  Class 1 Device Recall Customed see related information
Date Initiated by Firm May 20, 2014
Date Posted August 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall Number Z-2382-2014
Recall Event ID 68536
Product Classification Cesarean section kit - Product Code OHM
Product PREMIUM CESAREAN PACK- CUSTOMED-

(1)BABY BLANKET IMP. 30" X 30" POLY ABSORBENT
(3) TOWELS ABSORBENT 15" X 20" LIF
(1) BOWL PLASTIC 80oz WITH LID
(10) LAP SPONGE PREWASH XRD LIF
(2) PAD OBSTETRICAL X-LARGE ST.
(1) MAYO STAND COVER REINFORCED LIF
(1) DRAPE SHEET 41 " X 58" SMS LIF
(1) TUBE SUCTION CONNECT. X" X 12' LIF
(1) EAR ULCER SYRINGE 2oz LIF
(1) YANKAUER SUCTION TUBE WITHOUT VENT LIF
(1) CORD UMBILICAL CLAMP LIF
(1) DRESSING ISLAND 4" X 10"
(1) BLADE SURGICAL#10 CARBON STEEL
(1) BLADE SURGICAL#20 CARBON STEEL
(1) TABLE COVER REINFORCED 50" X 90" LIF
(1) C-SECTION DRAPE WITH POUCH
(1) GOWN IMPERVIOUS REINFORCED X-LARGE SMS AAMI III
(1) GOWN SURG. REINFORCED LARGE AAMI III TOWEL WRAP

EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information Product code 900-1294, 18 lots: 111030560 111030815 112020527 112114630 112124933 113015688 113026042 113036408 113057858 113089182 113109717 131110166 131210694 140211638 140312063 140412594 140513078 140513477 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information Contact Rosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
Action Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce 18 lots; 730 units (multiple units per lot)
Distribution Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-