 Class 1 Device Recall Customedsee related information
Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2308-2014
Recall Event ID 68536
Product Classification General surgery tray (kit) - Product Code LRO
ProductCustomed Femoral Pack , convenience pack include multiple components: Syringe 3cc, 5cc, 10cc, 20cc; swabstick chlorascrub, needles, gauze/sponges, table cover, absrobent towels, halsted Mosquito CVD, gown, bowls, wrapper, scalpel safety, Mosquito forcep, cloth huck, clamp towel, fluorocover, medicine cup, femoral angiodrape, mayo tray, counter & labels. EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents
Code Information Product code 900-164, 14 lots: 110061393 113046935 113047213 113047335 113057700 113068153 113109743 131110191 131110600 140111164 140211648 140311924 140513392 140513453 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce14 lots; 855 units (multiple units per lot)
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-