|
Class 2 Device Recall Lamp, surgical |
|
Date Initiated by Firm |
February 23, 2009 |
Date Posted |
April 12, 2009 |
Recall Status1 |
Terminated 3 on September 29, 2009 |
Recall Number |
Z-1154-2009 |
Recall Event ID |
51297 |
510(K)Number |
K882613
|
Product Classification |
ALM SURGICAL LAMPS - Product Code FTD
|
Product |
Alm Prismatic Surgical Light 5004 SF; Maquet SA.
ALM PRISMATIC (PRC) Surgical Lights are used to provide visible illumination for the surgical field or for the examination of the patient. PRC is designed to eliminate shadow, reduce reflected heat, and illuminate surgical incisions of all depths during surgical procedures. |
Code Information |
Article Number: 567030999; Serial Numbers: 6041, 6042, 6043, 6044, 6045, 6046, 6047, 6048, 6049, 6050, 6051, 6052, 6053, 6054, 6057, 6058, 6059, 6060, 6061 and 6062. |
Recalling Firm/ Manufacturer |
MAQUET Inc. 1140 US Highway 22 East Suite 202 Bridgewater NJ 08807-2958
|
For Additional Information Contact |
Jamie Yieh 732-667-1896
|
Manufacturer Reason for Recall |
The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.
|
FDA Determined Cause 2 |
Device Design |
Action |
An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative. |
Quantity in Commerce |
54 of all units, both article numbers |
Distribution |
Nationwide Distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FTD and Original Applicant = ALM SURGICAL EQUIPMENT, INC.
|
|
|
|