|
Class 2 Device Recall NeuViz Dual series Computed Tomography (CT) Scanner System |
|
Date Initiated by Firm |
July 21, 2008 |
Date Posted |
August 19, 2010 |
Recall Status1 |
Terminated 3 on August 23, 2010 |
Recall Number |
Z-2237-2010 |
Recall Event ID |
49012 |
510(K)Number |
K062451 K071308
|
Product Classification |
Computed Tomography X-Ray System - Product Code JAK
|
Product |
NeuViz Dual series Computed Tomography (CT) Scanner System, X-Ray System. Part Number: 989605651321.
Intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles. |
Code Information |
Serial Number: 400562, 400563, 400626, NDH009EI, NDH011El, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR08002, and NDHRO80003. |
Recalling Firm/ Manufacturer |
Neuisys, LLC 1500 Pinecroft Rd Ste 212 Greensboro NC 27407-3808
|
For Additional Information Contact |
Kevin Zhang 877-299-9052
|
Manufacturer Reason for Recall |
The reference lines for image generated for a Surview scan may appear in the incorrect position in the "FILM" display mode.
|
FDA Determined Cause 2 |
Software design |
Action |
On 07/21/2008, NEUISYS Imaging Systems Solutions distributor in Greensboro, NC began sending the URGENT DEVICE CORRECTION letter on behalf of Philips and NeuSoft Medical Systems (PNMS), Co., Ltd in China.
The letter informs the consignees that there is potential safety problem in the NeuViz Dual series CT scanners due to the malfunction of "Incorrect Reference Line in the Film". When a "Head first, Couch out" or a "Feet first, Couch in" Surview scan is started then paused midstream, the image generated is what the doctor expects, but the reference lines appear in the incorrect position in "FILM" display mode. The acquired images accurately represent the exact positions planned on the original view. The incorrect/unmatched reference line in film may result in incorrect diagnosis.
The consignees are advised not to stop or pause during Surview acquisition while PNMS in China will provide the updated software patch to all affected systems free of charge.
The Field Change Order #FCOP-08-CT-NMS-008 SP4 will be implemented by NEUISYS Imaging Systems Solutions in Greensboro, NC., who is a distributor for PNMS.
Consignees are advised to e-mail Service Support Department of PNMS at helpdesk@pnms.neusoft.com regarding the recall. |
Quantity in Commerce |
19 units in the US |
Distribution |
Nationwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = NEUSOFT MEDICAL SYSTEMS CO., LTD.
|
|
|
|