|
Class 2 Device Recall Solara CRTP MRI |
|
Date Initiated by Firm |
June 04, 2020 |
Date Posted |
July 10, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2513-2020 |
Recall Event ID |
85816 |
PMA Number |
P010015 |
Product Classification |
Pulse generator, permanent, implantable - Product Code NVZ
|
Product |
Solara CRT-P MRI, Model Numbers: a) W1TR03, b) W1TR06, c) W4TR03, d) W4TR06 |
Code Information |
a) W1TR03, serial numbers: RNS200027H RNS200037H RNS200043H RNS200044H RNS200045H RNS200052H RNS200059H RNS200063H RNS200068H RNS200076H RNS200077H RNS200079H RNS200081H RNS200088H RNS200089H RNS200090H RNS200094H RNS200098H RNS200100H RNS200005H RNS200007H RNS200016H RNS200020H RNS200022H RNS200025H RNS200026H RNS200028H RNS200030H RNS200032H RNS200033H RNS200034H RNS200038H RNS200042H RNS200050H RNS200051H RNS200054H RNS200060H RNS200061H RNS200064H RNS200065H RNS200067H RNS200071H RNS200072H RNS200074H RNS200080H RNS200082H RNS200084H RNS200085H RNS200086H RNS200087H RNS200091H RNS200093H RNS200099H RNS200101H RNS200053H RNS200104H RNS200109H RNS200111H RNS200116H RNS200118H RNS200019H RNS200040H RNS200092H RNS200035H RNS200102H RNS200103H RNS200105H RNS200107H RNS200108H RNS200110H RNS200113H RNS200115H RNS200117H RNS200119H RNS200120H RNS200121H RNS200122H RNS200123H RNS200125H RNS200126H RNS200127H RNS200048H RNS200128H RNS200130H RNS200140H RNS200124H RNS200129H RNS200143H RNS200144H RNS200131H RNS200138H RNS200132H RNS200136H RNS200137H RNS200133H RNS200134H RNS200135H RNS200148H RNS200149H RNS200150H RNS200151H RNS200154H RNS200155H RNS200156H RNS200158H RNS200159H RNS200160H RNS200161H RNS200162H RNS200163H RNS200164H RNS200165H RNS200167H RNS200168H RNS200169H RNS200170H RNS200171H RNS200172H RNS200173H RNS200174H RNS200175H RNS200176H RNS200177H RNS200178H RNS200179H RNS200181H RNS200185H RNS200193H RNS200194H RNS200195H RNS200196H RNS200198H RNS200199H RNS200202H RNS200203H RNS200204H RNS200206H RNS200207H RNS200210H RNS200212H RNS200213H RNS200214H RNS200215H RNS200216H RNS200217H RNS200220H RNS200223H RNS200224H RNS200226H RNS200227H RNS200228H RNS200230H RNS200232H RNS200182H RNS200183H RNS200186H RNS200191H RNS200200H RNS200205H RNS200229H RNS200231H RNS200142H RNS200218H RNS200233H RNS200234H RNS200235H RNS200236H RNS200237H RNS200239H RNS200240H RNS200243H RNS200245H RNS200246H RNS200247H RNS200248H RNS200251H RNS200252H RNS200253H RNS200255H RNS200256H RNS200258H RNS200259H RNS200260H RNS200242H RNS200238H RNS200296H RNS200299H RNS200301H RNS200302H RNS200307H RNS200335H RNS200340H RNS200350H RNS200359H RNS200369H RNS200187H RNS200309H RNS200311H RNS200313H RNS200314H RNS200316H RNS200318H RNS200321H RNS200327H RNS200329H RNS200330H RNS200331H RNS200332H RNS200333H RNS200334H RNS200336H RNS200337H RNS200345H RNS200346H RNS200347H RNS200348H RNS200349H RNS200353H RNS200355H RNS200356H RNS200360H RNS200363H RNS200364H RNS200367H RNS200368H RNS200266H RNS200268H RNS200273H RNS200275H RNS200281H RNS200283H RNS200286H RNS200287H RNS200289H RNS200291H RNS200293H RNS200324H RNS200325H RNS200339H RNS200343H RNS200357H RNS200361H RNS200010H RNS200262H RNS200263H RNS200264H RNS200265H RNS200267H RNS200270H RNS200274H RNS200276H RNS200278H RNS200279H RNS200280H RNS200288H RNS200290H RNS200303H RNS200315H RNS200317H RNS200319H RNS200326H RNS200370H RNS200284H RNS200285H RNS200294H RNS200298H RNS200306H RNS200310H RNS200341H RNS200354H RNS200365H RNS200391H RNS200392H RNS200393H RNS200394H RNS200398H RNS200401H RNS200402H RNS200405H RNS200376H RNS200380H RNS200382H RNS200383H RNS200385H RNS200389H RNS200399H RNS200372H RNS200373H RNS200374H RNS200377H RNS200381H RNS200384H RNS200388H RNS200390H RNS200395H RNS200396H RNS200397H RNS200400H RNS200297H RNS200145H RNS200407H RNS200408H RNS200410H RNS200412H RNS200413H RNS200414H RNS200420H RNS200421H RNS200422H RNS200423H RNS200424H RNS200430H RNS200406H RNS200409H RNS200411H RNS200415H RNS200418H RNS200425H RNS200426H RNS200427H RNS200428H RNS200432H RNS200440H RNS200449H RNS200451H RNS200454H RNS200434H RNS200435H RNS200444H RNS200446H RNS200448H RNS200450H RNS200455H RNS200456H RNS200433H RNS200436H RNS200437H RNS200438H RNS200439H RNS200441H RNS200442H RNS200443H RNS200445H RNS200453H RNS200338H RNS200459H RNS200461H RNS200462H RNS200463H RNS200464H RNS200465H RNS200466H RNS200467H RNS200468H RNS200469H RNS200471H RNS200472H RNS200473H RNS200474H RNS200475H RNS200476H RNS200477H RNS200479H RNS200480H RNS200481H RNS200482H RNS200484H RNS200485H RNS200486H RNS200487H RNS200488H RNS200489H RNS200490H RNS200491H RNS200492H RNS200494H RNS200495H RNS200457H RNS200493H RNS200496H RNS200497H RNS200498H RNS200501H RNS200505H RNS200506H RNS200507H RNS200503H RNS200504H RNS200510H RNS200508H RNS200509H RNS200511H RNS200512H RNS200521H RNS200525H RNS200528H RNS200532H RNS200564H RNS200571H RNS200523H RNS200526H RNS200527H RNS200563H RNS200567H RNS200515H RNS200514H RNS200516H RNS200517H RNS200519H RNS200531H RNS200536H RNS200553H RNS200558H RNS200561H RNS200565H RNS200566H RNS200529H RNS200533H RNS200534H RNS200537H RNS200541H RNS200546H RNS200548H RNS200551H RNS200555H RNS200568H RNS200569H RNS200572H RNS200556H RNS200573H RNS200578H RNS200581H RNS200588H RNS200591H RNS200597H RNS200600H RNS200610H RNS200615H RNS200618H RNS200621H RNS200623H RNS200624H RNS200628H RNS200629H RNS200547H RNS200574H RNS200576H RNS200579H RNS200582H RNS200583H RNS200585H RNS200587H RNS200589H RNS200592H RNS200594H RNS200595H RNS200598H RNS200599H RNS200601H RNS200603H RNS200604H RNS200605H RNS200606H RNS200607H RNS200609H RNS200613H RNS200616H RNS200619H RNS200626H RNS200627H RNS200631H RNS200584H RNS200611H RNS200617H RNS200539H RNS200540H RNS200544H RNS200545H RNS200550H RNS200552H RNS200554H RNS200559H RNS200560H RNS200622H RNS200640H RNS200642H RNS200643H RNS200655H RNS200656H RNS200667H RNS200634H RNS200637H RNS200645H RNS200649H RNS200652H RNS200668H RNS200633H RNS200635H RNS200647H RNS200648H RNS200650H RNS200660H RNS200666H RNS200636H RNS200638H RNS200646H RNS200657H RNS200658H RNS200663H RNS200669H RNS200670H RNS200641H RNS200662H RNS200664H RNS200673H RNS200674H RNS200677H RNS200678H RNS200679H RNS200680H RNS200682H RNS200684H RNS200685H RNS200686H RNS200687H RNS200690H RNS200676H RNS200672H RNS200688H RNS200697H RNS200705H RNS200691H RNS200692H RNS200693H RNS200694H RNS200695H RNS200696H RNS200698H RNS200699H RNS200701H RNS200702H RNS200703H RNS200700H RNS200706H RNS200709H RNS200712H RNS200713H RNS200714H RNS200715H RNS200716H RNS200718H RNS200719H RNS200721H RNS200722H RNS200725H RNS200727H RNS200728H RNS200729H RNS200731H RNS200732H RNS200733H RNS200734H RNS200735H RNS200737H RNS200738H RNS200739H RNS200740H RNS200742H RNS200744H RNS200745H RNS200746H RNS200747H RNS200748H RNS200749H RNS200750H RNS200751H RNS200752H RNS200754H RNS200707H RNS200710H RNS200726H RNS200743H RNS200753H RNS200723H RNS200763H RNS200764H RNS200771H RNS200772H RNS200777H RNS200778H RNS200785H RNS200755H RNS200756H RNS200758H RNS200759H RNS200765H RNS200766H RNS200770H RNS200774H RNS200775H RNS200776H RNS200779H RNS200780H RNS200781H RNS200782H RNS200783H RNS200757H RNS200768H RNS200773H RNS200730H RNS200761H RNS200799H RNS200792H RNS200797H RNS200800H RNS200806H RNS200812H RNS200815H RNS200818H RNS200788H RNS200795H RNS200801H RNS200802H RNS200808H RNS200809H RNS200813H RNS200814H RNS200816H RNS200817H RNS200819H RNS200820H RNS200821H RNS200823H RNS200790H RNS200826H RNS200789H RNS200828H RNS200831H RNS200832H RNS200838H RNS200839H RNS200843H RNS200850H RNS200851H RNS200852H RNS200854H RNS200855H RNS200857H RNS200858H RNS200860H RNS200861H RNS200862H RNS200865H RNS200866H RNS200867H RNS200868H RNS200869H RNS200870H RNS200871H RNS200872H RNS200874H RNS200883H RNS200886H RNS200887H RNS200888H RNS200890H RNS200891H RNS200892H RNS200827H RNS200840H RNS200844H RNS200877H RNS200881H RNS200922H RNS200968H RNS200836H RNS200841H RNS200845H RNS200846H RNS200859H RNS200863H RNS200875H RNS200876H RNS200880H RNS200882H RNS200885H RNS200894H RNS200895H RNS200897H RNS200902H RNS200906H RNS200910H RNS200924H RNS200927H RNS200931H RNS200934H RNS200935H RNS200936H RNS200937H RNS200942H RNS200943H RNS200945H RNS200953H RNS200956H RNS200957H RNS200960H RNS200964H RNS200966H RNS200967H RNS200969H RNS200974H RNS200975H RNS200977H RNS200978H RNS200981H RNS200982H RNS200983H RNS200986H RNS200989H RNS200990H RNS200991H RNS200994H RNS200995H RNS200996H RNS200999H RNS201000H RNS201003H RNS201004H RNS201005H RNS201008H RNS201009H RNS200842H RNS200879H RNS200893H RNS200899H RNS200904H RNS200908H RNS200918H RNS200928H RNS200938H RNS200940H RNS200950H RNS200962H RNS200971H RNS200973H RNS200985H RNS201006H RNS200830H RNS200896H RNS200898H RNS200900H RNS200903H RNS200905H RNS200913H RNS200915H RNS200921H RNS200923H RNS200925H RNS200926H RNS200932H RNS200933H RNS200944H RNS200949H RNS200951H RNS200955H RNS200961H RNS200976H RNS200980H RNS200984H RNS200988H RNS200993H RNS201001H RNS201007H RNS200998H RNS200972H RNS200997H RNS201002H RNS201015H RNS201016H RNS201018H RNS201020H RNS201022H RNS201023H RNS201024H RNS201026H RNS201029H RNS201030H RNS201031H RNS201010H RNS201013H RNS201011H RNS201012H RNS201014H RNS201028H RNS200965H RNS201033H RNS200833H RNS200970H RNS201040H RNS201041H RNS201045H RNS201052H RNS201053H RNS201054H RNS201055H RNS201059H RNS201060H RNS201062H RNS201068H RNS201044H RNS201049H RNS201058H RNS201061H RNS201035H RNS201037H RNS201038H RNS201063H RNS201065H RNS201039H RNS200834H RNS201046H RNS201112H RNS201114H RNS201117H RNS201078H RNS201082H RNS201083H RNS201085H RNS201086H RNS201088H RNS201091H RNS201099H RNS201101H RNS201105H RNS201108H RNS201110H RNS201119H RNS201121H RNS201070H RNS201071H RNS201098H RNS201104H RNS201118H RNS201084H RNS201096H RNS201102H RNS201103H RNS201109H RNS201115H RNS201123H RNS201124H RNS201126H RNS201069H RNS201076H RNS201089H RNS201090H RNS201093H RNS201094H RNS201095H RNS201097H RNS201113H RNS201125H RNS201073H RNS201077H RNS201051H RNS201081H RNS201056H RNS201072H RNS201079H RNS201143H RNS201141H RNS201150H RNS201131H RNS201133H RNS201151H RNS201130H RNS201132H RNS201134H RNS201137H RNS201148H RNS201158H RNS201164H RNS201159H RNS201171H RNS201173H RNS201184H RNS201153H RNS201156H RNS201161H RNS201167H RNS201170H RNS201174H RNS201181H RNS201154H RNS201160H RNS201162H RNS201165H RNS201168H RNS201169H RNS201176H RNS201177H RNS201178H RNS201179H RNS201180H RNS201166H RNS201182H RNS201185H RNS201187H RNS201188H RNS201199H RNS201200H RNS201207H RNS201209H RNS201210H RNS201212H RNS201214H RNS201217H RNS201219H RNS201227H RNS201229H RNS201189H RNS201201H RNS201205H RNS201208H RNS201220H RNS201221H RNS201225H RNS201186H RNS201190H RNS201192H RNS201194H RNS201211H RNS201224H RNS201230H RNS201231H RNS201232H RNS201233H RNS201234H RNS201236H RNS201237H RNS201238H RNS201240H RNS201241H RNS201242H RNS201243H RNS201246H RNS201249H RNS201251H RNS201155H RNS201191H RNS201195H RNS201197H RNS201206H RNS201216H RNS201235H RNS201247H RNS201248H RNS201252H RNS201254H RNS201244H RNS201245H RNS201253H RNS201226H RNS201222H RNS201257H RNS201258H RNS201260H RNS201262H RNS201263H RNS201265H RNS201266H RNS201267H RNS201268H RNS201264H RNS201273H RNS201274H RNS201288H RNS201291H RNS201293H RNS201294H RNS201300H RNS201301H RNS201302H RNS201303H RNS201269H RNS201270H RNS201272H RNS201281H RNS201283H RNS201286H RNS201289H RNS201298H RNS201299H RNS201304H RNS201305H RNS201306H RNS201308H RNS201311H RNS201314H RNS201322H RNS201323H RNS201309H RNS201310H RNS201317H RNS201320H RNS201326H RNS201319H RNS201321H RNS201330H RNS201332H RNS201335H RNS201327H RNS201259H RNS201328H RNS201329H RNS201331H RNS201334H RNS201333H RNS201280H RNS201032H RNS201292H RNS201342H RNS201337H RNS201338H RNS201339H RNS201345H RNS201346H RNS201350H RNS201353H RNS201354H RNS201344H RNS201355H RNS201364H RNS201368H RNS201372H RNS201352H RNS201360H RNS201375H RNS201378H RNS201357H RNS201363H RNS201369H RNS201370H RNS201358H RNS201376H RNS201361H RNS201362H RNS201371H RNS201377H RNS201382H RNS201388H RNS201392H RNS201394H RNS201395H RNS201380H RNS201381H RNS201383H RNS201385H RNS201386H RNS201389H RNS201390H RNS201391H RNS201393H RNS201396H RNS201398H RNS201400H RNS201402H RNS201412H RNS201417H RNS201422H RNS201423H RNS201426H RNS201428H RNS201415H RNS201420H RNS201424H RNS201427H RNS201397H RNS201406H RNS201410H RNS201411H RNS201413H RNS201416H RNS201418H RNS201419H RNS201408H RNS201425H RNS201443H RNS201446H RNS201448H RNS201454H RNS201460H RNS201463H RNS201455H RNS201459H RNS201401H RNS201414H RNS201431H RNS201435H RNS201436H RNS201438H RNS201440H RNS201441H RNS201458H RNS201462H RNS201433H RNS201437H RNS201439H RNS201442H RNS201444H RNS201449H RNS201457H RNS201461H RNS201465H RNS201432H RNS201464H RNS201469H RNS201468H RNS201473H RNS201477H RNS201481H RNS201486H RNS201488H RNS201491H RNS201492H RNS201493H RNS201494H RNS201495H RNS201497H RNS201501H RNS201505H RNS201511H RNS201536H RNS201503H RNS201504H RNS201507H RNS201508H RNS201509H RNS201510H RNS201512H RNS201517H RNS201524H RNS201527H RNS201531H RNS201533H RNS201534H RNS201513H RNS201514H RNS201532H RNS201525H RNS201529H RNS201519H RNS201499H RNS201518H RNS201520H RNS201522H RNS201523H RNS201535H RNS201521H RNS201434H RNS201456H RNS201587H RNS201595H RNS201596H RNS201599H RNS201600H RNS201601H RNS201602H RNS201608H RNS201619H RNS201621H RNS201622H RNS201625H RNS201626H RNS201630H RNS201631H RNS201634H RNS201588H RNS201594H RNS201604H RNS201606H RNS201607H RNS201617H RNS201598H RNS201614H RNS201615H RNS201628H RNS201633H RNS201487H RNS201639H RNS201675H RNS201553H RNS201539H RNS201545H RNS201546H RNS201563H RNS201564H RNS201565H RNS201573H RNS201574H RNS201575H RNS201576H RNS201581H RNS201584H RNS201586H RNS201638H RNS201666H RNS201677H RNS201678H RNS201681H RNS201641H RNS201656H RNS201659H RNS201667H RNS201670H RNS201570H RNS201496H RNS201549H RNS201589H RNS201651H RNS201652H RNS201665H RNS201668H RNS201671H RNS201676H RNS201544H RNS201554H RNS201566H RNS201568H RNS201648H RNS201655H RNS201474H RNS201642H RNS201644H RNS201650H RNS201660H RNS201662H RNS201680H RNS201542H RNS201555H RNS201556H RNS201636H RNS201637H RNS201640H RNS201643H RNS201646H RNS201683H RNS201679H RNS201551H RNS201558H RNS201579H RNS201645H RNS201647H RNS201653H RNS201663H RNS201672H RNS201674H RNS201541H RNS201543H RNS201547H RNS201548H RNS201561H RNS201580H RNS201682H RNS201538H RNS201560H RNS201635H RNS201684H RNS201686H RNS201706H RNS201708H RNS201727H RNS201728H RNS201685H RNS201687H RNS201688H RNS201689H RNS201690H RNS201691H RNS201698H RNS201707H RNS201713H RNS201714H RNS201716H RNS201721H RNS201725H RNS201726H RNS201730H RNS201731H RNS201733H RNS201693H RNS201700H RNS201704H RNS201705H RNS201718H RNS201720H RNS201732H RNS201692H RNS201694H RNS201702H RNS201709H RNS201710H RNS201711H RNS201719H RNS201723H RNS201724H RNS201735H RNS201737H RNS201739H RNS201743H RNS201748H RNS201749H RNS201751H RNS201752H RNS201754H RNS201755H RNS201756H RNS201757H RNS201764H RNS201766H RNS201771H RNS201772H RNS201773H RNS201776H RNS201779H RNS201780H RNS201781H RNS201782H RNS201794H RNS201798H RNS201803H RNS201814H RNS201815H RNS201824H RNS201828H RNS201734H RNS201741H RNS201744H RNS201745H RNS201746H RNS201747H RNS201753H RNS201759H RNS201770H RNS201774H RNS201775H RNS201783H RNS201784H RNS201801H RNS201807H RNS201808H RNS201810H RNS201816H RNS201817H RNS201795H RNS201813H RNS201793H RNS201797H RNS201806H RNS201818H RNS201821H RNS201833H RNS201761H RNS201792H RNS201827H RNS201831H RNS201785H RNS201788H RNS201791H RNS201802H RNS201811H RNS201819H RNS201823H RNS201829H RNS201490H RNS201482H RNS201777H RNS201758H RNS201837H RNS201838H RNS201840H RNS201841H RNS201842H RNS201846H RNS201847H RNS201849H RNS201850H RNS201851H RNS201858H RNS201864H RNS201867H RNS201868H RNS201869H RNS201873H RNS201845H RNS201835H RNS201843H RNS201848H RNS201852H RNS201855H RNS201856H RNS201860H RNS201861H RNS201862H RNS201871H RNS201872H RNS201875H RNS201853H RNS201857H RNS201881H RNS201885H RNS201889H RNS201900H RNS201901H RNS201904H RNS201916H RNS201879H RNS201884H RNS201888H RNS201891H RNS201893H RNS201894H RNS201896H RNS201897H RNS201898H RNS201902H RNS201903H RNS201906H RNS201907H RNS201910H RNS201912H RNS201915H RNS201919H RNS201922H RNS201925H RNS201892H RNS201895H RNS201886H RNS201887H RNS201917H RNS201920H RNS201921H RNS201905H RNS201918H RNS201926H RNS201876H RNS201877H RNS201938H RNS201941H RNS201943H RNS201944H RNS201945H RNS201946H RNS201961H RNS201962H RNS201935H RNS201936H RNS201947H RNS201948H RNS201949H RNS201950H RNS201927H RNS201952H RNS201953H RNS201955H RNS201956H RNS201958H RNS201960H RNS201964H RNS201966H RNS201967H RNS201924H RNS201929H RNS201930H RNS201931H RNS201951H RNS201959H RNS201963H RNS201954H RNS201968H RNS201970H RNS201972H RNS201971H RNS201973H RNS201977H RNS201979H RNS201986H RNS201975H RNS201978H RNS201988H RNS201994H RNS202006H RNS202008H RNS202014H RNS202025H RNS202026H RNS202027H RNS202033H RNS202035H RNS201993H RNS201995H RNS201996H RNS201997H RNS201999H RNS202001H RNS202003H RNS202005H RNS202015H RNS202019H RNS202020H RNS202021H RNS202029H RNS201987H RNS202017H RNS202028H RNS202032H RNS202010H RNS202031H RNS202036H RNS202000H RNS202002H RNS202004H RNS202007H RNS202011H RNS202016H RNS202022H RNS202023H RNS202034H RNS202024H RNS201985H RNS202037H RNS202041H RNS202042H RNS202045H RNS202050H RNS202058H RNS202059H RNS202066H RNS202070H RNS202038H RNS202040H RNS202048H RNS202049H RNS202052H RNS202060H RNS202065H RNS202068H RNS202047H RNS202043H RNS202044H RNS202056H RNS202098H RNS202102H RNS202092H RNS202099H RNS202100H RNS202101H RNS202108H RNS202075H RNS202105H RNS202111H RNS202091H RNS202097H RNS202104H RNS202114H RNS202083H RNS202087H RNS202115H RNS202119H RNS202120H RNS202076H RNS202080H RNS202081H RNS202082H RNS202085H RNS202089H RNS202093H RNS202094H RNS202096H RNS202121H RNS202113H RNS202132H RNS202133H RNS202141H RNS202125H RNS202126H RNS202135H RNS202136H RNS202145H RNS202147H RNS202148H RNS202149H RNS202150H RNS202151H RNS202152H RNS202153H RNS202156H RNS202157H RNS202162H RNS202163H RNS202164H RNS202165H RNS202166H RNS202167H RNS202168H RNS202169H RNS202170H RNS202118H RNS202123H RNS202124H RNS202131H RNS202137H RNS202139H RNS202142H RNS202144H RNS202155H RNS202160H RNS202130H RNS202185H RNS202187H RNS202171H RNS202175H RNS202190H RNS202172H RNS202173H RNS202176H RNS202178H RNS202179H RNS202181H RNS202186H RNS202192H RNS202194H RNS202204H RNS202205H RNS202236H RNS202195H RNS202198H RNS202199H RNS202201H RNS202206H RNS202214H RNS202226H RNS202227H RNS202240H RNS202241H RNS202223H RNS202231H RNS202212H RNS202213H RNS202234H RNS202221H RNS202235H RNS202207H RNS202215H RNS202218H RNS202222H RNS202225H RNS202229H RNS202237H RNS202239H RNS202256H RNS202261H RNS202268H RNS202273H RNS202278H RNS202255H RNS202257H RNS202264H RNS202272H RNS202274H RNS202280H RNS202250H RNS202254H RNS202271H RNS202279H RNS202245H RNS202252H RNS202266H RNS202277H RNS202285H RNS202289H RNS202290H RNS202292H RNS202293H RNS202299H RNS202300H RNS202301H RNS202283H RNS202288H RNS202295H RNS202297H RNS202298H RNS202304H RNS202305H RNS202309H RNS202317H RNS202319H RNS202282H RNS202284H RNS202306H RNS202308H RNS202310H RNS202312H RNS202315H RNS202316H RNS202321H RNS202323H RNS202328H RNS202330H RNS202331H RNS202333H RNS202334H RNS202335H RNS202336H RNS202339H RNS202340H RNS202341H RNS202344H RNS202348H RNS202325H RNS202332H RNS202351H RNS202354H RNS202355H RNS202358H RNS202359H RNS202357H RNS202353H RNS202356H RNS202374H RNS202376H RNS202378H RNS202370H RNS202382H RNS202390H RNS202393H RNS202394H RNS202400H RNS202401H RNS202380H RNS202387H RNS202398H RNS202369H RNS202404H RNS202405H RNS202367H RNS202381H RNS202383H RNS202384H RNS202389H RNS202395H RNS202402H RNS202407H RNS202388H RNS202326H RNS202409H RNS202410H RNS202413H RNS202424H RNS202425H RNS202427H RNS202437H RNS202446H RNS202408H RNS202412H RNS202415H RNS202422H RNS202430H RNS202431H RNS202432H RNS202433H RNS202438H RNS202439H RNS202440H RNS202442H RNS202444H RNS202411H RNS202423H RNS202435H RNS202436H RNS202483H RNS202491H RNS202457H RNS202465H RNS202476H RNS202478H RNS202480H RNS202481H RNS202490H RNS202447H RNS202449H RNS202450H RNS202461H RNS202470H RNS202472H RNS202473H RNS202474H RNS202477H RNS202479H RNS202494H RNS202456H RNS202462H RNS202469H RNS202471H RNS202458H RNS202460H RNS202482H RNS202459H RNS202467H RNS202475H RNS202493H RNS202505H RNS202498H RNS202500H RNS202503H RNS202497H RNS202502H RNS202506H RNS202518H RNS202528H RNS202520H RNS202543H RNS202544H RNS202545H RNS202549H RNS202559H RNS202548H RNS202533H RNS202536H RNS202538H RNS202540H RNS600103S RNS600104S RNS600102S RNS600109S RNS600110S RNS600111S RNS600115S RNS600116S RNS600117S RNS600119S RNS600121S RNS600130S RNS600132S RNS600133S RNS600134S RNS600135S RNS600138S RNS600139S RNS600141S RNS600142S RNS600143S RNS600147S RNS600148S RNS600149S RNS600150S RNS600153S RNS600154S RNS600157S RNS600164S RNS600165S RNS600167S RNS600171S RNS600173S RNS600174S RNS600175S RNS600176S RNS600177S RNS600180S RNS600182S RNS600184S RNS600185S RNS600186S RNS600187S RNS600188S RNS600189S RNS600193S RNS600195S RNS600197S RNS600199S RNS600203S RNS600206S RNS600207S RNS600210S RNS600112S RNS600114S RNS600123S RNS600128S RNS600129S RNS600131S RNS600145S RNS600151S RNS600156S RNS600163S RNS600169S RNS600172S RNS600178S RNS600183S RNS600194S RNS600196S RNS600198S RNS600200S RNS600201S RNS600208S RNS600212S RNS600213S RNS600191S RNS600202S RNS600205S RNS600118S RNS600137S RNS600215S RNS600217S RNS600218S RNS600219S RNS600221S RNS600222S RNS600223S RNS600224S RNS600225S RNS600226S RNS600228S RNS600233S RNS600235S RNS600239S RNS600240S RNS600241S RNS600242S RNS600243S RNS600252S RNS600256S RNS600255S RNS600245S RNS600253S RNS600263S RNS600246S RNS600247S RNS600248S RNS600251S RNS600254S RNS600257S RNS600258S RNS600261S RNS600265S RNS600266S RNS600267S RNS600268S RNS600271S RNS600273S RNS600309S RNS600280S RNS600284S RNS600286S RNS600287S RNS600293S RNS600296S RNS600297S RNS600298S RNS600299S RNS600300S RNS600311S RNS600282S RNS600285S RNS600295S RNS600310S RNS600319S RNS600321S RNS600324S RNS600325S RNS600326S RNS600327S RNS600330S RNS600332S RNS600337S RNS600339S RNS600350S RNS600355S RNS600340S RNS600361S RNS600364S RNS600363S RNS600360S b) W1TR06, serial numbers: RNY600012S RNY600014S RNY600015S RNY600016S RNY600020S RNY600025S RNY600028S RNY600032S RNY600033S RNY600036S RNY600037S RNY600038S RNY600042S RNY600043S RNY600045S RNY600047S RNY600049S RNY600052S RNY600054S RNY600055S RNY600056S RNY600058S RNY600059S RNY600062S RNY600063S RNY600064S RNY600065S RNY600066S RNY600067S RNY600068S RNY600069S RNY600070S RNY600071S RNY600072S RNY600074S RNY600075S RNY600076S RNY600078S RNY600079S RNY600081S RNY600082S RNY600083S RNY600084S RNY600085S RNY600086S RNY600095S RNY600099S RNY600100S RNY600104S RNY600105S RNY600106S RNY600107S RNY600110S RNY600111S RNY600115S RNY600118S RNY600119S RNY600121S RNY600122S RNY600127S RNY600129S RNY600133S RNY600135S RNY600138S RNY600140S RNY600141S RNY600143S RNY600145S RNY600147S RNY600148S RNY600150S RNY600151S RNY600152S RNY600153S RNY600155S RNY600157S RNY600158S RNY600161S RNY600162S RNY600171S RNY600172S RNY600178S RNY600179S RNY600180S RNY600181S RNY600185S RNY600187S RNY600189S RNY600190S RNY600191S RNY600192S RNY600201S RNY600207S RNY600208S RNY600213S RNY600214S RNY600218S RNY600219S RNY600226S RNY600227S RNY600230S RNY600233S RNY600235S RNY600237S RNY600239S RNY600013S RNY600044S RNY600073S RNY600087S RNY600114S RNY600116S RNY600195S RNY600228S RNY600234S RNY600206S RNY600212S RNY600229S RNY600231S RNY600238S RNY600017S RNY600018S RNY600019S RNY600021S RNY600022S RNY600024S RNY600026S RNY600027S RNY600029S RNY600030S RNY600031S RNY600035S RNY600039S RNY600041S RNY600046S RNY600048S RNY600050S RNY600051S RNY600057S RNY600060S RNY600089S RNY600090S RNY600091S RNY600092S RNY600097S RNY600102S RNY600103S RNY600108S RNY600109S RNY600112S RNY600113S RNY600117S RNY600120S RNY600128S RNY600136S RNY600142S RNY600149S RNY600166S RNY600167S RNY600168S RNY600170S RNY600174S RNY600175S RNY600176S RNY600182S RNY600184S RNY600188S RNY600193S RNY600194S RNY600198S RNY600199S RNY600203S RNY600209S RNY600210S RNY600217S RNY600220S RNY600223S RNY600224S RNY600236S RNY600061S RNY600077S RNY600080S RNY600094S RNY600098S RNY600130S RNY600132S RNY600160S RNY600177S RNY600183S RNY600186S RNY600196S RNY600200S RNY600204S RNY600221S RNY600222S RNY600225S RNY600040S RNY600088S RNY600126S RNY600131S RNY600144S RNY600156S RNY600159S RNY600197S RNY600101S RNY600034S RNY600093S RNY600124S RNY600154S RNY600240S RNY600241S RNY600242S RNY600243S RNY600244S RNY600246S RNY600247S RNY600250S RNY600251S RNY600252S RNY600253S RNY600254S RNY600255S RNY600256S RNY600257S RNY600258S RNY600259S RNY600260S RNY600261S RNY600262S RNY600264S RNY600265S RNY600269S RNY600248S RNY600263S RNY600266S RNY600270S RNY600271S RNY600272S RNY600273S RNY600274S RNY600275S RNY600276S RNY600277S RNY600278S RNY600279S RNY600280S RNY600281S RNY600283S RNY600284S RNY600286S RNY600288S RNY600289S RNY600290S RNY600291S RNY600292S RNY600293S RNY600294S RNY600295S RNY600296S RNY600297S RNY600299S RNY600300S RNY600301S RNY600302S RNY600303S RNY600304S RNY600305S RNY600306S RNY600307S RNY600308S RNY600309S RNY600310S RNY600312S RNY600313S RNY600314S RNY600316S RNY600317S RNY600318S RNY600319S RNY600321S RNY600322S RNY600323S RNY600324S RNY600325S RNY600327S RNY600328S RNY600330S RNY600331S RNY600332S RNY600333S RNY600334S RNY600335S RNY600336S RNY600337S RNY600339S RNY600341S RNY600342S RNY600343S RNY600344S RNY600345S RNY600346S RNY600347S RNY600348S RNY600349S RNY600350S RNY600351S RNY600352S RNY600353S RNY600354S RNY600355S RNY600356S RNY600357S RNY600359S RNY600360S RNY600361S RNY600362S RNY600363S RNY600364S RNY600365S RNY600367S RNY600368S RNY600369S RNY600370S RNY600371S RNY600372S RNY600373S RNY600374S RNY600375S RNY600329S RNY600376S RNY600377S RNY600378S RNY600379S RNY600380S RNY600381S RNY600382S RNY600383S RNY600384S RNY600385S RNY600386S RNY600387S RNY600388S RNY600395S RNY600405S RNY600409S RNY600412S RNY600398S RNY600406S RNY600407S RNY600408S RNY600410S RNY600413S RNY600414S RNY600416S RNY600417S RNY600418S RNY600419S RNY600420S RNY600421S RNY600422S RNY600423S RNY600424S RNY600425S RNY600426S RNY600431S RNY600433S RNY600435S RNY600437S RNY600439S RNY600438S RNY600430S RNY600432S RNY600434S RNY600173S RNY600491S RNY600492S RNY600494S RNY600495S RNY600496S RNY600497S RNY600498S RNY600499S RNY600500S RNY600501S RNY600502S RNY600503S RNY600504S RNY600505S RNY600506S RNY600507S RNY600508S RNY600509S RNY600512S RNY600513S RNY600514S RNY600515S RNY600516S RNY600517S RNY600518S RNY600519S RNY600520S RNY600521S RNY600522S RNY600525S RNY600526S RNY600527S RNY600528S RNY600529S RNY600530S RNY600510S RNY600532S RNY600533S RNY600539S RNY600541S RNY600542S RNY600544S RNY600545S RNY600546S RNY600548S RNY600549S RNY600554S RNY600555S RNY600556S RNY600557S RNY600558S RNY600559S RNY600560S RNY600561S RNY600562S RNY600563S RNY600564S RNY600565S RNY600567S RNY600568S RNY600569S RNY600570S RNY600571S RNY600572S RNY600573S RNY600574S RNY600575S RNY600577S RNY600580S RNY600581S RNY600582S RNY600583S RNY600584S RNY600585S RNY600586S RNY600587S RNY600588S RNY600589S RNY600590S RNY600591S RNY600593S RNY600594S RNY600595S RNY600596S RNY600597S RNY600598S RNY600599S RNY600600S RNY600601S RNY600602S RNY600603S RNY600604S RNY600605S RNY600606S RNY600608S RNY600610S RNY600611S RNY600612S RNY600613S RNY600614S RNY600616S RNY600617S RNY600619S RNY600627S RNY600534S RNY600540S RNY600543S RNY600553S RNY600620S RNY600626S RNY600628S RNY600547S RNY600631S RNY600636S RNY600648S RNY600658S RNY600662S RNY600629S RNY600630S RNY600632S RNY600633S RNY600634S RNY600635S RNY600637S RNY600638S RNY600639S RNY600640S RNY600641S RNY600642S RNY600643S RNY600644S RNY600645S RNY600647S RNY600649S RNY600650S RNY600651S RNY600654S RNY600655S RNY600657S RNY600659S RNY600660S RNY600661S RNY600663S RNY600664S RNY600665S RNY600666S RNY600670S RNY600671S RNY600672S RNY600673S RNY600674S RNY600675S RNY600676S RNY600677S RNY600678S RNY600652S RNY600653S RNY600668S RNY600669S RNY600551S RNY600427S RNY600428S RNY600429S RNY600441S RNY600442S RNY600443S RNY600444S RNY600445S RNY600446S RNY600447S RNY600448S RNY600449S RNY600450S RNY600451S RNY600452S RNY600453S RNY600454S RNY600455S RNY600456S RNY600457S RNY600458S RNY600460S RNY600461S RNY600462S RNY600463S RNY600464S RNY600465S RNY600466S RNY600467S RNY600468S RNY600469S RNY600470S RNY600471S RNY600472S RNY600473S RNY600474S RNY600475S RNY600476S RNY600477S RNY600478S RNY600479S RNY600480S RNY600481S RNY600482S RNY600483S RNY600484S RNY600485S RNY600486S RNY600487S RNY600488S RNY600489S RNY600490S RNY600679S RNY600681S RNY600682S RNY600683S RNY600684S RNY600685S RNY600686S RNY600687S RNY600688S RNY600689S RNY600690S RNY600691S RNY600692S RNY600693S RNY600695S RNY600696S RNY600697S RNY600698S RNY600699S RNY600701S RNY600704S RNY600706S RNY600709S RNY600710S RNY600711S RNY600713S RNY600715S RNY600716S RNY600718S RNY600719S RNY600723S RNY600724S RNY600727S RNY600729S RNY600735S RNY600736S RNY600741S RNY600744S RNY600746S RNY600747S RNY600748S RNY600751S RNY600752S RNY600753S RNY600760S RNY600761S RNY600778S RNY600785S RNY600790S RNY600792S RNY600793S RNY600795S RNY600802S RNY600803S RNY600804S RNY600812S RNY600813S RNY600816S RNY600817S RNY600818S RNY600680S RNY600694S RNY600700S RNY600702S RNY600703S RNY600705S RNY600707S RNY600708S RNY600712S RNY600714S RNY600717S RNY600720S RNY600721S RNY600722S RNY600725S RNY600731S RNY600738S RNY600739S RNY600740S RNY600742S RNY600743S RNY600749S RNY600750S RNY600754S RNY600755S RNY600756S RNY600757S RNY600758S RNY600759S RNY600762S RNY600763S RNY600764S RNY600765S RNY600766S RNY600767S RNY600768S RNY600769S RNY600770S RNY600771S RNY600772S RNY600773S RNY600774S RNY600775S RNY600776S RNY600777S RNY600779S RNY600780S RNY600781S RNY600782S RNY600783S RNY600784S RNY600786S RNY600787S RNY600788S RNY600789S RNY600791S RNY600794S RNY600796S RNY600797S RNY600798S RNY600800S RNY600801S RNY600805S RNY600806S RNY600807S RNY600808S RNY600809S RNY600810S RNY600814S RNY600815S RNY600819S RNY600820S RNY600821S RNY600822S RNY600824S RNY600825S RNY600826S RNY600827S RNY600828S RNY600829S RNY600832S RNY600833S RNY600834S RNY600835S RNY600838S RNY600842S RNY600846S RNY600847S RNY600849S RNY600853S RNY600862S RNY600863S RNY600864S RNY600865S RNY600867S RNY600869S RNY600830S RNY600831S RNY600837S RNY600839S RNY600840S RNY600841S RNY600843S RNY600844S RNY600845S RNY600848S RNY600850S RNY600852S RNY600857S RNY600858S RNY600859S RNY600861S RNY600866S RNY600873S RNY600875S RNY600877S RNY600878S RNY600879S RNY600880S RNY600882S RNY600883S RNY600884S RNY600885S RNY600886S RNY600887S RNY600888S RNY600889S RNY600890S RNY600891S RNY600667S RNY600892S RNY600854S RNY600860S RNY600876S RNY600881S RNY600902S RNY600903S RNY600904S RNY600905S RNY600906S RNY600907S RNY600909S RNY600912S RNY600914S RNY600915S RNY600917S RNY600919S RNY600922S RNY600923S RNY600924S RNY600925S RNY600926S RNY600927S RNY600930S RNY600931S RNY600932S RNY600913S RNY600918S RNY600920S RNY600921S RNY600911S RNY600934S RNY600935S RNY600936S RNY600937S RNY600938S RNY600939S RNY600940S RNY600941S RNY600942S RNY600945S RNY600948S RNY600950S RNY600951S RNY600957S RNY600965S RNY600929S RNY600943S RNY600944S RNY600946S RNY600947S RNY600949S RNY600952S RNY600953S RNY600955S RNY600956S RNY600958S RNY600960S RNY600961S RNY600962S RNY600963S RNY600964S RNY600967S RNY600968S RNY600969S RNY600970S RNY600971S RNY600972S RNY600959S RNY600966S RNY600973S RNY600974S RNY600975S RNY600976S RNY600977S RNY600979S RNY600928S RNY600980S RNY600978S RNY600983S RNY600985S RNY600986S RNY600987S RNY600988S RNY600989S RNY600990S RNY600991S RNY600992S RNY600993S RNY600994S RNY600995S RNY600996S RNY600997S RNY600998S RNY601000S RNY601001S RNY600535S RNY600536S RNY600537S RNY600538S RNY600984S RNY600999S RNY601005S RNY601011S RNY601015S RNY601022S RNY601023S RNY601024S RNY601025S RNY601029S RNY601032S RNY601033S RNY601034S RNY601039S RNY601041S RNY601042S RNY601043S RNY601048S RNY601049S RNY601052S RNY601054S RNY601058S RNY601059S RNY601061S RNY601063S RNY601064S RNY601065S RNY601066S RNY601067S RNY601068S RNY601069S RNY601070S RNY601002S RNY601003S RNY601004S RNY601006S RNY601007S RNY601008S RNY601009S RNY601010S RNY601012S RNY601013S RNY601014S RNY601016S RNY601017S RNY601018S RNY601019S RNY601020S RNY601021S RNY601026S RNY601027S RNY601030S RNY601031S RNY601035S RNY601036S RNY601037S RNY601038S RNY601040S RNY601044S RNY601045S RNY601046S RNY601047S RNY601050S RNY601051S RNY601053S RNY601055S RNY601056S RNY601057S RNY601060S RNY601062S RNY601071S RNY601074S RNY601075S RNY601077S RNY601079S RNY601084S RNY601085S RNY601094S RNY601095S RNY601096S RNY601098S RNY601101S RNY601104S RNY601105S RNY601106S RNY601107S RNY601108S RNY601109S RNY601111S RNY601076S RNY601082S RNY601087S RNY601088S RNY601089S RNY601091S RNY601092S RNY601093S RNY601099S RNY601100S RNY601102S RNY601103S RNY601112S RNY601115S RNY601118S RNY601078S RNY601080S RNY601086S RNY601090S RNY601097S RNY601113S RNY601116S RNY601117S RNY601119S RNY601121S RNY601114S RNY601120S RNY601126S RNY601135S RNY601139S RNY601122S RNY601123S RNY601124S RNY601127S RNY601128S RNY601129S RNY601130S RNY601131S RNY601134S RNY601136S RNY601138S RNY601140S RNY601141S RNY601142S RNY601143S RNY601144S RNY601145S RNY601146S RNY601147S RNY601148S RNY601149S RNY601150S RNY601151S RNY601152S RNY601153S RNY601154S RNY601155S RNY601156S RNY601157S RNY601158S RNY601159S RNY601160S RNY601161S RNY601162S RNY601163S RNY601164S RNY601165S RNY601167S RNY601168S RNY601169S RNY601083S RNY601125S RNY601137S RNY601170S RNY601171S RNY601172S RNY601173S RNY601174S RNY601175S RNY601176S RNY601180S RNY601181S RNY601182S RNY601183S RNY601184S RNY601185S RNY601186S RNY601187S RNY601188S RNY601189S RNY601179S RNY601190S RNY601193S RNY601196S RNY601198S RNY601194S RNY601195S RNY601197S RNY601200S RNY601201S RNY601202S RNY601203S RNY601204S RNY601205S RNY601206S RNY601207S RNY601208S RNY601209S RNY601212S RNY601213S RNY601218S RNY601210S RNY601211S RNY601214S RNY601215S RNY601216S RNY601217S RNY601219S RNY601220S RNY601221S RNY601222S RNY601223S RNY601224S RNY601225S RNY601226S RNY601227S RNY601228S RNY601229S RNY601230S RNY601231S RNY601232S RNY601233S RNY601234S RNY601238S RNY601239S RNY601235S RNY601258S RNY601240S RNY601241S RNY601242S RNY601243S RNY601244S RNY601245S RNY601246S RNY601247S RNY601248S RNY601249S RNY601250S RNY601251S RNY601252S RNY601253S RNY601254S RNY601255S RNY601256S RNY601257S RNY601259S RNY601260S RNY601271S RNY601282S RNY601283S RNY601284S RNY601287S RNY601288S RNY601261S RNY601262S RNY601263S RNY601264S RNY601265S RNY601266S RNY601267S RNY601268S RNY601269S RNY601270S RNY601272S RNY601273S RNY601274S RNY601275S RNY601276S RNY601278S RNY601279S RNY601280S RNY601281S RNY601285S RNY601286S RNY601289S RNY601291S RNY601292S RNY601293S RNY601294S RNY601295S RNY601296S RNY601297S RNY601298S RNY601300S RNY601301S RNY601302S RNY601303S RNY601305S RNY601306S RNY601307S RNY601309S RNY601310S RNY601311S RNY601312S RNY601314S RNY601315S RNY601316S RNY601317S RNY601318S RNY601319S RNY601320S RNY601321S RNY601322S RNY601323S RNY601324S RNY601325S RNY601313S RNY601326S RNY601327S RNY601330S RNY601332S RNY601333S RNY601334S RNY601335S RNY601328S RNY601304S RNY601336S RNY601337S RNY601338S RNY601339S RNY601341S RNY601343S RNY601344S RNY601351S RNY601072S RNY601073S RNY601342S RNY601345S RNY601346S RNY601347S RNY601348S RNY601349S RNY601350S RNY601352S RNY601354S RNY601355S RNY601356S RNY601358S RNY601359S RNY601360S RNY601361S RNY601362S RNY601363S RNY601364S RNY601365S RNY601366S RNY601367S RNY601368S RNY601369S RNY601370S RNY601371S RNY601372S RNY601373S RNY601375S RNY601376S RNY601377S RNY601378S RNY601379S RNY601380S RNY601381S RNY601382S RNY601383S RNY601384S RNY601385S RNY601386S RNY601387S RNY601388S RNY601389S RNY601390S RNY601391S RNY601392S RNY601393S RNY601396S RNY601397S RNY601398S RNY601399S RNY601400S RNY601401S RNY601402S RNY601403S RNY601405S RNY601406S RNY601407S RNY601408S RNY601409S RNY601410S RNY601411S RNY601412S RNY601413S RNY601414S RNY601404S RNY601416S RNY601417S RNY601418S RNY601419S RNY601420S RNY601421S RNY601422S RNY601423S RNY601424S RNY601425S RNY601426S RNY601427S RNY601428S RNY601430S RNY601431S RNY601432S RNY601433S RNY601434S RNY601436S RNY601437S RNY601438S RNY601439S RNY601440S RNY601441S RNY601442S RNY601443S RNY601444S RNY601445S RNY601448S RNY601449S RNY601451S RNY601450S RNY601452S RNY601453S RNY601454S RNY601455S RNY601456S RNY601458S RNY601459S RNY601460S RNY601462S RNY601463S RNY601464S RNY601468S RNY601470S RNY601472S RNY601473S RNY601475S RNY601477S RNY601478S RNY601479S RNY601481S RNY601482S RNY601484S RNY601485S RNY601486S RNY601488S RNY601492S RNY601493S RNY601494S RNY601495S RNY601461S RNY601465S RNY601466S RNY601467S RNY601469S RNY601474S RNY601476S RNY601483S RNY601489S RNY601490S RNY601457S RNY601480S RNY601487S RNY601496S RNY601497S RNY601498S RNY601499S RNY601500S RNY601501S RNY601502S RNY601503S RNY601504S RNY601505S RNY601506S RNY601507S RNY601508S RNY601509S RNY601511S RNY601512S RNY601513S RNY601514S RNY601515S RNY601516S RNY601517S RNY601518S RNY601519S RNY601520S RNY601521S RNY601522S RNY601523S RNY601524S RNY601525S RNY601526S RNY601527S RNY601528S RNY601529S RNY601530S RNY601531S RNY601532S RNY601533S RNY601534S RNY601535S RNY601536S RNY601537S RNY601538S RNY601539S RNY601540S RNY601541S RNY601542S RNY601543S RNY601544S RNY601545S RNY601546S RNY601547S RNY601548S RNY601550S RNY601551S RNY601554S RNY601555S RNY601556S RNY601557S RNY601558S RNY601561S RNY601563S RNY601565S RNY601566S RNY601567S RNY601568S RNY601569S RNY601571S RNY601572S RNY601573S RNY601574S RNY601549S RNY601552S RNY601576S RNY601577S RNY601578S RNY601579S RNY601580S RNY601581S RNY601582S RNY601585S RNY601586S RNY601588S RNY601589S RNY601591S RNY601592S RNY601593S RNY601594S RNY601595S RNY601596S RNY601597S RNY601598S RNY601601S RNY601606S RNY601608S RNY601609S RNY601610S RNY601611S RNY601612S RNY601613S RNY601614S RNY601615S RNY601616S RNY601617S RNY601618S RNY601619S RNY601620S RNY601622S RNY601623S RNY601625S RNY601626S RNY601627S RNY601628S RNY601629S RNY601630S RNY601631S RNY601632S RNY601633S RNY601634S RNY601635S RNY601637S RNY601638S RNY601639S RNY601640S RNY601641S RNY601642S RNY601643S RNY601645S RNY601647S RNY601648S RNY601649S RNY601650S RNY601651S RNY601653S RNY601654S RNY601655S RNY601587S RNY601605S RNY601636S RNY601652S RNY601599S RNY601656S RNY601657S RNY601658S RNY601659S RNY601660S RNY601661S RNY601662S RNY601663S RNY601664S RNY601665S RNY601666S RNY601667S RNY601668S RNY601669S RNY601670S RNY601671S RNY601672S RNY601673S RNY601674S RNY601675S RNY601676S RNY601677S RNY601678S RNY601679S RNY601680S RNY601682S RNY601684S RNY601685S RNY601687S RNY601688S RNY601689S RNY601690S RNY601692S RNY601693S RNY601694S RNY601696S RNY601697S RNY601698S RNY601699S RNY601700S RNY601701S RNY601703S RNY601704S RNY601705S RNY601706S RNY601708S RNY601709S RNY601710S RNY601712S RNY601713S RNY601714S RNY601715S RNY601716S RNY601718S RNY601719S RNY601720S RNY601721S RNY601723S RNY601724S RNY601725S RNY601726S RNY601727S RNY601729S RNY601730S RNY601731S RNY601732S RNY601733S RNY601734S RNY601736S RNY601737S RNY601738S RNY601739S RNY601740S RNY601741S RNY601743S RNY601744S RNY601745S RNY601746S RNY601747S RNY601748S RNY601749S RNY601750S RNY601751S RNY601752S RNY601753S RNY601755S RNY601756S RNY601757S RNY601758S RNY601759S RNY601760S RNY601761S RNY601762S RNY601763S RNY601764S RNY601765S RNY601767S RNY601768S RNY601769S RNY601766S RNY601770S RNY601771S RNY601772S RNY601773S RNY601774S RNY601775S RNY601777S RNY601778S RNY601779S RNY601780S RNY601782S RNY601783S RNY601784S RNY601787S RNY601788S RNY601789S RNY601575S RNY601735S RNY601776S RNY601711S RNY601781S RNY601786S RNY601790S RNY601791S RNY601792S RNY601793S RNY601794S RNY601795S RNY601797S RNY601799S RNY601801S RNY601802S RNY601805S RNY601807S RNY601808S RNY601809S RNY601810S RNY601811S RNY601812S RNY601813S RNY601814S RNY601815S RNY601817S RNY601818S RNY601819S RNY601821S RNY601822S RNY601823S RNY601824S RNY601825S RNY601826S RNY601830S RNY601831S RNY601832S RNY601833S RNY601835S RNY601836S RNY601837S RNY601838S RNY601839S RNY601840S RNY601842S RNY601843S RNY601844S RNY601845S RNY601846S RNY601847S RNY601848S RNY601849S RNY601722S RNY601796S RNY601800S RNY601803S RNY601816S RNY601827S RNY601859S RNY601861S RNY601863S RNY601871S RNY601873S RNY601876S RNY601877S RNY601878S RNY601879S RNY601881S RNY601883S RNY601885S RNY601886S RNY601887S RNY601888S RNY601890S RNY601892S RNY601893S RNY601894S RNY601895S RNY601897S RNY601899S RNY601850S RNY601851S RNY601852S RNY601853S RNY601854S RNY601855S RNY601856S RNY601857S RNY601858S RNY601860S RNY601864S RNY601865S RNY601866S RNY601867S RNY601869S RNY601870S RNY601872S RNY601874S RNY601875S RNY601880S RNY601882S RNY601884S RNY601889S RNY601891S RNY601896S RNY601898S RNY601590S RNY601600S RNY601602S RNY601603S RNY601604S RNY601841S RNY601903S RNY601904S RNY601905S RNY601906S RNY601907S RNY601908S RNY601909S RNY601910S RNY601911S RNY601912S RNY601913S RNY601916S RNY601919S RNY601920S RNY601921S RNY601922S RNY601924S RNY601925S RNY601926S RNY601927S RNY601928S RNY601929S RNY601930S RNY601931S RNY601932S RNY601933S RNY601935S RNY601936S RNY601937S RNY601938S RNY601939S RNY601940S RNY601941S RNY601942S RNY601943S RNY601944S RNY601945S RNY601946S RNY601947S RNY601948S RNY601949S RNY601950S RNY601951S RNY601952S RNY601953S RNY601954S RNY601955S RNY601956S RNY601957S RNY601958S RNY601959S RNY601960S RNY601961S RNY601962S RNY601964S RNY601965S RNY601968S RNY601969S RNY601970S RNY601971S RNY601972S RNY601973S RNY601974S RNY601975S RNY601977S RNY601978S RNY601979S RNY601980S RNY601981S RNY601806S RNY601834S RNY601917S RNY601923S RNY601976S RNY601982S RNY601983S RNY601984S RNY601985S RNY601986S RNY601987S RNY601988S RNY601989S RNY601990S RNY601992S RNY601993S RNY601994S RNY601997S RNY601998S RNY601999S RNY602000S RNY602001S RNY602003S RNY602004S RNY602005S RNY602006S RNY602007S RNY602008S RNY602009S RNY602010S RNY602012S RNY602013S RNY602014S RNY602015S RNY602016S RNY602017S RNY602018S RNY602019S RNY602021S RNY602022S RNY601966S RNY601991S RNY602002S RNY601914S RNY601996S RNY602024S RNY602025S RNY602026S RNY602027S RNY602028S RNY602030S RNY602031S RNY602032S RNY602034S RNY602035S RNY602036S RNY602037S RNY602038S RNY602039S RNY602040S RNY602041S RNY602042S RNY602043S RNY602044S RNY602045S RNY602046S RNY602047S RNY602048S RNY602049S RNY602050S RNY602051S RNY602052S RNY602054S RNY602055S RNY602056S RNY602057S RNY602058S RNY602059S RNY602060S RNY602061S RNY602062S RNY602063S RNY602066S RNY602067S RNY602068S RNY602096S RNY602099S RNY602064S RNY602065S RNY602069S RNY602070S RNY602072S RNY602073S RNY602074S RNY602075S RNY602077S RNY602078S RNY602079S RNY602080S RNY602081S RNY602082S RNY602083S RNY602084S RNY602085S RNY602086S RNY602087S RNY602088S RNY602089S RNY602091S RNY602092S RNY602093S RNY602094S RNY602095S RNY602098S RNY602090S RNY602100S RNY602101S RNY602102S RNY602104S RNY602105S RNY602106S RNY602107S RNY602108S RNY602110S RNY602112S RNY602113S RNY602114S RNY602115S RNY602117S RNY602118S RNY602121S RNY602122S RNY602123S RNY602124S RNY602126S RNY602127S RNY602128S RNY602129S RNY602130S RNY602131S RNY602132S RNY602133S RNY602134S RNY602135S RNY602136S RNY602137S RNY602138S RNY602139S RNY602103S RNY602116S RNY602144S RNY602146S RNY602148S RNY602151S RNY602153S RNY602155S RNY602158S RNY602159S RNY602161S RNY602166S RNY602182S RNY602143S RNY602145S RNY602147S RNY602149S RNY602150S RNY602152S RNY602156S RNY602157S RNY602160S RNY602163S RNY602164S RNY602165S RNY602169S RNY602170S RNY602171S RNY602174S RNY602175S RNY602176S RNY602178S RNY602180S RNY602167S RNY602177S RNY602125S RNY602183S RNY602184S RNY602185S RNY602186S RNY602187S RNY602188S RNY602189S RNY602190S RNY602191S RNY602192S RNY602193S RNY602194S RNY602195S RNY602196S RNY602199S RNY602200S RNY602201S RNY602202S RNY602204S RNY602209S RNY602213S RNY602222S RNY602225S RNY602228S RNY602229S RNY602231S RNY602233S RNY602234S RNY602237S RNY602238S RNY602239S RNY602243S RNY602244S RNY602245S RNY602246S RNY602247S RNY602248S RNY602249S RNY602250S RNY602251S RNY602253S RNY602254S RNY602255S RNY602256S RNY602257S RNY602258S RNY602261S RNY602205S RNY602206S RNY602207S RNY602208S RNY602210S RNY602211S RNY602214S RNY602216S RNY602217S RNY602218S RNY602219S RNY602226S RNY602227S RNY602230S RNY602232S RNY602236S RNY602242S RNY602259S RNY602221S RNY602224S RNY602168S RNY602263S RNY602264S RNY602265S RNY602266S RNY602267S RNY602269S RNY602270S RNY602272S RNY602273S RNY602274S RNY602275S RNY602276S RNY602277S RNY602279S RNY602281S RNY602282S RNY602283S RNY602287S RNY602288S RNY602289S RNY602290S RNY602292S RNY602293S RNY602294S RNY602295S RNY602298S RNY602300S RNY602301S RNY602304S RNY602307S RNY602308S RNY602313S RNY602314S RNY602315S RNY602317S RNY602320S RNY602323S RNY602325S RNY602326S RNY602327S RNY602330S RNY602333S RNY602335S RNY602336S RNY602337S RNY602342S RNY602262S RNY602268S RNY602271S RNY602278S RNY602280S RNY602284S RNY602285S RNY602291S RNY602296S RNY602297S RNY602302S RNY602303S RNY602305S RNY602306S RNY602311S RNY602312S RNY602318S RNY602324S RNY602328S RNY602331S RNY602332S RNY602339S RNY602340S RNY602310S RNY602172S RNY602179S RNY602140S RNY602141S RNY602142S RNY602344S RNY602345S RNY602346S RNY602347S RNY602348S RNY602349S RNY602350S RNY602351S RNY602352S RNY602353S RNY602354S RNY602355S RNY602356S RNY602357S RNY602358S RNY602359S RNY602361S RNY602363S RNY602365S RNY602366S RNY602367S RNY602368S RNY602369S RNY602370S RNY602372S RNY602373S RNY602375S RNY602377S RNY602378S RNY602379S RNY602380S RNY602382S RNY602383S RNY602384S RNY602385S RNY602386S RNY602387S RNY602388S RNY602389S RNY602390S RNY602391S RNY602392S RNY602393S RNY602394S RNY602395S RNY602396S RNY602397S RNY602398S RNY602399S RNY602400S RNY602401S RNY602403S RNY602405S RNY602406S RNY602407S RNY602409S RNY602410S RNY602411S RNY602412S RNY602415S RNY602416S RNY602417S RNY602418S RNY602419S RNY602420S RNY602421S RNY602364S RNY602374S RNY602376S RNY602408S RNY602422S RNY602423S RNY602426S RNY602431S RNY602433S RNY602434S RNY602436S RNY602437S RNY602442S RNY602445S RNY602446S RNY602461S RNY602465S RNY602466S RNY602468S RNY602469S RNY602470S RNY602471S RNY602473S RNY602474S RNY602477S RNY602478S RNY602480S RNY602482S RNY602483S RNY602484S RNY602485S RNY602486S RNY602487S RNY602488S RNY602490S RNY602493S RNY602495S RNY602496S RNY602498S RNY602499S RNY602501S RNY602502S RNY602508S RNY602509S RNY602513S RNY602514S RNY602521S RNY602522S RNY602523S RNY602530S RNY602531S RNY602536S RNY602539S RNY602540S RNY602546S RNY602547S RNY602548S RNY602550S RNY602551S RNY602552S RNY602554S RNY602556S RNY602561S RNY602563S RNY602565S RNY602567S RNY602568S RNY602571S RNY602572S RNY602576S RNY602577S RNY602581S RNY602585S RNY602595S RNY602598S RNY602602S RNY602603S RNY602425S RNY602427S RNY602429S RNY602430S RNY602435S RNY602439S RNY602440S RNY602443S RNY602444S RNY602447S RNY602448S RNY602449S RNY602450S RNY602451S RNY602452S RNY602453S RNY602454S RNY602455S RNY602456S RNY602457S RNY602458S RNY602459S RNY602460S RNY602462S RNY602467S RNY602472S RNY602475S RNY602476S RNY602479S RNY602489S RNY602491S RNY602492S RNY602494S RNY602500S RNY602503S RNY602504S RNY602505S RNY602506S RNY602507S RNY602510S RNY602516S RNY602518S RNY602520S RNY602524S RNY602525S RNY602528S RNY602529S RNY602532S RNY602533S RNY602534S RNY602538S RNY602541S RNY602542S RNY602543S RNY602545S RNY602557S RNY602559S RNY602560S RNY602562S RNY602564S RNY602570S RNY602573S RNY602574S RNY602575S RNY602578S RNY602579S RNY602580S RNY602582S RNY602583S RNY602584S RNY602586S RNY602587S RNY602588S RNY602589S RNY602590S RNY602591S RNY602592S RNY602593S RNY602594S RNY602599S RNY602600S RNY602604S RNY602605S RNY602607S RNY602608S RNY602544S RNY602558S RNY602609S RNY602610S RNY602612S RNY602613S RNY602617S RNY602619S RNY602621S RNY602517S RNY602519S RNY602481S RNY602623S RNY602624S RNY602627S RNY602628S RNY602629S RNY602630S RNY602631S RNY602633S RNY602635S RNY602636S RNY602637S RNY602639S RNY602642S RNY602644S RNY602646S RNY602648S RNY602649S RNY602650S RNY602651S RNY602622S RNY602625S RNY602634S RNY602638S RNY602640S RNY602641S RNY602643S RNY602652S RNY602653S RNY602654S RNY602655S RNY602656S RNY602658S RNY602660S RNY602661S RNY602662S RNY602663S RNY602664S RNY602665S RNY602666S RNY602667S RNY602671S RNY602672S RNY602673S RNY602675S RNY602676S RNY602678S RNY602679S RNY602680S RNY602681S RNY602682S RNY602683S RNY602684S RNY602685S RNY602687S RNY602688S RNY602689S RNY602690S RNY602691S RNY602692S RNY602693S RNY602694S RNY602695S RNY602697S RNY602698S RNY602699S RNY602700S RNY602701S RNY602702S RNY602703S RNY602704S RNY602706S RNY602707S RNY602708S RNY602710S RNY602711S RNY602712S RNY602713S RNY602715S RNY602718S RNY602696S RNY602720S RNY602722S RNY602723S RNY602724S RNY602726S RNY602729S RNY602731S RNY602732S RNY602733S RNY602734S RNY602735S RNY602736S RNY602737S RNY602738S RNY602739S RNY602741S RNY602743S RNY602744S RNY602745S RNY602747S RNY602719S RNY602725S RNY602727S RNY602728S RNY602740S RNY602746S RNY602721S RNY602748S RNY602749S RNY602750S RNY602751S RNY602752S RNY602753S RNY602755S RNY602756S RNY602757S RNY602758S RNY602759S RNY602761S RNY602762S RNY602763S RNY602765S RNY602766S RNY602767S RNY602768S RNY602769S RNY602770S RNY602771S RNY602772S RNY602773S RNY602774S RNY602775S RNY602776S RNY602777S RNY602778S RNY602779S RNY602780S RNY602781S RNY602782S RNY602783S RNY602784S RNY602785S RNY602786S RNY602787S RNY602790S RNY602792S RNY602793S RNY602794S RNY602795S RNY602796S RNY602797S RNY602798S RNY602799S RNY602800S RNY602801S RNY602802S RNY602803S RNY602804S RNY602805S RNY602807S RNY602808S RNY602810S RNY602811S RNY602813S RNY602814S RNY602815S RNY602816S RNY602818S RNY602819S RNY602820S RNY602821S RNY602822S RNY602823S RNY602824S RNY602825S RNY602826S RNY602827S RNY602828S RNY602829S RNY602830S RNY602831S RNY602832S RNY602833S RNY602834S RNY602835S RNY602837S RNY602838S RNY602839S RNY602840S RNY602841S RNY602842S RNY602843S RNY602844S RNY602846S RNY602857S RNY602864S RNY602849S RNY602853S RNY602854S RNY602855S RNY602856S RNY602858S RNY602861S RNY602862S RNY602866S RNY602867S RNY602868S RNY602869S RNY602870S RNY602871S RNY602873S RNY602874S RNY602875S RNY602876S RNY602877S RNY602879S RNY602880S RNY602881S RNY602882S RNY602884S RNY602886S RNY602887S RNY602888S RNY602889S RNY602891S RNY602892S RNY602894S RNY602895S RNY602896S RNY602897S RNY602898S RNY602899S RNY602900S RNY602901S RNY602902S RNY602903S RNY602904S RNY602905S RNY602686S RNY602872S RNY602906S RNY602908S RNY602883S RNY602909S RNY602913S RNY602914S RNY602915S RNY602920S RNY602924S RNY602926S RNY602929S RNY602932S RNY602939S RNY602947S RNY602957S RNY602959S RNY602963S RNY602965S RNY602967S RNY602968S RNY602910S RNY602911S RNY602912S RNY602916S RNY602919S RNY602925S RNY602927S RNY602928S RNY602930S RNY602933S RNY602934S RNY602937S RNY602938S RNY602942S RNY602943S RNY602945S RNY602946S RNY602948S RNY602949S RNY602950S RNY602951S RNY602952S RNY602953S RNY602956S RNY602958S RNY602961S RNY602966S RNY602917S RNY602921S RNY602931S RNY602935S RNY602936S RNY602944S RNY602954S RNY602960S RNY602962S RNY602964S RNY602181S RNY602515S RNY602971S RNY602977S RNY602979S RNY602980S RNY602983S RNY602984S RNY602988S RNY602989S RNY602990S RNY602991S RNY602994S RNY602995S RNY602976S RNY602985S RNY602986S RNY602987S RNY602993S RNY602998S RNY602999S RNY603002S RNY603003S RNY603004S RNY602969S RNY602970S RNY602972S RNY602973S RNY602974S RNY602975S RNY602978S RNY602981S RNY602992S RNY603008S RNY603031S RNY603032S RNY603036S RNY603037S RNY603038S RNY603048S RNY603050S RNY603052S RNY603055S RNY603057S RNY603058S RNY603067S RNY603010S RNY603011S RNY603015S RNY603020S RNY603022S RNY603023S RNY603028S RNY603029S RNY603034S RNY603035S RNY603041S RNY603047S RNY603054S RNY602120S RNY603009S RNY603013S RNY603024S RNY603025S RNY603026S RNY603030S RNY603040S RNY603043S RNY603044S RNY603046S RNY603049S RNY603051S RNY603053S RNY603056S RNY603059S RNY603060S RNY603062S RNY603063S RNY603064S RNY603066S RNY603068S RNY603014S RNY603021S RNY603027S RNY603069S RNY603070S RNY603071S RNY603072S RNY603073S RNY603074S RNY603075S RNY603076S RNY603077S RNY603078S RNY603079S RNY603080S RNY603081S RNY603083S RNY603084S RNY603085S RNY603088S RNY603089S RNY603090S RNY603091S RNY603092S RNY603093S RNY603094S RNY603095S RNY603096S RNY603097S RNY603098S RNY603099S RNY603100S RNY603101S RNY603102S RNY603103S RNY603104S RNY603105S RNY603106S RNY603107S RNY602463S RNY603087S RNY602286S RNY602154S RNY603086S RNY602299S RNY603115S RNY603116S RNY603121S RNY603122S RNY603123S RNY603124S RNY603125S RNY603127S RNY603128S RNY603129S RNY603130S RNY603131S RNY603133S RNY603134S RNY603135S RNY603137S RNY603138S RNY603139S RNY603140S RNY603109S RNY603110S RNY603111S RNY603112S RNY603114S RNY603119S RNY603132S RNY603136S RNY603141S RNY603143S RNY603146S RNY603145S RNY603120S RNY603147S RNY603148S RNY603149S RNY603150S RNY603151S RNY603152S RNY603154S RNY603155S RNY603156S RNY603157S RNY603158S RNY603159S RNY603160S RNY603161S RNY603162S RNY603163S RNY603164S RNY603165S RNY603166S RNY603167S RNY603168S RNY603170S RNY603172S RNY603176S RNY603178S RNY603180S RNY603184S RNY603187S RNY603198S RNY603212S RNY603214S RNY603217S RNY603219S RNY603221S RNY603222S RNY603223S RNY603224S RNY603226S RNY603227S RNY603250S RNY603253S RNY603266S RNY603289S RNY603290S RNY603291S RNY603292S RNY603126S RNY603153S RNY603171S RNY603173S RNY603174S RNY603175S RNY603177S RNY603179S RNY603181S RNY603182S RNY603183S RNY603185S RNY603186S RNY603188S RNY603189S RNY603190S RNY603191S RNY603193S RNY603194S RNY603197S RNY603200S RNY603201S RNY603202S RNY603203S RNY603204S RNY603205S RNY603209S RNY603210S RNY603211S RNY603213S RNY603215S RNY603216S RNY603218S RNY603225S RNY603228S RNY603229S RNY603230S RNY603231S RNY603232S RNY603233S RNY603234S RNY603235S RNY603236S RNY603237S RNY603238S RNY603239S RNY603240S RNY603241S RNY603242S RNY603243S RNY603244S RNY603245S RNY603246S RNY603247S RNY603248S RNY603249S RNY603251S RNY603254S RNY603255S RNY603256S RNY603257S RNY603258S RNY603259S RNY603260S RNY603261S RNY603262S RNY603263S RNY603264S RNY603265S RNY603267S RNY603268S RNY603269S RNY603270S RNY603271S RNY603272S RNY603274S RNY603275S RNY603276S RNY603277S RNY603278S RNY603279S RNY603280S RNY603281S RNY603282S RNY603283S RNY603284S RNY603285S RNY603286S RNY603287S RNY603142S RNY603293S RNY603294S RNY603295S RNY603296S RNY603297S RNY603298S RNY603300S RNY603301S RNY603302S RNY603303S RNY603304S RNY603308S RNY603309S RNY603311S RNY603312S RNY603313S RNY603314S RNY603315S RNY603317S RNY603318S RNY603319S RNY603320S RNY603321S RNY603322S RNY603323S RNY603324S RNY603325S RNY603326S RNY603327S RNY603328S RNY603329S RNY603330S RNY603331S RNY603332S RNY603333S RNY603334S RNY603335S RNY603337S RNY603338S RNY603339S RNY603340S RNY603341S RNY603342S RNY603343S RNY603344S RNY603345S RNY603346S RNY603347S RNY603348S RNY603349S RNY603350S RNY603351S RNY603352S RNY603353S RNY603354S RNY603355S RNY603356S RNY603357S RNY603359S RNY603361S RNY603362S RNY603363S RNY603364S RNY603365S RNY603366S RNY603367S RNY603368S RNY603369S RNY603370S RNY603371S RNY603372S RNY603373S RNY603374S RNY603375S RNY603376S RNY603377S RNY603378S RNY603379S RNY603380S RNY603381S RNY603382S RNY603385S RNY603192S RNY603196S RNY603383S RNY603387S RNY603388S RNY603389S RNY603391S RNY603392S RNY603393S RNY603395S RNY603397S RNY603404S RNY603407S RNY603408S RNY603409S RNY603410S RNY603413S RNY603415S RNY603416S RNY603417S RNY603421S RNY603422S RNY603423S RNY603424S RNY603425S RNY603426S RNY603427S RNY603435S RNY603438S RNY603441S RNY603386S RNY603390S RNY603396S RNY603398S RNY603400S RNY603412S RNY603414S RNY603418S RNY603420S RNY603429S RNY603430S RNY603431S RNY603432S RNY603433S RNY603434S RNY603436S RNY603437S RNY603440S RNY603442S RNY603428S RNY603445S RNY603394S RNY603486S RNY603487S RNY603488S RNY603494S RNY603495S RNY603497S RNY603498S RNY603499S RNY603504S RNY603509S RNY603492S RNY603505S RNY603506S RNY603507S RNY603511S RNY603513S RNY603514S RNY603515S RNY603516S RNY603517S RNY603520S RNY603521S RNY603522S RNY603523S RNY603524S RNY603525S RNY603527S RNY603528S RNY603531S RNY603532S RNY603535S RNY603537S RNY603542S RNY603543S RNY603545S RNY603447S RNY603448S RNY603455S RNY603459S RNY603461S RNY603462S RNY603465S RNY603468S RNY603470S RNY603471S RNY603472S RNY603475S RNY603478S RNY603484S RNY603485S RNY603489S RNY603490S RNY603491S RNY603493S RNY603496S RNY603500S RNY603501S RNY603503S RNY603508S RNY603510S RNY603512S RNY603519S RNY603530S RNY603540S RNY603544S RNY603547S RNY603548S RNY603446S RNY603449S RNY603450S RNY603451S RNY603452S RNY603453S RNY603454S RNY603456S RNY603457S RNY603458S RNY603460S RNY603463S RNY603464S RNY603466S RNY603467S RNY603469S RNY603473S RNY603474S RNY603476S RNY603477S RNY603479S RNY603480S RNY603483S RNY603518S RNY603482S RNY603526S RNY603529S RNY603534S RNY603536S RNY603538S RNY603541S RNY603546S RNY603533S RNY603553S RNY603555S RNY603556S RNY603557S RNY603558S RNY603559S RNY603560S RNY603561S RNY603562S RNY603565S RNY603567S RNY603568S RNY603569S RNY603570S RNY603572S RNY603574S RNY603575S RNY603576S RNY603580S RNY603581S RNY603582S RNY603583S RNY603584S RNY603585S RNY603588S RNY603589S RNY603591S RNY603593S RNY603595S RNY603596S RNY603600S RNY603601S RNY603602S RNY603603S RNY603604S RNY603608S RNY603577S RNY603579S RNY603586S RNY603587S RNY603592S RNY603594S RNY603597S RNY603605S RNY603607S RNY603610S RNY603399S RNY603613S RNY603614S RNY603615S RNY603617S RNY603622S RNY603623S RNY603625S RNY603627S RNY603651S RNY603578S RNY603620S RNY603621S RNY603624S RNY603628S RNY603629S RNY603630S RNY603631S RNY603632S RNY603633S RNY603635S RNY603636S RNY603637S RNY603638S RNY603639S RNY603640S RNY603641S RNY603642S RNY603644S RNY603648S RNY603649S RNY603650S RNY603645S RNY603612S RNY603616S RNY603643S RNY603647S RNY603652S RNY603654S RNY603655S RNY603657S RNY603661S RNY603664S RNY603666S RNY603667S RNY603668S RNY603669S RNY603671S RNY603676S RNY603677S RNY603678S RNY603683S RNY603684S RNY603686S RNY603690S RNY603653S RNY603656S RNY603662S RNY603663S RNY603665S RNY603670S RNY603672S RNY603673S RNY603679S RNY603680S RNY603681S RNY603682S RNY603687S RNY603688S RNY603689S RNY603658S RNY603659S RNY603660S RNY603685S RNY603691S RNY603693S RNY603694S RNY603695S RNY603697S RNY603698S RNY603699S RNY603700S RNY603701S RNY603702S RNY603703S RNY603705S RNY603706S RNY603707S RNY603709S RNY603710S RNY603696S RNY603711S RNY603712S RNY603714S RNY603716S RNY603717S RNY603718S RNY603720S RNY603721S RNY603722S RNY603723S RNY603724S RNY603725S RNY603726S RNY603728S RNY603729S RNY603730S RNY603731S RNY603732S RNY603733S RNY603735S RNY603738S RNY603739S RNY603740S RNY603741S RNY603742S RNY603744S RNY603748S RNY603749S RNY603750S RNY603752S RNY603756S RNY603764S RNY603767S RNY603779S RNY603780S RNY603784S RNY603788S RNY603789S RNY603791S RNY603793S RNY603796S RNY603801S RNY603808S RNY603713S RNY603727S RNY603734S RNY603745S RNY603763S RNY603800S RNY603803S RNY603747S RNY603751S RNY603753S RNY603754S RNY603755S RNY603757S RNY603758S RNY603759S RNY603761S RNY603762S RNY603766S RNY603768S RNY603775S RNY603776S RNY603777S RNY603778S RNY603782S RNY603790S RNY603792S RNY603795S RNY603798S RNY603799S RNY603804S RNY603805S RNY603806S RNY603760S RNY603770S RNY603771S RNY603773S RNY603774S RNY603781S RNY603785S RNY603786S RNY603794S RNY603797S RNY603765S RNY603787S RNY603809S RNY603502S RNY603810S RNY603826S RNY603829S RNY603834S RNY603843S RNY603845S RNY603846S RNY603847S RNY603858S RNY603811S RNY603813S RNY603814S RNY603815S RNY603816S RNY603817S RNY603818S RNY603819S RNY603820S RNY603821S RNY603822S RNY603823S RNY603824S RNY603825S RNY603827S RNY603828S RNY603830S RNY603831S RNY603832S RNY603833S RNY603835S RNY603836S RNY603837S RNY603838S RNY603839S RNY603840S RNY603841S RNY603842S RNY603844S RNY603848S RNY603849S RNY603850S RNY603851S RNY603852S RNY603853S RNY603854S RNY603855S RNY603857S RNY603772S RNY603859S RNY603861S RNY603863S RNY603865S RNY603866S RNY603867S RNY603869S RNY603872S RNY603873S RNY603874S RNY603875S RNY603877S RNY603878S RNY603879S RNY603881S RNY603882S RNY603883S RNY603884S RNY603885S RNY603886S RNY603887S RNY603888S RNY603889S RNY603890S RNY603891S RNY603892S RNY603893S RNY603894S RNY603895S RNY603896S RNY603897S RNY603880S RNY603860S RNY603936S RNY603626S RNY603898S RNY603899S RNY603901S RNY603902S RNY603903S RNY603904S RNY603905S RNY603906S RNY603907S RNY603909S RNY603910S RNY603911S RNY603912S RNY603913S RNY603914S RNY603915S RNY603916S RNY603917S RNY603918S RNY603919S RNY603920S RNY603921S RNY603922S RNY603923S RNY603924S RNY603925S RNY603926S RNY603927S RNY603928S RNY603929S RNY603930S RNY603931S RNY603932S RNY603933S RNY603934S RNY603935S RNY603937S RNY603938S RNY603939S RNY603940S RNY603941S RNY603942S RNY603943S RNY603944S RNY603945S RNY603946S RNY603947S RNY603948S RNY603950S RNY603951S RNY603952S RNY603953S RNY603954S RNY603955S RNY603956S RNY603957S RNY603958S RNY603959S RNY603960S RNY603961S RNY603962S RNY603963S RNY603965S RNY603966S RNY603967S RNY603968S RNY603969S RNY603970S RNY603971S RNY603972S RNY603973S RNY603974S RNY603975S RNY603976S RNY603977S RNY603984S RNY603986S RNY603988S RNY603989S RNY603990S RNY603994S RNY604005S RNY604008S RNY604014S RNY604016S RNY604017S RNY604018S RNY604019S RNY604020S RNY604042S RNY604043S RNY604045S RNY604046S RNY604048S RNY604051S RNY603979S RNY603980S RNY603982S RNY603983S RNY603987S RNY603991S RNY603992S RNY603993S RNY603996S RNY603998S RNY603999S RNY604000S RNY604001S RNY604002S RNY604003S RNY604004S RNY604010S RNY604011S RNY604012S RNY604013S RNY604015S RNY604022S RNY604023S RNY604024S RNY604025S RNY604026S RNY604027S RNY604028S RNY604029S RNY604030S RNY604031S RNY604033S RNY604034S RNY604035S RNY604036S RNY604037S RNY604038S RNY604039S RNY604041S RNY604047S RNY604049S RNY604052S RNY604053S RNY604054S RNY604055S RNY604056S RNY604057S RNY604058S RNY603985S RNY604006S RNY604059S RNY604060S RNY604061S RNY604062S RNY604063S RNY604065S RNY604067S RNY604068S RNY604069S RNY604070S RNY604071S RNY604072S RNY604073S RNY604074S RNY604075S RNY604076S RNY604077S RNY604078S RNY604080S RNY604081S RNY604082S RNY604083S RNY604084S RNY604085S RNY604086S RNY604087S RNY604089S RNY604091S RNY604095S RNY604096S RNY604097S RNY604098S RNY604094S RNY604099S RNY604100S RNY604101S RNY604102S RNY604103S RNY604104S RNY604106S RNY604107S RNY604108S RNY604109S RNY604110S RNY604111S RNY604113S RNY604114S RNY604115S RNY604116S RNY604117S RNY604118S RNY604119S RNY604120S RNY604121S RNY604122S RNY604123S RNY604125S RNY604126S RNY604127S RNY604129S RNY604130S RNY604131S RNY604132S RNY604133S RNY604134S RNY604135S RNY604139S RNY604140S RNY604141S RNY604146S RNY604147S RNY604149S RNY604157S RNY604158S RNY604142S RNY604144S RNY604145S RNY604150S RNY604152S RNY604153S RNY604154S RNY604155S RNY604156S RNY604151S RNY604148S RNY604159S RNY604161S RNY604162S RNY604163S RNY604164S RNY604165S RNY604166S RNY604173S RNY604176S RNY604179S RNY604181S RNY604182S RNY604183S RNY604184S RNY604187S RNY604189S RNY604190S RNY604191S RNY604192S RNY604193S RNY604195S RNY604196S RNY604197S RNY604198S RNY604160S RNY604167S RNY604178S RNY604186S RNY604199S RNY604200S RNY604202S RNY604203S RNY604206S RNY604209S RNY604216S RNY604217S RNY604220S RNY604221S RNY604222S RNY604223S RNY604225S RNY604226S RNY604228S RNY604229S RNY604230S RNY604232S RNY604234S RNY604235S RNY604238S RNY604239S RNY604240S RNY604244S RNY604249S RNY604251S RNY604252S RNY604253S RNY604256S RNY604257S RNY604112S RNY604205S RNY604218S RNY604267S RNY604268S RNY604271S RNY604272S RNY604277S RNY604278S RNY604282S RNY604283S RNY604284S RNY604285S RNY604286S RNY604287S RNY604288S RNY604289S RNY604290S RNY604291S RNY604292S RNY604294S RNY604295S RNY604300S RNY604259S RNY604260S RNY604261S RNY604262S RNY604263S RNY604264S RNY604265S RNY604266S RNY604269S RNY604270S RNY604273S RNY604274S RNY604275S RNY604276S RNY604281S RNY604297S RNY604298S RNY604299S RNY604301S RNY604302S RNY604303S RNY604304S RNY604305S RNY604306S RNY604307S RNY604308S RNY604309S RNY604312S RNY604313S RNY604314S RNY604315S RNY604316S RNY604317S RNY604318S RNY604319S RNY604320S RNY604321S RNY604322S RNY604324S RNY604325S RNY604326S RNY604327S RNY604328S RNY604330S RNY604331S RNY604332S RNY604333S RNY604334S RNY604335S RNY604336S RNY604337S RNY604338S RNY604339S RNY604340S RNY604341S RNY604342S RNY604343S RNY604344S RNY604345S RNY604346S RNY604347S RNY604348S RNY604349S RNY604350S RNY604351S RNY604353S RNY604354S RNY604355S RNY604356S RNY604357S RNY604358S RNY604359S RNY604361S RNY604362S RNY604363S RNY604364S RNY604366S RNY604368S RNY604369S RNY604371S RNY604372S RNY604373S RNY604374S RNY604375S RNY604376S RNY604378S RNY604379S RNY604380S RNY604381S RNY604382S RNY604383S RNY604370S RNY604385S RNY604387S RNY604390S RNY604392S RNY604393S RNY604394S RNY604395S RNY604399S RNY604400S RNY604402S RNY604398S RNY604401S RNY604403S RNY604404S RNY604405S RNY604406S RNY604407S RNY604408S RNY604413S RNY604414S RNY604415S RNY604416S RNY604417S RNY604418S RNY604420S RNY604421S RNY604423S RNY604425S RNY604426S RNY604427S RNY604429S RNY604431S RNY604432S RNY604433S RNY604434S RNY604435S RNY604436S RNY604437S RNY604438S RNY604439S RNY604440S RNY604442S RNY604443S RNY604444S RNY604445S RNY604446S RNY604448S RNY604449S RNY604450S RNY604451S RNY604452S RNY604453S RNY604454S RNY604455S RNY604456S RNY604457S RNY604458S RNY604459S RNY604460S RNY604461S RNY604462S RNY604464S RNY604465S RNY604466S RNY604467S RNY604468S RNY604469S RNY604470S RNY604471S RNY604472S RNY604473S RNY604474S RNY604475S RNY604476S RNY604477S RNY604478S RNY604479S RNY604480S RNY604481S RNY604482S RNY604483S RNY604484S RNY604485S RNY604486S RNY604487S RNY604489S RNY604490S RNY604491S RNY604492S RNY604493S RNY604495S RNY604496S RNY604498S RNY604499S RNY604500S RNY604501S RNY604422S RNY604430S RNY604447S RNY604424S RNY604503S RNY604504S RNY604506S RNY604507S RNY604508S RNY604509S RNY604510S RNY604511S RNY604512S RNY604513S RNY604514S RNY604515S RNY604516S RNY604518S RNY604520S RNY604521S RNY604523S RNY604525S RNY604529S RNY604532S RNY604533S RNY604535S RNY604537S RNY604538S RNY604539S RNY604541S RNY604412S RNY604528S RNY604536S RNY604540S RNY604542S RNY604544S RNY604545S RNY604546S RNY604547S RNY604548S RNY604549S RNY604550S RNY604551S RNY604554S RNY604555S RNY604562S RNY604563S RNY604564S RNY604566S RNY604567S RNY604568S RNY604569S RNY604570S RNY604571S RNY604572S RNY604573S RNY604574S RNY604575S RNY604576S RNY604577S RNY604578S RNY604579S RNY604580S RNY604581S RNY604582S RNY604583S RNY604585S RNY604587S RNY604596S RNY604598S RNY604603S RNY604608S RNY604609S RNY604610S RNY604611S RNY604612S RNY604614S RNY604616S RNY604617S RNY604620S RNY604621S RNY604622S RNY604552S RNY604553S RNY604556S RNY604619S RNY604654S RNY604655S RNY604657S RNY604658S RNY604659S RNY604623S RNY604627S RNY604629S RNY604632S RNY604634S RNY604643S RNY604644S RNY604645S RNY604649S RNY604628S RNY604630S RNY604631S RNY604635S RNY604636S RNY604637S RNY604638S RNY604640S RNY604642S RNY604646S RNY604647S RNY604650S RNY604653S RNY604660S RNY604662S RNY604519S RNY604663S RNY604664S RNY604665S RNY604666S RNY604667S RNY604668S RNY604669S RNY604670S RNY604672S RNY604673S RNY604674S RNY604678S RNY604679S RNY604680S RNY604681S RNY604682S RNY604683S RNY604684S RNY604685S RNY604686S RNY604687S RNY604688S RNY604689S RNY604690S RNY604691S RNY604693S RNY604694S RNY604695S RNY604697S RNY604698S RNY604699S RNY604700S RNY604701S RNY604702S RNY604703S RNY604704S RNY604705S RNY604706S RNY604707S RNY604708S RNY604709S RNY604710S RNY604711S RNY604712S RNY604714S RNY604715S RNY604716S RNY604717S RNY604718S RNY604719S RNY604720S RNY604721S RNY604723S RNY604724S RNY604725S RNY604726S RNY604727S RNY604728S RNY604729S RNY604731S RNY604732S RNY604733S RNY604736S RNY604737S RNY604740S RNY604742S RNY604743S RNY604744S RNY604747S RNY604748S RNY604760S RNY604761S RNY604762S RNY604763S RNY604764S RNY604765S RNY604766S RNY604767S RNY604769S RNY604774S RNY604772S RNY604789S RNY604790S RNY604802S RNY604792S RNY604808S RNY604809S RNY604810S RNY604812S RNY604813S RNY604814S RNY604815S RNY604824S RNY604825S RNY604829S RNY604830S RNY604831S RNY604832S RNY604833S RNY604834S RNY604835S RNY604836S RNY604837S RNY604839S RNY604841S RNY604805S RNY604806S RNY604845S RNY604850S RNY604851S RNY604852S RNY604853S RNY604854S RNY604859S RNY604860S RNY604861S RNY604864S RNY604866S RNY604879S RNY604880S RNY604885S RNY604886S RNY604888S RNY604890S RNY604891S RNY604893S RNY604895S RNY604896S RNY604844S RNY604857S RNY604865S RNY604867S RNY604868S RNY604869S RNY604871S RNY604878S RNY604884S RNY604898S RNY604899S RNY604900S RNY604901S RNY604902S c) W4TR03, serial numbers: RNT200066H RNT200004H RNT200006H RNT200007H RNT200008H RNT200009H RNT200010H RNT200011H RNT200012H RNT200013H RNT200014H RNT200015H RNT200016H RNT200021H RNT200023H RNT200024H RNT200025H RNT200026H RNT200027H RNT200028H RNT200029H RNT200030H RNT200031H RNT200034H RNT200035H RNT200036H RNT200038H RNT200039H RNT200040H RNT200041H RNT200042H RNT200043H RNT200044H RNT200047H RNT200048H RNT200051H RNT200052H RNT200053H RNT200054H RNT200055H RNT200056H RNT200062H RNT200065H RNT200067H RNT200068H RNT200069H RNT200070H RNT200071H RNT200072H RNT200073H RNT200074H RNT200076H RNT200077H RNT200078H RNT200080H RNT200081H RNT200082H RNT200083H RNT200085H RNT200086H RNT200087H RNT200088H RNT200090H RNT200091H RNT200092H RNT200094H RNT200095H RNT200096H RNT200097H RNT200098H RNT200099H RNT200101H RNT200102H RNT200103H RNT200104H RNT200105H RNT200106H RNT200107H RNT200108H RNT200109H RNT200110H RNT200084H RNT200111H RNT200116H RNT200117H RNT200119H RNT200120H RNT200123H RNT200126H RNT200130H RNT200132H RNT200133H RNT200134H RNT200140H RNT200142H RNT200144H RNT200145H RNT200146H RNT200147H RNT200148H RNT200152H RNT200153H RNT200154H RNT200156H RNT200158H RNT200162H RNT200163H RNT200164H RNT200165H RNT200166H RNT200167H RNT200169H RNT200170H RNT200171H RNT200174H RNT200179H RNT200180H RNT200182H RNT200185H RNT200192H RNT200195H RNT200198H RNT200205H RNT200217H RNT200218H RNT200100H RNT200112H RNT200122H RNT200124H RNT200125H RNT200129H RNT200131H RNT200136H RNT200141H RNT200157H RNT200186H RNT200200H RNT200201H RNT200113H RNT200115H RNT200118H RNT200127H RNT200128H RNT200135H RNT200138H RNT200149H RNT200151H RNT200155H RNT200159H RNT200160H RNT200161H RNT200168H RNT200173H RNT200175H RNT200176H RNT200177H RNT200178H RNT200181H RNT200184H RNT200190H RNT200193H RNT200194H RNT200196H RNT200197H RNT200199H RNT200202H RNT200203H RNT200204H RNT200206H RNT200207H RNT200208H RNT200209H RNT200210H RNT200211H RNT200212H RNT200214H RNT200215H RNT200216H RNT200005H RNT200058H RNT200183H RNT200079H RNT200341H RNT200346H RNT200349H RNT200350H RNT200351H RNT200372H RNT200324H RNT200327H RNT200343H RNT200344H RNT200348H RNT200365H RNT200368H RNT200374H RNT200379H RNT200389H RNT200220H RNT200221H RNT200222H RNT200223H RNT200224H RNT200225H RNT200226H RNT200227H RNT200228H RNT200229H RNT200230H RNT200231H RNT200232H RNT200233H RNT200234H RNT200237H RNT200238H RNT200239H RNT200240H RNT200241H RNT200242H RNT200243H RNT200245H RNT200246H RNT200247H RNT200248H RNT200250H RNT200251H RNT200252H RNT200253H RNT200254H RNT200255H RNT200256H RNT200257H RNT200258H RNT200259H RNT200260H RNT200261H RNT200262H RNT200263H RNT200264H RNT200265H RNT200266H RNT200267H RNT200268H RNT200269H RNT200270H RNT200271H RNT200272H RNT200273H RNT200274H RNT200275H RNT200278H RNT200279H RNT200280H RNT200281H RNT200283H RNT200284H RNT200285H RNT200286H RNT200287H RNT200288H RNT200289H RNT200290H RNT200291H RNT200292H RNT200293H RNT200294H RNT200295H RNT200297H RNT200298H RNT200299H RNT200300H RNT200302H RNT200303H RNT200304H RNT200305H RNT200306H RNT200307H RNT200309H RNT200310H RNT200311H RNT200312H RNT200313H RNT200314H RNT200315H RNT200316H RNT200317H RNT200318H RNT200319H RNT200320H RNT200322H RNT200323H RNT200325H RNT200326H RNT200328H RNT200329H RNT200330H RNT200331H RNT200333H RNT200335H RNT200336H RNT200337H RNT200338H RNT200339H RNT200340H RNT200353H RNT200354H RNT200355H RNT200356H RNT200357H RNT200358H RNT200359H RNT200360H RNT200361H RNT200362H RNT200363H RNT200364H RNT200367H RNT200369H RNT200370H RNT200371H RNT200373H RNT200375H RNT200376H RNT200378H RNT200380H RNT200382H RNT200383H RNT200384H RNT200385H RNT200386H RNT200387H RNT200388H RNT200390H RNT200391H RNT200236H RNT200244H RNT200431H RNT200432H RNT200433H RNT200434H RNT200435H RNT200437H RNT200438H RNT200439H RNT200440H RNT200345H RNT200395H RNT200398H RNT200400H RNT200401H RNT200402H RNT200404H RNT200405H RNT200406H RNT200407H RNT200408H RNT200409H RNT200413H RNT200415H RNT200417H RNT200419H RNT200420H RNT200422H RNT200424H RNT200425H RNT200427H RNT200428H RNT200430H RNT200441H RNT200049H RNT200050H RNT200213H RNT200347H RNT200392H RNT200394H RNT200399H RNT200403H RNT200410H RNT200411H RNT200412H RNT200416H RNT200429H RNT200393H RNT200442H RNT200444H RNT200445H RNT200446H RNT200447H RNT200448H RNT200449H RNT200450H RNT200452H RNT200454H RNT200455H RNT200456H RNT200457H RNT200458H RNT200459H RNT200461H RNT200462H RNT200463H RNT200464H RNT200465H RNT200466H RNT200467H RNT200469H RNT200470H RNT200473H RNT200474H RNT200475H RNT200478H RNT200479H RNT200480H RNT200481H RNT200482H RNT200483H RNT200484H RNT200485H RNT200486H RNT200487H RNT200488H RNT200491H RNT200492H RNT200493H RNT200494H RNT200495H RNT200496H RNT200497H RNT200498H RNT200499H RNT200508H RNT200521H RNT200529H RNT200502H RNT200503H RNT200504H RNT200505H RNT200506H RNT200507H RNT200511H RNT200514H RNT200524H RNT200527H RNT200501H RNT200509H RNT200517H RNT200520H RNT200532H RNT200533H RNT200536H RNT200537H RNT200538H RNT200545H RNT200547H RNT200548H RNT200554H RNT200556H RNT200561H RNT200563H RNT200567H RNT200569H RNT200574H RNT200579H RNT200582H RNT200585H RNT200587H RNT200589H RNT200020H RNT200489H RNT200510H RNT200513H RNT200515H RNT200516H RNT200518H RNT200519H RNT200522H RNT200523H RNT200528H RNT200530H RNT200534H RNT200535H RNT200539H RNT200540H RNT200541H RNT200542H RNT200549H RNT200550H RNT200551H RNT200552H RNT200553H RNT200557H RNT200558H RNT200559H RNT200560H RNT200562H RNT200564H RNT200565H RNT200566H RNT200568H RNT200570H RNT200572H RNT200573H RNT200575H RNT200577H RNT200583H RNT200584H RNT200586H RNT200588H RNT200590H RNT200500H RNT200555H RNT200591H RNT200592H RNT200605H RNT200612H RNT200617H RNT200423H RNT200593H RNT200595H RNT200597H RNT200598H RNT200599H RNT200600H RNT200601H RNT200602H RNT200603H RNT200604H RNT200606H RNT200607H RNT200609H RNT200610H RNT200611H RNT200613H RNT200614H RNT200616H RNT200618H RNT200647H RNT200657H RNT200658H RNT200659H RNT200662H RNT200664H RNT200667H RNT200669H RNT200672H RNT200676H RNT200596H RNT200622H RNT200624H RNT200625H RNT200626H RNT200630H RNT200633H RNT200634H RNT200635H RNT200640H RNT200641H RNT200642H RNT200644H RNT200645H RNT200648H RNT200653H RNT200654H RNT200655H RNT200656H RNT200660H RNT200668H RNT200670H RNT200671H RNT200674H RNT200675H RNT200709H RNT200621H RNT200623H RNT200627H RNT200628H RNT200629H RNT200631H RNT200632H RNT200636H RNT200638H RNT200643H RNT200646H RNT200650H RNT200651H RNT200652H RNT200677H RNT200678H RNT200680H RNT200706H RNT200707H RNT200708H RNT200704H RNT200682H RNT200683H RNT200685H RNT200686H RNT200687H RNT200688H RNT200689H RNT200690H RNT200691H RNT200692H RNT200693H RNT200694H RNT200695H RNT200696H RNT200697H RNT200699H RNT200700H RNT200701H RNT200703H RNT200649H RNT200576H RNT200712H RNT200713H RNT200714H RNT200717H RNT200718H RNT200720H RNT200721H RNT200724H RNT200725H RNT200728H RNT200731H RNT200733H RNT200736H RNT200737H RNT200738H RNT200739H RNT200742H RNT200743H RNT200744H RNT200745H RNT200746H RNT200747H RNT200748H RNT200750H RNT200751H RNT200753H RNT200754H RNT200756H RNT200758H RNT200723H RNT200752H RNT200191H RNT200735H RNT200187H RNT200188H RNT200189H RNT200726H RNT200727H RNT200730H RNT200732H RNT200740H RNT200741H RNT200757H RNT200759H RNT200352H RNT200711H RNT200763H RNT200766H RNT200767H RNT200772H RNT200773H RNT200779H RNT200780H RNT200783H RNT200761H RNT200762H RNT200765H RNT200768H RNT200774H RNT200775H RNT200778H RNT200782H RNT200784H RNT200785H RNT200786H RNT200792H RNT200804H RNT200808H RNT200811H RNT200823H RNT200764H RNT200770H RNT200776H RNT200781H RNT200788H RNT200789H RNT200790H RNT200791H RNT200793H RNT200795H RNT200797H RNT200801H RNT200803H RNT200806H RNT200809H RNT200810H RNT200812H RNT200815H RNT200816H RNT200817H RNT200818H RNT200819H RNT200820H RNT200821H RNT200822H RNT200824H RNT200825H RNT200800H RNT200799H RNT200798H RNT200844H RNT200845H RNT200846H RNT200847H RNT200848H RNT200849H RNT200850H RNT200852H RNT200854H RNT200856H RNT200857H RNT200859H RNT200863H RNT200826H RNT200827H RNT200828H RNT200829H RNT200830H RNT200831H RNT200832H RNT200833H RNT200834H RNT200836H RNT200837H RNT200838H RNT200840H RNT200841H RNT200842H RNT200855H RNT200858H RNT200860H RNT200861H RNT200864H RNT200866H RNT200882H RNT200885H RNT200892H RNT200895H RNT200901H RNT200921H RNT200851H RNT200870H RNT200879H RNT200898H RNT200919H RNT200868H RNT200873H RNT200875H RNT200880H RNT200881H RNT200886H RNT200889H RNT200891H RNT200894H RNT200899H RNT200904H RNT200908H RNT200909H RNT200911H RNT200865H RNT200869H RNT200876H
RNT200890H RNT200893H RNT200897H RNT200912H RNT200914H RNT200916H RNT200922H RNT200867H RNT200896H RNT200918H RNT200883H RNT200917H RNT200949H RNT200955H RNT200966H RNT200967H RNT200978H RNT200981H RNT200982H RNT200923H RNT200927H RNT200943H RNT200950H RNT200957H RNT200963H RNT200964H RNT200973H RNT200929H RNT200930H RNT200932H RNT200933H RNT200936H RNT200939H RNT200941H RNT200945H RNT200948H RNT200951H RNT200952H RNT200954H RNT200962H RNT200974H RNT200924H RNT200926H RNT200931H RNT200937H RNT200946H RNT200960H RNT200961H RNT200976H RNT201183H RNT201184H RNT201198H RNT201199H RNT201200H RNT201201H RNT201206H RNT201209H RNT201214H RNT201216H RNT201230H RNT201231H RNT201233H RNT201234H RNT201235H RNT201236H RNT201238H RNT201239H RNT201240H RNT201242H RNT201243H RNT201245H RNT201247H RNT201248H RNT201252H RNT201254H RNT201255H RNT201256H RNT201259H RNT201260H RNT201182H RNT201185H RNT201186H RNT201187H RNT201188H RNT201189H RNT201192H RNT201194H RNT201197H RNT201202H RNT201204H RNT201205H RNT201207H RNT201210H RNT201211H RNT201212H RNT201218H RNT201224H RNT201225H RNT201227H RNT201190H RNT201193H RNT201196H RNT201208H RNT201213H RNT201215H RNT201217H RNT201219H RNT201221H RNT201222H RNT201228H RNT201237H RNT201241H RNT201244H RNT201249H RNT201250H RNT201251H RNT201257H RNT201268H RNT201261H RNT201262H RNT201263H RNT201264H RNT201265H RNT201266H RNT201267H RNT201269H RNT201270H RNT201271H RNT201273H RNT201275H RNT201276H RNT201277H RNT201279H RNT201281H RNT201282H RNT201284H RNT201285H RNT201286H RNT201287H RNT201289H RNT201291H RNT201292H RNT201309H RNT201310H RNT201313H RNT201314H RNT201315H RNT201316H RNT201317H RNT201318H RNT201319H RNT201320H RNT201321H RNT201322H RNT201323H RNT201324H RNT201325H RNT201326H RNT201327H RNT201328H RNT200716H RNT200871H RNT200872H RNT200874H RNT200878H RNT200887H RNT200888H RNT200913H RNT200915H RNT200925H RNT200969H RNT201229H RNT201293H RNT201296H RNT201297H RNT201299H RNT201302H RNT201303H RNT201304H RNT201305H RNT201307H RNT201308H RNT201311H RNT201331H RNT201332H RNT201333H RNT201336H RNT201337H RNT201338H RNT201339H RNT201340H RNT201341H RNT201343H RNT201348H RNT201349H RNT201350H RNT201294H RNT201345H RNT201346H RNT201351H RNT200920H RNT201354H RNT201362H RNT201386H RNT201388H RNT201398H RNT201383H RNT201385H RNT201071H RNT201358H RNT201363H RNT201366H RNT201370H RNT201372H RNT201373H RNT201374H RNT201381H RNT201384H RNT201389H RNT201393H RNT201394H RNT201399H RNT201403H RNT201408H RNT201409H RNT201353H RNT201355H RNT201360H RNT201367H RNT201371H RNT201375H RNT201382H RNT201387H RNT201391H RNT201392H RNT201396H RNT201400H RNT201401H RNT201402H RNT201404H RNT201405H RNT201377H RNT201380H RNT201412H RNT201421H RNT201423H RNT201424H RNT201431H RNT201438H RNT201450H RNT201451H RNT201463H RNT201466H RNT201330H RNT201357H RNT201368H RNT201369H RNT201376H RNT201379H RNT201418H RNT201432H RNT201435H RNT201441H RNT201442H RNT201446H RNT201454H RNT201457H RNT201459H RNT201416H RNT201417H RNT201434H RNT201443H RNT201444H RNT201452H RNT201413H RNT201422H RNT201426H RNT201433H RNT201437H RNT201445H RNT201455H RNT201464H RNT201469H RNT201411H RNT201429H RNT201436H RNT201470H RNT201472H RNT201473H RNT201474H RNT201475H RNT201476H RNT201480H RNT201481H RNT201487H RNT201496H RNT201497H RNT201498H RNT201499H RNT201504H RNT201506H RNT201510H RNT201514H RNT201519H RNT201520H RNT201521H RNT201523H RNT201524H RNT201525H RNT201527H RNT201528H RNT201414H RNT201415H RNT201427H RNT201428H RNT201447H RNT201449H RNT201456H RNT201462H RNT201467H RNT201471H RNT201477H RNT201478H RNT201482H RNT201483H RNT201484H RNT201485H RNT201486H RNT201488H RNT201490H RNT201491H RNT201492H RNT201493H RNT201495H RNT201500H RNT201501H RNT201502H RNT201503H RNT201505H RNT201507H RNT201512H RNT201513H RNT201515H RNT201516H RNT201517H RNT201522H RNT201526H RNT201530H RNT201531H RNT201532H RNT201535H RNT201539H RNT201540H RNT201542H RNT201543H RNT201545H RNT201546H RNT201547H RNT201548H RNT201549H RNT201551H RNT201554H RNT201556H RNT201557H RNT201558H RNT201560H RNT201564H RNT201565H RNT201567H RNT201568H RNT201569H RNT201575H RNT201577H RNT201361H RNT201533H RNT201534H RNT201536H RNT201537H RNT201538H RNT201550H RNT201552H RNT201555H RNT201562H RNT201566H RNT201570H RNT201571H RNT201572H RNT201573H RNT201574H RNT201576H RNT201578H RNT201591H RNT201597H RNT201600H RNT201603H RNT201604H RNT201606H RNT201607H RNT201608H RNT201609H RNT201610H RNT201611H RNT201612H RNT201613H RNT201615H RNT201617H RNT201605H RNT201563H RNT201579H RNT201582H RNT201586H RNT201587H RNT201593H RNT201594H RNT201598H RNT201599H RNT201580H RNT201581H RNT201584H RNT201585H RNT201589H RNT201649H RNT201342H RNT201618H RNT201619H RNT201620H RNT201621H RNT201622H RNT201624H RNT201625H RNT201627H RNT201629H RNT201630H RNT201631H RNT201632H RNT201633H RNT201634H RNT201635H RNT201636H RNT201637H RNT201638H RNT201639H RNT201640H RNT201642H RNT201644H RNT201646H RNT201647H RNT201650H RNT201651H RNT201654H RNT201655H RNT201656H RNT201623H RNT201628H RNT201641H RNT201645H RNT201648H RNT201590H RNT201614H RNT201195H RNT201420H RNT201489H RNT201592H RNT201602H RNT201657H RNT201662H RNT201686H RNT201698H RNT201701H RNT201703H RNT201708H RNT201728H RNT201739H RNT201659H RNT201660H RNT201661H RNT201663H RNT201664H RNT201665H RNT201668H RNT201669H RNT201671H RNT201672H RNT201673H RNT201674H RNT201675H RNT201677H RNT201678H RNT201679H RNT201681H RNT201682H RNT201683H RNT201687H RNT201688H RNT201689H RNT201690H RNT201691H RNT201692H RNT201693H RNT201695H RNT201696H RNT201697H RNT201700H RNT201702H RNT201704H RNT201705H RNT201709H RNT201710H RNT201711H RNT201712H RNT201713H RNT201714H RNT201715H RNT201716H RNT201717H RNT201720H RNT201722H RNT201723H RNT201727H RNT201732H RNT201733H RNT201737H RNT201742H RNT201747H RNT201750H RNT201752H RNT201754H RNT201595H RNT201694H RNT201721H RNT201756H RNT201652H RNT201719H RNT201724H RNT201725H RNT201734H RNT201736H RNT201738H RNT201740H RNT201743H RNT201744H RNT201749H RNT201751H RNT201753H RNT201718H RNT201748H RNT201731H RNT201448H RNT201761H RNT201763H RNT201765H RNT201766H RNT201770H RNT201771H RNT201773H RNT201775H RNT201777H RNT201779H RNT201780H RNT201782H RNT201783H RNT201785H RNT201786H RNT201787H RNT201789H RNT201792H RNT201794H RNT201796H RNT201803H RNT201806H RNT201809H RNT201810H RNT201760H RNT201764H RNT201768H RNT201769H RNT201776H RNT201784H RNT201790H RNT201791H RNT201799H RNT201801H RNT201802H RNT201804H RNT201805H RNT201807H RNT201808H RNT201812H RNT201813H RNT201757H RNT201772H RNT201774H RNT201778H RNT201781H RNT201788H RNT201811H RNT201887H RNT201888H RNT201892H RNT201893H RNT201897H RNT201898H RNT201902H RNT201890H RNT201891H RNT201896H RNT201819H RNT201825H RNT201834H RNT201836H RNT201839H RNT201842H RNT201843H RNT201844H RNT201851H RNT201853H RNT201854H RNT201817H RNT201818H RNT201821H RNT201822H RNT201823H RNT201824H RNT201826H RNT201832H RNT201833H RNT201835H RNT201837H RNT201838H RNT201840H RNT201841H RNT201845H RNT201847H RNT201849H RNT201850H RNT201943H RNT201944H RNT201956H RNT201962H RNT201964H RNT201968H RNT201979H RNT201987H RNT201990H RNT201991H RNT201999H RNT202000H RNT201942H RNT201946H RNT201947H RNT201948H RNT201949H RNT201950H RNT201951H RNT201952H RNT201953H RNT201954H RNT201957H RNT201958H RNT201959H RNT201960H RNT201963H RNT201966H RNT201967H RNT201969H RNT201970H RNT201971H RNT201972H RNT201973H RNT201974H RNT201975H RNT201977H RNT201978H RNT201980H RNT201981H RNT201982H RNT201983H RNT201984H RNT201985H RNT201986H RNT201988H RNT201989H RNT201992H RNT201993H RNT201994H RNT201995H RNT201996H RNT201998H RNT202001H RNT202002H RNT201916H RNT201918H RNT201927H RNT201906H RNT201907H RNT201909H RNT201910H RNT201914H RNT201917H RNT201919H RNT201920H RNT201922H RNT201923H RNT201926H RNT201931H RNT201933H RNT201936H RNT201937H RNT201941H RNT201912H RNT201934H RNT201938H RNT201955H RNT201908H RNT201939H RNT201828H RNT201820H RNT202004H RNT202005H RNT202013H RNT202023H RNT202024H RNT202025H RNT202031H RNT202036H RNT202040H RNT202043H RNT202048H RNT202065H RNT202066H RNT202077H RNT202079H RNT202081H RNT202003H RNT202027H RNT202029H RNT202037H RNT202053H RNT202069H RNT202075H RNT202009H RNT202026H RNT202028H RNT202030H RNT202060H RNT202061H RNT202073H RNT202008H RNT202018H RNT202034H RNT202041H RNT202070H RNT202071H RNT202074H RNT202014H RNT202021H RNT202033H RNT202035H RNT202044H RNT202047H RNT202052H RNT202056H RNT202057H RNT202006H RNT202007H RNT202010H RNT202015H RNT202016H RNT202017H RNT202019H RNT202020H RNT202022H RNT202045H RNT202046H RNT202049H RNT202055H RNT202058H RNT202059H RNT202063H RNT202072H RNT202076H RNT202078H RNT202080H RNT202094H RNT202100H RNT202119H RNT202130H RNT202133H RNT202135H RNT202137H RNT201440H RNT202082H RNT202091H RNT202093H RNT202096H RNT202098H RNT202099H RNT202101H RNT202103H RNT202108H RNT202113H RNT202118H RNT202121H RNT202123H RNT202124H RNT202129H RNT202139H RNT202083H RNT202084H RNT202105H RNT202109H RNT202110H RNT202115H RNT202122H RNT202131H RNT201406H RNT202085H RNT202086H RNT202089H RNT202090H RNT202097H RNT202102H RNT202104H RNT202106H RNT202107H RNT202111H RNT202112H RNT202114H RNT202120H RNT202125H RNT202127H RNT202134H RNT202136H RNT202142H RNT202116H RNT202128H RNT202087H RNT202177H RNT202178H RNT202180H RNT202181H RNT202183H RNT202184H RNT202187H RNT202188H RNT202190H RNT202195H RNT202197H RNT202199H RNT202200H RNT202201H RNT202202H RNT202204H RNT202205H RNT202208H RNT202210H RNT202212H RNT202214H RNT202215H RNT202217H RNT202218H RNT202220H RNT202221H RNT202225H RNT202226H RNT202227H RNT202230H RNT202231H RNT202233H RNT202234H RNT202235H RNT202237H RNT202238H RNT202239H RNT202240H RNT202241H RNT202182H RNT202191H RNT202198H RNT202206H RNT202223H RNT202224H RNT202228H RNT202232H RNT202236H RNT202088H RNT202147H RNT202148H RNT202150H RNT202153H RNT202155H RNT202156H RNT202157H RNT202158H RNT202160H RNT202161H RNT202162H RNT202163H RNT202168H RNT202169H RNT202170H RNT202172H RNT202179H RNT202186H RNT202192H RNT202196H RNT202203H RNT202207H RNT202209H RNT202211H RNT202213H RNT202216H RNT202219H RNT202247H RNT202248H RNT202250H RNT202251H RNT202252H RNT202255H RNT202256H RNT202258H RNT202259H RNT202260H RNT202261H RNT202262H RNT202267H RNT202268H RNT202270H RNT202271H RNT202275H RNT202278H RNT202282H RNT202283H RNT202285H RNT202286H RNT202288H RNT202289H RNT202143H RNT202144H RNT202151H RNT202154H RNT202159H RNT202165H RNT202167H RNT202171H RNT202173H RNT202175H RNT202243H RNT202244H RNT202253H RNT202254H RNT202257H RNT202263H RNT202264H RNT202265H RNT202272H RNT202273H RNT202279H RNT202281H RNT202287H RNT202290H RNT202269H RNT202274H RNT201746H RNT201735H RNT202229H RNT202194H RNT202291H RNT200796H RNT202266H RNT201461H RNT202305H RNT202327H RNT202328H RNT202331H RNT202349H RNT202352H RNT202355H RNT202361H RNT202362H RNT202363H RNT202366H RNT202369H RNT202373H RNT202380H RNT202385H RNT202422H RNT202304H RNT202306H RNT202310H RNT202313H RNT202314H RNT202317H RNT202319H RNT202322H RNT202324H RNT202326H RNT202336H RNT202346H RNT202347H
RNT202351H RNT202353H RNT202354H RNT202357H RNT202359H RNT202368H RNT202370H RNT202371H RNT202372H RNT202374H RNT202376H RNT202377H RNT202379H RNT202381H RNT202382H RNT202383H RNT202386H RNT202387H RNT202308H RNT202309H RNT202318H RNT202321H RNT202329H RNT202333H RNT202367H RNT202415H RNT202416H RNT202420H RNT202425H RNT202426H RNT202429H RNT202437H RNT202445H RNT202448H RNT202451H RNT202452H RNT202454H RNT202459H RNT202311H RNT202316H RNT202323H RNT202330H RNT202398H RNT202412H RNT202413H RNT202414H RNT202417H RNT202418H RNT202421H RNT202423H RNT202424H RNT202428H RNT202430H RNT202431H RNT202433H RNT202434H RNT202436H RNT202440H RNT202442H RNT202443H RNT202444H RNT202447H RNT202449H RNT202450H RNT202453H RNT202455H RNT202457H RNT202458H RNT202460H RNT202461H RNT202463H RNT202464H RNT202465H RNT202466H RNT202467H RNT202468H RNT202469H RNT202470H RNT202471H RNT202472H RNT202473H RNT202474H RNT202475H RNT202419H RNT202481H RNT202375H RNT202389H RNT202390H RNT202391H RNT202393H RNT202395H RNT202397H RNT202399H RNT202404H RNT202407H RNT202409H RNT202410H RNT202295H RNT202296H RNT202297H RNT202302H RNT202303H RNT202334H RNT202344H RNT202345H RNT202489H RNT202496H RNT202502H RNT202504H RNT202513H RNT202526H RNT202530H RNT202542H RNT202553H RNT202293H RNT202298H RNT202300H RNT202301H RNT202335H RNT202337H RNT202339H RNT202342H RNT202343H RNT202498H RNT202312H RNT202356H RNT202360H RNT202402H RNT202403H RNT202406H RNT202476H RNT202478H RNT202479H RNT202480H RNT202482H RNT202483H RNT202486H RNT202487H RNT202488H RNT202490H RNT202491H RNT202492H RNT202497H RNT202499H RNT202500H RNT202501H RNT202503H RNT202505H RNT202506H RNT202507H RNT202508H RNT202509H RNT202510H RNT202512H RNT202515H RNT202518H RNT202519H RNT202520H RNT202521H RNT202523H RNT202524H RNT202525H RNT202527H RNT202528H RNT202531H RNT202532H RNT202533H RNT202534H RNT202535H RNT202536H RNT202537H RNT202538H RNT202540H RNT202541H RNT202543H RNT202545H RNT202546H RNT202547H RNT202548H RNT202549H RNT202552H RNT202554H RNT202555H RNT202556H RNT202557H RNT202558H RNT202559H RNT202560H RNT202562H RNT202563H RNT202564H RNT202565H RNT202566H RNT202567H RNT202568H RNT202569H RNT202570H RNT202571H RNT202572H RNT202573H RNT202574H RNT202575H RNT202577H RNT202578H RNT202579H RNT202580H RNT202582H RNT202584H RNT202586H RNT202587H RNT202588H RNT202589H RNT202591H RNT202592H RNT202593H RNT202594H RNT202595H RNT202596H RNT202597H RNT202598H RNT202599H RNT202600H RNT202601H RNT202602H RNT202603H RNT202604H RNT202606H RNT202607H RNT202608H RNT202609H RNT202610H RNT202611H RNT202613H RNT202614H RNT202615H RNT202616H RNT202617H RNT202619H RNT202620H RNT202621H RNT202622H RNT202623H RNT202624H RNT202625H RNT202626H RNT202627H RNT202628H RNT202629H RNT202630H RNT202631H RNT202633H RNT202634H RNT202635H RNT202637H RNT202638H RNT202639H RNT202640H RNT202641H RNT202643H RNT202644H RNT202645H RNT202646H RNT202647H RNT202551H RNT202576H RNT202658H RNT202687H RNT202695H RNT202696H RNT202698H RNT202699H RNT202700H RNT202701H RNT202400H RNT202411H RNT202494H RNT202517H RNT202650H RNT202654H RNT202655H RNT202656H RNT202657H RNT202661H RNT202662H RNT202665H RNT202666H RNT202669H RNT202670H RNT202673H RNT202674H RNT202676H RNT202677H RNT202679H RNT202682H RNT202685H RNT202686H RNT202693H RNT202694H RNT202697H RNT202702H RNT202659H RNT202648H RNT202649H RNT202652H RNT202653H RNT202660H RNT202663H RNT202664H RNT202667H RNT202668H RNT202671H RNT202678H RNT202681H RNT202684H RNT202690H RNT202692H RNT202581H RNT202605H RNT202612H RNT202689H RNT201583H RNT202651H RNT202704H RNT202708H RNT202713H RNT202715H RNT202716H RNT202717H RNT202719H RNT202720H RNT202721H RNT202703H RNT202706H RNT202707H RNT202709H RNT202710H RNT202712H RNT202725H RNT202705H RNT202711H RNT202722H RNT202068H RNT201899H RNT200940H RNT200942H RNT200965H RNT202741H RNT202742H RNT202743H RNT202744H RNT202746H RNT202747H RNT202750H RNT202759H RNT202728H RNT202729H RNT202730H RNT202732H RNT202734H RNT202735H RNT202736H RNT202738H RNT202751H RNT202752H RNT202755H RNT202733H RNT202737H RNT202739H RNT202740H RNT202745H RNT202748H RNT202753H RNT202754H RNT202756H RNT202758H RNT202432H RNT202618H RNT202766H RNT202845H RNT202851H RNT202876H RNT202878H RNT202883H RNT202770H RNT202835H RNT202879H RNT202881H RNT202760H RNT202763H RNT202764H RNT202765H RNT202768H RNT202808H RNT202825H RNT202847H RNT202848H RNT202849H RNT202852H RNT202877H RNT202880H RNT202884H RNT202769H RNT202793H RNT202814H RNT202842H RNT202844H RNT202846H RNT202882H RNT202885H RNT202886H RNT202803H RNT202857H RNT202861H RNT202864H RNT202868H RNT202785H RNT202786H RNT202787H RNT202788H RNT202790H RNT202791H RNT202794H RNT202796H RNT202797H RNT202798H RNT202800H RNT202802H RNT202806H RNT202807H RNT202822H RNT202823H RNT202831H RNT202832H RNT202834H RNT202841H RNT202853H RNT202855H RNT202860H RNT202862H RNT202784H RNT202801H RNT202805H RNT202810H RNT202815H RNT202820H RNT202824H RNT202826H RNT202827H RNT202850H RNT202858H RNT202865H RNT202772H RNT202774H RNT202781H RNT202854H RNT202866H RNT202870H RNT202894H RNT202896H RNT202897H RNT202898H RNT202908H RNT202909H RNT202914H RNT202915H RNT202916H RNT202917H RNT202918H RNT202930H RNT202933H RNT202938H RNT202941H RNT202782H RNT202771H RNT202777H RNT202778H RNT202780H RNT202873H RNT202890H RNT202899H RNT202902H RNT202911H RNT202913H RNT202919H RNT202922H RNT202924H RNT202925H RNT202927H RNT202928H RNT202929H RNT202932H RNT202935H RNT202936H RNT202937H RNT202939H RNT202940H RNT202945H RNT202775H RNT202776H RNT202888H RNT202931H RNT202944H RNT202819H RNT202875H RNT202903H RNT202912H RNT202921H RNT202839H RNT202869H RNT202833H RNT202837H RNT202972H RNT202978H RNT202983H RNT203000H RNT202792H RNT202969H RNT202970H RNT202982H RNT203001H RNT202986H RNT202995H RNT202996H RNT202999H RNT202950H RNT202951H RNT202953H RNT202955H RNT202956H RNT202957H RNT202958H RNT202959H RNT202961H RNT202962H RNT202963H RNT202965H RNT202976H RNT202977H RNT202988H RNT202991H RNT202994H RNT203003H RNT203007H RNT203008H RNT203009H RNT203010H RNT203011H RNT203013H RNT203017H RNT203020H RNT203021H RNT203024H RNT203025H RNT203026H RNT203027H RNT203030H RNT203036H RNT203039H RNT203041H RNT203042H RNT203046H RNT203048H RNT203051H RNT203052H RNT203053H RNT203056H RNT203058H RNT203060H RNT203062H RNT203066H RNT203069H RNT203070H RNT203072H RNT203077H RNT203079H RNT203083H RNT203084H RNT203086H RNT203088H RNT203089H RNT203090H RNT203091H RNT202952H RNT202954H RNT202960H RNT202964H RNT202966H RNT202967H RNT202973H RNT202974H RNT202975H RNT202980H RNT202981H RNT202984H RNT202987H RNT202989H RNT202990H RNT202992H RNT202993H RNT202997H RNT203002H RNT203004H RNT203006H RNT203015H RNT203018H RNT203019H RNT203022H RNT203028H RNT203032H RNT203035H RNT203054H RNT203063H RNT203064H RNT203065H RNT203067H RNT203068H RNT203073H RNT203075H RNT203076H RNT203078H RNT203081H RNT203087H RNT203092H RNT203094H RNT203043H RNT203049H RNT202892H RNT202968H RNT203037H RNT203040H RNT203047H RNT203059H RNT203080H RNT202817H RNT203120H RNT203122H RNT203121H RNT203128H RNT203129H RNT203131H RNT203133H RNT203138H RNT203144H RNT202836H RNT203119H RNT203124H RNT203125H RNT203126H RNT203127H RNT203130H RNT203135H RNT203140H RNT203155H RNT203167H RNT203095H RNT203096H RNT203104H RNT203106H RNT203108H RNT203116H RNT203136H RNT203147H RNT203148H RNT203151H RNT203152H RNT203153H RNT203154H RNT203157H RNT203158H RNT203161H RNT203162H RNT203166H RNT203168H RNT203099H RNT203101H RNT203102H RNT203107H RNT203109H RNT203110H RNT203111H RNT203112H RNT203114H RNT203115H RNT203117H RNT203118H RNT203150H RNT203159H RNT203163H RNT203160H RNT203113H RNT203074H RNT203098H RNT203165H RNT203174H RNT203177H RNT203183H RNT203185H RNT203186H RNT203198H RNT203202H RNT203203H RNT203181H RNT203182H RNT203192H RNT203195H RNT203196H RNT203170H RNT203187H RNT203193H RNT203204H RNT203201H RNT203210H RNT203221H RNT203240H RNT203175H RNT203176H RNT203178H RNT203179H RNT203184H RNT203188H RNT203190H RNT203194H RNT203197H RNT203205H RNT203209H RNT203211H RNT203213H RNT203217H RNT203222H RNT203225H RNT203239H RNT203244H RNT203245H RNT203249H RNT203171H RNT203214H RNT203215H RNT203226H RNT203227H RNT203228H RNT203229H RNT203230H RNT203231H RNT203232H RNT203233H RNT203241H RNT203242H RNT203247H RNT203248H RNT203250H RNT202284H RNT203207H RNT203216H RNT203223H RNT203235H RNT203238H RNT203246H RNT203219H RNT203224H RNT202811H RNT203218H RNT203220H RNT203286H RNT203295H RNT203297H RNT203300H RNT203301H RNT203303H RNT203304H RNT203285H RNT203288H RNT203290H RNT203294H RNT203298H RNT203302H RNT203287H RNT203289H RNT203296H RNT203305H RNT203256H RNT203258H RNT203259H RNT203263H RNT203265H RNT203276H RNT203277H RNT203278H RNT203254H RNT203257H RNT203260H RNT203261H RNT203262H RNT203267H RNT203268H RNT203269H RNT203270H RNT203271H RNT203272H RNT203273H RNT203280H RNT203281H RNT203283H RNT203306H RNT202979H RNT203071H RNT203134H RNT203137H RNT203252H RNT203284H RNT203307H RNT203310H RNT203311H RNT203312H RNT203313H RNT203314H RNT203315H RNT203324H RNT203327H RNT203329H RNT203330H RNT203335H RNT203338H RNT203339H RNT203123H RNT203143H RNT203156H RNT203251H RNT203253H RNT203266H RNT203274H RNT203282H RNT203308H RNT203316H RNT203328H RNT203332H RNT203337H RNT203325H RNT203345H RNT203351H RNT203344H RNT203349H RNT203350H RNT203354H RNT203359H RNT203360H RNT203343H RNT203346H RNT203347H RNT203348H RNT203356H RNT203357H RNT203361H RNT203363H RNT203326H RNT203373H RNT203352H RNT203355H RNT203362H RNT203364H RNT203365H RNT203366H RNT203367H RNT203368H RNT203370H RNT203375H RNT203376H RNT203358H RNT203380H RNT203381H RNT203383H RNT201869H RNT203236H RNT203379H RNT201859H RNT201860H RNT201875H RNT201880H RNT203318H RNT201855H RNT201858H RNT201861H RNT201878H RNT201868H RNT203387H RNT203394H RNT203400H RNT203384H RNT203385H RNT203405H RNT201879H RNT203386H RNT203389H RNT203391H RNT203395H RNT203396H RNT203397H RNT203398H RNT203404H RNT201870H RNT201884H RNT201904H RNT203388H RNT203392H RNT203393H RNT203399H RNT203406H RNT203409H RNT203418H RNT203419H RNT203422H RNT203423H RNT203412H RNT201864H RNT203407H RNT203408H RNT203410H RNT203411H RNT203413H RNT203414H RNT203416H RNT203417H RNT203420H RNT203421H RNT203424H RNT203425H RNT203426H RNT203427H RNT203428H RNT203429H RNT203402H RNT203434H RNT203438H RNT203441H RNT203452H RNT203457H RNT203460H RNT203432H RNT203436H RNT203459H RNT203466H RNT203484H RNT203431H RNT203443H RNT203447H RNT203449H RNT203451H RNT203455H RNT203456H RNT203461H RNT203473H RNT203485H RNT203480H RNT203483H RNT202276H RNT203433H RNT203435H RNT203444H RNT203448H RNT203453H RNT203458H RNT203467H RNT201857H RNT201901H RNT201905H RNT202485H RNT203491H RNT203493H RNT203496H RNT203497H RNT203503H RNT203504H RNT203508H RNT203514H RNT203515H RNT203516H RNT203517H RNT203518H RNT203520H RNT203523H RNT203524H RNT203525H RNT203527H RNT203529H RNT203531H RNT203532H RNT203533H RNT203539H RNT203540H RNT203542H RNT203546H RNT203547H RNT203488H RNT203506H RNT203507H RNT203512H RNT203526H RNT203534H RNT203536H RNT203537H RNT203543H
RNT203544H RNT203554H RNT203562H RNT203566H RNT203567H RNT203572H RNT203573H RNT203574H RNT203579H RNT203580H RNT203582H RNT203584H RNT203586H RNT203589H RNT203598H RNT203602H RNT203622H RNT203626H RNT203628H RNT203634H RNT203494H RNT203499H RNT203500H RNT203501H RNT203505H RNT203509H RNT203511H RNT203521H RNT203528H RNT203541H RNT203545H RNT203549H RNT203550H RNT203551H RNT203557H RNT203564H RNT203568H RNT203570H RNT203583H RNT203585H RNT203588H RNT203596H RNT203617H RNT203627H RNT203639H RNT203552H RNT203556H RNT203560H RNT203561H RNT203563H RNT203581H RNT203587H RNT203592H RNT203593H RNT203594H RNT203599H RNT203603H RNT203605H RNT203611H RNT203612H RNT203615H RNT203618H RNT203619H RNT203621H RNT203635H RNT203644H RNT203553H RNT203558H RNT203559H RNT203575H RNT203590H RNT203591H RNT203608H RNT203614H RNT203620H RNT203636H RNT203641H RNT203600H RNT203610H RNT203654H RNT203666H RNT203714H RNT201872H RNT202539H RNT203650H RNT203651H RNT203662H RNT203664H RNT203669H RNT203673H RNT203682H RNT203683H RNT203684H RNT203685H RNT203687H RNT203688H RNT203689H RNT203692H RNT203695H RNT203696H RNT203697H RNT203700H RNT203705H RNT203706H RNT203707H RNT203709H RNT203711H RNT203719H RNT203722H RNT203725H RNT203729H RNT203731H RNT203732H RNT203736H RNT203737H RNT201867H RNT203647H RNT203648H RNT203653H RNT203663H RNT203668H RNT203672H RNT203674H RNT203675H RNT203686H RNT203690H RNT203698H RNT203699H RNT203703H RNT203710H RNT203713H RNT201862H RNT201871H RNT201876H RNT203595H RNT203638H RNT203646H RNT203649H RNT203655H RNT203656H RNT203660H RNT203661H RNT203667H RNT203676H RNT203677H RNT203678H RNT203679H RNT203701H RNT203702H RNT203712H RNT203735H RNT203623H RNT203578H RNT203630H RNT203597H RNT203750H RNT203768H RNT203771H RNT203784H RNT203758H RNT203763H RNT203778H RNT203788H RNT203790H RNT201873H RNT203741H RNT203779H RNT203780H RNT203787H RNT201877H RNT201886H RNT203637H RNT203693H RNT203745H RNT203749H RNT203759H RNT203760H RNT203762H RNT203774H RNT203791H RNT203793H RNT203740H RNT203764H RNT203766H RNT203770H RNT203773H RNT203748H RNT203751H RNT203753H RNT201882H RNT201900H RNT203752H RNT203795H RNT203754H RNT203755H RNT203756H RNT203743H RNT203800H RNT203799H RNT203807H RNT203808H RNT203818H RNT203819H RNT203824H RNT203825H RNT203829H RNT203830H RNT203847H RNT203804H RNT203805H RNT203806H RNT203809H RNT203814H RNT203817H RNT203821H RNT203827H RNT203831H RNT203840H RNT203852H RNT202947H RNT203462H RNT203801H RNT203802H RNT203803H RNT203813H RNT203815H RNT203820H RNT203826H RNT203835H RNT203836H RNT203837H RNT203841H RNT203850H RNT203853H RNT203828H RNT203832H RNT203833H RNT203785H RNT203812H RNT203846H RNT203848H RNT203864H RNT203839H RNT203844H RNT203870H RNT203849H RNT203866H RNT203871H RNT203867H RNT203796H RNT203855H RNT203863H RNT203874H RNT203856H RNT203862H RNT203873H RNT203875H RNT203876H RNT203881H RNT203882H RNT203885H RNT203889H RNT203891H RNT203896H RNT203879H RNT203880H RNT203883H RNT203886H RNT203887H RNT203888H RNT203895H RNT203897H RNT203898H RNT203890H RNT203865H RNT203899H RNT203900H RNT203901H RNT203908H RNT203914H RNT203915H RNT203918H RNT203923H RNT203929H RNT203932H RNT203933H RNT203935H RNT203902H RNT203905H RNT203906H RNT203910H RNT203913H RNT203917H RNT203921H RNT203922H RNT203924H RNT203925H RNT203926H RNT203927H RNT203928H RNT203931H RNT203934H RNT203936H RNT203940H RNT203941H RNT203911H RNT203912H RNT203955H RNT203961H RNT203981H RNT203991H RNT203992H RNT204001H RNT204009H RNT204012H RNT204014H RNT204022H RNT204027H RNT204038H RNT203903H RNT203919H RNT203938H RNT203963H RNT203988H RNT203990H RNT203993H RNT203994H RNT204004H RNT204008H RNT204034H RNT203966H RNT204021H RNT204035H RNT204003H RNT204015H RNT204016H RNT204030H RNT203943H RNT203957H RNT203986H RNT203987H RNT204019H RNT203950H RNT203972H RNT203907H RNT203942H RNT203944H RNT203948H RNT203951H RNT203953H RNT203954H RNT203956H RNT203959H RNT203964H RNT203965H RNT203968H RNT203973H RNT203974H RNT203979H RNT203983H RNT203984H RNT203985H RNT203995H RNT203997H RNT203998H RNT204002H RNT204005H RNT204006H RNT204010H RNT204011H RNT204018H RNT204023H RNT204031H RNT204032H RNT203945H RNT203946H RNT203947H RNT203949H RNT203960H RNT203962H RNT203975H RNT203978H RNT203989H RNT204000H RNT204007H RNT204025H RNT204028H RNT204036H RNT203930H RNT204063H RNT204064H RNT204067H RNT204070H RNT204072H RNT204078H RNT204079H RNT204080H RNT204081H RNT204084H RNT204086H RNT204138H RNT204164H RNT204171H RNT204178H RNT204173H RNT204181H RNT203976H RNT204044H RNT204047H RNT204051H RNT204057H RNT204062H RNT204069H RNT204140H RNT204154H RNT204155H RNT204159H RNT204160H RNT204166H RNT204168H RNT204172H RNT204175H RNT204177H RNT204186H RNT204191H RNT204192H RNT204193H RNT204194H RNT204195H RNT204196H RNT204197H RNT204199H RNT204200H RNT204201H RNT204202H RNT204208H RNT204214H RNT204218H RNT204221H RNT204223H RNT204120H RNT204124H RNT204126H RNT204143H RNT204147H RNT204148H RNT204162H RNT204167H RNT204180H RNT204183H RNT204189H RNT204204H RNT204209H RNT204217H RNT204049H RNT204053H RNT204058H RNT204089H RNT204090H RNT204091H RNT204114H RNT204119H RNT204127H RNT204132H RNT204133H RNT204061H RNT204065H RNT204071H RNT204074H RNT204075H RNT204077H RNT204082H RNT204085H RNT204100H RNT204101H RNT204104H RNT204108H RNT204110H RNT204125H RNT204137H RNT204153H RNT204163H RNT204169H RNT204220H RNT203937H RNT203969H RNT204040H RNT204042H RNT204046H RNT204048H RNT204050H RNT204052H RNT204055H RNT204056H RNT204060H RNT204083H RNT204087H RNT204093H RNT204094H RNT204103H RNT204105H RNT204111H RNT204115H RNT204116H RNT204117H RNT204130H RNT204142H RNT204144H RNT204145H RNT204170H RNT204179H RNT204182H RNT204184H RNT204185H RNT204187H RNT204149H RNT204207H RNT204215H RNT204222H RNT204227H RNT204228H RNT204232H RNT204233H RNT204068H RNT204123H RNT204206H RNT204216H RNT204219H RNT204225H RNT204226H RNT204229H RNT204230H RNT204231H RNT204106H RNT204107H RNT204113H RNT204121H RNT204122H RNT204128H RNT204150H RNT204151H RNT204152H RNT204249H RNT204252H RNT204255H RNT204256H RNT204258H RNT204259H RNT204261H RNT204262H RNT204263H RNT204264H RNT204265H RNT204267H RNT204269H RNT204270H RNT204271H RNT204272H RNT204274H RNT204276H RNT204277H RNT204278H RNT204280H RNT204281H RNT204282H RNT204283H RNT204253H RNT204268H RNT204273H RNT204212H RNT204241H RNT204243H RNT204244H RNT204246H RNT204250H RNT204251H RNT204254H RNT204260H RNT204224H RNT204237H RNT204248H RNT204242H RNT204257H RNT204240H RNT204238H RNT204239H RNT204247H RNT204157H RNT204109H RNT204284H RNT204287H RNT204288H RNT204289H RNT204290H RNT204293H RNT204295H RNT204298H RNT204300H RNT204301H RNT204302H RNT204303H RNT204304H RNT204306H RNT204307H RNT204309H RNT204312H RNT204313H RNT204316H RNT204318H RNT204321H RNT204322H RNT204324H RNT204326H RNT204329H RNT204330H RNT204332H RNT204333H RNT204336H RNT204338H RNT204339H RNT204353H RNT204355H RNT204357H RNT204358H RNT204364H RNT204371H RNT204374H RNT204375H RNT204376H RNT204387H RNT204388H RNT204389H RNT204391H RNT204395H RNT204396H RNT204397H RNT204399H RNT204401H RNT204403H RNT204404H RNT204407H RNT204408H RNT204412H RNT204414H RNT204425H RNT204429H RNT204430H RNT204432H RNT204286H RNT204291H RNT204296H RNT204310H RNT204317H RNT204319H RNT204325H RNT204331H RNT204337H RNT204341H RNT204342H RNT204345H RNT204348H RNT204372H RNT204380H RNT204382H RNT204384H RNT204392H RNT204394H RNT204400H RNT204405H RNT204406H RNT204411H RNT204413H RNT204426H RNT204433H RNT204292H RNT204297H RNT204305H RNT204314H RNT204315H RNT204320H RNT204323H RNT204340H RNT204343H RNT204344H RNT204346H RNT204349H RNT204350H RNT204351H RNT204352H RNT204356H RNT204360H RNT204361H RNT204363H RNT204365H RNT204367H RNT204368H RNT204370H RNT204373H RNT204378H RNT204379H RNT204381H RNT204383H RNT204390H RNT204393H RNT204409H RNT204415H RNT204417H RNT204427H RNT204294H RNT204418H RNT204422H RNT204428H RNT204419H RNT204420H RNT204423H RNT204440H RNT204443H RNT204446H RNT204447H RNT204448H RNT204449H RNT204450H RNT204451H RNT204452H RNT204453H RNT204454H RNT204455H RNT204456H RNT204457H RNT204460H RNT204463H RNT204464H RNT204466H RNT204468H RNT204369H RNT204434H RNT204445H RNT204462H RNT204465H RNT204470H RNT204438H RNT204461H RNT204467H RNT204474H RNT204385H RNT204421H RNT204424H RNT204435H RNT204436H RNT204439H RNT204441H RNT204442H RNT204459H RNT204476H RNT204477H RNT204478H RNT204479H RNT204480H RNT204437H RNT204444H RNT204485H RNT204492H RNT204504H RNT204506H RNT204507H RNT204514H RNT204516H RNT204472H RNT204488H RNT204489H RNT204491H RNT204493H RNT204494H RNT204495H RNT204496H RNT204497H RNT204498H RNT204499H RNT204509H RNT204510H RNT204511H RNT204515H RNT204517H RNT204518H RNT204520H RNT204523H RNT204527H RNT204528H RNT204505H RNT204525H RNT204531H RNT204512H RNT204521H RNT204522H RNT204524H RNT204482H RNT204487H RNT204530H RNT204503H RNT204508H RNT204486H RNT204502H RNT204519H RNT204526H RNT204555H RNT204563H RNT204573H RNT204481H RNT204533H RNT204535H RNT204546H RNT204547H RNT204552H RNT204554H RNT204556H RNT204558H RNT204559H RNT204561H RNT204566H RNT204571H RNT204577H RNT203535H RNT204500H RNT204532H RNT204536H RNT204537H RNT204542H RNT204543H RNT204544H RNT204545H RNT204548H RNT204549H RNT204562H RNT204565H RNT204567H RNT204575H RNT204576H RNT204534H RNT204553H RNT204580H RNT204551H RNT204568H RNT204570H RNT204574H RNT204593H RNT204594H RNT204596H RNT204599H RNT204601H RNT204603H RNT204615H RNT204647H RNT204649H RNT204650H RNT204651H RNT204656H RNT204658H RNT204660H RNT204585H RNT204643H RNT204661H RNT204669H RNT204583H RNT204595H RNT204602H RNT204646H RNT204659H RNT204664H RNT204674H RNT204612H RNT204655H RNT204665H RNT204680H RNT204675H RNT204588H RNT204590H RNT204591H RNT204606H RNT204607H RNT204613H RNT204618H RNT204620H RNT204624H RNT204625H RNT204627H RNT204629H RNT204642H RNT204670H RNT204676H RNT204586H RNT204589H RNT204592H RNT204608H RNT204609H RNT204635H RNT204640H RNT204666H RNT204671H RNT204678H RNT204667H RNT204604H RNT204610H RNT204617H RNT204626H RNT204648H RNT204720H RNT204724H RNT204605H RNT204621H RNT204628H RNT204645H RNT204673H RNT204708H RNT204725H RNT204722H RNT204723H RNT204728H RNT204308H RNT204631H RNT204632H RNT204633H RNT204636H RNT204637H RNT204639H RNT204641H RNT204644H RNT204681H RNT204683H RNT204684H RNT204686H RNT204687H RNT204688H RNT204689H RNT204690H RNT204696H RNT204703H RNT204705H RNT204710H RNT204712H RNT204713H RNT204715H RNT204716H RNT204719H RNT204721H RNT204726H RNT204727H RNT204729H RNT204327H RNT204682H RNT204685H RNT204691H RNT204692H RNT204694H RNT204695H RNT204707H RNT204697H RNT204700H RNT204702H RNT204714H RNT204638H RNT204701H RNT204709H RNT204717H RNT204584H RNT204600H RNT204653H RNT203353H RNT203631H RNT203613H RNT203982H RNT204245H RNT204777H RNT204766H RNT204773H RNT204734H RNT204735H RNT204736H RNT204737H RNT204741H RNT204752H RNT204756H RNT204757H RNT204758H RNT204769H RNT204774H RNT204779H RNT204793H RNT204795H RNT204798H RNT204813H RNT204822H RNT204770H RNT204730H RNT204739H RNT204746H RNT204750H RNT204753H RNT204760H RNT204761H RNT204763H
RNT204764H RNT204765H RNT204768H RNT204772H RNT204781H RNT204789H RNT204790H RNT204792H RNT204794H RNT204799H RNT204805H RNT204806H RNT204815H RNT204820H RNT204821H RNT204823H RNT204731H RNT204732H RNT204733H RNT204742H RNT204743H RNT204744H RNT204745H RNT204751H RNT204754H RNT204759H RNT204775H RNT204782H RNT204784H RNT204786H RNT204791H RNT204797H RNT204800H RNT204802H RNT204808H RNT204809H RNT204812H RNT204817H RNT204819H RNT204824H RNT204788H RNT204778H RNT204780H RNT204796H RNT204804H RNT204814H RNT204755H RNT204867H RNT204868H RNT204909H RNT204914H RNT204915H RNT204916H RNT204919H RNT204920H RNT204922H RNT204836H RNT204852H RNT204891H RNT204912H RNT204921H RNT204842H RNT204854H RNT204866H RNT204871H RNT204883H RNT204884H RNT204901H RNT204905H RNT204831H RNT204833H RNT204834H RNT204835H RNT204847H RNT204849H RNT204850H RNT204851H RNT204862H RNT204863H RNT204864H RNT204869H RNT204870H RNT204874H RNT204875H RNT204877H RNT204880H RNT204882H RNT204885H RNT204886H RNT204888H RNT204890H RNT204893H RNT204896H RNT204897H RNT204898H RNT204899H RNT204900H RNT204902H RNT204903H RNT204906H RNT204907H RNT204908H RNT204913H RNT204881H RNT204825H RNT204826H RNT204827H RNT204828H RNT204829H RNT204830H RNT204837H RNT204838H RNT204840H RNT204841H RNT204844H RNT204846H RNT204848H RNT204879H RNT204894H RNT204895H RNT204911H RNT204855H RNT204857H RNT204955H RNT204956H RNT204963H RNT204966H RNT204969H RNT204927H RNT204930H RNT204947H RNT204949H RNT204950H RNT204951H RNT204952H RNT204953H RNT204954H RNT204957H RNT204964H RNT204965H RNT204967H RNT204924H RNT204925H RNT204933H RNT204942H RNT204945H RNT204960H RNT204928H RNT204929H RNT204937H RNT204940H RNT204941H RNT204944H RNT204946H RNT204959H RNT204962H RNT204923H RNT204936H RNT204938H RNT204961H RNT204974H RNT204975H RNT204978H RNT204979H RNT204980H RNT204982H RNT204986H RNT204988H RNT204990H RNT204991H RNT204995H RNT204996H RNT204998H RNT205001H RNT205002H RNT205003H RNT205007H RNT205008H RNT205009H RNT204999H RNT205000H RNT204935H RNT204939H RNT204843H RNT205012H RNT205024H RNT205027H RNT205034H RNT205045H RNT205049H RNT205057H RNT205066H RNT205068H RNT205073H RNT205076H RNT205078H RNT205079H RNT205081H RNT205087H RNT205100H RNT205020H RNT205026H RNT205036H RNT205038H RNT205042H RNT205044H RNT205054H RNT205055H RNT205058H RNT205059H RNT205070H RNT205072H RNT205075H RNT205077H RNT205080H RNT205082H RNT205083H RNT205085H RNT205086H RNT205088H RNT205089H RNT205090H RNT205091H RNT205092H RNT205093H RNT205094H RNT205102H RNT205107H RNT205064H RNT205067H RNT205069H RNT205025H RNT205030H RNT205046H RNT205050H RNT205060H RNT205062H RNT205071H RNT205084H RNT205099H RNT205103H RNT205010H RNT205011H RNT205014H RNT205015H RNT205016H RNT205019H RNT205021H RNT205029H RNT205031H RNT205035H RNT205037H RNT205039H RNT205040H RNT205043H RNT205048H RNT205051H RNT205052H RNT205053H RNT205056H RNT205063H RNT205097H RNT205101H RNT205104H RNT205105H RNT205106H RNT205018H RNT205028H RNT205022H RNT205114H RNT205133H RNT205135H RNT205137H RNT205140H RNT205142H RNT205148H RNT205109H RNT205115H RNT205129H RNT205134H RNT205136H RNT205138H RNT205139H RNT205141H RNT205144H RNT205146H RNT205154H RNT205157H RNT205110H RNT205117H RNT205127H RNT205145H RNT205147H RNT205158H RNT205111H RNT205116H RNT205118H RNT205119H RNT205120H RNT205121H RNT205123H RNT205124H RNT205125H RNT205126H RNT205128H RNT205130H RNT205132H RNT205143H RNT205149H RNT205150H RNT205160H RNT205161H RNT205162H RNT205163H RNT205164H RNT205165H RNT205166H RNT205167H RNT205168H RNT205169H RNT205170H RNT205171H RNT205172H RNT205173H RNT205174H RNT205175H RNT205176H RNT205177H RNT205182H RNT205184H RNT205185H RNT205186H RNT205187H RNT205200H RNT205203H RNT205188H RNT205189H RNT205190H RNT205191H RNT205193H RNT205194H RNT205195H RNT205196H RNT205199H RNT205201H RNT205204H RNT205205H RNT205074H RNT205256H RNT205264H RNT205270H RNT205249H RNT205251H RNT205255H RNT205257H RNT205259H RNT205260H RNT205262H RNT205267H RNT205268H RNT205271H RNT205159H RNT204973H RNT204976H RNT204977H RNT204972H RNT204997H RNT204971H RNT204987H RNT204993H RNT205275H RNT205279H RNT205280H RNT205285H RNT204983H RNT205276H RNT205277H RNT205278H RNT205283H RNT205211H RNT205212H RNT205222H RNT205227H RNT205236H RNT205207H RNT205208H RNT205214H RNT205219H RNT205220H RNT205223H RNT205225H RNT205228H RNT205231H RNT205233H RNT205234H RNT205221H RNT205226H RNT205215H RNT205218H RNT205287H RNT205292H RNT205305H RNT205312H RNT205314H RNT205315H RNT205318H RNT205320H RNT205322H RNT205325H RNT205330H RNT205331H RNT205332H RNT205333H RNT205334H RNT205338H RNT205340H RNT205341H RNT205342H RNT205343H RNT205347H RNT205348H RNT205352H RNT205355H RNT205356H RNT205357H RNT205361H RNT205362H RNT205363H RNT205364H RNT205365H RNT205367H RNT205368H RNT205374H RNT205375H RNT205377H RNT205381H RNT205382H RNT205386H RNT205390H RNT205395H RNT205398H RNT205399H RNT205400H RNT205379H RNT205389H RNT205396H RNT205289H RNT205309H RNT205311H RNT205313H RNT205316H RNT205317H RNT205321H RNT205323H RNT205326H RNT205328H RNT205344H RNT205345H RNT205354H RNT205358H RNT205359H RNT205369H RNT205370H RNT205371H RNT205372H RNT205373H RNT205383H RNT205384H RNT205387H RNT205391H RNT205392H RNT205394H RNT205397H RNT205290H RNT205291H RNT205302H RNT205366H RNT205376H RNT205378H RNT205380H RNT205307H RNT205293H RNT205294H RNT205297H RNT205303H RNT205304H RNT205306H RNT205360H RNT205385H RNT205393H RNT205351H RNT205401H RNT205402H RNT205407H RNT205408H RNT205409H RNT205410H RNT205414H RNT205415H RNT205416H RNT205417H RNT205418H RNT205419H RNT205404H RNT205411H RNT205412H RNT205420H RNT205413H RNT205444H RNT205449H RNT205424H RNT205425H RNT205426H RNT205430H RNT205432H RNT205433H RNT205434H RNT205446H RNT205455H RNT205456H RNT205457H RNT205459H RNT205437H RNT205448H RNT205421H RNT205427H RNT205428H RNT205429H RNT205435H RNT205436H RNT205439H RNT205443H RNT205450H RNT205454H RNT205458H RNT205557H RNT205475H RNT205477H RNT205490H RNT205507H RNT205511H RNT205560H RNT205563H RNT205565H RNT205570H RNT205577H RNT205584H RNT205489H RNT205493H RNT205501H RNT205506H RNT205508H RNT205510H RNT205515H RNT205519H RNT205522H RNT205525H RNT205526H RNT205528H RNT205529H RNT205532H RNT205534H RNT205535H RNT205538H RNT205540H RNT205553H RNT205554H RNT205559H RNT205561H RNT205564H RNT205566H RNT205567H RNT205568H RNT205571H RNT205572H RNT205573H RNT205574H RNT205575H RNT205576H RNT205579H RNT205580H RNT205581H RNT205583H RNT205585H RNT205587H RNT205588H RNT205243H RNT205254H RNT205273H RNT205461H RNT205487H RNT205496H RNT205518H RNT205242H RNT205244H RNT205246H RNT205248H RNT205253H RNT205258H RNT205266H RNT205462H RNT205463H RNT205464H RNT205465H RNT205468H RNT205469H RNT205471H RNT205474H RNT205478H RNT205481H RNT205483H RNT205485H RNT205486H RNT205488H RNT205492H RNT205494H RNT205498H RNT205499H RNT205500H RNT205512H RNT205516H RNT205520H RNT205521H RNT205524H RNT205536H RNT205541H RNT205543H RNT205545H RNT205547H RNT205548H RNT205549H RNT205551H RNT205552H RNT205556H RNT205589H RNT205274H RNT205467H RNT205473H RNT205531H RNT205546H RNT205238H RNT205240H RNT205241H RNT205250H RNT205263H RNT205239H RNT205590H RNT205606H RNT205614H RNT205620H RNT205622H RNT205625H RNT205627H RNT205593H RNT205608H RNT205613H RNT205615H RNT205621H RNT205628H RNT205596H RNT205597H RNT205602H RNT205610H RNT205591H RNT205594H RNT205595H RNT205598H RNT205600H RNT205612H RNT205616H RNT205617H RNT205618H RNT205623H RNT205624H RNT205631H RNT205636H RNT205645H RNT205647H RNT205648H RNT205655H RNT205662H RNT205665H RNT205666H RNT205630H RNT205633H RNT205635H RNT205639H RNT205643H RNT205659H RNT205664H RNT205671H RNT205674H RNT205676H RNT205632H RNT205678H RNT205673H RNT205677H RNT205637H RNT205640H RNT205644H RNT205651H RNT205656H RNT205657H RNT205658H RNT205661H RNT205663H RNT205667H RNT205669H RNT205670H RNT205681H RNT205683H RNT205691H RNT205699H RNT205700H RNT205716H RNT205685H RNT205688H RNT205694H RNT205695H RNT205697H RNT205703H RNT205707H RNT205708H RNT205709H RNT205710H RNT205711H RNT205713H RNT204811H RNT205693H RNT205696H RNT205714H RNT205715H RNT205717H RNT205723H RNT205729H RNT205733H RNT205731H RNT205736H RNT205730H RNT205734H RNT205719H RNT205728H RNT205727H RNT205718H RNT205722H RNT205735H RNT205737H RNT205720H RNT205721H RNT205724H RNT205726H RNT205750H RNT205762H RNT205764H RNT205765H RNT205773H RNT205787H RNT205792H RNT205795H RNT205782H RNT205761H RNT205763H RNT205766H RNT205772H RNT205777H RNT205778H RNT205788H RNT205767H RNT205776H RNT205791H RNT205751H RNT205758H RNT205775H RNT205780H RNT205794H RNT205808H RNT205820H RNT205825H RNT205806H RNT205809H RNT205810H RNT205814H RNT205816H RNT205821H RNT205824H RNT205823H RNT205739H RNT205801H RNT205740H RNT205805H RNT205748H RNT205749H RNT205759H RNT205741H RNT205754H RNT205742H RNT205747H RNT205783H RNT205797H RNT205800H RNT205834H RNT205839H RNT205841H RNT205745H RNT205784H RNT205803H RNT205836H RNT205840H RNT205827H RNT205828H RNT205830H RNT205833H RNT205842H RNT205702H RNT205845H RNT205846H RNT205853H RNT205854H RNT205862H RNT205863H RNT205864H RNT205857H RNT205858H RNT205847H RNT205860H RNT205871H RNT205865H RNT205866H RNT205870H RNT205876H RNT205877H RNT205899H RNT205895H RNT205891H RNT205892H RNT205893H RNT205952H RNT205904H RNT205915H RNT205917H RNT205919H RNT204582H RNT205901H RNT205906H RNT205918H RNT205921H RNT600008S RNT600009S RNT600010S RNT600011S RNT600012S RNT600013S RNT600014S RNT600015S RNT600016S RNT600018S RNT600019S RNT600020S RNT600021S RNT600022S RNT600023S RNT600024S RNT600025S RNT600026S RNT600027S RNT600029S RNT600030S RNT600031S RNT600032S RNT600033S RNT600034S RNT600036S RNT600037S RNT600039S RNT600040S RNT600041S RNT600042S RNT600044S RNT600045S RNT600047S RNT600048S RNT600049S RNT600051S RNT600052S RNT600053S RNT600054S RNT600055S RNT600056S RNT600059S RNT600060S RNT600061S RNT600062S RNT600063S RNT600064S RNT600065S RNT600066S RNT600068S RNT600070S RNT600071S RNT600072S RNT600073S RNT600074S RNT600075S RNT600076S RNT600077S RNT600078S RNT600079S RNT600080S RNT600082S RNT600083S RNT600084S RNT600085S RNT600086S RNT600087S RNT600088S RNT600089S RNT600092S RNT600093S RNT600094S RNT600095S RNT600096S RNT600097S RNT600098S RNT600100S RNT600101S RNT600102S RNT600103S RNT600104S RNT600105S RNT600106S RNT600107S RNT600108S RNT600109S RNT600110S RNT600111S RNT600113S RNT600114S RNT600115S RNT600116S RNT600117S RNT600118S RNT600119S RNT600121S RNT600122S RNT600123S RNT600124S RNT600125S RNT600126S RNT600127S RNT600128S RNT600035S RNT600038S RNT600007S RNT600169S RNT600170S RNT600175S RNT600176S RNT600177S RNT600178S RNT600180S RNT600181S RNT600183S RNT600184S RNT600185S RNT600187S RNT600190S RNT600191S RNT600193S RNT600194S RNT600195S RNT600197S RNT600199S RNT600200S RNT600204S RNT600208S RNT600209S RNT600210S RNT600215S RNT600217S RNT600218S RNT600219S RNT600220S RNT600221S RNT600223S RNT600224S RNT600225S RNT600226S RNT600227S RNT600229S RNT600230S RNT600231S RNT600242S RNT600243S RNT600253S RNT600254S RNT600129S RNT600130S RNT600131S RNT600132S RNT600134S
RNT600135S RNT600136S RNT600138S RNT600140S RNT600142S RNT600143S RNT600144S RNT600145S RNT600146S RNT600147S RNT600149S RNT600150S RNT600151S RNT600152S RNT600153S RNT600154S RNT600155S RNT600156S RNT600157S RNT600159S RNT600161S RNT600166S RNT600167S RNT600168S RNT600171S RNT600172S RNT600173S RNT600174S RNT600179S RNT600186S RNT600188S RNT600192S RNT600196S RNT600198S RNT600201S RNT600202S RNT600205S RNT600206S RNT600211S RNT600212S RNT600213S RNT600216S RNT600235S RNT600236S RNT600237S RNT600238S RNT600239S RNT600240S RNT600241S RNT600244S RNT600245S RNT600246S RNT600248S RNT600249S RNT600251S RNT600252S RNT600133S RNT600139S RNT600222S RNT600255S RNT600256S RNT600257S RNT600258S RNT600259S RNT600260S RNT600264S RNT600265S RNT600266S RNT600267S RNT600268S RNT600269S RNT600271S RNT600274S RNT600275S RNT600276S RNT600277S RNT600278S RNT600279S RNT600280S RNT600281S RNT600282S RNT600283S RNT600284S RNT600285S RNT600286S RNT600287S RNT600289S RNT600292S RNT600293S RNT600294S RNT600295S RNT600296S RNT600297S RNT600298S RNT600299S RNT600300S RNT600302S RNT600304S RNT600306S RNT600308S RNT600309S RNT600311S RNT600312S RNT600313S RNT600314S RNT600317S RNT600318S RNT600319S RNT600322S RNT600323S RNT600328S RNT600330S RNT600331S RNT600335S RNT600337S RNT600338S RNT600339S RNT600340S RNT600341S RNT600342S RNT600343S RNT600345S RNT600347S RNT600350S RNT600352S RNT600354S RNT600355S RNT600357S RNT600360S RNT600362S RNT600365S RNT600367S RNT600371S RNT600376S RNT600378S RNT600379S RNT600381S RNT600384S RNT600387S RNT600389S RNT600390S RNT600394S RNT600316S RNT600320S RNT600325S RNT600327S RNT600329S RNT600334S RNT600348S RNT600351S RNT600361S RNT600364S RNT600366S RNT600369S RNT600370S RNT600374S RNT600380S RNT600382S RNT600386S RNT600388S RNT600391S RNT600395S RNT600396S RNT600399S RNT600400S RNT600401S RNT600332S RNT600349S RNT600141S RNT600393S RNT600402S RNT600404S RNT600409S RNT600422S RNT600428S RNT600435S RNT600403S RNT600405S RNT600408S RNT600410S RNT600412S RNT600413S RNT600417S RNT600418S RNT600419S RNT600421S RNT600423S RNT600425S RNT600426S RNT600429S RNT600430S RNT600431S RNT600433S RNT600434S RNT600436S RNT600437S RNT600438S RNT600439S RNT600440S RNT600442S RNT600443S RNT600444S RNT600445S RNT600446S RNT600447S RNT600427S RNT600449S RNT600207S RNT600416S RNT600463S RNT600450S RNT600453S RNT600454S RNT600455S RNT600456S RNT600458S RNT600459S RNT600460S RNT600461S RNT600464S RNT600465S RNT600466S RNT600467S RNT600468S RNT600471S RNT600472S RNT600473S RNT600474S RNT600475S RNT600477S RNT600479S RNT600481S RNT600487S RNT600488S RNT600490S RNT600492S RNT600494S RNT600495S RNT600496S RNT600498S RNT600525S RNT600538S RNT600543S RNT600546S RNT600548S RNT600553S RNT600609S RNT600611S RNT600612S RNT600616S RNT600476S RNT600478S RNT600480S RNT600482S RNT600483S RNT600484S RNT600485S RNT600486S RNT600493S RNT600500S RNT600502S RNT600503S RNT600505S RNT600507S RNT600508S RNT600510S RNT600512S RNT600513S RNT600514S RNT600516S RNT600519S RNT600520S RNT600521S RNT600522S RNT600524S RNT600528S RNT600530S RNT600532S RNT600533S RNT600534S RNT600535S RNT600536S RNT600537S RNT600539S RNT600540S RNT600541S RNT600542S RNT600545S RNT600547S RNT600549S RNT600550S RNT600551S RNT600552S RNT600557S RNT600558S RNT600560S RNT600561S RNT600562S RNT600563S RNT600564S RNT600565S RNT600566S RNT600568S RNT600569S RNT600571S RNT600572S RNT600573S RNT600574S RNT600575S RNT600576S RNT600577S RNT600578S RNT600579S RNT600582S RNT600583S RNT600586S RNT600589S RNT600590S RNT600591S RNT600592S RNT600595S RNT600596S RNT600597S RNT600598S RNT600599S RNT600602S RNT600603S RNT600605S RNT600606S RNT600608S RNT600610S RNT600621S RNT600624S RNT600462S RNT600511S RNT600517S RNT600518S RNT600526S RNT600544S RNT600555S RNT600584S RNT600620S RNT600631S RNT600633S RNT600634S RNT600635S RNT600646S RNT600642S RNT600648S RNT600655S RNT600656S RNT600657S RNT600658S RNT600660S RNT600666S RNT600668S RNT600670S RNT600678S RNT600679S RNT600680S RNT600641S RNT600647S RNT600677S RNT600681S RNT600683S RNT600685S RNT600691S RNT600693S RNT600694S RNT600682S RNT600688S RNT600692S RNT600695S RNT600696S RNT600704S RNT600706S RNT600707S RNT600712S RNT600713S RNT600718S RNT600702S RNT600721S RNT600723S RNT600726S RNT600732S RNT600734S RNT600738S RNT600745S RNT600747S RNT600748S RNT600752S RNT600755S RNT600757S RNT600759S RNT600724S RNT600730S RNT600733S RNT600742S RNT600763S RNT600778S RNT600781S RNT600784S RNT600785S RNT600786S c) W4TR06, serial numbers: RNZ600013S RNZ600014S RNZ600015S RNZ600016S RNZ600017S RNZ600018S RNZ600019S RNZ600020S RNZ600022S RNZ600023S RNZ600024S RNZ600025S RNZ600026S RNZ600029S RNZ600030S RNZ600031S RNZ600032S RNZ600034S RNZ600035S RNZ600039S RNZ600040S RNZ600041S RNZ600042S RNZ600045S RNZ600046S RNZ600049S RNZ600050S RNZ600053S RNZ600054S RNZ600055S RNZ600056S RNZ600057S RNZ600059S RNZ600061S RNZ600062S RNZ600063S RNZ600064S RNZ600065S RNZ600066S RNZ600067S RNZ600068S RNZ600069S RNZ600070S RNZ600072S RNZ600074S RNZ600075S RNZ600077S RNZ600078S RNZ600079S RNZ600081S RNZ600083S RNZ600084S RNZ600085S RNZ600086S RNZ600089S RNZ600090S RNZ600091S RNZ600094S RNZ600095S RNZ600096S RNZ600098S RNZ600099S RNZ600101S RNZ600102S RNZ600103S RNZ600105S RNZ600107S RNZ600108S RNZ600109S RNZ600110S RNZ600111S RNZ600112S RNZ600113S RNZ600114S RNZ600117S RNZ600119S RNZ600120S RNZ600121S RNZ600122S RNZ600123S RNZ600124S RNZ600125S RNZ600126S RNZ600127S RNZ600128S RNZ600129S RNZ600130S RNZ600136S RNZ600137S RNZ600138S RNZ600141S RNZ600146S RNZ600147S RNZ600150S RNZ600151S RNZ600157S RNZ600158S RNZ600159S RNZ600160S RNZ600161S RNZ600162S RNZ600163S RNZ600164S RNZ600165S RNZ600166S RNZ600167S RNZ600168S RNZ600073S RNZ600076S RNZ600080S RNZ600082S RNZ600088S RNZ600093S RNZ600104S RNZ600106S RNZ600115S RNZ600134S RNZ600139S RNZ600140S RNZ600154S RNZ600155S RNZ600021S RNZ600028S RNZ600051S RNZ600087S RNZ600133S RNZ600027S RNZ600118S RNZ600058S RNZ600071S RNZ600092S RNZ600142S RNZ600143S RNZ600144S RNZ600145S RNZ600148S RNZ600152S RNZ600153S RNZ600169S RNZ600170S RNZ600171S RNZ600173S RNZ600175S RNZ600177S RNZ600178S RNZ600180S RNZ600181S RNZ600184S RNZ600185S RNZ600186S RNZ600187S RNZ600188S RNZ600174S RNZ600179S RNZ600183S RNZ600189S RNZ600190S RNZ600191S RNZ600192S RNZ600194S RNZ600195S RNZ600196S RNZ600198S RNZ600200S RNZ600202S RNZ600203S RNZ600205S RNZ600207S RNZ600208S RNZ600209S RNZ600215S RNZ600219S RNZ600221S RNZ600204S RNZ600206S RNZ600216S RNZ600172S RNZ600210S RNZ600211S RNZ600212S RNZ600213S RNZ600222S RNZ600223S RNZ600224S RNZ600225S RNZ600226S RNZ600227S RNZ600228S RNZ600229S RNZ600231S RNZ600232S RNZ600233S RNZ600234S RNZ600235S RNZ600236S RNZ600237S RNZ600238S RNZ600176S RNZ600214S RNZ600253S RNZ600254S RNZ600255S RNZ600256S RNZ600257S RNZ600258S RNZ600259S RNZ600260S RNZ600261S RNZ600264S RNZ600267S RNZ600245S RNZ600248S RNZ600250S RNZ600252S RNZ600263S RNZ600266S RNZ600269S RNZ600272S RNZ600276S RNZ600279S RNZ600280S RNZ600281S RNZ600251S RNZ600278S RNZ600283S RNZ600286S RNZ600287S RNZ600288S RNZ600289S RNZ600296S RNZ600297S RNZ600298S RNZ600307S RNZ600322S RNZ600361S RNZ600291S RNZ600293S RNZ600294S RNZ600299S RNZ600300S RNZ600301S RNZ600302S RNZ600303S RNZ600304S RNZ600305S RNZ600306S RNZ600309S RNZ600311S RNZ600312S RNZ600314S RNZ600315S RNZ600316S RNZ600317S RNZ600318S RNZ600320S RNZ600321S RNZ600324S RNZ600325S RNZ600326S RNZ600327S RNZ600328S RNZ600329S RNZ600330S RNZ600331S RNZ600332S RNZ600333S RNZ600337S RNZ600338S RNZ600340S RNZ600349S RNZ600353S RNZ600354S RNZ600355S RNZ600358S RNZ600359S RNZ600360S RNZ600362S RNZ600363S RNZ600365S RNZ600368S RNZ600369S RNZ600370S RNZ600371S RNZ600372S RNZ600374S RNZ600377S RNZ600379S RNZ600382S RNZ600383S RNZ600385S RNZ600386S RNZ600387S RNZ600388S RNZ600135S RNZ600380S RNZ600367S RNZ600343S RNZ600450S RNZ600492S RNZ600494S RNZ600495S RNZ600498S RNZ600500S RNZ600503S RNZ600507S RNZ600509S RNZ600512S RNZ600518S RNZ600496S RNZ600497S RNZ600501S RNZ600502S RNZ600506S RNZ600514S RNZ600515S RNZ600516S RNZ600517S RNZ600519S RNZ600520S RNZ600522S RNZ600525S RNZ600526S RNZ600527S RNZ600529S RNZ600530S RNZ600531S RNZ600533S RNZ600534S RNZ600535S RNZ600539S RNZ600541S RNZ600542S RNZ600543S RNZ600544S RNZ600545S RNZ600546S RNZ600548S RNZ600554S RNZ600555S RNZ600556S RNZ600557S RNZ600559S RNZ600562S RNZ600499S RNZ600504S RNZ600505S RNZ600511S RNZ600532S RNZ600561S RNZ600563S RNZ600564S RNZ600565S RNZ600566S RNZ600567S RNZ600568S RNZ600569S RNZ600524S RNZ600570S RNZ600571S RNZ600572S RNZ600573S RNZ600574S RNZ600575S RNZ600576S RNZ600577S RNZ600578S RNZ600579S RNZ600580S RNZ600581S RNZ600582S RNZ600583S RNZ600584S RNZ600585S RNZ600586S RNZ600587S RNZ600588S RNZ600589S RNZ600590S RNZ600591S RNZ600592S RNZ600593S RNZ600594S RNZ600595S RNZ600596S RNZ600597S RNZ600598S RNZ600599S RNZ600600S RNZ600601S RNZ600602S RNZ600603S RNZ600604S RNZ600605S RNZ600606S RNZ600607S RNZ600608S RNZ600609S RNZ600611S RNZ600612S RNZ600613S RNZ600614S RNZ600616S RNZ600617S RNZ600618S RNZ600619S RNZ600620S RNZ600621S RNZ600622S RNZ600623S RNZ600624S RNZ600625S RNZ600626S RNZ600627S RNZ600628S RNZ600629S RNZ600630S RNZ600631S RNZ600632S RNZ600633S RNZ600634S RNZ600635S RNZ600636S RNZ600637S RNZ600638S RNZ600639S RNZ600640S RNZ600641S RNZ600642S RNZ600643S RNZ600645S RNZ600646S RNZ600647S RNZ600649S RNZ600650S RNZ600651S RNZ600652S RNZ600653S RNZ600666S RNZ600644S RNZ600654S RNZ600655S RNZ600657S RNZ600658S RNZ600659S RNZ600660S RNZ600661S RNZ600662S RNZ600663S RNZ600664S RNZ600665S RNZ600667S RNZ600668S RNZ600510S RNZ600521S RNZ600523S RNZ600536S RNZ600538S RNZ600540S RNZ600547S RNZ600549S RNZ600551S RNZ600552S RNZ600558S RNZ600560S RNZ600528S RNZ600550S RNZ600656S RNZ600610S RNZ600669S RNZ600670S RNZ600679S RNZ600674S RNZ600712S RNZ600713S RNZ600716S RNZ600720S RNZ600730S RNZ600732S RNZ600733S RNZ600734S RNZ600736S RNZ600741S RNZ600742S RNZ600671S RNZ600672S RNZ600673S RNZ600676S RNZ600677S RNZ600678S RNZ600680S RNZ600682S RNZ600683S RNZ600684S RNZ600685S RNZ600691S RNZ600704S RNZ600706S RNZ600707S RNZ600708S RNZ600709S RNZ600710S RNZ600715S RNZ600717S RNZ600718S RNZ600719S RNZ600721S RNZ600722S RNZ600723S RNZ600724S RNZ600726S RNZ600727S RNZ600728S RNZ600729S RNZ600731S RNZ600735S RNZ600737S RNZ600738S RNZ600739S RNZ600740S RNZ600743S RNZ600744S RNZ600745S RNZ600746S RNZ600747S RNZ600748S RNZ600750S RNZ600751S RNZ600752S RNZ600755S RNZ600756S RNZ600757S RNZ600758S RNZ600759S RNZ600760S RNZ600761S RNZ600762S RNZ600763S RNZ600764S RNZ600765S RNZ600766S RNZ600767S RNZ600768S RNZ600681S RNZ600687S RNZ600690S RNZ600693S RNZ600700S RNZ600701S RNZ600702S RNZ600703S RNZ600705S RNZ600711S RNZ600714S RNZ600725S RNZ600513S RNZ600553S RNZ600375S RNZ600390S RNZ600391S RNZ600392S RNZ600393S RNZ600394S RNZ600395S RNZ600396S RNZ600397S RNZ600398S RNZ600400S RNZ600401S RNZ600402S RNZ600403S RNZ600404S RNZ600405S RNZ600406S RNZ600407S RNZ600409S RNZ600410S RNZ600411S RNZ600412S RNZ600413S RNZ600414S RNZ600415S RNZ600416S RNZ600417S RNZ600418S RNZ600419S RNZ600420S RNZ600422S RNZ600423S RNZ600424S RNZ600425S RNZ600426S RNZ600427S RNZ600429S RNZ600430S RNZ600431S RNZ600432S RNZ600433S RNZ600434S RNZ600435S RNZ600436S RNZ600437S RNZ600438S RNZ600439S RNZ600440S RNZ600441S RNZ600442S RNZ600443S RNZ600444S RNZ600445S RNZ600447S RNZ600448S RNZ600449S RNZ600451S RNZ600452S RNZ600455S RNZ600456S RNZ600457S RNZ600458S RNZ600459S RNZ600460S RNZ600461S RNZ600462S RNZ600463S RNZ600464S RNZ600465S RNZ600466S RNZ600467S RNZ600468S RNZ600469S RNZ600470S RNZ600472S RNZ600473S RNZ600474S RNZ600475S RNZ600476S RNZ600477S RNZ600478S RNZ600479S RNZ600481S RNZ600482S RNZ600483S RNZ600484S RNZ600485S RNZ600487S RNZ600488S RNZ600489S RNZ600490S RNZ600688S RNZ600769S RNZ600770S RNZ600771S RNZ600772S RNZ600773S RNZ600775S RNZ600776S RNZ600777S RNZ600778S RNZ600779S RNZ600780S RNZ600781S RNZ600782S RNZ600783S RNZ600784S RNZ600785S RNZ600786S RNZ600787S RNZ600788S RNZ600789S RNZ600790S RNZ600791S RNZ600793S RNZ600794S RNZ600796S RNZ600797S RNZ600798S RNZ600799S RNZ600480S RNZ600774S RNZ600792S RNZ600800S RNZ600801S RNZ600802S RNZ600803S RNZ600804S RNZ600805S RNZ600806S RNZ600807S RNZ600809S RNZ600810S RNZ600811S RNZ600812S RNZ600813S RNZ600816S RNZ600818S RNZ600821S RNZ600808S RNZ600814S RNZ600815S RNZ600817S RNZ600819S RNZ600820S RNZ600822S RNZ600823S RNZ600824S RNZ600825S RNZ600826S RNZ600827S RNZ600828S RNZ600829S RNZ600830S RNZ600831S RNZ600832S RNZ600833S RNZ600834S RNZ600835S RNZ600836S RNZ600837S RNZ600838S RNZ600839S RNZ600840S RNZ600841S RNZ600842S RNZ600843S RNZ600844S RNZ600845S RNZ600846S RNZ600847S RNZ600848S RNZ600849S RNZ600850S RNZ600851S RNZ600852S RNZ600853S RNZ600854S RNZ600855S RNZ600856S RNZ600857S RNZ600858S RNZ600859S RNZ600861S RNZ600862S RNZ600863S RNZ600864S RNZ600866S RNZ600867S RNZ600869S RNZ600876S RNZ600890S RNZ600893S RNZ600901S RNZ600860S RNZ600865S RNZ600871S RNZ600872S RNZ600873S RNZ600874S RNZ600875S RNZ600877S RNZ600878S RNZ600879S RNZ600880S RNZ600881S RNZ600882S RNZ600883S RNZ600884S RNZ600885S RNZ600886S RNZ600888S RNZ600889S RNZ600892S RNZ600894S RNZ600895S RNZ600896S RNZ600897S RNZ600898S RNZ600899S RNZ600900S RNZ600902S RNZ600903S RNZ600904S RNZ600905S RNZ600906S RNZ600907S RNZ600753S RNZ600870S RNZ600891S RNZ600910S RNZ600911S RNZ600912S RNZ600913S RNZ600914S RNZ600915S RNZ600916S RNZ600918S RNZ600919S RNZ600920S RNZ600921S RNZ600922S RNZ600923S RNZ600924S RNZ600926S RNZ600927S RNZ600928S RNZ600929S RNZ600930S RNZ600931S RNZ600932S RNZ600933S RNZ600934S RNZ600935S RNZ600937S RNZ600938S RNZ600942S RNZ600754S RNZ600868S RNZ600943S RNZ600944S RNZ600945S RNZ600946S RNZ600947S RNZ600949S RNZ600950S RNZ600951S RNZ600953S RNZ600954S RNZ600955S RNZ600956S RNZ600957S RNZ600958S RNZ600960S RNZ600961S RNZ600962S RNZ600964S RNZ600966S RNZ600968S RNZ600969S RNZ600971S RNZ600973S RNZ600974S RNZ600978S RNZ600980S RNZ600981S RNZ600983S RNZ600985S RNZ600986S RNZ600987S RNZ600990S RNZ600992S RNZ600993S RNZ600995S RNZ600996S RNZ601000S RNZ601002S RNZ600952S RNZ600959S RNZ600965S RNZ600967S RNZ600970S RNZ600972S RNZ600988S RNZ600936S RNZ600963S RNZ601004S RNZ601006S RNZ601007S RNZ601021S RNZ601023S RNZ601025S RNZ601028S RNZ601029S RNZ601030S RNZ601031S RNZ601032S RNZ601033S RNZ601034S RNZ601036S RNZ601037S RNZ601039S RNZ601040S RNZ601041S RNZ601042S RNZ601043S RNZ601044S RNZ601045S RNZ601046S RNZ601047S RNZ601048S RNZ601049S RNZ601051S RNZ601052S RNZ601054S RNZ601056S RNZ601057S RNZ601058S RNZ601059S RNZ601060S RNZ601061S RNZ601062S RNZ601063S RNZ601064S RNZ601065S RNZ601066S RNZ601067S RNZ601068S RNZ601069S RNZ601070S RNZ601071S RNZ601072S RNZ601073S RNZ601074S RNZ601075S RNZ601076S RNZ601077S RNZ601078S RNZ601079S RNZ601082S RNZ601083S RNZ601085S RNZ601005S RNZ601009S RNZ601010S RNZ601011S RNZ601012S RNZ601013S RNZ601014S RNZ601015S RNZ601016S RNZ601017S RNZ601018S RNZ601019S RNZ601022S RNZ601024S RNZ601026S RNZ601027S RNZ601035S RNZ601038S RNZ601050S RNZ601053S RNZ601055S RNZ601080S RNZ601081S RNZ601086S RNZ601087S RNZ601088S RNZ601089S RNZ601090S RNZ601091S RNZ601092S RNZ601093S RNZ601094S RNZ601095S RNZ601096S RNZ601098S RNZ601099S RNZ601100S RNZ601101S RNZ601102S RNZ601103S RNZ601104S RNZ601105S RNZ601106S RNZ601107S RNZ601108S RNZ601109S RNZ601110S RNZ601111S RNZ601112S RNZ601113S RNZ601114S RNZ601115S RNZ601116S RNZ601117S RNZ601118S RNZ601119S RNZ601120S RNZ601121S RNZ601122S RNZ601123S RNZ601124S RNZ601125S RNZ601126S RNZ601127S RNZ601128S RNZ601129S RNZ601130S RNZ601131S RNZ601008S RNZ601132S RNZ601133S RNZ601134S RNZ601135S RNZ601138S RNZ601139S RNZ601140S RNZ601141S RNZ601142S RNZ601146S RNZ601147S RNZ601149S RNZ601151S RNZ601137S RNZ601150S RNZ601152S RNZ601136S RNZ601143S RNZ601144S RNZ601145S RNZ601148S RNZ601153S RNZ601154S RNZ601155S RNZ601156S RNZ601157S RNZ601158S RNZ601159S RNZ601160S RNZ601161S RNZ601162S RNZ601163S RNZ601165S RNZ601166S RNZ601167S RNZ601168S RNZ601169S RNZ601170S RNZ601171S RNZ601172S RNZ601173S RNZ601174S RNZ601175S RNZ601176S RNZ601177S RNZ601178S RNZ601180S RNZ601181S RNZ601182S RNZ601183S RNZ601184S RNZ601185S RNZ601186S RNZ601187S RNZ601188S RNZ601190S RNZ601191S RNZ601195S RNZ601201S RNZ601203S RNZ601207S RNZ601189S RNZ601192S RNZ601193S RNZ601194S RNZ601197S RNZ601198S RNZ601199S RNZ601200S RNZ601204S RNZ601205S RNZ601206S RNZ601208S RNZ601209S RNZ601210S RNZ601211S RNZ601202S RNZ601179S RNZ601213S RNZ601214S RNZ601215S RNZ601216S RNZ601217S RNZ601219S RNZ601222S RNZ601223S RNZ601225S RNZ601226S RNZ601228S RNZ601218S RNZ601221S RNZ601224S RNZ601227S RNZ601230S RNZ601231S RNZ601232S RNZ601233S RNZ601234S RNZ601235S RNZ601237S RNZ601238S RNZ601239S RNZ601241S RNZ601242S RNZ601243S RNZ601244S RNZ601245S RNZ601246S RNZ601247S RNZ601248S RNZ601249S RNZ601250S RNZ601251S RNZ601252S RNZ601253S RNZ601254S RNZ601255S RNZ601256S RNZ601257S RNZ601258S RNZ601259S RNZ601260S RNZ601261S RNZ601265S RNZ601266S RNZ601267S RNZ601268S RNZ601269S RNZ601270S RNZ601271S RNZ601274S RNZ601280S RNZ601282S RNZ601288S RNZ601293S RNZ601294S RNZ601298S RNZ601301S RNZ601262S RNZ601275S RNZ601276S RNZ601277S RNZ601278S RNZ601281S RNZ601283S RNZ601284S RNZ601286S RNZ601287S RNZ601289S RNZ601290S RNZ601291S RNZ601292S RNZ601295S RNZ601296S RNZ601297S RNZ601299S RNZ601300S RNZ601263S RNZ601315S RNZ601317S RNZ601318S RNZ601319S RNZ601272S RNZ601273S RNZ601302S RNZ601303S RNZ601305S RNZ601306S RNZ601307S RNZ601308S RNZ601309S RNZ601310S RNZ601311S RNZ601313S RNZ601314S RNZ601316S RNZ601320S RNZ601321S RNZ601323S RNZ601325S RNZ601326S RNZ601328S RNZ601331S RNZ601333S RNZ601334S RNZ601337S RNZ601340S RNZ601212S RNZ601322S RNZ601324S RNZ601329S RNZ601332S RNZ601335S RNZ601338S RNZ601339S RNZ601341S RNZ601312S RNZ601327S RNZ601220S RNZ601229S RNZ601330S RNZ601342S RNZ601343S RNZ601346S RNZ601348S RNZ601354S RNZ601363S RNZ601381S RNZ601386S RNZ601430S RNZ601431S RNZ601432S RNZ601436S RNZ601437S RNZ601443S RNZ601445S RNZ601446S RNZ601455S RNZ601388S RNZ601411S RNZ601421S RNZ601429S RNZ601433S RNZ601336S RNZ601344S RNZ601345S RNZ601347S RNZ601349S RNZ601350S RNZ601351S RNZ601352S RNZ601353S RNZ601355S RNZ601356S RNZ601357S RNZ601358S RNZ601359S RNZ601360S RNZ601361S RNZ601362S RNZ601365S RNZ601366S RNZ601367S RNZ601368S RNZ601369S RNZ601370S RNZ601371S RNZ601372S RNZ601373S RNZ601375S RNZ601378S RNZ601379S RNZ601380S RNZ601382S RNZ601383S RNZ601384S RNZ601385S RNZ601387S RNZ601389S RNZ601390S RNZ601391S RNZ601392S RNZ601393S RNZ601394S RNZ601395S RNZ601396S RNZ601397S RNZ601398S RNZ601399S RNZ601400S RNZ601401S RNZ601402S RNZ601403S RNZ601404S RNZ601405S RNZ601406S RNZ601407S RNZ601408S RNZ601409S RNZ601410S RNZ601412S RNZ601413S RNZ601414S RNZ601415S RNZ601416S RNZ601417S RNZ601418S RNZ601419S RNZ601420S RNZ601422S RNZ601423S RNZ601424S RNZ601425S RNZ601426S RNZ601427S RNZ601428S RNZ601434S RNZ601435S RNZ601438S RNZ601439S RNZ601440S RNZ601441S RNZ601442S RNZ601444S RNZ601447S RNZ601448S RNZ601449S RNZ601451S RNZ601452S RNZ601453S RNZ601456S RNZ601457S RNZ601458S RNZ601459S RNZ601460S RNZ601364S RNZ601377S RNZ601454S RNZ601461S RNZ601462S RNZ601463S RNZ601464S RNZ601465S RNZ601466S RNZ601468S RNZ601469S RNZ601471S RNZ601472S RNZ601473S RNZ601474S RNZ601475S RNZ601476S RNZ601477S RNZ601478S RNZ601480S RNZ601481S RNZ601482S RNZ601483S RNZ601484S RNZ601485S RNZ601486S RNZ601487S RNZ601488S RNZ601489S RNZ601490S RNZ601491S RNZ601492S RNZ601493S RNZ601494S RNZ601495S RNZ601496S RNZ601497S RNZ601498S RNZ601499S RNZ601500S RNZ601470S RNZ601479S RNZ601501S RNZ601502S RNZ601503S RNZ601504S RNZ601505S RNZ601506S RNZ601507S RNZ601508S RNZ601509S RNZ601510S RNZ601511S RNZ601512S RNZ601513S RNZ601514S RNZ601516S RNZ601517S RNZ601518S RNZ601519S RNZ601520S RNZ601521S RNZ601522S RNZ601523S RNZ601524S RNZ601525S RNZ601526S RNZ601527S RNZ601528S RNZ601529S RNZ601531S RNZ601532S RNZ601533S RNZ601534S RNZ601536S RNZ601537S RNZ601538S RNZ601541S RNZ601542S RNZ601543S RNZ601544S RNZ601545S RNZ601546S RNZ601547S RNZ601548S RNZ601549S RNZ601550S RNZ601551S RNZ601552S RNZ601553S RNZ601554S RNZ601555S RNZ601556S RNZ601557S RNZ601558S RNZ601560S RNZ601539S RNZ601540S RNZ601467S RNZ601561S RNZ601562S RNZ601563S RNZ601564S RNZ601565S RNZ601566S RNZ601567S RNZ601568S RNZ601569S RNZ601570S RNZ601571S RNZ601572S RNZ601573S RNZ601574S RNZ601575S RNZ601576S RNZ601577S RNZ601578S RNZ601579S RNZ601580S RNZ601581S RNZ601582S RNZ601583S RNZ601584S RNZ601585S RNZ601586S RNZ601587S RNZ601588S RNZ601589S RNZ601590S RNZ601591S RNZ601592S RNZ601593S RNZ601594S RNZ601595S RNZ601596S RNZ601597S RNZ601598S RNZ601599S RNZ601600S RNZ601602S RNZ601604S RNZ601606S RNZ601608S RNZ601609S RNZ601612S RNZ601613S RNZ601614S RNZ601616S RNZ601617S RNZ601618S RNZ601619S RNZ601620S RNZ601621S RNZ601623S RNZ601624S RNZ601625S RNZ601626S RNZ601628S RNZ601629S RNZ601630S RNZ601631S RNZ601633S RNZ601635S RNZ601642S RNZ601645S RNZ601649S RNZ601662S RNZ601674S RNZ601675S RNZ601603S RNZ601605S RNZ601610S RNZ601611S RNZ601615S RNZ601622S RNZ601627S RNZ601632S RNZ601634S RNZ601636S RNZ601637S RNZ601638S RNZ601639S RNZ601640S RNZ601641S RNZ601643S RNZ601644S RNZ601646S RNZ601647S RNZ601651S RNZ601652S RNZ601653S RNZ601655S RNZ601656S RNZ601658S RNZ601659S RNZ601660S RNZ601661S RNZ601663S RNZ601664S RNZ601665S RNZ601666S RNZ601668S RNZ601669S RNZ601670S RNZ601671S RNZ601672S RNZ601673S RNZ601676S RNZ601677S RNZ601678S RNZ601679S RNZ601680S RNZ601681S RNZ601682S RNZ601683S RNZ601684S RNZ601685S RNZ601686S RNZ601687S RNZ601688S RNZ601689S RNZ601690S RNZ601691S RNZ601692S RNZ601693S RNZ601694S RNZ601695S RNZ601696S RNZ601697S RNZ601698S RNZ601699S RNZ601700S RNZ601701S RNZ601654S RNZ601657S RNZ601667S RNZ601650S RNZ601702S RNZ601703S RNZ601704S RNZ601705S RNZ601706S RNZ601707S RNZ601708S RNZ601709S RNZ601710S RNZ601711S RNZ601712S RNZ601713S RNZ601714S RNZ601715S RNZ601716S RNZ601717S RNZ601718S RNZ601719S RNZ601720S RNZ601721S RNZ601722S RNZ601724S RNZ601725S RNZ601726S RNZ601727S RNZ601728S RNZ601729S RNZ601731S RNZ601732S RNZ601733S RNZ601734S RNZ601735S RNZ601736S RNZ601737S RNZ601738S RNZ601739S RNZ601740S RNZ601741S RNZ601742S RNZ601743S RNZ601744S RNZ601745S RNZ601746S RNZ601747S RNZ601748S RNZ601749S RNZ601750S RNZ601751S RNZ601752S RNZ601753S RNZ601754S RNZ601756S RNZ601757S RNZ601758S RNZ601760S RNZ601761S RNZ601762S RNZ601768S RNZ601785S RNZ601786S RNZ601787S RNZ601788S RNZ601793S RNZ601794S RNZ601795S RNZ601796S RNZ601799S RNZ601801S RNZ601755S RNZ601759S RNZ601763S RNZ601764S RNZ601765S RNZ601766S RNZ601767S RNZ601769S RNZ601770S RNZ601771S RNZ601772S RNZ601773S RNZ601774S RNZ601775S RNZ601776S RNZ601777S RNZ601778S RNZ601779S RNZ601780S RNZ601781S RNZ601782S RNZ601783S RNZ601790S RNZ601797S RNZ601798S RNZ601800S RNZ601723S RNZ601784S RNZ601791S RNZ601802S RNZ601803S RNZ601804S RNZ601805S RNZ601806S RNZ601807S RNZ601808S RNZ601809S RNZ601810S RNZ601811S RNZ601812S RNZ601813S RNZ601814S RNZ601815S RNZ601816S RNZ601817S RNZ601818S RNZ601819S RNZ601820S RNZ601822S RNZ601823S RNZ601824S RNZ601825S RNZ601827S RNZ601828S RNZ601829S RNZ601830S RNZ601831S RNZ601832S RNZ601833S RNZ601834S RNZ601836S RNZ601837S RNZ601838S RNZ601835S RNZ601840S RNZ601841S RNZ601842S RNZ601843S RNZ601844S RNZ601845S RNZ601846S RNZ601847S RNZ601848S RNZ601850S RNZ601851S RNZ601852S RNZ601853S RNZ601854S RNZ601855S RNZ601856S RNZ601857S RNZ601858S RNZ601859S RNZ601860S RNZ601861S RNZ601862S RNZ601863S RNZ601864S RNZ601865S RNZ601866S RNZ601867S RNZ601868S RNZ601869S RNZ601870S RNZ601871S RNZ601872S RNZ601873S RNZ601875S RNZ601876S RNZ601877S RNZ601879S RNZ601880S RNZ601881S RNZ601882S RNZ601883S RNZ601884S RNZ601885S RNZ601886S RNZ601887S RNZ601888S RNZ601889S RNZ601890S RNZ601891S RNZ601892S RNZ601893S RNZ601897S RNZ601900S RNZ601901S RNZ601902S RNZ601904S RNZ601905S RNZ601906S RNZ601907S RNZ601908S RNZ601909S RNZ601910S RNZ601911S RNZ601912S RNZ601913S RNZ601914S RNZ601915S RNZ601916S RNZ601917S RNZ601918S RNZ601919S RNZ601920S RNZ601921S RNZ601922S RNZ601923S RNZ601924S RNZ601925S RNZ601926S RNZ601927S RNZ601928S RNZ601929S RNZ601930S RNZ601931S RNZ601932S RNZ601933S RNZ601934S RNZ601938S RNZ601939S RNZ601941S RNZ601942S RNZ601943S RNZ601944S RNZ601946S RNZ601947S RNZ601948S RNZ601949S RNZ601951S RNZ601952S RNZ601953S RNZ601954S RNZ601955S RNZ601956S RNZ601957S RNZ601959S RNZ601960S RNZ601962S RNZ601963S RNZ601964S RNZ601965S RNZ601967S RNZ601968S RNZ601969S RNZ601970S RNZ601935S RNZ601936S RNZ601937S RNZ601945S RNZ601950S RNZ601958S RNZ601961S RNZ601966S RNZ601895S RNZ601971S RNZ601972S RNZ601973S RNZ601975S RNZ601976S RNZ601977S RNZ601978S RNZ601979S RNZ601980S RNZ601981S RNZ601982S RNZ601983S RNZ601984S RNZ601986S RNZ601987S RNZ601989S RNZ601991S RNZ601996S RNZ601997S RNZ602002S RNZ602005S RNZ601974S RNZ601985S RNZ601988S RNZ601992S RNZ601995S RNZ601998S RNZ601999S RNZ602000S RNZ602001S RNZ602003S RNZ602004S RNZ602008S RNZ602009S RNZ602011S RNZ602013S RNZ602014S RNZ602015S RNZ602017S RNZ602020S RNZ601990S RNZ601993S RNZ602006S RNZ602007S RNZ602010S RNZ602012S RNZ602016S RNZ602019S RNZ602018S RNZ602021S RNZ602022S RNZ602023S RNZ602024S RNZ602025S RNZ602026S RNZ602027S RNZ602028S RNZ602029S RNZ602030S RNZ602031S RNZ602032S RNZ602033S RNZ602034S RNZ602035S RNZ602036S RNZ602037S RNZ602038S RNZ602040S RNZ602041S RNZ602042S RNZ602043S RNZ602044S RNZ602045S RNZ602046S RNZ602047S RNZ602048S RNZ602049S RNZ602050S RNZ602053S RNZ602054S RNZ602055S RNZ602057S RNZ602058S RNZ602059S RNZ602060S RNZ602074S RNZ602052S RNZ602056S RNZ602061S RNZ602062S RNZ602065S RNZ602067S RNZ602068S RNZ602069S RNZ602070S RNZ602071S RNZ602072S RNZ602073S RNZ602075S RNZ602076S RNZ602077S RNZ602078S RNZ602079S RNZ602080S RNZ602081S RNZ602082S RNZ602083S RNZ602084S RNZ602085S RNZ602086S RNZ602087S RNZ602088S RNZ602089S RNZ602090S RNZ602091S RNZ602092S RNZ602093S RNZ602094S RNZ602095S RNZ602096S RNZ602097S RNZ602098S RNZ602099S RNZ602100S RNZ602101S RNZ602102S RNZ602103S RNZ602105S RNZ602107S RNZ602108S RNZ602109S RNZ602110S RNZ602111S RNZ602112S RNZ602113S RNZ602114S RNZ602115S RNZ602116S RNZ602117S RNZ602118S RNZ602119S RNZ602120S RNZ602121S RNZ602122S RNZ602123S RNZ602124S RNZ602125S RNZ602126S RNZ602127S RNZ602128S RNZ602129S RNZ602130S RNZ602131S RNZ602132S RNZ602133S RNZ602134S RNZ602135S RNZ602136S RNZ602137S RNZ602138S RNZ602139S RNZ602063S RNZ602064S RNZ602066S RNZ602104S RNZ602140S RNZ602141S RNZ602145S RNZ602146S RNZ602147S RNZ602149S RNZ602150S RNZ602151S RNZ602152S RNZ602153S RNZ602154S RNZ602155S RNZ602156S RNZ602157S RNZ602158S RNZ602160S RNZ602161S RNZ602162S RNZ602163S RNZ602164S RNZ602166S RNZ602167S RNZ602168S RNZ602169S RNZ602171S RNZ602172S RNZ602173S RNZ602174S RNZ602175S RNZ602176S RNZ602177S RNZ602178S RNZ602179S RNZ602180S RNZ602181S RNZ602182S RNZ602183S RNZ602184S RNZ602185S RNZ602186S RNZ602187S RNZ602188S RNZ602189S RNZ602190S RNZ602191S RNZ602192S RNZ602193S RNZ602194S RNZ602195S RNZ602196S RNZ602199S RNZ602165S RNZ602170S RNZ602198S RNZ602200S RNZ602201S RNZ602202S RNZ602203S RNZ602204S RNZ602205S RNZ602206S RNZ602207S RNZ602208S RNZ602209S RNZ602210S RNZ602211S RNZ602212S RNZ602213S RNZ602214S RNZ602215S RNZ602216S RNZ602217S RNZ602218S RNZ602219S RNZ602220S RNZ602221S RNZ602223S RNZ602224S RNZ602225S RNZ602228S RNZ602229S RNZ602230S RNZ602231S RNZ602232S RNZ602233S RNZ602235S RNZ602236S RNZ602237S RNZ602143S RNZ602238S RNZ602239S RNZ602242S RNZ602243S RNZ602244S RNZ602245S RNZ602246S RNZ602248S RNZ602249S RNZ602252S RNZ602253S RNZ602254S RNZ602255S RNZ602256S RNZ602257S RNZ602258S RNZ602240S RNZ602241S RNZ602251S RNZ602259S RNZ602260S RNZ602261S RNZ602262S RNZ602263S RNZ602264S RNZ602265S RNZ602266S RNZ602267S RNZ602269S RNZ602271S RNZ602272S RNZ602273S RNZ602268S RNZ602270S RNZ602274S RNZ602275S RNZ602276S RNZ602278S RNZ602279S RNZ602280S RNZ602281S RNZ602282S RNZ602283S RNZ602284S RNZ602285S RNZ602286S RNZ602287S RNZ602288S RNZ602289S RNZ602290S RNZ602291S RNZ602292S RNZ602293S RNZ602294S RNZ602295S RNZ602296S RNZ602297S RNZ602298S RNZ602339S RNZ602348S RNZ602349S RNZ602351S RNZ602353S RNZ602358S RNZ602300S RNZ602302S RNZ602303S RNZ602304S RNZ602305S RNZ602306S RNZ602307S RNZ602309S RNZ602310S RNZ602311S RNZ602312S RNZ602313S RNZ602314S RNZ602315S RNZ602316S RNZ602318S RNZ602320S RNZ602321S RNZ602322S RNZ602323S RNZ602324S RNZ602325S RNZ602326S RNZ602327S RNZ602329S RNZ602330S RNZ602331S RNZ602332S RNZ602333S RNZ602334S RNZ602335S RNZ602336S RNZ602337S RNZ602338S RNZ602340S RNZ602343S RNZ602344S RNZ602345S RNZ602346S RNZ602347S RNZ602350S RNZ602352S RNZ602354S RNZ602355S RNZ602356S RNZ602357S RNZ602367S RNZ602368S RNZ602369S RNZ602370S RNZ602371S RNZ602372S RNZ602373S RNZ602374S RNZ602375S RNZ602376S RNZ602377S RNZ602378S RNZ602380S RNZ602381S RNZ602383S RNZ602385S RNZ602386S RNZ602384S RNZ602387S RNZ602388S RNZ602389S RNZ602391S RNZ602392S RNZ602393S RNZ602394S RNZ602396S RNZ602398S RNZ602399S RNZ602400S RNZ602401S RNZ602402S RNZ602403S RNZ602404S RNZ602405S RNZ602406S RNZ602407S RNZ602408S RNZ602409S RNZ602410S RNZ602411S RNZ602413S RNZ602414S RNZ602415S RNZ602416S RNZ602417S RNZ602418S RNZ602419S RNZ602420S RNZ602421S RNZ602422S RNZ602423S RNZ602424S RNZ602425S RNZ602426S RNZ602427S RNZ602428S RNZ602429S RNZ602430S RNZ602431S RNZ602432S RNZ602433S RNZ602434S RNZ602435S RNZ602436S RNZ602437S RNZ602438S RNZ602439S RNZ602440S RNZ602441S RNZ602442S RNZ602443S RNZ602444S RNZ602445S RNZ602446S RNZ602447S RNZ602448S RNZ602449S RNZ602450S RNZ602451S RNZ602452S RNZ602453S RNZ602454S RNZ602456S RNZ602457S RNZ602458S RNZ602148S RNZ602459S RNZ602460S RNZ602461S RNZ602462S RNZ602463S RNZ602465S RNZ602467S RNZ602468S RNZ602469S RNZ602470S RNZ602471S RNZ602472S RNZ602473S RNZ602474S RNZ602475S RNZ602478S RNZ602476S RNZ602477S RNZ602455S RNZ602226S RNZ602397S RNZ602482S RNZ602483S RNZ602484S RNZ602485S RNZ602486S RNZ602487S RNZ602488S RNZ602489S RNZ602490S RNZ602492S RNZ602493S RNZ602494S RNZ602495S RNZ602497S RNZ602498S RNZ602499S RNZ602501S RNZ602502S RNZ602503S RNZ602504S RNZ602505S RNZ602506S RNZ602507S RNZ602508S RNZ602509S RNZ602510S RNZ602511S RNZ602512S RNZ602513S RNZ602514S RNZ602515S RNZ602516S RNZ602517S RNZ602518S RNZ602496S RNZ602519S RNZ602523S RNZ602524S RNZ602525S RNZ602527S RNZ602528S RNZ602530S RNZ602531S RNZ602532S RNZ602534S RNZ602535S RNZ602536S RNZ602579S RNZ602581S RNZ602586S RNZ602590S RNZ602591S RNZ602592S RNZ602521S RNZ602522S RNZ602526S RNZ602529S RNZ602533S RNZ602538S RNZ602539S RNZ602541S RNZ602543S RNZ602545S RNZ602546S RNZ602547S RNZ602548S RNZ602549S RNZ602550S RNZ602551S RNZ602552S RNZ602553S RNZ602554S RNZ602555S RNZ602556S RNZ602557S RNZ602558S RNZ602559S RNZ602560S RNZ602561S RNZ602562S RNZ602563S RNZ602564S RNZ602565S RNZ602566S RNZ602567S RNZ602568S RNZ602569S RNZ602570S RNZ602572S RNZ602573S RNZ602574S RNZ602575S RNZ602578S RNZ602580S RNZ602582S RNZ602583S RNZ602584S RNZ602585S RNZ602587S RNZ602588S RNZ602589S RNZ602594S RNZ602596S RNZ602597S RNZ602598S RNZ602540S RNZ602542S RNZ602544S RNZ602520S RNZ602599S RNZ602600S RNZ602601S RNZ602602S RNZ602603S RNZ602604S RNZ602605S RNZ602606S RNZ602607S RNZ602608S RNZ602609S RNZ602610S RNZ602612S RNZ602613S RNZ602614S RNZ602615S RNZ602616S RNZ602617S RNZ602618S RNZ602619S RNZ602621S RNZ602622S RNZ602623S RNZ602624S RNZ602625S RNZ602626S RNZ602627S RNZ602628S RNZ602629S RNZ602630S RNZ602631S RNZ602632S RNZ602633S RNZ602634S RNZ602635S RNZ602636S RNZ602637S RNZ602319S RNZ602491S RNZ602638S RNZ602639S RNZ602640S RNZ602641S RNZ602642S RNZ602643S RNZ602644S RNZ602645S RNZ602647S RNZ602648S RNZ602650S RNZ602651S RNZ602652S RNZ602653S RNZ602654S RNZ602655S RNZ602656S RNZ602657S RNZ602658S RNZ602659S RNZ602662S RNZ602663S RNZ602664S RNZ602667S RNZ602668S RNZ602669S RNZ602670S RNZ602671S RNZ602672S RNZ602673S RNZ602674S RNZ602675S RNZ602676S RNZ602677S RNZ602678S RNZ602679S RNZ602680S RNZ602681S RNZ602682S RNZ602683S RNZ602684S RNZ602685S RNZ602686S RNZ602687S RNZ602688S RNZ602689S RNZ602690S RNZ602691S RNZ602692S RNZ602693S RNZ602694S RNZ602695S RNZ602696S RNZ602697S RNZ602698S RNZ602699S RNZ602700S RNZ602701S RNZ602702S RNZ602704S RNZ602705S RNZ602706S RNZ602661S RNZ602712S RNZ602714S RNZ602715S RNZ602718S RNZ602719S RNZ602721S RNZ602722S RNZ602723S RNZ602724S RNZ602725S RNZ602727S RNZ602729S RNZ602731S RNZ602732S RNZ602733S RNZ602735S RNZ602738S RNZ602739S RNZ602740S RNZ602741S RNZ602744S RNZ602745S RNZ602749S RNZ602753S RNZ602754S RNZ602755S RNZ602756S RNZ602764S RNZ602767S RNZ602769S RNZ602665S RNZ602711S RNZ602713S RNZ602716S RNZ602717S RNZ602720S RNZ602728S RNZ602734S RNZ602736S RNZ602737S RNZ602742S RNZ602743S RNZ602746S RNZ602747S RNZ602748S RNZ602750S RNZ602751S RNZ602752S RNZ602757S RNZ602758S RNZ602759S RNZ602760S RNZ602761S RNZ602762S RNZ602765S RNZ602766S RNZ602768S RNZ602649S RNZ602660S RNZ602703S RNZ602726S RNZ602730S RNZ602770S RNZ602771S RNZ602772S RNZ602773S RNZ602774S RNZ602775S RNZ602776S RNZ602777S RNZ602780S RNZ602781S RNZ602782S RNZ602783S RNZ602784S RNZ602785S RNZ602786S RNZ602787S RNZ602788S RNZ602789S RNZ602790S RNZ602791S RNZ602792S RNZ602793S RNZ602794S RNZ602795S RNZ602796S RNZ602797S RNZ602798S RNZ602799S RNZ602800S RNZ602801S RNZ602802S RNZ602803S RNZ602804S RNZ602805S RNZ602806S RNZ602807S RNZ602808S RNZ602809S RNZ602811S RNZ602812S RNZ602813S RNZ602814S RNZ602815S RNZ602816S RNZ602817S RNZ602818S RNZ602819S RNZ602820S RNZ602821S RNZ602822S RNZ602823S RNZ602824S RNZ602825S RNZ602778S RNZ602810S RNZ602826S RNZ602827S RNZ602828S RNZ602829S RNZ602830S RNZ602831S RNZ602832S RNZ602836S RNZ602838S RNZ602839S RNZ602841S RNZ602842S RNZ602843S RNZ602844S RNZ602845S RNZ602846S RNZ602847S RNZ602848S RNZ602849S RNZ602850S RNZ602851S RNZ602852S RNZ602853S RNZ602854S RNZ602855S RNZ602856S RNZ602857S RNZ602858S RNZ602860S RNZ602861S RNZ602862S RNZ602863S RNZ602864S RNZ602865S RNZ602867S RNZ602868S RNZ602870S RNZ602871S RNZ602873S RNZ602874S RNZ602875S RNZ602876S RNZ602877S RNZ602878S RNZ602879S RNZ602881S RNZ602882S RNZ602883S RNZ602884S RNZ602885S RNZ602887S RNZ602888S RNZ602889S RNZ602891S RNZ602892S RNZ602893S RNZ602894S RNZ602895S RNZ602896S RNZ602899S RNZ602900S RNZ602901S RNZ602902S RNZ602903S RNZ602904S RNZ602905S RNZ602906S RNZ602907S RNZ602908S RNZ602909S RNZ602910S RNZ602911S RNZ602912S RNZ602913S RNZ602914S RNZ602915S RNZ602916S RNZ602917S RNZ602918S RNZ602919S RNZ602920S RNZ602835S RNZ602859S RNZ602866S RNZ602890S RNZ602921S RNZ602880S RNZ602922S RNZ602923S RNZ602924S RNZ602928S RNZ602929S RNZ602930S RNZ602931S RNZ602932S RNZ602933S RNZ602934S RNZ602935S RNZ602936S RNZ602937S RNZ602938S RNZ602939S RNZ602940S RNZ602941S RNZ602942S RNZ602943S RNZ602944S RNZ602946S RNZ602947S RNZ602948S RNZ602949S RNZ602951S RNZ602952S RNZ602954S RNZ602955S RNZ602956S RNZ602957S RNZ602958S RNZ602959S RNZ602962S RNZ602963S RNZ602964S RNZ602965S RNZ602966S RNZ602967S RNZ602968S RNZ602969S RNZ602970S RNZ602971S RNZ602972S RNZ602973S RNZ602975S RNZ602976S RNZ602977S RNZ602978S RNZ602979S RNZ602980S RNZ602981S RNZ602983S RNZ602984S RNZ602985S RNZ602986S RNZ602953S RNZ602974S RNZ602991S RNZ602992S RNZ602993S RNZ602996S RNZ602997S RNZ602999S RNZ603002S RNZ603003S RNZ603006S RNZ603007S RNZ603010S RNZ603011S RNZ603012S RNZ603013S RNZ603014S RNZ603016S RNZ603022S RNZ603023S RNZ603025S RNZ603026S RNZ603032S RNZ603033S RNZ603038S RNZ603043S RNZ603044S RNZ603046S RNZ603051S RNZ603052S RNZ603054S RNZ603057S RNZ603059S RNZ603065S RNZ603066S RNZ602987S RNZ602989S RNZ602990S RNZ602994S RNZ602995S RNZ602998S RNZ603000S RNZ603001S RNZ603004S RNZ603008S RNZ603009S RNZ603018S RNZ603019S RNZ603020S RNZ603027S RNZ603028S RNZ603029S RNZ603030S RNZ603031S RNZ603035S RNZ603036S RNZ603040S RNZ603041S RNZ603042S RNZ603045S RNZ603047S RNZ603048S RNZ603049S RNZ603050S RNZ603053S RNZ603056S RNZ603058S RNZ603060S RNZ603061S RNZ603062S RNZ603064S RNZ603067S RNZ603068S RNZ603069S RNZ603070S RNZ603072S RNZ603073S RNZ603074S RNZ603075S RNZ603083S RNZ603084S RNZ603086S RNZ603087S RNZ603088S RNZ603090S RNZ603091S RNZ603092S RNZ603095S RNZ603096S RNZ603097S RNZ603098S RNZ603099S RNZ603100S RNZ603102S RNZ603103S RNZ603105S RNZ603106S RNZ603108S RNZ603110S RNZ603111S RNZ603113S RNZ603114S RNZ603115S RNZ603116S RNZ603117S RNZ603118S RNZ603119S RNZ603120S RNZ603121S RNZ603122S RNZ603123S RNZ603124S RNZ603125S RNZ603126S RNZ603034S RNZ603079S RNZ603109S RNZ603071S RNZ603089S RNZ603127S RNZ603131S RNZ603132S RNZ603134S RNZ603135S RNZ603136S RNZ603137S RNZ603139S RNZ603140S RNZ603141S RNZ603142S RNZ603143S RNZ603144S RNZ603145S RNZ603147S RNZ603148S RNZ603150S RNZ603151S RNZ603152S RNZ603154S RNZ603155S RNZ603156S RNZ603157S RNZ603158S RNZ603159S RNZ603161S RNZ603162S RNZ603163S RNZ603164S RNZ603128S RNZ603129S RNZ603130S RNZ603133S RNZ603138S RNZ603146S RNZ603149S RNZ603153S RNZ603160S RNZ603107S RNZ603165S RNZ603166S RNZ603168S RNZ603169S RNZ603171S RNZ603172S RNZ603174S RNZ603177S RNZ603178S RNZ603179S RNZ603181S RNZ603182S RNZ603183S RNZ603184S RNZ603185S RNZ603186S RNZ603187S RNZ603188S RNZ603189S RNZ603190S RNZ603191S RNZ603192S RNZ603193S RNZ603194S RNZ603195S RNZ603196S RNZ603197S RNZ603198S RNZ603199S RNZ603200S RNZ603202S RNZ603203S RNZ603206S RNZ603207S RNZ603208S RNZ603209S RNZ603210S RNZ603211S RNZ603212S RNZ603213S RNZ603214S RNZ603216S RNZ603217S RNZ603218S RNZ603219S RNZ603220S RNZ603221S RNZ603222S RNZ603224S RNZ603225S RNZ603226S RNZ603227S RNZ603228S RNZ603229S RNZ603180S RNZ603232S RNZ603234S RNZ603237S RNZ603231S RNZ603233S RNZ603235S RNZ603236S RNZ603239S RNZ603241S RNZ603242S RNZ603244S RNZ603245S RNZ603246S RNZ603247S RNZ603248S RNZ603238S RNZ603240S RNZ603243S RNZ603250S RNZ603252S RNZ603254S RNZ603255S RNZ603256S RNZ603257S RNZ603258S RNZ603259S RNZ603260S RNZ603261S RNZ603262S RNZ603263S RNZ603264S RNZ603265S RNZ603266S RNZ603267S RNZ603268S RNZ603269S RNZ603270S RNZ603271S RNZ603272S RNZ603274S RNZ603275S RNZ603276S RNZ603277S RNZ603278S RNZ602926S RNZ602927S RNZ603273S RNZ603279S RNZ602925S RNZ603251S RNZ603280S RNZ603281S RNZ603282S RNZ603283S RNZ603285S RNZ603286S RNZ603287S RNZ603288S RNZ603289S RNZ603291S RNZ603292S RNZ603294S RNZ603295S RNZ603296S RNZ603297S RNZ603298S RNZ603299S RNZ603300S RNZ603301S RNZ603302S RNZ603303S RNZ603305S RNZ603306S RNZ603308S RNZ603309S RNZ603310S RNZ603311S RNZ603313S RNZ603314S RNZ603315S RNZ603316S RNZ603317S RNZ603318S RNZ603319S RNZ603320S RNZ603321S RNZ603322S RNZ603323S RNZ603324S RNZ603325S RNZ603326S RNZ603327S RNZ603328S RNZ603293S RNZ602364S RNZ603329S RNZ603330S RNZ603331S RNZ603332S RNZ603333S RNZ603335S RNZ603336S RNZ603337S RNZ603338S RNZ603339S RNZ603340S RNZ603341S RNZ603342S RNZ603343S RNZ603344S RNZ603347S RNZ603348S RNZ603350S RNZ603351S RNZ603354S RNZ603356S RNZ603357S RNZ603358S RNZ603360S RNZ603361S RNZ603362S RNZ603364S RNZ603366S RNZ603367S RNZ603369S RNZ603370S RNZ603371S RNZ603373S RNZ603377S RNZ603379S RNZ603380S RNZ603381S RNZ603382S RNZ603384S RNZ603385S RNZ603386S RNZ603387S RNZ603388S RNZ603389S RNZ603390S RNZ603391S RNZ603392S RNZ603394S RNZ603396S RNZ603397S RNZ603399S RNZ603400S RNZ603401S RNZ603402S RNZ603403S RNZ603404S RNZ603405S RNZ603406S RNZ603407S RNZ603408S RNZ603409S RNZ603411S RNZ603412S RNZ603413S RNZ603414S RNZ603416S RNZ603417S RNZ603419S RNZ603420S RNZ603421S RNZ603422S RNZ603423S RNZ603424S RNZ603425S RNZ603433S RNZ603447S RNZ603449S RNZ603452S RNZ603453S RNZ603455S RNZ603458S RNZ603352S RNZ603353S RNZ603378S RNZ603383S RNZ603410S RNZ603418S RNZ602360S RNZ602363S RNZ603253S RNZ603346S RNZ603349S RNZ603355S RNZ603359S RNZ603398S RNZ603426S RNZ603427S RNZ603428S RNZ603431S RNZ603437S RNZ603440S RNZ603441S RNZ603442S RNZ603443S RNZ603444S RNZ603445S RNZ603446S RNZ603450S RNZ603454S RNZ603456S RNZ603457S RNZ603460S RNZ603462S RNZ603463S RNZ602365S RNZ602366S RNZ603464S RNZ603465S RNZ603466S RNZ603467S RNZ603468S RNZ603471S RNZ603473S RNZ603474S RNZ603475S RNZ603478S RNZ603480S RNZ603481S RNZ603485S RNZ603486S RNZ603487S RNZ603489S RNZ603491S RNZ603492S RNZ603494S RNZ603470S RNZ603472S RNZ603476S RNZ603477S RNZ603493S RNZ603469S RNZ603482S RNZ603483S RNZ603484S RNZ603495S RNZ603496S RNZ603497S RNZ603498S RNZ603500S RNZ603501S RNZ603502S RNZ603507S RNZ603508S RNZ603510S RNZ603511S RNZ603512S RNZ603513S RNZ603514S RNZ603515S RNZ603516S RNZ603517S RNZ603518S RNZ603521S RNZ603522S RNZ603523S RNZ603526S RNZ603529S RNZ603530S RNZ603532S RNZ603533S RNZ603534S RNZ603535S RNZ603536S RNZ603538S RNZ603539S RNZ603540S RNZ603541S RNZ603542S RNZ603543S RNZ603544S RNZ603545S RNZ603546S RNZ603547S RNZ603549S RNZ603550S RNZ603552S RNZ603560S RNZ603580S RNZ603593S RNZ603596S RNZ603598S RNZ603599S RNZ603519S RNZ603527S RNZ603528S RNZ603531S RNZ603537S RNZ603548S RNZ603553S RNZ603554S RNZ603555S RNZ603556S RNZ603557S RNZ603558S RNZ603559S RNZ603562S RNZ603563S RNZ603564S RNZ603565S RNZ603566S RNZ603567S RNZ603568S RNZ603569S RNZ603570S RNZ603571S RNZ603572S RNZ603573S RNZ603574S RNZ603575S RNZ603576S RNZ603577S RNZ603578S RNZ603579S RNZ603584S RNZ603585S RNZ603587S RNZ603588S RNZ603589S RNZ603591S RNZ603592S RNZ603594S RNZ603595S RNZ603597S RNZ603601S RNZ603603S RNZ603604S RNZ603605S RNZ603606S RNZ603607S RNZ603608S RNZ603609S RNZ603610S RNZ603611S RNZ603612S RNZ603613S RNZ603614S RNZ603615S RNZ603616S RNZ603618S RNZ603620S RNZ603625S RNZ603626S RNZ603627S RNZ603629S RNZ603630S RNZ603632S RNZ603633S RNZ603635S RNZ603637S RNZ603638S RNZ603640S RNZ603641S RNZ603642S RNZ603643S RNZ603647S RNZ603648S RNZ603653S RNZ603617S RNZ603622S RNZ603623S RNZ603624S RNZ603628S RNZ603631S RNZ603636S RNZ603639S RNZ603644S RNZ603649S RNZ603651S RNZ603652S RNZ603656S RNZ603658S RNZ603659S RNZ603660S RNZ603661S RNZ603664S RNZ603665S RNZ603666S RNZ603672S RNZ603673S RNZ603655S RNZ603657S RNZ603662S RNZ603663S RNZ603667S RNZ603669S RNZ603670S RNZ603671S RNZ603645S RNZ603674S RNZ603675S RNZ603676S RNZ603677S RNZ603678S RNZ603679S RNZ603680S RNZ603681S RNZ603682S RNZ603683S RNZ603684S RNZ603685S RNZ603686S RNZ603687S RNZ603688S RNZ603689S RNZ603690S RNZ603691S RNZ603692S RNZ603693S RNZ603694S RNZ603695S RNZ603696S RNZ603697S RNZ603698S RNZ603699S RNZ603700S RNZ603701S RNZ603702S RNZ603703S RNZ603704S RNZ603705S RNZ603706S RNZ603707S RNZ603708S RNZ603709S RNZ603710S RNZ603711S RNZ603714S RNZ603715S RNZ603716S RNZ603719S RNZ603723S RNZ603724S RNZ603725S RNZ603727S RNZ603728S RNZ603729S RNZ603730S RNZ603732S RNZ603733S RNZ603734S RNZ603736S RNZ603738S RNZ603739S RNZ603740S RNZ603741S RNZ603742S RNZ603744S RNZ603746S RNZ603748S RNZ603751S RNZ603752S RNZ603753S RNZ603668S RNZ603713S RNZ603754S RNZ603755S RNZ603761S RNZ603762S RNZ603763S RNZ603770S RNZ603771S RNZ603773S RNZ603774S RNZ603784S RNZ603759S RNZ603776S RNZ603777S RNZ603778S RNZ603779S RNZ603780S RNZ603781S RNZ603782S RNZ603783S RNZ603785S RNZ603788S RNZ603792S RNZ603793S RNZ603795S RNZ603796S RNZ603798S RNZ603799S RNZ603800S RNZ603801S RNZ603802S RNZ603803S RNZ603804S RNZ603805S RNZ603807S RNZ603808S RNZ603809S RNZ603810S RNZ603812S RNZ603758S RNZ603760S RNZ603767S RNZ603768S RNZ603816S RNZ603819S RNZ603820S RNZ603821S RNZ603823S RNZ603825S RNZ603826S RNZ603827S RNZ603830S RNZ603834S RNZ603835S RNZ603838S RNZ603840S RNZ603843S RNZ603844S RNZ603845S RNZ603846S RNZ603848S RNZ603849S RNZ603850S RNZ603851S RNZ603756S RNZ603757S RNZ603764S RNZ603765S RNZ603766S RNZ603769S RNZ603772S RNZ603775S RNZ603814S RNZ603815S RNZ603817S RNZ603822S RNZ603824S RNZ603828S RNZ603829S RNZ603831S RNZ603833S RNZ603836S RNZ603841S RNZ603842S RNZ603847S RNZ603852S RNZ603854S RNZ603855S RNZ603856S RNZ603857S RNZ603858S RNZ603859S RNZ603860S RNZ603861S RNZ603863S RNZ603864S RNZ603866S RNZ603868S RNZ603869S RNZ603871S RNZ603872S RNZ603873S RNZ603874S RNZ603877S RNZ603879S RNZ603887S RNZ603896S RNZ603897S RNZ603898S RNZ603899S RNZ603904S RNZ603908S RNZ603909S RNZ603910S RNZ603911S RNZ603912S RNZ603914S RNZ603915S RNZ603917S RNZ603918S RNZ603920S RNZ603921S RNZ603929S RNZ603950S RNZ603951S RNZ603955S RNZ603957S RNZ603962S RNZ603963S RNZ603971S RNZ603853S RNZ603862S RNZ603865S RNZ603867S RNZ603875S RNZ603876S RNZ603878S RNZ603880S RNZ603881S RNZ603882S RNZ603883S RNZ603884S RNZ603885S RNZ603886S RNZ603889S RNZ603890S RNZ603891S RNZ603892S RNZ603893S RNZ603895S RNZ603902S RNZ603907S RNZ603913S RNZ603916S RNZ603919S RNZ603923S RNZ603930S RNZ603931S RNZ603933S RNZ603934S RNZ603935S RNZ603937S RNZ603938S RNZ603939S RNZ603940S RNZ603941S RNZ603942S RNZ603943S RNZ603944S RNZ603945S RNZ603946S RNZ603947S RNZ603948S RNZ603949S RNZ603952S RNZ603953S RNZ603954S RNZ603956S RNZ603958S RNZ603961S RNZ603964S RNZ603965S RNZ603966S RNZ603968S RNZ603969S RNZ603970S RNZ603973S RNZ603975S RNZ603976S RNZ603978S RNZ603980S RNZ603981S RNZ603982S RNZ603984S RNZ603985S RNZ603987S RNZ603994S RNZ603995S RNZ603996S RNZ603998S RNZ604002S RNZ604003S RNZ604004S RNZ604010S RNZ604011S RNZ604012S RNZ603979S RNZ604013S RNZ604014S RNZ604015S RNZ604016S RNZ604017S RNZ604018S RNZ604019S RNZ604021S RNZ604023S RNZ604024S RNZ604027S RNZ604028S RNZ604029S RNZ604030S RNZ604032S RNZ604034S RNZ604038S RNZ604039S RNZ604041S RNZ604044S RNZ604045S RNZ604047S RNZ604053S RNZ604054S RNZ604055S RNZ604056S RNZ604057S RNZ604058S RNZ604059S RNZ604060S RNZ604061S RNZ604062S RNZ604063S RNZ604064S RNZ604065S RNZ604066S RNZ604067S RNZ604068S RNZ604069S RNZ604070S RNZ604071S RNZ604072S RNZ604073S RNZ604075S RNZ604076S RNZ604077S RNZ604078S RNZ604079S RNZ604082S RNZ604083S RNZ604085S RNZ604025S RNZ604031S RNZ604033S RNZ604035S RNZ604036S RNZ604040S RNZ604042S RNZ604043S RNZ604046S RNZ604049S RNZ604050S RNZ604080S RNZ604081S RNZ604037S RNZ604048S RNZ604086S RNZ604089S RNZ604090S RNZ604091S RNZ604092S RNZ604093S RNZ604094S RNZ604095S RNZ604096S RNZ604097S RNZ604098S RNZ604099S RNZ604100S RNZ604101S RNZ604102S RNZ604103S RNZ604104S RNZ604105S RNZ604106S RNZ604107S RNZ604108S RNZ604109S RNZ604110S RNZ604111S RNZ604112S RNZ604115S RNZ604116S RNZ604117S RNZ604118S RNZ604119S RNZ604120S RNZ604121S RNZ604122S RNZ604123S RNZ604124S RNZ604125S RNZ604126S RNZ604127S RNZ604128S RNZ604129S RNZ604130S RNZ604131S RNZ604132S RNZ604133S RNZ604134S RNZ604135S RNZ604136S RNZ604138S RNZ604139S RNZ604140S RNZ604141S RNZ604142S RNZ604143S RNZ604144S RNZ604145S RNZ604147S RNZ604148S RNZ604149S RNZ604150S RNZ604151S RNZ604152S RNZ604153S RNZ604154S RNZ604155S RNZ604156S RNZ604157S RNZ604158S RNZ604159S RNZ604160S RNZ604161S RNZ604162S RNZ604163S RNZ604164S RNZ604165S RNZ604006S RNZ604074S RNZ603204S RNZ603490S RNZ603787S RNZ604146S RNZ604169S RNZ604171S RNZ604172S RNZ604179S RNZ604180S RNZ604183S RNZ604184S RNZ604185S RNZ604186S RNZ604187S RNZ604189S RNZ604190S RNZ604191S RNZ604192S RNZ604193S RNZ604194S RNZ604195S RNZ604196S RNZ604198S RNZ604200S RNZ604201S RNZ604202S RNZ604208S RNZ604209S RNZ604215S RNZ604220S RNZ604221S RNZ604222S RNZ604173S RNZ604197S RNZ604203S RNZ604204S RNZ604205S RNZ604216S RNZ604175S RNZ604178S RNZ603176S RNZ604167S RNZ604188S RNZ604199S RNZ604206S RNZ604211S RNZ604212S RNZ604218S RNZ604166S RNZ604170S RNZ604174S RNZ604176S RNZ604177S RNZ604181S RNZ604182S RNZ604210S RNZ604213S RNZ604214S RNZ604217S RNZ604219S RNZ604223S RNZ604224S RNZ604232S RNZ604233S RNZ604236S RNZ604237S RNZ604240S RNZ604241S RNZ604242S RNZ604245S RNZ604248S RNZ604250S RNZ604251S RNZ604254S RNZ604255S RNZ604256S RNZ604261S RNZ604263S RNZ604264S RNZ604268S RNZ604270S RNZ604272S RNZ604281S RNZ604283S RNZ604284S RNZ604285S RNZ604225S RNZ604226S RNZ604227S RNZ604228S RNZ604229S RNZ604230S RNZ604231S RNZ604234S RNZ604238S RNZ604239S RNZ604243S RNZ604246S RNZ604247S RNZ604249S RNZ604252S RNZ604257S RNZ604259S RNZ604262S RNZ604265S RNZ604266S RNZ604267S RNZ604271S RNZ604273S RNZ604275S RNZ604277S RNZ604278S RNZ604280S RNZ604282S RNZ604286S RNZ604287S RNZ604289S RNZ604290S RNZ604291S RNZ604292S RNZ604293S RNZ604294S RNZ604295S RNZ604296S RNZ604297S RNZ604298S RNZ604299S RNZ604300S RNZ604301S RNZ604302S RNZ604303S RNZ604304S RNZ604305S RNZ604306S RNZ604307S RNZ604308S RNZ604309S RNZ604310S RNZ604311S RNZ604312S RNZ604313S RNZ604315S RNZ604316S RNZ604318S RNZ604320S RNZ604321S RNZ604322S RNZ604323S RNZ604324S RNZ604326S RNZ604327S RNZ604328S RNZ604329S RNZ604330S RNZ604332S RNZ604333S RNZ604334S RNZ604335S RNZ604336S RNZ604337S RNZ604338S RNZ604339S RNZ604340S RNZ604341S RNZ604288S RNZ604325S RNZ604276S RNZ604342S RNZ604343S RNZ604344S RNZ604345S RNZ604346S RNZ604347S RNZ604348S RNZ604349S RNZ604351S RNZ604352S RNZ604353S RNZ604355S RNZ604356S RNZ604357S RNZ604358S RNZ604359S RNZ604361S RNZ604363S RNZ604364S RNZ604365S RNZ604367S RNZ604368S RNZ604371S RNZ604372S RNZ604373S RNZ604374S RNZ604376S RNZ604378S RNZ604379S RNZ604380S RNZ604381S RNZ604382S RNZ604383S RNZ604384S RNZ604385S RNZ604386S RNZ604319S RNZ604360S RNZ604362S RNZ604366S RNZ604369S RNZ604370S RNZ604375S RNZ604377S RNZ604354S RNZ604387S RNZ604388S RNZ604389S RNZ604390S RNZ604391S RNZ604392S RNZ604393S RNZ604394S RNZ604396S RNZ604397S RNZ604398S RNZ604399S RNZ604400S RNZ604401S RNZ604402S RNZ604404S RNZ604405S RNZ604406S RNZ604408S RNZ604410S RNZ604411S RNZ604412S RNZ604409S RNZ604415S RNZ604413S RNZ604417S RNZ604418S RNZ604419S RNZ604420S RNZ604421S RNZ604422S RNZ604423S RNZ604424S RNZ604425S RNZ604426S RNZ604427S RNZ604428S RNZ604429S RNZ604430S RNZ604431S RNZ604432S RNZ604433S RNZ604435S RNZ604436S RNZ604438S RNZ604439S RNZ604440S RNZ604441S RNZ604444S RNZ604445S RNZ604448S RNZ604451S RNZ604452S RNZ604453S RNZ604454S RNZ604455S RNZ604456S RNZ604457S RNZ604458S RNZ604459S RNZ604460S RNZ604461S RNZ604462S RNZ604463S RNZ604464S RNZ604465S RNZ604467S RNZ604469S RNZ604470S RNZ604472S RNZ604475S RNZ604476S RNZ604434S RNZ604437S RNZ604443S RNZ604442S RNZ602834S RNZ602837S RNZ602840S RNZ603175S RNZ603503S RNZ603506S RNZ603509S RNZ604113S RNZ604114S RNZ604450S RNZ604481S RNZ604486S RNZ604487S RNZ604489S RNZ604490S RNZ604491S RNZ604492S RNZ604493S RNZ604497S RNZ604498S RNZ604503S RNZ604478S RNZ604496S RNZ604504S RNZ604512S RNZ604515S RNZ604519S RNZ604522S RNZ604524S RNZ604531S RNZ604532S RNZ604537S RNZ604538S RNZ604541S RNZ604542S RNZ604545S RNZ604546S RNZ604547S RNZ604477S RNZ604479S RNZ604482S RNZ604484S RNZ604488S RNZ604494S RNZ604495S RNZ604499S RNZ604501S RNZ604509S RNZ604511S RNZ604513S RNZ604514S RNZ604520S RNZ604526S RNZ604530S RNZ604533S RNZ604544S RNZ604548S RNZ604549S RNZ604550S RNZ604551S RNZ604552S RNZ604555S RNZ604500S RNZ604510S RNZ604521S RNZ604525S RNZ604528S RNZ604534S RNZ604536S RNZ604539S RNZ604553S RNZ604556S RNZ604558S RNZ604561S RNZ604563S RNZ604566S RNZ604567S RNZ604569S RNZ604572S RNZ604573S RNZ604578S RNZ604580S RNZ604581S RNZ604583S RNZ604585S RNZ604587S RNZ604618S RNZ604627S RNZ604628S RNZ604629S RNZ604630S RNZ604631S RNZ604632S RNZ604635S RNZ604636S RNZ604637S RNZ604638S RNZ604641S RNZ604642S RNZ604644S RNZ604645S RNZ604646S RNZ604647S RNZ604650S RNZ604656S RNZ604659S RNZ604662S RNZ604663S RNZ604664S RNZ604665S RNZ604669S RNZ604671S RNZ604672S RNZ604673S RNZ604560S RNZ604570S RNZ604582S RNZ604584S RNZ604586S RNZ604589S RNZ604590S RNZ604591S RNZ604592S RNZ604593S RNZ604594S RNZ604595S RNZ604596S RNZ604597S RNZ604598S RNZ604599S RNZ604600S RNZ604601S RNZ604602S RNZ604603S RNZ604605S RNZ604608S RNZ604619S RNZ604634S RNZ604648S RNZ604655S RNZ604554S RNZ604559S RNZ604564S RNZ604565S RNZ604571S RNZ604579S RNZ604588S RNZ604607S RNZ604609S RNZ604610S RNZ604611S RNZ604612S RNZ604615S RNZ604616S RNZ604617S RNZ604623S RNZ604625S RNZ604649S RNZ604651S RNZ604652S RNZ604653S RNZ604654S RNZ604657S RNZ604667S RNZ604670S RNZ604568S RNZ604613S RNZ604614S RNZ604621S RNZ604624S RNZ604485S RNZ604676S RNZ604677S RNZ604678S RNZ604679S RNZ604681S RNZ604684S RNZ604686S RNZ604687S RNZ604688S RNZ604689S RNZ604690S RNZ604691S RNZ604692S RNZ604693S RNZ604694S RNZ604715S RNZ604718S RNZ604719S RNZ604722S RNZ604725S RNZ604726S RNZ604733S RNZ604743S RNZ604696S RNZ604697S RNZ604698S RNZ604699S RNZ604700S RNZ604701S RNZ604703S RNZ604704S RNZ604705S RNZ604706S RNZ604707S RNZ604708S RNZ604710S RNZ604711S RNZ604712S RNZ604713S RNZ604714S RNZ604716S RNZ604717S RNZ604721S RNZ604723S RNZ604724S RNZ604727S RNZ604728S RNZ604729S RNZ604730S RNZ604731S RNZ604734S RNZ604735S RNZ604736S RNZ604737S RNZ604739S RNZ604740S RNZ604741S RNZ604742S RNZ604745S RNZ604746S RNZ604747S RNZ604748S RNZ604749S RNZ604750S RNZ604751S RNZ604752S RNZ604753S RNZ604754S RNZ604695S RNZ604744S RNZ604505S RNZ604702S RNZ604666S RNZ604762S RNZ604771S RNZ604772S RNZ604773S RNZ604774S RNZ604777S RNZ604779S RNZ604780S RNZ604782S RNZ604783S RNZ604785S RNZ604786S RNZ604787S RNZ604788S RNZ604790S RNZ604791S RNZ604792S RNZ604732S RNZ604755S RNZ604756S RNZ604757S RNZ604759S RNZ604763S RNZ604764S RNZ604765S RNZ604766S RNZ604767S RNZ604768S RNZ604795S RNZ604796S RNZ604797S RNZ604798S RNZ604799S RNZ604800S RNZ604801S RNZ604802S RNZ604804S RNZ604805S RNZ604806S RNZ604807S RNZ604808S RNZ604809S RNZ604810S RNZ604811S RNZ604812S RNZ604813S RNZ604814S RNZ604815S RNZ604816S RNZ604817S RNZ604818S RNZ604819S RNZ604820S RNZ604821S RNZ604822S RNZ604823S RNZ604824S RNZ604825S RNZ604826S RNZ604827S RNZ604828S RNZ604829S RNZ604830S RNZ604831S RNZ604832S RNZ604738S RNZ604758S RNZ604760S RNZ604761S RNZ604769S RNZ604770S RNZ604775S RNZ604784S RNZ604789S RNZ604793S RNZ604794S RNZ604833S RNZ604557S RNZ604575S RNZ604577S RNZ604834S RNZ604835S RNZ604836S RNZ604840S RNZ604841S RNZ604843S RNZ604849S RNZ604854S RNZ604857S RNZ604858S RNZ604859S RNZ604860S RNZ604861S RNZ604862S RNZ604864S RNZ604866S RNZ604871S RNZ604875S RNZ604876S RNZ604877S RNZ604878S RNZ604881S RNZ604882S RNZ604883S RNZ604884S RNZ604885S RNZ604886S RNZ604887S RNZ604888S RNZ604889S RNZ604890S RNZ604891S RNZ604892S RNZ604893S RNZ604837S RNZ604838S RNZ604839S RNZ604842S RNZ604844S RNZ604845S RNZ604846S RNZ604847S RNZ604848S RNZ604851S RNZ604853S RNZ604855S RNZ604856S RNZ604863S RNZ604865S RNZ604867S RNZ604868S RNZ604869S RNZ604870S RNZ604872S RNZ604873S RNZ604874S RNZ604880S RNZ604900S RNZ604894S RNZ604896S RNZ604897S RNZ604898S RNZ604899S RNZ604901S RNZ604903S RNZ604904S RNZ604905S RNZ604906S RNZ604907S RNZ604908S RNZ604909S RNZ604910S RNZ604911S RNZ604912S RNZ604913S RNZ604914S RNZ604915S RNZ604917S RNZ604918S RNZ604919S RNZ604920S RNZ604921S RNZ604922S RNZ604923S RNZ604924S RNZ604925S RNZ604926S RNZ604927S RNZ604929S RNZ604930S RNZ604931S RNZ604932S RNZ604933S RNZ604934S RNZ604935S RNZ604936S RNZ604937S RNZ604939S RNZ604940S RNZ604941S RNZ604942S RNZ604943S RNZ604916S RNZ604950S RNZ604958S RNZ604967S RNZ604968S RNZ604969S RNZ604973S RNZ604974S RNZ604975S RNZ604945S RNZ604947S RNZ604948S RNZ604949S RNZ604952S RNZ604953S RNZ604954S RNZ604955S RNZ604956S RNZ604959S RNZ604961S RNZ604962S RNZ604963S RNZ604964S RNZ604965S RNZ604970S RNZ604971S RNZ604972S RNZ604976S RNZ604977S RNZ604978S RNZ604979S RNZ604980S RNZ604981S RNZ604982S RNZ604983S RNZ604944S RNZ604984S RNZ604987S RNZ604995S RNZ604990S RNZ604991S RNZ604992S RNZ604994S RNZ604998S RNZ604999S RNZ605000S RNZ605002S RNZ604562S RNZ604960S RNZ604997S RNZ605005S RNZ605006S RNZ605007S RNZ605008S RNZ605010S RNZ605011S RNZ605012S RNZ605013S RNZ605016S RNZ605017S RNZ605019S RNZ605022S RNZ605023S RNZ605024S RNZ605025S RNZ605026S RNZ605027S RNZ605028S RNZ605029S RNZ605031S RNZ605032S RNZ605033S RNZ605034S RNZ605035S RNZ605038S RNZ605039S RNZ605040S RNZ605041S RNZ605043S RNZ605044S RNZ605045S RNZ605046S RNZ605047S RNZ605049S RNZ605051S RNZ605052S RNZ605053S RNZ605054S RNZ605056S RNZ605057S RNZ605058S RNZ605059S RNZ605061S RNZ605062S RNZ605063S RNZ605065S RNZ605066S RNZ605067S RNZ605068S RNZ605071S RNZ605072S RNZ605064S RNZ605070S RNZ605073S RNZ605075S RNZ605076S RNZ605077S RNZ605078S RNZ605079S RNZ605080S RNZ605082S RNZ605083S RNZ605084S RNZ605085S RNZ605086S RNZ605088S RNZ605089S RNZ605090S RNZ605091S RNZ605092S RNZ605093S RNZ605094S RNZ605095S RNZ605096S RNZ605097S RNZ605098S RNZ605099S RNZ605100S RNZ605101S RNZ605102S RNZ605104S RNZ605105S RNZ605106S RNZ605107S RNZ605108S RNZ605109S RNZ605110S RNZ605111S RNZ605112S RNZ605113S RNZ604902S RNZ605015S RNZ605115S RNZ605116S RNZ605117S RNZ605119S RNZ605120S RNZ605122S RNZ605123S RNZ605124S RNZ605125S RNZ605126S RNZ605127S RNZ605128S RNZ605129S RNZ605130S RNZ605131S RNZ605132S RNZ605133S RNZ605134S RNZ605135S RNZ605139S RNZ605140S RNZ605141S RNZ605142S RNZ605143S RNZ605118S RNZ605121S RNZ605136S RNZ605137S RNZ605144S RNZ605145S RNZ605146S RNZ605147S RNZ605148S RNZ605149S RNZ605150S RNZ605151S RNZ605152S RNZ605138S RNZ605153S RNZ605154S RNZ605161S RNZ605162S RNZ605163S RNZ605165S RNZ605166S RNZ605168S RNZ605169S RNZ605170S RNZ605172S RNZ605187S RNZ605188S RNZ605191S RNZ605156S RNZ605157S RNZ605158S RNZ605159S RNZ605160S RNZ605164S RNZ605167S RNZ605171S RNZ605173S RNZ605174S RNZ605175S RNZ605176S RNZ605177S RNZ605179S RNZ605180S RNZ605181S RNZ605182S RNZ605183S RNZ605184S RNZ605185S RNZ605186S RNZ605189S RNZ605190S RNZ605192S RNZ605178S RNZ605194S RNZ605195S RNZ605196S RNZ605197S RNZ605199S RNZ605200S RNZ605201S RNZ605204S RNZ605205S RNZ605206S RNZ605207S RNZ605208S RNZ605209S RNZ605210S RNZ605213S RNZ605214S RNZ605215S RNZ605216S RNZ605217S RNZ604658S RNZ604781S RNZ605193S RNZ605202S RNZ605203S RNZ605211S RNZ605212S RNZ605219S RNZ605221S RNZ605231S RNZ605232S RNZ605234S RNZ605235S RNZ605236S RNZ605238S RNZ605239S RNZ605240S RNZ605241S RNZ605242S RNZ605243S RNZ605244S RNZ605245S RNZ605247S RNZ605248S RNZ605250S RNZ605253S RNZ605255S RNZ605258S RNZ605260S RNZ605261S RNZ605262S RNZ605263S RNZ605264S RNZ605265S RNZ605267S RNZ605268S RNZ605271S RNZ605266S RNZ605270S RNZ605272S RNZ605273S RNZ605275S RNZ605281S RNZ605284S RNZ605285S RNZ605287S RNZ605289S RNZ605290S RNZ605291S RNZ605292S RNZ605293S RNZ605294S RNZ605295S RNZ605296S RNZ605298S RNZ605299S RNZ605305S RNZ605331S RNZ605332S RNZ605333S RNZ605336S RNZ605338S RNZ605346S RNZ605349S RNZ605350S RNZ605352S RNZ605353S RNZ605354S RNZ605356S RNZ605357S RNZ605303S RNZ605307S RNZ605309S RNZ605310S RNZ605311S RNZ605314S RNZ605315S RNZ605316S RNZ605317S RNZ605318S RNZ605319S RNZ605320S RNZ605323S RNZ605351S RNZ605361S RNZ605362S RNZ605365S RNZ605366S RNZ605367S RNZ605368S RNZ605369S RNZ605370S RNZ605371S RNZ605372S RNZ605373S RNZ605374S RNZ605375S RNZ605376S RNZ605377S RNZ605378S RNZ605379S RNZ605380S RNZ605301S RNZ605302S RNZ605304S RNZ605306S RNZ605312S RNZ605313S RNZ605322S RNZ605325S RNZ605327S RNZ605328S RNZ605329S RNZ605340S RNZ605342S RNZ605343S RNZ605347S RNZ605348S RNZ605400S RNZ605401S RNZ605408S RNZ605420S RNZ605421S RNZ605422S RNZ605423S RNZ605425S RNZ605426S RNZ605427S RNZ605428S RNZ605430S RNZ605431S RNZ605432S RNZ605433S RNZ605434S RNZ605435S RNZ605436S RNZ605437S RNZ605438S RNZ605439S RNZ605441S RNZ605442S RNZ605443S RNZ605444S RNZ605445S RNZ605446S RNZ605447S RNZ605448S RNZ605449S RNZ605450S RNZ605451S RNZ605452S RNZ605453S RNZ605454S RNZ605455S RNZ605456S RNZ605381S RNZ605382S RNZ605383S RNZ605384S RNZ605385S RNZ605386S RNZ605387S RNZ605388S RNZ605389S RNZ605391S RNZ605392S RNZ605393S RNZ605394S RNZ605396S RNZ605397S RNZ605398S RNZ605403S RNZ605404S RNZ605405S RNZ605406S RNZ605409S RNZ605410S RNZ605411S RNZ605412S RNZ605413S RNZ605414S RNZ605415S RNZ605416S RNZ605417S RNZ605418S RNZ605419S RNZ605399S RNZ605440S RNZ605460S RNZ605461S RNZ605464S RNZ605469S RNZ605470S RNZ605472S RNZ605473S RNZ605475S RNZ605476S RNZ605478S RNZ605480S RNZ605482S RNZ605483S RNZ605484S RNZ605489S RNZ605490S RNZ605503S RNZ605505S RNZ605507S RNZ605515S RNZ605462S RNZ605465S RNZ605466S RNZ605467S RNZ605468S RNZ605471S RNZ605477S RNZ605481S RNZ605485S RNZ605487S RNZ605488S RNZ605491S RNZ605492S RNZ605494S RNZ605495S RNZ605496S RNZ605497S RNZ605498S RNZ605499S RNZ605500S RNZ605502S RNZ605506S RNZ605508S RNZ605509S RNZ605510S RNZ605511S RNZ605512S RNZ605516S RNZ605517S RNZ605518S RNZ605519S RNZ605521S RNZ605522S RNZ605523S RNZ605524S RNZ605525S RNZ605526S RNZ605527S RNZ605528S RNZ605530S RNZ605531S RNZ605532S RNZ605533S RNZ605534S RNZ605536S RNZ605538S RNZ605520S RNZ605540S RNZ605541S RNZ605542S RNZ605543S RNZ605545S RNZ605546S RNZ605547S RNZ605548S RNZ605551S RNZ605553S RNZ605556S RNZ605559S RNZ605539S RNZ605544S RNZ605552S RNZ605554S RNZ605555S RNZ605557S RNZ605561S RNZ605563S RNZ605564S RNZ605565S RNZ605566S RNZ605567S RNZ605568S RNZ605569S RNZ605570S RNZ605571S RNZ605572S RNZ605573S RNZ605574S RNZ605575S RNZ605578S RNZ605579S RNZ605580S RNZ605581S RNZ605582S RNZ605583S RNZ605585S RNZ605586S RNZ605587S RNZ605588S RNZ605589S RNZ605592S RNZ605593S RNZ605594S RNZ605595S RNZ605596S RNZ605597S RNZ605598S RNZ605549S RNZ605584S RNZ605599S RNZ605602S RNZ605603S RNZ605604S RNZ605605S RNZ605606S RNZ605607S RNZ605608S RNZ605609S RNZ605610S RNZ605611S RNZ605612S RNZ605613S RNZ605614S RNZ605615S RNZ605616S RNZ605617S RNZ605618S RNZ605620S RNZ605622S RNZ605623S RNZ605624S RNZ605625S RNZ605626S RNZ605627S RNZ605628S RNZ605630S RNZ605631S RNZ605632S RNZ605633S RNZ605635S RNZ605636S RNZ605637S RNZ605638S RNZ605639S RNZ605640S RNZ605641S RNZ605643S RNZ605646S RNZ605629S RNZ605513S RNZ605621S RNZ605649S RNZ605657S RNZ605661S RNZ605651S RNZ605652S RNZ605655S RNZ605659S RNZ605660S RNZ605662S RNZ605663S RNZ605664S RNZ605665S RNZ605666S RNZ605671S RNZ605672S RNZ605673S RNZ605674S RNZ605675S RNZ605676S RNZ605677S RNZ605678S RNZ605680S RNZ605691S RNZ605692S RNZ605699S RNZ605700S RNZ605703S RNZ605705S RNZ605708S RNZ605709S RNZ605712S RNZ605714S RNZ605718S RNZ605719S RNZ605720S RNZ605721S RNZ605722S RNZ605732S RNZ605738S RNZ605744S RNZ605710S RNZ605713S RNZ605717S RNZ605724S RNZ605725S RNZ605729S RNZ605736S RNZ605740S RNZ605741S RNZ605749S RNZ605750S RNZ605751S RNZ605752S RNZ605754S RNZ605757S RNZ605758S RNZ605759S RNZ605760S RNZ605761S RNZ605762S RNZ605763S RNZ605764S RNZ605788S RNZ605756S RNZ605770S RNZ605772S RNZ605773S RNZ605783S RNZ605784S RNZ605785S RNZ605786S RNZ605789S RNZ605790S RNZ605791S RNZ605792S RNZ605793S RNZ605794S RNZ605795S RNZ605796S RNZ605797S RNZ605798S RNZ605799S RNZ605801S RNZ605802S RNZ605803S RNZ605804S RNZ605806S RNZ605807S RNZ605808S RNZ605809S RNZ605810S RNZ605811S RNZ605812S RNZ605813S RNZ605814S RNZ605815S RNZ605816S RNZ605817S RNZ605818S RNZ605819S RNZ605820S RNZ605821S RNZ605822S RNZ605823S RNZ605825S RNZ605828S RNZ605831S RNZ605837S RNZ605838S RNZ605843S RNZ605844S RNZ605845S RNZ605846S RNZ605847S RNZ605848S RNZ605868S RNZ605869S RNZ605870S RNZ605871S RNZ605872S RNZ605873S RNZ605875S RNZ605876S RNZ605877S RNZ605880S RNZ605881S RNZ605882S RNZ605890S RNZ605893S RNZ605899S RNZ605905S RNZ605906S RNZ605924S RNZ605849S RNZ605850S RNZ605851S RNZ605852S RNZ605853S RNZ605855S RNZ605857S RNZ605860S RNZ605862S RNZ605863S RNZ605864S RNZ605865S RNZ605928S RNZ605858S RNZ605829S RNZ605936S RNZ605937S RNZ605943S RNZ605947S RNZ605951S RNZ605952S RNZ605954S RNZ605959S RNZ605960S RNZ605966S RNZ605972S RNZ605975S RNZ605977S RNZ605980S RNZ605981S RNZ605988S RNZ605992S RNZ605997S RNZ605998S RNZ606003S RNZ606015S RNZ606018S RNZ606025S RNZ606027S RNZ605945S RNZ605946S RNZ605963S RNZ605964S RNZ605965S RNZ605982S RNZ605983S RNZ605994S RNZ605995S RNZ605999S RNZ606002S RNZ606004S RNZ606007S RNZ606010S RNZ606011S RNZ606014S RNZ606016S RNZ606017S RNZ606020S RNZ605827S RNZ605958S RNZ605993S RNZ606001S RNZ606005S RNZ606006S RNZ606026S RNZ606028S RNZ606032S RNZ606033S RNZ606034S RNZ606035S RNZ606036S RNZ606037S RNZ606038S RNZ606039S RNZ606040S RNZ606041S RNZ606044S RNZ606045S RNZ606046S RNZ606048S RNZ605969S RNZ605985S |
Recalling Firm/ Manufacturer |
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) 8200 Coral Sea St Ne Mounds View MN 55112-4391
|
For Additional Information Contact |
Technical Services 800-638-1991
|
Manufacturer Reason for Recall |
There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions. |
Quantity in Commerce |
16667 devices |
Distribution |
worldwide |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database |
PMAs with Product Code = NVZ and Original Applicant = MEDTRONIC Inc.
|
|
|
|