|
Class 3 Device Recall Medtronic Restore Neurostimulator |
|
Date Initiated by Firm |
November 22, 2005 |
Date Posted |
May 25, 2006 |
Recall Status1 |
Terminated 3 on December 26, 2006 |
Recall Number |
Z-0905-06 |
Recall Event ID |
34225 |
PMA Number |
P840001S074 |
Product Classification |
Stimulator, Spinal-Cord, Implanted (Pain Relief) - Product Code GZB
|
Product |
Medtronic Model 37711 Restore Neurostimulator. The Restore Neurostimulator is an implanted, rechargeable neurostimulator which provides pain management therapy. |
Code Information |
NJH701951H thru NJH801963H, NJH701965H, NJH701967H thru NJH701969H, NJH701971H thru NJH701975H, NJH701978H thru NJH701986H, NJH702007H, NJH702008H, NJH702028H, NJH702029H, NJH702032H, NJH702033H, NJH702035H, NJH702036H, NJH702038H, NJH702039H, NJH702044H, NJH702045H, NJH702144H, NJH702145H, NJH702146H, NJH702150H, NJH702152H, NJH702153H, NJH702157H, NJH702158H, NJH702159H, NJH702160H, NJH702162H, NJH702163H, NJH702164H, NJH702166H, NJH702168H, NJH702169H, NJH702171H thru NJH702175H, NJH702183H, NJH702184H, NJH702213H, NJH702215H, NJH702216H, NJH702263H, NJH702324H, NJH702425H, NJH702427H, NJH702428H, NJH702429H, NJH702434H, NJH702436H, NJH702438H, NJH702439H, NJH702441H, NJH702443H, NJH702447H, NJH702451H, NJH702452H, NJH702455H, NJH702458H, NJH702487H, NJH702765H, NJH702772H, NJH702809H, NJH702834H, NJH702852, NJH702862H, NJH702863H, NJH702919H, NJH702980, NJH702982H, NJH702990H, NJH702993H, NJH702999H, NJH703001H |
Recalling Firm/ Manufacturer |
Medtronic Inc, Neurological & Spinal Division 800 53rd Ave Ne Columbia Heights MN 55421-1241
|
For Additional Information Contact |
763-514-4000
|
Manufacturer Reason for Recall |
A limited number of Model 37711 Restore Neurostimulators have an incorrect internal memory parameter that causes the device to indicate it is fully recharged when it may only be 80% to 90% recharged. This anomaly does not affect patient safety. This anomoly can be corrected by reprogramming the Restore Neurostimulator using specific software.
|
FDA Determined Cause 2 |
Other |
Action |
A 'Customer Communication' letter was sent to Medtronic field personnel to hand-carry into the affected accounts; customer contact began on 11/22/2005. The letter explains the patients implanted with these devices likely will only be able to recharge their neurostimulator to about 80-90% of a full charge. This anomaly does not affect device safety or effectiveness but Medtronic will re-program them upon request for customer satisfaction. The letter will be followed by scheduled on-site visit to reset the internal memory parameter of the neurostimulator. |
Quantity in Commerce |
101 devices affected |
Distribution |
Nationwide (AZ, CA, FL, ID, IL, IN, KS, KY, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, and WI). |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|