Class 2 Device Recall IntelliVue MX40 Patient Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall IntelliVue MX40 Patient Monitor see related information
Date Initiated by Firm September 19, 2016
Create Date October 20, 2016
Recall Status1 Terminated 3 on May 25, 2018
Recall Number Z-0293-2017
Recall Event ID 75209
510(K)Number K113125  
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
Product Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 802.11a/b/g
Product Number: 865352

Exchange part:
453564262491
453564262511
453564615311
453564615331
453564262571
453564262591

Exchange part:
453564262531
453564262551
453564262571
453564262591
453564615351
453564615371
453564467761 (Korea)
453564467781 (Korea)
453564451791 (Japan, Taiwan, Indonesia)
453564451811 (Japan, Taiwan, Indonesia)
453564467821 (Mexico)
453564467841 (China)
453564467861 (China)

Worldwide
SW Revisions: B.05.28, B.05.29, and B.05.32
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals
Code Information SW Revisions: B.05.28, B.05.29, and B.05.32
Recalling Firm/
Manufacturer
Philips Electronics North America Corporation
3000 Minuteman Rd
Andover MA 01810-1032
For Additional Information Contact
978-659-3000
Manufacturer Reason
for Recall
Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX
FDA Determined
Cause 2
Software design
Action The firm, Philips Healthcare, sent a "Field Safety Notice (FSN)" dated 9/14/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to make sure that the MX40 Patient Wearable Monitor local configuration setting 'Alarms On' is configured to its factory default until your software is upgraded; and follow the instructions listed in the letter. A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.
Quantity in Commerce 2,212 units
Distribution Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MHX and Original Applicant = PHILIPS MEDICAL SYSTEMS
-
-