Date Initiated by Firm |
July 10, 2015 |
Date Posted |
September 03, 2015 |
Recall Status1 |
Terminated 3 on April 12, 2017 |
Recall Number |
Z-2579-2015 |
Recall Event ID |
71721 |
Product Classification |
Neurological tray - Product Code OJG
|
Product |
NEURO SPINE PACK 3/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. . |
Code Information |
Code:9001300 Lots: 141015765 exp. 11/30/15 150116761 exp. 11/30/15 150418292 exp. 5/31/16 150518636 exp. 5/31/16 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall |
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
|
FDA Determined Cause 2 |
Packaging |
Action |
Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer. |
Quantity in Commerce |
123 packs |
Distribution |
Distributed Only in Puerto Rico. |