| | Class 2 Recall Vector TAS Modular Driver |  |
| Date Posted |
November 05, 2012 |
| Recall Number |
Z-0205-2013 |
| Product |
Vector TAS Modular Driver, a manual driver for Vector TAS dental screws. |
| Code Information |
Part Numbers 601-0007 and 601-0010, all lots
|
Recalling Firm/ Manufacturer |
Ormco/Sybronendo 1332 S Lone Hill Ave Glendora, California 91740-5339 |
| Consumer Instructions |
Contact the recalling firm for information |
Reason for Recall |
The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilization temperature should be listed as 132¿C and/or 270¿F.
|
| Action |
"Urgent - Medical Device Correction" letters dated 12/12/2011 were sent via USPS 1st class mail to all customers and again on 2/21/2012 for recall expansion. The device issue was described and recommended actions were provided. Customers were instructed to acknowledge and complete the Return Form. Questions were referred to (800) 854-1741. |
| Quantity in Commerce |
4,543 |
| Distribution |
Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique. |
| |
|
|
|
|