309 1061441312 1061451309 1061451312 1061461309 1061461312 1061471309 1061471312 1061481309 1061491309 1061491312 1061501309 1061511309 1061511312 1061521309 1061531309 1061531312 1061541309 1061551312 1061561309 1061561312 1061571309 1061571312 1061591312 1061601312 1061611312 1061621312 1061631312 1061641309 1061641312 1061651312 1061661309 1061661312 1061671312 1061681312 1061701309 1061701312 1061711312 1061721309 1061721312 1061731309 1061731312 1061741309 1061741312 1061751309 1061751312 1061761312 1061771312 1061781312 1061791312 1061801312 1061811312 1061821312 1061841312 1061861312 1061881312 1061901312 1061921312 1061931312 1061941312 1061951312 1061961312 1061981312 1062001312 1062011312 1062021312 1062031312 1062041312 1062061312 1062081312 1062101312 1062121312 1062181312 1062201312 1062221312 1062231312 1062241312 1062251312 1062261312 1062271312 1062281312 1062291312 1062301312 1062311312 1062321312 1062331312 1062341312 1062361312 1062381312 1062401312 1062421312 1062431312 1062441312 1062451312 1067191311 1067211311 1067231311 1067271311 1067291311 1067711311 1067721311 1067741311 1067751311 1067761311 1067771311 1067781311 1067791311 1067801311 1067811311 1067821311 1067831311 1067851311 1067871311 1067891311 1067911311 1067931311 1067951311 1067971311 1067981311 1067991311 1068001311 1068011311 1068021311 1068031311 1068041311 1068051311 1068061311 1068081311 1068101311 1068121311 1073641309 1073651309 1073661309 1073671309 1073681309 1073691309 1073701309 1073711309 1073721309 1073731309 1073741309 1073751309 1073761309 1073771309 1073781309 1073791309 1073801309 1073811309 1073821309 1073831309 1073841309 1073851309 1073861309 1073871309 1073881309 1073891309 1073901309 1073911309 1073921309 1073931309 1073941309 1073951309 1073961309 1073971309 1073981309 1073991309 1074001309 1074011309 1074021309 1074031309 1074041309 1074051309 1074061309 1074071309 1074081309 1074091309 1074101309 1074111309 1074121309 1074131309 1074141309 1074151309 1074161309 1074171309 1074181309 1074191309 1074201309 1074211309 1074221309 1074231309 1074241309 1074261309 1074271309 1074281309 1074301309 1074321309 1074331309 1074341309 1074351309 1074361309 1074371309 1074381309 1074391309 1074401309 1074411309 1074421309 1074431309 1074441309 1074451309 1074461309 1074471309 1074481309 1074491309 1074501309 1074511309 1074521309 1074531309 1074541309 1074551309 1074561309 1074571309 1074581309 1074591309 1074601309 1074611309 1074621309 1074631309 1074641309 1074651309 1074661309 1074671309 1074681309 1074691309 1074701309 1074711309 1074721309 1074731309 1074741309 1074751309 1074761309 1074771309 1074781309 1075741311 1075761311 1075781311 1075801311 1075811311 1075821311 1075841311 1075851311 1075861311 1075871311 1075881311 1075891311 1075901311 1075911311 1075921311 1075931311 1075941311 1075951311 1075961311 1075971311 1075981311 1075991311 1076001311 1076011311 1076021311 1076031311 1076041311 1076061311 1076081311 1076101311 1076121311 1076141311 1076161311 1076181311 1076201311 1076221311 1076231311 1076241311 1076261311 1076271311 1076281311 1076291311 1076301311 1076311311 1076321311 1076331311 1076341311 1076351311 1076361311 1076371311 1076381311 1076391311 1076401311 1076411311 1076421311 1076431311 1076441311 1076451311 1076461311 1076471311 1076481311 1076491311 1076501311 1076511311 1076521311 1076531311 1076551311 1076561311 1076571311 1076591311 1077101310 1077121310 1077141310 1077151310 1077161310 1077171310 1077181310 1077191310 1077201310 1077211310 1077221310 1077231310 1077241310 1077251310 1077261310 1077271310 1077281310 1077301310 1077321310 1077341310 1077361310 1077381310 1077391310 1077401310 1077411310 1077421310 1077431310 1077441310 1077451310 1077461310 1077471310 1077481310 1077491310 1077501310 1077511310 1077531310 1077541310 1077561310 1077581310 1077591310 1077601310 1077611310 1077621310 1077631310 1077641310 1077651310 1077661310 1077671310 1077681310 1077691310 1077701310 1077721310 1077741310 1077761310 1077781310 1077801310 1077821310 1077841310 1077861310 1077881310 1077901310 1078131310 1078151310 1078171310 1078191310 1078211310 1078231310 1083331311 1083341311 1083351311 1083361311 1083371311 1083381311 1083391311 1083401311 1083411311 1083421311 1083431311 1083441311 1083451311 1083461311 1083471311 1083481311 1083491311 1083501311 1083511311 1083521311 1083531311 1083541311 1083551311 1083561311 1083571311 1083581311 1083591311 1083601311 1083611311 1083621311 1083631311 1083641311 1083651311 1083661311 1083671311 1083681311 1083691311 1083701311 1083711311 1083721311 1083731311 1083741311 1083751311 1083761311 1083771311 1083781311 1083791311 1083801311 1083811311 1083821311 1083831311 1083841311 1083851311 1083861311 1083871311 1083881311 1083891311 1083901311 1083911311 1083921311 1083931311 1083941311 1083951311 1083961311 1083971311 1083981311 1083991311 1084001311 1084011311 1084021311 1084031311 1084041311 1084051311 1084061311 1084071311 1084081311 1084091311 1084101311 1084111311 1084121311 1084131311 1084141311 1084151311 1084161311 1084171311 1084181311 1084191311 1084201311 1084211311 1084221311 1084231311 1084251311 1084271311 1084291311 1084301311 1084311311 1084321311 1084331311 1084341311 1084351311 1084361311 1084371311 1084381311 1084391311 1084401311 1084411311 1084421311 1084431311 1084441311 1084451311 1084941310 1084951310 1084961310 1084971310 1084981310 1085001310 1085021310 1085041310 1085061310 1085081310 1085091310 1085101310 1085111310 1085121310 1085131310 1085141310 1085151310 1085161310 1085171310 1085181310 1085191310 1085201310 1085211310 1085221310 1085231310 1085241310 1085251310 1085261310 1085271310 1085281310 1085291310 1085301310 1085311310 1085321310 1085331310 1085341310 1085351310 1085361310 1085371310 1085381310 1085401310 1085421310 1085441310 1085461310 1085481310 1085501310 1085521310 1085541310 1085561310 1085581310 1085601310 1085621310 1085641310 1085661310 1085681310 1085701310 1085711310 1085721310 1085731310 1085741310 1085751310 1085761310 1085771310 1085781310 1085791310 1085801310 1085811310 1085821310 1085831310 1085841310 1085851310 1085861310 1085871310 1085881310 1085891310 1085901310 1085911310 1085921310 1085931310 1085941310 1085961310 1086001310 1086021310 1086061310 1086081310 1088781311 1088791311 1088801311 1088811311 1088821311 1088831311 1088841311 1088851311 1088861311 1088871311 1088891311 1088911311 1088931311 1088951311 1088971311 1088991311 1089011311 1089031311 1089051311 1089071311 1089091311 1089111311 1089131311 1089151311 1089161311 1089171311 1089181311 1089191311 1089201311 1089211311 1089221311 1089231311 1089241311 1089261311 1089271311 1089281311 1089291311 1089301311 1089311311 1089321311 1089331311 1089351311 1089371311 1089391311 1089411311 1089431311 1089451311 1089471311 1089491311 1089511311 1089531311 1089551311 1089571311 1089581311 1089591311 1089601311 1089611311 1089621311 1089631311 1089641311 1089651311 1089661311 1089671311 1089681311 1089691311 1089701311 1089711311 1089721311 1089731311 1089741311 1089751311 1089761311 1089771311 1089781311 1089791311 1089801311 1089811311 1089821309 1089821311 1089831309 1089831311 1089841309 1089841311 1089851309 1089851311 1089861309 1089861311 1089871309 1089871311 1089881309 1089881311 1089891311 1089901309 1089901311 1089911311 1089921309 1089941309 1089961309 1090061309 1090081309 1090111309 1090131309 1090151309 1090171309 1090191309 1090211309 1090231309 1090251309 1090271309 1090281309 1090291309 1090301309 1090311309 1090321309 1090331309 1090341309 1090351309 1090361309 1090371309 1090381309 1090391309 1090401309 1090411309 1090421309 1090431309 1090441309 1090451309 1090481309 1090501309 1090521309 1090541309 1090561309 1090571309 1090581309 1090591309 1090601309 1090611309 1090621309 1090631309 1090641309 1090661309 1090761309 1090781309 1090801309 1090821309 1090841309 1090861309 1090881309 1090901309 1090911309 1090921309 1091551310 1091571310 1091591310 1091611310 1091631310 1091651310 1091671310 1091681310 1091691310 1091701310 1091711310 1091721310 1091731310 1091741310 1091751310 1091771310 1091791310 1091801310 1091811310 1091821310 1091831310 1091841310 1091851310 1091861310 1091871310 1091881310 1092231310 1092241310 1092251310 1092261310 1092271310 1092281310 1092291310 1092301310 1092321310 1092341310 1092361310 1092381310 1092401310 1092421310 1092441310 1092461310 1092481310 1092501310 1092521310 1092541310 1092571310 1108221310 1108241310 1108531310 1109031310 1109051310 1109071310 1109091310 1109111310 1109131310 1109151310 1109171310 1109191310 1109211310 1109231310 1109251310 1119071310 1119081310 1119091310 1119101310 1119111310 1119121310 1119131310 1119141310 1119151310 1119161310 1119171310 1119181310 1119191310 1119201310 1119211310 1119221310 1119231310 1119251310 1119271310 1119281310 1119291310 1119301310 1119311310 1119321310 1119331310 1119341310 1119351310 1119361310 1119371310 1119381310 1119391310 1119401310 1119411310 1119421310 1119431310 1119441310 1119451310 1119461310 1119471310 1119481310 1119491310 1119501310 1119511310 1119521310 1119531310 1119541310 1119551310 1119561310 1119571310 1119581310 1119591310 1119601310 1119611310 1119621310 1119631310 1119641310 1119651310 1119661310 1119671310 1119681310 1119691310 1119701310 1119711310 1119721310 1119731310 1119741310 1119751310 1119761310 1119771310 1119781310 1119791310 1119801310 1119811310 1119821310 1119831310 1119841310 1119851310 1119861310 1119871310 1119881310 1119891310 1119901310 1119921310 1119941310 1119951310 1119961310 1119971310 1119981310 1119991310 1120001310 1120021310 1120031310 1120041310 1120051310 1120061310 1121021311 1121041311 1121061311 1121081311 1121101311 1121121311 1121141311 1121161311 1121181311 1121201311 1121211311 1121221311 1121231311 1121241311 1121251311 1121261311 1121271311 1121281311 1121291311 1121301311 1121311311 1121321311 1121331311 1121341311 1121351311 1121371311 1121391311 1121401311 1121411311 1121421311 1121431311 1121441311 1121451311 1121461311 1121471311 1121481311 1121491311 1121501311 1121511311 1121521311 1121531311 1121541311 1121561311 1121581311 1121601311 1121621311 1121641311 1121661311 1121681311 1121701311 1121721311 1121741311 1121761311 1121821311 1121861311 1122011311 1122031311 1122051311 1122071311 1122361310 1122371310 1122381310 1122391310 1122401310 1122411310 1122421310 1122431310 1122441310 1122451310 1122461310 1122471310 1122481310 1122491310 1122501310 1122511310 1122521310 1122531310 1122541310 1122551310 1122561310 1122571310 1122581310 1122591310 1122601310 1122611310 1122621310 1122631310 1122641310 1122651310 1122661310 1122671310 1122681310 1122691310 1122701310 1122711310 1122721310 1122731310 1122741310 1122751310 1122761310 1122771310 1122781310 1122791310 1122801310 1122811310 1122821310 1122831310 1122841310 1122851310 1122861310 1122871310 1122881310 1122891310 1122901310 1122911310 1122921310 1122931310 1122941310 1122951310 1122961310 1122971310 1122981310 1122991310 1123001310 1123011310 1123021310 1123031310 1123041310 1123051310 1123061310 1123071310 1123081310 1123091310 1123101310 1123111310 1123121310 1123131310 1123141310 1123151310 1123171310 1123181310 1123191310 1123201310 1123211310 1123221310 1123231310 1123241310 1123251310 1123261310 1123271310 1123281310 1123291310 1123301310 1123311310 1123321310 1123331310 1123341310 1123351310 1123361310 1123371310 1123381310 1123391310 1123401310 1123411310 1123421310 1123431310 1123441310 1123451310 1123461310 1123471310 1123481310 1123491310 1133081311 1133091311 1133111311 1133131311 1133151311 1133171311 1133191311 1133211311 1133231311 1133251311 1133261311 1133271311 1133281311 1133291311 1133301311 1133311311 1133401311 1133421311 1133431311 1133441311 1133461311 1133471311 1133481311 1133491311 1133501311 1133511311 1133531311 1133551311 1133571311 1133591311 1133611311 1133631311 1133651311 1133671311 1133691311 1133711311 1133731311 1133751311 1133771311 1133791311 1133811311 1133831311 1133851311 1133941311 1133961311 1133981311 1134001311 1136321311 1136331311 1136341311 1136351311 1136361311 1136371311 1136381311 1136391311 1136411311 1136421311 1136431311 1136441311 1136451311 1136461311 1136471311 1136481311 1136491311 1136501311 1136511311 1136521311 1136541311 1136561311 1136601311 1136621311 1136631311 1136641311 1136651311 1136661311 1136671311 1136681311 1136691311 1136701311 1136711311 1136721311 1136731311 1136741311 1136751311 1136761311 1136781311 1136791311 1136801311 1136811311 1136821311 1136831311 1136841311 1136851311 1136861311 1136871311 1136881311 1136891311 1136901311 1136911311 1136921311 1136931311 1136941311 1136951311 1136961311 1136971311 1136981311 1136991311 1137001311 1137011311 1137021311 1137031311 1137041311 1137051311 1137061311 1137071311 1137081311 1137091311 1137101311 1137111311 1137121311 1137131311 1137141311 1137151311 1137161311 1137171311 1137181311 1137191311 1137201311 1137211311 1137221311 1137231311 1137241311 1137251311 1137261311 1137271311 1137281311 1137291311 1137301311 1137311311 1137321311 1137331311 1137341311 1137351311 1137361311 1137371311 1137381311 1137391311 1137401311 1152251309 1152271309 1152291309 1152311309 1152331309 1152351309 1152371309 1152391309 1152411309 1152431309 1152451309 1152471309 1152491309 1152511309 1152521309 1152531309 1152541309 1152551309 1152561309 1152571309 1152581309 1152591309 1152601309 1152611309 1152621309 1152631309 1152641309 1152651309 1152661309 1152671309 1152681309 1152691309 1152701309 1152711309 1152721309 1152731309 1152741309 1152751309 1152761309 1152771309 1152781309 1152791309 1152801309 1152811309 1152821309 1152831309 1152841309 1152851309 1152861309 1152871309 1152881309 1152901309 1152921309 1152941309 1152961309 1152971309 1152981309 1152991309 1153001309 1153011309 1153021309 1153041309 1153231309 1153251309 1153271309 1155491309 1155511309 1155521309 1155531309 1155541309 1155551309 1155561309 1155571309 1155581309 1155591309 1155601309 1155611309 1155621309 1155631309 1155641309 1155651309 1155661309 1155671309 1155681309 1155691309 1155701309 1155711309 1155721309 1155731309 1155741309 1155751309 1155761309 1155771309 1155781309 1155791309 1155801309 1155811309 1155821309 1155831309 1155841309 1155851309 1155861309 1155871309 1155881309 1155891309 1155901309 1155911309 1155921309 1155931309 1155941309 1155951309 1155961309 1155971309 1155981309 1155991309 1156001309 1156011309 1156021309 1156031309 1156041309 1156051309 1156061309 1156071309 1156081309 1156091309 1156101309 1156111309 1156121309 1156131309 1156141309 1156151309 1156161309 1156171309 1156181309 1156191309 1156201309 1156211309 1156221309 1156231309 1156241309 1156251309 1156261309 1156271309 1156281309 1156291309 1156301309 1156311309 1156321309 1156331309 1156341309 1156351309 1156361309 1156371309 1156381309 1156391309 1156401309 1156411309 1156421309 1156431309 1156441309 1156451309 1156461309 1156481309 1156491309 1156501309 1156511309 1156521309 1156531309 1156541309 1156551309 1156561309 1156571309 1156581309 1157661309 1157671309 1157681309 1157691309 1157701309 1157711309 1157721309 1157731309 1157741309 1157751309 1157761309 1157771309 1157781309 1157791309 1157801309 1157811309 1157821309 1157831309 1157851309 1157871309 1157891309 1157911309 1157931309 1157951309 1157961309 1157971309 1157981309 1158001309 1158011309 1158021309 1158031309 1158041309 1158061309 1158081309 1158091309 1158101309 1158111309 1158121309 1158131309 1158141309 1158151309 1158161309 1158171309 1158181309 1158191309 1158201309 1158211309 1158221309 1158231309 1158241309 1158251309 1158261309 1158271309 1158281309 1158291309 1158301309 1158311309 1158321309 1158331309 1158341309 1158351309 1158361309 1158371309 1158381309 1158391309 1158401309 1158411309 1158421309 1158431309 1158441309 1158451309 1158461309 1158471309 1158481309 1158491309 1158501309 1158511309 1158521309 1158531309 1158541309 1158551309 1158561309 1158571309 1158581309 1158591309 1158611309 1158621309 1158631309 1158641309 1158651309 1158661309 1158671309 1158681309 1158691309 1158701309 1158711309 1158721309 1158731309 1158741309 1158751309 1158761309 1158771309 1158781309 1158791309 1158801309 1158811309 1158831309 1158851309 1158871309 1159141309 1159171309 1159191309 1159211309 1159221309 1159231309 1159241309 1159251309 1159261309 1159281309 1159291309 1159301309 1159321309 1159341309 1159361309 1159381309 1159401309 1159421309 1159431309 1159441309 1159451309 1159461309 1159471309 1159481309 1159491309 1159501309 1159511309 1159521309 1159531309 1159541309 1159551309 1159561309 1159571309 1159581309 1159591309 1159601309 1159611309 1159621309 1159631309 1159641309 1159651309 1159661309 1159671309 1159681309 1159691309 1159701309 1159711309 1159721309 1159731309 1159741309 1159751309 1159761309 1159771309 1159781309 1159791309 1159801309 1159811309 1159821309 1159831309 1159841309 1159851309 1159861309 1159871309 1159881309 1159891309 1159901309 1159911309 1159921309 1159931309 1159941309 1159951309 1160841310 1160851310 1160861310 1160871310 1160891310 1160911310 1160931310 1160951310 1160991310 1161011310 1161031310 1161291310 1161311310 1161331310 1161351310 1161371310 1161431310 1161451310 1161471310 1161491310 1161511310 1161531310 1161551310 1161591310 1161611310 1161651310 1161671310 1170301310 1170321310 1170341310 1170361310 1170401310 1170421310 1170441310 1170461310 1170481310 1170681310 1170701310 1170721310 1170741310 1170761310 1170781310 1170801310 1170821310 1170841310 1170861310 1170881310 1170901310 1170921310 1170941310 1170961310 1170981310 1171141310 1171151310 1171161310 1171171310 1171181310 1171191310 1171201310 1171211310 1171231310 1171241310 1171251310 1171261310 1171271310 1171301310 1171321310 1171341310 1171361310 1171371310 1171381310 1171391310 1171411310 1192371310 1192381310 1192391310 1192401310 1192421310 1192441310 1192461310 1192481310 1192501310 1192521310 1192541310 1192561310 1192581310 1192601310 1192621310 1192641310 1192661310 1192681310 1192701310 1192721310 1192741310 1192761310 1192781310 1192801310 1192811310 1192821310 1192831310 1192841310 1192851310 1192861310 1192871310 1192881310 1192891310 1192901310 1192911310 1192921310 1192931310 1192941310 1192951310 1192961310 1192971310 1192981310 1192991310 1193001310 1193011310 1193021310 1193031310 1193041310 1193051310 1193061310 1193071310 1193081310 1193091310 1193101310 1193111310 1193121310 1193131310 1193141310 1193151310 1193161310 1193171310 1193181310 1193191310 1193201310 1193221310 1193231310 1193241310 1193251310 1193271310 1193291310 1193311310 1193331310 1193351310 1193371310 1193391310 1193411310 1193421310 1193431310 1193441310 1204501309 1204521309 1204541309 1204561309 1204581309 1204601309 1204621309 1204641309 1204661309 1204681309 1204701309 1204721309 1204741309 1204761309 1204781309 1204801309 1204821309 1204841309 1204861309 1204881309 1204901309 1205221309 1205241309 1205261309 1205281309 1205301309 1205321309 1205401309 1205471309 1205491309 1205511309 1205531309 1205551309 1205571309 1205591309 1205611309 1205721311 1205731311 1205781311 1205801311 1205811311 1205821311 1205831311 1205841311 1205851311 1205861311 1205881311 1205891311 1205901311 1205911311 1205921311 1205931311 1205941311 1205951311 1205961311 1205971311 1205981311 1205991311 1206001311 1206011311 1206021311 1206031311 1206041311 1206051311 1206061311 1206081311 1206091311 1206281311 1206301311 1206361311 1206381311 1206401311 1206421311 1206441311 1206461311 1206481311 1206501311 1206521311 1206541311 1211501310 1211511310 1211521310 1211531310 1211541310 1211551310 1211561310 1211571310 1211581310 1211591310 1211601310 1211611310 1211621310 1211631310 1211641310 1211651310 1211661310 1211671310 1211681310 1211691310 1211701310 1211711310 1211721310 1211731310 1211741310 1211751310 1211761310 1211771310 1211781310 1211791310 1211801310 1211811310 1211821310 1211831310 1211841310 1211851310 1211861310 1211871310 1211881310 1211891310 1211901310 1211911310 1211921310 1211931310 1211941310 1211951310 1211961310 1211971310 1211981310 1211991310 1212001310 1212011310 1212021310 1212031310 1212041310 1212051310 1212061310 1212071310 1212081310 1212091310 1212101310 1212111310 1212121310 1212131310 1212141310 1212151310 1212161310 1212171310 1212181310 1212191310 1212201310 1212211310 1212221310 1212231310 1212241310 1212251310 1212261310 1212271310 1212281310 1212291310 1212301310 1212311310 1212321310 1212331310 1212341310 1212351310 1212361310 1212371310 1212381310 1212391310 1212401310 1212411310 1212421310 1212431310 1212441310 1212451310 1212461310 1212471310 1212481310 1212491310 1212501310 1212511310 1212521310 1212531310 1212541310 1212551310 1212561310 1212571310 1212581310 1212591310 1212601310 1212611310 1212621310 1212631310 1227111311 1227121311 1227131311 1227141311 1227151311 1227161311 1227171311 1227181311 1227191311 1227201311 1227211311 1227221311 1227231311 1227241311 1227251311 1227261311 1227271311 1227281311 1227291311 1227301311 1227311311 1227321311 1227331311 1227341311 1227351311 1227361311 1227371311 1227381311 1227391311 1227401311 1227411311 1227421311 1227431311 1227441311 1227451311 1227461311 1227471311 1227481311 1227491311 1227501311 1227511311 1227521311 1227531311 1227541311 1227551311 1227561311 1227571311 1227581311 1227591311 1227601311 1227611311 1227621311 1227631311 1227641311 1227651311 1227661311 1227671311 1227681311 1227691311 1227701311 1227711311 1227721311 1227731311 1227741311 1227751311 1227761311 1227771311 1227781311 1227791311 1227801311 1227811311 1227821311 1227831311 1227841311 1227851311 1227861311 1227871311 1227881311 1227891311 1227901311 1227911311 1227921311 1227931311 1227941311 1227951311 1227961311 1227971311 1227981311 1227991311 1228001311 1228011311 1228021311 1228031311 1228041311 1228051311 1228061311 1228071311 1228081311 1228091311 1228101311 1228111311 1228121311 1228131311 1228141311 1228151311 1228161311 1228171311 1228181311 1229901309 1229911309 1229921309 1229931309 1229941309 1229951309 1229961309 1229971309 1229981309 1229991309 1230001309 1230011309 1230021309 1230031309 1230041309 1230051309 1230061309 1230071309 1230081309 1230091309 1230101309 1230111309 1230121309 1230131309 1230151309 1230171309 1230191309 1230211309 1230231309 1230241309 1230251309 1230271309 1230291309 1230311309 1230321309 1230331309 1230341309 1230351309 1230361309 1230371309 1230381309 1230391309 1230401309 1230411309 1230421309 1230431309 1230441309 1230451309 1230461309 1230471309 1230481309 1230501309 1230521309 1230531309 1230541309 1230551309 1230561309 1230571309 1230581309 1230591309 1230601309 1230611309 1230621309 1230631309 1230641309 1230651309 1230661309 1230671309 1230681309 1230691309 1230701309 1230711309 1230721309 1230731309 1230741309 1230751309 1230761309 1230771309 1230781309 1230791309 1230801309 1230811309 1230821309 1230831309 1230841309 1230851309 1230861309 1230871309 1230881309 1230891309 1230901309 1237801311 1237811311 1237821311 1237831311 1237841311 1237851311 1237861311 1237871311 1237891311 1237911311 1237931311 1237951311 1237971311 1237991311 1238011311 1238031311 1238051311 1238071311 1238081311 1238101311 1238121311 1238131311 1238141311 1238151311 1238161311 1238171311 1238181311 1238201311 1238211311 1238221311 1238231311 1238241311 1238251311 1238261311 1238271311 1238281311 1238291311 1238301311 1238311311 1238321311 1238331311 1238341311 1238351311 1238361311 1238371311 1238381311 1238391311 1238401311 1238411311 1238421311 1238431311 1238441311 1238451311 1238461311 1238471311 1238481311 1238491311 1238501311 1238511311 1238521311 1238531311 1238541311 1238551311 1238561311 1238571311 1238581311 1238591311 1238601311 1238611311 1238621311 1238641311 1238661311 1238681311 1238701311 1238721311 1238741311 1238761311 1238781311 1238801311 1238821311 1238831311 1238841311 1238851311 1238861311 1238871311 1238881311 1238891311 1238901311 1238911311 1238921311 1248291309 1248301309 1248311309 1248321309 1248331309 1248341309 1248351309 1248361309 1248371309 1248381309 1248391309 1248401309 1248411309 1248421309 1248431309 1248441309 1248451309 1248461309 1248471309 1248481309 1248491309 1248501309 1248511309 1248521309 1248531309 1248541309 1248551309 1248561309 1248571309 1248581309 1248591309 1248611309 1248631309 1248651309 1248671309 1248691309 1248711309 1248731309 1248751309 1248771309 1248791309 1248801309 1248811309 1248821309 1248831309 1248841309 1248851309 1248861309 1248871309 1248881309 1248891309 1248901309 1248911309 1248921309 1248931309 1248941309 1248951309 1248971309 1248991309 1249011309 1249031309 1249051309 1249071309 1249091309 1249111309 1249131309 1249151309 1249161309 1249171309 1249191309 1249201309 1249211309 1249221309 1249231309 1249241309 1249251309 1249261309 1249271309 1249281309 1249291309 1249301309 1249311309 1249321309 1249331309 1249341309 1249351309 1249361309 1249371309 1249381309 1249391309 1249401309 1249411309 1249421309 1249431309 1249441309 1249451309 1249461309 1249471309 1258021310 1258031310 1258041310 1258051310 1258061310 1258071310 1258081310 1258091310 1258101310 1258111310 1258121310 1258131310 1258141310 1258151310 1258161310 1258171310 1258181310 1258191310 1258201310 1258211310 1258221310 1258231310 1258251310 1258271310 1258291310 1258311310 1258331310 1258351310 1258371310 1258491310 1258511310 1258531310 1258541310 1258551310 1258561310 1258571310 1258581310 1258591310 1258601310 1258611310 1258621310 1258631310 1258641310 1258651310 1258661310 1258671310 1258681310 1258691310 1258701310 1258711310 1258721310 1258731310 1258741310 1258751310 1258761310 1258771310 1258781310 1258801310 1258821310 1258841310 1258861310 1258901310 1258921310 1258941310 1258961310 1262231308 1262241308 1262251308 1262271308 1262291308 1262311308 1262321308 1262331308 1262341308 1262351308 1262361308 1262391308 1262411308 1262431308 1262451308 1262491308 1262561308 1262581308 1262601308 1262621308 1262641308 1262651308 1262761308 1262781308 1262801308 1262821308 1262841308 1262861308 1262881308 1262941308 1262951308 1262971308 1262991308 1263001308 1263011308 1263021308 1263031308 1263041308 1263051308 1263061308 1263071308 1263081308 1263091308 1263101308 1263111308 1263121308 1263131308 1263141308 1263151308 1263161308 1281051308 1281061308 1281071308 1281081308 1281091308 1281101308 1281111308 1281121308 1281131308 1281141308 1281151308 1281161308 1281171308 1281181308 1281191308 1281201308 1281211308 1281221308 1281231308 1281241308 1281251308 1281261308 1281271308 1281281308 1281291308 1281301308 1281311308 1281321308 1281331308 1281341308 1281351308 1281361308 1281371308 1281381308 1281391308 1281401308 1281411308 1281431308 1281441308 1281451308 1281461308 1281481308 1281491308 1281501308 1281511308 1281521308 1281531308 1281541308 1281551308 1281561308 1281571308 1281581308 1281591308 1281601308 1281611308 1281621308 1281631308 1281641308 1281651308 1281661308 1281671308 1281681308 1281691308 1281701308 1281711308 1281721308 1281731308 1281741308 1281751308 1281771308 1281791308 1281801308 1281811308 1281821308 1281831308 1281851308 1281871308 1281881308 1281891308 1281901308 1281911308 1281921308 1281931308 1281941308 1281951308 1281961308 1281971308 1003551309 1003571309 1003591309 1003611309 1003631309 1003651309 1003671309 1003691309 1003711309 1003731309 1003751309 1003771309 1003791309 1003811309 1008001311 1008021311 1008041311 1008141311 1008161311 1008201311 1008221311 1008261311 1008281311 1008301311 1008321311 1008341311 1008361311 1008381311 1008401311 1008421311 1008441311 1008461311 1008481311 1008501311 1008511311 1008521311 1008531311 1008541311 1008551311 1008561311 1008581311 1008601311 1008611311 1008621311 1008641311 1008661311 1008681311 1008721311 1008741311 1008761311 1008781311 1008831311 1008891311 1008931311 1008971311 1008991311 1009011311 1009031311 1009051311 1009071311 1009091311 1009111311 1013501311 1013521311 1013531311 1013541311 1013551311 1013561311 1013571311 1013581311 1013591311 1013601311 1013611311 1013621311 1013631311 1013641311 1013651311 1013661311 1013671311 1013681311 1013691311 1013701311 1013711311 1013721311 1013731311 1013741311 1013751311 1013761311 1013781311 1013791311 1013801311 1013821311 1013841311 1013871311 1013891311 1013901311 1013911311 1013921311 1013931311 1013941311 1013951311 1013961311 1013971311 1013981311 1013991311 1014001311 1014011311 1014021311 1014031311 1014041311 1014051311 1014061311 1014071311 1014081311 1014091311 1014101311 1014111311 1014121311 1014131311 1014171311 1014181311 1014191311 1014201311 1014211311 1014221311 1014231311 1014241311 1014251311 1014261311 1014271311 1014291311 1014301311 1014311311 1014321311 1014331311 1014341311 1014351311 1014361311 1014371311 1014381311 1014391311 1014401311 1014411311 1014421311 1014431311 1014441311 1014451311 1014461311 1014471311 1014481311 1014491311 1014501311 1014511311 1014521311 1014531311 1014541311 1014551311 1014571311 1014591311 1014611311 1018661310 1018671310 1018681310 1018691310 1018701310 1018721310 1018741310 1018761310 1018781310 1018801310 1018811310 1018821310 1018831310 1018841310 1018861310 1018881310 1018901310 1018911310 1018921310 1018931310 1018941310 1018961310 1018981310 1018991310 1019001310 1019011310 1019031310 1019041310 1019051310 1019061310 1019071310 1019081310 1019101310 1019111310 1019121310 1019131310 1019161310 1019171310 1019181310 1019201310 1019221310 1019231310 1019241310 1019261310 1019271310 1019281310 1019301310 1019321310 1019331310 1019341310 1019351310 1019361310 1019371310 1019381310 1019401310 1019411310 1019491310 1019501310 1019511310 1019521310 1019541310 1019551310 1019561310 1019571310 1019581310 1019591310 1019601310 1019611310 1019621310 1019631310 1019641310 1019651310 1019661310 1019671310 1019681310 1019691310 1019701310 1019711310 1019721310 1019731310 1019741310 1019751310 1019761310 1019771310 1019781310 1019791310 1024201310 1024211310 1024221310 1024231310 1024251310 1024261310 1024271310 1024281310 1024291310 1024301310 1024311310 1024321310 1024331310 1024341310 1024351310 1024361310 1024371310 1024391310 1024461310 1024481310 1024501310 1024521310 1024531310 1024541310 1024551310 1024571310 1024591310 1024611310 1024621310 1024631310 1024641310 1024651310 1024661310 1024671310 1024681310 1024691310 1024701310 1024711310 1024721310 1024731310 1024741310 1024751310 1024761310 1024771310 1024781310 1024791310 1024801310 1024821310 1024831310 1024841310 1024851310 1024861310 1024871310 1024881310 1024891310 1024901310 1024911310 1024921310 1024931310 1024941310 1024951310 1024961310 1024971310 1024981310 1024991310 1025001310 1025011310 1025021310 1025031310 1025041310 1025051310 1025061310 1025071310 1025081310 1025091310 1025101310 1025111310 1025121310 1025131310 1025141310 1025151310 1025161310 1025171310 1025181310 1025191310 1025201310 1025211310 1025221310 1025231310 1025241310 1025251310 1025261310 1025281310 1034051310 1034071310 1034091310 1034111310 1034131310 1034151310 1034171310 1034191310 1034211310 1034231310 1034251310 1034271310 1034291310 1034311310 1034331310 1034351310 1034371310 1034391310 1034411310 1034431310 1034451310 1034471310 1034491310 1034511310 1034531310 1034551310 1034571310 1034591310 1034611310 1034631310 1034651310 1034671310 1034691310 1034711310 1034731310 1034751310 1034791310 1034801310 1034811310 1034831310 1034841310 1034851310 1034861310 1034871310 1034881310 1034891310 1034901310 1034911310 1034921310 1034931310 1034961310 1034981310 1035001310 1035011310 1035021310 1035031310 1035041310 1035051310 1035061310 1035071310 1035081310 1035091310 1035101310 1035111310 1035121310 1035141310 1035161310 1035171310 1035181310 1035191310 1035201310 1046911310 1046921310 1046931310 1046941310 1046951310 1046961310 1046971310 1046981310 1046991310 1047001310 1047011310 1047021310 1047031310 1047041310 1047051310 1047061310 1047071310 1047081310 1047091310 1047101310 1047111310 1047121310 1047131310 1047151310 1047171310 1047191310 1047211310 1047221310 1047231310 1047241310 1047251310 1047261310 1047271310 1047281310 1047291310 1047301310 1047311310 1047321310 1047341310 1047361310 1047381310 1047401310 1047421310 1047441310 1047461310 1047471310 1047481310 1047491310 1047501310 1047511310 1047521310 1047531310 1047541310 1047551310 1047561310 1047571310 1047581310 1047601310 1047621310 1047641310 1047661310 1047681310 1047701310 1047721310 1047741310 1047751310 1047761310 1047781310 1047791310 1047801310 1047811310 1047821310 1047831310 1047841310 1047851310 1047861310 1047871310 1047881310 1047891310 1047901310 1047911310 1047961310 1047981310 1071501309 1071511309 1071521309 1071531309 1071541309 1071551309 1071561309 1071571309 1071581309 1071591309 1071601309 1071611309 1071621309 1071631309 1071641309 1071651309 1071661309 1071671309 1071681309 1071691309 1071701309 1071711309 1071721309 1071731309 1071741309 1071751309 1071761309 1071771309 1071781309 1071791309 1071801309 1071811309 1071821309 1071831309 1071841309 1071851309 1071861309 1071871309 1071881309 1071891309 1071901309 1071911309 1071921309 1071931309 1071941309 1071951309 1071961309 1071971309 1071981309 1071991309 1072001309 1072011309 1072021309 1072031309 1072041309 1072051309 1072061309 1072071309 1072081309 1072091309 1072101309 1072111309 1072121309 1072131309 1072141309 1072151309 1072161309 1072171309 1072181309 1072191309 1072201309 1072211309 1072221309 1072231309 1072241309 1072251309 1072261309 1072271309 1072281309 1072291309 1072301309 1072311309 1072321309 1072331309 1072341309 1072351309 1072361309 1072371309 1072381309 1072391309 1072401309 1072411309 1072421309 1072431309 1072451309 1072461309 1072471309 1072481309 1072491309 1072511309 1072521309 1072531309 1072541309 1072551309 1072561309 1073471311 1073491311 1073511311 1073531311 1073571311 1073591311 1073611311 1073631311 1073651311 1073671311 1073691311 1073711311 1073731311 1073751311 1073771311 1073791311 1073811311 1073831311 1073851311 1073871311 1073881311 1073891311 1073911311 1073931311 1074391311 1074401311 1074411311 1074421311 1074431311 1074441311 1074461311 1074481311 1074501311 1074521311 1074541311 1074561311 1074581311 1074601311 1074621311 1074631311 1090471310 1090481310 1090491310 1090501310 1090511310 1090521310 1090531310 1090541310 1090551310 1090561310 1090571310 1090581310 1090591310 1090601310 1090611310 1090621310 1090631310 1090641310 1090651310 1090661310 1090681310 1090701310 1090741310 1090761310 1090901310 1090921310 1090941310 1090961310 1090981310 1091001310 1091021310 1091041310 1091061310 1091081310 1091101310 1091121310 1091141310 1091161310 1091181310 1091201310 1091221310 1091241310 1091261310 1091281310 1091301310 1091361310 1091501310 1091521310 1098631310 1098651310 1098671310 1098671311 1098691310 1098691311 1098711311 1098721310 1098731311 1098741310 1098751310 1098751311 1098761311 1098771310 1098771311 1098781311 1098791310 1098791311 1098801311 1098811311 1098821311 1098831311 1098841311 1098851311 1098861311 1098871310 1098871311 1098881311 1098891311 1098901311 1098911310 1098911311 1098921311 1098931310 1098931311 1098941311 1098951311 1098961311 1098971311 1098981311 1098991311 1099001311 1099011311 1099021311 1099031311 1099041311 1099051311 1099061311 1099071311 1099081311 1099091311 1099101311 1099111311 1099121311 1099131311 1099141311 1099151311 1099161311 1099171311 1099181311 1099201311 1099221311 1099241311 1099261311 1099281311 1099301311 1099321311 1099341311 1099361311 1099381311 1099401311 1099421311 1099441311 1099461311 1099481310 1099481311 1099501310 1099501311 1099521310 1099541310 1099561310 1099621310 1099631310 1099631311 1099641310 1099661310 1099661311 1099681311 1099701311 1099721311 1102821311 1103131311 1103151311 1103171311 1103191311 1103211311 1103231311 1103481311 1103501311 1103521311 1103541311 1103561311 1103581311 1103601311 1103621311 1103641311 1103661311 1103681311 1103701311 1103721311 1103731311 1103741311 1103751311 1103761311 1103771311 1103781311 1103791311 1103801311 1103811311 1103821311 1103831311 1105151309 1105161309 1105171309 1105181309 1105191309 1105201309 1105211309 1105221309 1105231309 1105241309 1105251309 1105261309 1105271309 1105281309 1105291309 1105301309 1105311309 1105321309 1105331309 1105341309 1105351309 1105361309 1105371309 1105381309 1105391309 1105401309 1105411309 1105421309 1105431309 1105441309 1105451309 1105461309 1105471309 1105481309 1105491309 1105501309 1105511309 1105521309 1105531309 1105541309 1105551309 1105561309 1105571309 1105581309 1105591309 1105601309 1105611309 1105621309 1105631309 1105641309 1105651309 1105661309 1105671309 1105681309 1105691309 1105701309 1105711309 1105721309 1105731309 1105741309 1105751309 1105761309 1105771309 1105781309 1105791309 1105801309 1105811309 1105821309 1105831309 1105841309 1105851309 1105861309 1105871309 1105881309 1105891309 1105901309 1105911309 1105921309 1105931309 1105941309 1105951309 1105961309 1105971309 1105981309 1105991309 1106001309 1106011309 1106021309 1106031309 1106041309 1106051309 1106061309 1106071309 1106081309 1106091309 1106101309 1106111309 1106121309 1106131309 1106141309 1106151309 1106161309 1106171309 1106181309 1106191309 1106201309 1106211309 1106221309 1106231309 1106241309 1106251309 1106261309 1106271309 1110571309 1110591309 1110601309 1110611309 1110621309 1110631309 1110641309 1110651309 1110661309 1110671309 1110681309 1110691309 1110711309 1110721309 1110731309 1110741309 1110751309 1110761309 1110771309 1110781309 1110791309 1110801309 1110811309 1110821309 1110831309 1110841309 1110851309 1110861309 1110871309 1110881309 1110891309 1110901309 1110911309 1110921309 1110931309 1110941309 1110951309 1110971309 1110981309 1110991309 1111001309 1111011309 1111021309 1111031309 1111041309 1111061309 1111081309 1111101309 1111121309 1111141309 1111161309 1111181309 1111201309 1111221309 1111241309 1111261309 1111281309 1111301309 1111321309 1111341309 1111361309 1111381309 1111401309 1111421309 1111431309 1111441309 1111451309 1111461309 1111471309 1111481309 1111491309 1111501309 1111511309 1111521309 1111531309 1111541309 1111551309 1111561309 1111571309 1111581309 1111591309 1111601309 1111611309 1111621309 1111641309 1111651309 1111661309 1111671309 1111681309 1111691309 1111701309 1111711309 1111731309 1133621309 1133631309 1133641309 1133651309 1133661309 1133671309 1133681309 1133691309 1133701309 1133711309 1133721309 1133731309 1133741309 1133751309 1133761309 1133771309 1133781309 1133791309 1133801309 1133811309 1133821309 1133831309 1133841309 1133851309 1133861309 1133871309 1133881309 1133901309 1133911309 1133921309 1133931309 1133941309 1133951309 1133961309 1133971309 1133981309 1133991309 1134011309 1134021309 1134031309 1134041309 1134061309 1134081309 1134091309 1134101309 1134121309 1134141309 1134161309 1134181309 1134441309 1134461309 1134481309 1134501309 1134521309 1134531309 1134541309 1134551309 1134561309 1134571309 1134581309 1134591309 1134601309 1134621309 1134641309 1134661309 1134671309 1134681309 1134691309 1134711309 1134721309 1140861311 1140881311 1140901311 1140921311 1140941311 1140961311 1140981311 1141001311 1141021311 1141041311 1141061311 1141071311 1141081311 1141091311 1141101311 1141111311 1141121311 1141131311 1141141311 1141151311 1141161311 1141171311 1141181311 1141191311 1141201311 1141211311 1141221311 1141231311 1141241311 1141251311 1141261311 1141271311 1141281311 1141291311 1141301311 1141311311 1141331311 1141351311 1141371311 1141391311 1141411311 1141431311 1141451311 1141471311 1141491311 1141511311 1143501309 1143511309 1143521309 1143531309 1143541309 1143551309 1143561309 1143571309 1143581309 1143591309 1143601309 1143611309 1143621309 1143631309 1143641309 1143651309 1143661309 1143671309 1143691309 1143701309 1143711309 1143721309 1143731309 1143731311 1143741309 1143751309 1143751311 1143761309 1143771309 1143771311 1143781309 1143791309 1143791311 1143801309 1143811309 1143811311 1143831309 1143831311 1143841309 1143851309 1143851311 1143861309 1143871309 1143871311 1143881309 1143881311 1143891309 1143901311 1143911309 1143911311 1143921311 1143931309 1143931311 1143951309 1143951311 1143971309 1143971311 1143991309 1143991311 1144011309 1144011311 1144031309 1144031311 1144041309 1144051309 1144051311 1144061309 1144071309 1144071311 1144081309 1144091309 1144091311 1144101309 1144111309 1144111311 1144121309 1144131309 1144131311 1144141309 1144151309 1144151311 1144161309 1144171309 1144171311 1144181309 1144191309 1144191311 1144201309 1144211309 1144211311 1144221309 1144231309 1144231311 1144241309 1144251309 1144251311 1144261309 1144271309 1144271311 1144281309 1144291309 1144291311 1144301309 1144311309 1144321309 1144331309 1144331311 1144341309 1144351309 1144351311 1144361309 1144371309 1144371311 1144381309 1144391309 1144391311 1144401309 1144411309 1144411311 1144421309 1144431309 1144431311 1144441309 1144451309 1144451311 1144461309 1144471309 1144471311 1144481309 1144491309 1144491311 1144501309 1144511309 1144511311 1144521309 1144531309 1144531311 1144541309 1144551311 1144561309 1144571309 1144571311 1144581309 1144591309 1144591311 1144601309 1144611309 1144611311 1144621309 1144631309 1144631311 1144641309 1144651309 1144651311 1144661309 1144671309 1144671311 1144691311 1144711311 1144721311 1144731311 1144741311 1144751311 1144761311 1144781311 1144791311 1144841311 1151811310 1151821310 1151831310 1151841310 1151851310 1151861310 1151871310 1151881310 1151891310 1151901310 1151911310 1151921310 1151931310 1151941310 1151951310 1151961310 1151971310 1151981310 1151991310 1152001310 1152011310 1152021310 1152031310 1152041310 1152051310 1152061310 1152071310 1152081310 1152091310 1152101310 1152111310 1152121310 1152131310 1152141310 1152151310 1152161310 1152171310 1152181310 1152191310 1152201310 1152211310 1152221310 1152231310 1152241310 1152251310 1152261310 1152271310 1152281310 1152291310 1152301310 1152311310 1152321310 1152331310 1152341310 1152351310 1152361310 1152371310 1152381310 1152391310 1152401310 1152411310 1152421310 1152431310 1152441310 1152451310 1152461310 1152471310 1152481310 1152491310 1152501310 1152511310 1152521310 1152531310 1152541310 1152551310 1152561310 1152571310 1152581310 1152591310 1152601310 1152611310 1152621310 1152631310 1152641310 1152651310 1152661310 1152671310 1152681310 1152691310 1152701310 1152711310 1152721310 1152731310 1152741310 1152751310 1152761310 1152771310 1153841310 1153851310 1153861310 1153891310 1153911310 1153931310 1153951310 1153971310 1153991310 1154011310 1154031310 1154051310 1154061310 1154071310 1154081310 1154091310 1154101310 1154111310 1154121310 1154131310 1154141310 1154151310 1154161310 1154171310 1154181310 1154191310 1154201310 1154211310 1154221310 1154231310 1154241310 1154251310 1154261310 1154271310 1154281310 1154291310 1154301310 1154311310 1154321310 1154331310 1154341310 1154351310 1154361310 1154371310 1154381310 1154391310 1154401310 1154411310 1154421310 1154431310 1154441310 1154451310 1154461310 1154471310 1154481310 1154491310 1154501310 1154511310 1154521310 1154531310 1154541310 1154551310 1154561310 1154571310 1154581310 1154591310 1154601310 1154611310 1154621310 1154631310 1154641310 1154651310 1154661310 1154671310 1154691310 1154711310 1154771310 1154781310 1154791310 1154801310 1154811310 1154821310 1154831310 1154841310 1154851310 1154861310 1154871310 1154881310 1154891310 1154901310 1154911310 1154921310 1154931310 1154941310 1154951310 1154961310 1159191311 1159211311 1159351311 1159361311 1159371311 1159391311 1159411311 1159431311 1159441311 1159451311 1159461311 1159471311 1159481311 1159491311 1159511311 1159531311 1159541311 1159551311 1159561311 1159571311 1159581311 1159591311 1159601311 1159611311 1159621311 1159631311 1159641311 1159651311 1159661311 1159671311 1159681311 1159691311 1159701311 1159711311 1159721311 1159741311 1159761311 1159781311 1159801311 1159821311 1159841311 1159921311 1159941311 1159961311 1159981311 1160001311 1160021311 1168021310 1168031310 1168041310 1168051310 1168061310 1168071310 1168081310 1168091310 1168101310 1168111310 1168121310 1168131310 1168141310 1168151310 1168161310 1168171310 1168181310 1168191310 1168201310 1168211310 1168221310 1168231310 1168241310 1168251310 1168261310 1168271310 1168281310 1168291310 1168301310 1168311310 1168321310 1168331310 1168341310 1168351310 1168361310 1168371310 1168381310 1168391310 1168401310 1168411310 1168421310 1168431310 1168451310 1168471310 1168481310 1168491310 1168501310 1168511310 1168521310 1168531310 1168541310 1168551310 1168561310 1168571310 1168581310 1168591310 1168611310 1168631310 1168651310 1168671310 1168691310 1168701310 1168711310 1168731310 1168751310 1168761310 1168771310 1168781310 1168791310 1168801310 1168811310 1168821310 1168831310 1168841310 1168851310 1168861310 1168871310 1168881310 1168891310 1168901310 1168911310 1168921310 1168931310 1168941310 1168951310 1168961310 1168981310 1168991310 1169001310 1169011310 1169021310 1169031310 1169041310 1169051310 1169061310 1169071310 1169081310 1169091310 1169101310 1169111310 1169121310 1169131310 1169141310 1188171309 1188181309 1188191309 1188201309 1188211309 1188231309 1188251309 1188271309 1188291309 1188311309 1188331309 1188351309 1188371309 1188391309 1188411309 1188431309 1188451309 1188471309 1188491309 1188501309 1188511309 1188521309 1188531309 1188541309 1188561309 1188571309 1188581309 1188591309 1188601309 1188611309 1188631309 1188651309 1188671309 1188681309 1188691309 1188701309 1188711309 1188721309 1188731309 1188741309 1188751309 1188761309 1188771309 1188781309 1188791309 1188801309 1188811309 1188821309 1188831309 1188841309 1188851309 1188861309 1188871309 1188881309 1188891309 1188901309 1188911309 1188921309 1188931309 1188941309 1188961309 1188981309 1189001309 1189011309 1189031309 1189051309 1189071309 1189091309 1189111309 1189131309 1189151309 1189171309 1189181309 1189191309 1189211309 1189231309 1189251309 1194821309 1194841309 1194861309 1194881309 1194901309 1194921309 1194941309 1194961309 1194981309 1195001309 1195021309 1195041309 1195061309 1195081309 1195101309 1195191309 1195211309 1195231309 1195251309 1195271309 1195291309 1195311309 1195331309 1195351309 1195391309 1195411309 1195431309 1195451309 1195471309 1195491309 1195511309 1195531309 1195551309 1195571309 1195581309 1195591309 1195601309 1195611309 1195621309 1195631309 1195641309 1195661309 1195671309 1195681309 1195701309 1195721309 1195741309 1195751309 1195761309 1195771309 1195781309 1195791309 1195801309 1195811309 1195821309 1195831309 1195841309 1195851309 1195861309 1195871309 1195881309 1195891309 1195911309 1195921309 1195931309 1195951309 1206431310 1206451310 1206471310 1206491310 1206501310 1206511310 1206531310 1206551310 1206571310 1206581310 1206591310 1206601310 1206611310 1206621310 1206631310 1206641310 1206651310 1206661310 1206671310 1206681310 1206691310 1206701310 1206711310 1206721310 1206731310 1206741310 1206751310 1206761310 1206771310 1206781310 1206801310 1206821310 1206841310 1206851310 1206861310 1206871310 1206881310 1206891310 1206901310 1206911310 1206921310 1206931310 1206941310 1206951310 1206971310 1206991310 1207001310 1207011310 1207021310 1207031310 1207041310 1207051310 1207061310 1207071310 1207091310 1207111310 1207131310 1207151310 1207171310 1207191310 1207211310 1207231310 1207251310 1207261310 1207281310 1207301310 1207321310 1207341310 1207361310 1207381310 1207401310 1207421310 1207441310 1207461310 1207481310 1207501310 1207521310 1210741310 1210751310 1210761310 1210771310 1210781310 1210791310 1210811310 1211131310 1211151310 1211171310 1211191310 1211211310 1211231310 1211251310 1211271310 1211291310 1211311310 1211331310 1211351310 1211371310 1211391310 1211411310 1211431310 1211451310 1211471310 1211491310 1212091311 1212111311 1212131311 1212151311 1212171311 1212191311 1212211311 1212231311 1212251311 1212291311 1212311311 1212331311 1212341311 1212351311 1212361311 1212371311 1212381311 1212391311 1212401311 1212411311 1212421311 1212431311 1212441311 1212451311 1212461311 1212471311 1212481311 1212491311 1212501311 1212511311 1212531311 1212541311 1212561311 1212571311 1212581311 1212601311 1212621311 1212641311 1212661311 1212681311 1212701311 1212721311 1212741311 1212761311 1212771311 1212781311 1212801311 1212821311 1212841311 1212861311 1212881311 1212901311 1212921311 1212941311 1212951311 1212961311 1212971311 1212991311 1213001311 1213021311 1213041311 1213061311 1213081311 1213101311 1213121311 1213131311 1213141311 1213151311 1213161311 1213171311 1213181311 1213191311 1213201311 1217621311 1217731311 1217751311 1217771311 1217781311 1217791311 1217801311 1217811311 1217821311 1217831311 1217841311 1217851311 1217861311 1217871311 1217881311 1217891311 1217901311 1217911311 1217921311 1217931311 1217941311 1217951311 1217961311 1217971311 1217981311 1217991311 1218001311 1218011311 1218021311 1218031311 1218041311 1218051311 1218061311 1218071311 1218081311 1218091311 1218101311 1218111311 1218121311 1218141311 1218161311 1218181311 1218201311 1218221311 1218241311 1218261311 1218281311 1218441311 1218461311 1218471311 1218481311 1218501311 1218521311 1218531311 1218561311 1218711311 1218731311 1238931311 1238941311 1238951311 1238961311 1238971311 1238981311 1238991311 1239001311 1239011311 1239021311 1239031311 1239041311 1239051311 1239061311 1239071311 1239081311 1239091311 1239101311 1239121311 1239141311 1239151311 1239161311 1239171311 1239181311 1239191311 1239201311 1239211311 1239221311 1239231311 1239241311 1239251311 1239261311 1239271311 1239281311 1239291311 1239301311 1239311311 1239321311 1239331311 1239341311 1239351311 1239361311 1239371311 1239381311 1239401311 1239421311 1239441311 1239461311 1239481311 1239501311 1239521311 1239541311 1239561311 1239581311 1239591311 1239601311 1239611311 1239621311 1239631311 1239641311 1239651311 1239661311 1239671311 1239681311 1239691311 1239701311 1239711311 1239721311 1239731311 1239741311 1239751311 1239761311 1239771311 1239781311 1239791311 1239801311 1239811311 1239821311 1239831311 1239841311 1239851311 1239861311 1239871311 1239881311 1239891311 1239901311 1239911311 1239921311 1239931311 1239941311 1239951311 1239961311 1239971311 1239981311 1240001311 1240021311 1253871310 1253891310 1253911310 1253931310 1253951310 1254081310 1254101310 1254121310 1254131310 1254141310 1254151310 1254161310 1254171310 1254181310 1254201310 1254221310 1254241310 1254261310 1254281310 1254301310 1254321310 1254341310 1254361310 1254381310 1254401310 1254421310 1254441310 1254461310 1254481310 1254501310 1254521310 1254541310 1254561310 1254581310 1254601310 1254621310 1254641310 1254661310 1254681310 1254701310 1254721310 1254741310 1254761310 1254781310 1254801310 1254821310 1254841310 1254861310 1254881310 1255851308 1255861308 1255871308 1255881308 1255901308 1255911308 1255921308 1255931308 1255941308 1255951308 1255961308 1255971308 1255981308 1255991308 1256001308 1256011308 1256021308 1256031308 1256041308 1256051308 1256061308 1256071308 1256081308 1256091308 1256101308 1256111308 1256121308 1256131308 1256141308 1256151308 1256161308 1256171308 1256181308 1256191308 1256201308 1256211308 1256221308 1256231308 1256241308 1256251308 1256261308 1256271308 1256281308 1256291308 1256301308 1256311308 1256321308 1256331308 1256341308 1256351308 1256361308 1256371308 1256381308 1256391308 1256401308 1256411308 1256421308 1256431308 1256441308 1256451308 1256461308 1256471308 1256481308 1256491308 1256501308 1256511308 1256521308 1256531308 1256541308 1256551308 1256561308 1256571308 1256581308 1256591308 1256601308 1256611308 1256621308 1256631308 1256641308 1256651308 1256661308 1256671308 1256681308 1256691308 1256701308 1256711308 1256721308 1256731308 1256741308 1256751308 1256761308 1256771308 1256781308 1256791308 1256801308 1256811308 1256821308 1256831308 1256841308 1256851308 1256861308 1256871308 1256881308 1258991308 1259001308 1259011308 1259021308 1259031308 1259041308 1259051308 1259061308 1259071308 1259081308 1259091308 1259101308 1259111308 1259121308 1259131308 1259141308 1259151308 1259161308 1259171308 1259181308 1259191308 1259201308 1259211308 1259221308 1259231308 1259241308 1259251308 1259261308 1259271308 1259281308 1259291308 1259301308 1259311308 1259321308 1259331308 1259341308 1259351308 1259361308 1259371308 1259381308 1259391308 1259401308 1259411308 1259421308 1259431308 1259441308 1259451308 1259461308 1259471308 1259481308 1259491308 1259501308 1259511308 1259521308 1259531308 1259541308 1259551308 1259561308 1259571308 1259581308 1259591308 1259601308 1259611308 1259621308 1259631308 1259641308 1259651308 1259661308 1259671308 1259681308 1259691308 1259701308 1259711308 1259721308 1259731308 1259741308 1259751308 1259761308 1259771308 1259781308 1259791308 1259801308 1259811308 1259821308 1259831308 1259841308 1259851308 1259861308 1259871308 1259881308 1259891308 1259901308 1259911308 1259921308 1259931308 1259941308 1259951308 1259961308 1259971308 1259981308 1259991308 1260001308 1260011308 1260021308 1260031308 1260041308 1260051308 1260061308 1260071308 1260081308 1260091308 1260101308 1260111308 1271411310 1271421310 1271431310 1271441310 1271451310 1271461310 1271471310 1271481310 1271491310 1271501310 1271521310 1271531310 1271561310 1271571310 1271581310 1271631310 1271681310 1271701310 1271721310 1271741310 1271751310 1271771310 1271781310 1271791310 1271801310 1271811310 1271851310 1271871310 1271891310 1271971310 1271991310 1272011310 1272031310 1272051310 1272241310 1272261310 1272301310 1272321310 1272341310 1272361310 1272371310 1272381310 1272391310 1272401310 1272411310 1272421310 1272431310 1272441310 1272451310 1272461310 1272471310 1272481310 1272491310 1272501310 1282081310 1282101310 1282111310 1282121310 1282141310 1282151310 1282161310 1282171310 1282181310 1282191310 1282211310 1282221310 1282501310 1282521310 1282541310 1282561310 1282581310 1282591310 1282601310 1282611310 1282621310 1282631310 1282641310 1282661310 1282681310 1282701310 1282721310 1282741310 1282761310 1282771310 1282781310 1282791310 1282801310 1282811310 1282821310 1282831310 1282851310 1282871310 1282891310 1282911310 1282931310 1282951310 1282971310 1282991310 1283011310 1283031310 1283051310 1286181308 1286201308 1286221308 1286241308 1286261308 1286281308 1286301308 1286321308 1286341308 1286361308 1286381308 1286401308 1286411308 1286421308 1286431308 1286441308 1286451308 1286461308 1286471308 1286481308 1286491308 1286501308 1286511308 1286521308 1286541308 1286611308 1286631308 1286651308 1286671308 1286691308 1286711308 1286731308 1286751308 1286771308 1286791308 1286811308 1286831308 1286851308 1286871308 1286891308 1286981308 1287001308 1287021308 1287061308 1287081308 1287101308 1298251308 1298271308 1298291308 1298311308 1298331308 1298351308 1298371308 1298391308 1298411308 1298431308 1298451308 1298471308 1298491308 1298511308 1298531308 1298551308 1298571308 1298581308 1298591308 1298601308 1298611308 1298621308 1298631308 1298641308 1298661308 1298681308 1298701308 1298721308 1298741308 1298761308 1298781308 1298801308 1298821308 1298831308 1298841308 1298851308 1298861308 1298871308 1298881308 1298891308 1298901308 1298911308 1298921308 1298931308 1298941308 1298951308 1298961308 1298971308 1298981308 1298991308 1299001308 1299011308 1299021308 1299031308 1299041308 1299051308 1299061308 1299071308 1299081308 1299091308 1299101308 1299111308 1299121308 1299131308 1299141308 1299151308 1299161308 1299171308 1299181308 1299201308 1299221308 1299241308 1299261308 1299281308 1299291308 1299301308 1299311308 1299321308 1299331308 1299341308 1299351308 1311841308 1311981308 1312591308 1312611308 1312631308 1312651308 1312671308 1312691308 1312711308 1312731308 1312751308 1312761308 1312771308 1312781308 1312791308 1312801308 1312811308 1312831308 1312851308 1312871308 1312891308 1312911308 1312921308 1312931308 1000441312 1000461312 1000481312 1000501312 1000521312 1000541312 1000561312 1000581312 1000851310 1000871310 1000891310 1000911310 1000931310 1000951310 1000971310 1000991310 1001011310 1001031310 1001051310 1001071310 1001091310 1001111310 1001131310 1001151310 1001171310 1001181310 1001191310 1001201310 1001221310 1001241310 1001261310 1001281310 1001301310 1001321310 1001331310 1001341310 1001351310 1001361310 1001371310 1001381310 1001391310 1001401310 1001411310 1001421310 1001431310 1001441310 1001451310 1001461310 1001471310 1001481310 1001491310 1001501310 1001521310 1001531310 1001541310 1001551310 1001561310 1001571310 1001591310 1001631310 1001651310 1001671310 1001721310 1001741310 1001751310 1001761310 1001781310 1001811310 1001831310 1001841310 1001851310 1001861310 1001871310 1001881310 1001901310 1001921310 1011201311 1011221311 1011241311 1011261311 1011291311 1011311311 1011351311 1011371311 1011381311 1011391311 1011401311 1011411311 1011421311 1011441311 1011461311 1011481311 1011521311 1011541311 1011561311 1011571311 1011581311 1011591311 1011601311 1011611311 1011621311 1011631311 1011641311 1011651311 1011661311 1011671311 1011681311 1011691311 1011701311 1011711311 1011721311 1011731311 1011741311 1011751311 1011761311 1011771311 1011781311 1011791311 1011801311 1011811311 1011821311 1011831311 1011841311 1011851311 1011861311 1011871311 1011881311 1011901311 1011921311 1011941311 1011961311 1011981311 1012001311 1012021311 1012041311 1012051311 1012061311 1012071311 1012081311 1012091311 1012101311 1012111311 1012121311 1012131311 1012141311 1012151311 1012161311 1012171311 1012181311 1012191311 1012201311 1012211311 1012221311 1012231311 1012241311 1012261311 1012271311 1012281311 1012291311 1012301311 1012311311 1017531312 1017711312 1017731312 1017751312 1017771312 1018111312 1018131312 1018151312 1018171312 1031891310 1031911310 1031931310 1031951310 1031971310 1031991310 1032011310 1032031310 1032051310 1032071310 1032091310 1032111310 1032131310 1032151310 1032191310 1032211310 1032231310 1032251310 1032271310 1032291310 1032311310 1032331310 1032351310 1032361310 1032371310 1032381310 1032391310 1032401310 1032411310 1032431310 1032441310 1032451310 1032461310 1032471310 1032481310 1032491310 1032501310 1032511310 1032521310 1032531310 1032541310 1032551310 1032581310 1032601310 1032621310 1032641310 1032661310 1032681310 1032701310 1032741310 1032761310 1032781310 1032791310 1032811310 1032831310 1032851310 1032871310 1032881310 1032891310 1032901310 1032911310 1032921310 1032931310 1032951310 1038231311 1038241311 1038251311 1038261311 1038271311 1038281311 1038291311 1038301311 1038311311 1038321311 1038331311 1038341311 1038351311 1038361311 1038371311 1038381311 1038391311 1038401311 1038411311 1038421311 1038431311 1038441311 1038451311 1038461311 1038471311 1038481311 1038491311 1038501311 1038511311 1038521311 1038531311 1038541311 1038551311 1038561311 1038571311 1038581311 1038591311 1038601311 1038611311 1038621311 1038631311 1038641311 1038651311 1038661311 1038671311 1038681311 1038691311 1038701311 1038711311 1038721311 1038731311 1038741311 1038751311 1038761311 1038771311 1038781311 1038791311 1038801311 1038811311 1038821311 1038831311 1038841311 1038851311 1038861311 1038871311 1038881311 1038891311 1038901311 1038911311 1038921311 1038931311 1038941311 1038951311 1038961311 1038971311 1038981311 1038991311 1039001311 1039011311 1039021311 1039031311 1039041311 1039051311 1039061311 1039071311 1039091311 1039111311 1039121311 1039131311 1039141311 1039151311 1039161311 1039171311 1039191311 1063421312 1078261310 1078281310 1078301310 1078321310 1078341310 1078381310 1078401310 1078421310 1078441310 1078491310 1078511310 1078521310 1078531310 1078541310 1078551310 1078561310 1078571310 1078581310 1078591310 1078601310 1078611310 1078621310 1078631310 1078641310 1078651310 1078661310 1078671310 1078681310 1078691310 1078701310 1078711310 1078721310 1078731310 1078741310 1078751310 1078761310 1078771310 1078781310 1078791310 1078801310 1078811310 1078821310 1078831310 1078841310 1078851310 1078861310 1078871310 1078881310 1078891310 1078901310 1078911310 1078931310 1078951310 1078971310 1078991310 1079061310 1079081310 1079101310 1079121310 1079141310 1079161310 1079181310 1079201310 1079221310 1079231310 1079241310 1079261310 1079271309 1079271310 1079281309 1079281310 1079291309 1079291310 1079301309 1079301310 1079311309 1079311310 1079321309 1079331309 1079331310 1079341309 1079351310 1079361309 1079371309 1079371310 1079381309 1079391309 1079401309 1079411309 1079421309 1079431309 1079441309 1079451309 1079461309 1079471309 1079491309 1079511309 1079531309 1079551309 1079571309 1079581309 1079591309 1079611309 1079631309 1079641309 1079651309 1079661309 1079671309 1079681309 1079691309 1079701309 1079711309 1079721309 1079731309 1079741309 1079751309 1079761309 1079771309 1079781309 1079791309 1079801309 1079811309 1079831309 1079851309 1079871309 1079891309 1079911309 1079931309 1079941309 1079951309 1079961309 1079981309 1080001309 1080021309 1080041309 1080061309 1080081309 1080101309 1080121309 1080141309 1080161309 1080181309 1080271309 1080291309 1080311309 1080331309 1080351309 1082761310 1082781310 1082801310 1082821310 1082841310 1082851310 1082861310 1082881310 1082891310 1082901310 1082921310 1082941310 1082961310 1082981310 1083031310 1083211310 1083231310 1083241310 1083251310 1083271310 1083291310 1083301310 1083311310 1083331310 1083341310 1083351310 1083361310 1083371310 1083381310 1083391310 1083401310 1083411310 1083421310 1083431310 1083451310 1083471310 1083491310 1083511310 1083531310 1083541310 1083571310 1083591310 1083611310 1083631310 1083651310 1083671310 1083691310 1083711310 1083731310 1083751310 1083771310 1083791310 1083811310 1083831310 1083851310 1083861310 1083871310 1083891310 1085511311 1085521311 1085531311 1085541311 1085551311 1085561311 1085571311 1085581311 1085591311 1085601311 1085611311 1085621311 1085631311 1085641311 1085651311 1085661311 1085671311 1085681311 1085691311 1085701311 1085721311 1085731311 1085741311 1085751311 1085761311 1085771311 1085781311 1085791311 1085801311 1085811311 1085821311 1085831311 1085841311 1085851311 1085861311 1085871311 1085881311 1085891311 1085901311 1085911311 1085921311 1085931311 1085941311 1085951311 1085961311 1085971311 1085981311 1085991311 1086001311 1086011311 1086021311 1086031311 1086041311 1086051311 1086071311 1086081311 1086091311 1086101311 1086111311 1086121311 1086131311 1086141311 1086181311 1086201311 1086221311 1086261311 1086281311 1086301311 1086351311 1086391311 1086401311 1086421311 1086441311 1086451311 1086461311 1086481311 1086501311 1086521311 1086541311 1086561311 1086581311 1086601311 1087231310 1087241310 1087261310 1087281310 1087301310 1087381310 1087501310 1087521310 1087541310 1087561310 1087581310 1087601310 1087621310 1087641310 1087661310 1087671310 1087681310 1087691310 1087701310 1087721310 1087741310 1087761310 1087781310 1087801310 1087821310 1087841310 1087861310 1087881310 1087901310 1087921310 1087941310 1088041310 1088061310 1088081310 1088101310 1088121310 1088141310 1088151310 1088161310 1088171310 1088181310 1088191310 1088201310 1088211310 1088221310 1088231310 1088241310 1088251310 1088261310 1088281310 1091011311 1091021311 1091031311 1091041311 1091051311 1091061311 1091071311 1091081311 1091091311 1091111311 1091131311 1091141311 1091151311 1091171311 1091191311 1091211311 1091231311 1091251311 1091271311 1091291311 1091311311 1091331311 1091351311 1091371311 1091391311 1091871311 1091891311 1091911311 1094371311 1094391311 1094411311 1094431311 1094451311 1094471311 1094491311 1094511311 1094551311 1094561311 1094581311 1094591311 1094601311 1094621310 1094621311 1094641310 1094641311 1094651310 1094661310 1094661311 1094671310 1094681310 1094681311 1094691310 1094701310 1094701311 1094711310 1094721310 1094721311 1094731310 1094741310 1094741311 1094751310 1094761310 1094761311 1094771310 1094781310 1094781311 1094791310 1094791311 1094801310 1094801311 1094811310 1094811311 1094821310 1094821311 1094831310 1094831311 1094841310 1094841311 1094851310 1094851311 1094861310 1094861311 1094871311 1094881310 1094881311 1094891311 1094901310 1094901311 1094911311 1094921310 1094921311 1094931310 1094931311 1094941310 1094941311 1094951310 1094951311 1094961310 1094961311 1094971310 1094971311 1094981310 1094991310 1094991311 1095001310 1095011310 1095011311 1095021310 1095031310 1095031311 1095041310 1095051310 1095051311 1095061310 1095071310 1095071311 1095081310 1095091310 1095091311 1095101310 1095111310 1095111311 1095121310 1095131310 1095131311 1095141310 1095151310 1095151311 1095161310 1095171310 1095171311 1095181310 1095191310 1095191311 1095201310 1095211310 1095221310 1095231310 1095241310 1095251310 1095261310 1095271310 1095281310 1095291310 1095301310 1095311310 1095331310 1095361310 1095381310 1103841311 1103851311 1103861311 1103871311 1103881311 1103891311 1103901311 1103911311 1103921311 1103931311 1103961311 1103981311 1104001311 1104021311 1104031311 1104041311 1104061311 1104071311 1104081311 1104091311 1104101311 1104121311 1104131311 1104141311 1104151311 1104161311 1104171311 1104181311 1104191311 1104201311 1104211311 1104221311 1104231311 1104241311 1104251311 1104271311 1104281311 1104291311 1104301311 1104311311 1104321311 1104331311 1104341311 1104351311 1104361311 1104371311 1104381311 1104391311 1104411311 1104421311 1104431311 1104441311 1104451311 1104461311 1104471311 1104481311 1104491311 1104501311 1104521311 1104531311 1104541311 1104561311 1104571311 1104581311 1104591311 1104611311 1104631311 1104651311 1104661311 1104671311 1104681311 1104691311 1104711311 1104721311 1104741311 1104751311 1104761311 1104771311 1104781311 1104791311 1104801311 1104811311 1104821311 1104831311 1104841311 1104851311 1104861311 1104871311 1132041311 1132061311 1132101311 1132121311 1132141311 1132161311 1132181311 1132201311 1132221311 1132261311 1132281311 1132301311 1132311311 1132321311 1132331311 1132341311 1132351311 1132361311 1132371311 1132381311 1132391311 1132401311 1132411311 1132421311 1132431311 1132441311 1132451311 1132461311 1132471311 1132481311 1132491311 1132501311 1132511311 1132521311 1132531311 1132541311 1132551311 1132561311 1132571311 1132581311 1132591311 1132601311 1132611311 1132621311 1132631311 1132641311 1132651311 1132661311 1132671311 1132681311 1132691311 1132701311 1132711311 1132731311 1132781311 1132791311 1132801311 1132811311 1132821311 1132831311 1132841311 1132851311 1132861311 1132871311 1132881311 1132891311 1132901311 1132911311 1132921311 1132931311 1133011311 1133031311 1133051311 1133071311 1145971311 1146011311 1146071311 1146081311 1146501311 1146521311 1146911311 1146931311 1147061311 1147081311 1147101311 1147121311 1147141311 1158321310 1158331310 1158341310 1158351310 1158361310 1158371310 1158381310 1158391310 1158401310 1158411310 1158431310 1158441310 1158461310 1158471310 1158481310 1158491310 1158501310 1158511310 1158531310 1158551310 1158571310 1158591310 1158661310 1158691310 1158721310 1158741310 1158751310 1158761310 1158781310 1158791310 1158801310 1158811310 1158821310 1158831310 1158841310 1158851310 1158861310 1158881310 1158901310 1158911310 1158921310 1158931310 1158941310 1158951310 1158961310 1158971310 1158981310 1158991310 1159001310 1159011310 1159021310 1159031310 1159041310 1159051310 1159071310 1159091310 1159111310 1159131310 1159151310 1159161310 1159171310 1159191310 1159211310 1159231310 1159251310 1159271310 1159281310 1159311310 1159321310 1159341310 1159361310 1159381310 1159401310 1159421310 1159441310 1159461310 1159481310 1163691310 1163701310 1163721310 1163741310 1163751310 1163761310 1163771310 1163781310 1163791310 1163801310 1163811310 1163821310 1163831310 1163841310 1163851310 1163861310 1163871310 1163881310 1163891310 1163901310 1163911310 1163921310 1163931310 1163951310 1163961310 1163971310 1163981310 1163991310 1164001310 1164011310 1164021310 1164031310 1164041310 1164051310 1164061310 1164071310 1164081310 1164121310 1164131310 1164141310 1164151310 1164161310 1164171310 1164181310 1164191310 1164201310 1164211310 1164221310 1164231310 1164241310 1164251310 1164261310 1164271310 1164281310 1164291310 1164301310 1164311310 1164321310 1164331310 1164341310 1164351310 1164361310 1164371310 1164381310 1164391310 1164401310 1164421310 1164441310 1164461310 1164511310 1164521310 1164531310 1164541310 1164551310 1164561310 1164571310 1164581310 1164591310 1164601310 1164611310 1164621310 1164631310 1164641310 1164651310 1164661310 1164671310 1164681310 1164691310 1164701310 1164711310 1164721310 1164731310 1164741310 1164751310 1164761310 1164771310 1182741309 1182751309 1182761309 1182771309 1182781309 1182791309 1182801309 1182811309 1182821309 1182831309 1182841309 1182851309 1182861309 1182871309 1182881309 1182891309 1182901309 1182911309 1182921309 1182931309 1182941309 1182951309 1182961309 1182971309 1182981309 1182991309 1183001309 1183011309 1183021309 1183031309 1183041309 1183061309 1183081309 1183101309 1183121309 1183141309 1183151309 1183161309 1183171309 1183181309 1183191309 1183201309 1183211309 1183221309 1183231309 1183241309 1183251309 1183261309 1183271309 1183281309 1183301309 1183311309 1183321309 1183331309 1183341309 1183351309 1183361309 1183381309 1183401309 1183421309 1183431309 1183441309 1183451309 1183461309 1183471309 1183481309 1183491309 1183501309 1183511309 1183521309 1183531309 1183541309 1183551309 1183561309 1183571309 1183581309 1183591309 1183601309 1183611309 1183621309 1183631309 1183641309 1183651309 1183661309 1183671309 1183691309 1183711309 1183731309 1183751309 1183771309 1183791309 1183811309 1183821309 1183831309 1183841309 1183851309 1183861309 1183871309 1183881309 1183891309 1183901309 1183911309 1183921309 1183931309 1183941309 1183951309 1183961309 1183971309 1183981309 1183991309 1184001309 1184011309 1184021309 1184031309 1184041309 1184051309 1184061309 1184071309 1184081309 1184091309 1184101309 1184111309 1184131309 1184141309 1184151309 1184161309 1184171309 1184181309 1184191309 1184201309 1184211309 1184231309 1184251309 1184271309 1184291309 1184311309 1184331309 1184351309 1184371309 1184391309 1184411309 1184431309 1184451309 1184471309 1184491309 1184511309 1184531309 1184551309 1184571309 1184591309 1184611309 1184631309 1184651309 1184661309 1184671309 1184691309 1184701309 1184711309 1184731309 1184771309 1184811309 1184851309 1184911309 1245941310 1245951310 1245961310 1245971310 1245981310 1245991310 1246001310 1246011310 1246021310 1246031310 1246041310 1246051310 1246061310 1246071310 1246081310 1246091310 1246101310 1246111310 1246121310 1246131310 1246141310 1246151310 1246161310 1246171310 1246181310 1246191310 1246201310 1246211310 1246221310 1246231310 1246241310 1246251310 1246261310 1246271310 1246281310 1246291310 1246301310 1246311310 1246321310 1246341310 1246361310 1246381310 1246401310 1246421310 1246441310 1246461310 1246481310 1246501310 1246521310 1246541310 1246561310 1246581310 1246601310 1246611310 1246621310 1246641310 1246661310 1246681310 1246701310 1246721310 1246741310 1246751310 1246761310 1246771310 1246781310 1246791310 1246801310 1246811310 1246821310 1246831310 1246851310 1246871310 1246891310 1246911310 1246921310 1246931310 1246941310 1246951310 1246961310 1246971310 1246981310 1246991310 1247001310 1247011310 1247021310 1247031310 1247041310 1247061310 1247081310 1247091310 1257831309 1257851309 1257871309 1257881309 1257891309 1257901309 1257911309 1257921309 1257931309 1257941309 1257951309 1257981309 1257991309 1258011309 1258021309 1258031309 1258041309 1258051309 1258061309 1258071309 1258081309 1258091309 1258101309 1258111309 1258121309 1258131309 1258141309 1258151309 1258161309 1258171309 1258191309 1258211309 1258231309 1258251309 1258271309 1258291309 1258311309 1258321309 1258331309 1258341309 1258351309 1258361309 1258371309 1258381309 1258391309 1258401309 1258411309 1258421309 1258431309 1258441309 1258451309 1258461309 1258471309 1258481309 1258491309 1258501309 1258511309 1258521309 1258531309 1258541309 1258551309 1258561309 1258571309 1258581309 1258591309 1258601309 1258611309 1258621309 1258631309 1258641309 1258651309 1258661309 1258671309 1258681309 1258691309 1258701309 1258711309 1258721309 1258731309 1258741309 1258751309 1258761309 1258771309 1266481308 1266491308 1266501308 1266511308 1266521308 1266531308 1266541308 1266551308 1266571308 1266581308 1266591308 1266601308 1266611308 1266621308 1266631308 1266641308 1266651308 1266661308 1266671308 1266681308 1266691308 1266701308 1266711308 1266721308 1266731308 1266741308 1266751308 1266761308 1266771308 1266781308 1266791308 1266801308 1266811308 1266821308 1266831308 1266851308 1266861308 1266871308 1266881308 1266891308 1266911308 1266921308 1266931308 1266951308 1266971308 1266991308 1267011308 1267021308 1267031308 1267041308 1267051308 1267061308 1267071308 1267081308 1267091308 1267101308 1267111308 1267121308 1267131308 1267141308 1267151308 1267161308 1267171308 1267181308 1267191308 1267201308 1267211308 1267221308 1267231308 1267241308 1267251308 1267261308 1267271308 1267281308 1267291308 1267301308 1267311308 1267321308 1267331308 1267351308 1267361308 1267371308 1267381308 1267391308 1267401308 1267411308 1267421308 1267431308 1267441308 1267451308 1267461308 1267471308 1267481308 1267491308 1267501308 1267511308 1267521308 1267531308 1268111310 1268131310 1268151310 1268171310 1268191310 1268271310 1268291310 1268311310 1268331310 1268351310 1268371310 1268461310 1268481310 1268501310 1268521310 1268541310 1268561310 1268571310 1268581310 1268591310 1268611310 1268621310 1268631310 1268641310 1268651310 1268661310 1268671310 1268681310 1268691310 1268701310 1268711310 1268721310 1268731310 1268751310 1268771310 1268791310 1268811310 1268821310 1268831310 1268841310 1268851310 1268861310 1268871310 1268891310 1268901310 1268911310 1268921310 1268931310 1268941310 1268961310 1268981310 1269001310 1269021310 1269031310 1269041310 1269051310 1269061310 1269071310 1269081310 1269091310 1269101310 1269111310 1269121310 1269131310 1269141310 1269151310 1269161310 1291391308 1291411308 1291431308 1291451308 1291471308 1291491308 1291511308 1291531308 1291551308 1291571308 1291591308 1291611308 1291631308 1291651308 1291671308 1291691308 1291711308 1291731308 1291751308 1291761308 1291771308 1291781308 1291801308 1291811308 1291821308 1291831308 1291841308 1291851308 1291861308 1291871308 1291881308 1291891308 1291901308 1291911308 1291921308 1291931308 1291941308 1291951308 1291961308 1291971308 1291981308 1291991308 1292001308 1292011308 1292021308 1292031308 1292051308 1292071308 1292091308 1292111308 1292131308 1292151308 1292171308 1292191308 1292221308 1292251308 1292271308 1292281308 1292291308 1292351308 1292371308 1292401308 1292421308 1292441308 1292461308 1292501308 1302171308 1302181308 1302201308 1302221308 1302261308 1302281308 1302291308 1302301308 1302311308 1302321308 1302331308 1302341308 1302351308 1302361308 1302371308 1302381308 1302391308 1302401308 1302411308 1302421308 1302441308 1302451308 1302461308 1302471308 1302481308 1302491308 1302501308 1302511308 1302521308 1302531308 1302551308 1302561308 1302581308 1302591308 1302601308 1302611308 1302621308 1302631308 1302641308 1302651308 1302661308 1302671308 1302681308 1302691308 1302701308 1302711308 1302721308 1302731308 1302741308 1302751308 1302761308 1302771308 1302781308 1302791308 1302801308 1302811308 1302821308 1302831308 1302841308 1302861308 1302871308 1302881308 1302901308 1302921308 1302941308 1302961308 1302981308 1302991308 1303001308 1303011308 1303021308 1303031308 1303041308 1303051308 1303061308 1303071308 1303081308 1303091308 1303111308 1303121308 1303131308 1303151308 1303161308 1303171308 1303181308 1303191308 1303211308 1303221308 1303241308 1303251308 1303261308 1303271308 1303291308 1303301308 1303311308 1303331308 1307531308 1307541308 1307551308 1307561308 1307571308 1307581308 1307591308 1307601308 1307611308 1307631308 1307641308 1307651308 1307671308 1307681308 1307691308 1307711308 1307731308 1307741308 1307751308 1307761308 1307771308 1307781308 1307791308 1307801308 1307821308 1307831308 1307841308 1307861308 1307871308 1307881308 1307891308 1307901308 1307911308 1307921308 1307931308 1307941308 1307951308 1307961308 1307971308 1307981308 1307991308 1308001308 1308011308 1308021308 1308031308 1308041308 1308051308 1308061308 1308071308 1308081308 1308091308 1308101308 1308111308 1308121308 1308131308 1308141308 1308151308 1308161308 1308171308 1308181308 1308191308 1308201308 1308211308 1308221308 1308231308 1308241308 1308251308 1308261308 1308271308 1308281308 1308291308 1308301308 1308311308 1308321308 1308331308 1308341308 1308351308 1308361308 1308371308 1308381308 1308391308 1308401308 1308411308 1308421308 1308431308 1308441308 1308451308 1308461308 1308471308 1308481308 1308491308 1308501308 1308521308 1308531308 1308541308 1308561308 1308581308 1308601308 1308621308 1308641308 1308651308 1308661308 1308671308 1308681308 1308691308 |
Recalling Firm/ Manufacturer |
AMO Puerto Rico Manufacturing, Inc. Rd # 402, Km 4.2 Anasco PR 00610
|
Manufacturer Reason for Recall |
Units may be labeled with the incorrect diopter power.
|
FDA Determined Cause 2 |
Error in labeling |
Action |
AMO sent an Urgent Medical Device Recall dated May 13, 2014 to the affected customers. The letter identified the affected product, problem and actions to be taken. This recall only affects the specific Silicone Tecnis IOLs Model Z9002. Implanted IOLs with appropriate post-op refractive outcomes are not affected by this recall. Please underatake the following actions:
1. Stop using and remove from your inventory the affected product.
2.Complete and return the included Facsimile Form as AMO.
If you have inventory to return, please complete the Facsimile Form, and returned to AMO. For questions contact AMO Customer Service Representative 877-266-4543. |
Distribution |
Worldwide Distribution - Nationwide (USA and Puerto Rico), and foreign: Latin America, Asia Pacific, Canada, Europe, Africa and Middle East, and Japan. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = HQL and Original Applicant = Johnson & Johnson Surgical Vision, Inc.
|