• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Spacelabs Medical Patient Monitor equipped with the Perioperative Option

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Spacelabs Medical Patient Monitor equipped with the Perioperative Option see related information
Date Initiated by Firm August 26, 2014
Date Posted September 25, 2014
Recall Status1 Terminated 3 on June 26, 2015
Recall Number Z-2707-2014
Recall Event ID 69223
510(K)Number K102422  
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
Product Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387, and 91393 are sold with Perioperative Mode (option D), which includes Start case/End case functionality, Upgrade Kit 040-1548-00.

Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Key monitored parameters available on the models 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed
Code Information The same KIT PART Number (PN 040-1458-00) identified on 12 units distributed IN the US and on 3 units distributed OUTSIDE of the US. These 15 units will receive serial numbers after the software upgrade is installed.   THERE ARE 70 SERIAL NUMBERS ON UNITS DISTRIBUTED IN THE US:  040-1548-00, 1367-000033, 1369-001517, 1369-003051, 1369-003077, 1369-004247, 1369-005914, 1369-006877, 1369-007165, 1369-008519, 1369-010292, 1369-011637, 1369-017520, 1369-061217, 1370-000013, 1370-000014, 1370-000025, 1370-000095, 1370-000285, 1370-000449, 1370-000471, 1370-000579, 1370-000658, 1370-001700, 1370-002494, 1370-002510, 1370-003785, 1370-004176, 1370-004184, 1370-004275, 1370-004292, 1387-000465, 1387-000535, 1387-000543, 1387-001071, 1387-001677, 1387-006200, 1387-010824, 1387-010830, 1387-011914, 1387-014142, 1387-014159, 1387-014184, 1387-014186, 1387-014187, 1387-014188, 1387-014596, 1387-015721, 1387-015722, 1387-015723, 1387-016045, 1387-018424, 1387-018947, 1387-019766, 1387-019768, 1387-019769, 1387-019771, 1387-019772, 1387-100036, 1387-101646, 1387-104672, 1387-105801, 1387-105811, 1387-105812, 1387-105882, 1393-000442, 1393-000579, 1393-000580, 1393-000581, 1393-000582, and 387-002421.   THERE ARE 137 SERIAL NUMBERS IDENTIFIED ON UNITS DISTRIBUTED OUTSIDE OF THE US:  1369-000011, 1369-000165, 1369-000407, 1369-000928, 1369-002976, 1369-003388, 1369-005316, 1369-005334, 1369-006100, 1369-006281, 1369-007338, 1369-009142, 1369-009143, 1369-009144, 1369-009145, 1369-009146, 1369-009889, 1369-009890, 1369-011741, 1369-011742, 1369-011743, 1369-011744, 1369-011745, 1369-013721, 1369-016648, 1369-016649, 1369-017759, 1369-017791, 1369-017792, 1370-000026, 1370-000725, 1370-000771, 1370-002373, 1370-002378, 1370-002380, 1370-002381, 1370-002382, 1370-002383, 1370-002384, 1370-003809, 1370-004133, 1370-004277, 1370-004509, 1370-004510, 1370-004511, 1370-004516, 1370-004617, 1370-200366, 1370-200370, 1370-200374, 1370-200375, 1370-200376, 1370-200378, 1370-200380, 1370-200471, 1370-200473, 1370-200474, 1370-200475, 1370-200476, 1370-200691, 1387-000044, 1387-000447, 1387-000467, 1387-000548, 1387-001086, 1387-002157, 1387-005006, 1387-008003, 1387-008310, 1387-010742, 1387-013241, 1387-013716, 1387-014610, 1387-018104, 1387-018105, 1387-018171, 1387-018176, 1387-018277, 1387-018638, 1387-018817, 1387-019563, 1387-100148, 1387-100150, 1387-100151, 1387-100155, 1387-100906, 1387-101061, 1387-101062, 1387-101063, 1387-101064, 1387-101065, 1387-101066, 1387-101067, 1387-102222, 1387-103550, 1387-103659, 1387-104462, 1387-105831, 1387-105832, 1387-105833, 1387-105834, 1393-000013, 1393-000517, 1393-000518, 1393-000519, 1393-000520, 1393-000521, 1393-000522, 1393-000523, 1393-000524, 1393-000525, 1393-000808, 1393-000809, 1393-000822, 1393-000831, 1393-000833, 1393-000834, 1393-000851, 1393-000852, 1393-00235, 1393-00236, 1393-00237, 1393-00238, 1393-00239, 1393-00240, 1393-00241, 1393-00335, 1393-00346, 1393-00347, 1393-00348, 1393-00349, 1393-00350, 1393-00351, 1393-00352, 1393-00353, 1393-00354, AND 1393-00355.
Recalling Firm/
Manufacturer
Spacelabs Healthcare Inc
35301 SE Center St
Snoqualmie WA 98065-9216
For Additional Information Contact Technical Support
800-522-7025
Manufacturer Reason
for Recall
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
FDA Determined
Cause 2
Environmental control
Action Spacelabs sent an "URGENT - MEDICAL DEVICE CORRECTION (Expanded)" letter dated 5 September 2014 to all consignees in the US. This letter identified recalled model 91367, 91369, 91370, 91387, and 91393, Upgrade Kit 040-1548-00, Equipped with the Perioperative Option (-D), problem, and actions to be taken. On September 12, 2014, a recall notification letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. The notification letter stated that the Operating Room staff may not be aware that alarm tones are off if the device has been left in the END CASE state. If the customers make use of the Perioperative START CASE / END CASE feature, they should advise the clinicians to familiarize themselves with this feature as described in the Perioperative section of the Operations Manual. In particular, they should know how to confirm the state of the alarm tones prior to starting a case. The firm's Field Service Engineers will contact customers to schedule a time to come out to update affected monitors at no cost. Customers have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025.
Quantity in Commerce 82 units distributed in US and 140 units distributed outside US
Distribution Worldwide Distribution- US (nationwide) including the states of: FL, GA, IA, KY, LA, MA, MI, MN, MO, MS, NJ, NY, OH, OK, PA, SC, TN, TX, WA, Guam and Puerto Rico and to the following countries: Argentina, Australia, Canada, China, Czech Republic, Finland, France, Germany, Greece, India, Israel, Italy, Netherlands, New Zealand, Panama, Paraguay, Poland, Singapore, Spain, Sri Lanka, Switzerland, Tunisia, United Kingdom, and Uruguay.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MHX and Original Applicant = SPACELABS MEDICAL INC.
-
-