|
Class 2 Device Recall Aisys |
|
Date Initiated by Firm |
December 15, 2015 |
Date Posted |
January 21, 2016 |
Recall Status1 |
Terminated 3 on September 21, 2016 |
Recall Number |
Z-0680-2016 |
Recall Event ID |
72943 |
510(K)Number |
K110213
|
Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
Product |
GE Healthcare, Aisys, 1011-9000-000.
The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment. |
Code Information |
ANAU00153 ANAU00156 ANAU00152 ANAU00150 ANAU00151 ANAU00154 ANAU00155 ANAU00157 ANAU00158 ANAU00159 ANAU00160 ANAU00180 ANAU00181 |
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
GE Healthcare Service 800-437-1171
|
Manufacturer Reason for Recall |
The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
GE Healthcare sent an Urgent Medical Device Correction letter dated December 15, 2015, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers with questions were instructed to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative. |
Quantity in Commerce |
13 |
Distribution |
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY.
Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BSZ and Original Applicant = DATEX-OHMEDA, INC.
|
|
|
|