  Class 2 Device Recall Gemini system see related information
Date Initiated by Firm May 01, 2008
Date Posted November 02, 2015
Recall Status1 Terminated 3 on November 06, 2015
Recall Number Z-0197-2016
Recall Event ID 69082
510(K)Number K032036  K051170  K052640  
Product Classification System, x-ray, tomography, computed - Product Code JAK
Product GEMINI PET/CT System, a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Code Information Model Name Model NUmber and serial number GEMINI Dual 4535 679 00891 G097, G119, 101-103, 105-107, 109-117, 119-127, 131-144, 146-164, 1025, 2003506, 2003507, 1Z9Z591, 2Z9Z591, 3Z9Z591, 4Z9Z591, GY9Z591, HY9Z591, JY9Z591 GEMINI 16 Power 4535 679 58961 1002, 1004-1008, 1010-1015, 1017-1028, 1030-1034, 1035R, 1036A, 1037-1046, 101R, 1023a, G098, MNM4411 GEMINI GXL (6, 10, 16, and 16 Mobile) 4535 679 75691, 4535 679 81871, 4535 679 76581, 4535 679 71891 4001, 4003-4040, 4042-4089, 4091-4123, 4125-4127, 4129-4137, 4139-4143, 7011, 4097A, 4147a GEMINI TF 16 Slice (16, 64, 64 Mobile and 16 Base) 4535 679 83931, 4535 674 47551, 4535 679 94741, 4535 674 41711 7001, 7003-7006, 7008, 7009, 7013-7017, 7022, 7024, 7025, 7027, 7028 
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information Contact
440-483-7600
Manufacturer Reason
for Recall
The reason for this inspection is to determine if the Spindle Block Servo Controllers on the CT subsystem of potentially affected GEMINI PET/CT systems are susceptible to premature failure.
FDA Determined
Cause 2
Unknown/Undetermined by firm
Action Philips field service personnel will conduct customer visits in order to inspect systems. For further questions please call (440) 483-7600.
Quantity in Commerce 264 units
Distribution Worldwide Distribution - US Distribution to the states of : AL, AZ, CA, CO, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MO, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI, WV, and WY., and to the countries of : Canada, Argentina, Australia, Austria, Belgium, Brazil, China, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Luxembourg, Mexico, Netherlands, Poland, Portugal, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and UK.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JAK and Original Applicant = PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
-
-