• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical & Radiation Emitting Device Recalls

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
KING LAD SILI CONE
see related information
Date Posted November 21, 2012
Recall Number Z-0424-2013
Product KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD® Patient Size: >70kg Product Usage: Laryngeal airway
Code Information REF: LAD-705, Lot #K120413
Recalling Firm/
Manufacturer
King Systems Corp.
15011 Herriman Blvd
Noblesville, Indiana 46060
Reason for
Recall
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
Action King Systems sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated October 31, 2012 to all affected consignees via email. The letter identified affected product, problem and actions to be taken. The letter states that further use and/or distribution of any remaining product affected by this recall should cease immediately. The letter instructs customers to destroy all of the affected products and indicate the destruction by signing and returning the attached forms using the enclosed self-addressed envelope. For questions contact your King Systems customer service representative at 800-642-5464.
Quantity in Commerce 263 cases of product. (2,630 devices)
Distribution US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
 
-
-