|
Class 2 Device Recall Precision 500D R&F System |
|
Date Initiated by Firm |
June 12, 2015 |
Date Posted |
July 02, 2015 |
Recall Status1 |
Terminated 3 on December 21, 2015 |
Recall Number |
Z-1966-2015 |
Recall Event ID |
71572 |
510(K)Number |
K081091 K973039
|
Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
Product |
GE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems.
A universal diagnostic imaging system for radiographic and fluoroscopic examinations. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Legacy and Legacy D Table is intended for use in a diagnostic X-ray system to support patient during general purpose radiological procedures in the horizontal, vertical and Trendelenburg positions. |
Code Information |
Mfg Lot or Serial # Model, Catalogue or Code # 00000959002WK7 2258627 00000966338WK6 2258627 00000966611WK6 2258627 00000956387WK5 2258627 00000968741WK9 2258627 00000961510WK5 2258627 00000972529WK2 2258627 00000962527WK8 2258627 00000963526WK9 2258627 00000955594WK7 2258627 00000956587WK0 2258627 00000959354WK2 2258627 00000964963WK3 2258627 00000955822WK2 2258627 00000969028WK0 2258627 00000970010WK5 2258627 00000977743WK4 2258627 00000977741WK8 2258627 00000961823WK2 2258627 00000966949WK0 2258627 00000974520WK9 2258627 00000969300WK3 2258627 00000980632WK4 2258627 00000968416WK8 2258627 00000954452WK9 2258627 00000976569WK4 2258627 00000966357WK6 2258627 00000965746WK1 2258627 00000973833WK7 2258627 00000968411WK9 2258627 00000981014WK4 2258627 00000978814WK2 2258627 00000983374WK0 2258627 00000969352WK4 2258627 00000956745WK4 2258627 00000959683WK4 2258627 00000960138WK6 2258627 00000978807WK6 2258627 00000966622WK3 2258627 00000972384WK2 2258627 00000986313WK5 2258627 00000978457WK0 2258627 00000980889WK0 2258627 00000962460WK2 2258627 00000968783WK1 2258627 00000975678WK4 2258627 00000963491WK6 2258627 00000961786WK1 2258627 00000973156WK3 2258627 00000953812WK5 2258627 00000967126WK4 2258627 00000969297WK1 2258627 00000978314WK3 2258627 00000975431WK8 2258627 00000978316WK8 2258627 00000963305WK8 2258627 00000973016WK9 2258627 00000954777WK9 2258627 00000966068WK9 2258627 00001005270WK2 2258627 00000959073WK8 2258627 00000972383WK4 2258627 00000956964WK1 2258627 00000972706WK6 2258627 00000971136WK7 2258627 00000964804WK9 2258627 00000964266WK1 2258627 00000977455WK5 2258627 00000968409WK3 2258627 00000976669WK2 2258627 00000967926WK7 2258627 00000957269WK4 2258627 00000964398WK2 2258627 00000963525WK1 2258627 00000954518WK7 2258627 00000973621WK6 2258627 00000977456WK3 2258627 00000963944WK4 2258627 00000962678WK9 2258627 00000960132WK9 2258627 00000963153WK2 2258627 00000976624WK7 2258627 00000977748WK3 2258627 00000967770WK9 2258627 00000961102WK1 2258627 00000967162WK9 2258627 00000956746WK2 2258627 00000957833WK7 2258627 00000958809WK6 2258627 00000981461WK7 2258627 00000963738WK0 2258627 00000961659WK0 2258627 00000974519WK1 2258627 00000969627WK9 2258627 00000969663WK4 2258627 00000979197WK1 2258627 00000955947WK7 2258627 00000966773WK4 2258627 00000966069WK7 2258627 00000963182WK1 2258627 00000962674WK8 2258627 00000963359WK5 2258627 00000980499WK8 2258627 00000974170WK3 2258627 00000968417WK6 2258627 00000955630WK9 2258627 00000972386WK7 2258627 00000973490WK6 2258627 00000967849WK1 2258627 00000967749WK3 2258627 00000966126WK5 2258627 00000958577WK9 2258627 00000970853WK8 2258627 00000971022WK9 2258627 00000971647WK3 2258627 00000963268WK8 2258627 00000975597WK6 2258627 00000963293WK6 2258627 00000974524WK1 2258627 00000974525WK8 2258627 00000962333WK1 2258627 00000962369WK5 2258627 00000975576WK0 2258627 00000974350WK1 2258627 00000969628WK7 2258627 00000969619WK6 2258627 00000975429WK2 2258627 00000978458WK8 2258627 00000962420WK6 2258627 00000969308WK6 2258627 00000976623WK9 2258627 00000962031WK1 2258627 00000979786WK1 2258627 00000953898WK4 2258627 00000959784WK0 2258627 00000982420WK2 2258627 00000959414WK4 2258627 00000973188WK6 2258627 00000956847WK8 2258627 00000970377WK8 2258627 00000963897WK4 2258627 00000966070WK5 2258627 00000970171WK5 2258627 00000970785WK2 2258627 00000963310WK8 2258627 00000968044WK8 2258627 00000955368WK6 2258627 00000963898WK2 2258627 00000953103WK9 2258627 00000971138WK3 2258627 00000965553WK1 2258627 00000973234WK8 2258627 00000979616WK0 2258627 00000963795WK0 2258627 00000971649WK9 2258627 00000972819WK7 2258627 00000960280WK6 2258627 00000976673WK4 2258627 00000976869WK8 2258627 00000973231WK4 2258627 00000969285WK6 2258627 00000973835WK2 2258627 00000957800WK6 2258627 00000957841WK0 2258627 00000964580WK5 2258627 00000983490WK4 2258627 00000962169WK9 2258627 00000957327WK0 2258627 00000965905WK3 2258627 00000965310WK6 2258627 00000965330WK4 2258627 00000968048WK9 2258627 00000978315WK0 2258627 00000954154WK1 2258627 00000985033WK0 2258627 00000954100WK4 2258627 00000969215WK3 2258627 00000963637WK4 2258627 00000969664WK2 2258627 00000959134WK8 2258627 00000960066WK9 2258627 00000975577WK8 2258627 00000975652WK9 2258627 00000975912WK7 2258627 00000960819WK1 2258627 00000961165WK8 2258627 00000969574WK3 2258627 00000977147WK8 2258627 00000964803WK1 2258627 00000969536WK2 2258627 00000956034WK3 2258627 00000963658WK0 2258627 00000979787WK9 2258627 00000970714WK2 2258627 00000960804WK3 2258627 00000971021WK1 2258627 00000975149WK6 2258627 00000974902WK9 2258627 00000972388WK3 2258627 00000969502WK4 2258627 00000955254WK8 2258627 00000959852WK5 2258627 00000969865WK5 2258627 00000971677WK0 2258627 00000963354WK6 2258627 00000954749WK8 2258627 00000955540WK0 2258627 00000954009WK7 2258627 00000966433WK5 2258627 00000972820WK5 2258627 00000972380WK0 2258627 00000962630WK0 2258627 00000974527WK4 2258627 00000956253WK9 2258627 00000955150WK8 2258627 00000967364WK1 2258627 00000976518WK1 2258627 00000970643WK3 2258627 00000960279WK8 2258627 00000966515WK9 2258627 00000966123WK2 2258627 00000956075WK6 2258627 00000982419WK4 2258627 00000976993WK6 2258627 00000957655WK4 2258627 00000977045WK4 2258627 00000970749WK8 2258627 00000972102WK8 2258627 00000973711WK5 2258627 00000967777WK4 2258627 00000957419WK5 2258627 00000967924WK2 2258627 00000971903WK0 2258627 00000971825WK5 2258627 00000965309WK8 2258627 00000965584WK6 2258627 00000973813WK9 2258627 00000976666WK8 2258627 00000955118WK5 2258627 00000973838WK6 2258627 00000972229WK9 2258627 00000968173WK5 2258627 00000969204WK7 2258627 00000959328WK6 2258627 00000967751WK9 2258627 00000965983WK0 2258627 00000978808WK4 2258627 00000954383WK6 2258627 00000953219WK3 2258627 00000979585WK7 2258627 00000962377WK8 2258627 00000969322WK7 2258627 00000978805WK0 2258627 00000983498WK7 2258627 00000964195WK2 2258627 00000975428WK4 2258627 00000983334WK4 2258627 00000965500WK2 2258627 00000957454WK2 2258627 00000983144WK7 2258627 00000960518WK9 2258627 00000961773WK9 2258627 00000975146WK2 2258627 00000955239WK9 2258627 00000973715WK6 2258627 00000986958WK7 2258627 00000971529WK3 2258627 00000975421WK9 2258627 00000969027WK2 2258627 00000960544WK5 2258627 00000963090WK6 2258627 00000970438WK8 2258627 00000969088WK4 2258627 00000954382WK8 2258627 00000964575WK5 2258627 00000953474WK4 2258627 00000969537WK0 2258627 00000973186WK0 2258627 00000967990WK3 2258627 00000957588WK7 2258627 00000964536WK7 2258627 00000963118WK5 2258627 00000982417WK8 2258627 00000979617WK8 2258627 00000978670WK8 2258627 00000978806WK8 2258627 00000981015WK1 2258627 00000983757WK6 2258627 00000982589WK4 2258627 00000977149WK4 2258627 00000978809WK2 2258627 00000973836WK0 2258627 00000976675WK9 2258627 00000961852WK1 2258627 00000977742WK6 2258627 00000973834WK5 2258627 00000966516WK7 2258627 00000974346WK9 2258627 00000980273WK7 2258627 00000971826WK3 2258627 00000979789WK5 2258627 00000956274WK5 2258627 00000967778WK2 2258627 00000977461WK3 2258627 00000973233WK0 2258627 00000982421WK0 2258627 00000975683WK4 2258627 00000967161WK1 2258627 00000966866WK6 2258627 00000973839WK4 2258627 00000972169WK7 2258627 00000977747WK5 2258627 00000961360WK5 2258627 00000959586WK9 2258627 00000961257WK3 2258627 00000986314WK3 2258627 00000983204WK9 2258627 00000957713WK1 2258627 00000974349WK3 2258627 00000971305WK8 2258627 00000968047WK1 2258627 00000964113WK5 2258627 00000959377WK3 2258627 00000963961WK8 2258627 00000966562WK1 2258627 00000966599WK3 2258627 00000971396WK7 2258627 00000963221WK7 2258627 00000966468WK1 2258627 00000959255WK1 2258627 00000973622WK4 2258627 00000963490WK8 2258627 00000966024WK2 2258627 00000970543WK5 2258627 00000961718WK4 2258627 00000961735WK8 2258627 00000964094WK7 2258627 00000976667WK6 2258627 00000977148WK6 2258627 00000962434WK7 2258627 00000963629WK1 2258627 00000964350WK3 2258627 00000971530WK1 2258627 00000973158WK9 2258627 00000973623WK2 2258627 00000974903WK7 2258627 00000966247WK9 2258627 00000969721WK0 2258627 00000972817WK1 2258627 00000976674WK2 2258627 00000977745WK9 2258627 00000968174WK3 2258627 00000975653WK7 2258627 00000969405WK0 2258627 00000965497WK1 2258627 00000977740WK0 2258627 00000978183WK2 2258627 00000979788WK7 2258627 00000972103WK6 2258627 00000971607WK7 2258627 00000972167WK1 2258627 00000964482WK4 2258627 00000966674WK4 2258627 00000966673WK6 2258627 00000955304WK1 2258627 00000954233WK3 2258627 00000973714WK9 2258627 00000973235WK5 2258627 00000970748WK0 2258627 00000974904WK5 2258627 00000964268WK7 2258627 00000980890WK8 2258627 00000957700WK8 2258627 00000965270WK2 2258627 00000977044WK7 2258627 00000953843WK0 2258627 00000963809WK9 2258627 00000957167WK0 2258627 00000961413WK2 2258627 00000959491WK2 2258627 00000958433WK5 2258627 00000959074WK6 2258627 00000976671WK8 2258627 00000973841WK0 2258627 00000962617WK7 2258627 00000971824WK8 2258627 00000973712WK3 2258627 00000980275WK2 2258627 00000975142WK1 2258627 00000953793WK7 2258627 00000964972WK4 2258627 00000977744WK2 2258627 00000964962WK5 2258627 00000961273WK0 2258627 00000975129WK8 2258627 00000964742WK1 2258627 00000967340WK1 2258627 00000959679WK2 2258627 00000977042WK1 2258627 00000972385WK9 2258627 00000974348WK5 2258627 00000972818WK9 2258627 00000978185WK7 2258627 00000969351WK6 2258627 00000966642WK1 2258627 00000984282WK4 2258627 00000960648WK4 2258627 00000959130WK6 2258627 00000964525WK0 2258627 00000960602WK1 2258627 00000964264WK6 2258627 00000956792WK6 2258627 00000958222WK2 2258627 00000973837WK8 2258627 00000960935WK5 2258627 00000971674WK7 2258627 00000978318WK4 2258627 00000969026WK4 2258627 00000973187WK8 2258627 00000972168WK9 2258627 00000954215WK0 2258627 00000986512WK2 2258627 00000964802WK3 2258627 00000978456WK2 2258627 00000975145WK4 2258627 00000975575WK2 2258627 00000963056WK7 2258627 00000983202WK3 2258627 00000961660WK8 2258627 00000961739WK0 2258627 00000965604WK2 2258627 00000968782WK3 2258627 00000963063WK3 2258627 00000956295WK0 2258627 00000966946WK6 2258627 00000967103WK3 2258627 00000965747WK9 2258627 00000955971WK7 2258627 00000954610WK2 2258627 00000973840WK2 2258627 00000964138WK2 2258627 00000955452WK8 2258627 00000970713WK4 2258627 00000975911WK9 2258627 00000959065WK4 2258627 00000958644WK7 2258627 00000975141WK3 2258627 00000978460WK4 2258627 00000959502WK6 2258627 00000967416WK9 2258627 00000954803WK3 2258627 00000975682WK6 2258627 00000978313WK5 2258627 00000962910WK6 2258627 00000961772WK1 2258627 00000962611WK0 2258627 00000978813WK4 2258627 00000983500WK0 2258627 00000972705WK8 2258627 00000970172WK3 2258627 00000972104WK4 2258627 00000976065WK3 2258627 00000974347WK7 2258627 00000966029WK1 2258627 00000959121WK5 2258627 00000953176WK5 2258627 00000962651WK6 2258627 00000964096WK2 2258627 00000964161WK4 2258627 00000973230WK6 2258627 00000959001WK9 2258627 00000954262WK2 2258627 00000976870WK6 2258627 00000967417WK7 2258627 00000960008WK1 2258627 00000978810WK0 2258627 00000982588WK6 2258627 00000956507WK8 2258627 00000957937WK6 2258627 00000969618WK8 2258627 00000955220WK9 2258627 00000955387WK6 2258627 00000975430WK0 2258627 00000954441WK2 2258627 00000972232WK3 2258627 00000968641WK1 2258627 00000972105WK1 2258627 00000965614WK1 2258627 00000965681WK0 2258627 00000968046WK3 2258627 00000983335WK1 2258627 00000953452WK0 2258627 00000964806WK4 2258627 00000963015WK3 2258627 00000980633WK2 2258627 00000955595WK4 2258627 00000976437WK4 2258627 00000970175WK6 2258627 00000978455WK4 2258627 00000956645WK6 2258627 00000974462WK4 2258627 00000978182WK4 2258627 00000953277WK1 2258627 00000968410WK1 2258627 00000973812WK1 2258627 00000974171WK1 2258627 00000974521WK7 2258627 00000975128WK0 2258627 00000980500WK3 2258627 00000970330WK7 2258627 00000966006WK9 2258627 00000978812WK6 2258627 00000964112WK7 2258627 00000964149WK9 2258627 00000964961WK7 2258627 00000975150WK4 2258627 00000983499WK5 2258627 00000981459WK1 2258627 00000966030WK9 2258627 00000959544WK8 2258627 00000976868WK0 2258627 00000977043WK9 2258627 00000966774WK2 2258627 00000970160WK8 2258627 00000970544WK3 2258627 00000976407WK7 2258627 00000976625WK4 2258627 00000975419WK3 2258627 00000975148WK8 2258627 00000976324WK4 2258627 00000962642WK5 2258627 00000971606WK9 2258627 00000976668WK4 2258627 00000968175WK0 2258627 00000970851WK2 2258627 00000971676WK2 2258627 00000976408WK5 2258627 00000978181WK6 2258627 00000970174WK9 2258627 00000971104WK5 2258627 00000974169WK5 2258627 00000967748WK5 2258627 00000977749WK1 2258627 00000966010WK1 2258627 00000963212WK6 2258627 00000980274WK5 2258627 00000974893WK0 2258627 00000976670WK0 2258627 00000966621WK5 2258627 00000958916WK9 2258627 00000958445WK9 2258627 00009881460WK9 2258627 00000974523WK3 2258627 00000959506WK7 2258627 00000958294WK1 2258627 00000959853WK3 2258627 00000973504WK4 2258627 00000953744WK0 2258627 00000953647WK5 2258627 00000979790WK3 2258627 00000956473WK3 2258627 00000962079WK0 2258627 00000976520WK7 2258627 00000960561WK9 2258627 00000979791WK1 2258627 00000971531WK9 2258627 00000960884WK5 2258627 00000979195WK5 2258627 00000982401WK2 2258627 00000958345WK1 2258627 00000972382WK6 2258627 00000967925WK9 2258627 00000964265WK3 2258627 00000954928WK8 2258627 00000963055WK9 2258627 00000969503WK2 2258627 00000969724WK4 2258627 00000989626WK7 2258627 00000990025WK9 2258627 00000988711WK8 2258627 00000958696WK7 2258627 00000989625WK9 2258627 00000986506WK4 2258627 00000989627WK5 2258627 00000957389WK0 2258627 00000961277WK1 2258627 00000989885WK9 2258627 00000961619WK4 2258627 00000961630WK1 2258627 00000973158WK9 2258627 00000956861WK9 2258627 00000955704WK2 2258627 00000961927WK1 2258627 00000976119WK8 2258627 00000976325WK1 2258627 00000961757WK2 2258627 00000978317WK6 2258627 00000957151WK4 2258627 00000957938WK4 2258627 00000982307WK1 2305472 00000978653WK4 2305472 00000973600WK0 2305472 00000979189WK8 2305472 00000981005WK2 2305472 00000975153WK8 2305472 00000976622WK1 2305472 00000981007WK8 2305472 00000984704WK7 2305472 00000987749WK9 2305472 00000986295WK4 2305472 00000980260WK2 2305472 00000978075WK0 2305472 00000982981WK3 2305472 00000988661WK5 2305472 00000989101WK1 2305472 00000986515WK5 2305472 00000986291WK3 2305472 00000985428WK2 2305472 00000990129WK9 2305472 00000979501WK4 2305472 00000980260WK4 2305472 00000977708WK7 2305472 00000979191WK4 2305472 00000990839WK3 2305472 00000982892WK2 2305472 00000977716WK0 2305472 00000979781WK2 2305472 00000979920WK6 2305472 00000981453WK4 2305472 00000979784WK6 2305472 00000984702WK1 2305472 00000983783WK2 2305472 00000974036WK6 2305472 00000982579WK5 2305472 00000989785WK1 2305472 00000989097WK1 2305472 00000983855WK8 2305472 00000977348WK2 2305472 00000978500WK7 2305472 00000982894WK8 2305472 00000976055WK4 2305472 00000989781WK0 2305472 00000981295WK9 2305472 00000978649WK2 2305472 00000977041WK3 2305472 00000978652WK6 2305472 00000979503WK0 2305472 00000975954WK9 2305472 00000980841WK1 2305472 00000981750WK3 2305472 00000975953WK1 2305472 00000978177WK4 2305472 00000987753WK1 2305472 00000984284WK0 2305472 00000981790WK9 2305472 00000979188WK0 2305472 00000987733WK3 2305472 00000981458WK3 2305472 00000980625WK8 2305472 00000976801WK1 2305472 00000982368WK3 2305472 00000097962WK2 2305472 00000979394WK4 2305472 00000981291WK8 2305472 00000982399WK8 2305472 00000981791WK7 2305472 00000983134WK8 2305472 00000979502WK2 2305472 00000982742WK9 2305472 00000981074WK8 2305472 00000972840WK3 2305472 00000972514WK4 2305472 00000986118WK8 2305472 00000979357WK1 2305472 00000979557WK6 2305472 00000981454WK2 2305472 00000983133WK0 2305472 00000977710WK3 2305472 00000983137WK1 2305472 00000978174WK1 2305472 00000981456WK7 2305472 00000987751WK5 2305472 00000979618WK6 2305472 00000979575WK8 2305472 00000988666WK4 2305472 00000977376WK3 2305472 00000981582WK0 2305472 00000985423WK3 2305472 00981296WK7000 2305472 00000986115WK4 2305472 00000980628WK2 2305472 00000989931WK1 2305472 00000977232WK8 2305472 00000977715WK2 2305472 00000978798WK7 2305472 00000975292WK4 2305472 00000978499WK2 2305472 00000980264WK6 2305472 00000977040WK5 2305472 00000986120WK4 2305472 00000978175WK8 2305472 00000982031WK7 2305472 00000982398WK0 2305472 00000977230WK2 2305472 00000972896WK5 2305472 00000973599WK4 2305472 00000974037WK4 2305472 00000973973WK1 2305472 00000974343WK6 2305472 00000977231WK0 2305472 00000986821WK7 2305472 00000980495WK6 2305472 00000987737WK4 2305472 00000976865WK6 2305472 00000982034WK1 2305472 00000986518WK9 2305472 00000977374WK8 2305472 00000984225WK3 2305472 00000984283WK2 2305472 00000979192WK2 2305472 00000987734WK1 2305472 00000981581WK2 2305472 00000983035WK7 2305472 00000982795WK7 2305472 00000979621WK0 2305472 00000098100WK7 2305472 00000988660WK7 2305472 00000986292WK1 2305472 00000986935WK5 2305472 00000979190WK6 2305472 00000989777WK8 2305472 00000975910WK1 2305472 00000975951WK5 2305472 00000989784WK4 2305472 00000983859WK0 2305472 00000981457WK5 2305472 00000980187WK9 2305472 00000980497WK2 2305472 00000990840WK1 2305472 00000988659WK9 2305472 00000982800WK5 2305472 00000978650WK0 2305472 00000980498WK0 2305472 00000976620WK5 2305472 00000979785WK3 2305472 00000974459WK0 2305472 00000980263WK8 2305472 00000975950WK7 2305472 00000986516WK3 2305472 00000981584WK6 2305472 00000989786WK9 2305472 00000989096WK3 2305472 00000983225WK4 2305472 00000982891WK4 2305472 00000988664WK9 2305472 00000978076WK8 2305472 00000984285WK7 2305472 00000981789WK1 2305472 00000987750WK7 2305472 00000981583WK8 2305472 00000979919WK8 2305472 00000984287WK3 2305472 00000989779WK4 2305472 00000980118WK4 2305472 00000990128WK1 2305472 00000985028WK0 2305472 00000979921WK4 2305472 00000981069WK8 2305472 00000982893WK0 2305472 00000989778WK6 2305472 00000989939WK4 2305472 00000989938WK6 2305472 00000979619WK4 2305472 00000989715WK8 2305472 00000990126WK5 2305472 00000974342WK8 2305472 00000979360WK5 2305472 00000979780WK4 2305472 00000978497WK6 2305472 00000975908WK5 2305472 00000980629WK0 2305472 00000989782WK8 2305472 00000989783WK6 2305472 00000986119WK6 2305472 00000982311WK3 2305472 00000990831WK0 2305472 00000976621WK3 2305472 00000979783WK8 2305472 00000978498WK4 2305472 00000976057WK0 2305472 00000981078WK9 2305472 00000979359WK7 2305472 00000988657WK3 2305472 00000975152WK0 2305472 00000980630WK8 2305472 00000982799WK9 2305472 00000991211WK4 2305472 00000987061WK9 2305472 00000986290WK5 2305472 00000981916WK0 2305472 00000971110WK2 2305472 00000980840WK3 2305472 00000977711WK1 2305472 00000991213WK0 2305472 00000977039WK7 2305472 00000981754WK5 2305472 00000980624WK1 2305472 00000978797WK9 2305472 00000980036WK8 2305472 00000987729WK1 2305472 00000984223WK8 2305472 00000983535WK6 2305472 00000980265WK3 2305472 00000989099WK7 2305472 00000978179WK0 2305472 00000983422WK7 2305472 00000987728WK3 2305472 00000978180WK8 2305472 00000986936WK3 2305472 00000991214WK8 2305472 00000987755WK6 2305472 00000982037WK4 2305472 00000991234WK6 2305472 00000977712WK9 2305472 00000981751WK1 2305472 00000978804WK3 2305472 00000986519WK7 2305472 00000981290WK0 2305472 00000984229WK5 2305472 00000977713WK7 2305472 00000980842WK9 2305472 00000981914WK5 2305472 00000982577WK9 2305472 00000982308WK9 2305472 00000981915WK2 2305472 00000975952WK3 2305472 00000985029WK8 2305472 00000986117WK0 2305472 00000982310WK5 2305472 00000982890WK6 2305472 00000982371WK7 2305472 00000990127WK3 2305472 00000981292WK6 2305472 00000982372WK5 2305472 00000986514WK8 2305472 00000982576WK1 2305472 00000989935WK2 2305472 00000991232WK0 2305472 00000982578WK7 2305472 00000982797WK3 2305472 00000981072WK2 2305472 00000982798WK1 2305472 00000983699WK0 2305472 00000991210WK6 2305472 00000981294WK2 2305472 00000979622WK8 2305472 00000981071WK4 2305472 00000991231WK2 2305472 00000982030WK9 2305472 00000979500WK6 2305472 00000987736WK6 2305472 00000974344WK4 2305472 00000975509WK1 2305472 00000971696WK0 2305472 00000989102WK9 2305472 00000982188WK5 2305472 00000983533WK1 2305472 00000987062WK7 2305472 00000986296WK2 2305472 00000990837WK7 2305472 00000990834WK4 2305472 00000981455WK9 2305472 00000982396WK4 2305472 00000978129WK5 2305472 00000980626WK6 2305472 00000982312WK1 2305472 00000989934WK5 2305472 00000990838WK5 2305472 00000981293WK4 2305472 00000983229WK6 2305472 00000982747WK8 2305472 00000989714WK1 2305472 00000984701WK3 2305472 00000986937WK1 2305472 00000983856WK6 2305472 00000990130WK7 2305472 00000984224WK6 2305472 00000985426WK6 2305472 00000981077WK1 2305472 00000974761WK9 2305472 00000982033WK3 2305472 00000975909WK3 2305472 00000975416WK9 2305472 00000977345WK8 2305472 00000974460WK8 2305472 00000977347WK4 2305472 00000977346WK6 2305472 00000986521WK3 2305472 00000985427WK4 2305472 00000978803WK5 2305472 00000991233WK8 2305472 00000980034WK3 2305472 00000980117WK6 2305472 00000985424WK1 2305472 00000983228WK8 2305472 00000983230WK4 2305472 00000983857WK4 2305472 00000983532WK3 2305472 00000974760WK1 2305472 00000974759WK3 2305472 00000982036WK6 2305472 00000986116WK2 2305472 00000986293WK9 2305472 00000989103WK7 2305472 00000989104WK5 2305472 00000989095WK5 2305472 00000980033WK5 2305472 00000980165WK5 2305472 00000989932WK9 2305472 00000982746WK0 2305472 00000982796WK5 2305472 00000979193WK0 2305472 00000980164WK8 2305472 00000983534WK9 2305472 00000984286WK5 2305472 00000986289WK7 2305472 00000979918WK0 2305472 00000987752WK3 2305472 00000986294WK7 2305472 00000988658WK1 2305472 00000983132WK2 2305472 00000982373WK3 2305472 00000982395WK6 2305472 00000983136WK3 2305472 00000979397WK7 2305472 00000980119WK2 2305472 00000978495WK0 2305472 00000980496WK4 2305472 00000980032WK7 2305472 00000974206WK5 2305472 00000981070WK6 2305472 00000981753WK7 2305472 00000979393WK6 2305472 00000979392WK8 2305472 00000987067WK6 2305472 00000978800WK1 2305472 00000983531WK5 2305472 00000983775WK8 2305472 00000985030WK6 2305472 00000982187WK7 2305472 00000978796WK1 2305472 00000981752WK9 2305472 00000978802WK7 2305472 00000978651WK8 2305472 00000981788WK3 2305472 00000983858WK2 2305472 00000982982WK1 2305472 00000988665WK6 2305472 00000987065WK0 2305472 00000981004WK5 2305472 00000979499WK1 2305472 00000981075WK5 2305472 00000987756WK4 2305472 00000983536WK4 2305472 00000987064WK3 2305472 00000978795WK3 2305472 00000981792WK5 2305472 00000979545WK1 2305472 00000975293WK2 2305472 00000980262WK0 2305472 00000981289WK2 2305472 00000979361WK3 2305472 00000979396WK9 2305472 00000982743WK7 2305472 00000989937WK8 2305472 00000982397WK2 2305472 00000987066WK8 2305472 00000982035WK8 2305472 00000983135WK5 2305472 00000978799WK5 2305472 00000989936WK0 2305472 00000978178WK2 2305472 00000979395WK1 2305472 00000983423WK5 2305472 00000981918WK6 2305472 00000977375WK5 2305472 00000981006WK0 2305472 00000987754WK9 2305472 00000986517WK1 2305472 00000984220WK4 2305472 00000984222WK0 2305472 00000981749WK5 2305472 00000973963WK2 2305472 00000985796WK2 2305495 00000014344GP6 2305495 00000009804GP6 2305495 00000984436WK6 2305495 00000990179WK4 2305495 00000010388GP7 2305495 00000010958GP7 2305495 00000013048GP4 2305495 00000010959GP5 2305495 00000984899WK5 2305495 00000010546GP0 2305495 00000989488WK2 2305495 00000012826GP4 2305495 00000008471GP5 2305495 00000089726DP4 2343125 00000118441DP5 2343125 00000120258DP9 2343125 00000089719DP9 2343125 00000121252DP1 2343125 00000118443DP1 2343125 00000097765DP2 2343125 00000081801DP3 2343125 00000067515DP7 2343125 00000068784DP8 2343125 00000065980DP5 2343125 00000111602DP9 2343125 00000081599DP3 2343125 00000097764DP5 2343125 00000120793DP5 2343125 00000119459DP6 2343125 00000100420DP9 2343125 00000109235DP2 2343125 00000118446DP4 2343125 00000086753DP1 2343125 00000089721DP5 2343125 00000082084DP5 2343125 00000095789DP4 2343125 00000113676DP1 2343125 00000120792DP7 2343125 00000107133DP1 2343125 00000087393DP5 2343128 00000085037DP0 2343128 00000105787DP6 2343128 00000135163DP4 2343128 00000121170DP5 2343128 00000105789DP2 2343128 00000080562DP2 2343128 00000080558DP0 2343128 00000064764DP4 2343128 00000066019DP1 2343128 00000085030DP5 2343128 00000115370DP9 2343128 00000095802DP5 2343128 00000105788DP4 2343128 00000093978DP5 2343128 00000080067DP2 2343128 00000064144DP9 2343128 00000117092DP7 2343128 00000133526DP4 2343128 00000080527DP5 2343128 00000068322DP7 2343128 00000090028DP2 2343128 00000129711DP8 2343128 00000113685DP2 2343128 00000063938DP5 2343128 00000133205DP5 2343128 00000129949DP4 2343128 00000064162DP1 2343128 00000121600DP1 2343128 00000112670DP5 2343128 00000095805DP8 2343128 00000087392DP7 2343128 00000112374DP4 2343128 00000065991DP2 2343128 0000000B0125LE 2343129 00000112680DP4 2343129 00000097770DP2 2343129 00000122578DP8 2343129 00000066235DP3 2343129 00000120246DP4 2343129 00000104749DP7 2343129 00000064196DP9 2343129 00000118774DP9 2343129 00000066222DP1 2343129 00000091448DP1 2343129 00000112683DP8 2343129 00000120239DP9 2343129 00000066296DP5 2343130 00000066261DP9 2343130 00000066254DP4 2343130 00000066259DP3 2343130 00000998538WK3 2403791 00000991363WK3 2403791 00000991903WK6 2403791 00000991908WK5 2403791 00000991905WK1 2403791 00000993103WK1 2403791 00000993040WK5 2403791 00001001363WK9 2403791 00000998535WK9 2403791 00000996806WK6 2403791 00001008860WK7 2403791 00000996448WK7 2403791 00001008918WK3 2403791 979917WK2 2403791 00000994733WK4 2403791 00000991907WK7 2403791 00001002235WK8 2403791 00000997357WK9 2403791 00000992312WX9 2403791 00000996797WK7 2403791 00001007608WK1 2403791 00001008710WK4 2403791 00001009278WK1 2403791 00000993953WK9 2403791 00000998970WK8 2403791 00000997826WK3 2403791 00000993949WK7 2403791 00001008707WK0 2403791 00001005375WK9 2404103 00000005421M33 2404103 00000023357M37 2404103 00000012142M36 2404103 00000010127M39 2404103 00001005395WK7 2404103 00000022063M32 2404103 00000999917WK8 2404103 00000003456M31 2404103 00000012607M38 2404103 00000017483M39 2404103 00000008740M33 2404103 00000002874M36 2404103 00000993343WK3 2404103 00000004858M37 2404103 00000005274M36 2404103 00000002216M30 2404103 00000012808M32 2404103 00000005341M33 2404103 00000005342M31 2404103 00000994433WK1 2404103 00000012639M31 2404103 00000012920M35 2404103 00000012221M38 2404103 00001000383WK8 2404103 00000012380M32 2404103 00000012640M39 2404103 00000009119M39 2404103 00000004238M32 2404103 00000009712M31 2404103 00000997869WK3 2404103 00000011890M31 2404103 00000008862M35 2404103 00000011757M32 2404103 00000009313M38 2404103 00000009405M32 2404103 00000013012M30 2404103 00000012297M38 2404103 00000013942M38 2404103 00000001653M35 2404103 00000998425WK3 2404103 00000012185M35 2404103 00000997663WK0 2404103 00000006861M39 2404103 00000997109WK4 2404103 00000996484WK2 2404103 00000003827M33 2404103 00000011033M38 2404103 00000007304M39 2404103 00000992835WK9 2404103 00000016715M35 2404103 00000008197M36 2404103 00000994432WK3 2404103 00000012661M35 2404103 00000995343WK1 2404103 00000006689M34 2404103 00000999057WK3 2404103 00000999915WK2 2404103 00000009181M39 2404103 00000017846M37 2404103 995346WK4 2404103 00000010281M34 2404103 00001002912WK2 2404103 00001003370WK2 2404103 00000006095M34 2404103 00000007879M30 2404103 00001001865WK3 2404103 00000008656M31 2404103 00000997102WK9 2404103 00000013747M31 2404103 00000012220M30 2404103 00001002911WK4 2404103 00000019196M35 2404103 00000991636WK2 2404103 00000997149WK0 2404103 00000006507M38 2404103 00000011296M31 2404103 00000007191M30 2404103 00000995339WK9 2404103 00000023395M37 2404103 00000008278M34 2404103 00001004454WK3 2404103 00000999061WK5 2404103 00000004723M33 2404103 00000026028M31 2404103 00000009871M35 2404103 00000001431M36 2404103 00000013191M32 2404103 00000996488WK3 2404103 00000002772M32 2404103 00000017415M31 2404103 00000011032M30 2404103 00000002349M39 2404103 00000996158WK2 2404103 00000006332M31 2404103 00000010128M37 2404103 00000024493M39 2404103 00000010625M32 2404103 00000011818M32 2404103 00000012382M38 2404103 00000026259M32 2404103 00000019346M36 2404103 00000993167WK6 2404103 00000006559M39 2404103 00000006696M39 2404103 00001000993WK4 2404103 00000012270M35 2404103 00000014697M37 2404103 00000997867WK7 2404103 00000993347WK4 2404103 00000003056M39 2404103 00000002981M39 2404103 00000010935M35 2404103 00000002347M33 2404103 00000993350WK8 2404103 00000012164M30 2404103 00000006153M31 2404103 00000009669M33 2404103 00000008307M31 2404103 00000007947M35 2404103 00000021570M37 2404103 00000999388WK2 2404103 00000999387WK4 2404103 00000006149M39 2404103 00000011116M31 2404103 00000997110WK2 2404103 00000008562M31 2404103 00000020787M38 2404103 00000998432WK9 2404103 00000999386WK6 2404103 00000011658M32 2404103 00000021714M31 2404103 00000997665WK5 2404103 00000006254M37 2404103 00000024197M36 2404103 00000999914WK5 2404103 00000024065M35 2404103 00000007240M35 2404103 00000022255M34 2404103 00000023680M32 2404103 00000007826M31 2404103 00000002640M31 2404103 00000012560M39 2404103 00000025221M33 2404103 00000995342WK3 2404103 00000011549M33 2404103 00000013514M35 2404103 00000015869M31 2404103 00000024976M33 2404103 00000005944M34 2404103 00001001054WK4 2404103 00000009948M31 2404103 00000010490M31 2404103 00000004861M31 2404103 00000008279M32 2404103 00000007946M37 2404103 00001004954WK2 2404103 00001004969WK0 2404103 00001005388WK2 2404103 00000002643M35 2404103 00001005376WK7 2404103 00001000990WK0 2404103 00001001863WK8 2404103 00000994426WK5 2404103 00000014032M37 2404103 00000990408WK7 2404103 00000024697M35 2404103 00000010376M32 2404103 00000007384M31 2404103 00000999389WK0 2404103 00001000995WK9 2404103 00000017127M32 2404103 00000001433M32 2404103 00000004533M36 2404103 00000010626M30 2404103 00000011790M33 2404103 00000006558M31 2404103 00000005423M39 2404103 00000002773M30 2404103 00000003580M33 2404103 00000011115M33 2404103 00000006331M33 2404103 00000010881M31 2404103 00000009180M31 2404103 00000996480WK0 2404103 00000008007M37 2404103 00000997865WK1 2404103 00000002776M33 2404103 00001005391WK6 2404103 00000005425M34 2404103 00000003824M30 2404103 00000010882M39 2404103 00000008657M39 2404103 00001004451WK9 2404103 00000003581M36 2404103 00000022698M35 2404103 00000009670M31 2404103 00000021093M30 2404103 00000021712M35 2404103 00000010215M32 2404103 00000992837WK5 2404103 00000999920WK2 2404103 00000011864M36 2404103 00000002641M39 2404103 00000014008M37 2404103 00000011449M36 2404103 00000002466M31 2404103 00000003984M32 2404103 00000013349M36 2404103 00000992392WK1 2404103 00000009310M34 2404103 00000001271M39 2404103 00000003983M34 2404103 00000003053M36 2404103 00001003073WK2 2404103 00000017128M30 2404103 00001004653WK0 2404103 00000012331M35 2404103 00000004233M33 2404103 00000011569M31 2404103 00000012807M34 2404103 00000025249M34 2404103 00001006207WK3 2404103 00001001867WK9 2404103 00000001235M31 2404103 00000993163WK5 2404103 00000001189M30 2404103 00000002778M39 2404103 00000003199M37 2404103 00001002434WK7 2404103 00001002435WK4 2404103 00000003193M30 2404103 00000003055M31 2404103 00000995871WK1 2404103 00001006204WK0 2404103 00001003373WK6 2404103 00000004715M39 2404103 00000991653WK7 2404103 00000023455M39 2404103 00000001887M39 2404103 00000022747M30 2404103 00000001838M32 2404103 00001004458WK4 2404103 00000995876WK0 2404103 00000009820M32 2404103 00000005426M32 2404103 00000023454M32 2404103 00000022406M33 2404103 00000001447M32 2404103 00000012774M36 2404103 00001004963WK3 2404103 00000022699M33 2404103 00000019489M34 2404103 00000007107M36 2404103 00000008194M33 2404103 00001004959WK1 2404103 00000004863M37 2404103 00000005491M36 2404103 00000013303M33 2404103 00000007881M36 2404103 00000010738M33 2404103 0000O011837M32 2404103 00000011226M38 2404103 00000004968M34 2404103 00000013350M34 2404103 00000010809M32 2404103 00000008849M32 2404103 00001004453WK5 2404103 00000002464M36 2404103 00000005336M33 2404103 00000005156M35 2404103 00000005654M39 2404103 00000997660WK6 2404103 00001004970WK8 2404103 00001004240WK6 2404103 00000002871M32 2404103 00000009311M32 2404103 00000023358M35 2404103 00000007829M35 2404103 00001002914WK8 2404103 00000006859M33 2404103 00000002217M38 2404103 00000002353M31 2404103 00000009401M31 2404103 00000014566M34 2404103 00000005651M35 2404103 00000994425WK7 2404103 00000008658M37 2404103 00000004239M30 2404103 00000003200M33 2404103 00000009947M33 2404103 00000011943M38 2404103 00000005424M37 2404103 00000005942M38 2404103 00001004965WK8 2404103 00000021356M31 2404103 00000022015M32 2404103 00000022049M31 2404103 00000013903M30 2404103 00000003975M30 2404103 00000017255M31 2404103 00000017125M36 2404103 00000008776M37 2404103 00000009275M39 2404103 00000023457M35 2404103 00000023727M31 2404103 00001003372WK8 2404103 00000020300M30 2404103 00000010124M36 2404103 00000012662M33 2404103 00000004243M32 2404103 00000004227M35 2404103 00000010380M34 2404103 00000013902M32 2404103 00000011434M38 2404103 00000006692M38 2404103 00000005111M30 2404103 00000005498M31 2404103 00000012528M36 2404103 00000001837M34 2404103 00000025562M30 2404103 00000026330M31 2404103 00000991639WK6 2404103 00000999392WK4 2404103 00000009532M33 2404103 00000009120M37 2404103 00000002639M33 2404103 00000005813M31 2404103 00000001835M38 2404103 00000013872M37 2404103 00000013493M32 2404103 00000005335M35 2404103 00000004231M37 2404103 00001004244WK8 2404103 00000008074M37 2404103 00001003369WK4 2404103 00000007584M36 2404103 00000003578M32 2404103 00000001658M34 2404103 00000013748M39 2404103 00000994429WK9 2404103 00000994428WK1 2404103 00000993348WK2 2404103 00001001057WK7 2404103 00001004951WK8 2404103 00000017195M39 2404103 00000996487WK5 2404103 00000009277M35 2404103 00000012865M32 2404103 00000005275M33 2404103 00000012025M33 2404103 00000021270M34 2404103 00000007505M31 2404103 00000014033M35 2404103 00000001967M39 2404103 00000002124M36 2404103 00000010286M33 2404103 00000012381M30 2404103 00000005273M38 2404103 00000022356M30 2404103 00000015914M35 2404103 00000004535M31 2404103 00000004859M35 2404103 00000017943M32 2404103 00000004644M31 2404103 00000995879WK4 2404103 00000012771M32 2404103 00000997106WK0 2404103 00000012866M30 2404103 00000001654M33 2404103 00000009118M31 2404103 00000005943M36 2404103 00000004716M37 2404103 00000006560M37 2404103 00000005270M34 2404103 00000002774M38 2404103 00000992217WK0 2404103 00000996483WK4 2404103 00000013176M33 2404103 00000008198M34 2404103 00000992394WK7 2404103 00001001072WK6 2404103 00000995347WK2 2404103 00000014000M34 2404103 00000007939M32 2404103 00000998438WK6 2404103 00000998422WK0 2404103 00001002437WK0 2404103 00000008408M37 2404103 00000016799M39 2404103 00000012513M38 2404103 00000007773M35 2404103 00000011114M36 2404103 00000007336M31 2404103 00000008309M37 2404103 00000005276M31 2404103 00000991645WK3 2404103 00000007243M39 2404103 00000002117M30 2404103 00000997115WK1 2404103 00000007583M38 2404103 00000012660M37 2404103 00000009671M39 2404103 00000008774M32 2404103 00000008962M33 2404103 00000002869M36 2404103 00000002644M33 2404103 00000991630WK5 2404103 00000016402M30 2404103 00000022908M38 2404103 00000002120M34 2404103 00000004536M39 2404103 00001005374WK2 2404103 00000990547WK2 2404103 00000994430WK7 2404103 00000012332M33 2404103 00000008860M39 2404103 00000003582M34 2404103 00000005109M34 2404103 00000993787WK1 2404103 00000025397M31 2404103 00000017850M39 2404103 00000010739M31 2404103 00000008858M33 2404103 00000005334M38 2404103 00000008961M35 2404103 00000014273M37 2404103 00000024861M37 2404103 00000003583M32 2404103 00000997104WK5 2404103 00000003981M38 2404103 00000016641M33 2404103 00000004529M34 2404103 00000005269M36 2404103 00000005998M30 2404103 00000006697M37 2404103 00000012922M31 2404103 00000004648M32 2404103 00000010219M34 2404103 00000997103WK7 2404103 00001003380WK1 2404103 00000002638M35 2404103 00000005941M30 2404103 00000011182M33 2404103 00000003575M38 2404103 00000991628WK9 2404103 00000008193M35 2404103 00001001071WK8 2404103 00000006152M33 2404103 00000006335M34 2404103 00000991640WK4 2404103 00000002214M35 2404103 00000996162WK4 2404103 00000995344WK9 2404103 00000012773M38 2404103 00000011843M30 2404103 00000002779M37 2404103 00000005154M30 2404103 00000009402M39 2404103 00000005650M37 2404103 00000992390WK5 2404103 00000005653M31 2404103 00000002978M35 2404103 00001001073WK4 2404103 00000013129M32 2404103 00000020626M38 2404103 00000995872WK9 2404103 00000008138M30 2404103 00000013664M38 2404103 00000014589M36 2404103 00000011194M38 2404103 00000022652M32 2404103 00000013494M30 2404103 00000014275M32 2404103 00000001427M35 2404103 00000009122M33 2404103 00001006188WK5 2404103 00000004530M32 2404103 00000017485M34 2404103 00000017783M32 2404103 00000997114WK4 2404103 00000024278M34 2404103 00000005418M39 2404103 00000023220M37 2404103 00001003072WK4 2404103 00000025507M35 2404103 00000013274M36 2404103 00000020788M36 2404103 00000013830M35 2404103 00001005389WK0 2404103 00000100539WK2 2404103 00000020302M36 2404103 00000006329M37 2404103 00000011756M34 2404103 00000017941M36 2404103 00000003194M38 2404103 00000003826M35 2404103 00000026412M37 2404103 00000011725M39 2404103 00001005334WK6 2404103 00000022650M36 2404103 00000008600M39 2404103 00001004455WK0 2404103 00000997659WK8 2404103 00000006555M37 2404103 00000010080M30 2404103 00000008041M36 2404103 00000005496M35 2404103 00000005945M31 2404103 00001004960WK9 2404103 00000005092M32 2404103 00000009872M33 2404103 00000010377M30 2404103 00000012919M37 2404103 00000014476M36 2404103 00000022407M31 2404103 00000005419M37 2404103 00000008473M31 2404103 00000999058WK1 2404103 00000013014M36 2404103 00000005647M33 2404103 00000005110M32 2404103 00000017013M34 2404103 00000003461M31 2404103 00000005940M32 2404103 00000007941M38 2404103 00000004532M38 2404103 00000010811M38 2404103 00001005386WK6 2404103 00000991634WK7 2404103 00000006252M31 2404103 00000010216M30 2404103 00000005272M30 2404103 00000010460M34 2404103 00000992219WK6 2404103 00000991644WK6 2404103 00000022906M32 2404103 00000996159WK0 2404103 00000006504M35 2404103 00000998428WK7 2404103 00000012606M30 2404103 00000025725M33 2404103 00001006201WK6 2404103 00000001889M35 2404103 00000008660M33 2404103 00000001445M36 2404103 00000996479WK2 2404103 00000990542WK3 2404103 00000011195M35 2404103 00000999052WK4 2404103 00000993783WK0 2404103 00000991649WK5 2404103 00001005380WK9 2404103 00000997873WK5 2404103 00001006202WK4 2404103 00000004240M38 2404103 00000003980M30 2404103 00000007503M36 2404103 00000009008M34 2404103 00000010285M35 2404103 00000020301M38 2404103 00000008040M38 2404103 00000005427M30 2404103 00000005337M31 2404103 00000005210M30 2404103 00000006693M36 2404103 00000003457M39 2404103 00000003058M35 2404103 00000007242M31 2404103 00000998431WK1 2404103 00000992834WK2 2404103 00000994417WK4 2404103 00000994423WK2 2404103 00001001055WK1 2404103 00000013596M32 2404103 00000024798M31 2404103 00000010218M36 2404103 00000004229M31 2404103 00001004239WK8 2404103 00001006203WK2 2404103 00000999056WK5 2404103 00000994415WK8 2404103 00000005203M35 2404103 00001003376WK9 2404103 00000999919WK4 2404103 00000997639WK0 2404103 00000994421WK6 2404103 00000990533WK2 2404103 00000001434M30 2404103 00000002356M34 2404103 00000993352WK4 2404103 00000998427WK9 2404103 00001004958WK3 2404103 00000003052M38 2404103 00000002983M35 2404103 00000990535WK7 2404103 00000002123M38 2404103 00000002355M36 2404103 00000010736M37 2404103 00000014034M33 2404103 00000013128M34 2404103 00000002467M39 2404103 00000003977M36 2404103 00000004236M36 2404103 00000008280M30 2404103 00000011755M36 2404103 00001003374WK4 2404103 00000011636M38 2404103 00001004952WK6 2404103 00000002984M33 2404103 00000010126M31 2404103 00001004237WK2 2404103 00000008859M31 2404103 00000001436M35 2404103 00000011724M32 2404103 00000010627M38 2404103 00000011734M31 2404103 00000010807M36 2404103 00000011723M34 2404103 00000003825M37 2404103 00000011416M35 2404103 00000995346WK4 2404103 00000012478M34 2404103 00000008311M33 2404103 00000010737M35 2404103 00000012045M31 2404103 00000010280M36 2404103 00000023608M33 2404103 00000002122M30 2404103 00000011817M34 2404103 00000995878WK6 2404103 00001002441WK2 2404103 00000005268M38 2404103 00000008135M36 2404103 00000010985M30 2404103 00000007194M34 2404103 00000992836WK7 2404103 00000014567M32 2404103 00000997668WK9 2404103 00000005494M30 2404103 00000014121M38 2404103 00000013471M38 2404103 00001004967WK4 2404103 00001004652WK2 2404103 00000001429M30 2404103 00000004237M34 2404103 00000014706M36 2404103 00000013513M37 2404103 00000002121M32 2404103 00000008006M39 2404103 00000002636M39 2404103 00000003197M31 2404103 00000004969M32 2404103 00000005648M31 2404103 00001000997WK5 2404103 00000004693M38 2404103 00001004452WK7 2404103 00000001439M39 2404103 00000004646M36 2404103 00000990548WK0 2404103 00000007880M38 2404103 00000006256M32 2404103 00000013175M35 2404103 00000010282M32 2404103 00000006503M37 2404103 00000993788WK9 2404103 00000002979M33 2404103 00000008310M35 2404103 00000996155WK8 2404103 00000011484M33 2404103 00000991655WK2 2404103 00000010934M38 2404103 00001004649WK8 2404103 00000004531M30 2404103 00000002577M35 2404103 00000002576M37 2404103 00000001655M30 2404103 00000004183M30 2404103 00000002777M31 2404103 00000010081M38 2404103 00000003054M34 2404103 00000002985M30 2404103 00001000382WK0 2404103 00000021611M39 2404103 00000006860M31 2404103 00000009179M33 2404103 00001001070WK0 2404103 00001007261WK9 2404103 00001005377WK5 2404103 00000993786WK3 2404103 00000994424WK0 2404103 00000006150M37 2404103 00000006336M32 2404103 00000995340WK7 2404103 00000014024M34 2404103 00000012810M38 2404103 00001000992WK6 2404103 00000996765WK4 2404103 00000002982M37 2404103 00000023495M35 2404103 00000004527M38 2404103 00000009142M31 2404103 00001002919WK7 2404103 00000993344WK1 2404103 00000002642M37 2404103 00001002917WK1 2404103 00000006858M35 2404103 00000007774M33 2404103 00000010624M35 2404103 00000003460M33 2404103 00000992391WK3 2404103 00000005646M35 2404103 00000007658M38 2404103 00000991651WK1 2404103 00000005497M33 2404103 00000008661M31 2404103 00000004647M34 2404103 00001006192WK7 2404103 00000996157WK4 2404103 00000012250M37 2404103 00001001060WK1 2404103 00000990540WK7 2404103 00000991650WK3 2404103 00000991643WK8 2404103 00000013013M38 2404103 00000009276M37 2404103 28575M39 2404103 00001002436WK2 2404103 00001006190WK1 2404103 00000026025M37 2404103 00000005204M33 2404103 00001005379WK1 2404103 00001003379WK3 2404103 00000010284M38 2404103 00000017016M37 2404103 00000990416WK0 2404103 00001005390WK8 2404103 00000012923M39 2404103 00000011789M35 2404103 00000007704M30 2404103 00001002913WK0 2404103 00000008659M35 2404103 00000005207M36 2404103 00000024494M37 2404103 00000012143M34 2404103 00000007447M36 2404103 00000012714M32 2404103 00000021355M33 2404103 00000001890M33 2404103 00000023678M36 2404103 00000004719M31 2404103 00000010492M37 2404103 00000024194M33 2404103 00000020467M37 2404103 00001003367WK8 2404103 00000025566M31 2404103 00000996486WK7 2404103 00000008861M37 2404103 00000026936M35 2404103 00000990280WK0 2404103 00000021715M38 2404103 00000021571M35 2404103 00000020668M30 2404103 00000001430M38 2404103 00000012360M34 2404103 00000012184M38 2404103 00000019344M31 2404103 00000014081M34 2404103 00000007502M38 2404103 00000017784M30 2404103 00000018376M34 2404103 00000012772M30 2404103 00000009404M35 2404103 00000017507M35 2404103 00001003381WK9 2404103 00001003375WK1 2404103 00000022746M32 2404103 00000007239M37 2404103 00000007882M34 2404103 00001005385WK8 2404103 00000005814M39 2404103 00000003579M30 2404103 00000994420WK8 2404103 00000007446M38 2404103 00001004962WK5 2404103 00000998436WK0 2404103 00001006924WK3 2404103 00000011243M33 2404103 00000011486M38 2404103 00001006191WK9 2404103 00000003577M34 2404103 00000006322M32 2404103 00000007449M32 2404103 00001001861WK2 2404103 00000005655M36 2404103 00000012717M35 2404103 00000009949M39 2404103 00001000387WK9 2404103 00001000810WK0 2404103 00000997870WK1 2404103 00000007945M39 2404103 00000008741M31 2404103 00000001966M31 2404103 00000997100WK3 2404103 00000008848M34 2404103 00000001836M36 2404103 00000024495M34 2404103 00000012165M37 2404103 00000994434WK9 2404103 00000991629WK7 2404103 00000991652WK9 2404103 00000999922WK8 2404103 00000013944M34 2404103 00000005034M34 2404103 00000001432M34 2404103 00001003371WK0 2404103 00000004242M34 2404103 00000008038M32 2404103 00000997666WK3 2404103 00000008602M35 2404103 00000013515M32 2404103 00000006617M35 2404103 00000004857M39 2404103 00000006002M30 2404103 00000993351WK6 2404103 00000011878M36 2404103 00000006000M34 2404103 00000996485WK9 2404103 00000996341WK5 2404103 00000006093M39 2404103 00000004967M36 2404103 00000990543WK1 2404103 00000018055M34 2404103 00000003455M33 2404103 00000007386M36 2404103 00000994418WK2 2404103 00000002351M35 2404103 00000008560M35 2404103 00000007656M32 2404103 00000021224M31 2404103 00000991631WK3 2404103 00000013218M33 2404103 00000010283M30 2404103 00000011031M32 2404103 00000997108WK6 2404103 00000011835M36 2404103 00000999918WK6 2404103 00000004862M39 2404103 00000023038M33 2404103 00000016934M32 2404103 00000017257M37 2404103 00001003929WK5 2404103 00000004721M37 2404103 00000997111WK0 2404103 00000021354M36 2404103 00001006205WK7 2404103 00000004645M38 2404103 00000005811M35 2404103 00000991647WK9 2404103 00000023433M36 2404103 00000002354M39 2404103 00000999393WK2 2404103 00000009946M35 2404103 00000010078M34 2404103 00000019761M36 2404103 00000008308M39 2404103 00000018563M37 2404103 00000008136M34 2404103 00000010495M30 2404103 00000014007M39 2404103 00000010182M34 2404103 00000009874M39 2404103 00000013352M30 2404103 00000018111M35 2404103 00000008005M31 2404103 00000009711M33 2404103 00000998429WK5 2404103 00000021222M35 2404103 00000001428M32 2404103 00000023977M32 2404103 00000996676WK3 2404103 00001003937WK8 2404103 00000994132WK9 2404103 00000024664M35 2404103 00000003459M35 2404103 00000008469M39 2404103 00000003982M36 2404103 00000001222M39 2404103 00000002867M30 2404103 00000013873M35 2404103 00000026260M30 2404103 00000026260M30 2404103 00000007106M38 2404103 00001003378WK5 2404103 00000021829M37 2404103 00000006253M39 2404103 00000003458M37 2404103 00000009403M37 2404103 00001005387WK4 2404103 00000024604M31 2404103 00000023396M35 2404103 00000008199M32 2404103 00000002635M31 2404103 00000997871WK9 2404103 00000002126M31 2404103 00000004241M36 2404103 00000990412WK9 2404103 00000005338M39 2404103 00000011944M36 2404103 00000010735M39 2404103 00001007260WK1 2404103 00000014478M32 2404103 00000008559M37 2404103 00000996160WK8 2404103 00000996760WK5 2404103 00001001062WK7 2404103 00000001964M36 2404103 00000994131WK1 2404103 00000007195M31 2404103 00000007192M38 2404103 00001003364WK5 2404103 00000990545WK6 2404103 00000997101WK1 2404103 00000002870M34 2404103 00000005340M35 2404103 00000004235M38 2404103 00000011657M34 2404103 00000013749M37 2404103 00000013302M35 2404103 00000026216M32 2404103 00000993780WK6 2404103 00000010497M36 2404103 00000008008M35 2404103 00000999060WK7 2404103 00000010181M36 2404103 00000993781WK4 2404103 00000014803M31 2404103 00000007193M36 2404103 00000002578M33 2404103 00000004714M32 2404103 00000025475M35 2404103 00000008471M35 2404103 00000010378M38 2404103 00000002350M37 2404103 00001003365WK2 2404103 00000009182M37 2404103 00001002915WK5 2404103 00001002185WK5 2404103 00000011842M32 2404103 00001004246WK3 2404103 00000996481WK8 2404103 00000990539WK9 2404103 00000012359M36 2404103 00000992395WK4 2404103 00000005036M39 2404103 00000992840WK9 2404103 00001005392WK4 2404103 00000991657WK8 2404103 00000013217M35 2404103 00000003828M31 2404103 00000012298M36 2404103 00000007179M35 2404103 00001000812WK6 2404103 00001006193WK5 2404103 00001000380WK4 2404103 00000024967M32 2404103 00000995345WK6 2404103 00000013945M31 2404103 00000014502M39 2404103 00000999391WK6 2404103 00001000389WK5 2404103 00000005495M37 2404103 00000990407WK9 2404103 00000006001M32 2404103 00000993161WK9 2404103 00000014477M34 2404103 00000008775M39 2404103 00000013127M36 2404103 00000007786M37 2404103 00000004230M39 2404103 00000994419WK0 2404103 00000024556M33 2404103 00001004241WK4 2404103 00000005808M31 2404103 00000012023M38 2404103 00000999055WK7 2404103 00000007448M34 2404103 00000012296M30 2404103 00000991654WK5 2404103 00000990541WK5 2404103 00000007387M34 2404103 00000009058M39 2404103 00000004722M35 2404103 00000008772M36 2404103 00000006337M30 2404103 00000011245M38 2404103 00000992842WK5 2404103 00000992393WK9 2404103 00001004245WK5 2404103 00000024497M30 2404103 00000022929M34 2404103 00000008300M36 2404103 00000994137WK8 2404103 00000008409M35 2404103 00000006619M31 2404103 00000010217M38 2404103 00000006096M32 2404103 00000006097M30 2404103 00000005093M30 2404103 00000004720M39 2404103 00000012024M36 2404103 00000004965M30 2404103 00000021353M38 2404103 00000005420M35 2404103 00000002980M31 2404103 00000002872M30 2404103 00000008410M33 2404103 00000011295M33 2404103 00000008470M37 2404103 00000025844M32 2404103 00000013215M39 2404103 00000008739M35 2404103 00000004970M30 2404103 00000005153M32 2404103 00000004528M36 2404103 00000005812M33 2404103 00000009533M31 2404103 00000001437M33 2404103 00000003829M39 2404103 00000006698M35 2404103 00001000384WK6 2404103 00000017645M33 2404103 00000014122M36 2404103 00000997868WK5 2404103 00000009873M31 2404103 00000002352M33 2404103 00000010079M32 2404103 00000009822M38 2404103 00000013216M37 2404103 00000013305M38 2404103 00001004966WK6 2404103 00000005206M38 2404103 00000024020M30 2404103 00000010379M36 2404103 00001004953WK4 2404103 00001003377WK7 2404103 00001000811WK8 2404103 00001003385WK0 2404103 00000008773M34 2404103 00000001661M38 2404103 00000008601M37 2404103 00000011244M31 2404103 00000008738M37 2404103 00000013597M30 2404103 00000012248M31 2404103 00000007657M30 2404103 00000018056M32 2404103 00000010986M38 2404103 00000005649M39 2404103 00000012548M34 2404103 00000003195M35 2404103 00000990278WK4 2404103 00001004457WK6 2404103 00000991635WK4 2404103 00000997661WK4 2404103 00000999395WK7 2404103 00000003198M39 2404103 00000991633WK9 2404103 00000990546WK4 2404103 00000001963M38 2404103 00000008857M35 2404103 00000014498M30 2404103 00001000994WK2 2404103 00001004456WK8 2404103 00000009274M32 2404103 00001000999WK1 2404103 00000994427WK3 2404103 00000998437WK8 2404103 00000013351M32 2404103 00000997872WK7 2404103 00000007703M32 2404103 00000006561M35 2404103 00000005205M30 2404103 00000012108M37 2404103 00000005999M38 2404103 00000008965M36 2404103 00001004968WK2 2404103 00000008963M31 2404103 00000994422WK4 2404103 00000003201M31 2404103 00000023221M35 2404103 00000023607M35 2404103 00000007827M39 2404103 00000012659M39 2404103 00000025568M37 2404103 00000001886M31 2404103 00001000998WK3 2404103 00000007383M33 2404103 00000001839M30 2404103 00000008412M39 2404103 00000006151M35 2404103 00000023600M30 2404103 00000990277WK6 2404103 00001004459WK2 2404103 00000006556M35 2404103 00000011433M30 2404103 00000002866M32 2404103 00000004966M38 2404103 00000008195M30 2404103 00000005339M37 2404103 00000013598M38 2404103 00000999913WK7 2404103 00000018236M30 2404103 00000005208M34 2404103 00000011820M38 2404103 00000001438M31 2404103 00000002118M38 2404103 00000991646WK1 2404103 00000014500M33 2404103 00001003383WK5 2404103 00000008411M31 2404103 00000012867M38 2404103 00000006334M37 2404103 00000012658M31 2404103 00000003462M39 2404103 00000025395M35 2404103 00000010493M35 2404103 00000997664WK8 2404103 00000998426WK1 2404103 00000992843WK3 2404103 00000999390WK8 2404103 00000023609M31 2404103 00000008964M39 2404103 00000008563M39 2404103 00000001442M33 2404103 00000002637M37 2404103 00000009278M33 2404103 00000992397WK0 2404103 00000005267M30 2404103 00000999394WK0 2404103 00000013219M31 2404103 00000013192M30 2404103 00000007940M30 2404103 00000008281M38 2404103 00000022956M37 2404103 00000006691M30 2404103 00000010459M36 2404103 00000017944M30 2404103 00000008039M30 2404103 00000003979M32 2404103 00000006330M35 2404103 00000999921WK0 2404103 00000009672M37 2404103 00000002977M37 2404103 00000009945M37 2404103 00000997866WK9 2404103 00001002433WK9 2404103 00000006255M34 2404103 00000996154WK1 2404103 00000994134WK5 2404103 00000993789WK7 2404103 00000990411WK1 2404103 00000013901M34 2404103 00000019556M30 2404103 00000006333M39 2404103 00001000385WK3 2404103 00000004245M37 2404103 00001004238WK0 2404103 00000003830M37 2404103 00000993169WK2 2404103 00000001885M33 2404103 00001002916WK3 2404103 00000018535M35 2404103 00000022649M38 2404103 00000009007M36 2404103 00000997864WK4 2404103 00000012333M31 2404103 00000007241M33 2404103 00000013848M37 2404103 00001003930WK3 2404103 00000002119M36 2404103 00001001859WK6 2404103 00000006310M37 2404103 00000006862M37 2404103 00000996675WK5 2404103 00000011570M39 2404103 00000004860M33 2404103 00000011468M36 2404103 00000012107M39 2404103 00000002579M31 2404103 00001001061WK9 2404103 00000024662M39 2404103 00000998430WK3 2404103 00001000996WK7 2404103 00000005113M36 2404103 00001001864WK6 2404103 00000005035M31 2404103 00000992398WK8 2404103 00000010287M31 2404103 00000005422M31 2404103 00000013174M38 2404103 00001004242WK2 2404103 00001003366WK0 2404103 00000013662M32 2404103 00000006506M30 2404103 00000999916WK0 2404103 00000010496M38 2404103 00000998424WK6 2404103 00000009009M32 2404103 00000012535M31 2404103 00000024862M35 2404103 00000025056M33 2404103 00000024067M31 2404103 00000024966M34 2404103 00000999053WK2 2404103 00000025939M30 2404103 00000997107WK8 2404103 00000012641M37 2404103 00000001652M37 2404103 00000994416WK6 2404103 00000005997M32 2404103 00000006857M37 2404103 00000005155M37 2404103 00000006557M33 2404103 00000001435M37 2404103 00000004182M32 2404103 00000005094M38 2404103 00000005809M39 2404103 00000006505M32 2404103 00000007828M37 2404103 00000990409WK5 2404103 00000992838WK3 2404103 00001005396WK5 2404103 00001002438WK8 2404103 00001006926WK8 2404103 00001004650WK6 2404103 00000013304M31 2404103 00000019487M38 2404103 00000017918M34 2404103 00001000381WK2 2404103 00001005378WK3 2404103 00000006616M37 2404103 00000006690M32 2404103 00000997105WK2 2404103 00001001866WK1 2404103 00000020309M31 2404103 00000001660M30 2404103 00000002873M38 2404103 00000996766WK2 2404103 00001004450WK1 2404103 00000997112WK8 2404103 00000993345WK8 2404103 00000008137M32 2404103 00000993346WK6 2404103 00000022253M39 2404103 00000996759WK7 2404103 00000993782WK2 2404103 00000018806M30 2404103 00000025563M38 2404103 00000007655M34 2404103 00001003932WK9 2404103 00000005493M32 2404103 00000010623M37 2404103 00000991641WK2 2404103 00000003822M34 2404103 00000004717M35 2404103 00000004228M33 2404103 00000007450M30 2404103 00000004244M30 2404103 00000001659M32 2404103 00000024938M33 2404103 00000005488M32 2404103 00000016935M39 2404103 00000003978M34 2404103 00000003976M38 2404103 00000016714M38 2404103 00001003382WK7 2404103 00000010494M33 2404103 00000002463M38 2404103 00000003823M32 2404103 00000016074M37 2404103 00000020067M35 2404103 00000005810M37 2404103 00000993164WK3 2404103 00000007385M38 2404103 00000007451M38 2404103 00000007585M33 2404103 00000994135WK2 2404103 00000013663M30 2404103 00000002868M38 2404103 00000005490M38 2404103 00000012222M36 2404103 00000012642M35 2404103 00001023630WK5 2403791-3 00001023676WK8 2403791-3 00001022088WK7 2403791-3 00001023876WK4 2403791-3 00001023059WK7 2403791-3 00001018400WK0 2403791-3 00001021225WK6 2403791-3 00001023983WK8 2403791-3 00001012891WK6 2403791-3 00001017448WK0 2403791-3 00001023632WK1 2403791-3 00001021880WK8 2403791-3 00001022261WK0 2403791-3 00001014111WK7 2403791-3 00001021884WK0 2403791-3 00001011861WK0 2403791-3 00001020193WK7 2403791-3 00001020933WK6 2403791-3 00001020694WK4 2403791-3 00001011182WK1 2403791-3 MISSING_SN_GON2505184 2403791-3 00001020999WK7 2403791-3 00001023727WK9 2403791-3 00001023878WK0 2403791-3 00001012895WK7 2403791-3 00001019945WK3 2403791-3 00001010774WK6 2403791-3 00001015275WK9 2403791-3 00001014968WK0 2403791-3 00001010224WK2 2403791-3 00001030301WK4 2403791-32 00001028159WK0 2403791-32 00001030983WK9 2403791-32 00001029838WK8 2403791-32 00001024742WK7 2403791-32 00001027087WK4 2403791-32 00001027086WK6 2403791-32 1034387WK9 2403791-32 00001027082WK5 2403791-32 00001026242WK6 2403791-32 00001027866WK1 2403791-32 00001026639WK3 2403791-32 00001031628WK9 2403791-32 00001030700WK7 2403791-32 00001025253WK4 2403791-32 00001025013WK2 2403791-32 00001027864WK6 2403791-32 00001029227WK4 2403791-32 00001031257WK7 2403791-32 00001030119WK0 2403791-32 00001032073WK7 2403791-32 00001026026WK3 2403791-32 00001030898WK9 2403791-32 00001024517WK3 2403791-32 00001028822WK3 2403791-32 00001032853WK2 2403791-32 00001027426WK4 2403791-32 00001027425WK6 2403791-32 1035500WK6 2403791-32 00001028162WK4 2403791-32 1033581WK8 2403791-32 1034356WK4 2403791-32 1035225WK3 2403791-32 00001028425WK5 2403791-32 00001034486WK9 2403791-32 MISSING_SN_GON4204229 2403791-32 00001030582WK9 2403791-32 00001028424WK8 2403791-32 00001029717WK4 2403791-32 00001029427WK0 2403791-32 00001031577WK8 2403791-32 00001034485WK1 2403791-32 MISSING_SN_GON 4168970 2403791-32 00001031633WK9 2403791-32 00001024889WK6 2403791-32 00001024874WK8 2403791-32 00001024518WK1 2403791-32 1035411WK6 2403791-32 00001025366WK4 2403791-32 00001025806WK9 2403791-32 00001027080WK9 2403791-32 00001027063WK5 2403791-32 00001034945WK4 2403791-32 00001029140WK9 2403791-32 00001024725WK2 2403791-32 MISSING_SN_GON4222007 2403791-32 00001029007WK0 2403791-32 00001029136WK7 2403791-32 00001029430WK4 2403791-32 00001030458WK2 2403791-32 00001029714WK1 2403791-32 00001030982WK1 2403791-32 00001032200WK6 2403791-32 00001026240WK0 2403791-32 00001024576WK9 2403791-32 00001025012WK4 2403791-32 00001027085WK8 2403791-32 00001027084WK1 2403791-32 1035903WK2 2403791-32 00001035903WK2 2403791-32 MISSING_4205856 2403791-32 00001033967WK9 2403791-32 00001032856WK5 2403791-32 00001027594WK9 2403791-32 MISSING_SN_GON4200472 2403791-32 00001031258WK5 2403791-32 00001027591WK5 2403791-32 MISSING_SN_GON4191638 2403791-32 00001027868WK7 2403791-32 00001030121WK6 2403791-32 00001029713WK3 2403791-32 00001025004WK1 2403791-32 00001029425WK4 2403791-32 00001027589WK9 2403791-32 MISSING_SN_GON4236297 2403791-32 00001027590WK7 2403791-32 00001026637WK7 2403791-32 00001024743WK5 2403791-32 00001027482WK7 2403791-32 000000102558K2 2403791-32 00001025583WK4 2403791-32 00001024741WK9 2403791-32 00001030583WK7 2403791-32 00001027483WK5 2403791-32 00001025585WK9 2403791-32 00001030697WK5 2403791-32 1035409WK0 2403791-32 00001031848WK3 2403791-32 00001026636WK9 2403791-32 00001029139WK1 2403791-32 00001030123WK2 2403791-32 00001027592WK3 2403791-32 00001030299WK0 2403791-32 00001028426WK3 2403791-32 00001032072WK9 2403791-32 00001033791WK3 2403791-32 00001030701WK5 2403791-32 00001032085WK1 2403791-32 00001029006WK2 2403791-32 00001029432WK0 2403791-32 MISSING_SN_GON4094900 2403791-32 00001032084WK4 2403791-32 00001026972WK8 2403791-32 00001028824WK9 2403791-32 00001031846WK7 2403791-32 00001029426WK2 2403791-32 00001028422WK2 2403791-32 00001030584WK5 2403791-32 00001027588WK1 2403791-32 1033968WK7 2403791-32 00001028158WK2 2403791-32 MISSING_SN_GON4162806 2403791-32 00001028163WK2 2403791-32 00001029135WK9 2403791-32 00001030300WK6 2403791-32 00001027867WK9 2403791-32 00001027079WK1 2403791-32 00001028423WK0 2403791-32 00001026705WK2 2403791-32 00001026971WK0 2403791-32 MISSING_SN_GON4226119 2403791-32 00001029794WK3 2403791-32 00001027424WK9 2403791-32 00001030581WK1 2403791-32 00001024726WK0 2403791-32 00001026465WK3 2403791-32 00001026638WK5 2403791-32 00001029138WK3 2403791-32 00001030122WK4 2403791-32 00001026970WK2 2403791-32 00001031444WK1 2403791-32 MISSING_SN_GON4163027 2403791-32 00001031629WK7 2403791-32 00001029839WK6 2403791-32 00001028635WK9 2403791-32 00001029716WK6 2403791-32 00001025367WK2 2403791-32 00001026241WK8 2403791-32 00001031630WK5 2403791-32 00001031442WK5 2403791-32 00001025254WK2 2403791-32 00001026973WK6 2403791-32 00001027593WK1 2403791-32 00001030462WK4 2403791-32 00001027484WK3 2403791-32 00001027081WK7 2403791-32 00001024888WK8 2403791-32 00001026706WK0 2403791-32 00001032523WK1 2403791-32 00001031443WK3 2403791-32 00001027863WK8 2403791-32 00001027083WK3 2403791-32 00001028633WK4 2403791-32 00001026248WK3 2403791-32 MISSING_SN_GON4192035 2403791-32 00001025002WK5 2403791-32 00001031635WK4 2403791-32 00001025011WK6 2403791-32 00001024578WK5 2403791-32 00001026704WK5 2403791-32 00001033159WK3 2403791-32 00001030461WK6 2403791-32 00001031090WK2 2403791-32 00001030120WK8 2403791-32 00001032740WK1 2403791-32 1034602WK1 2403791-32 00001034602WK1 2403791-32 00001030981WK3 2403791-32 00001032855WK7 2403791-32 00001029228WK2 2403791-32 00001029226WK6 2403791-32 00001032653WK6 2403791-32 00001030580WK3 2403791-32 MISSING_SN_GON4189208 2403791-32 00001031092WK8 2403791-32 00001032737WK7 2403791-32 00001030460WK8 2403791-32 00001031847WK5 2403791-32 00001029005WK4 2403791-32 00001034603WK9 2403791-32 00001033792WK1 2403791-32 00001033330WK0 2403791-32 00001031441WK7 2403791-32 00001031091WK0 2403791-32 00001034104WK8 2403791-32 00001034036WK2 2403791-32 1033793WK9 2403791-32 00001031845WK9 2403791-32 00001029837WK0 2403791-32 00001033281WK5 2403791-32 00001025256WK7 2403791-32 00001032526WK4 2403791-32 00001032738WK5 2403791-32 MISSING_4197442 2403791-32 00001024577WK7 2403791-32 00001030459WK0 2403791-32 00001025591WK7 2403791-32 00001026424WK0 2403791-32 00001029429WK6 2403791-32 00001029137WK5 2403791-32 00001026243WK4 2403791-32 00001025255WK9 2403791-32 00001034389WK5 2403791-32 00001028823WK1 2403791-32 00001027862WK0 2403791-32 00001026466WK1 2403791-32 00001025586WK7 2403791-32 00001027485WK0 2403791-32 00001031632WK1 2403791-32 00001024744WK3 2403791-32 00001035905WK7 2403791-32 MISSING_SN_GON4163469 2403791-32 00001027587WK3 2403791-32 00001033969WK5 2403791-32 00001025587WK5 2403791-32 00001026426WK5 2403791-32 00001027865WK3 2403791-32 00001031637WK0 2403791-32 00001030698WK3 2403791-32 00001032524WK9 2403791-32 00001025261WK7 2403791-40 00001026147WK7 2403791-40 00001025574WK3 2403791-40 00001029710WK9 2403791-40 00001029831WK3 2403791-40 00001027857WK0 2403791-40 00001026858WK9 2403791-40 00001029437WK9 2403791-40 00001032518WK1 2403791-40 00001029560WK8 2403791-40 00001029562WK4 2403791-40 00001030707WK2 2403791-40 00001026976WK9 2403791-40 00001027431WK4 2403791-40 00001026861WK3 2403791-40 00001033578WK4 2403791-40 00001024676WK7 2403791-40 00001025372WK2 2403791-40 00001033575WK0 2403791-40 00001032521WK5 2403791-40 00001028637WK5 2403791-40 00001025010WK8 2403791-40 00001025049WK6 2403791-40 00001028602WK9 2403791-40 00001027858WK8 2403791-40 00001026260WK8 2403791-40 00001024573WK6 2403791-40 00001032959WK7 2403791-40 00001029711WK7 2403791-40 00001030463WK2 2403791-40 00001029440WK3 2403791-40 00001031314WK6 2403791-40 00001027622WK8 2403791-40 00001034874WK6 2403791-40 00001032062WK0 2403791-40 00001030188WK5 2403791-40 00001031985WK3 2403791-40 MISSING_SN_GON4053108 2403791-40 00001025579WK2 2403791-40 00001030465WK7 2403791-40 00001026980WK1 2403791-40 00001034582WK5 2403791-40 00001024519WK9 2403791-40 00001025888WK7 2403791-40 00001026467WK9 2403791-40 00001027873WK7 2403791-40 00001025507WK3 2403791-40 00001025009WK0 2403791-40 00001031445WK8 2403791-40 00001035246WK6 2403791-40 1035103WK9 2403791-40 00001034875WK3 2403791-40 00001029439WK5 2403791-40 00001030310WK5 2403791-40 00001025373WK0 2403791-40 00001025511WK5 2403791-40 00001025879WK6 2403791-40 00001025879WK6 2403791-40 1035012WK2 2403791-40 00001035014WK8 2403791-40 00001033537WK0 2403791-40 00001029836WK2 2403791-40 00001026643WK5 2403791-40 00001030126WK5 2403791-40 00001031638WK8 2403791-40 00001030709WK8 2403791-40 00001026860WK5 2403791-40 00001030306WK3 2403791-40 00001033339WK1 2403791-40 00001026255WK8 2403791-40 00001026256WK6 2403791-40 00001028911WK4 2403791-40 00001029833WK9 2403791-40 00001027477WK7 2403791-40 00001026712WK8 2403791-40 00001030988WK8 2403791-40 00001031312WK0 2403791-40 00001029234WK0 2403791-40 00001026975WK1 2403791-40 00001025260WK9 2403791-40 00001030710WK6 2403791-40 00001029148WK2 2403791-40 00001027620WK2 2403791-40 00001027875WK2 2403791-40 00001028821WK5 2403791-40 00001026899WK3 2403791-40 00001031835WK0 2403791-40 00001026252WK5 2403791-40 00001032067WK9 2403791-40 00001025054WK6 2403791-40 00001030577WK9 2403791-40 00001030704WK9 2403791-40 1033577WK6 2403791-40 00001031839WK2 2403791-40 00001027616WK0 2403791-40 1034250WK9 2403791-40 1034393WK7 2403791-40 MISSING_SN_GON4230872 2403791-40 00001029832WK1 2403791-40 1034394WK5 2403791-40 00001029712WK5 2403791-40 00001030466WK5 2403791-40 00001028431WK3 2403791-40 00001025878WK8 2403791-40 00001025375WK5 2403791-40 00001025887WK9 2403791-40 00001025886WK1 2403791-40 00001031837WK6 2403791-40 00001024575WK1 2403791-40 00001025258WK3 2403791-40 00001025508WK1 2403791-40 00001026031WK3 2403791-40 00001026264WK0 2403791-40 1034359WK8 2403791-40 00001032068WK7 2403791-40 00001029001WK3 2403791-40 00001029438WK7 2403791-40 00001028432WK1 2403791-40 00001025006WK6 2403791-40 MISSING_SN_GON4225310 2403791-40 00001026786WK2 2403791-40 00001025008WK2 2403791-40 00001034585WK8 2403791-40 00001025050WK4 2403791-40 00001030309WK7 2403791-40 00001028999WK9 2403791-40 00001034584WK1 2403791-40 00001029147WK4 2403791-40 00001033335WK9 2403791-40 00001034588WK2 2403791-40 00001030187WK7 2403791-40 00001034871WK2 2403791-40 00001027090WK8 2403791-40 1035710WK1 2403791-40 00001026473WK7 2403791-40 00001030705WK6 2403791-40 00001031643WK8 2403791-40 00001034397WK8 2403791-40 00001026148WK5 2403791-40 00001027619WK4 2403791-40 00001031836WK8 2403791-40 00001031196WK7 2403791-40 00001034396WK0 2403791-40 00001025988WK5 2403791-40 00001029433WK8 2403791-40 00001030711WK4 2403791-40 00001029002WK1 2403791-40 00001028817WK3 2403791-40 00001030308WK9 2403791-40 00001028167WK3 2403791-40 00001029232WK4 2403791-40 00001028169WK9 2403791-40 00001034491WK9 2403791-40 00001033795WK4 2403791-40 00001028818WK1 2403791-40 00001028819WK9 2403791-40 00001030127WK3 2403791-40 00001027093WK2 2403791-40 00001025005WK8 2403791-40 00001024746WK8 2403791-40 00001025509WK9 2403791-40 00001026471WK1 2403791-40 00001028636WK7 2403791-40 00001029319WK9 2403791-40 00001025051WK2 2403791-40 00001029233WK2 2403791-40 00001030891WK4 2403791-40 00001026859WK7 2403791-40 00001026787WK0 2403791-40 00001031096WK9 2403791-40 00001031094WK4 2403791-40 00001029829WK7 2403791-40 00001029830WK5 2403791-40 00001028436WK2 2403791-40 00001034586WK6 2403791-40 00001029559WK0 2403791-40 00001027618WK6 2403791-40 00001027872WK9 2403791-40 00001029708WK3 2403791-40 00001032957WK1 2403791-40 00001026713WK6 2403791-40 00001024572WK8 2403791-40 00001025371WK4 2403791-40 00001028171WK5 2403791-40 00001026640WK1 2403791-40 00001024724WK5 2403791-40 00001028639WK1 2403791-40 00001027855WK4 2403791-40 1034112WK1 2403791-40 00001028437WK0 2403791-40 00001032522WK3 2403791-40 00001033510WK7 2403791-40 00001028165WK7 2403791-40 00001032655WK1 2403791-40 00001024520WK7 2403791-40 00001026468WK7 2403791-40 00001030579WK5 2403791-40 00001031838WK4 2403791-40 00001027096WK5 2403791-40 00001028434WK7 2403791-40 1033511WK5 2403791-40 00001032943WK1 2403791-40 00001025575WK0 2403791-40 00001026709WK4 2403791-40 00001031260WK1 2403791-40 00001030985WK4 2403791-40 00001028601WK1 2403791-40 00001024745WK0 2403791-40 00001034918WK1 2403791-40 00001027095WK7 2403791-40 00001031313WK8 2403791-40 00001035404WK1 2403791-40 00001024878WK9 2403791-40 00001024881WK3 2403791-40 00001025052WK0 2403791-40 00001032942WK3 2403791-40 00001029320WK7 2403791-40 00001032065WK3 2403791-40 00001034490WK1 2403791-40 00001031095WK1 2403791-40 00001026257WK4 2403791-40 00001027237WK5 2403791-40 00001027428WK0 2403791-40 00001028166WK5 2403791-40 00001033794WK7 2403791-40 00001032733WK6 2403791-40 00001027874WK5 2403791-40 00001034917WK3 2403791-40 MISSING_SN_GON4218515 2403791-40 00001026254WK1 2403791-40 00001024574WK4 2403791-40 00001027617WK8 2403791-40 00001028170WK7 2403791-40 00001028172WK3 2403791-40 00001030128WK1 2403791-40 MISSING_SN_GON4233885 2403791-40 00001025374WK8 2403791-40 00001025978WK6 2403791-40 00001027854WK7 2403791-40 00001031197WK5 2403791-40 00001026469WK5 2403791-40 00001026027WK1 2403791-40 00001026146WK9 2403791-40 00001025007WK4 2403791-40 00001026978WK5 2403791-40 1035406WK6 2403791-40 00001027089WK0 2403791-40 MISSING_SN_GON4223403 2403791-40 00001026788WK8 2403791-40 00001031093WK6 2403791-40 00001032956WK3 2403791-40 00001029435WK3 2403791-40 00000102474WK4 2403791-40 00001028430WK5 2403791-40 00001033796WK2 2403791-40 00001033971WK1 2403791-40 00001025582WK6 2403791-40 00001026641WK9 2403791-40 00001026900WK9 2403791-40 00001026470WK3 2403791-40 00001027094WK0 2403791-40 MISSING_SN_GON4221505 2403791-40 00001027098WK1 2403791-40 00001033574WK3 2403791-40 00001028638WK3 2403791-40 00001034395WK2 2403791-40 00001034392WK9 2403791-40 MISSING_SN_GON4217156 2403791-40 00001031844WK2 2403791-40 00001030467WK3 2403791-40 00001027092WK4 2403791-40 00001027091WK6 2403791-40 00001027876WK0 2403791-40 1029561WK6 2403791-40 00001035758WK0 2403791-40 00001028083WK2 2403791-40 00001029000WK5 2403791-40 00000510363WK5 46-262751G6 00000504288WK2 46-262751G6 00000508070WK0 46-262751G6 00000512135WK5 46-262751G6 00000510900WK4 46-262751G6 00000508977WK6 46-262751G6 00000499668WK2 46-262751G6 00000518141WK7 46-262751G6 00000504046WK4 46-262751G6 00000493066WK5 46-262751G6 00000503355WK0 46-262751G6 00000507625WK2 46-262751G6 00000516439WK7 46-262751G6 00000518140WK9 46-262751G6 00000512458WK1 46-262751G6 00000515659WK1 46-262751G6 00000506731WK9 46-262751G6 00000518142WK5 46-262751G6 00000518884WK2 46-262751G6 00000514167WK6 46-262751G6 00000516878WK6 46-262751G6 00000501270WK3 46-262751G6 00000515664WK1 46-262751G6 00000512137WK1 46-262751G6 00000492902WK2 46-262751G6 00000510837WK8 46-262751G6 00000506727WK7 46-262751G6 00000502617WK4 46-262751G6 00000497794WK8 46-262751G6 00000512136WK3 46-262751G6 00000506734WK3 46-262751G6 00000506733WK0 46-262751G6 00000502618WK2 46-262751G6 00000499724WK3 46-262751G6 00000512627WK1 46-262751G6 00000501974WK0 46-262751G6 00000506504WK0 46-262751G6 00000506736WK8 46-262751G6 00000518885WK9 46-262751G6 00000510838WK6 46-262751G6 00000512626WK3 46-262751G6 00000504291WK6 46-262751G6 00000518883WK4 46-262751G6 00000499945WK4 46-262751G6 00000512138WK9 46-262751G6 00000506071WK0 46-262751G6 00000514169WK2 46-262751G6 00000501268WK7 46-262751G6 00000502619WK0 46-262751G6 00000508469WK4 46-262751G6 00000506732WK7 46-262751G6 00000508686WK3 46-262751G6 00000500385WK0 46-262751G6 00000509155WK8 46-262751G6 00000513394WK7 46-262751G6 00000515663WK3 46-262751G6 00000509154WK1 46-262751G6 00000509158WK2 46-262751G6 00000512455WK7 46-262751G6 00000512633WK9 46-262751G6 00000504871WK5 46-262751G6 00000510842WK8 46-262751G6 00000512454WK0 46-262751G6 00000510617WK4 46-262751G6 00000501849WK4 46-262751G6 00000504873WK1 46-262751G6 00000514172WK6 46-262751G6 00000506733WK5 46-262751G6 00000498817WK6 46-262751G6 00000514176WK7 46-262751G6 00000513396WK2 46-262751G6 00000504870WK7 46-262751G6 00000512457WK3 46-262751G6 00000504041WK5 46-262751G6 00000501272WK9 46-262751G6 00000502437WK7 46-262751G6 00000501080WK6 46-262751G6 00000504045WK6 46-262751G6 00000513395WK4 46-262751G6 00000516875WK2 46-262751G6 00000515661WK7 46-262751G6 00000515667WK4 46-262751G6 00000507626WK0 46-262751G6 00000525674WK8 46-262751G7 00000525026WK1 46-262751G7 00000525031WK1 46-262751G7 00000525030WK3 46-262751G7 00000531415WK8 46-262751G7 00000524481WK9 46-262751G7 00000527087WK1 46-262751G7 00000532966WK9 46-262751G7 00000517710WK0 46-262751G7 00000536551WK5 46-262751G7 00000517056WK8 46-262751G7 00000530159WK3 46-262751G7 00000530446WK4 46-262751G7 00000531588WK2 46-262751G7 00000517706WK8 46-262751G7 00000522090WK0 46-262751G7 00000531589WK0 46-262751G7 00000527091WK3 46-262751G7 00000525029WK5 46-262751G7 00000517707WK6 46-262751G7 00000523402WK6 46-262751G7 00000537111WK7 46-262751G7 00000537112WK5 46-262751G7 00000537849WK2 46-262751G7 00000525670WK6 46-262751G7 00000530842WK4 46-262751G7 00000519515WK1 46-262751G7 00000532971WK9 46-262751G7 00000529630WK6 46-262751G7 00000529192WK7 46-262751G7 00000524475WK1 46-262751G7 00000531358WK0 46-262751G7 00000531413WK3 46-262751G7 00000530161WK9 46-262751G7 00000532514WK7 46-262751G7 00000519909WK6 46-262751G7 00000517708WK4 46-262751G7 00000520793WK1 46-262751G7 00000525025WK3 46-262751G7 00000527085WK5 46-262751G7 00000530448WK0 46-262751G7 00000522605WK5 46-262751G7 00000532968WK5 46-262751G7 00000527093WK9 46-262751G7 00000524477WK7 46-262751G7 00000515275WK6 46-262751G7 00000522091WK8 46-262751G7 00000519905WK4 46-262751G7 00000519912WK0 46-262751G7 00000522612WK1 46-262751G7 00000531741WK7 46-262751G7 00000519907WK0 46-262751G7 00000520640WK4 46-262751G7 00000533549WK2 46-262751G7 00000532969WK3 46-262751G7 00000522094WK2 46-262751G7 00000522608WK9 46-262751G7 00000523398WK6 46-262751G7 00000521504WK1 46-262751G7 00000519517WK7 46-262751G7 00000517709WK2 46-262751G7 00000529942WK5 46-262751G7 00000524394WK4 46-262751G7 00000528266WK0 46-262751G7 00000533552WK6 46-262751G7 00000532516WK2 46-262751G7 00000521505WK8 46-262751G7 00000521498WK6 46-262751G7 00000529195WK0 46-262751G7 00000528267WK8 46-262751G7 00000528248WK8 46-262751G7 00000519519WK3 46-262751G7 00000528270WK2 46-262751G7 00000524472WK8 46-262751G7 00000530676WK6 46-262751G7 00000520643WK8 46-262751G7 00000537846WK8 46-262751G7 00000524395WK1 46-262751G7 00000530160WK1 46-262751G7 00000524476WK9 46-262751G7 00000521501WK7 46-262751G7 00000529629WK8 46-262751G7 00000521503WK3 46-262751G7 00000525668WK0 46-262751G7 00000530674WK1 46-262751G7 00000530675WK8 46-262751G7 00000525673WK0 46-262751G7 00000523403WK4 46-262751G7 00000532290WK4 46-262751G7 00000522603WK0 46-262751G7 00000519118WK4 46-262751G7 00000524479WK3 46-262751G7 00000531357WK2 46-262751G7 00000519903WK9 46-262751G7 00000530678WK2 46-262751G7 00000537847WK6 46-262751G7 00000532970WK1 46-262751G7 00000522606WK3 46-262751G7 00000531046WK1 46-262751G7 00000532515WK4 46-262751G7 00000524473WK6 46-262751G7 00000521502WK5 46-262751G7 00000522493WK6 46-262751G7 00000523401WK8 46-262751G7 00000519518WK5 46-262751G7 00000528269WK4 46-262751G7 00000537110WK9 46-262751G7 00000532293WK8 46-262751G7 00000537113WK3 46-262751G7 00000533548WK4 46-262751G7 00000525669WK8 46-262751G7 00000537108WK3 46-262751G7 00000521500WK9 46-262751G7 00000520644WK6 46-262751G7 00000530841WK6 46-262751G7 00000532517WK0 46-262751G7 00000527088WK9 46-262751G7 00000520641WK2 46-262751G7 00000522197WK3 46-262751G7 00000540271WK4 46-262751G8 00000540120WK3 46-262751G8 00000544933WK5 46-262751G8 00000550141WK6 46-262751G8 00000952934WK8 46-262751G8 00000542312WK4 46-262751G8 00000541728WK2 46-262751G8 00000543656WK3 46-262751G8 00000547748WK4 46-262751G8 00000549122WK0 46-262751G8 00000554358WK2 46-262751G8 00000554360WK8 46-262751G8 00000951784WK8 46-262751G8 00000540122WK9 46-262751G8 00000551387WK4 46-262751G8 00000554457WK2 46-262751G8 00000536131WK6 46-262751G8 00000546581WK0 46-262751G8 00000551392WK4 46-262751G8 00000540121WK1 46-262751G8 00000547501WK7 46-262751G8 00000552622WK3 46-262751G8 00000548414WK2 46-262751G8 00000552931WK8 46-262751G8 00000549920WK7 46-262751G8 00000550993WK0 46-262751G8 00000553944WK0 46-262751G8 00000540272WK2 46-262751G8 00000549174WK1 46-262751G8 00000542310WK8 46-262751G8 00000542729WK9 46-262751G8 00000543348WK7 46-262751G8 00000538300WK5 46-262751G8 00000549923WK1 46-262751G8 00000551389WK0 46-262751G8 00000550144WK0 46-262751G8 00000951166WK8 46-262751G8 00000549578WK3 46-262751G8 00000951422WK5 46-262751G8 00000538689WK1 46-262751G8 00000539273WK3 46-262751G8 00000551893WK1 46-262751G8 00000544936WK8 46-262751G8 00000534305WK8 46-262751G8 00000950887WK0 46-262751G8 00000552617WK3 46-262751G8 00000554357WK4 46-262751G8 00000552314WK7 46-262751G8 00000952658WK3 46-262751G8 00000544942WK6 46-262751G8 00000541503WK9 46-262751G8 00000543655WK5 46-262751G8 00000541501WK3 46-262751G8 00000552613WK2 46-262751G8 00000551895WK6 46-262751G8 00000547503WK3 46-262751G8 00000553949WK9 46-262751G8 00000553951WK5 46-262751G8 00000536107WK6 46-262751G8 00000554462WK2 46-262751G8 00000552619WK9 46-262751G8 00000952073WK5 46-262751G8 00000553714WK7 46-262751G8 00000541727WK4 46-262751G8 00000951505WK7 46-262751G8 00000540274WK8 46-262751G8 00000536174WK6 46-262751G8 00000546368WK2 46-262751G8 00000534086WK4 46-262751G8 00000535375WK0 46-262751G8 00000951414WK2 46-262751G8 00000546587WK7 46-262751G8 00000543346WK1 46-262751G8 00000538308WK8 46-262751G8 00000550354WK5 46-262751G8 00000952619WK5 46-262751G8 00000538934WK1 46-262751G8 00000534769WK5 46-262751G8 00000951554WK5 46-262751G8 00000951592WK5 46-262751G8 00000539917WK5 46-262751G8 00000951400WK1 46-262751G8 00000953104WK7 46-262751G8 00000540277WK1 46-262751G8 00000554463WK0 46-262751G8 00000550139WK0 46-262751G8 00000552925WK0 46-262751G8 00000952034WK7 46-262751G8 00000534302WK5 46-262751G8 00000553330WK2 46-262751G8 00000553331WK0 46-262751G8 00000551900WK4 46-262751G8 00000552927WK6 46-262751G8 00000552614WK0 46-262751G8 00000552316WK2 46-262751G8 00000552933WK4 46-262751G8 00000539282WK4 46-262751G8 00000539916WK7 46-262751G8 00000552315WK4 46-262751G8 00000951563WK6 46-262751G8 00000534303WK3 46-262751G8 00000547235WK2 46-262751G8 00000552620WK7 46-262751G8 00000549921WK5 46-262751G8 00000550356WK0 46-262751G8 00000552616WK5 46-262751G8 00000539910WK0 46-262751G8 00000952720WK1 46-262751G8 00000553328WK6 46-262751G8 00000951165WK0 46-262751G8 00000544039WK1 46-262751G8 00000540119WK5 46-262751G8 00000538936WK6 46-262751G8 00000546244WK5 46-262751G8 00000529625WK6 46-262751G8 00000552618WK1 46-262751G8 00000953108WK8 46-262751G8 00000536173WK8 46-262751G8 00000951167WK6 46-262751G8 00000552924WK3 46-262751G8 00000951602WK2 46-262751G8 00000548140WK3 46-262751G8 00000553713WK9 46-262751G8 00000548409WK2 46-262751G8 00000544743WK8 46-262751G8 00000551899WK8 46-262751G8 00000551894WK9 46-262751G8 00000550997WK1 46-262751G8 00000553329WK4 46-262751G8 00000541892WK6 46-262751G8 00000552615WK7 46-262751G8 00000544940WK0 46-262751G8 00000541893WK4 46-262751G8 00000541502WK1 46-262751G8 00000549577WK5 46-262751G8 00000536175WK3 46-262751G8 00000544041WK7 46-262751G8 00000542728WK1 46-262751G8 00000952017WK2 46-262751G8 00000540269WK8 46-262751G8 00000550357WK8 46-262751G8 00000535878WK3 46-262751G8 00000535876WK7 46-262751G8 00000544739WK6 46-262751G8 00000551896WK4 46-262751G8 00000541723WK3 46-262751G8 00000539911WK8 46-262751G8 00000547240WK2 46-262751G8 00000953009WK8 46-262751G8 00000553952WK3 46-262751G8 00000550140WK8 46-262751G8 00000547239WK4 46-262751G8 00000551391WK6 46-262751G8 00000951884WK6 46-262751G8 00000951816WK8 46-262751G8 00000951751WK7 46-262751G8 00000548411WK8 46-262751G8 00000951628WK7 46-262751G8 00000551897WK2 46-262751G8 00000538302WK1 46-262751G8 00000952242WK6 46-262751G8 00000535376WK8 46-262751G8 00000535875WK9 46-262751G8 00000535373WK5 46-262751G8 00000535377WK6 46-262751G8 00000547502WK5 46-262751G8 00000535370WK1 46-262751G8 00000534304WK1 46-262751G8 00000536172WK0 46-262751G8 00000952634WK4 46-262751G8 00000540270WK6 46-262751G8 00000550142WK4 46-262751G8 00000554466WK3 46-262751G8 00000534561WK6 46-262751G8 00000549579WK1 46-262751G8 00000951254WK2 46-262751G8 00000536171WK2 46-262751G8 00000535879WK1 46-262751G8 00000953107WK0 46-262751G8 00000543349WK5 46-262751G8 00000554460WK6 46-262751G8 00000535369WK3 46-262751G8 00000951975WK2 46-262751G8 00000952466WK1 46-262751G8 00000552929WK2 46-262751G8 00000539913WK4 46-262751G8 00000547238WK6 46-262751G8 00000540273WK0 46-262751G8 00000951883WK8 46-262751G8 00000552621WK5 46-262751G8 00000553712WK1 46-262751G8 00000543017WK8 46-262751G8 00000952137WK8 46-262751G8 00000542311WK6 46-262751G8 00000951872WK1 46-262751G8 00000546590WK1 46-262751G8 00000534771WK1 46-262751G8 00000543652WK2 46-262751G8 00000546584WK4 46-262751G8 00000553948WK1 46-262751G8 00000551898WK0 46-262751G8 00000551901WK2 46-262751G8 00000539274WK1 46-262751G8 00000539915WK9 46-262751G8 00000550998WK9 46-262751G8 00000550355WK2 46-262751G8 00000544046WK6 46-262751G8 00000534772WK9 46-262751G8 00000541891WK8 46-262751G8 00000541725WK8 46-262751G8 00000542076WK5 46-262751G8 00000544038WK3 46-262751G8 00000543653WK0 46-262751G8 00000544037WK5 46-262751G8 00000543924WK5 46-262751G8 00000543651WK4 46-262751G8 00000547494WK5 46-262751G8 00000541724WK1 46-262751G8 |
Recalling Firm/ Manufacturer |
GE Healthcare 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
GE Healthcare Service 800-437-1171
|
Manufacturer Reason for Recall |
The SFD/IDD safety mechanism may not engage properly at the lower range of SFD/IDD travel. This mechanism is installed to hold and prevent the SFD/IDD from falling due to a counterweight cable failure. The lower range of travel is 1.1 in (28mm) from the lowest point of normal SFD/IDD vertical compression. A fall of the SFD/IDD could result in an injury to a patient or operator.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
GE Healthcare sent an "Urgent Medical Device Correction" letter GEHC Ref#10902 dated June 12, 2015 to Hospital Administrators/Risk Managers, Managers of Radiology/Cardiology and Radiologists/Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
Quantity in Commerce |
4,053 (3,804 US; 249 OUS) |
Distribution |
Worldwide Distribution - US Nationwide including PR, GU and DC and the countries of: Afghanistan, Argentina, Australia, Austria, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Czech Republic, Dominican Republic, Ecuador, Egypt, France, Germany, Ghana, Guatemala, Honduras, Indonesia, Ireland, Israel, Italy, Jamaica, Jordan, Korea (Republic of), Kuwait, Libyan Arab Jamahiriya, Malaysia, Malta, Mexico, New Zealand, Panama, Romania, Russia, Saudi Arabia, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Vietnam, and Yemen. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAA and Original Applicant = GE HEALTHCARE 510(K)s with Product Code = JAA and Original Applicant = GE MEDICAL SYSTEMS
|
|
|
|