| Class 2 Device Recall Medtronic Virtuoso DR | |
Date Initiated by Firm | September 09, 2009 |
Date Posted | November 09, 2009 |
Recall Status1 |
Terminated 3 on May 29, 2012 |
Recall Number | Z-0128-2010 |
Recall Event ID |
53294 |
PMA Number | P980016S062 |
Product Classification |
dual chamber ICDS - Product Code LWS
|
Product | Medtronic Virtuoso DR, D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and ventricular therapies. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. This device is indicated to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias in patients with NYHA functional Class II/III heart failure. In addition, the device is indicated for use in the above patients with atrial tachyarrhythmias, or those patients who are at significant risk of developing atrial tachyarrhythmias. |
Code Information |
serial numbers: PUL400006H, PUL400007H, PUL400009H, PUL400010H, PUL400018H, PUL400022H, PUL400023H, PUL400024H, PUL400030H, PUL400031H, PUL400033H, PUL400034H, PUL400035H, PUL400043H, PUL400046H, PUL400048H, PUL400049H, PUL400051H, PUL400055H, PUL400056H, PUL400058H, PUL400059H, PUL400060H, PUL400062H, PUL400064H, PUL400066H, PUL400069H, PUL400070H, PUL400072H, PUL400074H, PUL400078H, PUL400080H, PUL400081H, PUL400082H, PUL400085H, PUL400086H, PUL400087H, PUL400090H, PUL400091H, PUL400093H, PUL400098H, PUL400102H, PUL400104H, PUL400106H, PUL400108H, PUL400110H, PUL400115H, PUL400118H, PUL400119H, PUL400124H, PUL400132H, PUL400134H, PUL400136H, PUL400141H, PUL400159H, PUL400161H, PUL400162H, PUL400163H, PUL400164H, PUL400170H, PUL400185H, PUL400189H, PUL400191H, PUL400196H, PUL400198H, PUL400203H, PUL400209H, PUL400215H, PUL400217H, PUL400218H, PUL400224H, PUL400225H, PUL400226H, PUL400227H, PUL400230H, PUL400232H, PUL400243H, PUL400245H, PUL400249H, PUL400253H, PUL400256H, PUL400260H, PUL400261H, PUL400263H, PUL400266H, PUL400270H, PUL400271H, PUL400273H, PUL400280H, PUL400281H, PUL400283H, PUL400284H, PUL400290H, PUL400301H, PUL400302H, PUL400303H, PUL400306H, PUL400310H, PUL400312H, PUL400317H, PUL400318H, PUL400319H, PUL400335H, PUL400339H, PUL400344H, PUL400347H, PUL400350H, PUL400351H, PUL400352H, PUL400357H, PUL400358H, PUL400360H, PUL400375H, PUL400378H, PUL400381H, PUL400385H, PUL400388H, PUL400401H, PUL400405H, PUL400406H, PUL400411H, PUL400412H, PUL400413H, PUL400415H, PUL400430H, PUL400434H, PUL400440H, PUL400444H, PUL400449H, PUL400451H, PUL400458H, PUL400460H, PUL400464H, PUL400479H, PUL400480H, PUL400481H, PUL400484H, PUL400488H, PUL400494H, PUL400496H, PUL400498H, PUL400499H, PUL400503H, PUL400504H, PUL400511H, PUL400522H, PUL400527H, PUL400529H, PUL400530H, PUL400533H, PUL400543H, PUL400544H, PUL400550H, PUL400555H, PUL400563H, PUL400566H, PUL400567H, PUL400569H, PUL400573H, PUL400576H, PUL400585H, PUL400600H, PUL400601H, PUL400602H, PUL400608H, PUL400610H, PUL400611H, PUL400613H, PUL400624H, PUL400628H, PUL400632H, PUL400635H, PUL400636H, PUL400637H, PUL400648H, PUL400657H, PUL400662H, PUL400665H, PUL400667H, PUL400668H, PUL400674H, PUL400675H, PUL400677H, PUL400684H, PUL400694H, PUL400696H, PUL400701H, PUL400709H, PUL400728H, PUL400729H, PUL400730H, PUL400740H, PUL400748H, PUL400751H, PUL400768H, PUL400775H, PUL400779H, PUL400788H, PUL400789H, PUL400791H, PUL400794H, PUL400804H, PUL400805H, PUL400807H, PUL400825H, PUL400833H, PUL400835H, PUL400841H, PUL400842H, PUL400846H, PUL400856H, PUL400884H, PUL400885H, PUL400891H, PUL400893H, PUL400895H, PUL400896H, PUL400900H, PUL400901H, PUL400907H, PUL400911H, PUL400915H, PUL400917H, PUL400919H, PUL400920H, PUL400928H, PUL400930H, PUL400935H, PUL400939H, PUL400943H, PUL400948H, PUL400953H, PUL400955H, PUL400956H, PUL400957H, PUL400966H, PUL400968H, PUL400972H, PUL400974H, PUL400975H, PUL400978H, PUL400979H, PUL400986H, PUL400998H, PUL400999H, PUL401002H, PUL401009H, PUL401013H, PUL401018H, PUL401021H, PUL401023H, PUL401026H, PUL401034H, PUL401035H, PUL401037H, PUL401043H, PUL401046H, PUL401047H, PUL401048H, PUL401049H, PUL401050H, PUL401053H, PUL401054H, PUL401055H, PUL401056H, PUL401057H, PUL401060H, PUL401066H, PUL401068H, PUL401069H, PUL401072H, PUL401074H, PUL401075H, PUL401076H, PUL401079H, PUL401083H, PUL401085H, PUL401088H, PUL401090H, PUL401095H, PUL401096H, PUL401097H, PUL401099H, PUL401102H, PUL401107H, PUL401114H, PUL401116H, PUL401117H, PUL401118H, PUL401120H, PUL401121H, PUL401123H, PUL401124H, PUL401125H, PUL401127H, PUL401128H, PUL401130H, PUL401131H, PUL401133H, PUL401134H, PUL401135H, PUL401137H, PUL401140H, PUL401146H, PUL401147H, PUL401148H, PUL401149H, PUL401152H, PUL401153H, PUL401154H, PUL401155H, PUL401156H, PUL401157H, PUL401158H, PUL401160H, PUL401162H, PUL401163H, PUL401165H, PUL401166H, PUL401168H, PUL401169H, PUL401170H, PUL401172H, PUL401173H, PUL401174H, PUL401180H, PUL401182H, PUL401183H, PUL401185H, PUL401186H, PUL401189H, PUL401191H, PUL401193H, PUL401195H, PUL401196H, PUL401197H, PUL401202H, PUL401203H, PUL401204H, PUL401206H, PUL401207H, PUL401212H, PUL401216H, PUL401217H, PUL401218H, PUL401219H, PUL401221H, PUL401222H, PUL401223H, PUL401227H, PUL401228H, PUL401229H, PUL401232H, PUL401234H, PUL401235H, PUL401237H, PUL401238H, PUL401239H, PUL401240H, PUL401241H, PUL401243H, PUL401245H, PUL401246H, PUL401250H, PUL401253H, PUL401254H, PUL401256H, PUL401257H, PUL401258H, PUL401260H, PUL401261H, PUL401262H, PUL401263H, PUL401264H, PUL401266H, PUL401267H, PUL401268H, PUL401270H, PUL401272H, PUL401273H, PUL401274H, PUL401275H, PUL401276H, PUL401277H, PUL401278H, PUL401285H, PUL401286H, PUL401288H, PUL401290H, PUL401293H, PUL401294H, PUL401296H, PUL401297H, PUL401298H, PUL401301H, PUL401305H, PUL401307H, PUL401308H, PUL401310H, PUL401311H, PUL401313H, PUL401315H, PUL401319H, PUL401322H, PUL401323H, PUL401326H, PUL401327H, PUL401328H, PUL401329H, PUL401330H, PUL401331H, PUL401336H, PUL401338H, PUL401339H, PUL401341H, PUL401342H, PUL401343H, PUL401344H, PUL401353H, PUL401354H, PUL401355H, PUL401359H, PUL401360H, PUL401361H, PUL401366H, PUL401367H, PUL401368H, PUL401369H, PUL401370H, PUL401371H, PUL401373H, PUL401375H, PUL401376H, PUL401378H, PUL401380H, PUL401381H, PUL401382H, PUL401385H, PUL401388H, PUL401390H, PUL401392H, PUL401396H, PUL401397H, PUL401400H, PUL401401H, PUL401402H, PUL401403H, PUL401404H, PUL401405H, PUL401406H, PUL401409H, PUL401411H, PUL401413H, PUL401414H, PUL401415H, PUL401416H, PUL401418H, PUL401419H, PUL401420H, PUL401421H, PUL401422H, PUL401426H, PUL401427H, PUL401428H, PUL401429H, PUL401430H, PUL401431H, PUL401432H, PUL401435H, PUL401436H, PUL401438H, PUL401439H, PUL401440H, PUL401443H, PUL401444H, PUL401446H, PUL401448H, PUL401452H, PUL401454H, PUL401455H, PUL401456H, PUL401457H, PUL401459H, PUL401460H, PUL401462H, PUL401463H, PUL401464H, PUL401465H, PUL401467H, PUL401469H, PUL401470H, PUL401472H, PUL401476H, PUL401477H, PUL401478H, PUL401479H, PUL401480H, PUL401483H, PUL401484H, PUL401485H, PUL401486H, PUL401487H, PUL401488H, PUL401489H, PUL401491H, PUL401494H, PUL401495H, PUL401497H, PUL401498H, PUL401500H, PUL401501H, PUL401502H, PUL401504H, PUL401505H, PUL401506H, PUL401507H, PUL401509H, PUL401510H, PUL401512H, PUL401513H, PUL401515H, PUL401516H, PUL401517H, PUL401519H, PUL401521H, PUL401522H, PUL401524H, PUL401525H, PUL401527H, PUL401528H, PUL401530H, PUL401532H, PUL401533H, PUL401534H, PUL401537H, PUL401539H, PUL401541H, PUL401542H, PUL401544H, PUL401547H, PUL401548H, PUL401549H, PUL401550H, PUL401551H, PUL401554H, PUL401557H, PUL401560H, PUL401561H, PUL401564H, PUL401565H, PUL401568H, PUL401569H, PUL401572H, PUL401575H, PUL401576H, PUL401577H, PUL401578H, PUL401581H, PUL401583H, PUL401585H, PUL401586H, PUL401589H, PUL401591H, PUL401592H, PUL401594H, PUL401595H, PUL401596H, PUL401597H, PUL401598H, PUL401600H, PUL401601H, PUL401603H, PUL401604H, PUL401605H, PUL401606H, PUL401607H, PUL401609H, PUL401610H, PUL401612H, PUL401613H, PUL401615H, PUL401616H, PUL401620H, PUL401621H, PUL401624H, PUL401628H, PUL401629H, PUL401630H, PUL401631H, PUL401633H, PUL401634H, PUL401636H, PUL401639H, PUL401640H, PUL401647H, PUL401650H, PUL401651H, PUL401652H, PUL401654H, PUL401657H, PUL401658H, PUL401660H, PUL401661H, PUL401663H, PUL401664H, PUL401666H, PUL401667H, PUL401668H, PUL401670H, PUL401671H, PUL401675H, PUL401676H, PUL401679H, PUL401680H, PUL401681H, PUL401682H, PUL401683H, PUL401687H, PUL401688H, PUL401690H, PUL401692H, PUL401694H, PUL401696H, PUL401700H, PUL401705H, PUL401708H, PUL401709H, PUL401712H, PUL401713H, PUL401714H, PUL401716H, PUL401717H, PUL401718H, PUL401724H, PUL401726H, PUL401727H, PUL401728H, PUL401731H, PUL401732H, PUL401734H, PUL401735H, PUL401737H, PUL401738H, PUL401741H, PUL401742H, PUL401743H, PUL401745H, PUL401746H, PUL401749H, PUL401751H, PUL401753H, PUL401754H, PUL401756H, PUL401757H, PUL401758H, PUL401759H, PUL401760H, PUL401761H, PUL401762H, PUL401763H, PUL401767H, PUL401768H, PUL401769H, PUL401771H, PUL401772H, PUL401773H, PUL401775H, PUL401777H, PUL401779H, PUL401780H, PUL401781H, PUL401782H, PUL401783H, PUL401784H, PUL401786H, PUL401787H, PUL401788H, PUL401789H, PUL401790H, PUL401792H, PUL401793H, PUL401795H, PUL401796H, PUL401797H, PUL401798H, PUL401800H, PUL401802H, PUL401804H, PUL401805H, PUL401807H, PUL401809H, PUL401810H, PUL401813H, PUL401821H, PUL401822H, PUL401825H, PUL401826H, PUL401829H, PUL401830H, PUL401832H, PUL401835H, PUL401836H, PUL401839H, PUL401841H, PUL401842H, PUL401843H, PUL401845H, PUL401847H, PUL401848H, PUL401850H, PUL401852H, PUL401853H, PUL401854H, PUL401855H, PUL401856H, PUL401857H, PUL401858H, PUL401859H, PUL401860H, PUL401861H, PUL401862H, PUL401864H, PUL401865H, PUL401866H, PUL401867H, PUL401870H, PUL401871H, PUL401872H, PUL401873H, PUL401875H, PUL401876H, PUL401877H, PUL401880H, PUL401882H, PUL401883H, PUL401884H, PUL401886H, PUL401887H, PUL401888H, PUL401891H, PUL401893H, PUL401896H, PUL401897H, PUL401899H, PUL401903H, PUL401904H, PUL401905H, PUL401908H, PUL401911H, PUL401913H, PUL401916H, PUL401917H, PUL401919H, PUL401921H, PUL401922H, PUL401923H, PUL401924H, PUL401925H, PUL401926H, PUL401927H, PUL401928H, PUL401929H, PUL401930H, PUL401932H, PUL401934H, PUL401935H, PUL401937H, PUL401938H, PUL401939H, PUL401940H, PUL401941H, PUL401945H, PUL401946H, PUL401947H, PUL401949H, PUL401950H, PUL401951H, PUL401953H, PUL401954H, PUL401955H, PUL401956H, PUL401957H, PUL401963H, PUL401966H, PUL401968H, PUL401969H, PUL401970H, PUL401971H, PUL401972H, PUL401973H, PUL401974H, PUL401976H, PUL401978H, PUL401979H, PUL401980H, PUL401981H, PUL401982H, PUL401985H, PUL401986H, PUL401987H, PUL401989H, PUL401991H, PUL401992H, PUL401994H, PUL401995H, PUL401996H, PUL401997H, PUL402000H, PUL402002H, PUL402004H, PUL402006H, PUL402009H, PUL402010H, PUL402011H, PUL402012H, PUL402013H, PUL402014H, PUL402016H, PUL402019H, PUL402020H, PUL402023H, PUL402024H, PUL402027H, PUL402028H, PUL402031H, PUL402032H, PUL402033H, PUL402034H, PUL402036H, PUL402038H, PUL402041H, PUL402043H, PUL402045H, PUL402046H, PUL402049H, PUL402051H, PUL402053H, PUL402054H, PUL402055H, PUL402058H, PUL402059H, PUL402060H, PUL402061H, PUL402062H, PUL402063H, PUL402064H, PUL402066H, PUL402068H, PUL402070H, PUL402073H, PUL402074H, PUL402076H, PUL402077H, PUL402079H, PUL402080H, PUL402082H, PUL402083H, PUL402084H, PUL402086H, PUL402087H, PUL402088H, PUL402089H, PUL402091H, PUL402093H, PUL402094H, PUL402095H, PUL402096H, PUL402097H, PUL402099H, PUL402102H, PUL402103H, PUL402104H, PUL402105H, PUL402107H, PUL402108H, PUL402109H, PUL402114H, PUL402116H, PUL402118H, PUL402119H, PUL402123H, PUL402124H, PUL402125H, PUL402127H, PUL402128H, PUL402129H, PUL402132H, PUL402133H, PUL402134H, PUL402136H, PUL402138H, PUL402140H, PUL402141H, PUL402143H, PUL402144H, PUL402145H, PUL402149H, PUL402150H, PUL402151H, PUL402152H, PUL402155H, PUL402156H, PUL402158H, PUL402159H, PUL402160H, PUL402161H, PUL402162H, PUL402163H, PUL402164H, PUL402165H, PUL402166H, PUL402167H, PUL402168H, PUL402169H, PUL402172H, PUL402173H, PUL402174H, PUL402177H, PUL402178H, PUL402179H, PUL402180H, PUL402183H, PUL402184H, PUL402186H, PUL402187H, PUL402188H, PUL402189H, PUL402190H, PUL402191H, PUL402194H, PUL402197H, PUL402201H, PUL402202H, PUL402203H, PUL402204H, PUL402205H, PUL402207H, PUL402209H, PUL402210H, PUL402211H, PUL402212H, PUL402213H, PUL402215H, PUL402218H, PUL402220H, PUL402222H, PUL402223H, PUL402225H, PUL402226H, PUL402229H, PUL402230H, PUL402231H, PUL402233H, PUL402234H, PUL402235H, PUL402237H, PUL402239H, PUL402241H, PUL402242H, PUL402245H, PUL402247H, PUL402248H, PUL402250H, PUL402251H, PUL402252H, PUL402253H, PUL402257H, PUL402260H, PUL402261H, PUL402262H, PUL402265H, PUL402266H, PUL402267H, PUL402268H, PUL402270H, PUL402271H, PUL402272H, PUL402275H, PUL402276H, PUL402278H, PUL402279H, PUL402280H, PUL402282H, PUL402285H, PUL402288H, PUL402291H, PUL402292H, PUL402294H, PUL402299H, PUL402300H, PUL402301H, PUL402303H, PUL402304H, PUL402306H, PUL402307H, PUL402308H, PUL402309H, PUL402310H, PUL402313H, PUL402314H, PUL402316H, PUL402318H, PUL402319H, PUL402320H, PUL402321H, PUL402323H, PUL402325H, PUL402326H, PUL402327H, PUL402328H, PUL402332H, PUL402335H, PUL402336H, PUL402338H, PUL402340H, PUL402341H, PUL402343H, PUL402345H, PUL402347H, PUL402348H, PUL402349H, PUL402351H, PUL402355H, PUL402357H, PUL402359H, PUL402360H, PUL402361H, PUL402363H, PUL402365H, PUL402373H, PUL402376H, PUL402379H, PUL402383H, PUL402385H, PUL402389H, PUL402391H, PUL402392H, PUL402396H, PUL402397H, PUL402400H, PUL402401H, PUL402402H, PUL402405H, PUL402407H, PUL402411H, PUL402412H, PUL402413H, PUL402414H, PUL402415H, PUL402417H, PUL402419H, PUL402425H, PUL402427H, PUL402428H, PUL402431H, PUL402432H, PUL402433H, PUL402434H, PUL402437H, PUL402438H, PUL402439H, PUL402441H, PUL402443H, PUL402444H, PUL402445H, PUL402448H, PUL402450H, PUL402451H, PUL402452H, PUL402454H, PUL402455H, PUL402457H, PUL402458H, PUL402460H, PUL402461H, PUL402462H, PUL402464H, PUL402467H, PUL402471H, PUL402473H, PUL402475H, PUL402480H, PUL402482H, PUL402483H, PUL402487H, PUL402488H, PUL402490H, PUL402492H, PUL402493H, PUL402495H, PUL402497H, PUL402498H, PUL402500H, PUL402502H, PUL402507H, PUL402509H, PUL402510H, PUL402513H, PUL402516H, PUL402517H, PUL402519H, PUL402520H, PUL402524H, PUL402525H, PUL402526H, PUL402529H, PUL402530H, PUL402531H, PUL402533H, PUL402535H, PUL402538H, PUL402540H, PUL402541H, PUL402544H, PUL402554H, PUL402555H, PUL402556H, PUL402557H, PUL402558H, PUL402561H, PUL402562H, PUL402565H, PUL402566H, PUL402567H, PUL402568H, PUL402571H, PUL402572H, PUL402573H, PUL402578H, PUL402579H, PUL402581H, PUL402582H, PUL402589H, PUL402592H, PUL402593H, PUL402595H, PUL402597H, PUL402599H, PUL402601H, PUL402604H, PUL402606H, PUL402608H, PUL402611H, PUL402613H, PUL402614H, PUL402615H, PUL402621H, PUL402622H, PUL402624H, PUL402630H, PUL402635H, PUL402636H, PUL402637H, PUL402638H, PUL402649H, PUL402650H, PUL402654H, PUL402655H, PUL402657H, PUL402659H, PUL402661H, PUL402662H, PUL402663H, PUL402665H, PUL402667H, PUL402671H, PUL402673H, PUL402674H, PUL402675H, PUL402676H, PUL402680H, PUL402681H, PUL402683H, PUL402684H, PUL402685H, PUL402689H, PUL402690H, PUL402692H, PUL402693H, PUL402694H, PUL402697H, PUL402698H, PUL402699H, PUL402700H, PUL402701H, PUL402702H, PUL402703H, PUL402704H, PUL402706H, PUL402707H, PUL402710H, PUL402712H, PUL402713H, PUL402714H, PUL402716H, PUL402717H, PUL402718H, PUL402719H, PUL402720H, PUL402723H, PUL402724H, PUL402726H, PUL402728H, PUL402731H, PUL402732H, PUL402734H, PUL402736H, PUL402737H, PUL402738H, PUL402739H, PUL402740H, PUL402741H, PUL402742H, PUL402743H, PUL402745H, PUL402746H, PUL402747H, PUL402748H, PUL402749H, PUL402753H, PUL402755H, PUL402756H, PUL402757H, PUL402758H, PUL402760H, PUL402763H, PUL402767H, PUL402769H, PUL402772H, PUL402776H, PUL402777H, PUL402779H, PUL402782H, PUL402785H, PUL402786H, PUL402787H, PUL402788H, PUL402790H, PUL402791H, PUL402792H, PUL402794H, PUL402795H, PUL402796H, PUL402802H, PUL402803H, PUL402805H, PUL402807H, PUL402808H, PUL402810H, PUL402811H, PUL402812H, PUL402813H, PUL402815H, PUL402818H, PUL402820H, PUL402821H, PUL402825H, PUL402826H, PUL402827H, PUL402828H, PUL402830H, PUL402832H, PUL402834H, PUL402835H, PUL402837H, PUL402838H, PUL402840H, PUL402841H, PUL402842H, PUL402843H, PUL402845H, PUL402847H, PUL402848H, PUL402849H, PUL402851H, PUL402852H, PUL402855H, PUL402857H, PUL402858H, PUL402860H, PUL402862H, PUL402864H, PUL402865H, PUL402866H, PUL402867H, PUL402868H, PUL402869H, PUL402870H, PUL402871H, PUL402872H, PUL402873H, PUL402874H, PUL402876H, PUL402878H, PUL402879H, PUL402880H, PUL402881H, PUL402882H, PUL402884H, PUL402885H, PUL402886H, PUL402887H, PUL402888H, PUL402890H, PUL402893H, PUL402895H, PUL402896H, PUL402897H, PUL402898H, PUL402900H, PUL402901H, PUL402902H, PUL402903H, PUL402904H, PUL402906H, PUL402907H, PUL402908H, PUL402909H, PUL402910H, PUL402911H, PUL402912H, PUL402915H, PUL402916H, PUL402917H, PUL402918H, PUL402919H, PUL402920H, PUL402922H, PUL402923H, PUL402924H, PUL402925H, PUL402926H, PUL402927H, PUL402928H, PUL402929H, PUL402930H, PUL402931H, PUL402932H, PUL402933H, PUL402936H, PUL402937H, PUL402938H, PUL402939H, PUL402940H, PUL402945H, PUL402947H, PUL402948H, PUL402949H, PUL402951H, PUL402952H, PUL402953H, PUL402954H, PUL402955H, PUL402956H, PUL402957H, PUL402959H, PUL402960H, PUL402961H, PUL402962H, PUL402964H, PUL402966H, PUL402967H, PUL402968H, PUL402969H, PUL402970H, PUL402971H, PUL402972H, PUL402975H, PUL402976H, PUL402977H, PUL402980H, PUL402981H, PUL402982H, PUL402983H, PUL402984H, PUL402990H, PUL402991H, PUL402992H, PUL402993H, PUL402995H, PUL402996H, PUL402998H, PUL402999H, PUL403000H, PUL403001H, PUL403002H, PUL403005H, PUL403006H, PUL403008H, PUL403011H, PUL403012H, PUL403013H, PUL403016H, PUL403018H, PUL403020H, PUL403021H, PUL403025H, PUL403026H, PUL403028H, PUL403029H, PUL403030H, PUL403031H, PUL403033H, PUL403034H, PUL403036H, PUL403037H, PUL403038H, PUL403039H, PUL403041H, PUL403042H, PUL403044H, PUL403045H, PUL403047H, PUL403048H, PUL403049H, PUL403052H, PUL403053H, PUL403054H, PUL403055H, PUL403057H, PUL403060H, PUL403061H, PUL403062H, PUL403063H, PUL403064H, PUL403065H, PUL403066H, PUL403068H, PUL403069H, PUL403070H, PUL403071H, PUL403073H, PUL403074H, PUL403075H, PUL403076H, PUL403077H, PUL403079H, PUL403080H, PUL403081H, PUL403082H, PUL403083H, PUL403084H, PUL403085H, PUL403086H, PUL403088H, PUL403089H, PUL403090H, PUL403091H, PUL403092H, PUL403093H, PUL403094H, PUL403095H, PUL403096H, PUL403097H, PUL403099H, PUL403100H, PUL403103H, PUL403104H, PUL403105H, PUL403106H, PUL403107H, PUL403108H, PUL403109H, PUL403110H, PUL403111H, PUL403112H, PUL403114H, PUL403116H, PUL403121H, PUL403123H, PUL403124H, PUL403126H, PUL403128H, PUL403130H, PUL403131H, PUL403133H, PUL403136H, PUL403138H, PUL403142H, PUL403143H, PUL403149H, PUL403150H, PUL403153H, PUL403154H, PUL403158H, PUL403160H, PUL403167H, PUL403168H, PUL403173H, PUL403175H, PUL403179H, PUL403181H, PUL403182H, PUL403186H, PUL403187H, PUL403189H, PUL403190H, PUL403191H, PUL403193H, PUL403196H, PUL403197H, PUL403199H, PUL403201H, PUL403202H, PUL403203H, PUL403207H, PUL403210H, PUL403211H, PUL403214H, PUL403215H, PUL403217H, PUL403218H, PUL403219H, PUL403220H, PUL403223H, PUL403225H, PUL403227H, PUL403229H, PUL403230H, PUL403231H, PUL403232H, PUL403233H, PUL403237H, PUL403238H, PUL403240H, PUL403241H, PUL403242H, PUL403243H, PUL403244H, PUL403245H, PUL403247H, PUL403248H, PUL403249H, PUL403251H, PUL403252H, PUL403253H, PUL403256H, PUL403257H, PUL403258H, PUL403259H, PUL403260H, PUL403263H, PUL403264H, PUL403265H, PUL403269H, PUL403271H, PUL403273H, PUL403275H, PUL403276H, PUL403278H, PUL403279H, PUL403280H, PUL403281H, PUL403283H, PUL403284H, PUL403285H, PUL403286H, PUL403288H, PUL403295H, PUL403300H, PUL403301H, PUL403306H, PUL403308H, PUL403311H, PUL403312H, PUL403313H, PUL403315H, PUL403318H, PUL403321H, PUL403323H, PUL403326H, PUL403329H, PUL403332H, PUL403333H, PUL403335H, PUL403337H, PUL403340H, PUL403341H, PUL403342H, PUL403343H, PUL403344H, PUL403350H, PUL403351H, PUL403352H, PUL403355H, PUL403356H, PUL403359H, PUL403360H, PUL403363H, PUL403364H, PUL403365H, PUL403366H, PUL403367H, PUL403370H, PUL403371H, PUL403373H, PUL403377H, PUL403379H, PUL403382H, PUL403384H, PUL403385H, PUL403386H, PUL403387H, PUL403388H, PUL403392H, PUL403393H, PUL403397H, PUL403401H, PUL403408H, PUL403409H, PUL403411H, PUL403413H, PUL403414H, PUL403415H, PUL403416H, PUL403417H, PUL403420H, PUL403422H, PUL403423H, PUL403424H, PUL403425H, PUL403427H, PUL403430H, PUL403432H, PUL403434H, PUL403435H, PUL403436H, PUL403438H, PUL403439H, PUL403440H, PUL403441H, PUL403444H, PUL403446H, PUL403447H, PUL403448H, PUL403450H, PUL403451H, PUL403453H, PUL403454H, PUL403455H, PUL403457H, PUL403458H, PUL403460H, PUL403462H, PUL403463H, PUL403464H, PUL403466H, PUL403467H, PUL403469H, PUL403470H, PUL403471H, PUL403473H, PUL403475H, PUL403476H, PUL403477H, PUL403478H, PUL403480H, PUL403481H, PUL403482H, PUL403483H, PUL403484H, PUL403485H, PUL403486H, PUL403488H, PUL403489H, PUL403492H, PUL403493H, PUL403495H, PUL403496H, PUL403498H, PUL403499H, PUL403500H, PUL403502H, PUL403503H, PUL403505H, PUL403506H, PUL403507H, PUL403508H, PUL403510H, PUL403511H, PUL403513H, PUL403514H, PUL403515H, PUL403517H, PUL403518H, PUL403519H, PUL403520H, PUL403521H, PUL403523H, PUL403524H, PUL403525H, PUL403526H, PUL403527H, PUL403528H, PUL403529H, PUL403530H, PUL403531H, PUL403532H, PUL403533H, PUL403535H, PUL403537H, PUL403540H, PUL403541H, PUL403542H, PUL403545H, PUL403546H, PUL403547H, PUL403548H, PUL403549H, PUL403551H, PUL403552H, PUL403554H, PUL403555H, PUL403556H, PUL403558H, PUL403559H, PUL403561H, PUL403563H, PUL403564H, PUL403565H, PUL403566H, PUL403568H, PUL403569H, PUL403570H, PUL403571H, PUL403572H, PUL403573H, PUL403574H, PUL403575H, PUL403576H, PUL403577H, PUL403578H, PUL403579H, PUL403580H, PUL403581H, PUL403582H, PUL403584H, PUL403585H, PUL403586H, PUL403587H, PUL403589H, PUL403590H, PUL403591H, PUL403593H, PUL403595H, PUL403597H, PUL403598H, PUL403599H, PUL403601H, PUL403602H, PUL403603H, PUL403605H, PUL403608H, PUL403609H, PUL403611H, PUL403613H, PUL403618H, PUL403620H, PUL403622H, PUL403623H, PUL403626H, PUL403629H, PUL403630H, PUL403631H, PUL403632H, PUL403634H, PUL403636H, PUL403637H, PUL403638H, PUL403642H, PUL403645H, PUL403646H, PUL403651H, PUL403654H, PUL403658H, PUL403659H, PUL403660H, PUL403662H, PUL403663H, PUL403664H, PUL403665H, PUL403666H, PUL403668H, PUL403669H, PUL403673H, PUL403675H, PUL403676H, PUL403678H, PUL403682H, PUL403683H, PUL403684H, PUL403685H, PUL403686H, PUL403687H, PUL403688H, PUL403690H, PUL403691H, PUL403692H, PUL403693H, PUL403694H, PUL403696H, PUL403701H, PUL403702H, PUL403704H, PUL403706H, PUL403708H, PUL403709H, PUL403711H, PUL403713H, PUL403716H, PUL403717H, PUL403718H, PUL403719H, PUL403720H, PUL403721H, PUL403722H, PUL403723H, PUL403724H, PUL403725H, PUL403726H, PUL403727H, PUL403729H, PUL403731H, PUL403733H, PUL403734H, PUL403735H, PUL403736H, PUL403737H, PUL403738H, PUL403739H, PUL403740H, PUL403743H, PUL403744H, PUL403745H, PUL403747H, PUL403749H, PUL403750H, PUL403752H, PUL403753H, PUL403754H, PUL403755H, PUL403756H, PUL403757H, PUL403759H, PUL403760H, PUL403763H, PUL403764H, PUL403766H, PUL403767H, PUL403768H, PUL403769H, PUL403771H, PUL403772H, PUL403773H, PUL403774H, PUL403775H, PUL403776H, PUL403777H, PUL403779H, PUL403781H, PUL403782H, PUL403783H, PUL403784H, PUL403785H, PUL403787H, PUL403788H, PUL403789H, PUL403790H, PUL403792H, PUL403793H, PUL403794H, PUL403795H, PUL403796H, PUL403797H, PUL403798H, PUL403799H, PUL403800H, PUL403801H, PUL403804H, PUL403806H, PUL403807H, PUL403809H, PUL403810H, PUL403811H, PUL403813H, PUL403814H, PUL403815H, PUL403816H, PUL403817H, PUL403819H, PUL403820H, PUL403821H, PUL403822H, PUL403824H, PUL403825H, PUL403827H, PUL403828H, PUL403829H, PUL403831H, PUL403832H, PUL403834H, PUL403836H, PUL403837H, PUL403838H, PUL403839H, PUL403840H, PUL403842H, PUL403843H, PUL403844H, PUL403845H, PUL403847H, PUL403848H, PUL403849H, PUL403850H, PUL403852H, PUL403854H, PUL403857H, PUL403858H, PUL403859H, PUL403860H, PUL403863H, PUL403866H, PUL403867H, PUL403868H, PUL403869H, PUL403870H, PUL403871H, PUL403872H, PUL403873H, PUL403874H, PUL403875H, PUL403876H, PUL403878H, PUL403879H, PUL403880H, PUL403882H, PUL403883H, PUL403884H, PUL403886H, PUL403887H, PUL403888H, PUL403889H, PUL403890H, PUL403891H, PUL403893H, PUL403894H, PUL403896H, PUL403897H, PUL403900H, PUL403902H, PUL403904H, PUL403905H, PUL403906H, PUL403907H, PUL403908H, PUL403909H, PUL403910H, PUL403911H, PUL403913H, PUL403914H, PUL403915H, PUL403917H, PUL403918H, PUL403919H, PUL403920H, PUL403922H, PUL403923H, PUL403924H, PUL403927H, PUL403928H, PUL403929H, PUL403931H, PUL403932H, PUL403934H, PUL403935H, PUL403936H, PUL403937H, PUL403938H, PUL403939H, PUL403940H, PUL403941H, PUL403942H, PUL403943H, PUL403945H, PUL403948H, PUL403949H, PUL403950H, PUL403951H, PUL403953H, PUL403955H, PUL403956H, PUL403957H, PUL403959H, PUL403960H, PUL403961H, PUL403962H, PUL403964H, PUL403965H, PUL403966H, PUL403967H, PUL403968H, PUL403969H, PUL403970H, PUL403971H, PUL403972H, PUL403973H, PUL403974H, PUL403975H, PUL403976H, PUL403977H, PUL403982H, PUL403983H, PUL403984H, PUL403987H, PUL403988H, PUL403989H, PUL403991H, PUL403992H, PUL403994H, PUL403995H, PUL403996H, PUL404002H, PUL404003H, PUL404011H, PUL404012H, PUL404013H, PUL404016H, PUL404018H, PUL404020H, PUL404022H, PUL404023H, PUL404025H, PUL404027H, PUL404029H, PUL404031H, PUL404035H, PUL404037H, PUL404039H, PUL404044H, PUL404045H, PUL404049H, PUL404051H, PUL404056H, PUL404057H, PUL404058H, PUL404062H, PUL404063H, PUL404064H, PUL404066H, PUL404067H, PUL404068H, PUL404071H, PUL404072H, PUL404073H, PUL404074H, PUL404079H, PUL404080H, PUL404081H, PUL404083H, PUL404085H, PUL404086H, PUL404087H, PUL404089H, PUL404090H, PUL404092H, PUL404093H, PUL404094H, PUL404097H, PUL404098H, PUL404100H, PUL404101H, PUL404105H, PUL404107H, PUL404109H, PUL404110H, PUL404112H, PUL404113H, PUL404114H, PUL404117H, PUL404118H, PUL404123H, PUL404129H, PUL404131H, PUL404133H, PUL404135H, PUL404141H, PUL404142H, PUL404143H, PUL404144H, PUL404145H, PUL404149H, PUL404150H, PUL404156H, PUL404157H, PUL404165H, PUL404172H, PUL404177H, PUL404178H, PUL404180H, PUL404181H, PUL404182H, PUL404183H, PUL404184H, PUL404185H, PUL404186H, PUL404187H, PUL404188H, PUL404189H, PUL404191H, PUL404193H, PUL404194H, PUL404195H, PUL404196H, PUL404197H, PUL404198H, PUL404199H, PUL404200H, PUL404201H, PUL404202H, PUL404205H, PUL404208H, PUL404209H, PUL404210H, PUL404211H, PUL404213H, PUL404217H, PUL404218H, PUL404219H, PUL404220H, PUL404221H, PUL404222H, PUL404223H, PUL404224H, PUL404225H, PUL404227H, PUL404228H, PUL404229H, PUL404231H, PUL404232H, PUL404234H, PUL404236H, PUL404238H, PUL404239H, PUL404240H, PUL404241H, PUL404242H, PUL404243H, PUL404244H, PUL404245H, PUL404246H, PUL404247H, PUL404248H, PUL404249H, PUL404250H, PUL404253H, PUL404254H, PUL404256H, PUL404257H, PUL404259H, PUL404260H, PUL404261H, PUL404262H, PUL404265H, PUL404266H, PUL404267H, PUL404269H, PUL404270H, PUL404271H, PUL404272H, PUL404273H, PUL404274H, PUL404275H, PUL404276H, PUL404278H, PUL404279H, PUL404280H, PUL404281H, PUL404282H, PUL404286H, PUL404287H, PUL404288H, PUL404289H, PUL404290H, PUL404291H, PUL404292H, PUL404295H, PUL404296H, PUL404297H, PUL404299H, PUL404300H, PUL404301H, PUL404302H, PUL404303H, PUL404304H, PUL404306H, PUL404308H, PUL404309H, PUL404310H, PUL404312H, PUL404313H, PUL404315H, PUL404318H, PUL404319H, PUL404321H, PUL404322H, PUL404323H, PUL404324H, PUL404325H, PUL404326H, PUL404327H, PUL404330H, PUL404332H, PUL404333H, PUL404334H, PUL404335H, PUL404336H, PUL404337H, PUL404338H, PUL404339H, PUL404340H, PUL404341H, PUL404343H, PUL404344H, PUL404345H, PUL404346H, PUL404347H, PUL404348H, PUL404349H, PUL404350H, PUL404351H, PUL404352H, PUL404354H, PUL404355H, PUL404357H, PUL404358H, PUL404362H, PUL404363H, PUL404365H, PUL404366H, PUL404369H, PUL404371H, PUL404373H, PUL404375H, PUL404376H, PUL404378H, PUL404379H, PUL404380H, PUL404381H, PUL404382H, PUL404383H, PUL404384H, PUL404385H, PUL404387H, PUL404388H, PUL404389H, PUL404390H, PUL404391H, PUL404393H, PUL404394H, PUL404395H, PUL404397H, PUL404399H, PUL404400H, PUL404402H, PUL404404H, PUL404405H, PUL404406H, PUL404407H, PUL404409H, PUL404410H, PUL404411H, PUL404412H, PUL404413H, PUL404414H, PUL404417H, PUL404418H, PUL404422H, PUL404425H, PUL404426H, PUL404428H, PUL404432H, PUL404433H, PUL404434H, PUL404435H, PUL404438H, PUL404439H, PUL404441H, PUL404442H, PUL404443H, PUL404445H, PUL404446H, PUL404447H, PUL404448H, PUL404449H, PUL404450H, PUL404452H, PUL404453H, PUL404454H, PUL404455H, PUL404456H, PUL404457H, PUL404458H, PUL404460H, PUL404462H, PUL404463H, PUL404465H, PUL404466H, PUL404467H, PUL404468H, PUL404469H, PUL404471H, PUL404472H, PUL404473H, PUL404474H, PUL404475H, PUL404476H, PUL404477H, PUL404481H, PUL404486H, PUL404487H, PUL404490H, PUL404492H, PUL404493H, PUL404494H, PUL404495H, PUL404496H, PUL404498H, PUL404499H, PUL404502H, PUL404503H, PUL404504H, PUL404505H, PUL404506H, PUL404509H, PUL404512H, PUL404515H, PUL404518H, PUL404519H, PUL404520H, PUL404521H, PUL404522H, PUL404523H, PUL404524H, PUL404525H, PUL404526H, PUL404527H, PUL404528H, PUL404530H, PUL404531H, PUL404533H, PUL404534H, PUL404537H, PUL404539H, PUL404540H, PUL404542H, PUL404544H, PUL404545H, PUL404547H, PUL404548H, PUL404549H, PUL404551H, PUL404556H, PUL404557H, PUL404558H, PUL404559H, PUL404560H, PUL404563H, PUL404564H, PUL404565H, PUL404566H, PUL404569H, PUL404570H, PUL404571H, PUL404572H, PUL404573H, PUL404574H, PUL404575H, PUL404576H, PUL404577H, PUL404578H, PUL404580H, PUL404581H, PUL404583H, PUL404584H, PUL404585H, PUL404587H, PUL404588H, PUL404590H, PUL404591H, PUL404592H, PUL404594H, PUL404595H, PUL404596H, PUL404597H, PUL404598H, PUL404605H, PUL404606H, PUL404609H, PUL404610H, PUL404611H, PUL404612H, PUL404613H, PUL404615H, PUL404616H, PUL404617H, PUL404620H, PUL404625H, PUL404626H, PUL404627H, PUL404629H, PUL404634H, PUL404637H, PUL404640H, PUL404642H, PUL404644H, PUL404652H, PUL404656H, PUL404658H, PUL404661H, PUL404662H, PUL404663H, PUL404671H, PUL404673H, PUL404674H, PUL404675H, PUL404676H, PUL404678H, PUL404681H, PUL404687H, PUL404689H, PUL404696H, PUL404697H, PUL404698H, PUL404702H, PUL404703H, PUL404704H, PUL404705H, PUL404706H, PUL404707H, PUL404708H, PUL404710H, PUL404711H, PUL404712H, PUL404713H, PUL404715H, PUL404716H, PUL404719H, PUL404720H, PUL404722H, PUL404723H, PUL404730H, PUL404731H, PUL404733H, PUL404734H, PUL404735H, PUL404738H, PUL404740H, PUL404741H, PUL404743H, PUL404744H, PUL404747H, PUL404748H, PUL404749H, PUL404750H, PUL404752H, PUL404755H, PUL404756H, PUL404757H, PUL404759H, PUL404760H, PUL404761H, PUL404762H, PUL404763H, PUL404764H, PUL404765H, PUL404766H, PUL404769H, PUL404770H, PUL404771H, PUL404772H, PUL404773H, PUL404774H, PUL404776H, PUL404778H, PUL404779H, PUL404781H, PUL404782H, PUL404783H, PUL404785H, PUL404786H, PUL404787H, PUL404788H, PUL404790H, PUL404791H, PUL404793H, PUL404794H, PUL404795H, PUL404797H, PUL404798H, PUL404799H, PUL404800H, PUL404802H, PUL404805H, PUL404807H, PUL404811H, PUL404813H, PUL404814H, PUL404815H, PUL404817H, PUL404819H, PUL404820H, PUL404826H, PUL404828H, PUL404830H, PUL404831H, PUL404832H, PUL404833H, PUL404834H, PUL404836H, PUL404838H, PUL404839H, PUL404840H, PUL404842H, PUL404843H, PUL404844H, PUL404846H, PUL404847H, PUL404849H, PUL404852H, PUL404853H, PUL404854H, PUL404856H, PUL404857H, PUL404859H, PUL404860H, PUL404861H, PUL404863H, PUL404864H, PUL404866H, PUL404868H, PUL404870H, PUL404871H, PUL404873H, PUL404874H, PUL404875H, PUL404876H, PUL404877H, PUL404878H, PUL404879H, PUL404880H, PUL404881H, PUL404886H, PUL404888H, PUL404891H, PUL404892H, PUL404893H, PUL404894H, PUL404895H, PUL404896H, PUL404897H, PUL404900H, PUL404901H, PUL404907H, PUL404909H, PUL404910H, PUL404914H, PUL404916H, PUL404918H, PUL404920H, PUL404921H, PUL404922H, PUL404924H, PUL404925H, PUL404926H, PUL404927H, PUL404928H, PUL404929H, PUL404931H, PUL404933H, PUL404935H, PUL404939H, PUL404944H, PUL404946H, PUL404949H, PUL404950H, PUL404951H, PUL404960H, PUL404961H, PUL404962H, PUL404963H, PUL404964H, PUL404965H, PUL404966H, PUL404967H, PUL404984H, PUL404994H, PUL404996H, PUL404997H, PUL404999H, PUL405001H, PUL405005H, PUL405008H, PUL405020H, PUL405024H, PUL405026H, PUL405028H, PUL405029H, PUL405030H, PUL405031H, PUL405032H, PUL405033H, PUL405034H, PUL405035H, PUL405037H, PUL405039H, PUL405041H, PUL405047H, PUL405048H, PUL405049H, PUL405050H, PUL405053H, PUL405054H, PUL405057H, PUL405059H, PUL405062H, PUL405065H, PUL405070H, PUL405075H, PUL405076H, PUL405087H, PUL405088H, PUL405090H, PUL405092H, PUL405095H, PUL405106H, PUL405109H, PUL405110H, PUL405111H, PUL405112H, PUL405115H, PUL405117H, PUL405118H, PUL405121H, PUL405124H, PUL405128H, PUL405130H, PUL405131H, PUL405137H, PUL405138H, PUL405143H, PUL405156H, PUL405158H, PUL405169H, PUL405173H, PUL405174H, PUL405180H, PUL405189H, PUL405190H, PUL405191H, PUL405193H, PUL405195H, PUL405199H, PUL405201H, PUL405206H, PUL405209H, PUL405211H, PUL405215H, PUL405216H, PUL405219H, PUL405220H, PUL405222H, PUL405235H, PUL405242H, PUL405247H, PUL405251H, PUL405253H, PUL405254H, PUL405257H, PUL405262H, PUL405267H, PUL405276H, PUL405281H, PUL405284H, PUL405286H, PUL405297H, PUL405300H, PUL405303H, PUL405305H, PUL405307H, PUL405310H, PUL405319H, PUL405320H, PUL405325H, PUL405331H, PUL405333H, PUL405340H, PUL405342H, PUL405355H, PUL405356H, PUL405357H, PUL405360H, PUL405361H, PUL405371H, PUL405378H, PUL405383H, PUL405396H, PUL405397H, PUL405402H, PUL405416H, PUL405422H, PUL405426H, PUL405427H, PUL405433H, PUL405443H, PUL405445H, PUL405447H, PUL405448H, PUL405451H, PUL405452H, PUL405455H, PUL405460H, PUL405461H, PUL405475H, PUL405481H, PUL405490H, PUL405493H, PUL405499H, PUL405509H, PUL405516H, PUL405520H, PUL405524H, PUL405527H, PUL405528H, PUL405529H, PUL405534H, PUL405535H, PUL405536H, PUL405538H, PUL405539H, PUL405540H, PUL405542H, PUL405544H, PUL405546H, PUL405549H, PUL405550H, PUL405551H, PUL405554H, PUL405556H, PUL405557H, PUL405558H, PUL405560H, PUL405561H, PUL405562H, PUL405564H, PUL405566H, PUL405567H, PUL405568H, PUL405569H, PUL405570H, PUL405573H, PUL405574H, PUL405576H, PUL405577H, PUL405578H, PUL405580H, PUL405582H, PUL405583H, PUL405585H, PUL405591H, PUL405592H, PUL405594H, PUL405595H, PUL405596H, PUL405598H, PUL405599H, PUL405602H, PUL405603H, PUL405605H, PUL405606H, PUL405607H, PUL405609H, PUL405613H, PUL405615H, PUL405618H, PUL405623H, PUL405624H, PUL405625H, PUL405629H, PUL405633H, PUL405637H, PUL405643H, PUL405647H, PUL405648H, PUL405652H, PUL405654H, PUL405658H, PUL405660H, PUL405661H, PUL405662H, PUL405664H, PUL405668H, PUL405669H, PUL405670H, PUL405671H, PUL405675H, PUL405676H, PUL405677H, PUL405679H, PUL405680H, PUL405681H, PUL405686H, PUL405689H, PUL405690H, PUL405691H, PUL405693H, PUL405695H, PUL405698H, PUL405700H, PUL405703H, PUL405704H, PUL405708H, PUL405709H, PUL405710H, PUL405714H, PUL405715H, PUL405716H, PUL405717H, PUL405719H, PUL405720H, PUL405727H, PUL405728H, PUL405734H, PUL405735H, PUL405737H, PUL405738H, PUL405739H, PUL405742H, PUL405748H, PUL405749H, PUL405753H, PUL405757H, PUL405763H, PUL405764H, PUL405768H, PUL405771H, PUL405772H, PUL405783H, PUL405788H, PUL405790H, PUL405809H, PUL405810H, PUL405813H, PUL405814H, PUL405816H, PUL405817H, PUL405819H, PUL405820H, PUL405821H, PUL405824H, PUL405825H, PUL405826H, PUL405827H, PUL405828H, PUL405833H, PUL405839H, PUL405843H, PUL405845H, PUL405847H, PUL405850H, PUL405852H, PUL405855H, PUL405856H, PUL405857H, PUL405859H, PUL405860H, PUL405861H, PUL405862H, PUL405863H, PUL405864H, PUL405867H, PUL405870H, PUL405872H, PUL405873H, PUL405874H, PUL405875H, PUL405876H, PUL405877H, PUL405879H, PUL405881H, PUL405882H, PUL405883H, PUL405885H, PUL405887H, PUL405890H, PUL405893H, PUL405897H, PUL405899H, PUL405902H, PUL405903H, PUL405906H, PUL405908H, PUL405909H, PUL405911H, PUL405912H, PUL405913H, PUL405914H, PUL405919H, PUL405926H, PUL405927H, PUL405931H, PUL405932H, PUL405939H, PUL405940H, PUL405941H, PUL405942H, PUL405947H, PUL405949H, PUL405952H, PUL405958H, PUL405963H, PUL405967H, PUL405969H, PUL405971H, PUL405972H, PUL405973H, PUL405982H, PUL406002H, PUL406007H, PUL406009H, PUL406010H, PUL406027H, PUL406028H, PUL406029H, PUL406030H, PUL406031H, PUL406034H, PUL406036H, PUL406038H, PUL406047H, PUL406049H, PUL406051H, PUL406053H, PUL406056H, PUL406057H, PUL406058H, PUL406061H, PUL406063H, PUL406071H, PUL406073H, PUL406074H, PUL406077H, PUL406079H, PUL406081H, PUL406086H, PUL406104H, PUL406107H, PUL406111H, PUL406114H, PUL406118H, PUL406124H, PUL406127H, PUL406137H, PUL406142H, PUL406143H, PUL406146H, PUL406148H, PUL406150H, PUL406155H, PUL406159H, PUL406163H, PUL406166H, PUL406168H, PUL406171H, PUL406177H, PUL406179H, PUL406187H, PUL406188H, PUL406190H, PUL406191H, PUL406195H, PUL406200H, PUL406203H, PUL406204H, PUL406205H, PUL406207H, PUL406213H, PUL406218H, PUL406226H, PUL406229H, PUL406231H, PUL406232H, PUL406237H, PUL406238H, PUL406248H, PUL406255H, PUL406271H, PUL406289H, PUL406312H, PUL406367H, PUL406376H, PUL406382H, PUL406384H, PUL406388H, PUL406391H, PUL406392H, PUL406398H, PUL406399H, PUL406400H, PUL406403H, PUL406405H, PUL406406H, PUL406407H, PUL406410H, PUL406411H, PUL406414H, PUL406415H, PUL406417H, PUL406418H, PUL406419H, PUL406420H, PUL406424H, PUL406432H, PUL406436H, PUL406437H, PUL406438H, PUL406439H, PUL406440H, PUL406446H, PUL406447H, PUL406453H, PUL406454H, PUL406456H, PUL406457H, PUL406462H, PUL406466H, PUL406467H, PUL406469H, PUL406470H, PUL406472H, PUL406475H, PUL406479H, PUL406483H, PUL406484H, PUL406486H, PUL406552H, PUL406587H, PUL406593H, PUL406594H, PUL406602H, PUL406603H, PUL406607H, PUL406609H, PUL406610H, PUL406614H, PUL406615H, PUL406621H, PUL406622H, PUL406623H, PUL406625H, PUL406630H, PUL406632H, PUL406633H, PUL406635H, PUL406638H, PUL406643H, PUL406652H, PUL406654H, PUL406658H, PUL406661H, PUL406662H, PUL406667H, PUL406672H, PUL406703H, PUL406706H, PUL406715H, PUL406716H, PUL406720H, PUL406730H, PUL406741H, PUL406749H, PUL406752H, PUL406754H, PUL406757H, PUL406760H, PUL406761H, PUL406764H, PUL406768H, PUL406769H, PUL406770H, PUL406771H, PUL406772H, PUL406777H, PUL406781H, PUL406782H, PUL406790H, PUL406798H, PUL406804H, PUL406806H, PUL406808H, PUL406809H, PUL406841H, PUL406842H, PUL406845H, PUL406848H, PUL406858H, PUL406863H, PUL406864H, PUL406867H, PUL406874H, PUL406879H, PUL406886H, PUL406896H, PUL406897H, PUL406898H, PUL406901H, PUL406902H, PUL406909H, PUL406913H, PUL406920H, PUL406925H, PUL406928H, PUL406931H, PUL406933H, PUL406939H, PUL406942H, PUL406947H, PUL406948H, PUL406949H, PUL406955H, PUL406958H, PUL406973H, PUL406974H, PUL406989H, PUL406992H, PUL407000H, PUL407001H, PUL407006H, PUL407008H, PUL407011H, PUL407013H, PUL407014H, PUL407029H, PUL407035H, PUL407040H, PUL407044H, PUL407045H, PUL407050H, PUL407052H, PUL407058H, PUL407068H, PUL407076H, PUL407081H, PUL407086H, PUL407090H, PUL407099H, PUL410353H, PUL412089H, PUL412881H, PUL415125H, PUL438812H, PUL440753H, PUL441763H. |
Recalling Firm/ Manufacturer |
Medtronic Inc. Cardiac Rhythm Disease Managment 8200 Coral Sea St. N.E. Saint Paul MN 55112
|
For Additional Information Contact | 763-514-4000 |
Manufacturer Reason for Recall | A subset of active implanted Concerto CRT-D and Virtuoso ICD devices worldwide (estimated at 6,300) may not meet expected device longevity due to gradually increasing current drain caused by low voltage capacitor degradation. This issue may present in the affected devices as reaching the Recommended Replacement Time (RRT) earlier than projected. This issue does not compromise device functionalit |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | An "Important Medical Device Information" letter, dated September 2009 was hand delivered by Medtronic Representatives beginning 09/09/09. The letter describes the issue, identifies affected devices and provides recommendations for patients with devices in the affected subset.. A Patient letter was mailed to patients beginning 09/28/09.
For additional information:
Doctors are advised to contact your local Medtronic Representative or Medtronic Technical Services at 800-723-4636;
Device tracking managers are advised to contact your Medtronic representative, or the Patient Registration Department at 1-800-328-2518 x41832; and
Patients are advised to contact Medtronic Patient Services number at 800-551-5544 ext. 41835. |
Quantity in Commerce | 3162 |
Distribution | Worldwide distribution (except Idaho and Wyoming) including countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Norway, Portugal, Saudi Arabi, Spain, Sweden, Switzerland, Turkey and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database | PMAs with Product Code = LWS PMAs with Product Code = LWS
|
|
|
|