• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Leica Microsystems Inc.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Leica Microsystems Inc.see related information
Date Initiated by FirmAugust 05, 2015
Date PostedSeptember 17, 2015
Recall Status1 Terminated 3 on January 05, 2017
Recall NumberZ-2779-2015
Recall Event ID 71910
Product Classification Processor, tissue, automated - Product Code IEO
ProductThe Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Code Information Product Code: Leica ASP200S/ASP300S with Serial Numbers: 5682; 5886; 5264; 5267; 5356; 5259; 5956; 5946; 5794; 5385; 5929; 5301; 5401; 5310; 5530; 5909; 5804; 5272; 5724; 5796; 5807; 5446; 5435; 5670; 5947; 5348; 5434; 5388; 5366; 6000; 6131; 5407; 6040; 5799; 5815; 5737; 5706; 5381; 5416; 5673; 5875; 5962; 6041; 6042; 5266; 5948; 5693; 6021; 5610; 5798; 5926; 5387; 5349; 5273; 6115; 5612; 5965; 5955; 5260; 5725; 5871; 5873; 5874; 5562; 6046; 5386; 5977 ;5889; 5398; 6062; 5609; 6015; 5734; 5403; 5289; 5426; 6149; 5663; 6045
Recalling Firm/
Manufacturer
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove IL 60089-6622
For Additional Information ContactKatya Magee
847-236-3747
Manufacturer Reason
for Recall
The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionLeica Biosystems sent an Urgent Medical Device Recall Notification letter dated August 4, 2015, to all affected customers to provide instructions on what the customer should do with the recalled product. These include 1) As an immediate countermeasure, the firm recommends that the customers do not use the remote alarm function on the Leica ASP200S/ASP300S in its current state. 2) The customers local Leica representative will contact them to arrange for a Leica service engineer to visit their facility to correct the internal wiring on their tissue processor and to check the remote alarm function. 3) The Medical Device Recall Notification needs to be passed on to all those who need to be aware within the custromers organisation or to any organisation where the potentially affected devices have been transferred. Please maintain awareness of this Medical Device Recall Notification and resulting action to ensure effectiveness of the corrective action. 4) The letter asks the customer use the enclosed Medical Device Recall Notification Response Form to confirm that they have received and understand the Medical Device Recall Notification. it states "Please confirm receipt of this notice by signing and dating the attached Acknowledgement Form and faxing it back to us at 1-847-236-3747, or scan and email it to LMGRA@leica-microsystems.com"
Quantity in Commerce901
DistributionWorldwide Distribution - US including U.S. including AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, TX, VA, WA, WI, WV and District of Columbia. Internationall to AE, AT, AU , BD , BE, BF, BG,BR, CA, CL, CN, CO, CR, DE, DK, EC, EG, ES, ET, FI, FR, GB, HU, ID, IN, IT, JM, JP, KR, LK, MA, MT, MY, NL, NZ, PA, PE, PH, PL, PT, RO, RU, SA, SE, SG, TH, TR, TW, VN and ZA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-