• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall medtronic see related information
Date Initiated by Firm May 10, 2023
Date Posted June 16, 2023
Recall Status1 Open3, Classified
Recall Number Z-1717-2023
Recall Event ID 92328
PMA Number P980016S653 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
Product ICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators

Code Information GTIN 00763000001483, Lot Serial Numbers: CWM603134S, CWM600439S, CWM601269S, CWM603840S, CWM603922S, CWM601037S, CWM602990S, CWM603292S, CWM602888S, CWM602567S, CWM604350S, CWM604400S, CWM604147S, CWM604447S, CWM603919S, CWM604293S, CWM603993S, CWM604453S, CWM604276S, CWM603779S, CWM603512S, CWM604434S, CWM604110S, CWM603974S, CWM604303S, CWM604451S, CWM604296S, CWM603643S, CWM604349S, CWM603935S, CWM603697S, CWM603897S, CWM604404S, CWM603551S, CWM601796S, CWM601798S, CWM601800S, CWM601801S, CWM601802S, CWM601905S, CWM601911S, CWM601931S, CWM602000S, CWM602003S, CWM602004S, CWM602005S, CWM602006S, CWM602097S, CWM602098S, CWM602124S, CWM602139S, CWM602709S, CWM602710S, CWM602711S, CWM602823S, CWM602836S, CWM602838S, CWM602847S, CWM602848S, CWM602883S, CWM602898S, CWM603276S, CWM600958S, CWM601164S, CWM601353S, CWM601410S, CWM601640S, CWM601831S, CWM602141S, CWM603581S, CWM603582S, CWM603627S, CWM603628S, CWM604291S, CWM604292S, CWM601211S, CWM601263S, CWM601339S, CWM601506S, CWM601641S, CWM601775S, CWM601830S, CWM601834S, CWM601838S, CWM601839S, CWM601969S, CWM601974S, CWM601975S, CWM601982S, CWM602156S, CWM602313S, CWM602566S, CWM602573S, CWM602580S, CWM602662S, CWM602663S, CWM602864S, CWM602921S, CWM603097S, CWM603100S, CWM603191S, CWM603193S, CWM603334S, CWM603336S, CWM603361S, CWM603421S, CWM603445S, CWM603446S, CWM603573S, CWM603574S, CWM603714S, CWM603717S, CWM603718S, CWM603719S, CWM603720S, CWM603721S, CWM603722S, CWM603723S, CWM603927S, CWM603928S, CWM603930S, CWM603933S, CWM603934S, CWM603937S, CWM603939S, CWM603940S, CWM604396S, CWM604397S, CWM604398S, CWM603776S, CWM603777S, CWM603799S, CWM603765S, CWM603734S, CWM603795S, CWM603797S, CWM603798S, CWM603775S, CWM602189S, CWM602287S, CWM602291S, CWM602828S, CWM602829S, CWM602833S, CWM602271S, CWM603033S, CWM603689S, CWM601460S, CWM602068S, CWM601398S, CWM601455S, CWM601835S, CWM602588S, CWM603187S, CWM602165S, CWM601754S, CWM602951S, CWM603416S, CWM601079S, CWM601128S, CWM602397S, CWM603979S, CWM601709S, CWM604028S, CWM603474S, CWM600962S, CWM601207S, CWM602680S, CWM603052S, CWM600294S, CWM600649S, CWM601208S, CWM602132S, CWM600054S, CWM600672S, CWM602014S, CWM603386S, CWM602061S, CWM601633S, CWM603039S, CWM602787S, CWM602220S, CWM602270S, CWM601387S, CWM603584S, CWM601192S, CWM600669S, CWM600712S, CWM601707S, CWM602211S, CWM602941S, CWM603079S, CWM603681S, CWM602945S, CWM602947S, CWM603881S, CWM602082S, CWM603085S, CWM603423S, CWM604111S, CWM600500S, CWM600768S, CWM601710S, CWM602157S, CWM603619S, CWM600523S, CWM600716S, CWM601168S, CWM602154S, CWM602514S, CWM602191S, CWM603391S, CWM603028S, CWM603274S, CWM603014S, CWM602568S, CWM601256S, CWM601624S, CWM602102S, CWM600356S, CWM603119S, CWM604038S, CWM601700S, CWM602430S, CWM603194S, CWM600455S, CWM602815S, CWM601954S, CWM603244S, CWM600680S, CWM600802S, CWM600993S, CWM601312S, CWM601521S, CWM603154S, CWM603889S, CWM600316S, CWM600520S, CWM600710S, CWM600729S, CWM600752S, CWM601788S, CWM602512S, CWM602513S, CWM603068S, CWM603363S, CWM600476S, CWM601008S, CWM601075S, CWM601229S, CWM602416S, CWM602776S, CWM603790S, CWM600416S, CWM600497S, CWM602417S, CWM602449S, CWM602451S, CWM603007S, CWM603947S, CWM600480S, CWM600561S, CWM600793S, CWM601432S, CWM601863S, CWM601980S, CWM603016S, CWM603733S, CWM602559S, CWM602011S, CWM602250S, CWM602340S, CWM602267S, CWM600576S, CWM600638S, CWM600691S, CWM600779S, CWM601070S, CWM601213S, CWM601708S, CWM601758S, CWM601762S, CWM602522S, CWM603074S, CWM603258S, CWM602487S, CWM600059S, CWM600061S, CWM600313S, CWM600427S, CWM600433S, CWM600436S, CWM600733S, CWM602163S, CWM603257S, CWM600725S, CWM601500S, CWM600397S, CWM600584S, CWM600648S, CWM600651S, CWM603125S, CWM601095S, CWM601570S, CWM602443S, CWM602790S, CWM602862S, CWM603708S, CWM603973S, CWM601516S, CWM603186S, CWM602302S, CWM603146S, CWM600749S, CWM600285S, CWM600467S, CWM600468S, CWM600522S, CWM600557S, CWM600658S, CWM600661S, CWM600723S, CWM600977S, CWM601020S, CWM601081S, CWM601400S, CWM601727S, CWM601853S, CWM601913S, CWM602484S, CWM602739S, CWM602937S, CWM603160S, CWM603309S, CWM603349S, CWM604123S, CWM600029S, CWM600030S, CWM600031S, CWM600034S, CWM600038S, CWM600041S, CWM600043S, CWM600079S, CWM600090S, CWM600092S, CWM600293S, CWM600309S, CWM600314S, CWM600337S, CWM600368S, CWM600378S, CWM600391S, CWM600392S, CWM600394S, CWM600426S, CWM600429S, CWM600430S, CWM600435S, CWM600454S, CWM600491S, CWM600554S, CWM600568S, CWM600666S, CWM600709S, CWM600717S, CWM600783S, CWM601028S, CWM601116S, CWM601177S, CWM601178S, CWM601188S, CWM601318S, CWM601699S, CWM601856S, CWM601994S, CWM602029S, CWM602221S, CWM603012S, CWM603069S, CWM603313S, CWM603417S, CWM603546S, CWM603547S, CWM600629S, CWM600503S, CWM600504S, CWM600688S, CWM601491S, CWM601787S, CWM602237S, CWM602666S, CWM602668S, CWM603107S, CWM603222S, CWM603518S, CWM603556S, CWM603793S, CWM603571S, CWM603692S, CWM603887S, CWM603888S, CWM600362S, CWM600369S, CWM600370S, CWM600421S, CWM600470S, CWM600510S, CWM600528S, CWM600635S, CWM600684S, CWM600742S, CWM600743S, CWM600745S, CWM600979S, CWM601223S, CWM601628S, CWM603102S, CWM603298S, CWM603558S, CWM603951S, CWM600010S, CWM600304S, CWM600339S, CWM600444S, CWM600621S, CWM602655S, CWM600437S, CWM601126S, CWM601348S, CWM601452S, CWM601594S, CWM601595S, CWM601828S, CWM602145S, CWM602146S, CWM603048S, CWM601071S, CWM601073S, CWM601074S, CWM601151S, CWM601390S, CWM601706S, CWM601756S, CWM601772S, CWM601773S, CWM601996S, CWM602697S, CWM602771S, CWM602772S, CWM602854S, CWM603057S, CWM603116S, CWM603362S, CWM603437S, CWM602316S, CWM602320S, CWM602465S, CWM602526S, CWM602569S, CWM602994S, CWM603056S, CWM603150S, CWM603460S, CWM603894S, CWM603895S, CWM604122S, CWM600017S, CWM600018S, CWM600320S, CWM600736S, CWM601009S, CWM601124S, CWM601127S, CWM602123S, CWM602423S, CWM603725S, CWM603783S, CWM600469S, CWM600501S, CWM600509S, CWM600517S, CWM600564S, CWM600730S, CWM600777S, CWM600990S, CWM601603S, CWM602252S, CWM602293S, CWM603377S, CWM602933S, CWM600978S, CWM601157S, CWM601240S, CWM601333S, CWM601403S, CWM601457S, CWM601829S, CWM602196S, CWM602774S, CWM602810S, CWM603343S, CWM603992S, CWM601179S, CWM602292S, CWM600325S, CWM600327S, CWM600328S, CWM600330S, CWM600374S, CWM600400S, CWM600438S, CWM600496S, CWM600519S, CWM600632S, CWM600668S, CWM600675S, CWM601047S, CWM601402S, CWM601465S, CWM601522S, CWM601618S, CWM601842S, CWM601949S, CWM602478S, CWM603022S, CWM603027S, CWM603086S, CWM604022S, CWM601317S, CWM601319S, CWM601345S, CWM601445S, CWM601453S, CWM601510S, CWM601906S, CWM602440S, CWM602834S, CWM602995S, CWM603078S, CWM603788S, CWM604135S, CWM601963S, CWM602100S, CWM604311S, CWM601205S, CWM601206S, CWM602166S, CWM602284S, CWM602899S, CWM603127S, CWM603424S, CWM603792S, CWM600800S, CWM600013S, CWM600014S, CWM600015S, CWM600387S, CWM600533S, CWM602404S, CWM602410S, CWM600690S, CWM600095S, CWM600284S, CWM600478S, CWM600512S, CWM600737S, CWM600805S, CWM600806S, CWM601005S, CWM601171S, CWM602084S, CWM603071S, CWM603152S, CWM603425S, CWM603635S, CWM603639S, CWM602236S, CWM600357S, CWM600295S, CWM600332S, CWM600472S, CWM600481S, CWM600524S, CWM600525S, CWM600634S, CWM601763S, CWM601765S, CWM602282S, CWM602283S, CWM600301S, CWM600302S, CWM600331S, CWM600335S, CWM600343S, CWM600350S, CWM600352S, CWM600384S, CWM600448S, CWM600463S, CWM600464S, CWM600490S, CWM600521S, CWM600560S, CWM600617S, CWM600622S, CWM600630S, CWM600683S, CWM600689S, CWM600721S, CWM600753S, CWM600780S, CWM600790S, CWM600801S, CWM600814S, CWM600960S, CWM601003S, CWM601012S, CWM601072S, CWM601077S, CWM601102S, CWM601107S, CWM601133S, CWM601135S, CWM601138S, CWM601139S, CWM601169S, CWM601184S, CWM601196S, CWM601320S, CWM601325S, CWM601392S, CWM601434S, CWM601578S, CWM601590S, CWM601598S, CWM601753S, CWM601883S, CWM601924S, CWM601961S, CWM602024S, CWM602025S, CWM602049S, CWM602107S, CWM602215S, CWM602275S, CWM602303S, CWM602305S, CWM602307S, CWM602374S, CWM602460S, CWM602574S, CWM602658S, CWM603041S, CWM603325S, CWM603957S, CWM602671S, CWM600359S, CWM600591S, CWM600592S, CWM600593S, CWM600594S, CWM600596S, CWM600597S, CWM600642S, CWM600643S, CWM600645S, CWM600646S, CWM600647S, CWM600652S, CWM600653S, CWM600655S, CWM600678S, CWM600679S, CWM600695S, CWM600765S, CWM600771S, CWM600774S, CWM600776S, CWM600781S, CWM600782S, CWM600784S, CWM600795S, CWM600799S, CWM600821S, CWM600975S, CWM600976S, CWM600982S, CWM600983S, CWM600985S, CWM601014S, CWM601015S, CWM601016S, CWM601017S, CWM601018S, CWM601026S, CWM601031S, CWM601035S, CWM601038S, CWM601039S, CWM601040S, CWM601041S, CWM601043S, CWM601044S, CWM601051S, CWM601054S, CWM601055S, CWM601056S, CWM601057S, CWM601058S, CWM601059S, CWM601060S, CWM601061S, CWM601063S, CWM601065S, CWM601066S, CWM601080S, CWM601092S, CWM601096S, CWM601098S, CWM601099S, CWM601101S, CWM601104S, CWM601105S, CWM601109S, CWM601111S, CWM601112S, CWM601113S, CWM601142S, CWM601143S, CWM601149S, CWM601165S, CWM601174S, CWM601180S, CWM601190S, CWM601191S, CWM601209S, CWM601224S, CWM601227S, CWM601257S, CWM601258S, CWM601259S, CWM601261S, CWM601271S, CWM601274S, CWM601276S, CWM601277S, CWM601322S, CWM601326S, CWM601327S, CWM601328S, CWM601329S, CWM601332S, CWM601334S, CWM601335S, CWM601343S, CWM601349S, CWM601352S, CWM601355S, CWM601363S, CWM601366S, CWM601368S, CWM601369S, CWM601372S, CWM601382S, CWM601384S, CWM601409S, CWM601411S, CWM601412S, CWM601493S, CWM601494S, CWM601495S, CWM601498S, CWM601503S, CWM601504S, CWM601505S, CWM601597S, CWM601637S, CWM601638S, CWM601701S, CWM601702S, CWM601770S, CWM601811S, CWM601852S, CWM601867S, CWM601891S, CWM601910S, CWM601912S, CWM601933S, CWM601997S, CWM602009S, CWM602010S, CWM602018S, CWM602019S, CWM602022S, CWM602027S, CWM602028S, CWM602042S, CWM602057S, CWM602058S, CWM602074S, CWM602083S, CWM602113S, CWM602217S, CWM602336S, CWM600492S, CWM600493S, CWM600494S, CWM600567S, CWM601273S, CWM601464S, CWM601468S, CWM600023S, CWM600626S, CWM600791S, CWM600961S, CWM602121S, CWM603291S, CWM604003S, CWM604039S, CWM600371S, CWM601407S, CWM601496S, CWM601501S, CWM603490S, CWM602071S, CWM603494S, CWM601013S, CWM602852S, CWM602855S, CWM603564S, CWM601951S, CWM600028S, CWM600036S, CWM600303S, CWM600329S, CWM600389S, CWM600432S, CWM600508S, CWM600549S, CWM600565S, CWM600713S, CWM600772S, CWM600989S, CWM601091S, CWM601175S, CWM601189S, CWM601490S, CWM601513S, CWM601589S, CWM601734S, CWM601959S, CWM601960S, CWM601991S, CWM602101S, CWM602137S, CWM602248S, CWM602673S, CWM602782S, CWM603046S, CWM603333S, CWM603521S, CWM603929S, CWM604438S, CWM604449S, CWM600338S, CWM600395S, CWM600422S, CWM601992S, CWM602241S, CWM600091S, CWM600363S, CWM600365S, CWM600551S, CWM600552S, CWM602017S, CWM602072S, CWM603493S, CWM603506S, CWM601957S, CWM600580S, CWM600582S, CWM600640S, CWM600759S, CWM601114S, CWM601215S, CWM601396S, CWM601454S, CWM601512S, CWM601602S, CWM601625S, CWM601726S, CWM601730S, CWM601783S, CWM601850S, CWM601896S, CWM601927S, CWM601976S, CWM602013S, CWM602040S, CWM602104S, CWM602111S, CWM602118S, CWM602429S, CWM602441S, CWM602446S, CWM602456S, CWM602520S, CWM602607S, CWM602608S, CWM602670S, CWM602735S, CWM602740S, CWM602747S, CWM602804S, CWM602814S, CWM602837S, CWM602843S, CWM603006S, CWM603132S, CWM603428S, CWM603529S, CWM603587S, CWM603633S, CWM603835S, CWM603898S, CWM603900S, CWM603972S, CWM604448S, CWM600595S, CWM600614S, CWM600615S, CWM601097S, CWM602466S, CWM603114S, CWM603570S, CWM600452S, CWM600453S, CWM600506S, CWM600507S, CWM600530S, CWM600531S, CWM600548S, CWM600550S, CWM601275S, CWM601330S, CWM602008S, CWM600068S, CWM600069S, CWM600307S, CWM600308S, CWM600310S, CWM600312S, CWM602021S, CWM602405S, CWM601351S, CWM604012S, CWM600390S, CWM600495S, CWM601048S, CWM601050S, CWM602853S, CWM604094S, CWM600024S, CWM600305S, CWM600355S, CWM600991S, CWM601818S, CWM601826S, CWM601900S, CWM601903S, CWM601939S, CWM602059S, CWM602414S, CWM600027S, CWM600033S, CWM601459S, CWM600319S, CWM600048S, CWM600407S, CWM600420S, CWM600460S, CWM600757S, CWM601234S, CWM601894S, CWM602439S, CWM602665S, CWM602769S, CWM602839S, CWM602885S, CWM603021S, CWM603128S, CWM603536S, CWM603695S, CWM603971S, CWM600451S, CWM600660S, CWM600685S, CWM600994S, CWM601089S, CWM601150S, CWM602866S, CWM604287S, CWM601108S, CWM601110S, CWM601216S, CWM601217S, CWM601218S, CWM601236S, CWM601237S, CWM601239S, CWM601242S, CWM601517S, CWM601576S, CWM601621S, CWM601868S, CWM602105S, CWM602610S, CWM603651S, CWM603654S, CWM603907S, CWM604008S, CWM604440S, CWM601001S, CWM601002S, CWM601878S, CWM602342S, CWM603685S, CWM603899S, CWM600058S, CWM600354S, CWM601152S, CWM600639S, CWM601577S, CWM601591S, CWM601616S, CWM601617S, CWM601623S, CWM602134S, CWM602135S, CWM602213S, CWM602214S, CWM603427S, CWM603429S, CWM600654S, CWM600726S, CWM600785S, CWM601134S, CWM601172S, CWM601220S, CWM601723S, CWM602455S, CWM602587S, CWM602919S, CWM603204S, CWM603948S, CWM600981S, CWM600995S, CWM602055S, CWM602946S, CWM602954S, CWM602996S, CWM603303S, CWM603305S, CWM600042S, CWM600044S, CWM600045S, CWM600046S, CWM600047S, CWM600050S, CWM600071S, CWM600073S, CWM600078S, CWM600298S, CWM600318S, CWM600336S, CWM600344S, CWM600367S, CWM600386S, CWM600413S, CWM600724S, CWM601033S, CWM601084S, CWM601085S, CWM601086S, CWM601088S, CWM601117S, CWM601118S, CWM601119S, CWM601121S, CWM601146S, CWM601147S, CWM601246S, CWM601268S, CWM601418S, CWM601419S, CWM601451S, CWM601629S, CWM601703S, CWM601704S, CWM601748S, CWM601832S, CWM601833S, CWM601925S, CWM602032S, CWM602033S, CWM602034S, CWM602195S, CWM603250S, CWM603632S, CWM601648S, CWM601650S, CWM600057S, CWM600340S, CWM600418S, CWM600499S, CWM600513S, CWM600541S, CWM600572S, CWM600727S, CWM600775S, CWM600804S, CWM600836S, CWM600841S, CWM601622S, CWM601950S, CWM603096S, CWM603879S, CWM602987S, CWM603431S, CWM603590S, CWM604040S, CWM600505S, CWM600707S, CWM601582S, CWM601737S, CWM602299S, CWM603070S, CWM603138S, CWM603589S, CWM600599S, CWM600601S, CWM600732S, CWM600734S, CWM600824S, CWM600827S, CWM600828S, CWM600829S, CWM600830S, CWM600831S, CWM600832S, CWM600833S, CWM600834S, CWM600835S, CWM600965S, CWM600966S, CWM600967S, CWM600968S, CWM600969S, CWM600970S, CWM600971S, CWM600972S, CWM600973S, CWM600974S, CWM602354S, CWM602356S, CWM602360S, CWM602361S, CWM602363S, CWM602364S, CWM602365S, CWM602366S, CWM602369S, CWM602370S, CWM602959S, CWM602960S, CWM602961S, CWM602962S, CWM602963S, CWM603326S, CWM603329S, CWM603331S, CWM603335S, CWM603337S, CWM603647S, CWM603648S, CWM603649S, CWM603650S, CWM603960S, CWM603961S, CWM603962S, CWM603963S, CWM603964S, CWM603966S, CWM603967S, CWM603968S, CWM603969S, CWM603970S, CWM602597S, CWM603304S, CWM603548S, CWM603572S, CWM603699S, CWM603876S, CWM604281S, CWM603449S, CWM600663S, CWM601187S, CWM601781S, CWM602200S, CWM602419S, CWM603062S, CWM603526S, CWM603694S, CWM603943S, CWM600032S, CWM600807S, CWM600808S, CWM600810S, CWM600811S, CWM600812S, CWM600844S, CWM600963S, CWM601156S, CWM601630S, CWM601917S, CWM601918S, CWM602038S, CWM602039S, CWM602091S, CWM602867S, CWM602870S, CWM602930S, CWM603136S, CWM603252S, CWM603646S, CWM601881S, CWM601899S, CWM602147S, CWM602148S, CWM602495S, CWM602500S, CWM602589S, CWM603092S, CWM603285S, CWM603906S, CWM601137S, CWM601226S, CWM601936S, CWM602950S, CWM603693S, CWM603980S, CWM604092S, CWM601154S, CWM600609S, CWM601646S, CWM604285S, CWM601971S, CWM601197S, CWM600566S, CWM601620S, CWM601722S, CWM602467S, CWM603035S, CWM603393S, CWM600471S, CWM600364S, CWM600380S, CWM600526S, CWM600538S, CWM600662S, CWM600673S, CWM601441S, CWM602115S, CWM602150S, CWM602310S, CWM602845S, CWM603032S, CWM601264S, CWM601729S, CWM602266S, CWM602338S, CWM602579S, CWM602942S, CWM603950S, CWM603978S, CWM604006S, CWM602242S, CWM603595S, CWM603759S, CWM600574S, CWM601163S, CWM602490S, CWM601032S, CWM601173S, CWM602398S, CWM603457S, CWM601010S, CWM603496S, CWM602319S, CWM600692S, CWM600502S, CWM600636S, CWM603023S, CWM602259S, CWM603617S, CWM601193S, CWM602458S, CWM602476S, CWM602477S, CWM602577S, CWM602944S, CWM603599S, CWM603676S, CWM603913S, CWM604131S, CWM601943S, CWM601214S, CWM602159S, CWM602326S, CWM602332S, CWM602334S, CWM602348S, CWM602600S, CWM602601S, CWM603011S, CWM603072S, CWM604432S, CWM604455S, CWM601858S, CWM602116S, CWM602335S, CWM601631S, CWM601078S, CWM601647S, CWM602219S, CWM602432S, CWM601746S, CWM601757S, CWM602687S, CWM602253S, CWM602254S, CWM602454S, CWM603044S, CWM603956S, CWM602119S, CWM602120S, CWM602450S, CWM602480S, CWM602481S, CWM602811S, CWM602813S, CWM601170S, CWM601255S, CWM601394S, CWM601636S, CWM601736S, CWM602222S, CWM602351S, CWM602533S, CWM602986S, CWM603248S, CWM603455S, CWM603677S, CWM604129S, CWM604130S, CWM602399S, CWM602688S, CWM604418S, CWM601076S, CWM601774S, CWM602349S, CWM602511S, CWM602736S, CWM602999S, CWM603236S, CWM603301S, CWM603347S, CWM603678S, CWM603896S, CWM604405S, CWM600479S, CWM603323S, CWM600578S, CWM600605S, CWM600665S, CWM601265S, CWM603954S, CWM601514S, CWM601515S, CWM603544S, CWM602210S, CWM603288S, CWM601019S, CWM602244S, CWM603527S, CWM603545S, CWM602424S, CWM602989S, CWM603624S, CWM601062S, CWM601449S, CWM602062S, CWM602412S, CWM602415S, CWM603282S, CWM601244S, CWM601735S, CWM603535S, CWM601769S, CWM602422S, CWM602788S, CWM603192S, CWM601499S, CWM601599S, CWM601749S, CWM602260S, CWM602604S, CWM603622S, CWM602886S, CWM603010S, CWM602558S, CWM603504S, CWM600664S, CWM603202S, CWM600537S, CWM600587S, CWM601225S, CWM601721S, CWM602103S, CWM601212S, CWM601424S, CWM601527S, CWM602158S, CWM603118S, CWM603925S, CWM604093S, CWM600819S, CWM601421S, CWM602337S, CWM600722S, CWM601948S, CWM603005S, CWM603095S, CWM600579S, CWM601993S, CWM602099S, CWM603108S, CWM603009S, CWM600598S, CWM602872S, CWM602519S, CWM602784S, CWM600442S, CWM602106S, CWM600641S, CWM602518S, CWM600514S, CWM600704S, CWM603352S, CWM601024S, CWM602602S, CWM602956S, CWM600532S, CWM600671S, CWM600813S, CWM600840S, CWM600842S, CWM601162S, CWM601405S, CWM601436S, CWM602317S, CWM602327S, CWM602698S, CWM602051S, CWM600769S, CWM600477S, CWM601626S, CWM601717S, CWM602527S, CWM602798S, CWM603273S, CWM603310S, CWM603549S, CWM601579S, CWM600778S, CWM600474S, CWM600475S, CWM601767S, CWM601887S, CWM601889S, CWM601967S, CWM603126S, CWM601235S, CWM600529S, CWM601030S, CWM603281S, CWM600536S, CWM602759S, CWM602682S, CWM603112S, CWM600404S, CWM600559S, CWM601130S, CWM601587S, CWM601644S, CWM601946S, CWM602255S, CWM602322S, CWM602463S, CWM602531S, CWM603198S, CWM603004S, CWM603008S, CWM602048S, CWM602434S, CWM602570S, CWM603597S, CWM603683S, CWM600516S, CWM600562S, CWM601153S, CWM601926S, CWM601999S, CWM602479S, CWM603370S, CWM600489S, CWM600518S, CWM600542S, CWM600547S, CWM600571S, CWM601383S, CWM601435S, CWM601511S, CWM601724S, CWM601860S, CWM602198S, CWM602491S, CWM602948S, CWM602949S, CWM603003S, CWM603110S, CWM603533S, CWM603534S, CWM603567S, CWM603944S, CWM603945S, CWM604103S, CWM604139S, CWM604409S, CWM600425S, CWM600534S, CWM601526S, CWM601797S, CWM602830S, CWM602482S, CWM602936S, CWM603998S, CWM604433S, CWM600845S, CWM603498S, CWM602065S, CWM600677S, CWM600987S, CWM601195S, CWM601508S, CWM601985S, CWM602924S, CWM603142S, CWM603221S, CWM603365S, CWM601262S, CWM602330S, CWM602681S, CWM602952S, CWM603025S, CWM604034S, CWM604037S, CWM600756S, CWM601199S, CWM600539S, CWM600604S, CWM601635S, CWM600446S, CWM600458S, CWM601167S, CWM601443S, CWM601492S, CWM600402S, CWM600457S, CWM600486S, CWM600618S, CWM602427S, CWM604001S, CWM600585S, CWM600711S, CWM601083S, CWM601203S, CWM604124S, CWM603203S, CWM601181S, CWM601254S, CWM603592S, CWM602620S, CWM602841S, CWM603983S, CWM601344S, CWM601386S, CWM601592S, CWM601983S, CWM603230S, CWM602060S, CWM601847S, CWM600449S, CWM600714S, CWM600837S, CWM601094S, CWM601720S, CWM602122S, CWM602796S, CWM600569S, CWM600650S, CWM600674S, CWM600687S, CWM600738S, CWM600770S, CWM601004S, CWM601090S, CWM601219S, CWM601401S, CWM601425S, CWM601426S, CWM601518S, CWM601529S, CWM601619S, CWM601715S, CWM601916S, CWM601987S, CWM601998S, CWM602041S, CWM602160S, CWM602161S, CWM602298S, CWM602400S, CWM602461S, CWM602485S, CWM602789S, CWM602860S, CWM603058S, CWM603225S, CWM603531S, CWM603579S, CWM601253S, CWM602715S, CWM602525S, CWM602840S, CWM602746S, CWM603162S, CWM602281S, CWM603080S, CWM602190S, CWM603295S, CWM603317S, CWM601200S, CWM602077S, CWM602081S, CWM602257S, CWM602890S, CWM603275S, CWM603332S, CWM603269S, CWM602667S, CWM602795S, CWM600686S, CWM600706S, CWM600748S, CWM601194S, CWM601198S, CWM601423S, CWM601461S, CWM601642S, CWM601728S, CWM601848S, CWM601866S, CWM602044S, CWM602235S, CWM603067S, CWM603517S, CWM603704S, CWM604116S, CWM603520S, CWM602344S, CWM603682S, CWM603949S, CWM601323S, CWM601393S, CWM600409S, CWM600456S, CWM600405S, CWM601952S, CWM601752S, CWM602261S, CWM602756S, CWM603781S, CWM603015S, CWM603353S, CWM602696S, CWM602794S, CWM602188S, CWM602431S, CWM603905S, CWM601166S, CWM602557S, CWM602889S, CWM603528S, CWM601249S, CWM602618S, CWM602506S, CWM603188S, CWM603440S, CWM604002S, CWM604114S, CWM601573S, CWM601989S, CWM602807S, CWM601336S, CWM602037S, CWM602197S, CWM602272S, CWM602343S, CWM603560S, CWM600815S, CWM604023S, CWM600005S, CWM600019S, CWM600074S, CWM600299S, CWM600322S, CWM600351S, CWM600577S, CWM600612S, CWM601228S, CWM601230S, CWM601241S, CWM601245S, CWM601272S, CWM601359S, CWM601944S, CWM603189S, CWM603578S, CWM604424S, CWM602706S, CWM602713S, CWM602714S, CWM602716S, CWM602718S, CWM602052S, CWM600667S, CWM601021S, CWM601159S, CWM601160S, CWM601161S, CWM604043S, CWM604393S, CWM604395S, CWM604492S, CWM604493S, CWM603348S, CWM602030S, CWM601955S, CWM602152S, CWM602488S, CWM603210S, CWM601313S, CWM601718S, CWM603050S, CWM601747S, CWM601523S, CWM602358S, CWM600573S, CWM600750S, CWM600751S, CWM600754S, CWM600755S, CWM600760S, CWM600762S, CWM600764S, CWM600766S, CWM601042S, CWM601046S, CWM601052S, CWM601321S, CWM601364S, CWM601759S, CWM601803S, CWM602001S, CWM602755S, CWM602757S, CWM603771S, CWM602269S, CWM602521S, CWM603977S, CWM604095S, CWM604096S, CWM604097S, CWM600631S, CWM601141S, CWM601347S, CWM602859S, CWM601610S, CWM602929S, CWM603550S, CWM600697S, CWM600746S, CWM600761S, CWM600818S, CWM600846S, CWM601007S, CWM601022S, CWM601053S, CWM601202S, CWM601231S, CWM601233S, CWM601248S, CWM601280S, CWM601337S, CWM601391S, CWM601427S, CWM601433S, CWM601593S, CWM601596S, CWM601713S, CWM601719S, CWM601966S, CWM601981S, CWM601990S, CWM602026S, CWM602047S, CWM602078S, CWM602090S, CWM602092S, CWM602308S, CWM602309S, CWM602353S, CWM602372S, CWM602373S, CWM602442S, CWM602445S, CWM602475S, CWM602725S, CWM602731S, CWM602799S, CWM602800S, CWM602803S, CWM602917S, CWM602940S, CWM603018S, CWM603019S, CWM603037S, CWM603104S, CWM603148S, CWM603390S, CWM603448S, CWM603488S, CWM603513S, CWM603514S, CWM603532S, CWM603542S, CWM603616S, CWM603713S, CWM603784S, CWM603916S, CWM603917S, CWM603953S, CWM603959S, CWM603986S, CWM604004S, CWM604005S, CWM604085S, CWM604087S, CWM604090S, CWM604443S, CWM604444S, CWM604445S, CWM604446S, CWM604423S, CWM600089S, CWM600297S, CWM600321S, CWM600583S, CWM600708S, CWM600796S, CWM601106S, CWM601115S, CWM601125S, CWM601953S, CWM603559S, CWM602216S, CWM603255S, CWM601185S, CWM601186S, CWM601509S, CWM602265S, CWM603508S, CWM603509S, CWM604430S, CWM601266S, CWM601995S, CWM601182S, CWM602704S, CWM603318S, CWM603226S, CWM603539S, CWM603629S, CWM601940S, CWM602758S, CWM602993S, CWM602295S, CWM602609S, CWM602770S, CWM602939S, CWM602572S, CWM600681S, CWM602472S, CWM604091S, CWM601456S, CWM602087S, CWM602926S, CWM601799S, CWM602856S, CWM603507S, CWM602012S, CWM602690S, CWM603196S, CWM600306S, CWM602153S, CWM603054S, CWM603272S, CWM603350S, CWM603576S, CWM602878S, CWM601855S, CWM603034S, CWM600718S, CWM600763S, CWM601314S, CWM601447S, CWM601462S, CWM601588S, CWM601609S, CWM601725S, CWM602043S, CWM602117S, CWM602325S, CWM602781S, CWM603211S, CWM602306S, CWM600358S, CWM602093S, CWM602925S, CWM602928S, CWM601148S, CWM601201S, CWM601311S, CWM601417S, CWM601732S, CWM602493S, CWM602918S, CWM603161S, CWM603322S, CWM601250S, CWM604015S, CWM604410S, CWM601442S, CWM600075S, CWM600076S, CWM600077S, CWM600080S, CWM600083S, CWM600084S, CWM600085S, CWM600087S, CWM600093S, CWM600623S, CWM600624S, CWM600625S, CWM600627S, CWM600628S, CWM600998S, CWM600999S, CWM601000S, CWM601502S, CWM601507S, CWM601519S, CWM601520S, CWM601524S, CWM601525S, CWM601571S, CWM601572S, CWM601583S, CWM601586S, CWM601600S, CWM601604S, CWM601608S, CWM601634S, CWM601643S, CWM601645S, CWM601782S, CWM601784S, CWM601785S, CWM601786S, CWM602421S, CWM602629S, CWM602998S, CWM601738S, CWM600996S, CWM601315S, CWM601316S, CWM603487S, CWM603580S, CWM603741S, CWM603742S, CWM603800S, CWM604019S, CWM600008S, CWM600445S, CWM600515S, CWM600613S, CWM600633S, CWM601252S, CWM601267S, CWM601837S, CWM601849S, CWM602311S, CWM602582S, CWM603388S, CWM603732S, CWM604356S, CWM602109S, CWM602110S, CWM602581S, CWM602656S, CWM603709S, CWM604025S, CWM603430S, CWM600787S, CWM600788S, CWM602504S, CWM601093S, CWM600287S, CWM600288S, CWM600289S, CWM600290S, CWM600291S, CWM600292S, CWM600334S, CWM600345S, CWM600348S, CWM600381S, CWM600382S, CWM600450S, CWM600482S, CWM600483S, CWM600346S, CWM600347S, CWM600349S, CWM600544S, CWM600558S, CWM601270S, CWM601331S, CWM601406S, CWM601463S, CWM601466S, CWM601467S, CWM601469S, CWM602070S, CWM602401S, CWM602402S, CWM602403S, CWM602406S, CWM602407S, CWM602408S, CWM602409S, CWM603491S, CWM603492S, CWM602224S, CWM603051S, CWM603053S, CWM603489S, CWM603522S, CWM603523S, CWM603524S, CWM603538S, CWM603591S, CWM603706S, CWM603731S, CWM602080S, CWM602114S, CWM603707S, CWM602225S, CWM601611S, CWM601612S, CWM601613S, CWM601614S, CWM601615S, CWM601740S, CWM601742S, CWM601743S, CWM601892S, CWM602561S, CWM602563S, CWM602575S, CWM602693S, CWM602857S, CWM603497S, CWM603724S, CWM603762S, CWM602453S, CWM600062S, CWM600070S, CWM600072S, CWM600082S, CWM600296S, CWM600376S, CWM600443S, CWM600459S, CWM600485S, CWM600611S, CWM600698S, CWM600700S, CWM600797S, CWM600798S, CWM602321S, CWM602676S, CWM602934S, CWM602935S, CWM603081S, CWM603555S, CWM604041S, CWM600067S, CWM600317S, CWM600440S, CWM600607S, CWM600682S, CWM600957S, CWM601731S, CWM601964S, CWM602056S, CWM602223S, CWM602315S, CWM602806S, CWM602892S, CWM602923S, CWM602932S, CWM603017S, CWM603495S, CWM600461S, CWM600462S, CWM601404S, CWM601438S, CWM602510S, CWM600037S, CWM600039S, CWM600053S, CWM600081S, CWM600373S, CWM600377S, CWM600586S, CWM600644S, CWM600728S, CWM600822S, CWM601011S, CWM601158S, CWM601210S, CWM601399S, CWM601601S, CWM601836S, CWM602088S, CWM602089S, CWM602149S, CWM602922S, CWM603908S, CWM601431S, CWM600286S, CWM600414S, CWM600012S, CWM600051S, CWM600415S, CWM600703S, CWM602226S, CWM602229S, CWM602230S, CWM602231S, CWM602232S, CWM602233S, CWM602669S, CWM602819S, CWM602820S, CWM602821S, CWM602822S, CWM602824S, CWM602965S, CWM603064S, CWM603066S, CWM603247S, CWM603367S, CWM603481S, CWM603482S, CWM603485S, CWM603568S, CWM603569S, CWM603588S, CWM604014S, CWM600341S, CWM600602S, CWM602339S, CWM602997S, CWM603228S, CWM601605S, CWM601606S, CWM601607S, CWM601639S, CWM601733S, CWM602396S, CWM603155S, CWM603158S, CWM603159S, CWM602802S, CWM603923S, CWM601864S, CWM601340S, CWM601341S, CWM601342S, CWM602208S, CWM602444S, CWM602816S, CWM603030S, CWM603031S, CWM603047S, CWM603055S, CWM603644S, CWM604301S, CWM603087S, CWM603088S, CWM603089S, CWM603093S, CWM603094S, CWM603099S, CWM603103S, CWM603105S, CWM603106S, CWM603109S, CWM603111S, CWM603113S, CWM603115S, CWM603120S, CWM603121S, CWM603124S, CWM603271S, CWM603302S, CWM603306S, CWM603307S, CWM603308S, CWM603311S, CWM603312S, CWM603314S, CWM603320S, CWM603321S, CWM603324S, CWM603327S, CWM603328S, CWM603330S, CWM603378S, CWM603381S, CWM603383S, CWM603384S, CWM603387S, CWM604053S, CWM604058S, CWM604059S, CWM604060S, CWM604061S, CWM604062S, CWM604063S, CWM604064S, CWM604065S, CWM604066S, CWM604067S, CWM604068S, CWM604069S, CWM604070S, CWM604071S, CWM604072S, CWM604073S, CWM604074S, CWM604075S, CWM604076S, CWM601103S, CWM602426S, CWM601705S, CWM601776S, CWM601778S, CWM601780S, CWM601897S, CWM601973S, CWM602818S, CWM603024S, CWM601140S, CWM601844S, CWM603218S, CWM603220S, CWM603264S, CWM603824S, CWM601370S, CWM601371S, CWM601373S, CWM601806S, CWM601840S, CWM603151S, CWM603805S, CWM603828S, CWM604057S, CWM603268S, CWM601977S, CWM603766S, CWM600311S, CWM600423S, CWM600535S, CWM600540S, CWM600546S, CWM600702S, CWM601789S, CWM601922S, CWM602578S, CWM602659S, CWM603575S, CWM603924S, CWM604136S, CWM604324S, CWM604325S, CWM604326S, CWM604411S, CWM604412S, CWM603137S, CWM604082S, CWM600484S, CWM602459S, CWM602471S, CWM602530S, CWM602686S, CWM603626S, CWM604024S, CWM604048S, CWM602023S, CWM603594S, CWM601792S, CWM602586S, CWM602108S, CWM602187S, CWM602274S, CWM602278S, CWM602324S, CWM602565S, CWM603379S, CWM603618S, CWM603736S, CWM603827S, CWM601845S, CWM600964S, CWM601941S, CWM604399S, CWM601176S, CWM600065S, CWM600066S, CWM600300S, CWM600465S, CWM600498S, CWM600739S, CWM601388S, CWM602079S, CWM602328S, CWM602413S, CWM602678S, CWM602894S, CWM602953S, CWM603060S, CWM603296S, CWM603686S, CWM600016S, CWM600086S, CWM600992S, CWM601029S, CWM601251S, CWM601408S, CWM601627S, CWM601716S, CWM601915S, CWM602053S, CWM602483S, CWM602612S, CWM602642S, CWM603525S, CWM603029S, CWM603213S, CWM602296S, CWM602683S, CWM602684S, CWM602694S, CWM602717S, CWM602858S, CWM600986S, CWM601036S, CWM602462S, CWM602066S, CWM602067S, CWM602085S, CWM602290S, CWM602436S, CWM602452S, CWM602126S, CWM602127S, CWM602128S, CWM602129S, CWM602130S, CWM602015S, CWM602691S, CWM603420S, CWM604426S, CWM604427S, CWM604428S, CWM604429S, CWM603811S, CWM603207S, CWM603214S, CWM603280S, CWM603254S, CWM603817S, CWM603253S, CWM603209S, CWM604364S, CWM603810S, CWM603815S, CWM604347S, CWM604370S, CWM602707S, CWM602880S, CWM602881S, CWM603131S, CWM603163S, CWM603165S, CWM603190S, CWM603205S, CWM603206S, CWM603249S, CWM603256S, CWM603747S, CWM603755S, CWM603382S, CWM600020S, CWM600021S, CWM600040S, CWM600324S, CWM600408S, CWM600705S, CWM600997S, CWM601221S, CWM601584S, CWM601585S, CWM602046S, CWM602069S, CWM602448S, CWM602831S, CWM603369S, CWM601711S, CWM602849S, CWM602865S, CWM602876S, CWM603130S, CWM603208S, CWM603259S, CWM603749S, CWM603753S, CWM603754S, CWM603808S, CWM603816S, CWM604346S, CWM604365S, CWM602744S, CWM604345S, CWM602712S, CWM602812S, CWM602884S, CWM602895S, CWM603212S, CWM603251S, CWM603283S, CWM603812S, CWM603813S, CWM604343S, CWM603744S, CWM602623S, CWM602630S, CWM601374S, CWM601377S, CWM601807S, CWM602699S, CWM603216S, CWM603267S, CWM600659S, CWM601414S, CWM601745S, CWM601869S, CWM601914S, CWM601945S, CWM602045S, CWM602155S, CWM602251S, CWM602438S, CWM603219S, CWM603577S, CWM603593S, CWM603873S, CWM604138S, CWM604277S, CWM602988S, CWM603117S, CWM603418S, CWM603516S, CWM603729S, CWM603791S, CWM603926S, CWM603756S, CWM603757S, CWM603838S, CWM600570S, CWM601865S, CWM601904S, CWM601937S, CWM601956S, CWM601970S, CWM602050S, CWM602209S, CWM602245S, CWM602247S, CWM602286S, CWM602329S, CWM602435S, CWM602437S, CWM602657S, CWM602817S, CWM602826S, CWM602844S, CWM602992S, CWM603000S, CWM603002S, CWM603020S, CWM603083S, CWM603217S, CWM601429S, CWM601851S, CWM603063S, CWM603139S, CWM603316S, CWM604344S, CWM604194S, CWM604195S, CWM603484S, CWM604211S, CWM602720S, CWM602748S, CWM604178S, CWM603502S, CWM603503S, CWM604205S, CWM603602S, CWM604042S, CWM604044S, CWM604190S, CWM604202S, CWM604203S, CWM604204S, CWM603466S, CWM603433S, CWM603436S, CWM604463S, CWM604469S, CWM601581S, CWM602615S, CWM604177S, CWM603478S, CWM604214S, CWM604460S, CWM604184S, CWM603465S, CWM603224S, CWM603237S, CWM603239S, CWM603357S, CWM603483S, CWM603486S, CWM603604S, CWM604215S, CWM601886S, CWM602641S, CWM602637S, CWM602777S, CWM603435S, CWM602502S, CWM602874S, CWM603227S, CWM603338S, CWM603439S, CWM603653S, CWM604174S, CWM604187S, CWM602614S, CWM602640S, CWM602507S, CWM603462S, CWM603463S, CWM603932S, CWM602498S, CWM602603S, CWM602584S, CWM603073S, CWM603339S, CWM604201S, CWM602634S, CWM603065S, CWM604179S, CWM604191S, CWM604252S, CWM602621S, CWM602722S, CWM603345S, CWM604185S, CWM602703S, CWM603355S, CWM603356S, CWM603359S, CWM603360S, CWM603443S, CWM603444S, CWM604355S, CWM604387S, CWM604388S, CWM604389S, CWM602517S, CWM602532S, CWM602555S, CWM602556S, CWM602562S, CWM603240S, CWM603242S, CWM603243S, CWM603245S, CWM603375S, CWM602468S, CWM602501S, CWM602728S, CWM602805S, CWM602808S, CWM604210S, CWM604216S, CWM604354S, CWM604392S, CWM603461S, CWM602652S, CWM602653S, CWM603456S, CWM601439S, CWM601243S, CWM601279S, CWM601824S, CWM602496S, CWM603471S, CWM603340S, CWM604196S, CWM604212S, CWM604352S, CWM602624S, CWM602760S, CWM603373S, CWM603374S, CWM604183S, CWM603472S, CWM601356S, CWM601444S, CWM601446S, CWM601450S, CWM601458S, CWM601771S, CWM601777S, CWM601779S, CWM601791S, CWM601793S, CWM601794S, CWM601810S, CWM601812S, CWM601813S, CWM601817S, CWM601819S, CWM601820S, CWM601821S, CWM601822S, CWM601823S, CWM601846S, CWM601854S, CWM601873S, CWM601874S, CWM601877S, CWM601879S, CWM601880S, CWM601898S, CWM601902S, CWM602094S, CWM602096S, CWM602125S, CWM602473S, CWM602635S, CWM602644S, CWM602645S, CWM602646S, CWM602647S, CWM602648S, CWM602649S, CWM602650S, CWM602702S, CWM602705S, CWM602719S, CWM602750S, CWM602875S, CWM602877S, CWM602879S, CWM603229S, CWM603231S, CWM603232S, CWM603342S, CWM603366S, CWM603371S, CWM603395S, CWM603473S, CWM603475S, CWM603501S, CWM603636S, CWM603637S, CWM603638S, CWM603642S, CWM604180S, CWM604182S, CWM604207S, CWM604208S, CWM604390S, CWM604391S, CWM604461S, CWM604462S, CWM603803S, CWM601929S, CWM602194S, CWM602199S, CWM603277S, CWM603293S, CWM604234S, CWM602470S, CWM602651S, CWM604206S, CWM602738S, CWM603769S, CWM602791S, CWM603358S, CWM603467S, CWM602515S, CWM602780S, CWM602927S, CWM603625S, CWM603286S, CWM604197S, CWM602193S, CWM603260S, CWM603263S, CWM603515S, CWM603565S, CWM603712S, CWM603914S, CWM601361S, CWM601362S, CWM601413S, CWM601415S, CWM601416S, CWM601751S, CWM601790S, CWM601814S, CWM601815S, CWM601859S, CWM601870S, CWM601875S, CWM601885S, CWM601888S, CWM601890S, CWM602516S, CWM602528S, CWM602593S, CWM602595S, CWM602619S, CWM602724S, CWM602726S, CWM602727S, CWM602732S, CWM602741S, CWM602751S, CWM602753S, CWM602783S, CWM602785S, CWM602786S, CWM602792S, CWM602793S, CWM602797S, CWM603122S, CWM603364S, CWM603441S, CWM603442S, CWM601825S, CWM603505S, CWM600406S, CWM600428S, CWM600431S, CWM600589S, CWM600600S, CWM600608S, CWM600610S, CWM600758S, CWM600817S, CWM600838S, CWM600988S, CWM601144S, CWM601145S, CWM601574S, CWM601575S, CWM601760S, CWM601761S, CWM601871S, CWM601872S, CWM601988S, CWM602201S, CWM602202S, CWM602277S, CWM602695S, CWM602832S, CWM602920S, CWM603026S, CWM603135S, CWM603140S, CWM603299S, CWM603300S, CWM603553S, CWM603554S, CWM603710S, CWM603711S, CWM603892S, CWM603902S, CWM603903S, CWM604031S, CWM604032S, CWM604052S, CWM604078S, CWM604439S, CWM602288S, CWM603101S, CWM603315S, CWM603319S, CWM602249S, CWM602938S, CWM602964S, CWM603091S, CWM604026S, CWM603606S, CWM603768S, CWM604046S, CWM603586S, CWM600064S, CWM600360S, CWM600361S, CWM600424S, CWM600447S, CWM600488S, CWM600731S, CWM600843S, CWM601129S, CWM601430S, CWM602016S, CWM602300S, CWM602529S, CWM602677S, CWM603036S, CWM603262S, CWM600049S, CWM600088S, CWM600279S, CWM600280S, CWM600281S, CWM600282S, CWM600283S, CWM600315S, CWM600434S, CWM600803S, CWM600984S, CWM601006S, CWM601131S, CWM601183S, CWM601932S, CWM602035S, CWM602775S, CWM603043S, CWM603989S, CWM603991S, CWM602075S, CWM602076S, CWM604036S, CWM602086S, CWM602020S, CWM603084S, CWM604314S, CWM602887S, CWM603082S, CWM603910S, CWM603912S, CWM603987S, CWM603988S, CWM604320S, CWM602036S, CWM603042S, CWM603519S, CWM604406S, CWM604407S, CWM602617S, CWM603234S, CWM603464S, CWM604199S, CWM604213S, CWM602633S, CWM602773S, CWM603770S, CWM602729S, CWM603459S, CWM601346S, CWM602497S, CWM603537S, CWM602723S, CWM602779S, CWM602801S, CWM603351S, CWM603605S, CWM604047S, CWM603767S, CWM603458S, CWM603985S, CWM603061S, CWM603372S, CWM601876S, CWM603141S, CWM603153S, CWM603738S, CWM603804S, CWM603822S, CWM602660S, CWM602661S, CWM603040S, CWM603238S, CWM603540S, CWM603541S, CWM604009S, CWM604010S, CWM604050S, CWM604081S, CWM602596S, CWM602598S, CWM603075S, CWM603341S, CWM603344S, CWM604200S, CWM602505S, CWM601958S, CWM600333S, CWM600581S, CWM601025S, CWM601120S, CWM601397S, CWM602433S, CWM602664S, CWM602893S, CWM602955S, CWM603750S, CWM603752S, CWM603807S, CWM604342S, CWM602835S, CWM603199S, CWM603751S, CWM604209S, CWM604329S, CWM604368S, CWM603735S, CWM603566S, CWM603621S, CWM602654S, CWM603453S, CWM603764S, CWM602031S, CWM602625S, CWM604158S, CWM604045S, CWM603149S, CWM602730S, CWM604417S, CWM602778S, CWM603819S, CWM603611S, CWM602733S, CWM602737S, CWM602745S, CWM602850S, CWM602863S, CWM602868S, CWM602869S, CWM602871S, CWM603468S, CWM603469S, CWM603470S, CWM603499S, CWM603500S, CWM603598S, CWM604334S, CWM602882S, CWM603157S, CWM604337S, CWM602842S, CWM603197S, CWM603200S, CWM603809S, CWM603246S, CWM603279S, CWM603290S, CWM603746S, CWM604338S, CWM604339S, CWM604340S, CWM602576S, CWM601978S, CWM601919S, CWM603278S, CWM603294S, CWM603830S, CWM603870S, CWM603871S, CWM603874S, CWM603875S, CWM603877S, CWM604033S, CWM604080S, CWM604121S, CWM604163S, CWM604181S, CWM604193S, CWM601930S, CWM601934S, CWM602063S, CWM602227S, CWM602289S, CWM602357S, CWM603284S, CWM603289S, CWM603562S, CWM604188S, CWM601923S, CWM603287S, CWM604331S, CWM604332S, CWM600959S, CWM602192S, CWM602768S, CWM600696S, CWM602499S, CWM603235S, CWM603346S, CWM603477S, CWM603479S, CWM604186S, CWM603133S, CWM601354S, CWM601422S, CWM601428S, CWM601437S, CWM601795S, CWM601804S, CWM601805S, CWM601808S, CWM601809S, CWM601816S, CWM601857S, CWM601861S, CWM601862S, CWM601893S, CWM602095S, CWM602138S, CWM602140S, CWM602203S, CWM602205S, CWM602206S, CWM602239S, CWM602469S, CWM602474S, CWM602486S, CWM602503S, CWM602509S, CWM602524S, CWM602585S, CWM602591S, CWM602592S, CWM602599S, CWM602605S, CWM602606S, CWM602611S, CWM602613S, CWM602616S, CWM602622S, CWM602631S, CWM602632S, CWM602636S, CWM602638S, CWM602639S, CWM602643S, CWM602721S, CWM602742S, CWM602743S, CWM602749S, CWM602754S, CWM602761S, CWM602494S, CWM603603S, CWM604217S, CWM601440S, CWM603878S, CWM603883S, CWM603884S, CWM604198S, CWM603354S, CWM603652S, CWM602627S, CWM603215S, CWM603422S, CWM603727S, CWM603728S, CWM603829S, CWM603630S, CWM603763S, CWM604054S, CWM603049S, CWM603147S, CWM603156S, CWM603241S, CWM603266S, CWM603297S, CWM603368S, CWM603376S, CWM603380S, CWM603385S, CWM603389S, CWM603392S, CWM603426S, CWM603438S, CWM603447S, CWM603452S, CWM603454S, CWM603510S, CWM603511S, CWM603552S, CWM603561S, CWM603583S, CWM603612S, CWM603620S, CWM603631S, CWM603634S, CWM603640S, CWM603641S, CWM603645S, CWM603679S, CWM603690S, CWM603696S, CWM603698S, CWM603700S, CWM603701S, CWM603702S, CWM603715S, CWM603716S, CWM603726S, CWM603730S, CWM603737S, CWM603739S, CWM603740S, CWM603745S, CWM603760S, CWM603761S, CWM603789S, CWM603801S, CWM603802S, CWM603806S, CWM603820S, CWM603821S, CWM603823S, CWM603826S, CWM603831S, CWM603834S, CWM603836S, CWM603837S, CWM603839S, CWM603844S, CWM603872S, CWM603890S, CWM603893S, CWM603901S, CWM603904S, CWM603909S, CWM603915S, CWM603920S, CWM603921S, CWM603931S, CWM603936S, CWM603938S, CWM603941S, CWM603946S, CWM603955S, CWM603958S, CWM603965S, CWM603981S, CWM603982S, CWM603984S, CWM603994S, CWM603996S, CWM603997S, CWM604000S, CWM604007S, CWM604013S, CWM604016S, CWM604017S, CWM604018S, CWM604020S, CWM604029S, CWM604030S, CWM604049S, CWM604051S, CWM604056S, CWM604079S, CWM604083S, CWM604084S, CWM604105S, CWM604112S, CWM604117S, CWM604118S, CWM604119S, CWM604120S, CWM604125S, CWM604126S, CWM604128S, CWM604132S, CWM604133S, CWM604137S, CWM604140S, CWM604141S, CWM604142S, CWM604146S, CWM604148S, CWM604159S, CWM604160S, CWM604161S, CWM604192S, CWM604239S, CWM604240S, CWM604244S, CWM604246S, CWM604247S, CWM604248S, CWM604250S, CWM604278S, CWM604307S, CWM604309S, CWM604312S, CWM604313S, CWM604322S, CWM604327S, CWM604328S, CWM604341S, CWM604348S, CWM604366S, CWM604367S, CWM604401S, CWM604403S, CWM604413S, CWM604415S, CWM604416S, CWM604425S, CWM604441S, CWM604442S, CWM604450S, CWM604452S, CWM604235S, CWM604245S, CWM604283S, CWM604284S, CWM604315S, CWM603748S, CWM603814S, CWM603818S, CWM603999S, CWM604175S, CWM604308S, CWM601324S; GTIN 00763000001490, Lot Serial Numbers: CWM600930S, CWM600119S, CWM600152S, CWM600864S, CWM600870S, CWM601541S, CWM601664S, CWM601686S, CWM602167S, CWM603859S, CWM600923S, CWM600938S, CWM600952S, CWM600956S, CWM601281S, CWM601478S, CWM600129S, CWM600188S, CWM600154S, CWM601536S, CWM601667S, CWM602388S, CWM602967S, CWM603667S, CWM601301S, CWM601302S, CWM601676S, CWM601297S, CWM601298S, CWM601307S, CWM601308S, CWM602538S, CWM602542S, CWM602908S, CWM603168S, CWM600227S, CWM600883S, CWM601660S, CWM601671S, CWM603663S, CWM603674S, CWM603855S, CWM604302S, CWM603862S, CWM600178S, CWM600197S, CWM600217S, CWM600225S, CWM600847S, CWM600856S, CWM600878S, CWM600890S, CWM600924S, CWM600937S, CWM601288S, CWM602378S, CWM602379S, CWM602537S, CWM602978S, CWM600097S, CWM600162S, CWM600214S, CWM600226S, CWM601290S, CWM602551S, CWM600268S, CWM600862S, CWM600888S, CWM600891S, CWM600892S, CWM600896S, CWM600903S, CWM600933S, CWM600936S, CWM601299S, CWM601306S, CWM601546S, CWM601548S, CWM601563S, CWM601692S, CWM602912S, CWM603661S, CWM600258S, CWM600135S, CWM601534S, CWM600198S, CWM600218S, CWM600236S, CWM600253S, CWM601294S, CWM602182S, CWM603170S, CWM604269S, CWM600113S, CWM600120S, CWM600126S, CWM600848S, CWM600865S, CWM600238S, CWM600244S, CWM603405S, CWM603659S, CWM603861S, CWM600203S, CWM600204S, CWM600205S, CWM600206S, CWM600213S, CWM600220S, CWM600237S, CWM600248S, CWM600850S, CWM600853S, CWM600854S, CWM600857S, CWM600858S, CWM600859S, CWM600874S, CWM600906S, CWM600911S, CWM600920S, CWM600943S, CWM601653S, CWM602172S, CWM602384S, CWM602545S, CWM602549S, CWM602554S, CWM602914S, CWM602974S, CWM602977S, CWM603867S, CWM600101S, CWM600106S, CWM600109S, CWM600112S, CWM600130S, CWM600133S, CWM600139S, CWM600159S, CWM600192S, CWM600231S, CWM600877S, CWM600926S, CWM601300S, CWM601542S, CWM601670S, CWM603869S, CWM602393S, CWM600944S, CWM601293S, CWM601685S, CWM601694S, CWM602178S, CWM600894S, CWM601550S, CWM602907S, CWM602909S, CWM602970S, CWM604263S, CWM600882S, CWM601483S, CWM601487S, CWM601689S, CWM602173S, CWM602392S, CWM602539S, CWM602902S, CWM602911S, CWM602973S, CWM602975S, CWM603401S, CWM603853S, CWM603854S, CWM603410S, CWM603411S, CWM600912S, CWM601309S, CWM601489S, CWM601673S, CWM602385S, CWM603660S, CWM600210S, CWM600907S, CWM601285S, CWM601554S, CWM601560S, CWM601682S, CWM600199S, CWM600887S, CWM600951S, CWM601544S, CWM601665S, CWM602183S, CWM603664S, CWM603673S, CWM603413S, CWM603172S, CWM600105S, CWM603415S, CWM600127S, CWM600866S, CWM600904S, CWM600918S, CWM602176S, CWM603850S, CWM603402S, CWM603412S, CWM603175S, CWM600180S, CWM600181S, CWM600182S, CWM601679S, CWM602981S, CWM600867S, CWM600876S, CWM600916S, CWM601662S, CWM601675S, CWM602552S, CWM601284S, CWM603180S, CWM604317S, CWM600267S, CWM600272S, CWM600872S, CWM601470S, CWM601566S, CWM604270S, CWM601551S, CWM601691S, CWM600849S, CWM600915S, CWM601471S, CWM601655S, CWM601672S, CWM600229S, CWM600240S, CWM600934S, CWM602903S, CWM603166S, CWM601680S, CWM600209S, CWM600216S, CWM600233S, CWM600243S, CWM600939S, CWM601287S, CWM601533S, CWM601657S, CWM600224S, CWM601473S, CWM602175S, CWM602389S, CWM603849S, CWM601556S, CWM601659S, CWM604261S, CWM604286S, CWM600110S, CWM600111S, CWM600116S, CWM600157S, CWM600158S, CWM600212S, CWM600873S, CWM600900S, CWM601656S, CWM602381S, CWM600278S, CWM600925S, CWM604305S, CWM600138S, CWM600141S, CWM600144S, CWM600255S, CWM600851S, CWM600860S, CWM600861S, CWM600880S, CWM600910S, CWM601539S, CWM602543S, CWM603176S, CWM603183S, CWM604262S, CWM600147S, CWM600160S, CWM602550S, CWM603404S, CWM600909S, CWM601305S, CWM601477S, CWM602390S, CWM602535S, CWM602553S, CWM602971S, CWM603181S, CWM600868S, CWM600869S, CWM600927S, CWM600929S, CWM600940S, CWM600941S, CWM601310S, CWM602382S, CWM603846S, CWM601283S, CWM602180S, CWM604260S, CWM600885S, CWM600931S, CWM600954S, CWM601282S, CWM601479S, CWM601480S, CWM601484S, CWM601553S, CWM601558S, CWM601562S, CWM601654S, CWM601666S, CWM601674S, CWM602174S, CWM602548S, CWM602969S, CWM603169S, CWM603852S, CWM601552S, CWM601693S, CWM601303S, CWM602380S, CWM604259S, CWM601695S, CWM602168S, CWM603409S, CWM602972S, CWM600914S, CWM601296S, CWM601549S, CWM601567S, CWM601568S, CWM601569S, CWM601649S, CWM602536S, CWM601292S, CWM602982S, CWM603655S, CWM603403S, CWM603406S, CWM601532S, CWM603173S, CWM604258S, CWM602976S, CWM601291S, CWM601555S, CWM602184S, CWM602546S, CWM602904S, CWM602966S, CWM604268S, CWM600908S, CWM601486S, CWM601545S, CWM603863S, CWM603864S, CWM604266S, CWM604267S, CWM604290S, CWM602913S, CWM603399S, CWM603845S; GTIN 00763000611910, Lot Serial Numbers: CWM604616S, CWM604596S, CWM604576S, CWM604588S, CWM604611S, CWM604545S, CWM604508S, CWM604627S, CWM604624S, CWM604518S, CWM604495S, CWM604513S, CWM604499S, CWM604546S, CWM604587S, CWM604502S, CWM604592S, CWM604494S, CWM604536S, CWM604537S, CWM604538S, CWM604539S, CWM604567S, CWM604572S, CWM604534S, CWM604574S, CWM604535S, CWM604540S, CWM604541S, CWM604583S, CWM604609S, CWM604602S, CWM604528S, CWM604599S, CWM604584S, CWM604585S, CWM604507S, CWM604548S, CWM604547S, CWM604511S, CWM604512S, CWM604542S, CWM604543S, CWM604497S, CWM604498S, CWM604500S, CWM604501S, CWM604503S, CWM604504S, CWM604505S, CWM604506S, CWM604509S, CWM604510S, CWM604514S, CWM604515S, CWM604517S, CWM604519S, CWM604524S, CWM604525S, CWM604529S, CWM604530S, CWM604531S, CWM604532S, CWM604533S, CWM604544S, CWM604549S, CWM604550S, CWM604551S, CWM604552S, CWM604553S, CWM604554S, CWM604555S, CWM604565S, CWM604566S, CWM604575S, CWM604581S, CWM604582S, CWM604586S, CWM604589S, CWM604590S, CWM604591S, CWM604593S, CWM604594S, CWM604595S, CWM604597S, CWM604603S, CWM604610S, CWM604612S, CWM604613S, CWM604617S, CWM604618S, CWM604619S, CWM604620S, CWM604625S, CWM604626S, CWM604628S, CWM604629S, CWM604630S, CWM604631S, CWM604632S, CWM604633S, CWM604635S, CWM604639S, CWM604640S, CWM604641S, CWM604655S, CWM604642S, CWM604643S, CWM604644S, CWM604645S, CWM604646S, CWM604653S, CWM604621S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information Contact Technical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
Action Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce 3801 units
Distribution US Nationwide - Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = LWS and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = LWS and Original Applicant = Medtronic, Inc.
-
-