| | Class 1 Recall Ventlab Rescue Med Infant Resuscitator |  |
| Date Posted |
November 01, 2012 |
| Recall Number |
Z-0179-2013 |
| Product |
Ventlab Rescue Med Infant Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028.
The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest. |
| Code Information |
Product Code: BVM700, Lot #101638.
|
Recalling Firm/ Manufacturer |
Ventlab Corporation 155 Boyce Dr Mocksville, North Carolina 27028-4187 |
| For Additional Information Contact |
James A. Cochie 336-753-5000
|
Reason for Recall |
Possible volume leakage through the inlet valve during compressions of the resuscitator bag.
|
| Action |
VentLab Corporation sent an Attention: Recall Notification letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product, complete and return the enclosed response form by fax to 336-753-5002, indicating whether or not their facility has any of the affected product. Customers can also scan and email the response form to CSR@Ventlab.com. For questions regarding this recall call 336-753-5000. |
| Distribution |
Nationwide Distribution including IL, OH, NY and TN. |
| |
|
|
|
|