| | Class 2 Recall R-20J, Diagnostic X-ray Beam Limiting Device |  |
| Date Posted |
November 08, 2012 |
| Recall Number |
Z-0261-2013 |
| Product |
R-20J , Diagnostic X-Ray Beam-Limiting Device.
In combination with any of below X-Ray tubes:
06/1.2P364DK-85, 0.6/1.2P324DK-125,
06/1.2P324DK-85, 0.6/1.2P38DE-85.
Product Usage:
The device is the beam-limiting device for X-ray Radiography. |
| Code Information |
Device Model #R-20J
|
Recalling Firm/ Manufacturer |
Shimadzu Medical Systems 20101 S Vermont Ave Torrance, California 90502-1328 |
| Consumer Instructions |
Contact the recalling firm for information |
Reason for Recall |
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
|
| Action |
Shimadzu sent an Urgent Voluntary Medical Device Recall - Field Safety Notice letters dated September 19, 2012 to all the affected customers. The letter identifed the description of non-compliance, how to identify the affected products, corrective action plan by Shimadzu and actions to be taken customers. Customers were instructed to contact their local Shimadzu representative or contact Shimadzu National Technical Support at Shimadzu Medical Systems National Technical Support (800) 228-1429. |
| Quantity in Commerce |
308 units |
| Distribution |
US Nationwide Distribution |
| |
|
|
|
|