Date Initiated by Firm |
February 19, 2016 |
Create Date |
January 20, 2017 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1066-2017 |
Recall Event ID |
73459 |
510(K)Number |
K050190
|
Product Classification |
Table, radiologic - Product Code KXJ
|
Product |
Apollo EZ: Model: 9784152036
Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures. |
Code Information |
13101261 |
Recalling Firm/ Manufacturer |
Villa Radiology Systems LLC 91 Willenbrock Rd Ste B1 Oxford CT 06478-1036
|
For Additional Information Contact |
Walter Schneider 203-262-8836
|
Manufacturer Reason for Recall |
The analysis showed that the incident was caused by a chain of four faults:
1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.
2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area.
3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation).
4.) Delayed activation of the emergency stop.
|
FDA Determined Cause 2 |
Device Design |
Action |
Consignees were notified of the recall via letter starting on 2/19/16. The letter explains the issue and actions to be taken by the consignee and manufacturer. |
Quantity in Commerce |
66 distributed in US |
Distribution |
AL
AZ
CA
CO
CT
DC
FL
IA
IL
IN
LA
MA
MD
MI
MN
MO
MS
NJ
NY
OH
PA
Puerto Rico
TN
TX
UT
WI |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = KXJ and Original Applicant = VILLA SISTEMI MEDICALI S.P.A.
|