Class 2 Device Recall Surgical Bur |
|
Date Initiated by Firm |
February 11, 2015 |
Date Posted |
July 24, 2015 |
Recall Status1 |
Terminated 3 on December 30, 2015 |
Recall Number |
Z-2247-2015 |
Recall Event ID |
71608 |
Product Classification |
Bur, surgical, general & plastic surgery - Product Code GFF
|
Product |
Medical Device Exchange 2.15MMX22MM Fluted Spiral Router for CRANI-A; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number A-CRN-S; EXP (repackager), Anspach (OEM)
Bur, surgical, general & plastic surgery |
Code Information |
OEM Catalog number A-CRN-S. |
Recalling Firm/ Manufacturer |
EXP Pharmaceutical Services Corp 48021Warm Springs Boulevard Fremont CA 94539
|
For Additional Information Contact |
Frederick R. Jorgenson, Esq. 336-770-3556
|
Manufacturer Reason for Recall |
EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.
|
FDA Determined Cause 2 |
Other |
Action |
EXP first notified their customers to quarantine EXP devices February 11-17, 2015. Customers were contacted through telephone, e-mail, and written communications. Customers were instructed to recover and quarantine all EXP products from their inventories. EXP believes that they have contacted all customers to whom products were shipped between April 26, 2014, and December 9, 2014.
A second letter is scheduled to be sent in July 2015 (date unknown at this time). This letter will advise users of the recall of due to lack of registration with the FDA. |
Quantity in Commerce |
4 |
Distribution |
Distributed in DC and the states of CO, CA, FL, TX, and AR. |