Class 2 Device Recall Fiber stripper |
|
Date Initiated by Firm |
May 28, 2015 |
Date Posted |
July 21, 2015 |
Recall Status1 |
Terminated 3 on January 19, 2016 |
Recall Number |
Z-2175-2015 |
Recall Event ID |
71534 |
Product Classification |
Powered laser surgical instrument - Product Code GEX
|
Product |
Tool, Stripper W/BLD, 12S, Autoclavable;
The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size. |
Code Information |
Model 80-10002-002; All product manufactured from April 8, 2012 to April 8, 2015 |
Recalling Firm/ Manufacturer |
American Medical Systems Innovation Center - Silicon Valley 3070 Orchard Dr San Jose CA 95134-2011
|
For Additional Information Contact |
Jennifer Mascioli-Tudor 408-456-3320 Ext. 6703
|
Manufacturer Reason for Recall |
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
|
FDA Determined Cause 2 |
Reprocessing Controls |
Action |
Urgent Medical Device Field notification letters were sent on June 9, 2015. Translation for countries out of the US will be conducted. Communications are being coordinated by the Prostate Health QA Director.
Urgent Medical Device Field notification Letters advised customers of the issue and list all affected products with name and model numbers. Any SureFlex Reuseable Fibers which have not been used but are in inventory and with the 3 year shelf life should be returned as per instructions enclosed.
Any of the SureFlex devices in inventory that have been used should be discarded or returned.
All Reuseable Stripper and Cleaver Accessories that have not been used should be returned. Used devices should be destroyed or returned. Questions to 1-408-456-3320. The Acknowledgement form should be completed and returned. |
Quantity in Commerce |
16,710 total affected devices |
Distribution |
Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa,
Australia, Germany, Netherlands, Spain,
Austria, Greece, New Zealand, Sri Lanka,
Belgium, Guatemala, Norway, Switzerland,
Bolivia, India, Panama, Thailand,
Brazil, Ireland, Peru, Turkey,
Canada, Israel, Poland, United Kingdom,
Chile, Italy, Portugal, United States,
Colombia, Korea, Qatar, Uruguay,
Costa Rica, Lebanon, Romania,
Finland, Malaysia and Saudi Arabia. |