|
Class 2 Device Recall DePuy Orthopaedics LCS COMPLETE RPS |
|
Date Initiated by Firm |
March 04, 2015 |
Date Posted |
May 26, 2015 |
Recall Status1 |
Terminated 3 on October 30, 2015 |
Recall Number |
Z-1672-2015 |
Recall Event ID |
71085 |
PMA Number |
P830055 |
Product Classification |
Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing - Product Code NJL
|
Product |
LCS COMPLETE RPS Femoral Implants, various sizes and orientation. Knee prosthesis component for orthopedic surgery. |
Code Information |
All Lots. Catalog numbers: 129411010 129411020 129411030 129411040 129411050 129411060 129411070 129412010 129412020 129412030 129412040 129412050 129412060 129412070 129495040 129495050 129495060 129496040 129496050 129496060 |
Recalling Firm/ Manufacturer |
DePuy Orthopaedics, Inc. 700 Orthopaedic Dr Warsaw IN 46582-3994
|
For Additional Information Contact |
Kimberly C. Earle 574-371-4917
|
Manufacturer Reason for Recall |
The LCS COMPLETE RPS Knee System has higher rates of revision in Australia when the native patella is not resurfaced when compared to other unresurfaced PS implants. If the LCS COMPLETE RPS Knee System is implanted and the native patella is not resurfaced, the patient may experience patello-femoral pain, potentially requiring a secondary procedure.
|
FDA Determined Cause 2 |
Unknown/Undetermined by firm |
Action |
On 3/4/2015, URGENT INFORMATION - DEVICE CORRECTION NOTICE for the LCS COMPLETE RPS Knee System and on 3/19/2015 REVISED URGENT INFORMATION - DEVICE CORRECTION NOTICE for the LCS COMPLETE RPS Knee System (Revised) notifications were sent to the affected distributors and medical professionals. All distributors were notified via electronic mail, and all medical professionals with affected inventory were also notified by distributor or Sales Consultant via in person delivery or mail or electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. to 5 p.m. EDT). |
Quantity in Commerce |
19,059 |
Distribution |
Worldwide Distribution-US (nationwide) and the countries of CANADA, AUSTRALIA, AUSTRIA, BELGIUM, UK, FRANCE, GREECE, IRELAND, ISRAEL, ITALY, KOREA, LUXEMBOURG, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, SOUTH AFRICA, SWITZERLAND, THAILAND. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = NJL and Original Applicant = DEPUY, INC.
|
|
|
|