 Class 2 Device Recall Skytron Infinity Series surgical light see related information
Date Posted February 22, 2007
Recall Status1 Terminated on November 18, 2009
Recall Number Z-0529-2007
Recall Event ID 37120
510(K)Number K915285 
Product Classification surgical light - Product Code FTD
Product Skytron Infinity Series surgical light, 5 bulbs in 1 - 22' diameter lighthead, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Models IF22, IF22B, IF22LH, IF22LHH-B, IF22M-S, IN22, IN22EL, IN22LH and IF22(EL).
Code Information Each lighthead bears a unique serial number. For units with model numbers beginning with IF22, all serial numbers with the following numbers in the third and fourth position of the numerical series: 95, 96, 97, 98, 99 or 00 (examples: xx97xxxx or xx00xxxB). For units with model numbers beginning with IN22 , all serial numbers with the following numbers in the third and fourth positions of the numerical series: 92, 93 or 94 (example: xx93xxxx).
Recalling Firm/
Manufacturer
Skytron, Div. The KMW Group, Inc
5000 36th St Se
Grand Rapids MI 49512-2008
For Additional Information Contact Mary Preston-Murphy
800-759-8766 Ext. 3048
Manufacturer Reason
for Recall
If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
FDA Determined
Cause 2
Other
Action The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Quantity in Commerce 5,000 for all models.
Distribution Nationwide and Canada.