• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Customed

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall Customed see related information
Date Initiated by Firm May 20, 2014
Date Posted August 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall Number Z-2303-2014
Recall Event ID 68536
Product Classification Obstetrical kit - Product Code OKV
Product Customed O.B. Pack
convenience pack include multiple components: set up cover, absorbent towel, ear ulcer syringe, pad, umbilical cord clamp, wash basin, baby blankets, table cover, gauze sponge, syringe 10CC w/o needle & 10CC with needle, feeding tube, drape, legging with cuff, underbuttock drape, gown, container with lid, mayo tray, receiving blanket, & baby beannies.

EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents
Code Information Product code 900-012, 24 lots: 113036306 113015685 112030796 112041172 112062299 112083042 112051659 112093835 113046992 113047150 113015422 113057828 112124902 131110142 113089148 113099380 113078658 131210664 140111130 140211621 140311909 140412444 140513051 140613549 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information Contact Rosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
Action Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce 24 lots, 904 units (multiple units per lot)
Distribution Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-