Class 2 Device Recall KimVent Closed Suction System for Adults, Elbow |
|
Date Initiated by Firm |
July 29, 2015 |
Date Posted |
August 31, 2015 |
Recall Status1 |
Terminated 3 on January 12, 2016 |
Recall Number |
Z-2503-2015 |
Recall Event ID |
71966 |
Product Classification |
Catheters, suction, tracheobronchial - Product Code BSY
|
Product |
KimVent* Closed Suction System for Adults, 12 F, Elbow (Product Code 216); KimVent* Closed Suction KimVent* Closed Suction System for Adults, 14 F, Elbow (Product Code 221); KimVent* Closed Suction System for Adults, 14 F, Elbow (Product Code 22113) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway. |
Code Information |
Product Code 216 - M5098T617, Product Code 221 - M5079T607, M5082T637, M5166T631; Product Code 22113 - M5089T610, M5166T622 |
Recalling Firm/ Manufacturer |
Halyard Health, Inc 5405 Windward Pkwy Alpharetta GA 30004-3894
|
For Additional Information Contact |
Thomas Kozma, PhD 470-448-5681
|
Manufacturer Reason for Recall |
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
|
FDA Determined Cause 2 |
Process control |
Action |
Halyard sent a Product Advisory Notice on July 29, 2015, via FEDEX with a second letter dated August 12, 2015, being issued via FEDEX. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to report any occurrence of a valve leak to Halyard Customer Service at 1-844-425-9273, option 1, option 3. |
Distribution |
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada |