|
Class 2 Device Recall Zenith Slide WIDE Long Term Care Bed |
|
Date Initiated by Firm |
July 21, 2014 |
Date Posted |
September 16, 2014 |
Recall Status1 |
Terminated 3 on August 11, 2015 |
Recall Number |
Z-2656-2014 |
Recall Event ID |
69057 |
Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
Product |
Zenith Slide W-I-D-E Long Term Care Bed, Model(s) APS98674, APS9867406, APS986741, APS986742, ZG58674, ZG78674, ZZ78674, ZZ98674, and ZZ98774.
Designed for use within an institutional healthcare environment (i.e. assisted living, skilled nursing, transitional care, rehabilitation centers, etc.). |
Code Information |
Model # Serial # APS98674 1161998674140123 APS98674 1169000522 APS98674 1161998674130104 APS98674 1161998674130105 APS98674 1169000642 APS98674 1169000643 APS98674 1161998674130049 APS98674 1161998674130002 APS98674 1161998674130003 APS98674 1161998674130004 APS98674 1161998674130005 APS98674 1161998674130006 APS98674 1161998674130007 APS98674 1161998674130008 APS98674 1161998674130009 APS98674 1161998674130010 APS98674 1161998674130011 APS98674 1161998674130012 APS98674 1161998674130013 APS98674 1161998674130014 APS98674 1161998674130015 APS98674 1161998674130016 APS98674 1161998674130017 APS98674 1161998674130018 APS98674 1161998674130019 APS98674 1161998674130020 APS98674 1161998674130021 APS98674 1161998674130022 APS98674 1161998674130023 APS98674 1161998674130024 APS98674 1161998674130025 APS98674 1161998674130026 APS98674 1161998674130027 APS98674 1161998674130028 APS98674 1161998674130029 APS98674 1161998674130030 APS98674 1161998674130031 APS98674 1161998674130032 APS98674 1161998674130033 APS98674 1161998674130034 APS98674 1161998674130035 APS98674 1161998674130036 APS98674 1161998674130037 APS98674 1161998674130038 APS98674 1161998674130039 APS98674 1161998674130040 APS98674 1161998674130041 APS98674 1161998674130042 APS98674 1161998674130043 APS98674 1161998674130044 APS98674 1161998674130046 APS98674 1161998674130047 APS98674 1161998674130048 APS98674 1161998674130103 APS98674 2626903676 APS98674 1161998674130001 APS98674 1169000889 APS98674 1169000890 APS98674 1169000891 APS98674 1169000892 APS98674 1169000893 APS98674 1169000894 APS98674 1169000895 APS98674 1169000896 APS98674 1169000897 APS98674 1169000898 APS98674 1169000899 APS98674 1169000900 APS98674 1169000901 APS98674 1169000902 APS98674 1169000903 APS98674 1169000826 APS98674 1169000905 APS98674 1161998674130106 APS98674 1161998674130107 APS98674 1161998674130108 APS98674 1161998674130109 APS98674 1161998674130110 APS98674 1161998674130111 APS98674 1161998674130112 APS98674 1161998674130113 APS98674 1161998674130114 APS98674 1161998674130115 APS98674 1161998674130116 APS98674 1161998674130117 APS98674 1161998674130118 APS98674 1161998674130119 APS98674 1161998674130120 APS98674 1161998674130121 APS98674 1161998674130045 APS98674 1161998674130051 APS98674 1161998674140124 APS98674 1161998674140125 APS98674 1169000494 APS98674 1161998674130122 APS98674 1169000904 APS9867406 116199810130001 APS9867406 116199810130110 APS9867406 116199810130111 APS9867406 116199810130112 APS9867406 116199810130113 APS9867406 116199810130114 APS9867406 116199810130115 APS9867406 116199810130116 APS9867406 116199810130117 APS9867406 116199810130118 APS9867406 116199810130119 APS986741 116199867411338 APS986741 116199867411343 APS986741 116199867411335 APS986741 116199867411301 APS986741 116199867411302 APS986741 116199867411303 APS986741 116199867411337 APS986741 116199867411339 APS986741 116199867411341 APS986741 116199867411342 APS986741 116199867411306 APS986741 116199867411307 APS986741 116199867411308 APS986741 116199867411309 APS986741 116199867411310 APS986741 116199867411311 APS986741 116199867411312 APS986741 116199867411313 APS986741 116199867411314 APS986741 116199867411315 APS986741 116199867411316 APS986741 116199867411317 APS986741 116199867411318 APS986741 116199867411319 APS986741 116199867411320 APS986741 116199867411321 APS986741 116199867411322 APS986741 116199867411323 APS986741 116199867411324 APS986741 116199867411325 APS986741 116199867411329 APS986741 116199867411330 APS986741 116199867411331 APS986741 116199867411332 APS986741 116199867411333 APS986741 1169000549 APS986741 1169000550 APS986741 1169000551 APS986741 1 APS986741 1169000220 APS986741 116199867411340 APS986742 11619986742140006 APS986742 11619986742140007 APS986742 11619986742140008 APS986742 11619986742140009 APS986742 11619986742140010 APS986742 11619986742140011 APS986742 11619986742140012 APS986742 11619986742140013 APS986742 11619986742140014 APS986742 11619986742140015 APS986742 11619986742140016 APS986742 11619986742140017 APS986742 11619986742140018 APS986742 11619986742140019 APS986742 11619986742140020 APS986742 11619986742140021 APS986742 11619986742140022 APS986742 11619986742140023 APS986742 11619986742140024 APS986742 11619986742140025 APS986742 11619986742130003 APS986742 11619986742140026 APS986742 1169000746 APS986742 11619986742130005 ZG58674 26075867141301 ZG78674 260778674140003 ZG78674 26077000026 ZG78674 26077000027 ZG78674 260778674140073 ZG78674 26077000029 ZG78674 260778674130001 ZZ78674 262678671430715 ZZ78674 262678671430716 ZZ78674 262678671430791 ZZ78674 262678671430792 ZZ78674 262678671430793 ZZ78674 26267867430064 ZZ78674 262678671430587 ZZ78674 262678671430598 ZZ78674 262678671430609 ZZ78674 262678671430610 ZZ78674 262678671430611 ZZ78674 262678671430612 ZZ78674 26267867430563 ZZ78674 26267867430564 ZZ78674 26267867430565 ZZ78674 26267867430566 ZZ78674 26267867430324 ZZ78674 26267867430325 ZZ78674 26267867430326 ZZ78674 26267867430327 ZZ78674 26267867430328 ZZ78674 26267867430381 ZZ78674 26267867430382 ZZ78674 26267867430383 ZZ78674 26267867430384 ZZ78674 26267867430385 ZZ78674 26267009193 ZZ78674 26267009194 ZZ78674 26267009195 ZZ78674 26267009196 ZZ78674 262678671430588 ZZ78674 262678671430589 ZZ78674 262678671430591 ZZ78674 262678671430592 ZZ78674 262678671430617 ZZ78674 262678671430618 ZZ78674 262678671430619 ZZ78674 262678671430620 ZZ78674 262678671430621 ZZ78674 262678671430622 ZZ78674 262678671430623 ZZ78674 262678671430624 ZZ78674 262678671430625 ZZ78674 262678671430626 ZZ78674 262678671430627 ZZ78674 262678671430628 ZZ78674 262678671430629 ZZ78674 262678671430630 ZZ78674 262678671430631 ZZ78674 262678671430632 ZZ78674 262678671430633 ZZ78674 262678671430634 ZZ78674 262678671430635 ZZ78674 262678671430636 ZZ78674 262678671430637 ZZ78674 262678671430638 ZZ78674 262678671430639 ZZ78674 262678671430640 ZZ78674 262678671430641 ZZ78674 262678671430642 ZZ78674 262678671430643 ZZ78674 262678671430644 ZZ78674 262678671430645 ZZ78674 262678671430646 ZZ78674 262678671430647 ZZ78674 262678671430676 ZZ78674 262678671430720 ZZ78674 262678671430725 ZZ78674 26267867430001 ZZ78674 26267867430002 ZZ78674 26267867430003 ZZ78674 26267867430004 ZZ78674 26267867430005 ZZ78674 26267867430006 ZZ78674 26267867430007 ZZ78674 26267867430008 ZZ78674 26267867430194 ZZ78674 26267867430195 ZZ78674 26267867430196 ZZ78674 26267867430197 ZZ78674 26267867430198 ZZ78674 26267867430199 ZZ78674 26267867430200 ZZ78674 26267867430201 ZZ78674 26267867430202 ZZ78674 26267867430203 ZZ78674 26267867430204 ZZ78674 26267867430205 ZZ78674 26267867430206 ZZ78674 26267867430207 ZZ78674 26267867430208 ZZ78674 26267867430209 ZZ78674 26267867430210 ZZ78674 26267867430211 ZZ78674 26267867430212 ZZ78674 26267867430213 ZZ78674 26267867430214 ZZ78674 26267867430215 ZZ78674 26267867430216 ZZ78674 26267867430217 ZZ78674 26267867430218 ZZ78674 26267867430219 ZZ78674 26267867430220 ZZ78674 26267867430221 ZZ78674 26267867430222 ZZ78674 26267867430223 ZZ78674 26267867430224 ZZ78674 26267867430225 ZZ78674 26267867430226 ZZ78674 26267867430227 ZZ78674 26267867430228 ZZ78674 26267867430229 ZZ78674 26267867430230 ZZ78674 26267867430231 ZZ78674 26267867430232 ZZ78674 26267867430233 ZZ78674 26267867430234 ZZ78674 26267867430235 ZZ78674 26267867430236 ZZ78674 26267867430237 ZZ78674 26267867430238 ZZ78674 26267867430239 ZZ78674 26267867430240 ZZ78674 26267867430241 ZZ78674 26267867430368 ZZ78674 26267867430412 ZZ78674 26267867430413 ZZ78674 26267867430414 ZZ78674 26267867430415 ZZ78674 26267867430416 ZZ78674 26267867430417 ZZ78674 26267867430418 ZZ78674 26267867430419 ZZ78674 26267867430420 ZZ78674 26267867430421 ZZ78674 26267867430422 ZZ78674 26267867430423 ZZ78674 26267867430425 ZZ78674 26267867430426 ZZ78674 26267867430428 ZZ78674 26267867430429 ZZ78674 26267867430430 ZZ78674 26267867430431 ZZ78674 26267867430432 ZZ78674 26267867430433 ZZ78674 26267867430434 ZZ78674 26267867430435 ZZ78674 26267867430436 ZZ78674 26267867430437 ZZ78674 26267867430438 ZZ78674 26267867430439 ZZ78674 26267867430440 ZZ78674 26267867430441 ZZ78674 26267867430442 ZZ78674 26267867430443 ZZ78674 26267867430444 ZZ78674 26267867430445 ZZ78674 26267867430446 ZZ78674 26267867430447 ZZ78674 26267867430448 ZZ78674 26267867430449 ZZ78674 26267867430450 ZZ78674 26267867430451 ZZ78674 26267867430452 ZZ78674 26267867430453 ZZ78674 26267867430454 ZZ78674 26267867430455 ZZ78674 26267867430456 ZZ78674 26267867430457 ZZ78674 26267867430458 ZZ78674 26267867430459 ZZ78674 26267867430502 ZZ78674 26267867430523 ZZ78674 26267867430560 ZZ78674 26267867430561 ZZ78674 26267867430562 ZZ78674 26267867430567 ZZ78674 26267867430568 ZZ78674 26267867430569 ZZ78674 26267867430570 ZZ78674 26267867430571 ZZ78674 26267867430572 ZZ78674 26267867430573 ZZ78674 26267867430574 ZZ78674 26267867430575 ZZ78674 26267867430576 ZZ78674 26267867430577 ZZ78674 26267867430578 ZZ78674 26267867430503 ZZ78674 26267867430504 ZZ78674 262678671430679 ZZ78674 262678671430663 ZZ78674 262678671430657 ZZ78674 262678671430794 ZZ78674 262678671430671 ZZ78674 262678671430672 ZZ78674 262678671430673 ZZ78674 262678671430674 ZZ78674 262678671430675 ZZ78674 26267867430133 ZZ78674 26267867430134 ZZ78674 26267867430135 ZZ78674 26267867430136 ZZ78674 26267867430137 ZZ78674 26267867430250 ZZ78674 26267867430251 ZZ78674 26267867430252 ZZ78674 26267867430253 ZZ78674 26267867430254 ZZ78674 26267867430257 ZZ78674 26267867430258 ZZ78674 26267867430259 ZZ78674 26267867430260 ZZ78674 26267867430261 ZZ78674 26267867430286 ZZ78674 26267867430287 ZZ78674 26267867430288 ZZ78674 26267867430289 ZZ78674 26267867430290 ZZ78674 26267867430303 ZZ78674 26267867430304 ZZ78674 26267867430305 ZZ78674 26267867430306 ZZ78674 26267867430307 ZZ78674 26267867430313 ZZ78674 26267867430314 ZZ78674 26267867430315 ZZ78674 26267867430316 ZZ78674 26267867430317 ZZ78674 26267867430342 ZZ78674 26267867430343 ZZ78674 26267867430344 ZZ78674 26267867430345 ZZ78674 26267867430346 ZZ78674 26267867430360 ZZ78674 26267867430361 ZZ78674 26267867430362 ZZ78674 26267867430363 ZZ78674 26267867430364 ZZ78674 26267867430397 ZZ78674 26267867430398 ZZ78674 26267867430399 ZZ78674 26267867430400 ZZ78674 26267867430401 ZZ78674 26267867430402 ZZ78674 26267867430403 ZZ78674 26267867430404 ZZ78674 26267867430405 ZZ78674 26267867430406 ZZ78674 26267867430579 ZZ78674 26267867430580 ZZ78674 26267867430581 ZZ78674 26267009303 ZZ78674 262678671430595 ZZ78674 26267867430546 ZZ78674 26267867430547 ZZ78674 26267867430548 ZZ78674 26267867430549 ZZ78674 26267867430550 ZZ78674 26267867430558 ZZ78674 26267867430559 ZZ78674 262678671430712 ZZ78674 26267867430062 ZZ78674 26267867430063 ZZ78674 262678671430728 ZZ78674 262678671430729 ZZ78674 26267867430193 ZZ78674 26267867430365 ZZ78674 26267867430366 ZZ78674 26267867430367 ZZ78674 26267867430061 ZZ78674 26267867430387 ZZ78674 26267867430388 ZZ78674 26267867430389 ZZ78674 26267867430390 ZZ78674 26267008151 ZZ78674 26267008391 ZZ78674 26267008152 ZZ78674 262678671430696 ZZ78674 26267867430312 ZZ78674 26267867430318 ZZ78674 262678671430680 ZZ78674 262678671430681 ZZ78674 262678671430682 ZZ78674 262678671430683 ZZ78674 262678671430684 ZZ78674 262678671430692 ZZ78674 262678671430693 ZZ78674 262678671430694 ZZ78674 262678671430695 ZZ78674 262678671430707 ZZ78674 262678671430711 ZZ78674 26267009293 ZZ78674 26267009294 ZZ78674 26267009295 ZZ78674 26267009296 ZZ78674 26267009297 ZZ78674 26267009305 ZZ78674 26267009306 ZZ78674 26267009307 ZZ78674 262678671430662 ZZ78674 26267009089 ZZ78674 26267867430551 ZZ78674 26267867430255 ZZ78674 26267867430276 ZZ78674 26267867430277 ZZ78674 26267867430278 ZZ78674 26267867430279 ZZ78674 26267867430280 ZZ78674 26267867430291 ZZ78674 26267867430010 ZZ78674 26267867430508 ZZ78674 26267867430509 ZZ78674 26267867430510 ZZ78674 26267867430511 ZZ78674 26267867430512 ZZ78674 26267867430513 ZZ78674 26267867430514 ZZ78674 26267867430515 ZZ78674 26267867430516 ZZ78674 26267867430517 ZZ78674 26267867430518 ZZ78674 26267867430519 ZZ78674 26267867430520 ZZ78674 26267867430521 ZZ78674 26267867430522 ZZ78674 262678671430705 ZZ78674 262678671430706 ZZ78674 26267009300 ZZ78674 26267867430009 ZZ78674 26267867430411 ZZ78674 26267009301 ZZ78674 26267008725 ZZ78674 26267867430018 ZZ78674 26267008153 ZZ78674 262678671430604 ZZ78674 262678671430605 ZZ78674 262678671430606 ZZ78674 262678671430607 ZZ78674 262678671430741 ZZ78674 262678671430742 ZZ78674 26267867430191 ZZ78674 26267867430243 ZZ78674 26267867430244 ZZ78674 26267867430245 ZZ78674 26267867430246 ZZ78674 262678671430795 ZZ78674 26267008150 ZZ78674 26267008733 ZZ78674 26267867430127 ZZ78674 26267867430248 ZZ78674 26267867430249 ZZ78674 26267867430175 ZZ78674 262678671430666 ZZ78674 262678671430667 ZZ78674 262678671430668 ZZ78674 262678671430669 ZZ78674 262678671430670 ZZ78674 262678671430661 ZZ78674 26267867430050 ZZ78674 26267867430060 ZZ78674 26267009090 ZZ78674 262678671430721 ZZ78674 262678671430599 ZZ78674 262678671430600 ZZ78674 262678671430601 ZZ78674 262678671430602 ZZ78674 262678671430613 ZZ78674 262678671430614 ZZ78674 262678671430615 ZZ78674 262678671430616 ZZ78674 262678671430714 ZZ78674 262678671430743 ZZ78674 262678671430776 ZZ78674 262678671430777 ZZ78674 262678671430778 ZZ78674 262678671430779 ZZ78674 262678671430780 ZZ78674 262678671430781 ZZ78674 262678671430782 ZZ78674 262678671430783 ZZ78674 262678671430784 ZZ78674 262678671430785 ZZ78674 26267867430321 ZZ78674 26267867430544 ZZ78674 26267008723 ZZ78674 26267008411 ZZ78674 26267008412 ZZ78674 262678671430590 ZZ78674 262678671430608 ZZ78674 262698674130160 ZZ78674 262698674130161 ZZ78674 262698674130182 ZZ78674 262698674130183 ZZ78674 26267867430019 ZZ78674 26267867430140 ZZ78674 26267867430141 ZZ78674 26267867430142 ZZ78674 26267867430143 ZZ78674 26267867430144 ZZ78674 26267867430153 ZZ78674 26267009069 ZZ78674 26267009070 ZZ78674 26267009071 ZZ78674 26267009072 ZZ78674 26267009073 ZZ78674 26267009074 ZZ78674 26267009075 ZZ78674 26267009076 ZZ78674 26267009077 ZZ78674 26267009078 ZZ78674 26267009079 ZZ78674 26267009080 ZZ78674 26267009081 ZZ78674 26267009082 ZZ78674 26267009083 ZZ78674 26267009084 ZZ78674 26267009085 ZZ78674 26267009086 ZZ78674 26267009087 ZZ78674 26267009088 ZZ78674 26267008402 ZZ78674 26267008403 ZZ78674 26267008404 ZZ78674 26267008728 ZZ78674 26267008729 ZZ78674 26267008730 ZZ78674 26267009068 ZZ78674 26267867430192 ZZ78674 26267867430500 ZZ78674 26267867430501 ZZ78674 26267008410 ZZ78674 26267009308 ZZ78674 26267009309 ZZ78674 26267867430051 ZZ78674 26267867430151 ZZ78674 26267867430152 ZZ78674 26267009064 ZZ78674 26267009065 ZZ78674 26267009066 ZZ78674 26267009067 ZZ78674 26267867430552 ZZ78674 26267867430553 ZZ78674 26267867430554 ZZ78674 26267867430555 ZZ78674 26267867430323 ZZ78674 26267867430150 ZZ78674 26267867430545 ZZ78674 26267867430268 ZZ78674 26267867430269 ZZ78674 26267867430270 ZZ78674 26267867430319 ZZ78674 26267867430170 ZZ78674 26267867430171 ZZ78674 26267867430172 ZZ78674 26267867430173 ZZ78674 26267867430174 ZZ78674 26267009092 ZZ78674 26267867430262 ZZ78674 26267867430263 ZZ78674 26267867430264 ZZ78674 26267867430265 ZZ78674 26267867430266 ZZ78674 26267867430271 ZZ78674 26267867430272 ZZ78674 26267867430273 ZZ78674 26267867430274 ZZ78674 26267867430275 ZZ78674 26267867430281 ZZ78674 26267867430282 ZZ78674 26267867430283 ZZ78674 26267867430284 ZZ78674 26267867430285 ZZ78674 26267867430292 ZZ78674 26267867430293 ZZ78674 26267867430294 ZZ78674 26267867430295 ZZ78674 26267867430296 ZZ78674 26267867430297 ZZ78674 26267867430298 ZZ78674 26267867430299 ZZ78674 26267867430300 ZZ78674 26267867430301 ZZ78674 26267867430302 ZZ78674 26267867430308 ZZ78674 26267867430309 ZZ78674 26267867430310 ZZ78674 26267867430311 ZZ78674 262678671430664 ZZ78674 262678671430665 ZZ78674 26267008724 ZZ78674 26267009304 ZZ78674 262678671430648 ZZ78674 262678671430649 ZZ78674 262678671430650 ZZ78674 262678671430651 ZZ78674 262678671430652 ZZ78674 262678671430659 ZZ78674 26267867430128 ZZ78674 26267867430129 ZZ78674 26267867430130 ZZ78674 26267867430131 ZZ78674 26267867430132 ZZ78674 26267867430145 ZZ78674 26267867430146 ZZ78674 26267867430147 ZZ78674 26267867430148 ZZ78674 26267867430149 ZZ78674 26267867430155 ZZ78674 26267867430156 ZZ78674 26267867430157 ZZ78674 26267867430158 ZZ78674 26267867430159 ZZ78674 26267867430160 ZZ78674 26267867430161 ZZ78674 26267867430162 ZZ78674 26267867430163 ZZ78674 26267867430164 ZZ78674 26267867430165 ZZ78674 26267867430166 ZZ78674 26267867430167 ZZ78674 26267867430168 ZZ78674 26267867430169 ZZ78674 26267867430176 ZZ78674 26267867430177 ZZ78674 26267867430178 ZZ78674 26267867430179 ZZ78674 26267867430180 ZZ78674 26267867430181 ZZ78674 26267867430182 ZZ78674 26267867430183 ZZ78674 26267867430184 ZZ78674 26267867430185 ZZ78674 26267867430186 ZZ78674 26267867430187 ZZ78674 26267867430188 ZZ78674 26267867430189 ZZ78674 26267867430190 ZZ78674 26267008385 ZZ78674 26267008386 ZZ78674 26267008387 ZZ78674 26267008388 ZZ78674 26267008389 ZZ78674 26267008405 ZZ78674 26267008406 ZZ78674 26267008407 ZZ78674 26267008408 ZZ78674 26267008409 ZZ78674 26267008413 ZZ78674 26267008414 ZZ78674 26267867430154 ZZ78674 26267867430242 ZZ78674 2626903767 ZZ78674 26267867430407 ZZ78674 26267867430408 ZZ78674 26267867430409 ZZ78674 26267867430410 ZZ78674 262678671430710 ZZ78674 26267867430533 ZZ78674 26267867430534 ZZ78674 26267867430535 ZZ78674 26267867430536 ZZ78674 26267007974 ZZ78674 26267009302 ZZ78674 26267867430054 ZZ78674 262678671430593 ZZ78674 262678671430660 ZZ78674 262678671430687 ZZ78674 262678671430689 ZZ78674 262678671430690 ZZ78674 262678671430691 ZZ78674 26267867430256 ZZ78674 26267867430267 ZZ78674 26267867430329 ZZ78674 26267867430330 ZZ78674 26267867430331 ZZ78674 26267867430334 ZZ78674 26267867430335 ZZ78674 26267867430505 ZZ78674 26267867430506 ZZ78674 26267867430507 ZZ78674 26267867430524 ZZ78674 26267867430525 ZZ78674 26267867430526 ZZ78674 26267867430539 ZZ78674 26267867430540 ZZ78674 26267867430541 ZZ78674 26267867430542 ZZ78674 26267867430543 ZZ78674 26267867430529 ZZ78674 26267867430530 ZZ78674 26267867430531 ZZ78674 26267867430373 ZZ78674 26267867430374 ZZ78674 26267867430375 ZZ78674 26267867430376 ZZ78674 26267867430377 ZZ78674 26267867430378 ZZ78674 26267867430379 ZZ78674 26267867430380 ZZ78674 26267867430386 ZZ78674 26267867430391 ZZ78674 26267867430392 ZZ78674 26267867430393 ZZ78674 26267867430394 ZZ78674 26267867430395 ZZ78674 26267867430396 ZZ78674 26267867430336 ZZ78674 26267867430332 ZZ78674 26267867430333 ZZ78674 26267867430337 ZZ78674 26267867430338 ZZ78674 26267867430339 ZZ78674 26267867430340 ZZ78674 26267867430341 ZZ78674 26267867430351 ZZ78674 26267867430353 ZZ78674 26267867430354 ZZ78674 26267867430357 ZZ78674 26267867430358 ZZ78674 26267867430359 ZZ78674 262678671430771 ZZ78674 262678671430772 ZZ78674 262678671430773 ZZ78674 262678671430774 ZZ78674 262678671430775 ZZ78674 262678671430786 ZZ78674 262678671430787 ZZ78674 262678671430788 ZZ78674 262678671430789 ZZ78674 262678671430790 ZZ78674 26267867430320 ZZ78674 26267008731 ZZ78674 26267009197 ZZ78674 26267867430247 ZZ78674 26267008390 ZZ78674 26267008393 ZZ78674 26267008394 ZZ78674 26267008395 ZZ78674 26267008396 ZZ78674 26267008397 ZZ78674 26267008398 ZZ78674 26267008399 ZZ78674 26267008400 ZZ78674 26267008401 ZZ78674 26267008726 ZZ78674 26267008727 ZZ78674 26267867430556 ZZ78674 26267867430557 ZZ78674 262678671430596 ZZ78674 262678671430597 ZZ78674 26267867430527 ZZ78674 26267867430528 ZZ78674 26267867430537 ZZ78674 26267867430538 ZZ78674 26267008732 ZZ78674 26267867430322 ZZ78674 26267867430347 ZZ78674 26267867430348 ZZ78674 26267867430349 ZZ78674 26267867430350 ZZ78674 26267867430352 ZZ78674 26267867430369 ZZ78674 26267867430370 ZZ78674 26267867430371 ZZ78674 26267867430372 ZZ78674 26267007964 ZZ78674 26267007965 ZZ78674 26267007966 ZZ78674 26267007967 ZZ78674 26267007968 ZZ78674 26267007969 ZZ78674 26267007970 ZZ78674 26267007971 ZZ78674 26267007972 ZZ78674 26267007973 ZZ78674 26267009298 ZZ78674 26267009299 ZZ78674 26267867430011 ZZ78674 26267867430012 ZZ78674 26267867430013 ZZ78674 26267867430014 ZZ78674 26267867430015 ZZ78674 26267867430016 ZZ78674 26267867430017 ZZ78674 26267867430020 ZZ78674 26267867430021 ZZ78674 26267867430022 ZZ78674 26267867430023 ZZ78674 26267867430024 ZZ78674 26267867430025 ZZ78674 26267867430026 ZZ78674 26267867430027 ZZ78674 26267867430028 ZZ78674 26267867430029 ZZ78674 26267867430030 ZZ78674 26267867430031 ZZ78674 26267867430032 ZZ78674 26267867430033 ZZ78674 26267867430034 ZZ78674 26267867430035 ZZ78674 26267867430036 ZZ78674 26267867430037 ZZ78674 26267867430038 ZZ78674 26267867430039 ZZ78674 26267867430040 ZZ78674 26267867430041 ZZ78674 26267867430042 ZZ78674 26267867430043 ZZ78674 26267867430044 ZZ78674 26267867430045 ZZ78674 26267867430046 ZZ78674 26267867430047 ZZ78674 26267867430048 ZZ78674 26267867430049 ZZ78674 26267867430052 ZZ78674 26267867430053 ZZ78674 26267867430055 ZZ78674 26267867430056 ZZ78674 26267867430057 ZZ78674 26267867430058 ZZ78674 26267867430059 ZZ78674 26267867430065 ZZ78674 26267867430066 ZZ78674 26267867430067 ZZ78674 26267867430068 ZZ78674 26267867430069 ZZ78674 26267867430070 ZZ78674 26267867430071 ZZ78674 26267867430072 ZZ78674 26267867430073 ZZ78674 26267867430074 ZZ78674 26267867430075 ZZ78674 26267867430076 ZZ78674 26267867430077 ZZ78674 26267867430078 ZZ78674 26267867430079 ZZ78674 26267867430080 ZZ78674 26267867430081 ZZ78674 26267867430082 ZZ78674 26267867430083 ZZ78674 26267867430084 ZZ78674 26267867430085 ZZ78674 26267867430086 ZZ78674 26267867430087 ZZ78674 26267867430088 ZZ78674 26267867430089 ZZ78674 26267867430090 ZZ78674 26267867430091 ZZ78674 26267867430092 ZZ78674 26267867430093 ZZ78674 26267867430094 ZZ78674 26267867430095 ZZ78674 26267867430096 ZZ78674 26267867430097 ZZ78674 26267867430098 ZZ78674 26267867430099 ZZ78674 26267867430100 ZZ78674 26267867430101 ZZ78674 26267867430102 ZZ78674 26267867430103 ZZ78674 26267867430104 ZZ78674 26267867430105 ZZ78674 26267867430106 ZZ78674 26267867430107 ZZ78674 26267867430108 ZZ78674 26267867430109 ZZ78674 26267867430110 ZZ78674 26267867430111 ZZ78674 26267867430112 ZZ78674 26267867430113 ZZ78674 26267867430114 ZZ78674 26267867430115 ZZ78674 26267867430116 ZZ78674 26267867430117 ZZ78674 26267867430118 ZZ78674 26267867430119 ZZ78674 26267867430120 ZZ78674 26267867430121 ZZ78674 26267867430122 ZZ78674 26267867430123 ZZ78674 26267867430124 ZZ78674 26267867430125 ZZ78674 26267867430126 ZZ78674 26267007975 ZZ78674 26267008392 ZZ78674 26267009091 ZZ78674 26267867430532 ZZ78674 262678671430594 ZZ98674 2626903412 ZZ98674 2626903690 ZZ98674 2626903691 ZZ98674 2626903692 ZZ98674 2626904056 ZZ98674 2626904057 ZZ98674 2626904058 ZZ98674 262698674130010 ZZ98674 262698674130011 ZZ98674 262698674130012 ZZ98674 262698674130072 ZZ98674 262698674130073 ZZ98674 262698674130074 ZZ98674 262698674130096 ZZ98674 262698674130097 ZZ98674 262698674130098 ZZ98674 262698674130099 ZZ98674 262698674130100 ZZ98674 262698674130278 ZZ98674 262698674130279 ZZ98674 2626903235 ZZ98674 262698674130619 ZZ98674 262698674130607 ZZ98674 2626904191 ZZ98674 2626903459 ZZ98674 2626903460 ZZ98674 2626903461 ZZ98674 2626903462 ZZ98674 2626903463 ZZ98674 2626903464 ZZ98674 2626903465 ZZ98674 262698674130611 ZZ98674 262698674140726 ZZ98674 262698674140727 ZZ98674 262698674140790 ZZ98674 262698674140791 ZZ98674 262698674140792 ZZ98674 262698674130567 ZZ98674 262698674130568 ZZ98674 262698674130569 ZZ98674 262698674140666 ZZ98674 262698674140667 ZZ98674 262698674140668 ZZ98674 2626904204 ZZ98674 2626904205 ZZ98674 2626904206 ZZ98674 26260902875 ZZ98674 2626903505 ZZ98674 262698674130612 ZZ98674 262698674130613 ZZ98674 2626904060 ZZ98674 262698674130017 ZZ98674 262698674130018 ZZ98674 262698674130019 ZZ98674 262698674130020 ZZ98674 262698674130021 ZZ98674 262698674130025 ZZ98674 262698674130026 ZZ98674 262698674130027 ZZ98674 262698674130028 ZZ98674 262698674130029 ZZ98674 262698674130030 ZZ98674 262698674130031 ZZ98674 262698674130032 ZZ98674 262698674130033 ZZ98674 262698674130034 ZZ98674 262698674130035 ZZ98674 262698674130036 ZZ98674 262698674130037 ZZ98674 262698674130038 ZZ98674 262698674130039 ZZ98674 262698674130040 ZZ98674 262698674130041 ZZ98674 262698674130042 ZZ98674 262698674130043 ZZ98674 262698674130044 ZZ98674 262698674130048 ZZ98674 262698674130049 ZZ98674 262698674130050 ZZ98674 262698674130051 ZZ98674 262698674130053 ZZ98674 262698674130054 ZZ98674 262698674130055 ZZ98674 262698674130109 ZZ98674 262698674130024 ZZ98674 262698674130045 ZZ98674 262698674130500 ZZ98674 262698674130501 ZZ98674 262698674130502 ZZ98674 262698674130503 ZZ98674 262698674130504 ZZ98674 262698674130505 ZZ98674 262698674130506 ZZ98674 262698674130507 ZZ98674 262698674130508 ZZ98674 262698674130514 ZZ98674 262698674130515 ZZ98674 262698674130516 ZZ98674 262698674130517 ZZ98674 262698674130518 ZZ98674 262698674130519 ZZ98674 262698674130520 ZZ98674 262698674130521 ZZ98674 262698674130522 ZZ98674 262698674130536 ZZ98674 262698674130537 ZZ98674 262698674130538 ZZ98674 262698674130539 ZZ98674 262698674130540 ZZ98674 262698674130541 ZZ98674 262698674130542 ZZ98674 262698674130543 ZZ98674 262698674130544 ZZ98674 262698674130545 ZZ98674 262698674130546 ZZ98674 262698674130547 ZZ98674 262698674130609 ZZ98674 262698674140640 ZZ98674 262698674140641 ZZ98674 262698674140642 ZZ98674 262698674140643 ZZ98674 262698674140644 ZZ98674 262698674140645 ZZ98674 262698674140646 ZZ98674 262698674140647 ZZ98674 262698674140648 ZZ98674 262698674140660 ZZ98674 262698674140661 ZZ98674 262698674140662 ZZ98674 262698674140663 ZZ98674 262698674140664 ZZ98674 262698674140704 ZZ98674 262698674140705 ZZ98674 262698674140706 ZZ98674 262698674140707 ZZ98674 262698674140708 ZZ98674 262698674130574 ZZ98674 262698674130606 ZZ98674 262698674140764 ZZ98674 2626904187 ZZ98674 2626904188 ZZ98674 262698674130553 ZZ98674 262698674130554 ZZ98674 262698674130555 ZZ98674 2626903279 ZZ98674 262698674130509 ZZ98674 262698674130510 ZZ98674 262698674130511 ZZ98674 262698674130512 ZZ98674 262698674130513 ZZ98674 262698674130525 ZZ98674 262698674130526 ZZ98674 262698674130527 ZZ98674 262698674130528 ZZ98674 262698674130529 ZZ98674 262698674130530 ZZ98674 262698674130531 ZZ98674 262698674130532 ZZ98674 262698674130533 ZZ98674 262698674130534 ZZ98674 262698674130548 ZZ98674 262698674130549 ZZ98674 262698674130556 ZZ98674 262698674140665 ZZ98674 262698674140670 ZZ98674 262698674140671 ZZ98674 262698674140673 ZZ98674 262698674140674 ZZ98674 262698674140675 ZZ98674 262698674140676 ZZ98674 262698674140677 ZZ98674 262698674140683 ZZ98674 262698674140684 ZZ98674 262698674140685 ZZ98674 262698674140686 ZZ98674 262698674140687 ZZ98674 262698674140688 ZZ98674 262698674140689 ZZ98674 262698674140690 ZZ98674 262698674140691 ZZ98674 262698674140692 ZZ98674 262698674140693 ZZ98674 262698674140694 ZZ98674 262698674140695 ZZ98674 262698674140696 ZZ98674 262698674140697 ZZ98674 262698674140698 ZZ98674 262698674140672 ZZ98674 262698674140759 ZZ98674 2626903468 ZZ98674 2626903469 ZZ98674 2626903561 ZZ98674 2626903763 ZZ98674 2626903901 ZZ98674 2626904128 ZZ98674 262698674130022 ZZ98674 262698674130023 ZZ98674 262698674130071 ZZ98674 262698674130105 ZZ98674 262698674130106 ZZ98674 262698674140714 ZZ98674 262698674140715 ZZ98674 262698674130257 ZZ98674 262698674140655 ZZ98674 262698674140656 ZZ98674 2626903280 ZZ98674 262698674130081 ZZ98674 262698674130082 ZZ98674 262698674130083 ZZ98674 262698674130084 ZZ98674 262698674130085 ZZ98674 262698674130086 ZZ98674 262698674130087 ZZ98674 262698674130088 ZZ98674 262698674130089 ZZ98674 262698674130090 ZZ98674 262698674130091 ZZ98674 262698674130092 ZZ98674 262698674130093 ZZ98674 262698674130094 ZZ98674 262698674130095 ZZ98674 262698674130103 ZZ98674 262698674130137 ZZ98674 262698674130155 ZZ98674 262698674130212 ZZ98674 2626904129 ZZ98674 262698674130101 ZZ98674 262698674140718 ZZ98674 262698674140719 ZZ98674 262698674140720 ZZ98674 262698674140750 ZZ98674 262698674140751 ZZ98674 262698674140774 ZZ98674 262698674140775 ZZ98674 262698674140776 ZZ98674 262698674140777 ZZ98674 262698674140778 ZZ98674 2626904094 ZZ98674 262698674130620 ZZ98674 262698674130621 ZZ98674 262698674130622 ZZ98674 262698674130623 ZZ98674 262698674130624 ZZ98674 262698674130593 ZZ98674 262698674130594 ZZ98674 262698674130595 ZZ98674 262698674130596 ZZ98674 262698674140649 ZZ98674 262698674130256 ZZ98674 262698674130610 ZZ98674 2626903282 ZZ98674 262698674140709 ZZ98674 262698674130173 ZZ98674 262698674130174 ZZ98674 2626903411 ZZ98674 2626903236 ZZ98674 2626903237 ZZ98674 2626903673 ZZ98674 2626903674 ZZ98674 2626903675 ZZ98674 2626903677 ZZ98674 2626903678 ZZ98674 2626903070 ZZ98674 2626904135 ZZ98674 2626904136 ZZ98674 2626904137 ZZ98674 1026904126 ZZ98674 2626904125 ZZ98674 2626904127 ZZ98674 2626904138 ZZ98674 2626904139 ZZ98674 2626904185 ZZ98674 2626904186 ZZ98674 2626904192 ZZ98674 2626904193 ZZ98674 2626904194 ZZ98674 2626904197 ZZ98674 2626904198 ZZ98674 2626904199 ZZ98674 2626904200 ZZ98674 2626904201 ZZ98674 2626904202 ZZ98674 2626904203 ZZ98674 2626904207 ZZ98674 2626904208 ZZ98674 2626904209 ZZ98674 262698674130003 ZZ98674 262698674130004 ZZ98674 262698674130005 ZZ98674 262698674130006 ZZ98674 262698674130007 ZZ98674 262698674130008 ZZ98674 262698674130264 ZZ98674 262698674130265 ZZ98674 262698674130266 ZZ98674 262698674130267 ZZ98674 262698674130268 ZZ98674 262698674130272 ZZ98674 262698674130273 ZZ98674 262698674130274 ZZ98674 262698674130275 ZZ98674 262698674130276 ZZ98674 262698674140752 ZZ98674 262698674140753 ZZ98674 262698674140754 ZZ98674 262698674140755 ZZ98674 262698674140756 ZZ98674 262698674130214 ZZ98674 262698674130215 ZZ98674 262698674130216 ZZ98674 262698674140784 ZZ98674 262698674140785 ZZ98674 262698674140669 ZZ98674 262698674130222 ZZ98674 262698674130223 ZZ98674 262698674130224 ZZ98674 262698674130225 ZZ98674 262698674130226 ZZ98674 262698674130227 ZZ98674 262698674130228 ZZ98674 262698674130229 ZZ98674 262698674130230 ZZ98674 262698674130231 ZZ98674 262698674130233 ZZ98674 262698674130234 ZZ98674 262698674130235 ZZ98674 262698674130236 ZZ98674 262698674130237 ZZ98674 262698674130238 ZZ98674 262698674130239 ZZ98674 262698674130240 ZZ98674 262698674130241 ZZ98674 262698674130242 ZZ98674 262698674130141 ZZ98674 262698674130145 ZZ98674 262698674130248 ZZ98674 262698674130559 ZZ98674 262698674130560 ZZ98674 262698674130578 ZZ98674 262698674130580 ZZ98674 262698674130581 ZZ98674 262698674130582 ZZ98674 262698674130584 ZZ98674 262698674130134 ZZ98674 262698674130110 ZZ98674 262698674130111 ZZ98674 262698674130112 ZZ98674 262698674130113 ZZ98674 262698674130114 ZZ98674 262698674130277 ZZ98674 262698674130146 ZZ98674 262698674140745 ZZ98674 262698674140725 ZZ98674 262698674130131 ZZ98674 262698674130132 ZZ98674 262698674130133 ZZ98674 262698674130153 ZZ98674 262698674130154 ZZ98674 262698674140739 ZZ98674 262698674130570 ZZ98674 262698674130571 ZZ98674 262698674130572 ZZ98674 262698674130575 ZZ98674 262698674130587 ZZ98674 262698674130588 ZZ98674 262698674130589 ZZ98674 262698674130590 ZZ98674 262698674130603 ZZ98674 262698674130628 ZZ98674 262698674130550 ZZ98674 262698674130551 ZZ98674 262698674130552 ZZ98674 262698674140738 ZZ98674 262698674140741 ZZ98674 262698674140736 ZZ98674 262698674140737 ZZ98674 262698674140713 ZZ98674 262698674140740 ZZ98674 262698674130116 ZZ98674 262698674130592 ZZ98674 262698674130602 ZZ98674 262698674140728 ZZ98674 262698674140762 ZZ98674 262698674130220 ZZ98674 262698674130221 ZZ98674 262698674130535 ZZ98674 262698674130142 ZZ98674 262698674130159 ZZ98674 262698674130243 ZZ98674 262698674130244 ZZ98674 262698674130172 ZZ98674 262698674130204 ZZ98674 262698674130205 ZZ98674 262698674140788 ZZ98674 262698674140789 ZZ98674 262698674140793 ZZ98674 262698674130104 ZZ98674 262698674140699 ZZ98674 262698674130143 ZZ98674 262698674140730 ZZ98674 2626903562 ZZ98674 2626903694 ZZ98674 262698674130047 ZZ98674 262698674130232 ZZ98674 262698674130608 ZZ98674 2626902821 ZZ98674 2626903740 ZZ98674 2626903741 ZZ98674 2626903742 ZZ98674 2626903743 ZZ98674 2626903759 ZZ98674 2626903760 ZZ98674 2626903761 ZZ98674 2626903762 ZZ98674 2626903764 ZZ98674 2626903765 ZZ98674 2626903841 ZZ98674 2626903842 ZZ98674 2626903843 ZZ98674 2626903844 ZZ98674 2626903845 ZZ98674 2626903846 ZZ98674 2626903982 ZZ98674 2626903983 ZZ98674 2626903984 ZZ98674 2626903985 ZZ98674 2626903986 ZZ98674 2626903987 ZZ98674 2626903988 ZZ98674 2626903989 ZZ98674 2626903990 ZZ98674 2626903991 ZZ98674 2626903992 ZZ98674 2626903993 ZZ98674 2626903994 ZZ98674 2626903995 ZZ98674 2626903996 ZZ98674 2626903997 ZZ98674 2626903998 ZZ98674 2626903999 ZZ98674 2626904000 ZZ98674 2626904001 ZZ98674 2626904002 ZZ98674 2626904003 ZZ98674 2626904004 ZZ98674 2626904005 ZZ98674 2626904006 ZZ98674 2626904007 ZZ98674 2626904008 ZZ98674 2626904009 ZZ98674 2626904010 ZZ98674 2626904011 ZZ98674 2626904012 ZZ98674 2626904013 ZZ98674 2626904014 ZZ98674 2626904015 ZZ98674 2626904016 ZZ98674 2626904017 ZZ98674 2626904018 ZZ98674 2626904019 ZZ98674 2626904020 ZZ98674 2626904021 ZZ98674 2626904022 ZZ98674 2626904023 ZZ98674 2626904024 ZZ98674 2626904025 ZZ98674 2626904026 ZZ98674 2626904027 ZZ98674 2626904028 ZZ98674 2626904029 ZZ98674 2626904030 ZZ98674 2626904031 ZZ98674 2626904032 ZZ98674 2626904033 ZZ98674 2626904034 ZZ98674 2626904035 ZZ98674 2626904036 ZZ98674 2626904037 ZZ98674 2626904038 ZZ98674 2626904039 ZZ98674 2626904040 ZZ98674 2626904041 ZZ98674 2626904042 ZZ98674 2626904043 ZZ98674 2626904044 ZZ98674 2626904045 ZZ98674 2626904046 ZZ98674 2626904047 ZZ98674 2626904048 ZZ98674 2626904049 ZZ98674 2626904050 ZZ98674 2626904051 ZZ98674 2626904052 ZZ98674 2626904053 ZZ98674 2626904054 ZZ98674 2626904055 ZZ98674 262698674130060 ZZ98674 262698674130151 ZZ98674 262698674130152 ZZ98674 262698674130148 ZZ98674 262698674130149 ZZ98674 2626904059 ZZ98674 262698674140724 ZZ98674 262698674140770 ZZ98674 2626903615 ZZ98674 2626903070 ZZ98674 262698674130577 ZZ98674 262698674140657 ZZ98674 262698674140658 ZZ98674 262698674140659 ZZ98674 262698674140742 ZZ98674 262698674140743 ZZ98674 262698674140744 ZZ98674 262698674130565 ZZ98674 262698674130566 ZZ98674 262698674130579 ZZ98674 262698674140758 ZZ98674 262698674130561 ZZ98674 0 ZZ98674 2626903070 ZZ98674 262698674130102 ZZ98674 262698674130523 ZZ98674 262698674130524 ZZ98674 262698674130258 ZZ98674 2626904130 ZZ98674 2626904131 ZZ98674 2626904132 ZZ98674 2626904133 ZZ98674 2626904134 ZZ98674 262698674130626 ZZ98674 2626902953 ZZ98674 262698674130119 ZZ98674 262698674130219 ZZ98674 2626903672 ZZ98674 262698674130250 ZZ98674 2626903744 ZZ98674 262698674130076 ZZ98674 262698674130077 ZZ98674 262698674130078 ZZ98674 262698674130079 ZZ98674 262698674130080 ZZ98674 262698674130120 ZZ98674 262698674130121 ZZ98674 262698674130122 ZZ98674 262698674130123 ZZ98674 262698674130124 ZZ98674 2626903685 ZZ98674 2626903686 ZZ98674 2626903687 ZZ98674 2626903688 ZZ98674 2626903689 ZZ98674 2626903695 ZZ98674 2626903696 ZZ98674 2626903697 ZZ98674 2626903698 ZZ98674 2626903699 ZZ98674 2626903700 ZZ98674 2626903701 ZZ98674 2626903732 ZZ98674 2626903733 ZZ98674 2626903734 ZZ98674 2626903735 ZZ98674 2626903736 ZZ98674 2626903737 ZZ98674 2626903738 ZZ98674 2626903739 ZZ98674 262698674130138 ZZ98674 262698674130139 ZZ98674 262698674130140 ZZ98674 262698674130156 ZZ98674 262698674130158 ZZ98674 262698674130162 ZZ98674 262698674130163 ZZ98674 262698674130164 ZZ98674 262698674130165 ZZ98674 262698674130166 ZZ98674 262698674130167 ZZ98674 262698674130168 ZZ98674 262698674130169 ZZ98674 262698674130170 ZZ98674 262698674130171 ZZ98674 262698674130177 ZZ98674 262698674130178 ZZ98674 262698674130179 ZZ98674 262698674130180 ZZ98674 262698674130181 ZZ98674 262698674130184 ZZ98674 262698674130185 ZZ98674 262698674130186 ZZ98674 262698674130187 ZZ98674 262698674130188 ZZ98674 262698674130189 ZZ98674 262698674130190 ZZ98674 262698674130191 ZZ98674 262698674130192 ZZ98674 262698674130193 ZZ98674 262698674130194 ZZ98674 262698674130195 ZZ98674 262698674130196 ZZ98674 262698674130197 ZZ98674 262698674130198 ZZ98674 262698674130199 ZZ98674 262698674130200 ZZ98674 262698674130201 ZZ98674 262698674130202 ZZ98674 262698674130203 ZZ98674 262698674130206 ZZ98674 262698674130150 ZZ98674 2626904061 ZZ98674 2626904092 ZZ98674 262698674130117 ZZ98674 262698674130118 ZZ98674 262698674130147 ZZ98674 262698674130213 ZZ98674 262698674130591 ZZ98674 262698674130597 ZZ98674 262698674130604 ZZ98674 262698674130605 ZZ98674 262698674130207 ZZ98674 262698674130208 ZZ98674 262698674130209 ZZ98674 262698674130210 ZZ98674 262698674130211 ZZ98674 262698674130175 ZZ98674 262698674130176 ZZ98674 262698674130599 ZZ98674 262698674130600 ZZ98674 262698674140729 ZZ98674 262698674130627 ZZ98674 262698674130052 ZZ98674 262698674130056 ZZ98674 262698674130057 ZZ98674 262698674130058 ZZ98674 262698674130059 ZZ98674 262698674140765 ZZ98674 262698174130765 ZZ98674 262698674130013 ZZ98674 2626903766 ZZ98674 262698674140760 ZZ98674 262698674140761 ZZ98674 262698674140766 ZZ98674 2626903839 ZZ98674 2626903840 ZZ98674 2626903847 ZZ98674 2626903848 ZZ98674 2626903849 ZZ98674 2626903850 ZZ98674 2626903851 ZZ98674 2626903852 ZZ98674 2626903853 ZZ98674 2626903854 ZZ98674 2626903238 ZZ98674 262698674130246 ZZ98674 262698674130247 ZZ98674 262698674130269 ZZ98674 262698674130270 ZZ98674 262698674130271 ZZ98674 262698674130014 ZZ98674 262698674130075 ZZ98674 262698674130601 ZZ98674 262698674130115 ZZ98674 2626904093 ZZ98674 2626903693 ZZ98674 2626903745 ZZ98674 2626903746 ZZ98674 2626903747 ZZ98674 2626903748 ZZ98674 2626903749 ZZ98674 2626903750 ZZ98674 2626903751 ZZ98674 2626903756 ZZ98674 2626903757 ZZ98674 2626903758 ZZ98674 262698674140763 ZZ98674 262698674130557 ZZ98674 262698674130558 ZZ98674 262698674130562 ZZ98674 262698674130563 ZZ98674 262698674130564 ZZ98674 262698674130585 ZZ98674 262698674130586 ZZ98674 262698674130107 ZZ98674 262698674130108 ZZ98674 2626903679 ZZ98674 262698674140716 ZZ98674 262698674140717 ZZ98674 262698674140721 ZZ98674 262698674140722 ZZ98674 262698674140723 ZZ98674 262698674140771 ZZ98674 262698674140772 ZZ98674 262698674140773 ZZ98674 262698674140779 ZZ98674 262698674140780 ZZ98674 262698674140781 ZZ98674 262698674140782 ZZ98674 2626903900 ZZ98674 2626903838 ZZ98674 2626904189 ZZ98674 2626904190 ZZ98674 262698674130598 ZZ98674 262698674130617 ZZ98674 262698674130618 ZZ98674 262698674140746 ZZ98674 262698674140757 ZZ98674 262698674130061 ZZ98674 262698674130062 ZZ98674 262698674130063 ZZ98674 262698674130064 ZZ98674 262698674130065 ZZ98674 262698674130066 ZZ98674 262698674130067 ZZ98674 262698674130068 ZZ98674 262698674130069 ZZ98674 262698674130070 ZZ98674 262698674140749 ZZ98674 262698674140783 ZZ98674 262698674130009 ZZ98674 262698674130015 ZZ98674 262698674140710 ZZ98674 262698674140711 ZZ98674 262698674140712 ZZ98674 262698674140678 ZZ98674 262698674140679 ZZ98674 262698674140680 ZZ98674 262698674140681 ZZ98674 262698674140682 ZZ98674 262698674140700 ZZ98674 262698674140701 ZZ98674 262698674140702 ZZ98674 262698674140703 ZZ98674 262698674130217 ZZ98674 262698674130218 ZZ98674 262698674130245 ZZ98674 262698674130614 ZZ98674 2626903466 ZZ98674 2626903467 ZZ98674 2626903152 ZZ98674 2626903680 ZZ98674 2626903681 ZZ98674 2626903682 ZZ98674 2626903683 ZZ98674 2626903684 ZZ98674 262698674140731 ZZ98674 262698674140732 ZZ98674 262698674140733 ZZ98674 262698674140734 ZZ98674 262698674140735 ZZ98674 262698674130016 ZZ98674 2626903616 ZZ98674 2626903504 ZZ98674 262698674130576 ZZ98674 262698674140768 ZZ98674 262698674140786 ZZ98674 262698674130046 ZZ98674 262698674130249 ZZ98674 262698674130255 ZZ98674 2626903323 ZZ98674 2626903324 ZZ98674 2626903325 ZZ98674 2626903326 ZZ98674 2626903327 ZZ98674 2626903328 ZZ98674 2626903329 ZZ98674 2626903330 ZZ98674 2626903331 ZZ98674 2626903332 ZZ98674 2626903333 ZZ98674 2626903334 ZZ98674 2626903335 ZZ98674 2626903336 ZZ98674 2626903337 ZZ98674 2626903338 ZZ98674 2626903339 ZZ98674 2626903340 ZZ98674 2626903341 ZZ98674 2626903342 ZZ98674 2626903343 ZZ98674 2626903344 ZZ98674 2626903345 ZZ98674 2626903346 ZZ98674 2626903347 ZZ98674 2626903348 ZZ98674 2626903349 ZZ98674 2626903350 ZZ98674 2626903351 ZZ98674 2626903352 ZZ98674 2626903353 ZZ98674 2626903354 ZZ98674 2626903355 ZZ98674 2626903356 ZZ98674 2626903357 ZZ98674 2626903358 ZZ98674 2626903359 ZZ98674 2626903360 ZZ98674 2626903361 ZZ98674 2626903362 ZZ98674 2626903363 ZZ98674 2626903364 ZZ98674 2626903365 ZZ98674 2626903366 ZZ98674 2626903367 ZZ98674 2626903368 ZZ98674 2626903369 ZZ98674 2626903370 ZZ98674 2626903371 ZZ98674 2626903372 ZZ98674 2626903373 ZZ98674 2626903374 ZZ98674 2626903375 ZZ98674 2626903376 ZZ98674 2626903377 ZZ98674 2626903378 ZZ98674 2626903379 ZZ98674 2626903380 ZZ98674 2626903381 ZZ98674 2626903382 ZZ98674 2626903383 ZZ98674 2626903384 ZZ98674 2626903385 ZZ98674 2626903386 ZZ98674 2626903387 ZZ98674 2626903388 ZZ98674 2626903389 ZZ98674 2626903390 ZZ98674 2626903391 ZZ98674 2626903392 ZZ98674 2626903393 ZZ98674 2626903394 ZZ98674 2626903395 ZZ98674 2626903396 ZZ98674 2626903397 ZZ98674 2626903398 ZZ98674 2626903399 ZZ98674 2626903400 ZZ98674 2626903401 ZZ98674 2626903402 ZZ98674 2626903403 ZZ98674 2626903404 ZZ98674 2626903405 ZZ98674 2626903406 ZZ98674 2626903407 ZZ98674 2626903408 ZZ98674 2626903409 ZZ98674 2626903410 ZZ98674 262698674140650 ZZ98674 262698674140651 ZZ98674 262698674140652 ZZ98674 262698674140653 ZZ98674 262698674140654 ZZ98674 262698674130625 ZZ98674 2626903281 ZZ98774 26260908070704143982 ZZ98774 2626903860 ZZ98774 2626903861 ZZ98774 2626903862 ZZ98774 2626903863 ZZ98774 2626903864 ZZ98774 2626903865 ZZ98774 2626903866 ZZ98774 2626903867 ZZ98774 2626903868 ZZ98774 2626903869 ZZ98774 2626903870 ZZ98774 2626903871 ZZ98774 2626903872 ZZ98774 2626903873 ZZ98774 2626903874 ZZ98774 2626903875 ZZ98774 2626903876 ZZ98774 2626903981 ZZ98774 2626903878 ZZ98774 2626903879 ZZ98774 2626903880 ZZ98774 2626903881 ZZ98774 2626903882 ZZ98774 2626903883 ZZ98774 2626903884 ZZ98774 2626903885 ZZ98774 2626903886 ZZ98774 2626903887 ZZ98774 2626903888 ZZ98774 2626903889 ZZ98774 2626903890 ZZ98774 2626903891 ZZ98774 2626903892 ZZ98774 2626903893 ZZ98774 2626903894 ZZ98774 2626903895 ZZ98774 2626903896 ZZ98774 2626903897 ZZ98774 2626903898 ZZ98774 2626903899 ZZ98774 2626903902 ZZ98774 2626903903 ZZ98774 2626903904 ZZ98774 2626903905 ZZ98774 2626903906 ZZ98774 2626903907 ZZ98774 2626903908 ZZ98774 2626903909 ZZ98774 2626903910 ZZ98774 2626903911 ZZ98774 2626903912 ZZ98774 2626903913 ZZ98774 2626903914 ZZ98774 2626903915 ZZ98774 2626903916 ZZ98774 2626903917 ZZ98774 2626903918 ZZ98774 2626903919 ZZ98774 2626903920 ZZ98774 2626903921 ZZ98774 2626903922 ZZ98774 2626903923 ZZ98774 2626903924 ZZ98774 2626903925 ZZ98774 2626903926 ZZ98774 2626903927 ZZ98774 2626903928 ZZ98774 2626903929 ZZ98774 2626903930 ZZ98774 2626903931 ZZ98774 2626903932 ZZ98774 2626903933 ZZ98774 2626903934 ZZ98774 2626903935 ZZ98774 2626903936 ZZ98774 2626903937 ZZ98774 2626903938 ZZ98774 2626903939 ZZ98774 2626903940 ZZ98774 2626903941 ZZ98774 2626903942 ZZ98774 2626903943 ZZ98774 2626903944 ZZ98774 2626903945 ZZ98774 2626903946 ZZ98774 2626903947 ZZ98774 2626903948 ZZ98774 2626903949 ZZ98774 2626903950 ZZ98774 2626903951 ZZ98774 2626903952 ZZ98774 2626903953 ZZ98774 2626903954 ZZ98774 2626903955 ZZ98774 2626903956 ZZ98774 2626903957 ZZ98774 2626903958 ZZ98774 2626903959 ZZ98774 2626903960 ZZ98774 2626903961 ZZ98774 2626903962 ZZ98774 2626903963 ZZ98774 2626903964 ZZ98774 2626903965 ZZ98774 2626903966 ZZ98774 2626903967 ZZ98774 2626903968 ZZ98774 2626903969 ZZ98774 2626903970 ZZ98774 2626903971 ZZ98774 2626903972 ZZ98774 2626903973 ZZ98774 2626903974 ZZ98774 2626903975 ZZ98774 2626903976 ZZ98774 2626903977 ZZ98774 2626903978 ZZ98774 2626903979 ZZ98774 2626903980 ZZ98774 2626903981 |
Recalling Firm/ Manufacturer |
GF Health Products, Inc./d.b.a.Basic American Metal Products 336 Trowbridge Dr Fond Du Lac WI 54937-9103
|
For Additional Information Contact |
Field Correction Coordinator 888-974-4347
|
Manufacturer Reason for Recall |
The bed head deck may fail to elevate as expected during normal operation. When the non-operation occurs, the head deck remains in the lowered horizontal / flat position and will no longer function as intended.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Consignees were sent a Basic American Medical Products "Product Field Correction Notice" dated July 21, 2014.
The letter provided the background information on the recall and the " Required Action" to be taken which includes the replacement of the head deck. Consignees need to contact 1-888-974-4347 to scheduled the replacement order. The replaced head decks should be disposed and disposal documented on the enclosed Letter of Disposal Agreement. The completed letter needs to be returned to Basic American Medical Products. For additional questions they can contact the Field Correction Coordinator at 1-888-974-4347 or electronically: fieldcorrection@grahamfield.com. |
Quantity in Commerce |
1941 |
Distribution |
Worldwide Distribution: USA (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KY, LA, ME, MD, MA,MI, MN, MS, MO,MT, NE, NH, NJ, NM, NY,M NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, LT, VA, WA, WV, and WI and PUERTO RICO and the country of CANADA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|