| | Class 2 Recall TriVascular Fill Polymer Kit |  |
| Date Posted |
November 13, 2012 |
| Recall Number |
Z-0288-2013 |
| Product |
TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only);
PRODUCT Usage:
The TriVascular Ovation Abdominal Stent Graft System is indicated in subjects diagnosed with an aneurysm in the abdominal aorta having vascular morphology suitable for endovascular repair, including: (1) Adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories, (2) Non-aneurysmal proximal aortic neck: - with a length of at least 7 mm proximal to the aneurysm, - with an inner wall diameter of no less than 16 mm and no greater than 30 mm, and - with an aortic angle of d 60 degrees if proximal neck is e 10 mm and d 45 degrees if proximal neck is < 10 mm, (3) Non-aneurysmal distal iliac landing zone: - with a length of at least 10 mm, - with an inner wall diameter of no less than 8 mm and no greater than 20 mm |
| Code Information |
The following lot numbers are in distributor inventory in Europe:
FF121211-04,
FF010912-01,
FF013112-01,
FF020912-01,
FF022212-03,
FF022912-01,
FF040412-01,
FF051012-01
|
Recalling Firm/ Manufacturer |
Trivascular, Inc 3910 Brickway Blvd Santa Rosa, California 95403 |
| For Additional Information Contact |
Meg Carr 707-543-8800
|
Reason for Recall |
Selected lots of the Fill Polymer Kit (European Model Number: TVFP14-
B) in European distributor inventory only that potentially contain leaking stopcocks due to
possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in
|
| Action |
Trivascular sent an Urgent Field Safety Notice letter dated October 19, 2012 to all affect consignees. A verbal (telephone) communication was also initiated. The forms communication identified the affected product, problem and actions to be taken. Consignees were instructed to immediately quarantine and return affected products. A return response form was included for consignees to complete and returned via the e-mail address or fax number provided. Consignees were instructed to further notify its (sub-level) consignees, and have them return affected product directly to the recalling firm. For questions call 1-707-543-8732. |
| Quantity in Commerce |
41 units |
| Distribution |
Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S. |
| |
|
|
|
|