| | Class 2 Recall AutoPulse Resuscitation System |  |
| Date Posted |
January 09, 2013 |
| Recall Number |
Z-0658-2013 |
| Product |
AutoPulse Resuscitation System Model 100,
Mfg by:Zoll Circulation
Sunnyvale, CA.
The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse. |
| Code Information |
all
|
Recalling Firm/ Manufacturer |
Zoll Circulation, Inc. 650 Almanor Ave Sunnyvale, California 94085-3513 |
| Consumer Instructions |
Contact the recalling firm for information |
| For Additional Information Contact |
Sam Nanavati 408-419-1030
|
Reason for Recall |
Complaints associated with an unexpected stop in AutoPulse compressions.
|
| Action |
Zoll sent Customer Notification letters dated August 8, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Zoll Technical Support Line at 1 (800) 348-9011 for further assistance. |
| Quantity in Commerce |
3881 in US, 1846 ROW. |
| Distribution |
Worldwide distribution. |
| |
|
|
|
|