• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall HLS Set Advanced

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall HLS Set Advanced see related information
Date Initiated by Firm July 22, 2022
Create Date September 19, 2022
Recall Status1 Open3, Classified
Recall Number Z-1766-2022
Recall Event ID 90740
510(K)Number K112360  
Product Classification Oxygenator, cardiopulmonary bypass - Product Code DTZ
Product BEQ-HLS 5050 USA;
HLS Set Advanced 5.0
Part No. 70106.9077.
A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.
Code Information UDI-DI: 04058863076355 Lots 3000192081, 3000193363, 3000196200, 3000208697, 3000216172
Recalling Firm/
Manufacturer
Maquet Medical Systems USA
45 Barbour Pond Dr
Wayne NJ 07470-2094
For Additional Information Contact Allison Jean Kaplan
973-709-7000
Manufacturer Reason
for Recall
The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).
FDA Determined
Cause 2
Package design/selection
Action On July 15, 2022, the firm notified affected customers via Urgent Medical Device Removal Expansion letter. The letter informed customers of the inclusion of additional product codes and lots into the earlier recall initiated on November 6, 2020. Customers were instructed to examine their inventory and identify any affected product on hand. Affected product in use may be continued to be used according to normal practices. Unused and unexpired affected product should be immediately returned to Getinge for credit. Please contact Getinge Customer Service at (888) 9GETUSA / (888) 943-8872 (press option 2) between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. At this time, the recalling firm has already implemented the more durable Tyvek accessory packaging in all product manufactured September 2021 and later.
Quantity in Commerce 6,871 US; 18,670 worldwide
Distribution Worldwide distribution - US Nationwide and the countries Canada and Australia.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = DTZ and Original Applicant = MAQUET CARDIOPULMONARY AG
-
-