| Class 2 Device Recall Stryker Orthopaedics | |
Date Initiated by Firm | August 15, 2014 |
Date Posted | May 19, 2015 |
Recall Status1 |
Terminated 3 on January 05, 2018 |
Recall Number | Z-1635-2015 |
Recall Event ID |
70994 |
510(K)Number | K042993 K070095 |
Product Classification |
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
|
Product | Stryker Orthopaedics Triathlon femoral component.
Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile.
For use in total reconstructive replacement of the knee joint |
Code Information |
Item number Lot number 5512-F-201 Lots JYET, KFVZ, KGLX ; 5512-F-202 Lots IPPOP, JYKX, KHZL, KJDN, KJHI, KJKY; 5512-F-301 Lots JTJA, KKAE, KKNB; 5512-F-302 Lots JXXF, KEYD, KFBY, KFTZ, KFXT, KHAB; 5512-F-401 Lots JWGY, KIED; 5512-F-402 Lots IUAJ, JJHT, JOMB, JWPF, JXXM, KGPE, KJMO, KLUP; 5512-F-501 Lots JXMV, KGKP; 5512-F-502 Lots KHWY, KJBK; 5512-F-601 Lots JIML, JIRX, KFAS, KIIY; 5512-F-602 Lots KEXE, KFOS, KFPI, KJAG, KKJW; 5512-F-701 Lot JKTF; 5512-F-801 Lots JWNL, JWPY; 5515-F-301 Lots IVXZD, JEPOD, JHZRA, JITNA, JITND,JJAFD, JJNYA, JJNYD, JKEND, JSNRD, JXBXA, JYFTD, JYGID, JYMTD,JYPJD, KFWHA, KFXID, KGAXD, KGJTD, KGXNA, KIBTA, KIEBA, KIGRD, KIYUD, KJABD, KJISD, KJRID; 5515-F-302 Lots IOLSD, IOLTD, IPRVD, IPTFD, IVGVA, IVLVD, JGIXD, JIGSD, JIYVD, JJSYA, JJWNA, JNVTA, JOWJD, JPFDA, JPHRA, JPIWA, JSZGD, JVYDD, JWPRA, JXBKD, JXOBA, JXXZD, JYBRA, JYBRD, JYKFD, JYKTA, JYMXA, JYNIA, JYNID, JYUED, JZDWD, JZEHD, JZFDD, KFNEA, KFOBA, KFOBD, KFOHD, KFVND, KGMFA, KGMFD, KGWGA, KGZFA, KHBGA, KIEXD, KJBFD, KJDLD, KJFXA, KJFXD, KJNPA, KKJOD; 5515-F-401 Lots JPDRA, JPIXD, JVLND, JVMTD, JXUDD, JYIDD, JYMUD, JZDSA, KABYA, KFDKD, KFMLA, KFPNA, KGAVD, KHEAA, KHKSA, KHZHA, KIATA, KIBVD, KIFMD, KIFOA, KIIFA, KJKBD, KKFYD; 5515-F-402 Lots JIFRD, JIGXD, JIMUD, JIRZD, JKESD, JLVSA, JLWTA, JOUGD, JPFEA, JPGKD, JYEIA, KAJID, KFSND, KGFLD, KGRBA, KGSHD, KGYOD, KJXDD, KKVHA, KLBHD, KLLBD; 5515-F-501 Lots HWHLA, IEXUA, IFDEA, IKEMD, JIAID, JILXA, JJBFD, JJTVD, JKGXD, JOXUA, JPMVD, JZRRA, KFSRA, KGIVD, KGKZA, KGSID, KGVHD, KGVPD, KHGPA, KHKAD, KHWRD, KHXWA, KJGBA, KKHFA, KKHFD, KLOSA; 5515-F-502 Lots IKRKA, JIDLA, JIEZD, JIVTA,JIYYD, JJWSA, JKAAD, JNVJA, JOPDD, JPFBD, JVUAD, JXBYD, JXFLA, JXTRD, JXYSA, JZHAD, JZKTD, KFNGD, KGTLA, KGTMA, KGYFA, KHGVA, KHGVA, KHGVD, KHOFD, KHPLA, KHRTA, KHWMA, KHWVA, KIDPA, KIJDA, KJGLD, KKMDD, KKOOD, KKPKA, KKWAA, KLBZA, KLMBA, KLSFA, KLSFD; 5515-F-601 Lots IPFTA, JIUMD, JJHZA, JMBTA, JWFHD, JWKTA, JXWXA, KABRD, KAGLA, KFISA, KFJXA, KGBVA, KGTDA, KHFSA, KHGNA, KHJAA, KHLIA, KHSWA, KHUAA, KHYED, KHZYA, KIGND, KJKOA; 5515-F-602 Lots JIYHD, JIZOA, JJBWD, JJWPD, JKDPD, JKERA, JPPKD, JJPSXD, JRBRA, JYTBD, KEZND, KFKDA, KFTMD, KFURA, KGIRA, KGNPD, KGRYD, KHJTD, KHPZA, KKXKA, KLKSA. |
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp. 325 Corporate Dr Mahwah NJ 07430-2006
|
For Additional Information Contact | Mr. Paul Jahnke 201-831-5826 |
Manufacturer Reason for Recall | The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | The firm, Stryker Orthopaedics, notified their Branches/Agencies via email on 8/15/2014 and sent an "Urgent Product Recall" letter dated 8/15/2014 with a Product Recall Acknowledgement Forms to their Branches/Agencies and consignees. The letter describes the product, issue, potential hazards and actions to be taken. The consignees were instructed to complete and return the attached Product Recall Acknowledgment Form within 5 days via fax to 855-251-3635; and return the affected product to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah NJ 07430.
If you have any questions, feel free to contact Sr. Regulatory Compliance Specialist to 201-831-5826. |
Quantity in Commerce | 1,147 in total |
Distribution | Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JWH
|
|
|
|