Date Initiated by Firm |
June 06, 2007 |
Date Posted |
July 03, 2007 |
Recall Status1 |
Terminated 3 on November 30, 2007 |
Recall Number |
Z-1001-2007 |
Recall Event ID |
38118 |
510(K)Number |
K062058
|
Product Classification |
Glucose Monitoring System - Product Code LFR
|
Product |
Ascensia Contour Blood Glucose Monitoring System ( Bayer )Diabetes Care System; Product 7152A.---NO USA Units are affected---. |
Code Information |
NO USA Units Affected: Serial numbers 1475139, 1688643, 1841620, 1852071, 2064303, 2064306, 2064329, 2064601, 2086449, 2203283, 2221308, 2263059, 2263260, 2280056, 2317037 and 2319010. |
Recalling Firm/ Manufacturer |
Bayer Healthcare LLC 430 S Beiger Street Mishawaka IN 46544-3207
|
For Additional Information Contact |
Roger Sonnenburg 574-256-3441
|
Manufacturer Reason for Recall |
Incorrect Unit of Measure displayed : The meters report results in the wrong unit of measurement for Canadian users.
|
FDA Determined Cause 2 |
Other |
Action |
The firm began calling Canadian customers on or about 6/06/07 to inform them of the problem, to tell them that a meter with the correct measurement units will be sent to them, and to request that they return the meter currently in their possession. |
Quantity in Commerce |
16-Canadian Units only are affected |
Distribution |
Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LFR and Original Applicant = BAYER HEALTHCARE
|