Class 2 Device Recall Anesthesia system |
|
Date Initiated by Firm |
March 26, 2010 |
Date Posted |
April 16, 2010 |
Recall Status1 |
Terminated 3 on August 23, 2011 |
Recall Number |
Z-1408-2010 |
Recall Event ID |
55177 |
510(K)Number |
K051629
|
Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
Product |
The BleaseSirius and BleaseFocus Anaesthesia Systems. |
Code Information |
Serial numbers: Siri 000116, Siri 000885, Siri 001100, Siri 001609, Siri 001610, Siri-001618, Siri-001626, Siri-001637, Siri-001640, Siri-001644, Siri-001658, Siri-001666, Siri-001667, Siri-001668, Siri-001674, AND Siri-001681. Focu-000735, Focu-000737, Focu-000738, Focu-000739, Focu-000740, Focu-000741, Focu-000742, Focu-000743, Focu-000744, Focu-000745, Focu-000746, Focu-000747, Focu-000748, Focu-000749, Focu-000750, Focu-000751, Focu-100001, Focu-100003, Focu-100004, Focu-100005, Focu-100006, Focu-100007, Focu-100008, AND Focu-100009. |
Recalling Firm/ Manufacturer |
Blease Medical Equipment, Ltd. Deansway, Chesham Bucks, England United Kingdom
|
For Additional Information Contact |
Al Van Houdt 425-657-7200
|
Manufacturer Reason for Recall |
Potential for delivery of a lower than set concentration of anaesthetic to a patient due to a sticking valve. The lower shuttle may become jammed, not permitting flow of anaesthesia gas through the vapouriser. The vapouriser back bar mounting valves will require exchange only on the systems identified in this matter, no other systems are affected.
|
FDA Determined Cause 2 |
Other |
Action |
On 2/26/2010 notification letters were sent to all international consignees. The 2 domestic consignees were sent notification letters via FEDEX on 3/26/2010. The domestic consignees were also contacted by phone on 3/26/2010.
The notification letter is titled URGENT MEDICAL DEVICE CORRECTION. It advises users of the issue and provides additional information which would permit the continued use of the Blease system until replacement back bar valves can be fitted by the recalling firm's service representatives.
For users who choose to continue to use the device, the manufacturer recommends the use of an Agent Monitor and user education; also checks that vapourisers have been mounted correctly and the system is leak free, and testing in advance of a procedure.
For additional questions, US Customers are directed to call Global Technical Support - +1 425-657-7200 ext: 5089.
Non US Customers, call Global Technical Support - +44 1494 784422. |
Quantity in Commerce |
40 units |
Distribution |
Distributed to 2 U.S. hospitals and to customers in the following countries:
Panama, Poland, Australia, Singapore, United Kingdom, Bangladesh, Nicaragua, Vietnam, Argentina, Finland, and Columbia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BSZ and Original Applicant = BLEASE USA, INC.
|
|
|
|