• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Signa HDX MR System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Signa HDX MR System
see related information
Date Posted September 11, 2008
Recall Status1 Open
Recall Number Z-2225-2008
Recall Event ID 48689
Premarket Notification
510(K) Number
K052293 
Product Classification System, Nuclear Magnetic Resonance Imaging - Product Code LNH
Product GE 1.5T and 3.0T Signa HDX MR System Model Number(s): 2226300, 2377062-61, 2395001, 2395001-2, 5107849, and 5127452. This system is indicated for use as a diagnostic imaging device to produce axial sagittula, coronal and oblique images, spectroscopic images, and or spectra, dynamic images of the internal structure or organs of the entire body,
Code Information 000000000R1597 000000000R5210 000000000R5871 00000007557YR2 00000007496YR3 00000010285YR5 00000008560YR5 00000009025YR8 0000000B523YR3 00000006697YR7 00000007562YR2 00000005539YR2 00000006662YR1 00000007568YR9 00000007108YR4 00000007497YR1 00000007542YR4 00000008116YR6 00000008322YR0 00000008384YR0 00000009538YR0 00000010300YR2 00000007114YR2 00000006604YR3 00000008325YR3 00000009872YR3 00000006557YR3 00000009534YR9 00000007563YR0 00000009854YR1 00000006722YR3 00000007541YR6 00000008376YR6 00000008317YR0 00000008099YR4 00000009006YR8 00000007126YR6 00000008351YR9 00000010283YR0 00000009009YR2 00000005981YR6 00000008100YR0 00000006710YR8 00000008098YR6 00000007890YR7 00000006594YR6 00000008539YR9 00000007871YR7 00000007564YR8 00000005992YR3 00000007556YR4 00000007498YR9 00000007878YR2 00000007565YR5 00000007877YR4 00000007100YR1 00000008993YR* 00000007097YR9 00000009436YR7 00000009419YR3 00000006698YR5 00000007530YR9 00000009858YR2 00000009082YR9 00000007551YR5 00000007870YR9 00000007508YR5 00000009539YR8 00000008404YR6 00000006713YR2 00000008339YR4 00000008396YR4 00000009773YR3 00000009576YR0 00000008352YR7 00000006707YR4 00000006678YR7 00000007582YR0 00000007882YR4 00000007887YR3 00000010511YR4 00000006558YR1 00000009566YR1 00000008197YR6 00000007087YR0 00000007859YR2 00000009060YR5 00000008387YR3 00000009461YR5 00000008342YR8 00000005564YRO 00000007133YR2 00000005537YR6 00000006737YR1 00000006757YR9 00000010160YR0 00000010396YR0 00000005564YR0 00000005532YR7 00000009052YR2 00000007532YR5 00000006690YR2 00000008355YR0 00000006691YR0 00000007099YR5 00000006574YR8 00000009545YR5 00000005970YR9 00000007858YR4 00000007856YR8 00000006682YR9 00000008338YR6 00000005927YR5 00000007549YR9 00000007122YR5 00000006696YR9 00000006715YR7 00000009979YR6 00000009553YR9 00000010525YR4 00000005222YR5 00000010132YR9 00000010432YR3 00000010284YR8 00000010048YR7 00000009618YR0 00000009421YR9 00000009848YR3 00000009578YR6 00000006738YR9 00000009084YR5 00000007866YR7 00000007865YR9 00000007514YR3 00000010045YR3 00000009587YR7 00000009986YR1 00000009766YR7 00000006721YR5 00000009531YR5 00000010028YR9 00000006600YR1 00000010395YR2 00000006727YR2 00000010134YR5 00000007544YR0 00000010289YR7 00000005548YR3 00000007547YR3 00000007555YR6 00000007853YR5 00000006572YR2 00000006670YR4 00000006673YR8 00000008332YR9 00000009867YR3 00000005974YR1 00000006669YR6 00000009982YR0 00000009774YR1 00000010163YR4 00000005215TR9 00000009049YR8 00000007117YR5 00000008378YR2 00000008373YR3 00000007594YR5 00000006716YR5 00000009617YR2 00000008391YR5 00000008059YR8 00000005216YR7 00000006566YR4 00000008337YR8 00000008392YR3 00000009007YR6 00000005536YR8 00000009775YR8 00000007572YR1 00000009599YR2 00000007862YR6 00000007507YR7 00000006570YR6 00000006708YR2 00000006714YR0 00000009845YR9 00000007591YR1 00000007561YR4 00000008335YR2 00000009438YR3 00000007579YR6 00000009547YR1 00000008327YR9 00000009028YR2 00000008061YR4 00000006667YR0 00000006556YR5 00000010059YR4 00000006660YR5 00000006568YR0 00000008359YR2 00000010153YR5 00000009571YR1 00000006743YR9 00000009990YR3 00000006733YR0 00000007884YR0 00000009455YR7 00000009070YR4 00000008380YR8 00000008385YR7 00000007116YR7 00000010144YR4 00000009019YR1 00000010063YR6 00000009062YR1 00000007109YR2 00000009593YR5 00000009564YR6 00000007867YR5 00000007539YR0 00000007520YR0 00000006705YR8 00000009031YR6 00000007090YR4 00000009081YR1 00000007864YR2 00000010302YR8 00000007863YR4 00000005990YR7 00000008372YR5 00000010393YR7 00000009973YR9 00000009627YR1 00000008530YR8 00000007554YR9 00000009847YR5 00000008318YR8 00000009057YR1 00000007869YR1 00000009026YR6 00000008367YR5 00000007892YR3 00000007540YR8 00000006665YR4 00000006562YR3 00000009053YR0 00000007115YR9 00000009865YR7 00000009642YR0 00000009846YR7 00000009459YR9 00000009573YR7 00000006577YR1 00000008364YR2 00000006661YR3 00000005525YR1 00000008394YR9 00000009080YR3 00000007516YR8 00000010140YR2 00000008319YR6 00000008528YR2 00000005559YR0 00000009565YR3 00000009442YR5 00000006726YR4 00000007577YR0 00000006573YR0 00000005220YR9 00000008354YR3 00000008347YR7 00000007588YR7 00000006555YR7 00000008377YR4 00000009581YR0 00000006583YR9 00000009018YR3 00000005543YR4 00000008103YR4 00000008537YR3 00000008534YR0 00000008314YR7 00000009030YR8 00000008561YR3 00000007575YR4 00000009047YR2 00000009083YR7 00000007845YR1 00000007534YR1 00000009454YR0 00000007107YR6 00000009572YR9 00000008123YR2 00000008542YR3 00000008323YR8 00000007505YR1 00000007518YR4 00000009037YR3 00000010168YR3 00000007531YR7 00000009837YR6 00000006595YR3 00000007101YR9 00000007105YR0 00000008991YR2 00000008311YR3 00000008549YR8 00000008532YR4 00000007134YR0 00000009450YR8 00000008358YR4 00000007570YR5 00000010060YR2 00000008366YR7 00000007566YR3 00000010035YR4 00000005982YR4 00000007098YR7 00000007528YR3 00000009001YR9 00000009763YR4 00000008362YR6 00000006694YR4 00000008097YR8 00000005555YR8 00000008402YR0 00000010627YR8 00000010157YR6 00000009984YR6 00000009864YR0 00000009619YR8 00000006725YR6 00000007088YR8 00000007112YR6 00000007111YR8 00000007123YR3 00000008122YR4 00000006672YR0 00000008350YR1 00000008527YR4 00000009032YR4 00000008992YR0 00000008371YR7 00000009027YR4 00000009579YR4 00000010037YR0 00000009561YR2 00000007504YR4 00000010031YR3 00000010487YR7 00000008557YR1 00000006706YR6 00000007092YR0 00000007574YR7 00000008195YR0 00000007503YR6 00000008986YR2 00000006754YR6 00000008990YR4 00000277444MR6 00000279066MR5 00000278393MR4 00000277641MR7 00000280066MR2 00000274886MR1 00000280365MR8 00000276657MR4 00000279414MR7 00000273383MRO 00000281158MR6 00000280778MR2 00000282865MR5 00000274740MR0 00000277535MR1 00000279464MR2 00000279447MR7 00000276184MR9 00000281828MR4 00000278605MR1 00000278800MR8 00000278554MR1 00000280921MR8 00000283517MR1 00000280253MR6 00000280251MR0 00000276680MR6 00000278041MR9 00000280252MR8 00000276457MR9 00000285493MR3 00000280065MR4 00000276656MR6 00000277917MR1 00000277534MR4 00000279400MR6 00000280364MR1 00000273632MR0 00000279910MR4 00000274742MR6 00000000000001 00000279067MR3 00000276245MR8 00000284034MR6 00000278391MR8 00000279166MR3 00000280775MR8 00000000000000 00000281825MR0 00000283346MR5 00000280249MR4 00000278606MR9 00000275537MR9 00000277779MR5 00000279909MR6 00000286339MR7 00000279882MR5 00000278164MR9 00000285490MR9 00000027810MR1 00000281863MR1 00000278799MR2 00000283458MR8 00000284493MR4 00000274887MR9 00000278103MR7 00000280857MR4 00000282383MR9 00000279908MR8 00000282508MR1 00000275526MR2 00000285589MR8 00000279461MR8 00000276163MR3 00000280777MR4 00000281467MR1 00000278223MR3 00000280328MR6 00000273631MR2 00000279563MR1 00000284872MR9 00000280325MR2 00000279616MR7 00000284562MR6 00000285496MR6 00000283516MR3 00000279398MR2 00000280068MR8 00000276243MR3 00000285167MR3 00000272922MR6 00000285437MR0 00000285166MR5 00000276244MR1 00000278039MR3 00000278801MR6 00000278556MR6 00000281375MR6 00000278315MR7 00000280366MR6 00000283773MR0 00000278221MR7 00000284650MR9 00000278674MR7 00000282003MR3 00000278676MR2 00000285674MR8 00000284869MR5 00000276265MR6 00000280673MR5 00000275647MR6 00000278851MR1 00000284234MR2 00000280000MR1 00000279068MR1 00000276266MR4 00000275967MR8 00000275648MR4 00000282549MR5 00000280781MR6 00000283745MR8 00000274744MR2 00000275046MR1 00000275968MR6 00000278397MR5 00000281341MR8 00000276434MR8 00000284261MR5 00000280780MR8 00000285053MR5 00000275465MR3 00000268369MR6 00000260651MR5 000000HDMR2329 000000HDMR2927 00000HDMR22498 00000HDMR10281 00000HDMR10273 00000HDMR20270 00000HDMR10462 000000HDMR2471 000000HDMR3334 000000HDMR2613 000000HDMR3288 000000HDMR2438 00000HDMR20181 000000HDMR2441 00000HDMR20228 000000HDMR2872 00000HDMR20034 00000276073MR4 00000001568422 000000HDMR2901 000000HDMR2799 00000HDMR20056 000000HDMR3407 00000DDDM10670 00000HDMR20044 00000HDMR20171 000000HDMR2444 00000HDMR10788 00000HDMR10322 000000HDMR2936 00000HDMR20294 000000HDMR2531 00000HDMR20219 00000HDMR20235 000000HDMR2829 000000HDMR2893 00000HDMR20018 000000HDMR2528 00000HDMR20061 00000HDMR10743 00000HDMR11028 000000HDMR2686 00000HDMR20206 000000HDMR2508 00000HDMR20142 00000HDMR20050 00000HDMR20151 000000HDMR2515 00000HDMR10404 000000HDMR2641 000000HDMR2888 00000HDMR20158 000000HDMR2998 00000275160MR0 000000HDMR3089 00000HDMR20276 00000HDMR20065 00000HDMR20053 000000HDMR2451 000000HDMR2463 000000HDMR2553 000000HDMR2689 000000HDMR2335 00000001516299 000000HDMR2482 00000275157MR6 00000HDMR20070 00000273807MR8 000000HDMR3373 000000HDMR1111 000000HDMR3298 00000HDMR10291 000000HDMR1005 00000HDMR10846 000000HDMR3048 000000HDMR3262 000000HDMR2591 000000HDMR2651 000000HDMR2675 00000HDMR10831 00000HDMR10334 00000HDMR10548 00000275162MR6 0000HDMR127452 000000HDMR2522 000000HDMR2344 000000HDMR2803 000000HDMR2626 00000HDMR20236 00000HDMR10739 000000HDMR1130 000000HDMR2993 00000273127MR1 00000HDMR20128 000000HDMR3088 00000HDMR20043 00000HDMR10368 00000274350MR8 00000275283MR0 00000HDMR10489 00000HDMR20084 00000HDMR10986 00000HDMR20059 000000HDMR1098 00000HDMR20068 000000HDMR3221 000000HDMR2404 00000HDMR10440 000000HDMR2560 00000HDMR10412 000000HDMR2822 00000275173MR3 000000HDMR3355 000000HDMR2448 00000275279MR8 000000HDMR2612 000000HDMR2660 00000HDMR10938 00000274359MR9 00000HDMR10764 000000HDMR2397 000000HDMR2523 00000HDMR20073 00000HDMR20289 000000HDMR2428 00000HDMR20248 000000HDMR2623 000000HDMR2756 00000276057MR7 000000HDMR2517 000000HDMR3109 000000HDMR2804 00000HDMR10351 000000HDMR3272 00000HDMR10719 00000HDMR20274 000000HDMR3299 00000HDMR20172 00000HDMR20293 000000HDMR3263 00000HDMR20047 000000HDMR2592 00000HDMR10471 00000HDMR20036 000000HDMR3050 000000HDMR2981 000000HDMR2758 000000HDMR2789 000000HDMR2524 000000HDMR2890 000000HDMR2773 000000HDMR3337 000000HDMR2873 000000HDMR2729 00000005127452 000000HDMR2904 00000HDMR10356 000030485C4228 000000HDMR2948 00000HDMR20291 000000HDMR3316 00000HDMR20269 00000HDMR10711 00000HDMR20189 000000HDMR2418 000000HDMR2755 000000HDMR3128 000000HDMR2831 00000HDMR20272 000000HDMR3199 000000HDMR3130 00000HDMR10477 000000HDMR2958 000000HDMR3459 00000HDMR20178 00000HDMR10878 00000HDMR20089 00000HDMR10787 00000HDMR20099 000000HDMR2769 000000HDMR2917 00000HDMR20290 00000HDMR19996 000000HDMR2395 000000HDMR3003 000000HDMR2483 000000HDMR2696 00000274392MR0 00000HDMR10343 000000HDMR2747 00000HDMR20246 00000HDMR10417 00000273839MR1 00000275236MR8 000000HDMR3301 000000HDMR3542 00000276031MR2 000000HDMR2313 00000273119MR8 000000HDMR2882 000000HDMR2902 00000275237MR6 00000274385MR4 00000HDMR20012 00000273832MR6 00000276038MR7 00000274383MR9 00000275168MR3 00000276042MR9 00000276102MR1 000000HDMR2572 000000HDMR2600 000000HDMR3028 000000HDMR3077 000000HDMR2835 000000HDMR2544 00000275151MR9 00000273132MR1 000000HDMR3452 00000275113MR9 00000275120MR4 00000275107MR1 00000HDMR10623 000000HDMR2521 000000HDMR2808 000000HDMR2779 000000HDMR2695
Recalling Firm/
Manufacturer
GE Healthcare
3000 N Grandview Blvd
Waukesha, Wisconsin 53188-1615
Manufacturer Reason
for Recall
Possible misdiagnosis or inappropriate treatment based on the image. Signal homogeneity may be affected when BRAVO applications are set with IR Prep off. The concern is that signal inhomogeneity can result in an erroneous diagnosis.
FDA Determined
Cause 2
CHANGE CONTROL (GMP - GOOD MANUFACTURING PRACTICE): Component Change Control
Action Customer Letter and Operator Manual addendum were sent directly to customers in August 2008. When the software correction is issued, GE Field Engineer will visit each site to implement the correction. (Not conducted yet).
Quantity in Commerce 720 units (360 US / 360 OUS) estimate
Distribution AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS,MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA,WI, and WV. Countries of Argentina Australia Austria Belgium Brazil Canada Chile China Czech Republic Denmark Finland France French Polynesia Germany Greece Hong Kong India Indonesia Ireland Israel Italy Japan Korea Kuwait Lebanon Malaysia Mexico Netherlands New Zealand Norway Peru Philippines Russia Saudi Arabia Singapore South Africa Spain Sweden Switzerland Taiwan Taiwan Thailand Turkey United Arab Emirates and United Kingdom.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS, LLC
-
-